Professional Documents
Culture Documents
1
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.
D 1179 – 04
sufficiently high purity to permit its use without adversely
affecting the precision and bias of the test method. Type II
water was specified at the time of round robin testing of this
test method.
6. Sampling
6.1 Collect the sample in accordance with Practice D 1066,
Specification D 1192, or Practice D 3370, as applicable.
7. Scope
7.1 This test method is applicable to the accurate determi-
nation of fluoride ion in water, including most wastewaters.
This test method may not be applicable to concentrated brines
and oily wastes.
7.2 This test method was tested on reagent water and
wastewater. It is the user’s responsibility to ensure the validity
of this test method for waters of untested matrices.
A—Heating mantle (quartz).
8. Summary of Test Method B—Round-bottom flask, 1000 mL.
8.1 The fluoride is distilled as hydrofluosilicic acid and is C—Adapter with thermometer opening.
D—Thermometer, 200°C.
determined by the ion selective electrode. E—Connecting tube.
F—Graham condenser, 300 mm.
9. Interferences G—Vessel, calibrated at 300 mL.
9.1 In sample distillation, interferences may be experienced FIG. 1 Distillation Assembly for Fluoride Isolation
due to the following factors.
9.1.1 Aluminum in excess of 300 mg/L and silicon dioxide
as colloidal silica in excess of 400 mg/L will hold up in the 11.2 Sodium Arsenite Solution (2 g/L)—Dissolve 2 g of
condenser to a certain extent, causing low results and acting as sodium arsenite (NaAsO 2) in water and dilute to 1 L.
a positive interference for subsequent samples of lower fluoride 11.3 Sodium Fluoride Solution, Standard (1.0 mL = 0.01
content. In these cases, the condenser should be flushed with mg F)—Dissolve 0.2210 g of sodium fluoride (NaF) in water
300 to 400 mL of water and the washwater added to the and dilute to 1.0 L. Dilute 100 mL of this solution to 1.0 L with
distillate. The distillate may then be diluted to 1.0 L. If the water. Store in borosilicate glass or polyethylene.
analyst prefers, a smaller sample aliquot diluted to 300 mL may 11.4 Sulfuric Acid (H2SO 4), Concentrated (sp gr 1:84).
be selected for distillation.
9.1.2 Sea water, brines, and generally samples of dissolved 12. Procedure
solids in excess of 2500 mg/L will cause bumping in the 12.1 Distillation:
distillation flask. Dilution of the sample with fluoride-free 12.1.1 Place 400 mL of water in the distilling flask and add
water to a lesser-dissolved solids concentration is an effective 200 mL of concentrated H2SO4(sp gr 1.84). Observe the usual
remedy to bumping. precautions while mixing the H2SO4 by slow addition of the
9.1.3 Samples containing oily matter which may result in a acid accompanied by constant swirling. Add sufficient boiling
two-phase distillate, an emulsion, or anything other than a clear stones6 and assemble the apparatus as shown in Fig. 1. Heat the
distillate may prevent accurate measurement of fluoride. Such solution in the flask, preferably with an electric heating mantle,
samples should be extracted initially with a suitable solvent until the temperature of the contents reaches exactly 180°C. (A
(such as ether, chloroform, benzene, and similar solvents) to quartz heating mantle is preferred in order to reach the required
remove the oily material, and then warmed on a steam bath to 180°C in a minimum time.) While heating, the tip of the
remove traces of the added solvent. thermometer must extend below the level of the liquid in the
flask. Discard the distillate. The procedure, to this step, serves
10. Apparatus to adjust the acid-water ratio for subsequent distillations.
10.1 Distillation Assembly—Glassware consisting of a 1-L, 12.1.2 Caution—Cool the acid-water mixture to below
round bottom, borosilicate boiling flask, an adapter with a 100°C, slowly add 300 mL of sample, and mix thoroughly
thermometer opening, a connecting tube, a condenser, and a before heating. Distill as described in 12.1.1, until the tempera-
thermometer reading to 200°C, assembled as shown in Fig. 1. ture reaches 180°C. Do not allow the temperature to exceed
Standard-taper or spherical ground glass joints shall be used 180°C; excessive carryover of sulfate occurs at temperatures of
throughout the apparatus.
11. Reagents 6
Glass beads must be of a soft glass (rather than borosilicate). Use about 12
11.1 Silver Sulfate (Ag2SO4), powder. beads. Soft beads will provide silica to the fluoride and protect the distillation flask.
2
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.
D 1179 – 04
TABLE 1 Determination of Precision—Final Statistics for Test TABLE 2 Determination of Bias for Test Method A
Method A
NOTE—Bias of Test Method A was determined from round robin data
NOTE—Precision of Test Method A was determined from round robin using distillation with ion selective electrode finish.
data using distillation with ion selective electrode finish. Amount Mean re- Statistically
Amount added, mg/L 0.150 0.560 0.840 2.600 added covery Bias % Bias significant
mg/L mg/L at 5 % level
Reagent Water
Concentration, x 0.147 0.558 0.818 A 2.520 Reagent Water
St 0.033 0.053 0.034 A 0.099 0.150 0.147 − 0.003 −2.0 % no
So 0.013 0.017 0.004 A 0.031 0.560 0.558 −0.002 −0.4 % no
Wastewater 0.840 0.818 − 0.022 −2.6 % yes
Concentration, x A 0.126 0.505 0.771 2.454 A 2.600 2.520 − 0.080 −3.1 % yes
St 0.048 0.068 0.092 0.070 A Wastewater
So 0.018 0.013 0.017 0.030 A 0.150 0.126 − 0.024 − 16.0 % no
A
0.560 0.505 − 0.055 −9.8 % yes
Calculated with outlier point removed from data base. 0.840 0.771 −0.069 −8.2 % yes
2.600 2.454 − 0.146 −5.7 % yes
3
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.
D 1179 – 04
TABLE 4 Determination of Precision Without Distillation using TABLE 5 Determination of Bias Without Distillation (Buffer A)
Buffers Amount Added, Amount Found,
6 Bias 6 % Bias
Buffer A Buffer B mg/L mg/L
19.1.1 For unusual waters containing high concentrations of 22. Precision and Bias 9
fluorides, the range of standards may be expanded up to 1000 22.1 The precision of this test method, using the buffer
mg/L, if necessary. solution described in 18.1 (see Table 4), was tested by 111
19.2 Pipet 50 mL of each standard into a 150-mL beaker. laboratories doing single analysis and found to be 0.030 at a
Using a pipet, add 50 mL of buffer. Mix each solution well concentration of 0.85 mg/L of fluoride.9
using a magnetic stirrer.
19.3 Calibration of pH Meter—Immerse the pH electrodes
in each standard solution, starting with the lowest concentra- 9
Supporting data have been filed at ASTM International Headquarters and may
tion, and measure the developed potential while mixing. The be obtained by requesting Research Report RR: D19–161.
4
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.
D 1179 – 04
TABLE 6 Determination of Bias Without Distillation (Buffer B) 23.3.3 Calculate the mean and standard deviation of the
Amount Added, Amount Found,
6 Bias 6 % Bias
seven values and compare to the acceptable ranges of bias in
mg/L mg/L Tables 1-6. This study should be repeated until the recoveries
0.25 0.257 + 0.007 + 2.8 are within the limits given in Tables 1-6. If an amount other
2.5 2.486 − 0.01 − 0.56
25 24.98 −0.02 − 0.08
than the recommended amount is used, refer to Practice D5847
250 245.75 − 4.25 −1.7 for information on applying the F test and t test in evaluating
the acceptability of the mean and standard deviation.
23.4 Laboratory Control Sample (LCS)
23.4.1 To ensure that the test method is in control, analyze
22.2 The precision of this test method (Table 5), using the a LCS containing a known amount of fluoride with each batch
buffer solution A described in 18.2, was tested by seven or 10 samples. If large numbers of samples are analyzed in the
operators in five laboratories doing three replicates of four batch, analyze the LCS after every 10 samples. The LCS must
concentrations. Each concentration was analyzed in the pres- be taken through all of the steps of the analytical method
ence of 20 mg/L of aluminum. The results were from analysis including sample preservation and pretreatment. The result
in reagent waters. It is the responsibility of the user to test the obtained for the LCS shall fall within 6 15 % of the known
application to other matrices. amount.
22.3 The precision of this test method (Table 6), using the 23.4.2 If the result is not within these limits, analysis of
buffer solution B described in 18.3 was tested by six operators samples is halted until the problem is corrected, and either all
in four laboratories doing three replicates of four concentra- the samples in the batch must be reanalyzed, or the results must
tions. Each concentration was analyzed in the presence of 20 be qualified with an indication that they do not fall within the
mg/L of aluminum. The results were from the analysis of performance criteria of the test method.
reagent water. It is the responsibility of the user to test the 23.5 Method Blank
application to other matrices. 23.5.1 Analyze a reagent water test blank with each batch.
22.4 This precision and bias statement conforms to Practice The amount of fluoride found in the blank should be less than
D 2777 – 77, which was in place at the time of round robin the analytical reporting limit. If the amount of fluoride is found
testing. Under the allowances made in 1.4 of D 2777–98, these above this level, analysis of samples is halted until the
precision and bias data do meet existing requirements for contamination is eliminated, and a blank shows no contamina-
interlaboratory studies of Committee D19 test methods. tion at or above this level, or the results must be qualified with
an indication that they do not fall within the performance
23. Quality Control criteria of the test method.
23.1 In order to be certain that analytical values obtained 23.6 Matrix Spike (MS)
using these test methods are valid and accurate within the 23.6.1 Fluoride is not an analyte that can be feasibly spiked
confidence limits of the test, the following QC procedures must into samples.
be followed when analyzing fluoride. 23.7 Duplicate
23.2 Calibration and Calibration Verification 23.7.1 To check the precision of sample analyses, analyze a
23.2.1 Either calibrate the pH meter using pH 4 and pH 7 sample in duplicate with each batch. The value obtained must
solutions or calibrate the selective-ion meter using 3 standards; fall within the control limits established by the laboratory.
or as recommended by the manufacturer. 23.7.2 Calculate the standard deviation of the duplicate
23.2.2 Verify instrument calibration after standardization by values and compare to the precision in the collaborative study
analyzing a sample with a known amount of the fluoride. The using an F test. Refer to 6.4.4 of Practice D 5847 for
amount of the sample should fall within 6 15 % of the known information on applying the F test.
concentration. 23.7.3 If the result exceeds the precision limit, the batch
23.2.3 If calibration cannot be verified, recalibrate the must be reanalyzed or the results must be qualified with an
instrument. indication that they do not fall within the performance criteria
23.3 Initial Demonstration of Laboratory Capability of the test method.
23.3.1 If a laboratory has not performed the test before, or if 23.8 Independent Reference Material (IRM)—
there has been a major change in the measurement system, for 23.8.1 In order to verify the quantitative value produced by
example, new analyst, new instrument, and so forth, a precision the test method, analyze an Independent Reference Material
and bias study must be performed to demonstrate laboratory (IRM) submitted as a regular sample (if practical) to the
capability. laboratory at least once per quarter. The amount of the IRM
23.3.2 Analyze seven replicates of a known solution pre- should be in the analytical range for the method chosen. The
pared from an Independent Reference Material containing a value obtained must fall within the control limits established by
known amount of fluoride. Each replicate must be taken the laboratory.
through the complete analytical test method including any
sample preservation and pretreatment steps. The replicates may 24. Keywords
be interspersed with samples. 24.1 analysis; electrode; fluoride; water
5
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.
D 1179 – 04
APPENDIX
(Nonmandatory Information)
X1.1 SPADNS Procedure, Photometric precision only if it is known that the concentration of certain
X1.1.1 This test method was discontinued in 1988. The test interfering substances will not cause a 0.1 mg/L error at 1.0 mg
method may be found in its entirety in the 1988 Annual Book F/L.
of ASTM Standards, Vol 11.01. X1.1.4 This test method was discontinued and replaced by a
X1.1.2 This test method is applicable to the accurate deter- distillation method which determines fluoride content in water
mination of fluoride ion in water, including most wastewaters. by the selective ion electrode. The SPADNS photometric
This test method may not be applicable to certain exceptions, procedure was discontinued because there were insufficient
such as concentrated brines and oily wastes. laboratories interested in participating in another collaborative
X1.1.3 The fluoride is distilled as hydrofluosilicic acid and study to obtain the necessary precision and bias as required by
is determined photometrically by its bleaching effect upon the Practice D 2777.
SPADNS dye. Distillation may be omitted without loss of
SUMMARY OF CHANGES
Committee D19 has identified the location of selected changes to this standard since the last issue
(D 1179 – 99) that may impact the use of this standard.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
6
Copyright by ASTM Int'l (all rights reserved); Fri Apr 17 15:39:36 EDT 2009
Downloaded/printed by
Universadad Santiago de Chile pursuant to License Agreement. No further reproductions authorized.