Professional Documents
Culture Documents
Abstract. C-reactive protein (CRP) is a major, acute-phase protein in dogs; however, there is a need for
automated assays to ensure in-time patient monitoring. Three automated immunoturbidimetric assays
(Randox, Thermo, and Wako) developed for human beings were evaluated for their ability to detect canine
CRP, including method validation, evaluation of diagnostic use, and establishment of exploratory reference
intervals. Sera from 36 healthy dogs and 82 diseased dogs were included for method comparison with the
enzyme-linked immunosorbent assay (ELISA; Tridelta) serving as the reference method. A nonparametric
estimate of the 1-sided 95% reference interval was established (n 5 36). Precision study revealed good intra-
assay coefficients of variation (CVs) of 1–10%, 0–9%, and 2–13% for the Randox, Thermo, and Wako assays,
respectively. Interassay CVs were 18%, 24%, and 19% respectively. Because of a low linear range, the Thermo
test was considered unsuitable for use with canine specimens. No significant differences were present between
the results obtained with the Randox and Wako assays with CRP concentrations less than 15 mg/l; however,
median CRP results differed significantly between the Thermo test and the ELISA (P 5 0.03). Bland–Altman
analysis detected a proportional bias of 0.28, 20.59, and 0.61 mg/l for the Randox, Thermo, and Wako assays,
respectively. For all tests, median CRP values were significantly different between healthy dogs and dogs with
neoplasia. The upper limit of the reference intervals were 8.2 and 9.9 mg/l for the Randox and Wako assays,
respectively. In contrast to the Thermo test, the Randox and Wako assays were suitable for detection of
abnormally high canine CRP concentrations; however, improvement of assay precision and evaluation of
accuracy are warranted before their clinical use with canine specimens.
Key words: C-reactive protein; dogs; human; immunoturbidimetry; validation.
determination in dogs, including 2 human tests concentrations of dogs with different diseases to those
obtained from Bayerb,16 and Olympus,c,3 as well as 2 obtained in healthy dogs. In the third part of the study,
canine-specific tests.9,10 Use of a species-specific test exploratory reference intervals were established if all
would be preferable, but unfortunately, one of the previous examinations were acceptable.
tests10 is no longer marketed commercially and the Method validation
other9 is not commercially available.
The Bayer CRP turbidimetric immunoassayb Linearity. Linearity was assessed by serial dilution of
(hereafter, Bayer assay) for human beings has been the following human and canine specimens with markedly
increased CRP concentrations: the human calibration
proven to reliably detect canine CRP,16 but Bayer no
standardh containing a highly abnormal CRP concentra-
longer produces this assay. The same antibody, tion (153 mg/l) and a canine serum sample with high CRP
however, is used in the Randox CRP testd (Randox concentration (.45 mg/l). Both specimens were serially
Laboratories Ltd., personal communication, 2010). diluted with saline (0.9%) to achieve samples with 6
In 2008, the Olympus human immunoturbidimetric different CRP concentrations (undiluted CRP; 1:1; 1:2;
assayc was used to detect CRP in dogs with 1:4; 1:16; saline 0.9%). All samples were analyzed in
hyperadrenocorticism.3 Although excellent correla- triplicate using the 3 human CRP assays.
tion to the canine-specific ELISAa and acceptable Intra-assay and interassay repeatability. A precision
intra-assay and interassay coefficients of variation study evaluated the random error by investigating intra-
(CV) of ,10% were reported for the Olympus assay,3 assay and interassay repeatability (CV). The repeatability
full assay validation was not detailed by the authors of a 20-run intra-assay was assessed by repeated measure-
because that was not the primary aim of their ments of patient serum samples with 3 different CRP
concentrations (high: .45 mg/l; medium: ,15 mg/l; low:
investigation. ,8 mg/l).
In the current study, 3 automated immunoturbidi- The 15 consecutive observations from serum samples
metry CRP assaysd–f originally developed for use with with similar high, medium, and low CRP levels were
human specimens were evaluated for use with canine included for assessment of interassay repeatability. Serum
serum samples. Test evaluation included method samples with 3 different CRP levels (high: .45 mg/l;
validation, assessment of clinical validity (i.e., the medium: ,15 mg/l; low: ,8 mg/l) were prepared with
capacity of the tests to differentiate between dogs specimens obtained from 1 healthy and 4 diseased dogs;
suffering from various diseases and healthy dogs), aliquots were frozen at 280uC until analysis. Only vials
and establishment of exploratory reference intervals needed for each analytical run were thawed to avoid
for the tests. repetitive freeze–thaw cycles, thus minimizing a source of
potential variation. For each level, standard deviation (SD)
Materials and methods was calculated as the square route of the estimated variance
components. In the next step, CV was calculated as (SD/
Study design global mean) 3 100 to evaluate imprecision of the assays.
The study was carried out between August 2007 and Method comparison. Serum samples from 118 dogs (36
April 2008. Three automated CRP assays developed for healthy, 82 diseased) were available for the method
human beings were evaluated in the present study. The first comparison part of the study. Samples were first measured
assay from Randox Laboratoriesd (hereafter, Randox with the 3 automated CRP assays and, afterward, were
assay) consisted of a high-linearity test using an unspeci- frozen at 280uC until ELISA analysis. For ELISA analysis,
fied–anti-human CRP antibody. The second assay from all samples were thawed, thoroughly mixed, and the ELISA
Thermo Clinical Labsystemse (hereafter, Thermo assay) was manually performed by one of the authors (SK) and by
used a goat–anti-human CRP antibody. The third test kit a trained technician as recommended by the manufacturer.
from Wako Chemicalsf (hereafter, Wako assay) used an A standard curve was prepared using serial dilutions of the
ovine–anti-CRP antibody. The manufacturers of the 3 tests calibrator with the standard diluent buffer (both provided
did not clarify whether a monoclonal or a polyclonal in the test kit) according to the manufacturers’ instruction.
antibody was used. All 3 CRP assays were run on an Absorbance was detected at 450 nm with an ELISA reader.i
automated biochemical analyzerg according to the manu-
facturers’ recommendations and were calibrated with test- Clinical validation
specific calibration material. Because no test-specific Eighty-two dogs presented to the Small Animal Clinic
material for internal quality control was available for the (Faculty of Veterinary Sciences, Justus-Liebig-University,
Wako assay, the Randox control was used. Giessen, Germany) with different diseases were included in
The study was divided into 3 parts, beginning with the study to evaluate CRP concentrations in various
method validation. Method validation included evaluation illnesses. Depending on the clinical signs, various diagnos-
of linearity and assay precision and comparison of the new tic tests and procedures were performed to establish the
methods to the reference method (canine CRP ELISAa), final diagnosis. These included biochemical profiles,
which was previously validated.18 In the second part of the radiography, ultrasonography, endocrine tests, microbial
study, clinical validity was tested by comparing CRP tests, bronchoalveolar lavage, analysis of cerebrospinal
546 Klenner, Bauer, Moritz
fluid, cytology, and biopsy. Diseased dogs were assigned to Establishment of the reference intervals. Data of the
the following groups based on their final diagnoses: healthy dogs were described by calculating medians, means,
neoplasia (n 5 16), inflammatory disease (n 5 17), and SDs, and minimums and maximums, as well as by depicting
miscellaneous diseases (n 5 48), those that did not match histograms. If the data were not normally distributed,
either of the previous diagnoses. logarithmic transformation was done.
Automated CRP assays were performed on the day of The 95% 1-sided reference interval was calculated using
sampling; subsequently, the serum samples were frozen at commercial software.m Possible differences between females
280uC until ELISA analysis. The dogs included for and males in the reference range population were investi-
calculation of reference intervals served as controls. gated using the Mann–Whitney U-test.
Clinical validation. A 1-way analysis of variance (AN-
Establishment of reference intervals OVA) and Tukey’s posttest were used for assessment of
Thirty-six clinically healthy, adult (.1 year old) dogs potential differences between the different disease groups
were sampled to establish reference intervals. The dogs when CRP was measured with the 3 assays. For analysis of
comprised the following breeds: 8 Beagles, 3 German the impact of the factors test and disease and the interaction
Shepherd Dogs, 2 Rottweilers, 2 Australian Shepherd between test and disease, a 2-way ANOVA was applied. All
Dogs, 2 Great Danes, 2 German Wirehaired Pointers, 2 tests were performed after logarithmic transformation of
Golden Retrievers, 1 Bernese Mountain Dog, 1 Saint the data.n A P-value of ,0.05 was considered significant.
Bernard, 1 Austrian black and tan hound (Austrian
brandlbracke), 1 Flat-Coated Retriever, 1 Kuvasz, 1
Results
Labrador Retriever, 1 Newfoundland, 1 German spaniel, Method validation
1 Weimaraner, and 5 mixed-breed dogs. Females were
All 3 CRP tests showed excellent linearity (r .
slightly overrepresented with 22 out of 36 dogs (15 intact, 7
spayed), whereas 14 out of 36 were males (9 intact, 5
0.99) within a calibration range of ,10–153 mg/ml for
neutered). The median age of the dogs was 2 years (range: serially diluted human specimens. The Randox and
1–8 years). Dogs were regularly vaccinated and dewormed Wako assays also showed linear calibration graphs
blood donors, staff-owned dogs, or healthy dogs presented with correlation coefficients of r . 0.98 in serially
for routine radiologic examination to screen for hereditary diluted canine samples; however, the Thermo assay
hip or elbow dysplasia at the Clinic for Small Animals had a linear range less than about 8.5 mg/l, with a
(Faculty of Veterinary Medicine, Justus-Liebig University, markedly lower correlation coefficient than the other
Giessen, Germany). Special care was taken in the selection assays (r 5 0.76). As shown in Figure 1, the Randox
of the healthy dogs for study. All dogs were clinically and Wako assays mildly overestimated CRP mea-
healthy and had no recent history of acute illness. As CRP surement in the lower range of concentrations (10–
is a major acute-phase protein, which increases in a wide 30 mg/l). Intra-assay CVs were 1–10%, 0–9%, and 2–
variability of acute inflammatory diseases, traditional
13% for the Randox, Thermo, and Wako assays,
markers of inflammation (i.e., white blood cell count
[WBC], erythrocyte sedimentation rate [ESR], and fibrin-
respectively. Interassay precision showed slightly
ogen) were determined to exclude dogs with underlying higher CVs at 8–18%, 9–24%, and 4–19% for the
inflammation that was not detected by routine clinical Randox, Thermo, and Wako assays, respectively
examination. Therefore, inclusion criteria for study were an (Tables 1, 2).
unremarkable physical examination; lack of increase in Sample volume of the sera available for the method
WBC, ESR, and fibrinogen concentration; and a hemato- comparison experiment was too low to perform
logic profile within reference range. Erythrocyte sedimen- duplicate measurements. After the ELISA analysis,
tation rate was performed via the Westergren methodol- 24 sera had to be excluded from the statistical analysis
ogy.j An automated hematology analyzerk was used for because results exceeded the upper limit of linearity of
hematological analysis. Fibrinogen concentration was the ELISA (.15 mg/l). Because of the low sample
determined by the Clauss method, using a coagulometer.l volume, dilution of the samples and reanalysis with
Statistical analysis
the ELISA was not possible. Thus, 94 samples were
finally included in the statistical analysis. There was
All data were evaluated for normal distribution using the no significant difference between the median results
Anderson–Darling test.m obtained with the Randox assay and the ELISA (P 5
Method validation. Descriptive statistics were per- 0.20) nor between the Wako assay and the ELISA (P
formed to calculate medians, means, SDs, and CVs.
5 0.36). In contrast, a significant difference was
Calculation of the Pearson’s product moment correlation
coefficient was used to evaluate the linearity of the tests. In
detected (P 5 0.03) between the median results
the method comparison experiment, possible differences achieved with the Thermo assay and the ELISA.
between the means of pairs were detected using the The Bland–Altman difference plot revealed a mean
Wilcoxon signed rank test. Additionally, Bland–Altman proportional bias of 0.28 mg/l for the Randox assay,
analysis was used to characterize the bias and to visually 20.59 mg/l for the Thermo assay, and 0.61 mg/l for
evaluate the difference plot.40 the Wako assay. For all tests, the 95% limits of
Automated assays for canine C-reactive protein detection 547
Figure 1. Linearity range of 3 immunoturbidimetric assays originally developed for human beings determined by serial dilution of
a human calibration standard with a C-reactive protein (CRP) concentration of 153 mg/l (A1, Randox; A2, Thermo; A3, Wako) and
serial dilution of a canine serum sample with a CRP concentration of 50 mg/l (B1, Randox; B2, Thermo; B3, Wako).
agreement were large, with a range of 212.64 to significant impact of the factor disease on results (P
13.21 mg/l (Randox vs. ELISA), 29.96 to 8.79 mg/l , 0.0001), as well as of the applied test (P 5
(Thermo), and 211.67 to 12.90 mg/l (Wako). 0.0136). There was no significant interaction be-
tween the factors (P 5 0.2129). As depicted in
Clinical validation Figure 2, all 3 CRP assays detected significantly
Data from the diseased dogs were not normally higher CRP values in dogs with neoplastic diseases
distributed. The 2-way ANOVA revealed a highly compared with the control. Both the Randox and
Table 1. Intra-assay repeatability of 3 automated immuno- Table 2. Interassay repeatability of 3 automated immuno-
turbidimetric assays (Randox, Thermo, and Wako) originally turbidimetric assays (Randox, Thermo, and Wako) originally
developed for human beings for detection of a high (.45 mg/l), developed for human beings for detection of a high (.45 mg/l),
medium (,15 mg/l), and low (,8 mg/l) concentrations of canine medium (,15 mg/l), and low (,8 mg/l) concentrations of canine
C-reactive protein.*{ C-reactive protein.*{
Randox Randox
Mean 6 SD (mg/l) 47.81 6 0.38 17.73 6 0.31 5.23 6 0.51 Mean 6 SD (mg/l) 44.35 6 2.79 15.75 6 1.27 8.01 6 1.42
CV (%) 1.0 2.0 10 CV (%) 6.0 8.0 18
Thermo Thermo
Mean 6 SD (mg/l) 16.0 6 0.0 12.3 6 0.47 8.65 6 0.81 Mean 6 SD (mg/l) 28.27 6 6.70 10.07 6 0.88 9.60 6 0.74
CV (%) 0.0 4.0 9.0 CV (%) 24 9.0 8.0
Wako Wako
Mean 6 SD (mg/l) 53.91 6 1.32 22.95 6 1.16 8.23 6 1.10 Mean 6 SD (mg/l) 45.80 6 1.81 16.29 6 3.16 9.30 6 1.35
CV (%) 2.0 5.0 13 CV (%) 4.0 19 15
* CV 5 coefficient of variation; SD 5 standard deviation. * CV 5 coefficient of variation; SD 5 standard deviation.
{ Mean, SD, and CV were calculated from 20 replicate { Mean, SD, and CV (coefficient of variation) were calculated
measurements of the different levels. from 15 replicate measurements of the different concentrations.
548 Klenner, Bauer, Moritz
Figure 2. C-reactive protein (CRP) concentration measured with the Randox, Thermo, and Wako assays in healthy dogs (n 5 36),
dogs with neoplasia (n 5 16), dogs with inflammatory diseases (n 5 17), and dogs with miscellaneous diseases (n 5 48). Black horizontal
lines express the median value.
Table 3. Median and range of the C-reactive protein (CRP) concentrations in healthy dogs and dogs with neoplastic, inflammatory,
and miscellaneous diseases.*
is still higher than the results obtained in the current and because it demonstrated the stability of the
study.16 This discrepancy between results is surprising material over a certain period of time (e.g., 1 year).48
because the Randox and Bayer assays are identical Moreover, test principals of the Wako and Randox
(Randox Laboratories Ltd., personal communication, assays were equal, which made the use of the Randox
2010). One explanation might be that because control possible. A similar principle is also applied in
different patient groups were studied, absolute CRP external quality assessment, where control material
concentrations were not directly comparable. Differ- from 1 company or reference laboratory is used for
ences between the reagent batches used in the present assays from different manufacturers.47 A further
study and the previous investigation might also limitation of the study is that the CRP concentration
contribute. A further possibility might be that of the canine serum used for the linearity study should
antibodies or other components of the test kit have have been approximately the same as the CRP
been changed by the manufacturer, although this is concentration of the human-calibration standard.
not the case according to the information provided by However, serum with a similar high canine CRP level
Randox. The presence of lower CRP results in human was not observed in the dogs evaluated in the current
CRP tests than in assays specifically designed for use study, even in patients with severe underlying
in canine specimens indicates a lower antibody- inflammatory diseases. A reason for this might be
binding capacity. that, even in the Wako and Randox assays, the cross-
The reference intervals established herein have been reactivity between the antibodies and canine CRP was
derived from a slightly lower number of individuals not as high as for human CRP.
(36 instead of 40 dogs) than recommended previ- For the ELISA performed in the method compar-
ously.14 Additionally, nonparametric statistics had to ison experiment, sera had to be frozen and analyzed in
be used, which would have required a sample size of n a batch. According to the literature, CRP is stable for
. 120.41 Unfortunately, it was not possible, econom- at least 4 months if frozen at 210uC.42 In accordance
ically, to include such a high number of healthy dogs. with other studies, a negative influence from freezing
Therefore, the reference intervals have to be seen as on CRP stability was, therefore, considered un-
preliminary, and the available data are reported likely.16,17 Simultaneous measurements of the auto-
graphically as suggested previously for small sam- mated assays and the ELISA would have been
ples.13 Similar to the slightly lower results obtained optimal. However, this procedure was not performed
from a similar low numbers of dogs of ,7.1 mg/l (n 5 in the current study because the preliminary results of
19),39 ,6.3 mg/l (n 5 8),18 and ,8.4 mg/l (n 5 14)16 the automated CRP assays were directly compared
have been described in the literature for healthy dogs. with the actual disease status of the patients to gain
No significant influence by sex on reference intervals further experience with the different new tests,
was detected. This is in accordance with other studies whereas the ELISA analyses had to be performed in
where no sex-related differences in CRP concentra- a batch. The ELISA was performed by 2 examiners
tions could be found in healthy Beagles.22,51 Median (SK and staff member); however, 1 person would have
age of the healthy control group in the current study been optimal to keep interassay variation as low as
was quite young, but significant age-related differ- possible. Interassay and intra-assay repeatability of
ences in CRP concentrations have not been found the ELISA could have been further reduced by using
previously.51 To the authors’ knowledge, only 1 study an automated ELISA processing system, but an
investigated breed-related differences in acute-phase automated system was not available in the authors’
proteins,7 and no significant difference in the CRP laboratory. For interpretation of the ELISA results,
concentration between retired racing Greyhounds and sera with high CRP concentrations that exceeded the
dogs of other breeds was found. Because 16 different upper limit of the ELISA’s linearity (.15 mg/l) could
breeds were included in the current study, breed- not be diluted and reanalyzed because of the low
specific differences in CRP plasma concentration sample volume. Thus, no information on method
were not examined. agreement in higher concentrations can be given.
The present study had several limitations. It had Further studies assessing the linear relationship
been recommended that canine-specific control ma- between the new automated tests and the ELISA as
terial be used, but the material was not available in the gold standard are recommended, especially for
the authors’ laboratory at the time of the experiment. higher CRP concentrations.
Additionally, test-specific control material could not In summary, the current study revealed that
be used for the Wako assay. The Randox control was automated CRP tests developed for human beings
used instead. It was considered to be adequate from Randox and Wako demonstrated cross-reactiv-
because it fulfilled the defined criteria of a control, ity with canine CRP and, thus, could be used to detect
including an acceptable stated CRP concentration, an increase in canine CRP concentration from
Automated assays for canine C-reactive protein detection 551
neoplasia or inflammation. Further studies are 5. Ceron JJ, Eckersall PD, Martynez-Subiela S: 2005, Acute
required to minimize the imprecision of the tests phase proteins in dogs and cats: current knowledge and future
perspectives. Vet Clin Pathol 34:85–99.
and to gain deeper insight into the accuracy of the 6. Chan DL, Rozanski EA, Freeman LM: 2009, Relationship
assays before their routine use in veterinary labora- among plasma amino acids, C-reactive protein, illness
tories. The Thermo assay cannot be recommended for severity, and outcome in critically ill dogs. J Vet Intern Med
use with canine specimens because of a low cross- 23:559–563.
reactivity with canine CRP. 7. Couto CG, Ceron JJ, Parra MD, et al.: 2009, Acute phase
protein concentrations in retired racing Greyhounds. Vet Clin
Acknowledgements Pathol 38:219–223.
8. Dabrowski R, Wawron W, Kostro K: 2007, Changes in CRP,
The authors wish to thank Dr. Klaus Failing, Institute SAA and haptoglobin produced in response to ovariohyster-
for Biomathematics and Data Processing, Justus-Liebig- ectomy in healthy bitches and those with pyometra. Therio-
University, Giessen, Germany, for his kind advice in genology 67:321–327.
statistical questions. The authors affirm that they do not 9. Eckersall PD, Conner JG, Harvie J: 1991, An immunoturbidi-
have any affiliations or financial involvement in any metric assay for canine C-reactive protein. Vet Res Commun
organization or entity with a direct financial interest in 15:17–24.
10. Fransson BA, Bergstrom A, Wardrop KJ, Hagman R: 2007,
the tests discussed in this manuscript.
Assessment of three automated assays for C-reactive protein
determination in dogs. Am J Vet Res 68:1281–1286.
Sources and manufacturers 11. Freeman LJ, Rahmani EY, Sherman S, et al.: 2009,
Oophorectomy by natural orifice transluminal endoscopic
a. PhaseTM Range canine CRP, Tridelta Development Ltd.,
surgery: feasibility study in dogs. Gastrointest Endosc
Maynooth, County Kildare, Ireland.
69:1321–1332.
b. Bayer CRP TIA, Bayer plc, Newbury, Berkshire, United
12. Ganapathi MK, Rzewnicki D, Samols D, et al.: 1991, Effect
Kingdom.
of combinations of cytokines and hormones on synthesis of
c. CRP OSR 6147, Olympus Life and Material Science Europe
serum amyloid A and C-reactive protein in Hep 3B cells. J
GmbH, Lismeehan, O’Callaghan’s Mills, Ireland.
Immunol 147:1261–1265.
d. High Linearity CRP (catalog CP7950), Randox Laboratories
13. Geffre A, Braun JP, Trumel C, Concordet D: 2009,
Ltd., Crumlin, United Kingdom.
Estimation of reference intervals from small samples: an
e. KonelabTM CRP (catalog 981699), Thermo Clinical Labsys-
example using canine plasma creatinine. Vet Clin Pathol
tems Oy, Vantaa, Finland.
38:477–484.
f. CRP-HS (catalog 419-22007), Wako Chemicals GmbH,
Neuss, Germany. 14. Geffre A, Friedrichs K, Harr K, et al.: 2009, Reference values:
a review. Vet Clin Pathol 38:288–298.
g. Horiba ABX Pentra 400, Horiba ABX, Montpellier, France.
h. CRP Plus Calibrator, Thermo Clinical Labsystems Oy, 15. Kjelgaard-Hansen M, Jensen AL, Houser GA, et al.: 2006,
Vantaa, Finland. Use of serum C-reactive protein as an early marker of
i. Dade Behring, Siemens Healthcare Diagnostics GmbH, inflammatory activity in canine type II immune-mediated
Eschborn, Germany. polyarthritis: case report. Acta Vet Scand 48:9.
j. SediplusH BSG/ESG/VS-Westergren, Sarstedt, Nürnbrecht, 16. Kjelgaard-Hansen M, Jensen AL, Kristensen AT: 2003,
Germany. Evaluation of a commercially available human C-reactive
k. ADVIA 2120, Siemens Healthcare Diagnostics GmbH, protein (CRP) turbidometric immunoassay for determination
Eschborn, Germany. of canine serum CRP concentration. Vet Clin Pathol 32:
l. Amelung-Coagulometer KC4A, Heinrich Amelung GmbH, 81–87.
Lemgo, Germany. 17. Kjelgaard-Hansen M, Jensen AL, Kristensen AT: 2004,
m. Analyse-it for Microsoft Excel (version 2.00), Analyse-it Internal quality control of a turbidimetric immunoassay for
Software Ltd., Leeds, United Kingdom. canine serum C-reactive protein based on pooled patient
n. BMDP Statistical Software 8.1., Statistical Solutions Ltd., samples. Vet Clin Pathol 33:139–144.
Cork, Ireland. 18. Kjelgaard-Hansen M, Kristensen AT, Jensen AL: 2003,
Evaluation of a commercially available enzyme-linked immu-
nosorbent assay (ELISA) for the determination of C-reactive
References protein in canine serum. J Vet Med A Physiol Pathol Clin
1. Bathen-Noethen A, Carlson R, Menzel D, et al.: 2008, Med 50:164–168.
Concentrations of acute-phase proteins in dogs with steroid 19. Kjelgaard-Hansen M, Mikkelsen LF, Kristensen AT, et al.:
responsive meningitis-arteritis. J Vet Intern Med 22: 2003, Study on biological variability of five acute phase
1149–1156. reactants in dogs. Comp Clin Pathol 12:69–74.
2. Bayramli G, Ulutas B: 2008, Acute phase protein response in 20. Kjelgaard-Hansen M, Stadler M, Jensen AL: 2008, Canine
dogs with experimentally induced gastric mucosal injury. Vet serum C-reactive protein detected by means of a near-patient
Clin Pathol 37:312–316. test for human C-reactive protein. J Small Anim Pract
3. Caldin M, Tasca S, Carli E, et al.: 2008, Serum acute phase 49:282–286.
protein concentrations in dogs with hyperadrenocorticism 21. Knapp A, Schrödl W, Grevel V, Oechtering G: 2009,
with and without concurrent inflammatory conditions. Vet Konzentration des C-reaktiven proteins (CRP) im Blutserum
Clin Pathol 38:2–12. bei Hunden mit postoperativen Wundheilungsstörungen [C-
4. Caspi D, Baltz ML, Snel F, et al.: 1984, Isolation and reactive protein (CRP) serum-concentration in dogs with
characterization of C-reactive protein from the dog. Immu- impaired postoperative wound healing]. Tierarztl Prax
nology 53:307–313. 37:21–27. In German. Abstract in English.
552 Klenner, Bauer, Moritz
22. Kuribayashi T, Shimada T, Matsumoto M, et al.: 2003, 37. Ng PM, Jin Z, Tan SS, et al.: 2004, C-reactive protein: a
Determination of serum C-reactive protein (CRP) in healthy predominant LPS-binding acute phase protein responsive to
beagle dogs of various ages and pregnant beagle dogs. Exp Pseudomonas infection. J Endotoxin Res 10:163–174.
Anim 52:387–390. 38. Nielsen L, Toft N, Eckersall PD, et al.: 2007, Serum C-
23. Lowrie M, Penderis J, Eckersall PD, et al.: 2009, The role of reactive protein concentration as an indicator of remission
acute phase proteins in diagnosis and management of steroid- status in dogs with multicentric lymphoma. J Vet Intern Med
responsive meningitis arteritis in dogs. Vet J 182:125–130. 21:1231–1236.
24. Martinez-Subiela S, Bernal LJ, Ceron JJ: 2003, Serum 39. Parra MD, Tuomola M, Cabezas-Herrera J, Ceron JJ: 2006,
concentrations of acute-phase proteins in dogs with leishman- Analytical and clinical validation of a time-resolved immuno-
iosis during short-term treatment. Am J Vet Res 64: fluorimetric assay (TR-IFMA) for canine C-reactive protein
1021–1026. in serum. Vet Res Commun 30:113–126.
25. Martinez-Subiela S, Ceron JJ: 2004, Validación analı́tica de 40. Petrie A, Watson P: 2006, Statistics for veterinary and animal
técnicas comerciales para la determinación de haptoglobina, science, 2nd ed. Blackwell Publishing, Oxford, U.K.
proteı́na C reactiva y amiloide A sérico en caninos [Analytical 41. Reed AH, Henry RJ, Mason WB: 1971, Influence of statistical
validation of commercial assays for the determination of method used on the resulting estimate of normal range. Clin
haptoglobin, C-reactive protein and serum amyloid A in dogs]. Chem 17:275–284.
Arch Med Vet 37:61–66. In Spanish. Abstract in English. 42. Riley RF, Zontine W: 1972, Further observations on the
26. Martinez-Subiela S, Tecles F, Eckersall PD, Ceron JJ: 2002, properties of dog C-reactive protein and the C-reactive protein
Serum concentrations of acute phase proteins in dogs with response in the dog. J Lab Clin Med 80:698–703.
leishmaniasis. Vet Rec 150:241–244. 43. Shimada T, Ishida Y, Shimizu M, et al.: 2002, Monitoring C-
27. Mastrorilli C, Dondi F, Agnoli C, et al.: 2007, Clinicopath- reactive protein in beagle dogs experimentally inoculated with
ologic features and outcome predictors of Leptospira inter- Ehrlichia canis. Vet Res Commun 26:171–177.
rogans Australis serogroup infection in dogs: a retrospective
44. Suzuki K, Wakabayashi H, Takahashi M, et al.: 2007, A
study of 20 cases (2001–2004). J Vet Intern Med 21:3–10.
possible treatment strategy and clinical factors to estimate the
28. Matijatko V, Mrljak V, Kis I, et al.: 2007, Evidence of an
treatment response in Babesia gibsoni infection. J Vet Med Sci
acute phase response in dogs naturally infected with Babesia
69:563–568.
canis. Vet Parasitol 144:242–250.
45. Tagata K, Yokoyama S, Ginbo T, et al.: 1996, Quantitative
29. McCann TM, Ridyard AE, Else RW, Simpson JW: 2007,
capillary reversed passive latex agglutination test for C-
Evaluation of disease activity markers in dogs with idiopathic
reactive protein (CRP) in the dog. Vet Res Commun 20:21–30.
inflammatory bowel disease. J Small Anim Pract 48:620–625.
30. McGrotty YL, Knottenbelt CM, Ramsey IK, et al.: 2004, 46. Taylor AW, Ku NO, Mortensen RF: 1990, Regulation of
Evaluation of a rapid assay for canine C-reactive protein. Vet cytokine-induced human C-reactive protein production by
Rec 154:175–176. transforming growth factor-beta. J Immunol 145:2507–2513.
31. Mischke R, Waterston M, Eckersall PD: 2007, Changes in C- 47. Tecles F, Spiranelli E, Bonfanti U, et al.: 2005, Preliminary
reactive protein and haptoglobin in dogs with lymphatic studies of serum acute-phase protein concentrations in
neoplasia. Vet J 174:188–192. hematologic and neoplastic diseases of the dog. J Vet Intern
32. Mold C, Baca R, Du Clos TW: 2002, Serum amyloid P Med 19:865–870.
component and C-reactive protein opsonize apoptotic cells for 48. Thomson M, Wood R: 1995, Harmonized guidelines for
phagocytosis through Fcgamma receptors. J Autoimmun 19: internal quality control in analytical chemistry laboratories
147–154. (technical report). Pure Appl Chem 67:649–666.
33. Mold C, Du Clos TW: 2006, C-reactive protein increases 49. Ulutas B, Bayramli G, Ulutas PA, Karagenc T: 2005, Serum
cytokine responses to Streptococcus pneumoniae through concentration of some acute phase proteins in naturally
interactions with Fc gamma receptors. J Immunol 176: occurring canine babesiosis: a preliminary study. Vet Clin
7598–7604. Pathol 34:144–147.
34. Mold C, Rodriguez W, Rodic-Polic B, Du Clos TW: 2002, C- 50. Watanabe M, Oikawa T, Hiraoka H, et al.: 2006, Experi-
reactive protein mediates protection from lipopolysaccharide mental inoculation of beagle dogs with Ehrlichia species
through interactions with Fc gamma R. J Immunol 169: detected from Ixodes ovatus. Vet Parasitol 136:147–154.
7019–7025. 51. Yamamoto S, Shida T, Okimura T, et al.: 1994, Determina-
35. Moritz A, Krämer F: 2009, Die kanine Leishmaniose des tion of C-reactive protein in serum and plasma from healthy
Hundes [The canine leishmaniasis of the dog]. Tierarztl Prax dogs and dogs with pneumonia by ELISA and slide reversed
37:290–294. In German. passive latex agglutination test. Vet Q 16:74–77.
36. Nakamura M, Takahashi M, Ohno K, et al.: 2008, C-reactive 52. Yamamoto S, Tagata K, Nagahata H, et al.: 1992, Isolation of
protein concentration in dogs with various diseases. J Vet Med canine C-reactive protein and characterization of its proper-
Sci 70:127–131. ties. Vet Immunol Immunopathol 30:329–339.