Operation
Manual
1. Presentation-----------------------------------------------------P1
2. Main Function and Technical Index-----------------------P1-3
2.1 Main function------------------------------------------P1-2
2.2 Technical index-----------------------------------------P2-3
3. Construction Illustration and Function--------------------P3-5
3.1 Construction illustration -------------------------------P3-4
3.2 Button function---------------------------------------------P5
3.3 Menu and remote controller------------------------------P5
4. The Operations------------------------------------------------P6-11
4.1 Turn on---------------------------------------------------P6
4.2 Turn off---------------------------------------------------P6-
4.3 Menu instruction--------------------------------------P6-11
5. Installation-------------------------------------------------P17-19
5.1 Installation position----------------------------------P17-18
5.2 Wire connection-------------------------------------P18-19
6. Sensor Module Replacement-------------------------------P20
7. Normal failure and solving----------------------------------P20
8. The spare parts------------------------------------------------P21
9. Attentions-------------------------------------------------- P21
1. Presentation-
The module gas detector( detector as below), it is a fixed
detector continuously monitoring the combustible gas,
toxic gas as well as oxygen. The detector works by
natural diffusion, adopts the imported electro-chemical
sensor for toxic gas and oxygen, catalytic sensor for
combustible gas, IR sensor and PID sensor, too. It has
good reproducibility and sensitivity, disturbing-proof
against temperature and humidity, long life-span and easy
operation.
The detector is with LED display screen showing the real
gas leakage concentration and once the gas leakage
concentration exceeds the preset alarm point (threshold) it will
start the acoustic-optical alarm on the central unit and drive the
external equipment like exhaust fans.
The signal output is standard 4-20mA or RS485. It is widely
used in refineries, petrol, chemistry, metallurgy, artifice, gas
transportation, biological medicine and water procession etc.
4.1 Turn on
The detector enters into normal detecting mode after it
is electrified and the 10s self-test countdown
finishes.When the read keeps steady, the read on the main
display window is the detected gas concentration.(It takes
different time for different gases to keep the read steady,
normal period is 2-5 minutes).
The detector is set with the standard value and quality
checked OK when it is ex-factory. If there is no special
requirement of the setting value, the detector can work
directly without any operation on the settings. Range
calibration is forbidden without the standard calibration
gas.
4.2 Turn off
In detecting mode, The detector turns off by powering
off.
4.3 Menu Instruction
In normal detecting mode, press confirmation button
“ ”3seconds or press the “menu”on the remote controller,
the display screen will show “ ”. Input the password
“1111” and then press confirmation button “ ”.The menu
items showing as below:
1. escape
2. baud rate 600 1200 2400 4800 9600
3. address code 001~127
4. Range calibration
5. Zero calibration
6. Hysteresis error of high alarm
7. High Alarm
8. Hysteresis error of low alarm
9. --Low alarm
In normal detecting mode, press UP button “ ”3seconds
or press the “F1”on the remote controller, the display
screen will show “ ”. Input the password “3088” and
then press confirmation button “ ”.The menu items
showing as below::
10. -escape
11. --reset (recover the ex-factory setting.)
12. The proofread of DA output 20mA.
13. The proofread of DA output 4mA.
4.3.1 Baud rate Set
Upon the menu instruction and choose “ ”baud rate set
menu, then press confirmation button “ ”to baud rate
value “e.g 9600”. Change the baud rate value by UP
and Down button, and confirmation button “ ”to
confirm. When the baud rate value is set successfully,
back to the baud rate menu. Press Up and Down
button till menu appears, finally exit by
confirmation button“ ”.
4.3.2 The Communication Address Set
Choose address set menu and press confirmation
button till the screen display address “e.g 0001”. Change the
communication address by UP and Down button,
and confirmation button “ ”to confirm.
4.3.3 Range Calibration
First, connect the calibration cover and sensor breathi ng
device. Put the standard calibration gas with known
concentration into the device, the calibration gas flux of
350ml---400ml is suggested. Second, press confirmation
button 3seconds or “menu” on the remote controller. It
displays“ ”, input the password “1111”and then press
Confirmation button to menu: choose calibration menu
by Up and Down button, and then press
confirmation button It displays calibration value e.g
“0200”. Revise the calibration value as the standard calibration
gas value by UP and Down button. Press “Confirmation
” button to remember this figure and calibrate it. 15s
countdown displays and the detector returns to normal status.
Note: 1. Please don’t carry the calibration when the detector is
working or there is no calibration gas.
2. The calibration can be repeated till the value steady.
4.3.4 Zero Calibration
Upon the menu instruction, choose Zero calibration menu
“ ” by Up and Down button. Press the
confirmation button and it displays “0000”, again press
the confirmation button , when the 10s countdown on
the screen finishes, the detector returns to the normal
status.
Notes: 1) Zero calibration must be done in the clean air. It
is forbidden to do zero calibration when the target gas
exists in the working place.
2) For O2 or N2 gas detector, the detector should be put
into pure N2 gas environment).
4.3.5 Hysteresis error of high alarm
Upon the menu instruction to choose “ ” Hysteresis
error of high alarm menu by Up and Down button.
Press the confirmation button and it displays
Hysteresis error value of high alarm “e.g 0003”. To revise
the value by Up and Down button. Then press
confirmation button to remember this figure and press
confirmation button to normal status.
4.3.6 High alarm setting
Upon the menu instruction, choose High alarm menu
“ ” by Up and Down button. Press the
confirmation button and it displays high alarm value,
“e.g 0150”. To revise the value by , Up and Down
button. press confirmation button to save this figure and
press confirmation button to normal status.
4.3.7 Hysteresis error of Low alarm
Upon the menu instruction, choose Hysteresis error of low
alarm menu “ ” by Up and Down button. Press
the confirmation button and it displays Hysteresis error
value of low alarm “e.g 0003”. To revise the value by
pressing Up and Down button. press confirmation
button to remember this figure and press confirmation
button to normal status.
4.3.8 Low alarm setting
Upon the menu instruction, choose Low alarm menu “ ”
by Up and Down button. Press the confirmation
button and it displays low alarm value, “e.g 0050”. To
revise the value by Up and Down button. press
confirmation button to save this figure and press
confirmation button to normal status.
4.3.9 Display Adjust
In the normal status, press Up button or “Display” button on
the remote controller, the screen displays the detector
parameter in sequence as below: low alarm value, Hysteresis
error value of low alarm, high alarm value, Hysteresis
error value of high alarm, calibration value, address code,
baud rate value, gas type, range.
10. Attentions
1. Avoid being fallen from high places and vibrated mightily.
2. Keep the detector away from the place where is rapid air
flow. Otherwise it will influence the testing result.
3. Before using this product, please read the following
instruction carefully and adhere to related requirement strictly.
4. It is forbidden to use the detector in the places where is
corrosive gas and other severe conditions with exorbitant or
over-low temperate, high humidity, electromagnetism
environment and strong sunshine.
Please clean the detector with soft cloth in case there is dirt on
the appearance for long usage. Don’t clean with corrosive
solvent or hard materials, which may hurt and break the
appearance of the detector.
Please re-calibrate the sensor once every 6 months. It’s
suggested to test or inspect the detector frequently, so as to
avoid any dust or unexpected painting on the sensor.
Any application and failure not mentioned in this instruction,
please contact us for solving.
The reference chart for gas detector choice
For the gas not covered here, please don’t hesitate to contact
us.