You are on page 1of 6

LESSON 2

• It reflects the ability of the method to detect true-


negatives with very few false-positive.
• System of ensuring accuracy and precision in
the laboratory by reagents (quality control) in
KINDS OF QUALITY CONTROL
every series of measurement
• Process of ensuring that analytical results are INTRALAB (Internal QC)
correct by testing known samples (control • Involves the analysis of control samples together
solution) that resembles patient samples with patient specimens.
• Involves the process of monitoring the • Important for the daily monitoring of accuracy
characteristics of the analytical processes and and precision of anlytical methods.
detects analytical errors during testing • Detects both random and systemic error.
• It is one component of the quality assurance
system. INTERLAB (External QC)
• Involves proficiency testing programs that
1. SENSITIVITY periodically provide samples of unknown
• The ability of an analytical method to measure concentration of analytes to participating
the smallest concentration of the analyte of laboratories
interest. • Important in maintaining long term accuracy of
2. SPECIFICITY the analytical method
• Difference of >2 in the results indicates that a
• Is the ability of an analytical method to measure
laboratory is not in agreement with the rest of the
only the analyte of interest.
laboratories included in the program.
3. ACCURACY
• Nearness or closeness of the assayed value to OBJECTIVES OF QUALITY CONTROL
the true or target value. 1. To check the stability of the machine
2. To check the quality of reagents
4. PRECISION OR REPRODUCIBILITY 3. To check technical errors
• Ability to give repeated results on the same
sample that agree with one another. CHARACTERISTICS OF AN IDEAL QC
MATERIAL
5. PRACTICABILITY
1. Resembles human sample
• Degree by which a method is easily repeated. 2. Inexpensive and stable for long periods
6. RELIABILITY 3. No communicable disease.
4. No matrix effects/known matrix effects.
• Ability of analytical method to maintain accuracy
5. With known analyte concentrations.
and precision over extended period of time
6. Convenient packaging for easy dispensing and
during which equipment, reagents and personnel
storage.
change.
7. DIAGNOSTIC SENSITIVITY NOTES TO REMEMBER
• Ability of the test to detect proportion of individual • Quality control materials should resemble human
with that disease who test positively with the test. sample and be available for a minimum of one
year (same lot number)
• Indicates the ability of the test to generate more
• Bovine control material – not the choice for
true-positive results and few false-negative
immunochemistry, dye-binding and certain
8. DIAGNOSTICS SPECIFICITY bilirubin assays.
• Control limits are calculated from the mean and
• Ability of the test to detect proportion of individual
standard deviation
with that disease who test negatively with the
test.

PASCUAL QUALITY MANAGEMENT | 5


POST-ANALYTICAL ERRORS
1. Unavailable or delayed laboratory results.
RANDOM ERROR
2. Incomplete laboratory results.
• Non repeating, occurs once 3. Wrong transcription of the patient’s data and
• The basis for varying differences between laboratory results.
repeated measurement - R
• It is due to instrument, operator, and
environmental conditions
o Pipeting errors, mislabeling of samples, • Is the science of gathering, analyzing,
temperature fluctuations, improper mixing of interpreting and presenting data.
sample and reagent.
MEAN
SYSTEMATIC ERROR • A measure of central tendency. It is associated
• Repeating, can be predicted with symmetrical or normal distribution.
• Often related to calibration problems,
deterioration of reagents and control materials, STANDARD DEVIATION
unstable and inadequate reagent blanks, • a measure of the dispersion of values from the
contaminated solutions, failing instrumentation mean.
and poorly written procedures. • It helps describe the normal curve.
• It is a measure of the agreement between the • A measure of the distribution range
measured quantity and the true value – A
COEFFICIENT OF VARIATION
a. Constant Error • A percentile expression of the mean
o refers to a difference between the target • Index of precision
value and the assay value.
o Independent of sample concentration VARIANCE
b. Proportional/Slope/Percent Error • Is called the standard deviation squared
o Results in greater deviation from the target • Measure of variability
value due to higher sample concentration.
TERMINOLOGIES
CLERICAL ERROR MEDIAN
• The highest frequency of clerical errors with the • Value of the observation that divides the
use of handwritten labels and request forms. observations into two groups, each containing
equal numbers of observation.
PRE-ANALYTIC ERRORS • It is a midpoint of distribution
1. Improper patient preparation. • 50th percentile
2. Mislabeled specimen.
MODE
3. Incorrect order of draw.
4. Incorrect patient identification • Is the most frequent observation
5. Wrong specimen container INFERENTIAL STATISTICS
6. Incorrect anticoagulant to blood ratio. • Are used to compare the means or standard
7. Improper mixing of sample and additives.
deviations of two groups of data.
8. Incorrect specimen preservation.
9. Incorrect used of tubes for blood collection. T-TEST
10. Mishandled specimen (transport and storage) • Is used to determine whether there is statistically
11. Missed or incorrectly interpreted laboratory significant difference between the means of two
requests. groups of data.

PASCUAL QUALITY MANAGEMENT | 6


F-TEST YOUDEN/TWIN PLOT
• Is used to determine whether there is statistically • It is used to compare results obtained on a high
significant difference between the standard and low control serum from different
deviations of two groups of data. laboratories.
• The points falling from a center but on the 45o
QUALITY CONTROL CHART line suggest a proportional error, and points
falling from the center but not on the 45 o line
GAUSSIAN CURVE
suggest a constant error.
• Bell-shaped curve
• It occurs when the data set can be accurately
described by the SD and mean
• It is obtained by plotting the values from multiple
analysis of a sample. Figure 2-3
• It occurs when the data elemnts are centered Youden/Twin
around the mean with most elements close to the Plot

mean.
• It focuses on the distribution of errors from the
analytical method rather than the values from a
healthy or patient population.

SHEWHART LEVEY-JENNING CHART


• Most widely used system in clinical laboratory
• Allows the laboratorians to apply multiple rules
Figure 2-1 without the aid of a computer
Gaussian Curve • A graphic representation of the acceptable limits
of variation in the results of an analytical method.

CUMULATIVE SUM GRAPH (CUSUM)


Figure 2-4
• It calculates the difference between QC results Levey-
and the target means. Jenning Chart
• V mask is the most common method – requires
computer implementation
• Identifies consistent bias problems.
• This plot will give the earliest indication of
systematic errors (trend)
• Results are out of control when the slope ERRORS WHICH CAN BE OBSERVED
exceeds 45o or a decision (+/- 2.7 SD) is a. Trend
exceeded. o Is formed by control values that either
increase or decrease for six consecutive
days.
o Main cause: deterioration of reagents

Figure 2-2
Cumulative Sum
Graph
Figure 2-5
Trends

PASCUAL QUALITY MANAGEMENT | 7


b. Shift
• Is formed by control values that distribute
• It recognized that the use of simple upper and
themselves on one side or either side of the
lower control limits are not enough to identify
mean for six consecutive days.
analytical problems.
• Shift in the reference range is due to transient
• Westgard used the term control rule to indicate if
instrument differences
the analytical process is out of control.
• Main cause: improper calibration of the
instrument.
CONTROL RULES
12s
• use as rejection or warning rule when one control
result exceeds the mean +/- 2SD; for screening
Figure 2-6
Shift
purposes.

c. Outliers
• Are control values thatt are far from the main
set of values.
• They are highly deviating values
• Cause by random or systematic errors

NOTES
• 95% confidence limit (+/- 2SD) – acceptable
range.
• Kurtosis- refers to the degree of flatness or Figure 2-7 12s Rule

sharpness in the peak of a set of values


having a gaussian distribution. 13s
• An analytical method is considered in • reject a run when one control result exceeds the
control when ther is symmetrical distribution mean +/- 3SD.
of control values about the mean and there
are few control values outside the 2s control
limits.
• If the analytical test results to control values
are not within the +/- 2SD confidence range,
run a new set of controls and repeat
specimen testing.
• A control value between 2s and 3s is a sign
of potential problem.
• A value outside 3s would require corrective
action.
• Continued QC failure requires more trouble
shooting – preparation of new reagents,
recalibration, instrument maintenance and
Figure 2-8 13s Rule
repair, and contacting the
dealer/manufacturer for technical support or
service.

PASCUAL QUALITY MANAGEMENT | 8


22s R4s
• reject a run when the last 2 control results (or 2 • reject a run if the range or difference between the
results from the same run); exceed either the highest and lowest control result within an
mean +/- 2SD. analytical run exceeds 4SD.

Figure 2-11 R4s Rule

10x
• reject a run when ten consecutive results are on
the same side of the target mean.

Figure 2-9 22s Rule

41s
• reject a run when the last four (or any four)
consecutive control results exceed either mean
+/- 1SD.

Figure 2-12 10x Rule


Figure 2-10 41s Rule

PASCUAL QUALITY MANAGEMENT | 9


TERMINOLOGIES
ANALYTICAL RUN
• a set of control and patient specimens
assayed, evaluated and reported together.
DELTA CHECK
• Is the most commonly used patient based-QC
technique
• It requires computerization of test data
• The difference between two consecutive
measurement of the same analytes on the same
individual.
POINT OF CARE TESTING
• Analytical testing performed outside the confines
of the central laboratory, usually by non
laboratorians personnel
• Use of portable whole blood glucose meters for
the management of patients with diabetes- most
commonly used POCT
• Other names: near-patient testing, decentralized
testing, bedside testing, alternate site testing.
REFERENCE LIMIT / REFERENCE INTERVAL /
REFERENCE VALUE
• A value obtained by observation or
measurement of a particular type of quantity on
a reference individual.
• Usual values for a healthy population that
represents 95% central tendency.
FIVE FACTORS WHEN ESTABLISHING
REFERENCE INTERVALS:
1. The composition of reference population
2. The criteria of excluding and including
individuals from the reference population.
3. The physiologic and environmental conditions of
the reference population
4. Specimen collection procedure, preparation for
testing
5. Analytical method used

NOTES
• Atleast 120 individuals should be tested for each
age and sex category
• However 20 reference individual need to be
sampled for analysis on the test instrument (not
more than 2 out of 20)
• Reference values vary slightly depending upon
method and specimen type.

PASCUAL QUALITY MANAGEMENT | 10

You might also like