Lab Advisor Diagnostic Result HPLC 1200 (LB-099)

G1315D Intensity Test Passed


DE64258593 The test scans the Intensity spectrum generated by the UV and VIS Lamp.
B.07.10[0004]
DAD

Started at 7/1/2020 7:33:04 AM by laboratorio-PC\laboratorio on LABORATORIO-PC


using Agilent Lab Advisor - Version B.02.08 [178] SP1 - Basic
Results
Item Value Result
Accumulated UV Lamp Burn Time 0.00 h Done
UV Lamp On-Time 0.00 h Done
Accumulated Vis Lamp Burn Time 1834.14 h Done
Vis Lamp On-Time 0.00 h Done
Lowest Intensity in Range 190 - 220 nm 7753 Counts
Lowest Intensity in Range 190 - 220 nm 2000 Counts Passed
Lowest Intensity in Range 221 - 350 nm 18763 Counts
Lowest Intensity in Range 221 - 350 nm 5000 Counts Passed
Lowest Intensity in Range 351 - 500 nm 14620 Counts
Lowest Intensity in Range 351 - 500 nm 2000 Counts Passed
Lowest Intensity in Range 501 - 950 nm 17974 Counts
Lowest Intensity in Range 501 - 950 nm 2000 Counts Passed
Highest Intensity in Range 190 - 350 nm 40354 Counts
Highest Intensity in Range 190 - 350 nm 2000 Counts Passed
Highest Intensity in Range 700 - 950 nm 79583 Counts
Highest Intensity in Range 700 - 950 nm 2000 Counts Passed
Highest Intensity for D2 Alpha Line (600 - 700 145084 Counts
nm)
Highest Intensity for D2 Alpha Line 2000 Counts Passed
Spectrum Integral 31678325 Done
UV Integral (190 - 349 nm) 3885644 Done

Finished at 7/1/2020 7:34:16 AM

Signals

Agilent Technologies Printed at: 7/1/2020 7:35:07 AM Page 1 of 2


Lab Advisor Diagnostic Result HPLC 1200 (LB-099)

Signature: ........................................................................

Agilent Technologies Printed at: 7/1/2020 7:35:07 AM Page 2 of 2