Professional Documents
Culture Documents
59]
Original Article
units. Central beam alignment was evaluated high tension transformer were measured using PMID: ***
using a collimator and beam alignment test tools. a pressurized ion chamber‑based survey meter. Quick Response Code:
Accuracy of tube accelerating voltage was tested Exposure rate at table top in fluoroscopy mode
for 45–150 kVp. Accuracy of timer was evaluated was also measured using the above mentioned
for set time 0.1–1 sat 60 kVp and 100 mA. Linearity survey meter. Image quality of simulator was
of mA loading were studied for set mA ranges access using a high contrast resolution test tools
from 20 to 320 mA with fixed kVp of 80 and and Leeds tools.
1056 Journal of Cancer Research and Therapeutics - April-June 2016 - Volume 12 - Issue 2
[Downloaded free from http://www.cancerjournal.net on Wednesday, August 2, 2017, IP: 179.61.164.59]
RESULTS AND DISCUSSION It can be inferred that they are in well agreement with
each other. Figure 3 shows the image of Leeds phantom
Figure 1 shows the image acquired for beam alignment test. acquired from the simulator. The 13th circle for low contrast
It can be inferred from the image that both ball are exactly resolution which is corresponding to the contrast value of
coinciding with each other and beam alignment found to be 0.022 and 10th line pattern for high contrast resolution which
well below the acceptable values of 1.5°. Figure 2 presents the is corresponding to the 1.4 lp/mm are visible in the image.
congruence between set and observed values of delineating Table 2 shows the test result of set and measured kVP values.
wire for field size ranges from 5 × 5 cm2 to 15 × 15 cm2. It can be observed that the variation between measured and
set kVP values are ranges from −0.1 to 1.1 kVp, which is
Table 1: Technical specification of indigenously developed well within the tolerance limit value of ± 5kVp. Table 3
radiotherapy simulator represents the variation between measured and set value
Gantry rotation at focus to axis ±185 degree of the timer. The tolerance values for the variation between
distance 100 cm
Isocenter height above the floor 128
measured and set timing is ± 10%, while the evaluated
Maximumgantry rotation speed 1RPMs variation was not detectable. Table 4 shows the test result
Anticollision logic Yes
FAD 80-120 cm Table 4: Test result for linearity of mA loading station
Collimator rotation ±95 degree
Field size (delineators) 45×45 cm2 Set Measured mR mR/mAs Coefficient Remarks
Independent delineators motions Yes mA (average) of linearity
Field size (shielding jaws) 0.5×0.5 to 50×50 cm2 20.0 74.1 3.703 0.004 Tolerance: Coefficient
Independent jaw motion Yes 50.0 186. 3.727 of linearity <0.1
Image intensifier arm motions ±20 cm (longitudinal), ±20 cm 100.0 373.2 3.732
(lateral), 0-60 cm (radial) 160.0 595.4 3.721 0.0064
Image intensifier auto recenter Yes 320.0 1,175.4 3.673
KV peak 40-125 KVp (radiography mode), Exposure parameters: kVp=80, s=1
40-150 KVp (fluoroscopy mode)
Milliampere 1-20 mA (radiography mode),
Table 5: Test result for linearity of timer
1-800 mA (fluoroscopy mode)
Focal spot size 0.4 mm, 0.8 mm Set time Measured mR/ Coefficient Remarks
Image intensifier size 290 mm (ms) mR (average) mAs of linearity
Couch motions 68-135 cm (vertical), 90cm 280 71.7 1.2807 0.006 Tolerance: Coefficient
(longitudinal), ±20 cm (lateral) 500 129.2 1.2924 of linearity <0.1
Table top Carbon fiber 800 204.4 1.2778
Software control Yes 1,000 255.4 1.2771
DICOM compatibility Yes Exposure parameters: kVp=50, mA=200
DICOM=Digital imaging and communications in medicine, FAD=Focus to axis
distance, KV=Kilo voltage
Table 6: Variation of set and measured mA values
Table 2: Accuracy of tube accelerating voltage (kVp) Set mA Measured mA % error
50.0 49.5 1
Set kVp Measured kVp Difference (kVp) 100.0 99.5 0.480
45 44.2 0.8
160.0 159.7 0.158
50 49 1
200 199.2 0.385
55 53.9 1.1
320 319.9 0.038
60 59.9 0.1
Exposure parameters: kVp=60, s=1
65 64.9 0.1
70 69.6 0.4
80 79.6 0.4
90 89 1
100 100.1 −0.1
110 109.4 0.6
120 119 1
130 129.7 0.3
140 139.3 0.7
150 148.9 1.1
kVp=Kilo voltage
Journal of Cancer Research and Therapeutics - April-June 2016 - Volume 12 - Issue 2 1057
[Downloaded free from http://www.cancerjournal.net on Wednesday, August 2, 2017, IP: 179.61.164.59]
CONCLUSIONS
1058 Journal of Cancer Research and Therapeutics - April-June 2016 - Volume 12 - Issue 2
[Downloaded free from http://www.cancerjournal.net on Wednesday, August 2, 2017, IP: 179.61.164.59]
Bhabha Atomic Research Centre, Mumbai. requirements for the safety of radiotherapy simulators,
3. European Commission. European guidelines on quality criteria for IEC‑ 60601‑2‑29 (1991).
diagnostic radiographic images. EUR 16260ISBN 92‑827‑7284‑5,
Brussels; 1996.
Cite this article as: Kumar R, Kar DC, Sharma SD, Ilpakurty R,
4. Quality assurance handbook. Middleton (WI): RMI; 1992. Subrahmanyam GV. Radiological safety features of indigenously
5. American Association of Physicists in Medicine Medical Physics developed radiotherapy simulator. J Can Res Ther 2016;12:1056-9.
Monograph 4; 1977.
Source of Support: Nil, Conflict of Interest: None declared.
6. International Electrotechnical Commission, Particular
Journal of Cancer Research and Therapeutics - April-June 2016 - Volume 12 - Issue 2 1059