Professional Documents
Culture Documents
IB035161
Feb 2020
User’s Manual
Sequence Creation Software
SD033-PLZ-5WH2
Wavy for PLZ-5WH2
Ver. 5.x
Contents
This manual applies to versions 5.x of SD033-PLZ-5WH2 Wavy for PLZ-5WH2. To check the version,
on the Help menu, click About Wavy.
Wavy for PLZ-5WH2 applies to PLZ-5WH2 series with firmware version 2.1x and later. For
instructions on how to view the firmware version, see the electronic load operation manual.
Intended readers
This document is intended for users who will use SD033-PLZ-5WH2 Wavy for PLZ-5WH2 to control
the PLZ-5WH2 Electronic Load series and instructors of such users.
The manual assumes that the reader has knowledge about power supplies.
Trademarks
Windows is a registered trademark or trademark of Microsoft Corporation in the United States and/or
other countries.
Other company names and product names used in this manual are generally trademarks or registered
trademarks of the respective companies.
• Sequence creation software SD033-PLZ-5WH2 Wavy for PLZ-5WH2 is also referred to as Wavy.
• In this manual, the PLZ1005WH2, PLZ2005WH2, PLZ4005WH2, PLZ12005WH2, or PLZ20005WH2
Electronic Load is also referred to as the PLZ-5WH2 or PLZ-5WH2 series.
• “PC” in this manual is a generic term for personal computers and workstations.
• The screens used in this manual may differ from the actual items.
• The following markings are used in this guide.
Copyright
Reproduction and reprinting of this operation manual, whole or partially, without our permission is
prohibited. The specifications of this product and the contents of this document are subject to change
without prior notice.
You need to be careful about the auto update function of Window 10. If you use the default Windows
Update settings, Windows may automatically restart while a sequence is running.
See “About Windows updates (Windows 10 only)” in the setup guide, and construct the most stable OS
environment possible before running a sequence.
4 Press Modify.
You can now edit the interface settings.
5 Use the rotary knob to select Bitrate under RS232C Settings, and then press Edit.
6 Use the rotary knob to select the bit rate, and then press ENTER.
7 Use the rotary knob to select Flow Control, and then press Edit.
8 Use the rotary knob to select CTS-RTS, and then press ENTER.
9 Press Apply, use the rotary knob to select OK, and then press ENTER.
This completes the setting.
Configuring Wavy
1 Start Wavy.
3 In the Instrument list, click the model name of the connected PLZ-5WH2.
4 Click RS232C.
5 In the Port list, click the COM port number that the PLZ-5WH2 is using.
To check the port number, from Windows Control Panel, open Device Manager. Expand Port (COM
& LPT) to view the port number in use.
6 In the Baud rate list, click the same value as the Bitrate setting on the PLZ-5WH2.
7 Click Test to check that Wavy can communicate with the PLZ-5WH2.
Protocol settings other than baudrate and flow control are fixed on the PLZ-5WH2. They are set as
follows:
Data bits: 8 bits; stop bits: 1 bit; parity bit: none
Wavy exchanges data according to the above protocol settings. There is no need to manually
change them on the PC side.
8 Click OK.
This completes the setting.
To use GPIB, you need a GPIB converter (PIA5100 firmware version 1.02 or later) sold
separately. For details, see the PIA5100 operation manual.
4 Refer to the GPIB converter manual, and configure the necessary settings.
Configuring Wavy
1 Start Wavy.
3 In the Instrument list, click the model name of the connected PLZ-5WH2.
4 Select GPIB.
5 Select the GPIB ID you want to use with the GPIB ID parameter.
Normally the GPIB ID is 0 (GPIB0), but if several GPIB boards, USB-GPIB adapter, or the like are
connected to the PC, the ID is used to distinguish them. The drop-down list shows the results of
searching the GPIB IDs registered in VISA.
If the GPIB ID you want to use does not appear in the drop-down list, see “When the GPIB ID is not
displayed”(p.8).
6 In the Address list, click the GPIB address that you assigned with the GPIB
converter.
7 Click Test to check that Wavy can communicate with the PLZ-5WH2.
8 Click OK.
This completes the setting.
You need to register the GPIB ID in VISA. The following procedure is for when KI-VISA is used for the
VISA library. If you are using another VISA library, see the corresponding manual.
1 From the Windows start menu, select Kikusui IO Software and then Instrument
Explorer.
If a GPIB ID is registered in VISA, a character string that starts with GPIB will appear on the left side
of the KI-VISA Instrument Explorer window, as shown below.
Display example for ID: 0,
4 Select the “Use GPIB instrument,” “Enable dynamic search for active controllers
and listeners,” and “Use this GPIB board” check boxes.
6 From the Model drop-down list, select the applicable model name.
7 Click OK.
The KI-VISA IO Config window closes.
8 Check that the GPIB ID that you registered is displayed on the left side of the KI-
VISA Instrument Explorer window.
The registered ID will also appear in the GPIB ID drop-down list in the Interface window.
To use the USB interface to control the PLZ-5WH2, a device driver that supports the USB
Test & Measurement Class (USBTMC) must be installed in the controller. The USBTMC
driver is installed automatically by the VISA library. For details on installing the VISA library,
see the included setup guide.
Configuring Wavy
1 Start Wavy.
3 From the Instrument list, selected the model name of the connected PLZ-5WH2.
4 Select USB.
5 From the Serial Number list, select the serial number of the PLZ-5WH2.
For the location of the serial number, see the PLZ-5WH2 Operation Manual.
If the serial number does not appear even with a proper connection, from Windows Control Panel,
open Device Manager, and check whether “USB Test and Measurement Device” is displayed. If it is
not displayed, reinstall the VISA library.
6 Click Test to check that Wavy can communicate with the PLZ-5WH2.
7 Click OK.
This completes the setting.
If a network problem occurs, an unexpected dangerous voltage may occur that may
WARNING cause electric shock, fire, physical damage to the DUT, and so on. If you are going to
remotely control the PLZ-5WH2 from a distance, install a Web camera or take other
measures to monitor the status.
Connection
Use a LAN cable (category 5 or better, straight) to connect the PLZ-5WH2 using method (a) or (b).
(a) Connect the PLZ-5WH2 to a corporate LAN or router.
(b) Connect the PLZ-5WH2 directly to the PC or through a switching hub.
When connected to the network, the LAN icon on the front panel turns green.
Setup
PC Configuration
Method (a)
There are no particular settings you need to make.
Method (b)
Set the Internet protocol version 4 (TCP/IPv4) properties as follows:
• Windows Start (right-click) > Network and Internet > Network and Sharing Center > Ethernet >
Properties > Internet Protocol Version 4 (TCP/IPv4) > Properties
When the PLZ-5WH2 shipped from the factory, the IP address assignment method is set to automatic.
Unless there is a special reason otherwise, you can simply use these settings for methods (a) and (b).
If you want to use a fixed IP address, on the PLZ-5WH2’s front panel, press SYSTEM > Interface > Modify
> Edit. Set Method to Static, and set IP Address, Subnet Mask, and Default Gateway as you like. For
details, see “System Configuration” the user’s manual.
To control the PLZ-5WH2 from Wavy through LAN, you need to know the PLZ-5WH2’s IP address.
If you are going to use the PLZ-5WH2 with the factory default settings, the PLZ-5WH2’s IP address will be
automatically assigned. To check the IP address, follow the procedure below.
Configuring Wavy
1 Start Wavy.
3 From the Instrument list, selected the model name of the connected PLZ-5WH2.
4 Click LAN.
6 Click Test to check that Wavy can communicate with the PLZ-5WH2.
7 Click OK.
This completes the setting.
If you are not using a fixed IP address, the PLZ-5WH2’s IP address may change, causing the PC to no
longer recognize the PLZ-5WH2. If this happens, check the IP address by referring to “Checking the PLZ-
5WH2’s IP address”(p.12).
In the Graph window, you can create steps by drawing a graph with the mouse (p.31).
In the Worksheet window, you can edit steps by entering values and conditions (p.35).
The status bar shows the following information.
SCPI error codes may be displayed while using Wavy. If an error code is displayed, see the PLZ-5WH2
Communication Interface Manual.
A sequence consists of programs and steps. A program is a collection of steps. Steps are executed in
order one at a time, starting from step 1. The completion of the last step signifies that the program has
been executed once. The same program can be executed for a specified number of times, or different
programs can be executed consecutively (only when in the same operation mode, using the chain
function). When all programs have been executed, the sequence stops.
Time
Program1 Program2 (2nd repetition) Program2 (2nd repetition)
Step 1 Step 2 Step 3 Step 1 Step 2 Step 3 Step 1 Step 2 Step 3
Sequence
Set a program for each operation mode. Up to 30 programs can be set.
Up to 10000 steps total can be used in all programs.
If transition method of step 1 is set to Ramp, the start current of step 1 will always be 0 A.
• Do not create more than 10000 steps. If the total number of steps exceeds 10000,
unexpected behavior may result.
• Wavy does not check the validity of the settings for the PLZ-5WH2. Do not set a step to a
value outside the PLZ-5WH2 range specifications. If you transfer a program containing
invalid values to the PLZ-5WH2, you will not be able to run it.
Main functions
The available load value transition methods are step and ramp. With step, the load value transitions step-
wise from the value of the previous step. With ramp, the load value transitions in a ramp from the value of
the previous step.
Example when the value is 10 A and the interval 100 s
Step Ramp
10 A 10 A
100 m s 100 ms
If step 1 is set to ramp transition, the start point of the load value is always 0 A / 0 S / 0 V /
0 W.
Set the operation mode, slew rate, and response speed for executing program.
You cannot change the operation mode when a step is entered in the Graph window or Worksheet
window, so set the mode when there is no entered steps.
Constant Cur- The operation mode is set to constant current. Enter the default step slew rate for cre-
rent ating steps. You can change the slew rate for each step when a step is created.
Constant The operation mode is set to constant resistance. Set the response speed to Normal or
Resistance Fast.
Constant Volt- The operation mode is set to constant voltage. Set the response speed to Normal or
age Fast.
Constant The operation mode is set to constant power.
Power
3 Click OK.
This completes the setting.
If you set the operation mode (p.18) to Constant Resistance, set the resistance range for executing pro-
gram.
3 Click OK.
This completes the setting.
You can set the unit of the step execution time to use when creating steps.
You cannot change the step execution time when a step is entered in the Graph window or Worksheet
window, so set the time when there is no entered steps.
Unit Interval setting range Resolution
ms 0.050 ms to 9999999.999 ms 0.001 ms
(milliseconds)
s (seconds) 0.001 s to 99999.999 s 0.001 s
min (minutes) 0.1 min to 999.9 min 0.1 min
h (hours) 0.1 h to 1000.0 h 0.1 h
2 In the Unit list, set the unit of step execution time to ms, s, min, or h.
3 Click OK.
This completes the setting.
This function limits the current when the voltage applied to the PLZ-5WH2 becomes equal to or less than
the UVP setting.
You can set this function when Limit for the undervoltage protection (UVP) is enabled in the hardware
protection settings (p.28).
The UVP Limit value is the same as the undervoltage protection value of the protection settings.
1 In the hardware protection settings (p.28), enable Limit for the undervoltage
protection.
4 Click OK.
This completes the setting.
You can control the timing for starting the program after clicking Run in the sequence run window (p.43).
Setting Value Description
Source — Set an event (trigger source) that will be used as a condition for starting program.
Immediate A trigger is applied immediately.
BUS A trigger is applied when the PLZ-5WH2 receives a *TRG command from a PC
or the like.
DIGITAL21 A trigger is applied when DIGITAL2 of the PLZ-5WH2’s EXT CONT connector
receives a signal.
MSync The trigger application timing is synced between the PLZ-5WH2 series that are
synchronized. A trigger is applied when you press MSync on the front panel of a
PLZ-5WH2 that is connected in sync or when a PLZ-5WH2 receives a
TRIG:TRAN:MSYN command from a PC or the like.
Delay 0 s to 100 s Set the delay time from trigger application until the start of program.
1. Only when Direction of Digital 2 is set to Input on the PLZ-5WH2.
3 Under Delay, enter the delay time from trigger application until the start of program.
4 Click OK.
This completes the setting.
You can set the load state (load on or off, load value, slew rate) after the program completion.
The last step function is applied to each program even when multiple programs are executed
consecutively (p.25). If a single program is repeated (p.26), the last step function is applied only to the last
step of that program.
Depending on the combination with the “Load OFF after Completion of Processing” function (p.27), the
operation after the completion of the sequence is as follows.
Last step: Enabled Last step: Disabled
Load OFF after Completion of The load is turned off by the “Load The load is turned off by the “Load
Processing1 : Enabled OFF after Completion of OFF after Completion of Processing”
Processing” function after the last function.
step function is executed.
Load OFF after Completion of Depends on the last step setting. Retains the last step value.
Processing: Disabled
1. The “Load OFF after Completion of Processing” function is executed after the sequence have been completed.
2 To enable, select the Last step ON check box. To disable, clear the check box.
3 If you enable the function, set the load state after program completion.
4 Click OK.
This completes the setting.
Cutoff function
This function turns the load off when the elapsed time from load-on, the voltage drop, the integrated
current, or the integrated power reaches the specified value. These four functions are collectively referred
to as cutoff.
For details on the cutoff function, see the PLZ-5WH2 user’s manual.
Cutoff item Description
Elapsed time (ETime) The load turns off when the elapsed time value reaches the specified
value.
Voltage drop1 (Voltage) The load turns off when the voltmeter value reaches the specified
value.
Integrated current The load turns off when the ampere-hour meter value reaches the
(Capacity) specified value.
Integrated power The load turns off when the watt-hour meter value reaches the specified
(Energy) value.
1. The same operation is performed using Trip of the low voltage protection (UVP), but whereas
when an UVP is activated, an alarm occurs, in the case of cutoff, an alarm does not occur.
3 Set State of the cutoff item you want to change to Enable or Disable.
4 If you enable an item, enter the value for activating the cutoff in the Level box.
For Capacity and Energy, you can change the unit.
5 Click Set.
The settings are applied to the PLZ-5WH2.
6 Click OK.
This completes the setting.
Chain function
This function executes programs registered in the PLZ-5WH2 consecutively in the order you like. You can
load programs from the PLZ-5WH2, and edit the programs and the execution order in the Wavy List of
Registration. The programs registered in the list are consecutively executed in order.
2 To enable the chain function, select the Chain Function ON check box. To disable,
clear the check box.
3 If you enable the function and want to load programs from the connected PLZ-
5WH2, click Program list will be loaded.
If a list of registration is saved, it is cleared, and the programs are loaded in Program List.
4 From Program List, add programs to List of Registration, and change the order of
the programs as you like.
Program List Programs loaded from the PLZ-5WH2 are displayed. Clicking ADD→ causes the
selected program to appear in List of Registration.
List of Registration A list of programs registered in Wavy is displayed. You can change the order of
the selected program by clicking UP or DOWN.
Clicking ← DELETE causes the selected program to be deleted from List of
Registration.
Set how many times to repeat the program when the sequence is executed.
2 In the Repetition Count box, enter the number of times that the program will repeat.
If you want to repeat it infinitely, select the Infinity check box.
3 Click Close.
To apply the settings to the PLZ-5WH2, write the program in the PLZ-5WH2 (p.41).
Set whether to turn off the load after the sequence have been completed (the “Load OFF after Completion
of Processing” function).
Depending on the combination with the last step setting (p.23), the operation after the completion of the
sequence is as follows.
Last step: Enabled Last step: Disabled
Load OFF after Completion of The load is turned off by the “Load The load is turned off by the “Load
Processing: Enabled OFF after Completion of OFF after Completion of Processing”
Processing” function after the last function.
step function is executed.
Load OFF after Completion of Depends on the last step setting. Retains the last step value.
Processing: Disabled
2 To turn the load off after sequence completion, select the Load OFF after
Completion of Processing check box. To disable, clear the check box.
Hardware settings
Hardware settings are protection (alarm) functions available on the PLZ-5WH2. You can set overcurrent
protection (OCP), overpower protection (OPP), undervoltage protection (UVP), and watchdog protection.
Software settings
These protection functions are activated by Wavy according to judgments made on the monitored
data. You can set overvoltage protection (OVP), overcurrent protection (OCP), undervoltage
protection (UVP), and Undercurrent protection (UCP).
2 Click Read.
The values are read from the connected PLZ-5WH2.
Overcurrent Enter the OCP value (0 % to 110 % of the rated current). To turn the load off when
protection OCP alarms occur, select the Load Off check box. To keep the load turned on and control
the current so that the OCP value is not exceeded, clear the check box.
Overpower Enter the OPP value (0 % to 110 % of the rated power). To turn the load off when
protection OPP alarms occur, select the Load Off check box. To keep the load turned on and control
the current so that the OPP value is not exceeded, clear the check box.
Undervoltage To enable UVP, select the Enable check box, and enter the value (0 V to 1000 V).
protection UVP To turn the load off when alarms occur, select the Trip check box. To keep the load
turned on and limit the current so that it does not fall below the UVP value, select
the Limit check box.
WDOG To enable watchdog protection, clear the check box. When enabled, the PLZ-
5WH2’s watchdog protection time value (delay) is set to 5 seconds.
4 Click Set.
The values are sent to the PLZ-5WH2 and applied to the protection settings for direct control
(p.60) and command control (p.64).
Just clicking Set does not apply them to the protection settings of each program. To apply them to
the protection settings a program, write the program in the PLZ-5WH2 (p.41).
5 Click OK.
This completes the setting.
If you set overvoltage protection (OVP) or overcurrent protection (OCP), a sequence stops when a Wavy’s
monitored value exceeds the set value. If you set undervoltage protection (UVP) or undercurrent
protection (UCP), a sequence stops when a Wavy’s monitored value falls below the set value.
2 Select the check box for the protection function that you want to set.
The text box for entering the protection value is enabled.
4 Click OK.
This completes the setting.
As the first monitored values are unstable, they may activate a protection function and stop a sequence. If
you select the “Ignore the first monitored value” check box, a sequence will continue its execution even
when the first monitored values exceed or fall below the protection values.
You can create and edit steps by drawing graphs with the mouse.
You can also create and edit steps by entering values (p.35).
Creating steps
Graph window
2 Drag the pointer to the time and the current or voltage that you want to set.
Step-division line
Drag
The setting is confirmed, the step-division line is displayed, and the step values are entered in the
first row of the worksheet. You can change the unit of step interval (p.20) and graph scale (p.54).
This completes the creation of step 1.
3 If you want to continue creating steps, move the mouse pointer over the end point
of the step that you have just created. When the mouse pointer becomes a
crosshair, drag the mouse pointer to the point that corresponds to the time and the
current or voltage that you want to set.
The value is confirmed, and the step values are entered in the first blank row of the Worksheet
window.
Repeat this procedure until you have set all the steps.
To change the X-axis and Y-axis scales according to the graph that you drew, on the Graph menu, click
Update Auto Scale, or right-click in the Graph window, and click Update Auto Scale.
You can set the time interval (X-axis) resolution and the value (Y-axis) resolution.
2 Click Resolution of time period or Resolution of setting value, and select the
resolution.
Resolution of time Select Default, 1st digit of integer, 1st digit of decimal place, or 2nd digit of
period decimal place. The default is three decimal places.
Resolution of setting Select Default, 1st digit of integer, 1st digit of decimal place, 2nd digit of decimal
value place, or 3rd digit of decimal place. The default is four decimal places.
Editing steps
3 Drag the pointer up or down to the value that you want to change to.
Drag
1 Double-click the vertical line (the step-division line) at the end point of the step that
you want to edit.
A black square is displayed at the top of the step-division line, and you can edit the step.
3 Drag the pointer left or right to the value that you want to change to.
Drag
2 On the Graph menu, click Transition. Or, right-click, and click Transition.
If you want the transition method to always be step when drawing graphs, on the Graph menu, click
Always draw as steps. Or, right-click in the Graph window, and click Always draw as steps.
Deleting a step
2 On the Graph menu, click Delete. Or, right-click in the Graph window, and click
Delete.
The step is deleted.
You can create steps by entering values in tables. You can also copy and delete steps using tables.
You can also create and edit steps by drawing graphs (p.31).
Creating steps
This section will explain how to create a step in the Worksheet window.
Step 1
Step 2
Step 3
You can use the Worksheet window to copy and delete the steps you have created.
2 On the Worksheet menu, click Copy. Or, right-click in the Worksheet window, and
click Copy.
3 Click OK.
5 On the Worksheet menu, click Paste. Or, right-click in the Worksheet window, and
click Paste.
6 Click Yes.
The copied step is inserted before the selected step, and the total number of steps increases by 1.
Deleting steps
You can delete steps (rows). You cannot delete the content of cells individually.
2 On the Worksheet menu, click Delete. Or, right-click, and then click Delete.
3 Click Yes.
The step is deleted.
Deleted steps are copied to the clipboard. If you insert a step immediately after deleting steps, the deleted
steps will be inserted.
You can undo the last copying or deleting of steps by clicking Undo on the Worksheet menu, or right-
clicking the Worksheet window and clicking Undo.
Keyboard shortcuts
Undo1 Z or Ctrl+Z
Select all rows A or Ctrl+A
1. A confirmation dialog appears before execution.
Saving a program
To overwrite a file, on the File menu, click Save. Or, click on the toolbar.The file name extension is
“.wvy.”
You can edit saved program data using Notepad or other text editor.
When editing program data, do not change the case of letters. If you do, the sequence will no
longer work.
• By default, data is saved as tab-separated values. Data can also be saved as comma-separated
values (p.59).
• The first row shows the name of the connected electronic load series (PLZ5WH2).
• The second row shows the following items.
Loading a program
To open the file you want to load, on the File menu, click Open. Or, click on the toolbar.
This section describes how to write (transfer) programs that you create in the PLZ-5WH2.
The following settings are also transferred along with the program.
• Repetition Count
• Load OFF after Completion of Processing
Program name To overwrite a program that you created on the PLZ-5WH2, select the name
from the upper list. To write a new program, select the lower list, and enter a file
name using up to 255 characters.
Erase All Programs Select this check box to erase all programs stored in the PLZ-5WH2 first and
then transfer the file. The programs that are deleted are those that correspond to
the operation mode of the program to be transferred.
Repetition Count Set the number of times that the program will repeat. If you want to repeat it
infinitely, select the Infinity check box.
Load OFF after Select this check box to turn the load off after the sequence have finished (p.27). If
Completion of you want to retain the value of the last step of the sequence or use the value of
Processing1 the last step (p.23), clear the check box.
1. If the Load OFF after Completion of Processing check box is selected, the load turns off after the
sequence have finished even when LOAD is set to ON in the last step settings (p.23).
It is also possible to load programs from the PLZ-5WH2 into Wavy. You can use this function to back up
programs that you create on the PLZ-5WH2 or to edit them.
The number of programs that can be loaded is 30. Loading more than 30 programs may
make the Wavy operation unstable.
2 Under Read, select the name of the program from the Program name list.
• Before executing any program that has been created on Wavy, be sure to write the
program in the PLZ-5WH2. (p.41) If you do not write the program first, the program
selected on the PLZ-5WH2 will be executed.
• If you want to save the monitored data during sequence execution, specify the monitored
data save setting in advance (p.49).
If you selected the “Display the monitor graph” check box in Monitoring Setup (p.47), the Monitor
Graph tab appears. In Monitor Graph, you can view in real time the voltage, current, and power
values during sequence execution.
If the chain function is enabled (p.25), “Chain Function ON” is displayed in the upper right of the
window. The programs registered by the chain function are executed.
2 Click Run.
The sequence is executed. To abort a sequence, click Stop.
To pause a sequence, click Pause. While a sequence is paused, clicking Release will release the
pause.
Status indicator
While a sequence is running, the menus and the toolbar are disabled. Further, the window cannot
be resized.
Item Description
Green line The approximate position of execution. For sequences that have a large number of
repetitions or those that take a long time to run, the position indicated by the green line
may appear offset from the actual position.
program The number of the program that is being executed.
Count The present repetition count.
step The step that is currently being executed.
Elapsed Time The elapsed time since the start of sequence execution.
Voltage, Current, If the Voltage, Current, and Power check boxes are selected in Monitoring Setup (p.46),
and Power the corresponding load values are displayed.
Status indicators
Status Description
Standing by.
A sequence is running.
When the monitor graph is displayed (p.47) , executing a long sequence will increase the
monitor graph data and put a heavy load on the PC. In such situations, you can reduce the
load by setting a maximum number of data entries of the monitor graph (p.58).
If an alarm occurs, remove the cause of the alarm, and click Alarm Clear/Load OFF to clear the alarm.
If you want to turn off the load, click Alarm Clear/Load OFF.
If communication with the PLZ-5WH2 is disconnected while a sequence is running, the status in the
sequence run window will change to “Error” or “End.”
The PLZ-5WH2 will continue executing the program even when communication is disconnected, because
the program is already transferred to the PLZ-5WH2. Even if communication is restored while the
sequence is running, Wavy will not be able to obtain monitored data.
An error will occur if you try to run the sequence again.
You can specify whether to display monitored values (current, voltage, power) and in the sequence run
window.
2 Under Select Items to be Monitored, select the check boxes for the monitored
values you want to display.
3 Click OK.
This completes the setting.
You can specify whether to show the monitor graph in the sequence run window. If you show the monitor
graph, you can view in real time on graphs the current, voltage, and power values during sequence
execution. You can also view the setting graph and the monitor graph in a single window.
You can change which graphs (current, voltage, power) to show in the monitor graph (p.46).
Click the tab to switch.
Monitor graph Move your mouse pointer over the Setting graph and monitor graph
graph to display the monitored
values.
When the monitor graph is displayed, executing a long sequence will increase the monitor
graph data and put a heavy load on the PC. In such situations, you can reduce the load by
setting a maximum number of data entries of the monitor graph (p.58).
2 To display the monitor graph, select the “Display the monitor graph” check box. To
not display it, clear the check box.
3 Click OK.
This completes the setting.
2 In the Monitoring Interval box, enter the interval (200 ms to 600000 ms).
3 Click OK.
This completes the setting.
You can save the monitored data obtained by executing sequences as a text file. Enabling this setting will
automatically save monitored values during sequence execution.
2 Under Save File, select Yes. Then, specify the save destination.
The file name consists of the operation mode, the execution start date and time, and the file name
extension. You can change the file name extension (the default is .txt) (p.59).
3 Under Unit of Elapsed Time for Output to File, select the output unit.
[s](S) Seconds
[h:min:s](H) Hours, minutes, seconds
The accuracy of the elapsed time varies depending on the PC environment that you are using.
4 Under “for Saved Data,” select the check boxes of the options you want to use.
Data right before Run The monitored values before sequence execution (monitored values at time
0 s) are written at the beginning of the saved data.
Data Immediately after The monitored values when the sequence is completed or aborted are
End/Stop written at the end of the saved data.
5 Use the settings under “Order for Output to File” to set the order that time, voltage,
current, and power are written to files.
6 Click OK.
This completes the setting.
You can view saved monitored data using Notepad or other text editor.
By default, data is saved as tab-separated values. You can also change to comma-separated values
(p.59).
Setting graph in the Run window Monitor graph in the Run window
To display the monitor graph, see “Showing or hiding the monitor graph” (p.47).
The items that you can set vary depending on the graph.
For the graphs in the Graph window and setting graph and monitor graph in the sequence run window, you
can show or hide the X-axis and Y-axis scales and set the number of scale lines and the line types (dotted
or solid).
Right-click the graph, and then click X-axis Scale Lines or Y-axis Scale Lines to show or hide scales.
For the Graph window, you can also show or hide the scales by clicking X-axis Scale Lines or Y-axis Scale
Lines on the Graph menu.
You can set the number of graph scale lines and the type of scale lines (solid or dotted).
Graph window
You can also right-click in the graph, and click Scale to display the Scale window.
Number of scale Enter the number of scale lines for the X-axis and Y-axis.
lines
Dot To use dotted scale lines, select the check box. To use solid scale lines, clear the
check box.
Number of Enter the number of scale lines for the X-axis and Y-axis.
scale lines
Dot To use dotted scale lines, select the check box. To use solid scale lines, clear the
check box.
You can change the line colors and background color of the graphs in the Graph window and the setting or
monitor graphs in the Run window. The items that you can set vary depending on the graph.
1 Right-click the graph, select Color, and select the item you want to change.
The color setting window appears.
For the Graph window, you can also display the color setting window by clicking Color on the
Graph menu and then selecting the item you want to change.
You can show or hide the division lines (vertical axis) between steps in the graphs in the Graph window
and the setting graph in the sequence run window.
Step-division line
Right-click the graph. Each time you click Vertical Axis, the step-division lines are shown or hidden.
For the Graph window, on the Graph menu, click Vertical Axis. Or, click on the toolbar. This will
show or hide the step-division lines.
You can set the X-axis and Y-axis scales on the graphs in the Graph window and the monitor graph in the
sequence run window.
Graph window
You can also right-click in the graph, and click Scale to display the Scale window.
3 To manually set the scales, enter the X-axis and Y-axis scales.
X-axis Maximum Time Enter the maximum time to display on the X-axis.
Minimum Time Enter the minimum time to display on the X-axis.
Y-axis Maximum Value Enter the maximum value to display on the Y-axis.
Minimum Value Enter the minimum value to display on the Y-axis.
4 Click OK.
This completes the setting.
3 If you select the Auto check box of the X-axis, select the graph scroll mode.
Roll mode The X-axis and Y-axis are scrolled at the same time according to the elapsed time
and load values. The display range is defined by subtracting the minimum time
value from the maximum time value.
Normal mode The minimum time is fixed. The maximum time is updated to the appropriate value.
If you want to scroll only when the monitored value exceeds the maximum X-axis
value, select the “Scrolling within the specified range” check box.
The amount of time that the graph will be scrolled by is determined by subtracting
the minimum time value and the set scroll time from the maximum time value.
For example, if the minimum time is set to 0 seconds, the maximum time to 23
seconds, and the scroll time to 10 seconds, the time range that will be scrolled is 13
seconds.
4 To manually set the scales, set the X-axis and Y-axis scales.
X-axis Minimum Time Enter the minimum time to display on the X-axis.
Maximum Time Enter the maximum time to display on the X-axis.
Y-axis Min current, Min voltage, Min power Enter the minimum value to display on the Y-axis.
Max current, Max voltage, Max power Enter the maximum value to display on the Y-axis.
You can show or hide current, voltage, and power lines on the monitor graph in the sequence run window.
Right-click the graph, select Line Display, and select the value you want to show. Each time you select,
show and hide toggle.
You can set the Y-axis value (current, voltage, or power) that is displayed when the mouse pointer is
placed on the Y-axis on the monitor graph in the sequence run window.
Right-click the graph, select Y-Axis Value, and select Current, Voltage, or Power.
You can change the X-axis (time-axis) unit (s or h:min:s) of the monitor graph in the sequence run window.
Right-click the monitor graph, click X-axis Unit and select “s” to set the unit to seconds. If you click X-axis
Unit and select “h:min:s,” the unit is set to hour:minute:second.
Each time you select Toolbar or Status Bar from the View menu, the toolbar or status bar switches
between show and hide.
Each time you select Save Position from the View menu, the setting for saving the window position and
size switches between on and off.
Arranging windows
You can change the way the Graph and Worksheet windows are displayed.
On the Window menu, click Cascade, Tile Vertically, or Tile Horizontally to change the way that windows
are displayed.
Executing a long sequence will increase the amount of monitor graph data on the sequence run window.
This puts a heavy load on the PC and may cause the PC to malfunction. In such situations, you can
reduce the load by setting a maximum number of data entries of the monitor graph.
Even if you set the maximum data size, all monitored data will be saved.
3 Click OK.
This completes the setting.
When you run a long sequence, we recommend that you use Task Manager to check the amount
of physical memory being used.
Set the number of significant decimal digits of the load values (current, conductance, voltage, power) when
creating steps. The actual number of significant decimal digits varies depending on the model and range of
the PLZ-5WH2.
3 Click OK.
This completes the setting.
You can change the file format for saving program and monitored data.
Sequence data file Click Tab Separation or Comma Separation1 to select the file format for saving
format programs and ARB mode settings.
Monitor output data Click Tab Separation or Comma Separation*1 to select the file format for saving
file format monitored data. Select the extension (txt, csv, log) or enter an extension of your
choice.
1. For regions in which comma separation cannot be used, select Tab Separation.
3 Click OK.
This completes the setting.
To copy the image of the graphs in the Graph window, click the graph, and on the Graph menu, click Copy.
You can set whether to display the program that was showing when Wavy was closed previously. Each
time you click Previous File on the View menu, show and hide toggle.
The Remote Control Panel window is used to control the PLZ-5WH2 directly.
On the Tool menu, click Direct Control to display the Remote Control Panel window.
(4)
(5) (6)
No. Description
(1) Entering operating conditions(p.61)
(2) Entering load values(p.61)
(3) Set the pulse function (p.62)and sine function(p.63)
(4) Displays execution results.
(5) Turn the load on and off, or clear alarms (p.63).
(6) Monitor function. Save monitored values (p.64).
Item Description
Operation mode Select the operation mode. If you select ARB, set the I-V characteristics in the ARB
mode window (p.65).
Resistance range Select the resistance range for CR mode. During parallel operation, the master unit’s
resistance range is displayed.
Response CV Select the response speed for CV mode.
Response CR Select the response speed for CR mode.
UVP Limit This function limits the current when the voltage applied to the product becomes equal
to or less than the UVP setting. Entering a voltage and clicking Set will apply the UVP
voltage to the PLZ-5WH2. You can set this function when Limit for the PLZ-5WH2’s
undervoltage protection (UVP) is enabled.
Slew rate Entering a slew rate and clicking Set will apply the slew rate to the PLZ-5WH2. To load
the slew rate set on the PLZ-5WH2, click Read.
Click the tabs of the relevant operation modes, enter the load values and the maximum and minimum load
values, and click Set. If you click Up or Down, the load value increases or decreases by the value entered
in the Step box.
This function repeatedly executes two settings. It is suitable for transient response characteristics testing
of large capacity power supplies and batteries. This function operates in CC and CR modes.
Select the Pulse Function tab.
Item Description
Pulse Function ON Select the check box to enable the pulse function.
Pulse amplitude (Depth) Enter the low-level current. The high-level current becomes the load value.
Frequency Set the pulse frequency.
Duty cycle Set the ratio of the high level in one cycle.
Read Reads the settings from the connected PLZ-5WH2.
Set Applies the settings to the PLZ-5WH2.
Frequency
Duty cycle
Current
Load value
This function varies the current sinusoidally. It is suitable for superposed ripple testing of large capacity
power supplies and batteries. This function operates in CC mode.
Select the Sine Function tab.
Item Description
Sine Function ON Select the check box to enable the sine function.
Sine amplitude Enter the sine amplitude. The center value of the amplitude becomes the load
(Amplitude) value.
Frequency Set the pulse frequency.
Read Reads the settings from the connected PLZ-5WH2.
Set Applies the settings to the PLZ-5WH2.
Current Frequency
Sine amplitude
Load value
0 Time
Item Description
Status This shows the load on/off status.
ON Turns the load on.
OFF Turns the load off.
Clear Clears the PLZ-5WH2 alarm.
Item Description
Max, Min Enter the maximum and minimum monitored values.
Voltage, Current, Select the check boxes for the values you want to monitor. The monitored values will be
Power, Energy displayed while monitoring is in progress. Clear the check boxes if you do not want to
monitor the values.
Status Shows the PLZ-5WH2 status (p.44) while monitoring is in progress.
Elapsed Time Displays the elapsed time since the start of monitoring.
Run Starts the monitor. You cannot close the Remote Control Panel window while monitoring
is in progress.
Stop Stops the monitor.
Interval Enter the monitoring interval (200 ms to 600000 ms).
Save File Select the check box to save monitored data. Click Folder to select the file save
destination, the unit of the elapsed time that will be output, and the order in which values
will be output.
The file name consists of “Wavy,” the monitor start date and time, and file name
extension. You can change the file name extension (the default is .txt) (p.59).
For details on commands, see the Communication Interface Manual on the CD-ROM included with the
PLZ-5WH2.
In ARB mode, arbitrary I-V characteristics can be set by registering multiple I-V characteristic points (pairs
of voltage and current values). Three to 100 points can be registered, and the space between two points is
linearly interpolated. The minimum voltage (0.00 V) and current (0.00 A) and the maximum voltage
(1010.00 V) are fixed. This mode can be used for simulation of LED loads and the like.
On the Tool menu, click Arb Mode to display the ARB Mode window.
(1)
(2)
(3)
No. Description
(1) Saving and loading settings (p.67)
(2) Setting area(p.66)
(3) The settings entered in the setting area are shown on a graph. When you move the mouse pointer over
the Y-axis or X-axis, the current or voltage is displayed. When you move the mouse pointer over the
graph line, the voltage and current settings are displayed.
In the setting area, enter the load voltage and current values.
Item Description
Table of voltages and Double-click the voltage and current cells, and enter the load values.
currents Change the number of rows in the Data num box.
The voltage and current on the first row and the voltage on the last row cannot be
changed.
For voltage, you can only enter a value greater than the value of the previous row.
To copy and paste a value, select the cell, right-click, and select Copy and Paste.
Response Set the response speed for ARB mode. The value is the filter response time. OFF is
for no filter.
(OFF/0.5 ms/1 ms/2 ms/5 ms/10 ms/20 ms/50 ms/100 ms)
Data num Set the number of rows of the table (3 to 100). If you change the number of points, a
row is added or deleted at the row above the last row.
Minimum voltage, The minimum and maximum voltages of ARB mode. Only the maximum voltage can
Maximum voltage be changed.
Default current value Enter the default value for the current column of the setting worksheet.
Update Applies the entered settings to the table.
Get Obtains ARB values from the connected PLZ-5WH2.
Set Applies the edited contents to the PLZ-5WH2.
Save The settings on the setting worksheet are saved to a text file (IV file (.iv)). You can
edit IV files (p.67).
Open Loads a saved IV file and applies the settings to the table.
You can save the ARB mode settings to a text file (IV file (.iv) or .csv). You can edit and load IV files.
Item Description
Open Loads a saved IV file.
Save Saves the entered ARB mode settings to an IV file.
You can edit saved IV files using a text editor. Edited IV files can be loaded from ARB mode setting.
• By default, data is saved as tab-separated values. Data can also be saved as comma-separated
values (p.59).
• In the first line, the minimum and maximum voltages are displayed. Do not change these values.
• In the following lines, a voltage is displayed first and then a current. Do not change the voltage and
current on the second line or the voltage on the last line.
If you set the operation mode to ARB in the Remote Control Panel window (p.60) or command control
window (p.64), you can control the PLZ-5WH2 in ARB mode.
You can set the measurement recording conditions after Run is clicked in the sequence run window (p.43).
For details on the recording conditions, see the PLZ-5WH2 user’s manual.
Setting Value Description
Source — Set an event (trigger source) that will be used as a measurement
recording condition.
Immediate A trigger is applied immediately.
BUS A trigger is applied when the PLZ-5WH2 receives a *TRG command
from a PC or the like.
DIGITAL21 A trigger is applied when DIGITAL2 of the PLZ-5WH2’s EXT CONT
connector receives a signal.
TALink If you set Trigger OUT (p.36) to TALink in a program step, a trigger is
applied at the start of the step.
MSync The trigger application timing is synced between the PLZ-5WH2 series
that are synchronized. A trigger is applied when you press MSync on the
front panel of a PLZ-5WH2 that is connected in sync or when a PLZ-
5WH2 receives a TRIG:TRAN:MSYN command from a PC or the like.
LOADOFF A trigger is applied when the load is turned off.
Count 1 to 65536 Sets the number of times measurement values are to be recorded.
Delay 0 s to 100 s Set the delay time from trigger application until measurement value
(resolution: 10 μs) recording.
Interval Disable/Enable Set whether to insert an interval between recordings when Count is 2 or
higher.
Interval Time 0.00001 s to 3600 s Set the recording interval time when Interval is set to Enable.
(resolution: 10 μs)
Sense 0.00001 s to 1 s Set the recording time period per iteration. The average over the time
Aperture (resolution: 10 μs) period is recorded.
1. Only when Direction of Digital 2 is set to Input on the PLZ-5WH2.
3 Enter or select the trigger settings with Source, Count, Delay, Interval, and Interval
Time.
4 Enter the recording time period per iteration in the Sense Aperture box.
5 Click Set.
The settings are applied to the PLZ-5WH2.
6 Click Close.
This completes the setting.
File
New1 Creates a new file.
Worksheet
Select All Selects all cells.
Undo 1 Undoes the previous operation.
Sequence Creation Mode1 Set the step interval unit, operation mode, resistance range,
undervoltage protection, program start trigger, and last step.
Monitoring Setup1 Configure monitor settings used during sequence execution.
Protection Setup Configure the hardware and software protection functions.
Interface Set the conditions that are used to connect the PC and PLZ-
5WH2.
CutOff Mode This function turns the load off when the elapsed time from load-
on, the voltage drop, the integrated current, or the integrated
power reaches the specified value. Set the conditions for
activating the cutoff function.
Environment Settings Set the file format for saving text files.
Options Set the number of significant digits for load values.
Chain mode Executes multiple programs consecutively in the order you like.
Load the programs registered in the PLZ-5WH2, and set the
programs and their order to be executed on Wavy.
Tool
Direct Control Directly control the PLZ-5WH2.
Command Control Enter commands to control the PLZ-5WH2.
Arb Mode Set arbitrary I-V characteristics in the PLZ-5WH2.
Acquire trigger You can set the measurement recording start conditions after Run
is clicked in the sequence run window.
Window
Cascade Displays cascaded windows.
Tile Vertically Displays horizontally tiled windows.
Tile Horizontally Displays vertically tiled windows.
Help
Help Topics Opens the help file.
About Wavy 1 Shows the program version and copyright information.