Professional Documents
Culture Documents
4 TECHNICAL 1. Detectors:
SPECIFICATIONS – CT Detectors: a. Data acquisition/recon system capable of
Scan a. Data acquisition system capable of acquiring 32 Slices or more per
acquiring 32 Slices or more per 360°rotation.
360°rotation.
7 TECHNICAL 5. Others:
SPECIFICATIONS – CT Others: 6. iii. deleted
Scan iii. A Barcode reader for entering patient
data from HIS/ RIS must be possible.
8 TECHNICAL Workstation: (At operator console room Workstation: (At operator console room with
SPECIFICATIONS – CT with two monitors: One at operator console one monitors and one at Radiologist chamber)
Scan room & One at Radiologist chamber) 7.
Should be USFDA & CE (Notified body) It Should be USFDA / EUCE (Notified body)
should be mains and battery operated. It should be mains and battery operated.
Should deliver at least 25 shocks with fully Should deliver at least 25 shocks with fully
charged battery with the following charged battery with the following accessories:
accessories: a. Adult and paediatric paddle : 1 each
f. Adult and paediatric paddle : 1 each b. Disposable pads : 10
g. Disposable pads : 10 c. ECG Cable- 5 lead : 1
h. ECG Cable- 5 lead : 1 d. ECG Electrodes : 100
i. ECG Electrodes : 100
12 TECHNICAL Single head Pressure Injector with the Single head Pressure Injector with the
SPECIFICATIONS – CT following: following:
Scan a. Should be CE (Notified body) approved.
b. Flow rate: 0.1-10 ml/sec, Volume- 1 ml to a. Should be EUCE (Notified body)
syringe capacity, programmable pressure approved.
limit of 325 psi with 200 ml disposable b. Flow rate: 0.1-10 ml/sec, Volume- 1 ml
sterile syringes. Syringe heater range 35 deg to syringe capacity, programmable
C+/- 5 degC. pressure limit of 325 psi with 200 ml
c. Should be provided with head mounting disposable sterile syringes. Syringe
device and integral IV pole. heater range 35 deg C+/- 5 degC.
d. 100 no‘s syringes with tubing‘s to be c. Should be provided with head mounting
provide with the machine. device and integral IV pole.
e. Unit will be provided with display d. 100 no’s syringes with tubing’s to be
monitor to provide Pressure Monitor graph, provide with the machine.
Flow Profile, Stop Watch Feature, Scan e. Unit will be provided with display
Display, multiphase capability and protocol
locking capabilities. monitor to provide Pressure Monitor
graph, Flow Profile, Stop Watch Feature,
Scan Display, multiphase capability and
protocol locking capabilities.
9.
13 TECHNICAL 125 kVA Diesel Set with fully Auto start/ 125 kVA Diesel Set with fully Auto start/ off
SPECIFICATIONS – CT off features features
Scan
a. Supply of 125 KVA DG set with Cummins
a. Supply of 125 KVA DG set with /Kirlosker (as per CPEB-2 norms) with
Cummins /Kirlosker /TMTL /Mahendra standard accessories and coupled with
/Yamaha / Honda or equivalent (as per STAMFORD make or equivalent standard
CPEB-2 norms) with standard accessories design alternator rated at 125 KVA,
and coupled with STAMFORD make or 0.8p.f(lag),415V, 3 phase,4 wire mounted on a
equivalent standard design alternator rated at common base iron frame, suitable rated battery
125 KVA, 0.8p.f(lag),415V, 3 phase,4 wire with leads, silencer, fuel tank, set of AVM
mounted on a common base iron frame, pads, composite type acoustic enclosure
suitable rated battery with leads, silencer, including transport to site and etc
fuel tank, set of AVM pads, composite type
acoustic enclosure including transport to site
and etc.
16 GCC 16.1 of special Payment shall be made in Indian Payment shall be made in Indian Rupees
condition of contract Rupees in the following manner: in the following manner:
(a) (i) Advance Payment: Nil (a) (i) Advance Payment: upto 10% of the
unit price against BG of equal amount
Repayment of advance:
Advance payment made will be deducted
from the first invoice.
Sd/-
(SMTI.AMENLA JAMIR)
Senior Accounts Officers
Directorate of Health & Family Welfare
Kohima, Nagaland