You are on page 1of 2

Sabre 180

Electrosurgical Generator

Sabre 180 shown with CONMED


®

1000 SES™ Smoke Evacuation System


Sabre 180 Electrosurgical Generator
®

Power Operating Frequency: Pure Cut, Blend 1, Blend 2: 417 KHz,


The Sabre 180 provides advanced electrosurgical features and performance Mono Coag: 540 KHz, Bipolar: 1.05 KHz
in a compact and affordable electrosurgical generator package. The Aspen
Return Monitor (ARM®) combines automatic feedback cirtuitry Pulse Repetition Frequency: Blend 1: 20 KHz, Blend 2: 20 KHz,
and visual displays to provide the most sensitive contact quality Coag: 33 KHz
monitoring system available today. In Pad Monitoring Mode, the ARM®
system continuously monitors the safety of the pad placement and automatically Input Voltage: 104-127 VRMS, 50/60 Hz, Consult CONMED for
disables electrosurgical output and alerts the surgical staff if a return fault other available voltages
condition occurs. The Assist Bargraph provides a visual display of the
resistance at the pad site and the manual Alarm Set Point eliminates Line Frequency (Per UL 544): Patient connected too neutral: less than 10
the possibility of pad site burns attributable to machine error. μA typical. Chassis to neutral: less than 25 μA typical.

Designed for Performance RF Isolation: Less than 100 mA RF Leakage

The Sabre 180 delivers impressive electrosurgical performance in both the CUT Return Monitor System: With Dual Foil Pads, the Aspen Return
and COAG Modes. The Sabre 180 delivers a thermal damage free CUT with Monitor will alarm above 150Ω or below 10Ω, and with an
patented Controlled Output Voltage. As the electrode glides through approximate 20% increase in resistance above the set point.
different tissue types in the CUT Mode, Sabre 180 constantly adjusts the With a Single Foil Pad, a two wire continuity detector alarms
voltage to provide superior cutting performance. Surgeons can select either above 10Ω.
Pure CUT or one of the 2 Blend Modes that add hemostasis while
dissecting. The Enhanced Bipolar Mode provides excellent bipolar Power Cord: AWG #18-3 9.3 ft. (3m) Detachable
coagulation with minimal tissue sticking.
Dimensions: Weight: 20lb (9.1kg), Height: 6 in. (153mm)
Designed for Ease of Use Width: 12 in. (305 mm), Depth: 16 in. (407 mm)

The Sabre 180 is designed for simple set up and use. The ergonomic touch Warranty: 2 Years, Parts and Labor. 7 Year Extended
panel is simple to set up and adjust, and the Previous Mode Memory Warranty Available.
stores the last set of power, mode and return pad selections to facilitate
user set up.
Ordering Information
Sabre 180 electrosurgical
®
60-5800-001 Sabre® 180 Electrosurgical Generator with 9.8 foot (3m)
Generator Specifications Power Cable and Operator’s Manual.
60-5480-001 Universal ESU Cart
Max. 60-5104-001 Monopolar Footswitch, Explosion Proof, Splash Proof,
Mode Max. Peak to Peak Crest with 10’ (3m) Cable.
Power Voltage Factor Load 60-5801-001 Sabre 180 Operator’s Manual
(Watts) (Volts) (Ohms) 60-5803-001 Sabre 180 Service Manual
60-6850-001 CONMED 1000 SESTM Smoke Evacuation System
Pure Cut 100 1,600 1.8 500 with 12’ (3.6m) Power Cord
60-6850-001 CONMED 1000 SES™ Smoke Evacuation System
Blend 1 65 1,650 2.3 500
with 12’ (3.6m) Power Cord
Blend 2 50 1,700 3.5 500
60-6875-001 Filter Cartridge
Mono Coag 80 6,500 7.0 500
60-6886-003 Remote Activator
Bipolar Coag 50 400 2.4 50
*Bipolar Coag crest factor decreases with increased power settings.

+U.S. Patent Nos. 4,569,345 & 4,617,927


++U.S. Patent Nos. 4,848,355 & 4,961,739
Specifications subject to change without notice.

For International orders or inquiries, please


contact CONMED International Sales
(315) 797-8375 • FAX (315) 735-6235

© CONMED Corporation, 7/05, printed in U.S.A., Control #7660 Rev. C