You are on page 1of 8

Ann. occup. Hyg., Vol. 47, No. 4, pp.

305–312, 2003
© 2003 British Occupational Hygiene Society
Published by Oxford University Press
DOI: 10.1093/annhyg/meg043

Permeation of 70% Isopropyl Alcohol Through


Surgical Gloves: Comparison of the Standard Methods
ASTM F739 and EN 374
ERJA A. MÄKELÄ*, SINIKKA VAINIOTALO and KIMMO PELTONEN†
Department of Industrial Hygiene and Toxicology, Finnish Institute of Occupational Health,

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


Topeliuksenkatu 41 a A, FIN-00250 Helsinki, Finland

Received 1 July 2002; in final form 17 December 2002

Standard test methods ASTM F739 and EN 374 were compared by assessing the permeation of
70% isopropyl alcohol (2-propanol) through seven brands of surgical gloves. The two standards
differ in the flow rates of the collection medium and in the chemical permeation rate at which
the breakthrough time (BTT) is detected, the EN detection level being 10 times higher than the
permeation rate used by ASTM. In a departure from the EN standard method, a 4 h testing
time was used instead of 8 h. All of the tested gloves were from the same manufacturer and were
made from either natural rubber (NR) (six brands) or chloroprene rubber (CR) (one brand).
Two of the NR glove brands were double layered. For the thin NR gloves (0.22, 0.28 and
0.27 mm) the permeation rates were higher throughout the tests with a flow rate of 474 ml/min
(EN) of the collection medium (nitrogen) compared with the permeation rates obtained with a
flow rate of 52 ml/min (ASTM). These resulted in BTTs of 4.6, 6.5 and 7.6 min (EN) and 4.8, 6.5
and 9.1 min (ASTM), respectively. No statistical difference could be observed between the BTT
values obtained with the two standard methods for any of the thin gloves. Thus, although the
ASTM standard has a lower criterion for the detection of permeation, it does not necessarily
produce shorter BTTs. For the better barriers the methods yielded more equivalent permeation
rate curves and thus the EN BTTs were longer than the ASTM BTTs: the EN results were 21,
80, 122 and >240 min compared with the ASTM results of 12, 32, 38 and 103 min for glove thick-
nesses of 0.37 (NR), 0.22 + 0.22 (double layered NR), 0.31 + 0.29 (double layered NR) and
0.19 mm (CR), respectively.

Keywords: gloves; isopropyl alcohol; permeation; 2-propanol; standards; surgical

INTRODUCTION Chemical permeation of a volatile test chemical is


measured periodically from the nitrogen or air collec-
The European standard EN 374 (European Commit-
tion medium.
tee for Standardization, 1994) and the American
Both standards have been developed in order to
Society of Testing and Materials (ASTM) standard
F739 (ASTM, 1999) describe similar test methods for assess how quickly and to what extent a chemical can
measuring chemical permeation through protective gain access to the skin through protective clothing.
materials. Both standards define the same test cell Breakthrough time (BTT) and steady-state perme-
with two chambers, a flow-through chamber for the ation rate (SSP) have been established as criteria of
collection medium and a chamber with an inlet for a the resistance to permeation. The BTT is the time
test chemical. A sample of the protective material between the application of the test chemical onto the
being studied is clamped between the chambers. outer surface of the sample material and detection of
the test chemical permeating the material with a
specified rate (Pb). In the ASTM standard, this defin-
ition applies to ‘the normalized breakthrough time’.
The SSP is the value from which the permeation rate
*Author to whom correspondence should be addressed.
E-mail: erja.makela@ttl.fi does not grow and only the ASTM standard requires
†Current address: National Veterinary and Food Research its measurement. The major differences in the two
Institute, PO Box 45, FIN-00580 Helsinki, Finland standards are the Pb and the flow rate of the gaseous

305
306 E. A. Mäkelä, S. Vainiotalo and K. Peltonen

collection medium. The EN standard has a Pb that is the skin to the systemic circulation (Jensen, 1981;
10 times higher than the ASTM standard. The flow Fiserova-Bergerova et al., 1990). It is also known to
rate in the EN standard (5 test cell vol/min corre- be able to permeate through rubber gloves. The danger
sponding to 420–520 ml/min) can also be 10 times in its permeation through the gloves to the skin is that
higher than in the ASTM (50–150 ml/min). it may transfer other hazardous components in solu-
The measurement of the maximum permeation rate tion. For industrial gloves, the permeation of 100%
demands that additional factors that can restrict the IPA has been frequently studied, but only a few
mass transfer must be minimized. A slow desorption permeation tests have been reported for surgical or
rate of a chemical from the material surface decreases examination gloves. BTTs as low as 1 min have been
the permeation rate. The desorption rate can in some reported (Mellström et al., 1992; Forsberg and Keith,
cases be improved by increasing the flow rate of the 1999). Due to the high expected permeability, short
gaseous collection medium (Schwope et al., 1988b; BTTs and the common use of diluted IPA, we
Que Hee, 1996; Anna et al., 1998). On the other selected 70% IPA for use in these tests instead of
hand, this leads to dilution of the test chemical in the 100%.

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


collection medium, thus affecting the detection limit The permeation of 70% IPA was studied through
of the permeation rate. There are a few previous single layer natural rubber (NR), double layer NR
studies about the influence of the flow rate on the and single layer chloroprene (CR) surgical gloves
permeation results (Mellström et al., 1989, 1991; according to the ASTM F739 and the EN 374 stand-
Mellström, 1991a,b,c; Zellers and Sulewski, 1993). ards. The permeation results were compared between
Only five tests in these studies (Anna et al., 1998) single and double layer gloves and gloves of different
have been carried out with the collection medium materials and thicknesses. As all the tests were
flow rates as defined by the two standards and the carried out by both standards, the effects of the
large standard test cell (glove sample diameter inside different flow rates could be evaluated for the perme-
the cell 5.1 cm). These studies clearly state that in ation of IPA.
some tests an increase in the flow rate can yield
higher SSP values, but the differences in BTTs have MATERIALS AND METHODS
not been considered significant. A theoretical evalu-
ation has shown that the BTTs are longer if the The seven brands of surgical gloves (Table 1) were
flow rates are increased (Schwope et al., 1988b). supplied by SSL International plc (Oldham, UK). Six
However, at the time of the Schwope study, calcula- of the glove materials had a polymeric Biogel® inner
tion of the BTT was based on the detection limit of coating. Two brands were double gloves with a dark
the analytical method. The current calculation is now green glove to be worn underneath a pale yellow one;
based on a fixed permeation rate (Pb) and thus a re- this allows holes in the gloves to be seen during
evaluation is needed. surgical operations. The test chemical was made by
Isopropyl alcohol (IPA) (2-propanol) is frequently mixing together isopropyl alcohol (p.a. grade; Merck)
used as a disinfectant or a solvent for other chemicals and HPLC grade water in the proportions 70:30 (v:v).
in a number of healthcare products. IPA can be an Test samples with an average diameter of 9.0 ±
irritant and can dry the skin or permeate through 0.1 cm were cut from the palms of the gloves. The

Table 1. Glove materials


Glove Material No. of test samples Thickness (mm) Weight per unit area
(mean ± SD) (g/m2) (mean ± SD)
Biogel® Super Sensitive NRa + BCb 40 0.22 ± 0.01 197 ± 6
Regent Surgical NR 40 0.28 ± 0.01 220 ± 7
Biogel® NR + BC 21 0.27 ± 0.02 244 ± 11
Biogel® Orthopaedic NR + BC 21 0.37 ± 0.01 302 ± 6
Biogel® Indicator NR + BC
Inner glove 21 0.22 ± 0.01 199 ± 5
Outer glove 21 0.22 ± 0.01 202 ± 5
Biogel® Reveal NR + BC
Inner glove 21 0.31 ± 0.01 251 ± 5
Outer glove 21 0.29 ± 0.01 241 ± 6
Skinsense™ N CRc + BC 40 0.19 ± 0.01 255 ± 8
aNaturalrubber.
bBiogel® coating.
cChloroprene rubber.
2-Propanol permeation through surgical gloves 307

diameters were measured in three directions for every first measuring point after which P has exceeded the
sample. Glove samples were weighed using an Pb.
analytical balance with a precision of 0.1 mg. The The ASTM standard requires the calculation of the
weight per unit area values were calculated. The average SSP, if the permeation rate reaches a
sample thicknesses were measured (Ames M034 E; constant level. The SSP is calculated from a few
Shirley Developments Ltd) using a pressure of 3.4 kPa points at the end of the three parallel tests.
and measuring diameter of 28.7 mm (Table 1). Three t-Tests (Microsoft® Excel 97 SR-2, T-test: Two-
values were measured for each sample. The samples Sample Assuming Equal Variances) with a signifi-
were not inspected for structural imperfections before cance level of 0.05 were applied for statistical
the chemical permeation testing took place. comparisons between the ASTM and the EN standard
As required by both standards, ASTM F739 and results or between results obtained with different
EN 374, a circular glove sample was mounted flow rates and the same permeation rate criteria.
between the two chambers of the standard test cell. Samples were injected (EC6W valve with valve
Collection medium was set to flow through the actuator E60; VICI Valco Instruments Co. Inc.,

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


chamber, which had the inner surface of the glove Houston, TX) from the selected collection medium
sample as one wall. Sampling of the collection medium flow line [six port valve SF6P (VICI Valco Instru-
for chemical analysis was then started. The test ments Co. Inc.) with a Smart Drive valve actuator
chemical was then injected into the chamber with the (Omnifit Ltd, Cambridge, UK)] into a gas chromato-
outer surface of the glove sample, at the starting time graph (Micromat HRGC 412, flame ionization
for the BTT determination. Sampling was terminated detector; HNU Nordion, Helsinki, Finland). Both
after 4 h, or earlier if the test chemical was found to valves were controlled by a computer program
permeate the glove material at a steady rate. For the written for the system (Valve Actuator Program, GC
EN standard, values in excess of a Pb of 1.0 µg/cm2/min Version 1.0, Matti Jussila). A HP-5 column (30 m ×
would have sufficed. The use of a more appropriate 0.53 mm, film thickness 2.65 µm; Hewlett Packard,
testing time considering the usage of disposable Little Falls, Wilmington, DE) was used at a constant
gloves, i.e. 4 instead of 8 h, was the only modification oven temperature of 45°C and a single chromato-
made to the standards in this study. The laboratory graphic run was used for the whole test. The sample
and the collection medium temperatures were kept at loop had been spliced from 1/16 inch and 1/8 inch
23.0 ± 1.0°C and the test chemical temperature was tubing (HNU Nordion) and its volume was ∼1 ml.
adjusted before the test to 23.0 ± 0.1°C using a water The flow rate of the helium carrier gas was 8.5 ml/min
bath. and an injection split of 1:3 was used (Pekari et al.,
1992). The system allows the testing of between one
BTT is the time between the application of the test
and three protective material samples at the same
chemical into the test cell and the time when the
time.
permeation rate exceeds a Pb of 0.1 µg/cm2/min
The collection medium was nitrogen gas, with flow
(ASTM) or 1.0 µg/cm2/min (EN). At least three
rates (Electronic flow meter 5182-0879; Hewlett
parallel tests were carried out and the final BTT result
Packard) of 52.0 ± 1.4 (ASTM) or 474 ± 2 ml/min
is the average of the three measurements. The perme-
(EN). The flows were split before the regulator valve
ation rate (P, µg/cm2/min) is calculated as follows in
to avoid a build up of pressure in the system. The
the open loop methods (i.e. without re-circulating the
splitting system consisted of plastic T-pieces and flow
collection medium).
restrictors (flow rate 474 ml/min, disposable 0.80 ×
38 mm needle; flow rate 52 ml/min, 0.45 × 12 mm
P = c × v/A (1) needle) for the free ends of the T-pieces. The tests of
rapid permeation (BTT < 20 min) were carried out
where c is the test chemical concentration (mg/l), v is with one glove sample at a time and the tests with
the collection medium flow rate and A is the exposed slower permeation with two glove samples at the
area of the test sample. According to the EN standard same time. In both cases the sampling frequency into
it is assumed that the entry of the test chemical into the gas chromatograph was once every 2 min. The
the collection medium is constant between two sampling programs included regular injections of
consecutive samplings. Thus the BTT (min) is calcu- pure nitrogen to ensure that the system had not
lated from the first degree equation: become contaminated.
For both the ASTM and EN methods, quantifica-
BTT = t1 + (t2 – t1) × (Pb – P1)/(P2 – P1) (2) tion was carried out using the same range of calibra-
tion samples. Isopropyl alcohol calibration samples
where t is the time from the injection of the test chem- were prepared in plastic laminated aluminum foil
ical into the test cell (min), index 1 refers to the first bags at concentrations of 30–1500 µg/l in nitrogen.
measuring point just before permeation exceeds the Calibration samples were introduced through the
Pb defined by the standard and index 2 refers to the valve system into the chromatograph with plastic
308 E. A. Mäkelä, S. Vainiotalo and K. Peltonen

Table 2. Isopropyl alcohol (70% v/v) breakthrough times measured for surgical gloves with two collection medium flow rates and
calculated at two permeation rate levels
Breakthrough time (min) (mean ± SD)
Standard used EN 374 ASTM F739
Flow rate (ml/min) 474 52 474 52
Permeation rate (µg/cm2/min1)a 1.0 0.1 0.1 1.0
Biogel® Super Sensitive 4.6 ± 0.6b 4.8 ± 0.9b 2.1 ± 0.1 9.6 ± 1.7
Regent® Surgical 6.5 ± 0.5b 6.5 ± 0.9b 3.4 ± 0.3 12 ± 1
Biogel® 7.6 ± 0.5b 9.1 ± 1.5b 3.5 ± 0.7 18 ± 3
Biogel® Orthopaedic 21 ± 2 12 ± 0c 9.8 ± 1.4c 25 ± 2
Biogel® Indicator 80 ± 13d 32 ± 2c 30 ± 5c 121 ± 26d
Biogel® Reveal 122 ± 6 38 ± 3c 37 ± 3c 153 ± 18
Skinsense™ N >240 103 ± 7c 97 ± 2c >240

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


The results which have been obtained under different test conditions and which do not differ statistically (P > 0.05) have been
indicated.
aThe level of the rate used for the determination of the BTT.
bThe ASTM result does not differ from the EN result statistically.
cThe result with the flow rate 52 ml/min does not differ statistically from the result with the flow rate 474 ml/min using

Pb = 0.1 µg/cm2/min.
dThe result with the flow rate 52 ml/min does not differ statistically from the result with the flow rate 474 ml/min using

Pb = 1.0 µg/cm2/min.

luer-lock syringes (10 ml, Discardit™ II; Becton of the EN standard in the 4 h test with the Skin-
Dickinson, Madrid, Spain). Regression analyses sense™ N gloves, but reached the ASTM Pb of
(Microsoft® Excel 97 SR-2) were applied to the cali- 0.1 µg/cm2/min after 103 min. The BTTs and the
bration responses and they produced linear lines corresponding standard deviations are shown in
which were forced to 0 (r = 0.999 ± 0.001, P = 3.2 ± Table 2. Figure 1 illustrates the increase in the perme-
0.1 × 10–15). Calibration was performed daily before ation rates during the tests. The points where the
testing and was checked after each test. curves cross the Pb of 1.0 µg/cm2/min can be
observed as the corresponding BTTs. In all of the EN
testing, the BTTs for the parallel tests were within the
RESULTS
required range ± 20% of the mean.
The thicknesses of the gloves were 0.22, 0.28, 0.27 The BTT results were calculated for both of the Pb
and 0.37 mm for the single layered NR gloves values and for both of the flow rates. When viewed
Biogel® Super Sensitive, Regent® Surgical, Biogel® with the same flow rate (Table 2 and Fig. 2), the Pb
and Biogel® Orthopaedic, respectively. Weight per of 1.0 µg/cm2/min yielded about double or higher
unit area values were 197, 220, 244 and 302 g/m2. BTT values than the Pb of 0.1 µg/cm2/min. When
The Biogel® Indicator and Biogel® Reveal gloves comparing the ASTM results with the corresponding
were double gloves, whose inner and outer gloves EN results, it was found that the results of the thin NR
had about the same thickness and weight per unit gloves (Biogel® Super Sensitive, Regent® Surgical
area. Both Indicator gloves were equal in their thick- and Biogel®) did not differ statistically even though
ness and weight per unit area to the Super Sensitive there was a 10-fold difference in their Pb values. On
gloves and both Reveal gloves were approximately the other hand, for the different flow rates but the
the same as the standard Biogel® gloves. The CR same Pb, a statistical difference was found between
gloves, Skinsense™ N, were the thinnest of all at the BTTs at the Pb of 0.1 µg/cm2/min for the above
0.19 mm, but they had a relatively high weight per mentioned thin NR gloves and at the Pb of 1.0
unit area value of 255 g/m2. Further information on µg/cm2/min for all the gloves except the Biogel®
the gloves is shown in Table 1. Indicator gloves.
The BTTs using EN 374 to determine the perme- SSP rates according to the ASTM standard were
ation of 70% IPA were 4.6, 6.5, 7.6, 21, 80 and 122 attained in the single layered NR testing. The average
min for the NR gloves Biogel® Super Sensitive, SSPs of 70% IPA were 4.2, 4.2, 3.1 and 2.7 µg/cm2/min
Regent® Surgical, Biogel®, Biogel® Orthopaedic, for Biogel® Super Sensitive, Regent® Surgical,
Biogel® Indicator and Biogel® Reveal, respectively. Biogel® and Biogel® Orthopaedic, respectively. The
Following the same order of increasing weight per remaining gloves did not reach the SSP during the 4 h
unit area, the ASTM test results for the NR gloves test. SSP values were also calculated for the tests
were 4.8, 6.5, 9.1, 12, 32 and 38 min. The permeation using the EN flow rate of 474 ml/min: the SSPs of
rate of IPA did not reach the Pb of 1.0 µg/cm2/min IPA were 7.2, 5.3, 5.3, 3.2 and 1.7 µg/cm2/min for
2-Propanol permeation through surgical gloves 309

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014

Fig. 1. Permeation of 70% isopropyl alcohol through surgical rubber gloves. The ASTM F739 test was made with a collecting
medium flow rate of 52 ml/min (crosses) and the EN 374 test with a rate of 474 ml/min (diamonds).
310 E. A. Mäkelä, S. Vainiotalo and K. Peltonen

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


Fig. 2. Breakthrough times of 70% isopropyl alcohol through surgical rubber gloves calculated for different flow rates and
breakthrough detection levels. The test conditions are indicated for each series of points as collection medium flow rate
(ml/min)/breakthrough detection level (µg/cm2/min). The error bars represent the standard deviations.

Table 3. Average (n = 3) steady-state permeation rates (SSPs) tion against hazardous chemicals, it should have a
for isopropyl alcohol (70% v/v) BTT of at least 10 min and twice the working time
SSP (µg/cm2/min) (mean ± SD) with the chemical. The difference between the
Standard used EN 374 ASTM F739 recommended time of use for the gloves and the BTT
Flow rate (ml/min) 474 52 arises from the differences in conditions between
Biogel® Super Sensitive 7.2 ± 0.8 4.2 ± 0.2 everyday actual use and the standardized test method
Regent Surgical 5.3 ± 0.3 4.2 ± 0.6 (Leinster et al., 1990; Leinster, 1994). Thus the tested
Biogel® 5.3 ± 0.2 3.1 ± 0.5
single layered NR gloves cannot be recommended
against 70% IPA solutions, except for occasional
Biogel® Orthopaedic 3.2 ± 0.0 2.7 ± 0.2
very brief tasks. Even then there should not be
Biogel® Indicator 1.7 ± 0.1 Not detected
dermally toxic substances present which could
The SSP values are required by the ASTM F739 standard but permeate along with the IPA. For a short period of
not by EN 374. time, the double layered and the CR gloves should
provide protection.
This study shows that flow rate is a critical factor in
Biogel® Super Sensitive, Regent® Surgical, Biogel®, tests in which a gaseous collection medium is used.
Biogel® Orthopaedic and Biogel® Indicator gloves, Compared with the flow rate of 474 ml/min, the
respectively (Table 3). The SSPs differed statistically lower flow rate of 52 ml/min yielded lower perme-
significantly for the tests with different flow rates. ation rates throughout the tests and thus made the
BTTs longer in five of the seven test pairs when
DISCUSSION detected at the Pb of 1.0 µg/cm2/min and in three of
the seven test pairs when detected at the Pb of 0.1
Surgical or other thin rubber gloves may often µg/cm2/min (Table 2). The differences in the BTTs
provide insufficient protection against small organic with the different flow rates at a Pb of 1.0 µg/cm2/min
molecules. If the glove is to have any protective func- were so marked that they would influence how three
2-Propanol permeation through surgical gloves 311

of the gloves (Regent® Surgical, Biogel® and Biogel® ness was the same as the Biogel® gloves. Thus there
Indicator) would be classified for protection against was no evidence that the Biogel® polymer layer made
IPA according to the EN standard. the NR material less permeable to IPA.
The BTTs for the three thin NR materials (Biogel® Skinsense™ N proved to be the best barrier against
Super Sensitive, Regent® Surgical and Biogel®) were IPA of the tested gloves. It was the thinnest of the
about the same when tested using the two different materials and the least permeable. CR surgical gloves
standards, despite the 10-fold difference in the Pb have also proved to be effective against meth-
values (Table 2 and Fig. 2). It seems that in the tests acrylates and formaldehyde (Schwope et al., 1988a;
of these three materials with the collection medium Mäkelä et al., 1999, 2003). However, the CR gloves
flow rate of 52 ml/min the vaporization of IPA meant for healthcare are only sold as sterilized
suffered from the low mixing efficiency in the test surgical gloves, although they could also have poten-
cell, which has been discussed by Anna et al. (1998). tial in providing protection against hazardous chem-
Thus their BTTs became about twice as long as those icals, e.g. for laboratory workers. At present their
measured with the higher flow rate, if calculated at price may hinder their more extensive use outside the

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


the same Pb. Statistical examination of the BTTs operating theatre.
calculated at the same Pb revealed that the BTTs of The BTT values for the double layered gloves were
the better barriers (Biogel® Orthopaedic, Biogel® about five times longer than for the single layered
Indicator, Biogel® Reveal and Skinsense™ N) did gloves of similar thicknesses in the ASTM test and
not differ between the two flow rates in all but two about 16 times longer in the EN test. Thus it can be
cases (Biogel® Orthopaedic and Biogel® Reveal, Pb concluded that irrespective of the testing method, the
1.0 µg/cm2/min). The slow permeation probably protection provided by double layered gloves against
allowed better mixing and vaporization. Thus, IPA is more than double the protection provided by
because of the difference in the Pb values, it was single layered gloves. This observation can be
possible that the ASTM method could yield clearly explained by the chemical having to adsorb, diffuse
shorter BTTs than the EN method for these good and desorb twice and the differences in the concentra-
barriers, as expected. tion gradients driving the chemical across the
Anna et al. (1998) examined the effect of the systems. The advantage of double gloving has been
collection medium flow rate in the ASTM F739 noted previously for protection against microorgan-
method. They performed 176 chemical permeation isms and chemicals (Mellström, 1991a; Gerberding
tests and concluded that since the SSP values et al., 1995; Heller and Greer, 1995; Page, 1997;
increase for many test chemicals when the flow rate Mäkelä et al., 1999, 2003).
is increased, the standard method must be changed to Even if the regulations require the use of a specific
determine the maximum SSP. They reported BTTs standard test method when certifying chemical pro-
for five chemical/glove pairs with collection medium tective gloves, they do not forbid the use of other
flow rates of 50 and 500 ml/min and noticed that the available test results when selecting the gloves. Both
higher flow rate appeared to shorten the BTT, the standards, ASTM F739 and EN 374, can be used
although they did not detect any actual statistically to differentiate totally unsuitable gloves from gloves
significant differences. The differences of our find- which provide adequate protection. Often there is a
ings compared with the Anna et al. study are due to shortage of information when gloves are to be
the different test chemicals and glove materials. In selected. Therefore, the instructions for use, in add-
other studies there is less evidence of the BTTs short- ition to the required standard information, should also
ening with an increase in the collection medium flow include information obtained with other test methods.
rate, but the tests have been performed under However, if the permeation test results are to be of
different test conditions (Mellström et al., 1989, value, the applied methods must also be described.
1991; Mellström, 1991a,b,c; Zellers and Sulewski, Both standards are meant to specify a test method
1993). In those studies a narrower range of collection which yields reproducible and comparable results
medium flow rates or much higher flow rates were that describe the resistance of a protective material to
used compared with the EN and the ASTM standard chemical permeation. The collection medium flow
test conditions, if the flow rates are calculated as the rate of the ASTM method is 100 ± 50 ml/min, which
test cell volume changes per minute. corresponds to 1.0 ± 0.5 chamber vol/min. This is
In the case of NR materials, the thicker the glove low and the range is also wide, which should be
material the longer the BTT values. The Regent® considered as these factors may influence the perme-
Surgical glove, which is manufactured differently, was ation results. On the other hand, the low flow rate
in line with the other gloves. Its SSP and BTT values and thus the high concentration of the test chemical
measured under the standard conditions did not differ in the collection medium allows the use of a low Pb
from the corresponding values of the Biogel® gloves. (0.1 µg/cm2/min). Thus permeation which cannot be
The weight per unit area was in between the Biogel® determined by the EN standard test can sometimes be
Super Sensitive and the Biogel® gloves and its thick- measured using the ASTM test. The higher flow rate
312 E. A. Mäkelä, S. Vainiotalo and K. Peltonen

of the EN test lessens the impact of the desorption editors. Protective gloves for occupational use. Boca Raton,
rate on the permeation results. The higher permeation FL: CRC Press, pp. 269–82.
Leinster P, Bonsall JL, Evans MJ, Lewis SJ. (1990) The appli-
rate produced by the higher collection medium flow cation of test data in the selection and use of gloves against
rate also results in a steeper slope of the permeation chemicals. Ann Occup Hyg; 34: 85–90.
rate versus time curve, which makes determination of Mäkelä EA, Väänänen V, Alanko K, Jolanki R, Estlander T,
the BTT more accurate. Thus the different conditions Kanerva L. (1999) Resistance of disposable gloves to perme-
of the two standards both have their advantages and ation by 2-hydroxyethyl methacrylate and triethyleneglycol
dimethacrylate. Occup Hyg; 5: 121–9.
without a large number of comparable test results, it Mäkelä EA, Vainiotalo S, Peltonen K. (2003) The permeability
is quite difficult to define which represents the better of surgical gloves to seven chemicals commonly used in
way of determining the BTT. hospitals. Ann Occup Hyg; 47: 313–23.
Mellström G. (1991a) Protective gloves of polymeric materials.
Experimental permeation testing and clinical study of side
Acknowledgements—The authors wish to express their grati- effects. Acta Derm Venereol Suppl (Stockh); 163: 1–54.
tude to Matti Jussila, the scientist, the technician and the Mellström GA. (1991b) Comparison of chemical permeation
computer programmer. The study was made possible by the data obtained with ASTM and ISO permeation test cells. I.

Downloaded from http://annhyg.oxfordjournals.org/ at Pennsylvania State University on March 3, 2014


kind support of SSL International plc, Oldham, UK. The ASTM standard test procedure. Ann Occup Hyg; 35:
153–66.
Mellström GA. (1991c) Comparison of chemical permeation
REFERENCES data obtained with ASTM and ISO permeation test cells. II.
The ISO/DIS standard test procedure. Ann Occup Hyg; 35:
ASTM. (1999) Standard test method for resistance of protec- 167–80.
tive clothing materials to permeation by liquids or gases Mellström GA, Landersjö L, Boman AS. (1989) Permeation of
under conditions of continuous contact, ASTM F 739-99a. neoprene protective gloves by acetone: comparison of three
West Conshohocken, PA: American Society of Testing and different permeation cells in an open-loop system. Am Ind
Materials. Hyg Assoc J; 50: 554–9.
Anna DH, Zellers ET, Sulewski R. (1998) ASTM F 739 Mellström GA, Landersjö L, Boman AS. (1991) Permeation
method for testing the permeation resistance of protective testing of protective gloves by using two different perme-
clothing materials: critical analysis with proposed changes in ation cells in an open-loop system (neoprene-toluene). Am
procedure and test-cell design. Am Ind Hyg Assoc J; 59: Ind Hyg Assoc J; 52: 309–14.
547–56. Mellström GA, Lindberg M, Boman A. (1992) Permeation and
European Committee for Standardization. (1994) European destructive effects of disinfectants on protective gloves.
standard EN 374-3: 1994. Protective gloves against chemi- Contact Dermatitis; 26: 163–70.
cals and micro-organisms—Part 3: Determination of resist- Page D. (1997) Study points to latex glove permeability.
ance to permeation by chemicals. Brussels: CEN. Nephrol. News Issues; 11: 28–30.
Fiserova-Bergerova V, Pierce JT, Droz PO. (1990) Dermal Pekari K, Vainiotalo S, Heikkilä P, Palotie A, Luotamo M,
absorption potential of industrial chemicals: criteria for skin Riihimäki V. (1992) Biological monitoring of occupational
notation. Am J Ind Med; 17: 617–35. exposure to low levels of benzene. Scand J Work Environ
Forsberg K, Keith LH. (1999) Chemical protective clothing— Health; 18: 317–22.
performance index, 2nd Edn. New York: John Wiley & Que Hee SS. (1996) Simple glove permeation models. Appl
Sons. Occup Environ Hyg; 11: 117–24.
Gerberding JL, Quebbeman EJ, Rhodes RS. (1995) Hand pro- Schwope A, Costas PP, Mond CR et al. (1988a) Gloves for pro-
tection. Surg Clin N Am; 75: 1133–9. tection from aqueous formaldehyde: permeation resistance
Heller ET, Greer CR. (1995) Glove safety: summary of recent and human factors analyses. Appl Ind Hyg; 3: 167–76.
findings and recommendations from health care regulators. Schwope A, Goydan R, Reid RC, Krishnamurthy S. (1988b)
South Med J; 88: 1093–8. State-of-the-art review of permeation testing and the inter-
Jensen O. (1981) Contact allergy to propylene oxide and iso- pretation of its results. Am Ind Hyg Assoc J; 49: 557–65.
propyl alcohol in a skin disinfectant swab. Contact Dermati- Zellers ET, Sulewski R. (1993) Modeling the temperature
tis; 7: 148–50. dependence of n-methylpyrrolidone permeation through
Leinster P. (1994) The selection and use of gloves against butyl- and natural-rubber gloves. Am Ind Hyg Assoc J; 54:
chemicals. In Mellström GA, Wahlberg JE and Maibach HI, 465–79.

You might also like