You are on page 1of 2

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

** Connecting Details TECHNICAL ROOM

PDB Positioner C2 Electronics Cabinet (positioner)


Bulkhead NOTE: Cable group either 4-2, either 4-3 and 4-4 could be extended with cable group 4-1 not shown.
20 EXAM ROOM CEILING 3/4 Monitors suspension
Room 3
Light
BULKHEAD SYSTEM
(Pre-Cabled)
Control TB3
21 Advantx LCA W/Bolus chasing 27083A P S Legend:
Ground Room 27682A S PDB On-Off CJ 512/1
A 14 Light
Switch
Dose Monitor Display
= Water Lines
A
(TSSC) 27686A S PDU On-Off CJ4 Downscan supply CJ1 27684A P S = Signal Cables W
15 S 10750A .
TB2 27685A V Downscan in CJ3
VCIM console CJ712 27676A S V = Video Cables
Shield 1 I = Internal Cable
27678A Without Fluoro UPS S
12 2
X-Ray ON RT Bus CJ301
13 3 Downscan out CJ2 27495A V
28259A
With Fluoro UPS
S H = High Voltage Cables
18 Shield 4
5 Live / Fluoro Monitor 34 = Cable Run Numbers
X-Ray Power ON S
Lamp 6 P = Power Cables
Modular PDU/C2 V 27563A VIDEO IN
19 Shield 7 C2 power 3 Phases
from PDB (For US only) 27660A
16 8 24 V G P Elegance power cable TB602 P Monitor Suspen Pwr Strip P = Field Supplied Cables
27661A
G = Ground Wires FS
X-Ray 9
3 Phase 400V and GND (Independent of
Lamp FL2/CB4 2 wires AC
Door G P 27675A P G
17 Switch 27680A power cable CB600 Power Cables)
11198A = MIS Numbers
C2 power 3 Phases
B from C1 (Not for US) P 27582A AC In /
P
Roadmap / Reference Monitor B
BULKHEAD TABLE

27683A S Debug J18 V 27564A VIDEO IN


or = Not included in GTC
100 wires signal table data #1 BJ400 27672A S
27679A S Ge Bus / Arcnet DJ23 *27673A - Not used for
100 wires signal table data #2 BJ201 27673A*
InnovaIQ Table
P or
Door Interlock TB2 25 wires UIF table power data BJ403 27674A P ECG Display Monitor (Optional) = Optional Selectable in GTC
See Connecting
24 wires Motor power AJ200
V 27568A Red Video Reserved
X Ray On Light TB2
Details ** 27669A P
15 wires signal cable AJ300 27670A S V 27569A Green Video Reserved CONTROL BOOTH
27681A G BOLT 100 wires signal cable data AJ 202 27671A V 27570A Blue Video Reserved Dose Monitor Control
Remote Access Modem Line
S 10750A DISPLAY
C Local Phone Walk Plug (DIRECT) 27587A V 27601A Video H Sync
C
S 27126A ION CHAMBER A
V 27602A Video V Sync

Hospital Main S 27126B ION CHAMBER B


Fluoro UPS I/F box (Option)
Fluoro UPS (Option) With Fluoro UPS
Power S 28259A J2 RT-bus C2 cabinet Bolus Wallbox
UPS control With Fluoro UPS DL Infrared ADVANTX LCA W/BOLUS
28257A S S 28257A J3 UPS control P S 27083A
3 x 380 to With Fluoro UPS Remote Receiver CHASING
S 28258A J4 PDB control
Ethernet 28262A S With Fluoro UPS 4-4 S 27581A S 27041A INNOVA W/BOLUS CHASING Control Booth Roadmap
480 VAC S 28260A J1 RT-bus C1 cabinet
Monitor
With Fluoro UPS
EPO/ON/OFF S G 28261A Bolt
(+Neutral for Fluoro 28256A
V 27500A VIDEO IN
UPS Option) AW Monitor #2
C1 Electronics Cabinet (system) P
Ethernet HUBS
V
Accessories Video Peripheral
D With Fluoro UPS
S 28262A Ethernet (Fluoro UPS) Ethernet RJ POS J5 27662A S
27732A
V
27495A VIDEO IN
D
Rectifier

By Pass

S
Output

P 27531 Mains HUB (UPS) RJ45 for FE Laptop J8 27505A S


EMI FILTER 27496A REMOTE CONTROL
(Fluoro UPS Option)
Modem with DC Bloc KVM
S 27587A Modem phone line J16 Keyboard, VGA, Mouse extender J6 27493A S AW Video Splitter
P G G P 27532 Mains Modem (non UPS) Live Monitor Option 27732A AW Workstation Central Unit
V
DL & Video Splitter and Fans 4-2 V P G 27501A VIDEO IN/AC/
DL video live monitor J26 27563A V
P G G P P P
V 27687A LCD Red J41
AW Power Strip P
DL video Rdp monitor J30 27564A V To 230Vac With keyboard and mouse
V 27688A LCD Green J42 4 OUTLETS
With Fluoro UPS
Fluoro UPS EPO/ON/OFF 28256A S LCD Blue J43
V 27689A DL Infra Red Receiver J15 27581A S V
Power V 27690A LCD H Synchro J44 Roadmap Monitor Option AW Workstation Monitor #1
Distribution 3rd Monitor Option J28 27501A V
E Box
PDB
control
28258A S
With Fluoro UPS
V 27691A LCD V Synchro J45 VGA
E
(Wall PDB) 4th Monitor Option J32 27502A V 4-3 V P G 27502A VIDEO IN/AC/
V 27685A P
VCR Video KX6 J24
Control Booth Power Strip
System ON/OFF Loop 27682A S P 27533 Mains Video splitter (non UPS)

P 27472A 5 OUTLETS P 29012A


P 27458 Mains DL (UPS)
Provided by installator
JEDI POWER 3 Phases P VCR Control J13 27496A S
P 29010A
P 27526 Mains Fan (UPS) 2
Provided by installator 27683A S Debug COM6 J40
Main Ground Wire G Video Main KX6 cable J27 27492A V P 29011A
27686A UPS Stop (Zephir Board)
DL Keypad J14 27494A S Secured Power 29024A
C2 power 3 Phases Provided by installator J10 GEMNET Plug Injector ACIST Injector/ 10906B/ Strip 3 RED OUTLETS P
(For US only) P G G
EXAM ROOM FLOOR LF Angio Illumina Pedestal Injector 10908B
PDU/C2 power 3 Phases RT Bus (Gitane Board) Video Slave KX6 Rdp J31 27500A V
(Not for US) Provided by installator
P G Universal Interface (LF Angio Illumina Pedestal Injector) 10907B S
PDU 2 Phases P
F (for US only) With Fluoro UPS
28260A RTAC F
Emergency UPS Stop 27561A S 29013A
Without Fluoro UPS Same as ECG Acquisition Head
27678A
S H1C RT Bus (Gitane board) J22
Fiber Channel J7 27512A S
Provided by installator 27679A S Ge Bus / Arcnet J23
S 27567A Data Reserved KVM - R Mouse
Chiller 3 Phase P G
27523A S Detector Chiller Control J17 Detector power CAN Bus J11 27514A S
1 S 27493A
P 27459 Mains RTAC (UPS) DL Keyboard
Vascular Table (Omega or Elegance)

Door Light S 27693 CAN & RT lines 15 pins J1


P 27660A Elegance power cable
Injector (J2) 10907B S DL Flat Panel
System ON 27661A
10906B/
JEDI S 27662A RJ45 Ethernet
L BUS BAR
10908B
G V 27687A J41 LCD Red
X-Ray tube stator cde DJ2 27664A S
* Connecting Details S 27693 CAN 15 pins J4 Innova 3100/4100 V 27688A J42 LCD Green
X-Ray tube temp switches DJ3 27665A G 27468A BUS BAR w/ Bolus chasing only 27041A P S P
S
G (Not for US)
27473A BUS BAR
V 27689A J43 LCD Blue G
3 Phase 400V Jedi X-Ray tube Ingrid DJ1 27663A ITU interface
TB02 V 27690A J44 LCD H Synchro
Anode I+ S 27672A J600 Table Data #1
3x 160A 27589A H J2 27452A P S J45 LCD V Synchro
AWG6
*27673A - Not used for V 27691A
27673A* J601 Table Data #2
2347573

AWG6 Cathode I- 27590A InnovaIQ Table


Jedi S 27677A X-Ray Tube Chiller Control DJ4
AWG6
RS232 J3 (Generic IF Board) 27701 P 27674A J605 Table Pwr Data
FUSE BOX LVPS 400 Relay command Oil Pump
PDB

P/N 2364241 Power 27675A AC IN / TB1 (1-2) L BUS BAR 27508A G Control Booth Live Monitor
STRIP UPS 27702
P
27694 P 27513A DET. POWER MAINS J1
16A
strip box V 27492A VIDEO IN
AWG10 AWG12
2315513

J47 Bulkhead S 27514A DET. PW CAN J2 L BUS BAR


JEDI RAIL 27470A G
2358531

AWG10 AWG12

STRIP NON UPS L BUS BAR


PDU P/N 2305956-2
28245
P 27722A DET. PW OUT J3 27471A
strip box VCIM On-Off Console with
AWG10
AWG10
AWG10

27550 27549

Opto-
Handswitch
H PDU & BUS BAR
Bridge 28254A S
LC Gantry S 27676A
H
J6 J2B Signals and 24Vdc
C1-
C1-CABINET G 28261A BUS BAR J5 Not Used J8 Not Used
27586A DETECTOR
Non UPS UPS
2350103 P 27586A Detector Supply L BUS BAR 27470A G
G 27468 BUS BAR
DL Keypad
AWG10

AWG10

AWG10

20A
J4A UPS Downscan supply 27684A P S 27512A Optic fiber L BUS BAR 27471A
w/ bulkhead
G BUS BAR from PDB
27722A S 27494A Serial Link with Power
J3A Detector AC 27513A P Detector water
Positioner power supply W 27511A
pipe in flow
C2 power 3 Phases SERVICE PWS J15
(Not for US) Detector water L BUS BAR 27030 G
C2-
C2-CABINET BUS BAR 27468A G 27510A
pipe out flow
27680A 2 Phase PDU
See Connecting Details * 27473A FE Laptop
BUS BAR
P TB02 3 Phase 400V CB1 TUBE

[
* Connecting Details X-Ray tube water S 27505A NO BULKHEAD (Free)
(For US only) J4B UPS/J2A Strip AC
27472A P G W 27666A
pipe out flow GND 27030 G
BUS BAR
S 27561A Emergency UPS Stop J17
Control Booth Power Strip 27667A X-Ray tube water
pipe in flow

[
Remote Box . Option
G 27681A BUS BAR J9 Strip AC P G
BUS BAR 27582A H 27589A ANODE I+
J TB02

3x 160A 3Ph to C2
Ceiling Power
27590A CATHODE I- P S 27452A J1
ECG Control Unit (Optional) J
AWG6 J1A 3rd MON OPTION BUS BAR 27501A P G
S S 27560A Data Reserved
2347573

AWG6
S 27686A 27664A X-ray tube stator cde J1
Jedi 27686A PDU ON/OFF J20
AWG6
J1B 4th MON OPTION BUS BAR 27502A P G 27665A X-ray tube temp status
V 27555A Red Video Reserved
FUSE BOX
P 27522A
Detector Conditioner Power J3B
V 27556A Green Video Reserved
G
PDB

P/N 2364241
27663A X-Ray tube ingrid 27508A
V 27557A Blue Video Reserved
16A
P 27669A J100 Power
AWG10 AWG12 motor V 27558A H SYNC Video Reserved
2315513
2358531

AWG10 AWG12
Detector Conditioner S 27670A J206 Collimator
V 27559A V SYNC Video Reserved
PDB

PDU P/N 2305956 - 2


WATER INLET 27510A W 27671A J205 Data LC
P 27522A Chiller AC power 3 wires
AWG10

AWG10

3 WATER OUTLET 27511A S 27126A ION CHAMBER A


C1 - CABINET S 27523A Chiller control RS232

S 27126B ION CHAMBER B


CAUTION: For proper operation and regulatory compliance, it is mandatory to follow the guidelines
K described in direction 46-019668, Advantx System Facility Power and Ground Requirements K
20A
X-Ray Tube Water Chiller RS232 28254A S
Positioner S 27677A Remote Control WATER OUTLET 27666A W
REVISIONS
MADE
APPROVALS YY/MM/DD TITLE Innova 2100IQ, 3100/3100IQ, 4100/4100IQ REV
REV MADE BY DATE PCN NO. DESCRIPTION 02/05/31
1 THEBAULT. C March 5, 2009 - SPR EURge88651
E. de SCHUTTELAERE System MISMAPS with CRT Monitors
C2 - CABINET G P 1TB - AC POWER & GROUND WATER INLET 27667A
NOTE: All sheets of this schematic will be at the same revision of this sheet. 2 .
CHECK
R. COLLOT - FIRST MADE FOR: Suspension option 1
3 APPR
-
The revision of this sheet will be the revision of the entire document. GE MEDICAL SYSTEMS 5160946-6-1EN TPH
4
OTHER BUC 1 2
- - SHEET OF

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
** Connecting Details TECHNICAL ROOM
C2 Electronics Cabinet (positioner) NOTE: Cable group either 4-2, either 4-3 and 4-4 could be extended with cable group 4-1 not shown.
PDB Positioner Connection 3/4 LCD Monitor suspension
Bulkhead 3 EXAM ROOM CEILING
Room
20 BULKHEAD SYSTEM
24 m 36 m Box
Light Advantx LCA W/Bolus chasing 27083A P S
Control TB3 27682A S PDB On-Off CJ 512/1 28172 28173
DOSE DISP 2
S Dose Monitor Display 2 Legend:
21
Downscan supply CJ1 27684A P
A 14
Ground Room
Light 27686A S PDU On-Off CJ4
.
A
Switch
S W = Water Lines
(TSSC)
V VCIM console CJ712 27676A S 29183 28194
DOSE DISP 1
S Dose Monitor Display 1 = Signal Cables
15 27685A Downscan in CJ3
TB2 .
Without Fluoro UPS V
Shield 1 27678A S Downscan out CJ2 27495A V = Video Cables I = Internal Cable
12 2
X-Ray ON With Fluoro UPS RT Bus CJ301 BW 1 Location
13 3
28259A S
3 mon 28066 28067 BW-1 H = High Voltage Cables
18 Shield 4 Modular PDU/C2 V Live Fluoro Monitor
5
C2 power 3 Phases
from PDB (For US only) 4 mon 28070 28071
34 = Cable Run Numbers
X-Ray Power ON
Lamp 6 G P Elegance power cable TB602
27660A
P 2nd shielding
P = Power Cables
3 Phase 400V and GND 27661A
19 Shield 7 SUSPENSION
GROUND
16 8 24 V 27680A G P FL2/CB4 2 wires AC
27675A P 3 / 4 mon 28189 28190 G = Ground Wires FS = Field Supplied Cables
9 power cable CB600 G
X-Ray
Lamp
C2 power 3 Phases
from C1 (Not for US)
4-4 MONITOR (Independent of
Door POWER
17 Switch 3 / 4 mon 28186 28187 P Monitor Suspen Pwr Strip Power Cables)
11198A = MIS Numbers
BULKHEAD TABLE
B BW 2 Location B
27683A S Debug J18 3 mon 28068 28069 BW-2
V Roadmap Monitor
100 wires signal table data #1 BJ400 27672A S 4 mon 28072 28073
27679A S Ge Bus / Arcnet DJ23
*27673A - Not used for 2nd shielding or = Not included in GTC
100 wires signal table data #2 BJ201 27673A*
IQ
Innova Table BW 3 / Color 1 Location
Door Interlock TB2 See Connecting 25 wires UIF table power data BJ403 27674A P HOSPITAL 3 / 4 mon POWER ECG
28120 28121 P ECG Display Monitor OR
Details ** UPS LINE
X Ray On Light TB2
24 wires Motor power AJ200 27669A P or = Optional Selectable in GTC
ECG Control 3 mon 28045 28046
Unit COLOR 1
15 wires signal cable AJ300 27670A S OR V AW Monitor OR
27681A G BOLT WARNING: Prior to AW Video
100 wires signal cable data AJ 202 Splitter 4 mon 28047 28048
Remote Access Modem Line
27671A each installation, CONTROL BOOTH
ensure that the ECG BW-3 Dose Monitor Control
DL Video
4 mon 28076 28077 V
BW3 Monitor - 4TH 29183
Local Phone Walk Plug (DIRECT) 27587A power cable is Splitter
S DISPLAY
J32 / J1B Option 28194
connected to a line that
C is protected against
nd
2 shielding
S 27126A ION CHAMBER A
C
short circuit hazards,
Fluoro UPS I/F box (Option) according to local BW 4 / Color 2 Location S 27126B ION CHAMBER B
Hospital Main Fluoro UPS (Option) With Fluoro UPS
S 28259A J2 RT-bus C2 cabinet regulations and that HOSPITAL 28120 28121
POWER ECG
P
4 mon ECG Display Monitor OR
With Fluoro UPS UPS LINE Bolus Wallbox
the ECG monitor is
Power UPS control 28257A S
With Fluoro UPS
S 28257A J3 UPS control
powered according to
ECG Control ADVANTX LCA W/BOLUS
S 28258A J4 PDB control Unit
4 mon
COLOR 2
V P S 27083A CHASING
OR 28049 28050
3 x 380 to Ethernet 28262A S With Fluoro UPS
S 28260A J1 RT-bus C1 cabinet
IEC60601-1 AW Video AW Monitor OR
requirements (leakage Splitter 27041A INNOVA W/BOLUS CHASING
S Control Booth Roadmap
480 VAC EPO/ON/OFF 28256A S
With Fluoro UPS
G 28261A Bolt current and BW-4
DL Video BW4 Monitor - 3RD Monitor
(+Neutral for Fluoro grounding). Splitter 4 mon 28074 28075 V
J28 / J1A Option V 27500A VIDEO IN
UPS Option) 2nd shielding
C1 Electronics Cabinet (system)
3 / 4 mon
28195 28196
I/R RECEIVER
S
DL Infrared
Ethernet HUBS
With Fluoro UPS
Ethernet RJ POS J5 27662A S Remote Receiver Accessories Video Peripheral
D S 28262A Ethernet (Fluoro UPS J48)
V D
Rectifier

By Pass

27495A VIDEO IN
Output

P 27531 Mains HUB (UPS) RJ45 for FE Laptop J8 27505A S


CAUTION: The second shieldings of BNC cables (BW-1 through BW-4) are connected as follows: S
EMI FILTER - Cabinet side: connected to the Ground BAR. 27496A REMOTE CONTROL
(Fluoro UPS Option) Modem with DC Bloc KVM - Suspension side: connected to the connection Box at the point labeled “Suspension Ground”.
AW Video Splitter
27587A S Modem phone line J16 Keyboard, VGA, Mouse extender J6 27493A S 28045 ...
...28050
P G G P 27532 Mains Modem (non UPS)
AW Workstation Central Unit
Live Monitor Option V
DL & Video Splitter and Fans 28066/28067
DL video live monitor J26
28070/28071
V With keyboard and mouse
4-2 V P G 27501A VIDEO IN/AC/
P G G P P P V 27687A LCD Red J41
28068/28069 AW Power Strip P
DL video Rdp monitor J30 V
V 27688A LCD Green J42 28072/28073 To 230Vac
4 OUTLETS
LCD Blue J43 28195
With Fluoro UPS V 27689A DL Infra Red Receiver J15
28196 S AW Workstation Monitor #2
Fluoro UPS EPO/ON/OFF 28256A S
V 27690A LCD H Synchro J44 27501A VGA
3rd Monitor Option J28 V
Power 28074/28075 Roadmap Monitor Option
E Distribution PDB With Fluoro UPS V 27691A LCD V Synchro J45 P
E
Box control
28258A S
27502A 4-3
(Wall PDB) V 27685A VCR Video KX6 J24 4th Monitor Option J32
28076/28077 V V P G 27502A VIDEO IN/AC/ AW Workstation Monitor #1
Control Booth Power Strip V VGA
P 27533 Mains Video splitter (non UPS)
System ON/OFF Loop 27682A S
P
P 27458 Mains DL (UPS)
P 27472A 5 OUTLETS P 29012A
Provided by installator VCR Control J13 27496A S
JEDI POWER 3 Phases P P 27526 Mains Fan (UPS) P 29010A
2
27683A S Debug COM6 J40 Video Main KX6 cable J27 27492A V
Provided by installator 29011A
G UPS Stop (Zephir Board) P
Main Ground Wire 27686A
DL Keypad J14 27494A S Secured Power
J10 GEMNET Plug
C2 power 3 Phases Provided by installator Strip 3 RED OUTLETS P 29024A
(For US only) P G Video Slave KX6 Rdp J31 27500A V EXAM ROOM FLOOR Injector ACIST Injector/ 10906B/
RT Bus (Gitane Board) G
PDU/C2 power 3 Phases LF Angio Illumina Pedestal Injector 10908B
(Not for US) Provided by installator With Fluoro UPS
PDU 2 Phases P G 28260A 10907B S
F (for US only) RTAC Universal Interface (LF Angio Illumina Pedestal Injector)
P F
Without Fluoro UPS
Same as
Emergency UPS Stop 27561A S 27678A
S H1C RT Bus (Gitane board) J22 29013A

27679A S Ge Bus / Arcnet J23


Fiber Channel J7 27512A S ECG Acquisition Head
Provided by installator
P G Detector Chiller Control J17 S 27567A Data Reserved KVM - R Mouse
Chiller 3 Phase 27523A S Detector power CAN Bus J11 27514A S
P 27459 Mains RTAC (UPS) S 27493A
1 DL Keyboard
Vascular Table (Omega or Elegance)
S 27693 CAN & RT lines 15 pins J1
Door Light P 27660A Elegance power cable
27661A DL Flat Panel
System ON JEDI X-Ray tube stator cde DJ2 27664A S S 27662A RJ45 Ethernet Injector (J2) 10907B S V 27687A J41 LCD Red
S 27693 CAN 15 pins J6 X-Ray tube temp switches DJ3 27665A
* Connecting Details S 10906B/ V 27688A J42 LCD Green
G 27468A BUS BAR L BUS BAR G
G (Not for US)
P X-Ray tube Ingrid DJ1 27663A
10908B
V 27689A J43 LCD Blue
P G
27473A BUS BAR
TB02 Innova 3100/4100 27041A P S
Anode I+ 27589A H w/ Bolus chasing only V 27690A J44 LCD H Synchro
3x 160A S 27672A J600 Table Data #1
AWG6
Cathode I- 27590A *27673A - Not used for ITU interface V 27691A J45 LCD V Synchro
27673A* J601 Table Data #2
2347573

AWG6
Jedi S 27677A X-Ray Tube Chiller Control DJ4 RS232 J3 (Generic IF Board) InnovaIQ Table
(Innova 2100-IQ only) 27701
AWG6
P 27674A J605 Table Pwr Data J2 27452A P S
LVPS 400 Relay command Oil Pump
FUSE BOX Power
PDB

P/N 2364241 27675A AC IN / TB1 (1-2)


STRIP UPS
27694 27702 L BUS BAR 27508A G Control Booth Live Monitor
P 27513A DET. POWER MAINS J1 P
16A strip box
J47 Bulkhead
AWG10 AWG12 JEDI RAIL V 27492A VIDEO IN
2315513

S 27514A DET. PW CAN J2 L BUS BAR 27470A G


2358531

AWG10 AWG12
STRIP NON UPS
28245 L BUS BAR
PDU P/N 2305956-2
strip box P 27722A DET. PW OUT J3 27471A
27550 27549
AWG10
AWG10
AWG10

27722A
VCIM On-Off Console with
PDU & BUS BAR Opto-
Bridge 28254A S Handswitch
H 28261A J6 J8 J2B LC Gantry H
G BUS BAR J5 Not Used UPS Not Used S 27676A Signals and 24Vdc
C1-
C1-CABINET Non UPS 28189/ 27586A DETECTOR
SUSPENSION 28190 G
2350103 G 27468 BUS BAR P 27586A Detector Supply L BUS BAR 27470A G
J4A Downscan supply 27684A
AWG10

AWG10

AWG10

20A S 27512A Optic fiber L BUS BAR 27471A DL Keypad


G BUS BAR from PDB w/ bulkhead
J3A Detector AC 27513A P
power supply Detector water S 27494A Serial Link with Power
C2 power 3 Phases SERVICE PWS J15 W 27511A
Positioner (Not for US) pipe in flow
BUS BAR 27468A G Detector water L BUS BAR 27030 G
C2- 2 Phase PDU 27510A
C2-CABINET 27680A
See Connecting Details * pipe out flow
BUS BAR 27473A
P TB02 3 Phase 400V CB1
TUBE FE Laptop

[
* Connecting Details J4B UPS/J2A Strip AC
BUS BAR 27472A P G W 27666A X-Ray tube water
GND
(For US only) pipe out flow 27030 G S 27505A NO BULKHEAD (Free)
S 27561A Emergency UPS Stop J17 Control Booth Power Strip

[
27667A X-Ray tube water Remote Box . Option
J9 Strip AC 28186/ pipe in flow
G 27681A BUS BAR BUS BAR P G
28187 H 27589A ANODE I+ P S 27452A J1
J TB02
3Ph to C2
Ceiling Power
ECG Control Unit (Optional)
J
3x 160A 27590A CATHODE I-
AWG6
S 27686A J1A 3rd MON OPTION BUS BAR
27501A P G G 27508A
27686A PDU ON/OFF J20 S S 27560A Data Reserved
2347573

AWG6 27664A X-ray tube stator cde J1


Jedi
AWG6
P 27522A J1B 4th MON OPTION BUS BAR 27502A P G 27665A X-ray tube temp status
Detector Conditioner Power J3B 24 m 36 m
FUSE BOX
PDB

P/N 2364241
27663A X-Ray tube ingrid SUSPENSION 3 mon 28045 28046
CONNECTION
BOX
16A
P 27669A J100 Power COLOR 1 4 mon 28047 28048 V VGA
AWG10 AWG12 motor
2315513

Detector Conditioner
2358531

AWG10 AWG12
S 27670A J206 Collimator
PDB

PDU P/N 2305956 - 2


WATER INLET 27510A W J205 Data LC SUSPENSION 4 mon 28049 28050
P 27522A Chiller AC power 3 wires 27671A CONNECTION
BOX
AWG10

AWG10

WATER OUTLET 27511A COLOR 2


S 27523A Chiller control RS232 S 27126A ION CHAMBER A
C1 - CABINET
S 27126B ION CHAMBER B
K 3 CAUTION: For proper operation and regulatory compliance, it is mandatory to follow the guidelines K
20A
X-Ray Tube Water Chiller described in direction 46-019668, Advantx System Facility Power and Ground Requirements
RS232 28254A S
Positioner S Remote Control
REVISIONS APPROVALS YY/MM/DD TITLE Innova 2100IQ, 3100/3100IQ, 4100/4100IQ REV
27677A WATER OUTLET 27666A W REV MADE BY DATE PCN NO. DESCRIPTION MADE
02/05/31
1 THEBAULT.C March 5, 2009 -
E. de SCHUTTELAERE System MISMAPS with LCD Monitor
C2 - CABINET G P 1TB - AC POWER & GROUND WATER INLET 27667A NOTE: All sheets of this schematic will be at the same revision of this sheet. 2
SPR EURge88651 CHECK
R. COLLOT - FIRST MADE FOR: Suspension option 1
3 APPR
-
The revision of this sheet will be the revision of the entire document. GE MEDICAL SYSTEMS 5160946-6-1EN TPH
4
OTHER BUC 2 2
5 - - SHEET OF