Optima XR646 HD

:Ceiling Height
Recommended Wall Circuit-Breaker Ratings
Voltage/Power 50KW 65KW 80KW Cable Size
Optima XR646 380 V 55A/600V 74A/600V 95A/600V 4 X16 mm  or 4 X 25 mm2
2

HD

Discovery IGS730

Exam Room Layout:

Configuration Length Width


Minimum 6600 mm (260 in) 5200 mm (205 in)
Typical 8500 mm (335 in) 7000 mm (276 in)
Maximum 10000 mm (393.7 in) 8000 mm (315 in)

The minimum door height (to accommodate Gantry on its dolly) is 2060 mm

Table Base Plate:


Ceiling CMS Plate :

Floor Duct:
Rails for Mavig suspension:

Cable Size and circuit breaker:

Cable # From To Size Circuit Breaker


1 Mains EMI Filter 5 X 35 mm²
2 EMI Filter PDB 5 X 35 mm²
3 PDB Fluoro UPS 5 X 10 mm²
100 A
4 Fluoro UPS PDB 3 X 10 mm²
5 PDB C1 cabinet 4 X 6 mm²
6 PDB Tube Chiller 4 X 6 mm²
Discovery RF180

Table – delivered in one piece


Table – delivered in two pieces
Room Dimensions

Recommended Value of the ceiling height From 300 to 319


Base Plate :

1-2-3-4-5-6-7-8-9: Standard fixing point.


a-b-c-d: Fixing point with adjusting washers.
(*) Cable in point.
Cable Size and circuit breaker:

Cable # From To Size Circuit Breaker


1 PDB Table Power supply cable 4 x 6 mm²
80A
2 PDB Generator PWS cable 4 x 16 mm²

Senographe Pristina
Electrical requirements for other X-Ray systems:

Systems Electrical specifications


Systems Voltage Cable Size Power Circuit Breaker UPS
20 KVA comes with
Cath-Lab 380 V to 400 V In System Section 150 KVA In System Section
the system
Mammography 220 V 2 X 4 mm2  5 KW 32 A 10KVA
Depends on
In System
Fluoroscopy 380 V to 400 V In System Section In System Section Generator Power
Section
selection
Depends on
Optima XR646 In System
In System Section In System Section In System Section Generator Power
HD Section
selection

GE Recommended Ground :

The ground wire impedance from the system disconnect (including the ground rod) measured to
earth, must not exceed 2 ohms