You are on page 1of 7

Acta Ophthalmologica 2021

Subjective and objective refractions in eyes with


extended-depth-of-focus intraocular lenses using
echelette optics: clinical and experimental study
Yuka Ota,1,2 Keiichiro Minami,1 Shinichi Oki,1 Hiroko Bissen-Miyajima,1 Keiichiro Okamoto,3
Masaya Nakashima3 and Kazuo Tsubota2
1
Department of Ophthalmology, Tokyo Dental Collage Suidobashi Hospital, Tokyo, Japan
2
Department of Ophthalmology, Keio University School of Medicine, Tokyo, Japan
3
Tomey Corporation, Nagoya, Japan

ABSTRACT. Introduction
Purpose: To evaluate differences in subjective and objective refractions in eyes
with extended-depth-of-focus intraocular lenses (EDOF IOLs) using echelette Measurement of the subjective refrac-
optics, and the effect of the light wavelength used during examinations. tions after implantation of presbyopia-
Methods: In the prospective study, subjective and objective refractions of 128 correction intraocular lenses (IOLs) is
critical for determining if the postop-
eyes of 64 patients were examined 3 months after implantation of the EDOF
erative refractions are within the range
IOLs (ZXR00V, Johnson & Johnson Surgival Vision). Objective refractions
anticipated preoperatively. Objective
were measured using an autorefractor with a near-infrared (NIR) light source. refractions measured with an autore-
Clinical differences in the spherical, cylindrical and spherical equivalent (SE) fractor are normally used as the initial
refractions between the subjective and objective refractions were evaluated. estimates of the subjective refractions
Then, lens powers of monofocal, EDOF and diffractive bifocal IOLs in the use of (Kinge et al. 1996; Bullimore et al.
a 850-nm light source were measured experimentally for using a lensmeter, and 1998). Although these refractions coin-
the differences from the monofocal IOLs were calculated. cide well for eyes with monofocal
Results: The mean objective refractions were more myopic (p < 0.001) than the IOLs, significant differences between
subjective refractions; the differences in the spherical, cylindrical and SE the refractive values were found with
refractions were 0.71, 0.26 and 0.84 dioptre, respectively. Experimental some refractive multifocal IOLs
investigation resulted that there was the mean difference of 0.83 D with the (Mu~ noz et al., 2007a, 2007b;
EDOF from monofocal IOLs at 850 nm, while the difference was 0.20 D with Albarr an-Diego et al. 2011; van der
bifocal IOLs. Linden et al. 2014). The lights used
Conclusions: The diffractive EDOF IOLs using echelette gratings inherently during the measurements are disruptive
by adding refractions to the IOLs. In
induced constant differences in the subjective and objective refractions, which
contrast, no such refractive differences
arose from the chromatic difference in IOL powers for the visible and NIR lights.
are observed with diffractive bifocal
IOLs (Bissen-Miyajima et al. 2010),
Key words: chromatic focal difference – echelette optics – extended-depth-of-focus intraocular
since the diffractive IOLs form the
lens – near infrared – objective refraction
distance focus with the refractive sur-
Conflict of interest: Minami: AMO Japan K.K, Japan (consulting); Alcon Japan Ltd, Japan (consulting); face of the IOL, which is the same as
Tomey Corporation, Japan (patent pending). Bissen-Miyajima: Alcon (consulting, lectures); Johnson & monofocal IOLs. Thus, the subjective
Johnson Surgical Vision (consulting, lectures); Zeiss (consulting); Santen (consulting, lectures); HOYA refractions can be measured in the
(consulting, lectures). Okamoto and Nakashima: Tomey Corporation, Japan (employment). The other
authors declare that they have no conflict of interest.
same manner as with monofocal IOLs.
A new concept of presbyopia-cor-
rection IOLs was developed by extend-
Acta Ophthalmol. 2021: 99: e837–e843 ing the depth of focus. The Symfony
ª 2020 The Authors. Acta Ophthalmologica published by John Wiley & Sons Ltd on behalf of Acta Ophthalmologica
Scandinavica Foundation
IOL (Johnson & Johnson Surgical
This is an open access article under the terms of the Creative Commons Attribution License, which permits use, Vision, Santa Ana, CA, USA) is a
distribution and reproduction in any medium, provided the original work is properly cited. diffractive extended-depth-of-focus
doi: 10.1111/aos.14660 (EDOF) IOL that uses echelette grat-
ings and produces the distance and

e837
Acta Ophthalmologica 2021

near foci with the first- and second- ocular surgery, or had chronic or formula (A constant: 119.3) and the
order diffractions, respectively (Weeber recurrent uveitis, acute ocular disease, Barrett Universal II ( LF: 2.04), and
et al. 2015; Millan & Vega 2017). With external/internal infection, diabetes the higher power was selected for
EDOF IOLs, a discrepancy between with retinal changes, glaucoma, exfoli- avoiding an hyperopic refractive error.
the subjective and objective refractions ation syndrome, pathological miosis, The target refraction was emmetropia
has been noticed in clinical practices as keratoconus, corneal endothelial dys- or 0.5 D in the staged implantation
reported by the manufacturer. The trophy, or abnormality in the capsule, procedure as described previously (Bis-
effect of the grating on the objective zonule, or pupil, were excluded. One sen-Miyajima et al. 2019).
refraction measurement is concerning. hundred twenty-eight eyes of 64
Visible and near-infrared (NIR) lights patients were enrolled. Table 1 shows Examinations
are used in measurements of the sub- the patient demographic data. The The subjective refractions were
jective and objective refractions, target refractions were emmetropia obtained 3 months postoperatively
respectively, and the power for the for 65 eyes and 0.5 D in 63 eyes. during measurement of the corrected
distance focus changes with the wave- distance visual acuity (CDVA) at 5 m.
lengths of light used for each examina- Implanted intraocular lenses Without the use of the objective
tion. To our knowledge, no study has The ZXR00V IOL was a one-piece, refraction results, the CDVA was
evaluated the difference between the hydrophobic acrylic, diffractive EDOF measured by increasing the spherical
subjective and objective refractions in IOL that is 6.0 mm in diameter, with powers in 0.25-D increments until the
eyes with diffractive EDOF IOLs. The aspheric optics on the front surface, a corrected visual acuity decreased from
purpose of the current study was to continuous sharp optic edge on the the best-corrected measurement, and
clinically evaluate the refractive differ- posterior surface, and anteriorly shifted the power before the decrease was
ences and experimentally determine the haptic designs. The EDOF function recorded as the subject’s spherical
etiology. was provided by using the echelette refraction. The cylindrical power and
grating on the posterior surface. The axis were examined using the cross-
add power for the near focus was +1.75 cylinder technique in increments of
Methods D. In the echelette gratings, the line 0.25 D. The uncorrected distance
Clinical and experimental evaluations spacing was larger and the efficacy for visual acuity (UDVA) also was mea-
were performed in this study. First, the higher order diffractions was more sured. The UDVA and CDVA were
differences were assessed between the optimized compared with the diffrac- converted to the logarithm of the
subjective refractions obtained from tive optics used in bifocal IOLs. Thus, minimum angle of resolution (log-
visual acuity examinations and objec- unlike forming a near focus with the MAR) for analysis.
tive refractions measured with an first-order diffraction in bifocal IOLs, Objective refractions were measured
autorefractor after implantation of the EDOF IOL was designed to form with the TONOREF II autorefractor
diffractive EDOF IOLs. In the use of the foci for the distance and slightly (Nidek, Gamagori, Japan). The instru-
diffractive bifocal IOL in which the near vision with the first- and second- ment utilized the Scheiner principle and
distance focus is formed with the order diffractions. The blazed structure projected an image of a circle on the
refractive element, while the EDOF increased the efficacy of the first- and retina, which was observed through the
IOL formed it with the grating element. second-order diffractions (Mill an & cornea that had been disrupted by the
Hence, the effect of chromatic differ- Vega 2017). ocular refraction. The light source used
ence in IOL power for a NIR light on The IOLs were implanted during a was a superluminescent diode emitting
the objective refractions was experi- femtosecond laser-assisted cataract sur- in a wavelength in the range of 870–
mentally examined. gery using the LenSx laser system 900 nm. The refractive measurement
(Alcon Laboratories, Ft. Worth, TX, areas were expanded within the entire
USA). The size of capsulotomy was pupil up to 4.0 mm in diameter. An
Difference between subjective and
5.0 mm. The IOL was inserted into the autofogging mechanism was used to
objective refractions
capsular bag using the injector through reduce the accommodative effect. The
The institutional review board of a 2.2-mm-wide corneal incision. With observed ring images were fit with
Tokyo Dental College (identifier, 800) biometry measurement using an IOL- ellipses, and the spherocylindrical
approved this clinical prospective Master 700 (Carl Zeiss Meditec, Inc., refractions were calculated. The spher-
study, which was conducted according Dublin, CA, USA), the IOL powers ical equivalent (SE) refraction also was
to the tenets of the Declaration of were calculated using the SRK/T calculated for analysis.
Helsinki. Patients who were scheduled
to undergo cataract surgery with
Table 1. Patient demographic data.
implantation of a diffractive EDOF
IOL (Symfony ZXR00V) were Mean (SD) Range
recruited, and all patients provided
written informed consent. The inclu- No. of eyes/patients 128/64
sion criteria were anticipated postop- Age (years) 67.6  6.5 51–82
Axial length (mm) 24.7  1.6 21.6–29.6
erative astigmatism of 1.25 dioptres
Averaged keratometry (D) 43.6  1.1 40.9–46.7
(D) or less and the conventional crite- Keratometric astigmatism (D) 0.68  0.38 0.05–1.72
ria for multifocal IOL implantation.
Eyes that had undergone a previous D = dioptre, SD = standard deviation.

e838
Acta Ophthalmologica 2021

Statistical analysis refraction measurement with NIR light 15.0D monofocal IOLs, respectively.
The differences and correlations in the could differ from the amount measured Based on geometric optics, the lens
sphere, cylinder and SE values were in visible light. The differences were powers (D) were negatively propor-
examined between the subjective and investigated experimentally. tional to the distance of the pinholes.
objective refraction examinations using Six distances between three points
a paired t-test and linear regression Power measurements at 850 nm around the centre were measured using
analysis, respectively. Systematic errors Monofocal, bifocal and EDOF one- the ImageJ software (version 1.52p;
between the subjective (at 5 m) and piece hydrophobic acrylic IOLs National Institutes of Health,
objective (at infinity) refractions, which (ZCB00V, ZMB00 and ZXR00V, Bethesda, MD, USA) and averaged.
corresponded to 0.2 D, were not cor- respectively; Johnson & Johnson Surgi- For bifocal IOLs, the points were split
rected in the same manner as the cal Vision) were used because they have towards the centre owing to the +4.0 D
previous works (Bissen-Miyajima the same lens and haptics platform, add power (Fig. 2C), so that the outer
et al. 2010; van der Linden et al. except for the optical element for pres- points were subject to measurement.
2014). Differences between the two byopia correction. IOLs with powers of For EDOF IOLs with a +1.75 D add
refraction examinations were systemat- 10.0, 15.0 and 20.0 D were used for power, the split was too small and the
ically examined using the Bland–Alt- measurements at a wavelength of image was slightly oval (Fig. 2D).
man plots (Bland & Altman 1999) for 850 nm (NIR). Three lenses of each Therefore, the distances between the
the three refractive values. Data are power were examined. A lensmeter (TL- outer ellipse foci were measured.
expressed as the mean  standard 7000, Tomey, Nagoya, Japan) based on
deviation, otherwise specified. the Hartmann sensor (Hartmann 1900) Statistical analysis
p < 0.05 was considered significant. was modified to accommodate a NIR Differences in the powers of the bifocal
LED (LSB855SL-550G; OPTRANS, and EDOF IOLs from those of the
Latham, NY, USA), which had a peak monofocal IOLs were evaluated. First,
Effect of chromatic difference in
wavelength of 850 nm, and a CMOS a calibration equation between the
intraocular lens power
camera (da1280-54uc; Basler AG, powers labelled and point distances
The diffractive EDOF IOL had an Ahrensburg, Germany). measured was obtained with data with
echelette optical element, and the first- shows a schematic diagram of the the monofocal IOL using regression
order diffraction constituted the dis- power measurement. Collimated light analysis. With the calibration equation,
tance focus. The 1.75-D add power was from the LED illuminated the IOLs the powers of the bifocal and EDOF
produced in visible light, while the light that were immersed in water. The IOLs at 850 nm were calculated from
used in the autorefractor normally has focused light passed through the Hart- the pinhole distance data. Differences
been NIR to retain the intraocular mann plate with 117 pinholes and in the power from the powers of the
permeability. The diffraction gratings condenser lenses to the CMOS camera. monofocal IOLs were examined using
induce negative chromatic aberration, The CMOS camera captured the pin- the Wilcoxon signed-rank test.
which was used to compensate chro- hole images.
matic aberration induced by monofocal The distance between the pinholes
refractive IOL (Weeber et al. 2015). decreased as the power of the IOL
Results
The negative chromatic aberration also increased, as the Hartmann pinholes
Differences between subjective and
increased the refractive powers of the were posterior to the IOL. Figure 2A,B
objective refractions
IOL with the wavelength (Ravikumar shows representative images of the
et al. 2014), so that the objective pinhole images for the 10.0-D and The postoperative UDVAs ranged
from 0.30 to 0.52 logMAR (6/20 to
20/10 in Snellen notation), and the
mean was 0.03  0.12. The CDVAs
ranged from 0.30 to 0.00 logMAR
(20/20 to 20/10 in Snellen notation)
with the mean of 0.15  0.05. All
eyes obtained a CDVA of 20/20 or
better. No intraoperative or postoper-
ative complications occurred. The
mean refractions and their differences
are shown in Table 2. There were
significant differences between the two
refractions in the spherical, cylindrical
and SE (p < 0.0001, paired t-test). The
mean differences showed that the
objective refractions shifted myopi-
cally, and the SE difference was 0.84
Fig. 1. Schematic diagram of the intraocular lens (IOL) power measurement. Lights from a near- D.
infrared LED (light emitting diode) illuminate the IOL in water after collimated with collimated The linear regression analysis
lenses (L1). Focused lights pass through the Hartmann plate with 117 pinholes and condenser between the refractions showed signif-
lenses (L2, L3). The pinhole images are captured by the CMOS camera. icant (p < 0.001) and strong

e839
Acta Ophthalmologica 2021

Fig. 2. Representative images of Hartmann pinholes with monofocal intraocular lenses (IOLs) of 10.0 D (A) and 20.0 D (B), bifocal IOL of 10.0 D
(C), and extended-depth-of-focus (EDOF) IOL of 10.0 D (D) at a wavelength of 850 nm. The distance between the pinholes decreased as the IOL
power increased (A and B). A split in the pinhole images (arrow in C) is observed with the bifocal IOL. Slight ovalization (arrow in D) is seen with the
EDOF IOL. D, dioptre.

Table 2. Mean subjective and objective refractions and their differences.

Subjective (D) Objective (D) p Value* Difference (D)

Spherical 0.14  0.42 ( 1.00 to 1.00) 0.85  0.46 ( 2.13 to 0.27) <0.0001 0.71  0.24 ( 1.38 to 0.14)
Cylindrical 0.45  0.42 ( 1.50 to 0.00) 0.72  0.41 ( 1.89 to 0.04) <0.0001 0.26  0.22 ( 1.29 to 0.15)
SE 0.37  0.43 ( 1.50 to 1.00) 1.21  0.44 ( 2.25 to 0.19) <0.0001 0.84  0.24 ( 1.51 to 0.32)

D = dioptre, SE = spherical equivalent.


The data are expressed as the mean  standard deviation (range).
* Paired t-test.

(R2 = 0.74, 0.73, and 0.73) correlations represented the mean and difference of data with monofocal IOLs) indicated
in the spherical, cylindrical and SE the subjective and objective refractions, that the distances of the pinholes were
refractions, respectively. The coeffi- respectively. Systematic differences were significantly and strongly corrected
cients of the resultant regression equa- found in the spherical and SE refrac- with the power labelled (p < 0.001,
tions, objective = Slope * subjec- tions. In the cylindrical refraction, the R2 > 0.999), and the distances
tive + Constant, are shown in objective refraction agreed clinically decreased by 5.7988 (95% confidence
Table 3. The slopes were over 0.83, with the subjective values. interval [CI], 5.9477 to 5.6499). The
which was close to unity. The constant powers of the other IOLs were calcu-
indicated the differences between the lated from the distances of pinholes
Effect of chromatic difference in
refractions. and the above regression result.
intraocular lens power
Figure 3 shows Bland–Altman plots Table 4 shows the calculated powers
for the spherical, cylindrical and SE The regression analysis of the data with at 850 nm of the bifocal and EDOF
refractions. The x- and y-axis monofocal IOLs (Table S1 for a list of IOLs and the differences from the

Table 3. Regression equations of objective spherical, cylindrical and (SE) refractions.

Slope Constant (D)

Spherical 0.945 (95% CI, 0.847 to 1.043) 0.721 (95% CI, 0.765 to 0.678)
Cylindrical 0.837 (95% CI, 0.747 to 0.926) 0.331 (95% CI, 0.386 to 0.275)
SE 0.868 (95% CI, 0.774 to 961) 0.891 (95% CI, 0.944 to 0.838)

CI = confidence interval, D = dioptre, SE = spherical equivalent.

e840
Acta Ophthalmologica 2021

Fig. 3. Bland–Altman plots for the spherical (A), cylindrical (B) and SE (C) refractions. The differences in refraction denote the differences between
the objective and subjective values. The solid and broken lines indicate the mean and 95% limits of agreement.

Table 4. Calculated powers and the differences from the labelled powers of the bifocal and
EDOF IOLs. 0.00 D and 1.50 D (Kohnen et al.
2019). To maximize the function of the
Wavelength = 850 nm EDOF IOLs, postoperative refractive
Labelled power (D) Calculated powers (D) Differences (D)
errors should be minimized. Ideally,
the refractions should be examined
Bifocal IOLs without the use of the objective refrac-
10.0 10.11, 9.58, 9.97 0.11, 0.42, 0.03 tion data as in the current study, but
15.0 14.69, 14.61, 14.77 0.31, 0.39, 0.23 the procedure is time consuming. With
20.0 19.66, 19.98, 19.77 0.34, 0.02, 0.23 the current results, the differences
EDOF IOLs
inherent in the subjective refraction
10.0 10.97, 10.70, 10.70 0.97, 0.70, 0.70
15.0 15.64, 15.78, 15.76 0.64, 0.78, 0.76
examination could be compensated
20.0 20.97, 21.00, 20.98 0.97, 1.00, 0.98 for easily.
Since the chromatic aberrations of
D = dioptre, EDOF = extended depth of focus, IOL = intraocular lens. the diffractive and refractive lenses are
opposite each other, diffractive lenses
have been used to compensate it inher-
powers of the monofocal IOLs. For the mean difference was 0.89 D in the SE ent in the refractive lens (Davidson
bifocal IOLs, the calculated powers refractions. The experimental results et al. 1993; Flores et al. 2004). In the
were around the labelled values, and indicated that the powers of the EDOF diffractive EDOF IOL, the refractive
the mean difference from the monofo- IOLs were underestimated when mea- and diffractive (first-order diffraction)
cal IOLs was 0.21  0.19 D. In sured at 850 nm, and the difference optical elements were combined for the
contrast, the calculated powers were from the monofocal IOLs was 0.83 D. compensation in visible light (Mill
an &
higher than the labelled powers, and These findings indicated that the clin- Vega 2017). The power of the diffrac-
the mean difference between the EDOF ical difference in eyes with the EDOF tive optical element is determined as
and monofocal IOLs was 0.83  0.14 IOLs would result from the chromatic 2mk/rm2 (D), where rm indicates the
D. difference in IOL power for NIR lights. radius of the mth zone in metres and k
Clinically, the objective refractions is the wavelength of light in metres
for eyes with EDOF IOLs were biased (Ravikumar et al. 2014). With the
by approximately 1.0 D and the bias calculation, the difference in the EDOF
Discussion did not change. The binocular defocus IOL powers at 550 and 870–900 nm
The current evaluations showed that curve of eyes with the Symfony was obtained as 1.08–1.11 D, which
there were differences between the EDOF IOLs indicated that the visual was close to the experimental result.
subjective and objective refractions in acuity ranges from 0.05 to 0.02 Thus, the current evaluations showed
eyes with diffractive EDOF IOLs. The logMAR could be obtained between that the chromatic difference induced

e841
Acta Ophthalmologica 2021

by the diffractive element resulted in technologies (Rozema et al. 2005; Had- diffractive multifocal intraocular lenses with
higher objective refraction measure- dad et al. 2020). The wavelengths of three add powers. Am J Ophthalmol 199:
ments using 800- to 850-nm lights. lights used in these measurements were 223–229.
Bland J & Altman D (1999): Measuring
The current results indicated that the red light (650–785 nm) and NIR (808–
agreement in method comparison studies.
difference between the subjective and 850 nm), so that the influence of the Stat Methods Med Res 8: 135–160.
objective refractions was inherent from diffractive optical element would alter. Bullimore MA, Fusaro RE & Adams CW
the use of the echelette grating element. Further examinations are necessary for (1998): The repeatability of automated and
For diffractive bifocal IOLs, such a confirmation. Next, there were system- clinician refraction. Optom Vis Sci 75: 617–
difference has not been found clinically atic errors between the subjective and 622.
(Bissen-Miyajima et al. 2010), since the objective refractions (0.2 D) owing to Davidson N, Friesem AA & Hasman E (1993):
distance focus is formed with the the 5-m measurement of subjective Analytic design of hybrid diffractive-refrac-
tive achromats. Appl Opt 32: 4770–4774.
refractive elements. However, in the refraction. The subjective examinations
Flores A, Wang MR & Yang JJ (2004):
EDOF IOL, the distance focus was for infinity are impractical. In a previ- Achromatic hybrid refractive-diffractive lens
formed with the first-order diffraction ous evaluation of bifocal IOLs, the with extended depth of focus. Appl Opt 43:
of the echelette grating element, so that mean differences in the sphere and SE 5618–5630.
the effect of the difference in IOL refraction were 0.12 and 0.28 D Haddad JS, Gouvea L, Ferreira JL, Ambr osio
power would be enhanced at the 800– (Bissen-Miyajima et al. 2010), whereas R Jr, Waring GO 4th & Rocha KM (2020):
850 nm wavelength. It was supposed the subjective refraction was examined Impact of a chromatic aberration-correcting
that the effect resulting from the refrac- in increments of 0.25 D. The effect intraocular lens on automated refraction. J
Refract Surg. 36: 334–339.
tive lens would be relatively small would be clinically negligible. In the
Hartmann J (1900): Bemerkungen uber den
because the Abbe number (55) and experiment, the areas of the clinical Bau und die Justirung von Spektrographen.
refractive index (1.47) of the IOLs were and experimental measurements were Z. Instrumentenkd 20: 17–27.
relatively low (Millan & Vega 2017). In not comparable. The experimental Kinge B, Midelfart A & Jacobsen G (1996):
autorefractors, the difference between measurement was analysed around Clinical evaluation of the Allergan Humphrey
the visible and NIR light should be nine central points, so that the results 500 autorefractor and the Nidek AR-1000
corrected for refractive lenses. How- would represent values on the optical autorefractor. Br J Ophthalmol 80: 35–39.
ever, diffractive lenses have been rarely axis. However, the autorefractor mea- Kohnen T, B€ ohm M, Hemkeppler E,
Sch€onbrunn S, DeLorenzo N, Petermann
used for pseudophakic eyes, so that the sured a wider area of 2–4 mm in
K & Herzog M (2019): Visual performance
difference found in the EDOF IOL diameter. The corneal keratometry of an extended depth of focus intraocular
would not have been addressed and would change from the centre to the lens for treatment selection. Eye 33: 1556–
corrected. The use of diffracted light to surroundings, and the difference 1563.
form the distance focus is also seen in involves an error in the objective van der Linden JW, Vrijman V, Al-Saady R,
diffractive trifocal IOLs. The PanOp- refraction. Corneal topographic data van der Meulen IJ, Mourits MP & Lapid-
tix IOL (Alcon) is based on a quadri- measured using anterior-segment opti- Gortzak R (2014): Autorefraction versus
focal design using three add powers, cal coherence tomography are neces- subjective refraction in a radially asymmet-
ric multifocal intraocular lens. Acta Oph-
one of which is designed for the sary to compensate.
thalmol 92: 764–768.
distance focus together with the refrac- In conclusion, the diffractive EDOF Millan MS & Vega F (2017): Extended depth
tive of the IOL base curve (Sudhir et al. IOLs using echelette gratings inher- of focus intraocular lens: chromatic perfor-
2019). Therefore, the effect on the ently induced constant differences mance. Biomed Opt Express 8: 4294–4309.
objective refraction is concerning. A between the subjective and objective Mu~ noz G, Albarran-Diego C & Sakla HF
clinical evaluation of 62 eyes that were refractions, and the differences resulted (2007): Validity of autorefraction after
implanted with the IOLs and examined because the first-order diffraction cataract surgery with multifocal ReZoom
with an auto kerato-refractometer KR- formed the distance focus and the intraocular lens implantation. J Cataract
Refract Surg 33: 1573–1578.
8800 (Topcon, Japan, NIR light powers at visible and NIR lights dif-
Mu~ noz G, Albarran-Diego C & Sakla HF
source) showed no difference in the fered due to the chromatic difference in (2007): Autorefraction after multifocal
spherical refraction between the sub- IOL power. IOLs. Ophthalmology 114: 2100.
jective and objective refractions Ravikumar S, Bradley A & Thibos LN (2014):
(Rementerıa-Capelo et al., 2020), Chromatic aberration and polychromatic
whereas the objective cylindrical and References image quality with diffractive multifocal
SE refractions shifted myopically intraocular lenses. J Cataract Refract Surg
(mean differences, 0.43 and 0.20 D, Albarran-Diego C, Mu~ noz G & Ferrer-Blasco 40: 1192–1204.
T (2011): Subjective refraction before Rementerıa-Capelo LA, Garcıa-Perez JL,
respectively). An examination of the
LASIK enhancement in bioptics procedures Contreras I, Blazquez V & Ruiz-Alcocer J
differences in the powers at 550 and (2020): Automated refraction after trifocal
800–900 nm is required to investigate with refractive multifocal intraocular lenses.
J Refract Surg 27: 556–557. and trifocal toric intraocular lens implanta-
the effect. tion. Eur J Ophthalmol. in press
Bissen-Miyajima H, Minami K, Yoshino M,
The current study had some limita- Nishimura M & Oki S (2010): Autorefrac- Rozema JJ, Van Dyck DE & Tassignon MJ
tions. Clinically, only the autorefractor tion after implantation of diffractive multi- (2005): Clinical comparison of 6 aberrome-
was used for the objective refraction, focal intraocular lenses. J Cataract Refract ters. Part 1: Technical specifications. J
while could have been measured using Surg 36: 553–556. Cataract Refract Surg 31: 1114–1127.
Bissen-Miyajima H, Ota Y, Nakamura K, Sudhir RR, Dey A, Bhattacharrya S & Bahu-
other instruments, such as a Shack–
Hirasawa M & Minami K (2019): Binocular layan A (2019): AcrySof IQ PanOptix
Hartmann aberrometer, automatic reti- intraocular lens versus extended depth of
noscopy, ray tracing and Tscheming visual function with staged implantation of

e842
Acta Ophthalmologica 2021

focus intraocular lens and trifocal intraocu- Chiyoda-ku (employment). The other authors declare that they
lar lens: a clinical overview. Asia Pac J Tokyo, 101-0061 have no conflict of interest.
Ophthalmol (Phila) 8: 335–349. Japan
Funding: The study was supported in part by
Weeber HA, Meijer ST & Piers PA (2015): Tel/Fax: +81-3-5275-1912
Johnson & Johnson Surgical Vision.
Extending the range of vision using diffrac- Email: minami@miyata-med.ne.jp
tive intraocular lens technology. J Cataract
Acknowledgements: We thank Mr. Young Hyun
Refract Surg 41: 2746–2754.
Kim of Tomey Corporation for assisting with the
modification of a lensmeter and measurements. Supporting Information
Conflict of interest: Minami: AMO Japan K.K, Additional Supporting Information
Received on June 5th, 2020.
Japan (consulting); Alcon Japan Ltd, Japan (con- may be found in the online version of
Accepted on October 11th, 2020.
sulting); Tomey Corporation, Japan (patent pend-
this article:
ing). Bissen-Miyajima: Alcon (consulting, lectures);
Correspondence:
Keiichiro Minami, PhD
Johnson & Johnson Surgical Vision (consulting, Table S1. Mean distances of three-
lectures); Zeiss (consulting); Santen (consulting, pinhole images with monofocal IOLs.
Department of Ophthalmology
lectures); HOYA (consulting, lectures). Okamoto
Tokyo Dental College Suidobashi Hospital
and Nakashima: Tomey Corporation, Japan
Kandamisaki-cho 2-9-18

e843

You might also like