User Guide

Digital scale
Type KPZ 2-02-4
Table of Contents
CE Certificate 3
Technical Certificate of scale 4
After-Sales Service and Warranty 5

General Information
Installation 6
Power Supply 6
Precautions Measures 6
Avoid contact with liquids 6

Optimum Operation Conditions 7


Instructions to avoid damage 7
Handling of scale 7

Operating of KPZ 2-02-4 8


Positioning of Load 8
Indicator self-test 8
Warm-up period 8
Control of accuracy 8
Reasons for possible incorrect weighing 8
Maintenance and Service 8
Storage for prolonged periods of time 8

The KPZ 2-02-4 9


Unpacking 9
Technical data on KPZ 2-02-4 9

page 2
CERTIFICATE
We hereby certify, that the scale of type:

KPZ 2-02-4

Corresponds to the requirements of the Guidelines 89/336/EG


with the application of EN – 55011 (1991) Scale A and EN – 45501
(1992).

Klaus-Peter Zander GmbH

page 3
CERTIFICATE
Manuf. No.: ...........

KPZ 2-02-4

In addition we hereby certify, that the above scal e has been


checked by us before delivery for all operational parameters
and that the scale is within the following tolerance range for
medium-scale division accuracy.

Testing of: Tolerance

Sensitivity ½d

Variation 1d

Fault under 50% of max. load 1d

Fault under max. load 2d

Klaus-Peter Zander GmbH

page 4
CAUTION !!!
Please read carefully the following Operating Instructions before commissioning this
scale. By careful reading of these Operating Instructions you will avoid any possible
damage caused by incorrect operation. In case of any further questions, please contact
your scale supplier or regional sales office.

All indicators of electronic scales require a warm up period from a cold start to stabilise
the internal circuitry. It may happen, that the indicator will appear unstable at initial
switch-on or that the display shows a weight without a load on the scale.

After reaching the correct operating temperature after approx. 2-3 minutes, push ZERO
and the scale will fully stabilise ready for normal weighing.

After-Sales Service and Warranty


KPZ confirms, that any scale produced and supplied by KPZ is free from defect with
regard to material and workmanship. Any warranty claims are restricted to those
components of the units, which prove defective in material and workmanship under
standard and professional use and under correct maintenance conditions. The period of
warranty is 6 months after delivery to the purchaser, provided the note of defect from the
purchaser to KPZ is effected immediately, together with sufficient proof. The respective
component/unit must be returned to KPZ postage and freight prepaid, so that the
defective component/unit can be inspected by KPZ.

The warranty does not cover any damage resulting from excessive loading or other
incorrect use, from non-conformity with the operating Instructions or any accidental
damage to the units special equipment or other accessories, which have not been
installed by KPZ or by a person authorised by KPZ or to which any modifications have
been made to change the specified purpose of use.

This warranty is automatically invalidated, if any persons not authorised by KPZ has
carried out any modifications, interventions or other repairs etc. to the system.

page 5
GENERAL INSTRUCTIONS

Installation
Install scale in a dry, well-lit and vibration-free location. Use the bubble level and
adjusting legs, if supplied, to properly level the KPZ 2-02-4, before commencing the
operation of the KPZ 2-02-4. Strictly follow the safety instructions specified with regard
to the power supply.

Power Supply
Check that the voltage requirements for the scale or for the charging device/power pack
are met by the on-site power supply (230/240V). When disconnecting the device, always
remove at the plug, never remove by pulling on the power cable.
Always use original KPZ cable or KPZ power pack/charging device. The use of any
other device may cause considerable damage to the electronics.
If the charging device/power pack is damaged by accident or if any other defect exists,
please check the output. If the output is affected in any way, immediately order a new
KPZ cable or KPZ charging device/power pack. Any damage to the charging
device/power pack may result in a reduction or failure of the unit, resulting in most cases
in damage to the scale.

Avoid using the same power socket to connect other devices. Try to lay the cable, so
that it can not be stepped on or pulled.
If the scale is equipped with a rechargeable battery, it is recommended to fully charge
the battery after the installation and before the first start-up of the scale. Connect the
KPZ 2-02-4 to the on-site power source (230/240V) and fully recharge the rechargeable
battery, with the indicator switched off.
Always keep the rechargeable battery at a high charging level by recharging it as often
as possible (daily); this will prolong the service life of the battery.

Precaution Measures
Do not try to eliminate any faults or to carry out any repairs, which are not specified in
these Instructions.
Under no circumstances, open the housing of the scale, thus removing the protection of
the sensitive electronic components. This will immediately invalidate the warranty.

CAUTION !
Do not remove the platform from the scale. Any material falling into the gap between
scale housing and platform support may affect the performance of the weighing sensor
and / or the internal electronics.

Avoid any Contact with Liquids


Only clean the scale as described in these operating Instructions. If any water or other
liquid penetrates into the housing, remove the mains plug and have unit checked by
qualified maintenance staff before re-commencing operation.

page 6
Optimum Operating Conditions
In order to achieve the most accurate results from your weighing
scale the unit should be used at location having the following conditions:
The base support must be level and horizontal
The base support should be stable and not be subjected to vibrations
Avoid any continuous, direct sunlight
Do not use in areas where corrosion gasses are present
Ensure a dust-free environment if possible
Ambient temperature –10°C to 40°C
Relative air humidity 40 to 70% (do not install close to air humidifiers!)
Do not use in the proximity of other electronic devices to avoid any electrical interference
Do not install close to heating units and outlet openings of air-conditioning units to avoid
any large temperature variations around the scale.

Instructions to Avoid Damage


The KPZ scale is a high-quality accurate scale. Thus, avoid any extreme applications.
When cleaning the unit, make sure, that water cannot penetrate into the electronic
housing.
This scale should not be subjected to extreme temperature influences below minus 10°C
or above plus 40°C. The load cell and also the scale may be damaged, if any impact,
overload or excessive point loads are exerted onto the scale. Contact KPZ in case of
extreme operating requirements.
Avoid any unprofessional handling of the scale and always consider, that the unit is a
high-quality measuring instrument.

Please note:
Only carry out those measures for maintenance and fault elimination, which are
described in this Manual.

Make sure, that the indicator is not subjected to water or moisture to avoid any risk of
fire or electrical short circuit.

Do not open the indicator housing to avoid risk of an electric shock!

Handling of Scale
Despite the sturdy design features of the scale, each unit contains sensitive electronic
components, which could be damaged through careless handling of the scale. Use both
hands to transport the scale, with at least one hand underneath the scale. (This will be
necessary, if the power supply, for example, is not close enough to the operating site
when recharging the rechargeable battery).

page 7
OPERATION OF KPZ 2-02-4

Positioning of Load
Always place the load carefully and central onto platform. Any sudden vibrations may
damage the scale sensor, so that it cannot be repaired. In general, the service life of the
sensor is prolonged by avoiding any vibrations. In order to achieve permanent, high
accuracy weighing results, do not leave loads on the platform for a long period of time
(e.g. overnight), this may deteriorate the performance of the weighing sensor.

Self-Test of Indicator
At initial switch-on, the scale will immediately activate its self-test. This test is visible via
the external countdown on the display
from –8888- to –0000-. Make sure, that all symbols are completely visible on the display
to avoid any subsequent recording of false measurement data. As soon as the Zero-
Balance status is displayed, the scale is ready for operation. If the scale does not
indicate “0”, then set the display to “0” again by using the ZERO key (0)
(if installed.)

Warm-up Stage
It is strongly recommended to allow for a warm-up period before using the scale.

Control of Accuracy
This scale has no mechanical wear parts. When using the scale properly and for the
intended purpose, there is normally no reduction in performance. However a prolonged
period of operation, external impacts or other various reasons may result in a deviation
of the accuracy. For your own safety, we recommend you make accuracy checks at
regular intervals by using fully calibrated weights.

Reasons for Possible Incorrect Weighing


The scale may display incorrect readings if:
◼ The power supply is not stable
◼ A load is put onto the scale after switching on during the self-test
◼ The platform is not freely movable
◼ The connection between indicator and platform is damaged

Maintenance and Servicing


Warning: Never use any acetone or any other ethereal solvents, such as thinner or
alcohol to clean the scale.
Daily Maintenance: Clean platform with soft, moist cloth and/or mild detergent.

Storage for Prolonged Periods of Time


Fully recharge the rechargeable battery (if supplied). Make sure, that no load is on
platform. Clean scale and store in dry, dust-free area. Recharge battery from time to
time, as the battery will discharge slowly during storage.

page 8
THE KPZ 2-02-4
Unpacking
You will find the following contents, when opening the shipment box:

Operating Instructions
The KPZ 2-02-4 scale
Platform
Charging set

Note:
Do not throw away box and packaging material. It might be useful for future
transportation/storage of the scale. This scale is fully operational and calibrated and
should not require any re-adjustment by the user.

Technical data

High Precision:
Type KPZ 2-02-4/30000 KPZ 2-02-4/6000
Capacity 30kg 6kg
Division 1g 0,2g
Tare range 100%
Power supply Re-chargeable battery
Mains 230VAC 50/60Hz
Operating Temp. 0° to 40°C
Article-no. 2033300000 2033060000

Standard:
Type KPZ 2-02-4/1500 KPZ 2-02-4/7500 KPZ 2-02-4/15k KPZ 2-02-4/30k
Capacity 1,5kg 7,5kg 15kg 30kg
Division 0,1g 0,5g 1g 2g
Tare range 100%
Power supply Re-chargeable battery / Mains 230VAC 50/60Hz
Operating Temp. 0° to 40°C
Article-no. 2034015000 2034075000 2034150000 203430000

page 9
Description of display
Scale is at zero point

Net Scale displays the nett weight.

Symbol appears, if the battery requires re-charging.

HIGH Symbol appears, if the indicated value is higher than the keyed in range

OK Symbol appears, if the indicated value is inside the keyed in range

LOW Symbol appears, if the indicated value is lower than the keyed in range

~ This symbol marked by an  in the display, indicates the weight is stable

PT This symbol marked by an  in the display, indicates if a pre-tare value was keyed in.

M+ This symbol marked by an  in the display, indicates if a value was added to the memory

 This symbol marked by an  in the display, indicates if the piece weight too small. The counting
function is still possible, but inaccuracies may be shown.

+ This symbol marked by an  in the display, indicates if the keyed in piece number is too small.
The counting function is still possible, but inaccuracies may be shown.

page 10
Description of keys

Use this key to call up a pre-tare value.


Use this key to Zero the scale.
Use this key to carry out TARE and ZERO.

Use this key to confirm entered quantity.


Use this key to recall the memory value ( e.g. addition total, pre-tare, peak value )

CE Use this key to clear the memory value ( e.g. addition total, pre-tare, peak value)
Pcs Use this key to select the reference piece counting number ( 10, 20, 50, 100 ).
M+ Use this key to add the piece or weight total into memory.
With key function 1. You increase the value shown at the display.
With key function 2. You select the backlight function ( 3 options )
With key function 1. You select the unit of measurement required.
With key function 2 you change from "kg/LB" to count mode.
Use this key to set up the peak value.

With key function 1. You transfer the date to RS 232 output. ( 4 options )
With key function 2. Option selection for RS 232 is confirmed.

page 11
On / Off Switch
The on/off switch for the scale is at the right side. It is marked with a 0 (off) and a 1 (on).

Zero
The scales should after switching on display 0.0, if not re-set the zero by pressing the button on the
indicator.

Tare
There are two possibilities to tare a weight (e.g. empty container), so that only the weight of the contents
are shown.
1) Weighing out
2) Key-in (pretare)

1) Weighing out
Place the container (e.g. empty dish) on the platform, wait until this value is stable and press the key .

2) Key in (Pre-tare)
Handling Display
You have a dish (e.g. 10g) with material inside (e.g. 90g) on the scale. 100,0 g
press key , the first digit is flashing 0 - - - -,P
Key in the tare value with the keys and (e.g. 10g). Note: the decimal point!
Press key 3 times 0000 -,P
Press key 0001 -,P
Press key 2 times 00010,0
Press key 90,0 g

Counting

Handling Display
Put the reference number (10, 20, 50 or 100) on the platform (e.g. 20 pieces @ 5g) 100,0 g
Press key Pcs C 10
Press key Pcs C 20
Press key to enter the chosen value 20 Pcs
The scale is ready to count
100,0 g
Press key to show the weight
Press key to show the count, the piece weight will be memorised. 20 Pcs

page 12
Memory
You can add weight-values or count-values to the memory. But only, either weight-values or count-values.

Notice!
An accidental double addition can only be possible if the weight changes plus or minus on the scales.

Addition
Handling Display
Put a weight on the platform (e.g. 150g) 150,0 g
Press key M+ . In the display will appear the following message „addition successful, then Add
the number of the addition then - 0001 -
the accumulative total sum 150,0 g
Remove the load 0,0 g
Put the second weight on the platform (e.g. 250g) 250,0 g
Press key M+. Add
- 0002 -
400,0 g

Recall Memory
Handling Display
Press key . In the display appears the additions number - 0002 -
then the sum 400 g
Following this the scale is ready to weight 0,0 g

Clear memory
Handling Display
Press key . - 0002 -
While the number of additions is shown, press key CE 0,0 g
The memory is cleared and the scale is ready to weight 0,0 g

page 13
Check function
There are 2 check points (high value and low value), providing 3 options:
The value above the maximum set point HIGH
The value between the maximum & minimum set points OK
The value below the minimum set point LOW

If both values are keyed in, the display indicates the current area. It is possible to choose in which area
the signal will be shown and the alarm operate. For this two code-numbers must keyed in.

Alarm Display symbol


1. digit 0 If display stable If display stable
1 If display stable If display stable
2 If display stable or unstable If display stable or unstable
2. digit 0 No
1 Within set range OK
2 At set value LOW or HIGH

Handling Display
0 - - - -,L
Press key , the first digit is flashing
Key in the low value with the keys and (e.g. 100g), Pay attention to the decimal point
Press key 2 times 000 - -,L
Press key 001 - -,L
Press key 3 times 00100,0
Press key 0 - - - -,H
Key in the high value with the keys and (e.g. 500g), Pay attention to the decimal point
Press key 2 times 000 - -,H
Press key 5 times 005 - -,H
Press key 3 times 00500,0
Press key Print 0-b
Key in the code-numbers you want with the keys and (e.g. audible signal and
symbols in the area OK , not necessary for the scale to be stable -> 21b)
Press key 2 times 2-b
Press key 20 b
Press key 21 b
Now the scale is ready to weight 0,0 g

Clear check point


Handling Display
0 - - - -,L
Press key , the fist digit is flashing
Press key L 100,0 g
Press key CE H 500,0 g
Press key CE 0,0 g
Now the scale is ready to weight

page 14
Backlight
The unit has 3 variations of backlight:

- no backlight bl. OFF


- permanent backlight bl. ON
- backlight, if the load is more than 10d is (1d = 1 division step) bl. AUtO
or if a key is pressed.
The backlight will switch off automatically after 5 second, if the load is less than 10d.

Press key until the option you want is shown in the display.

Change of weight measurement


If the scale is in weighing mode press the key to switch between kg and lb.

Serial interface RS 232 (Option)

Baud rate 2400


Data Bits 7
Parity 1 even
Stop Bit 1
Code ASCII

Weight set-up:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Head 1 , Head 2 , DATA measure CR LF
ment

Piece count set-up:


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Head 1 , Head 2 , DATA measurement CR LF

Head 1: OL = overload or underload


ST = stable
US = unstable

Head 2: NT = Nett weight


GS = Gross weight

Data Bits special characters Measurement

- (minus) = 2D (hex) kg = 6B, 67 (hex)


(space) = 20 (hex) LB = 6C, 62 (hex)
. (point) = 2E (hex) PCS = 50, 43, 53 (hex)

page 15
Set-up of data-transfer
Handling Display
Press and hold key and switch on the scale. rnP 0
Release key after the display shows „rnP 0“
Select with the key the mode you require
0 = no transfer
1 = one transfer, if a load on the platform and the display is stable
2 = permanent transfer
3 = transfer of the simple format by pressing key M+
4 = transfer of complete format by pressing key M+

(e.g. transfer of simple format by pressing key M+)


press key 3 times rnP 3
Press key to enter the chosen mode 0,0 g
The scale caries out a selftest then the set-up is ready

Press key: sample print-out


simple format
M+
S/N WT/kg
---------------
0001 100,0
M+ 0002 200,0
M+ 0003 300,0
---------------
0003 600,0

Press key: sample print-out


complete format
M+
TICKET NO.0001
G 100,0g G = gross weight
T 0,0g T = tare weight
N 100,0g N = Nett weight
M+
TICKET NO.0002
G 200,0g
T 0,0g
N 200,0g
M+
TICKET NO.0001
G 300,0g
T 0,0g
N 300,0g

TOTAL NUMBER
OF TICKETS 0003
TOTAL
NET 600,0g

page 16
Parameter Adjustment:
Parameter-List:
F0: Set Calibration-weight and do to adjust
F1: Set Capacity and Division (absolutly Zeropoint)
F2: Adjust the filters
F3: Adjust the automatical Zeroset
F4: Adjust the set point (like operation manual)
F5: RS 232 Interface format setting (like operation manual)
F6: Left Adjustment of parameter
F7: Indicate the intern value
F8: Adjust holdpoint for top level
F9: Adjust of factory-set

The numerical value is setting how below:


Numerical will go up
The Curser will go one Stepp to right
Confirm the adjustment-value
Delete the adjustment-value
(in calibration routine without function)

Select the Parameters:


Confirm Indicate
Key 0 push and hold during the scale switch on F0
Key  push for Stepp up adjusts F1... F2... F3... F1
Key push for stepp down adjusts F9 ... F8... F7... F9

F0 set calibration weight and adjust:


Confirm Indicate
Remove all weight from the platform F0
Key push and hold during the scale switch on
Key push ZEro
Key push 0 - - - -.C
Put in the weight, witch will for calibration routine. Please note, to use the decimal place. 1500.0
as sample:
Calibration weight: 30kg
Division: 2g
Adjustment value 00030,0
 push for Stepp up adjusts
The Curser will go one Stepp to right (S.3)
Put the calibration weight on the platform and wait until the scale is stabil. 1500.0
Key  push WAit
When 0030.0 indicate, the calibration routine is finish 1500.0

page 17
page 18
page 19
KPZ – Váhy, s.r.o.
Libošovice 76
CZ 507 44 Libošovice

Tel.: 493 571 111-3


e-mail: kpz@kpz-vahy.cz
www.kpzvahy.cz

www.kpzv ahy -esho p.cz

Váhy KPZ – KVALITNÍ, ROBUSTNÍ a LEVNÉ ! !