BD Kiestra™ TLA (Total Lab Automation)

Technical Specifications
BD Kiestra™ TLA (Total Lab Automation)
Technical Specifications
SorterA/ ReadA Reading
BarcodA InoqulA+™ TLA Ergonomic A Compact ProceedA Workstation SCU**
Max Heat *1365 *2730 *1365 *2730 NA *1365 *3750
Production (BTU/hr)
*Approximately
Electrical 208–240 Vac 208–240 Vac 208–240 Vac 208–240 Vac Dependent on 208–240 Vac 208–204 Vac
Requirements 50–60 Hz 50–60 Hz 50–60 Hz 50–60 Hz customizable track 50–60 Hz 50–60 Hz
1x16A 1x16A 1x16A 1x16A configuration 1x16A 2x group 16A
Power Consumption (W) 400 800 400 800 NA 400 1100
Pneumatic >8 bar minimum, >8 bar minimum, N/A >8 bar minimum, >8 bar minimum, N/A N/A
Requirements 100 L /min minimum 250 L /min minimum 80 L /min minimum 50 L /min minimum
Total pneumatic Moisture and oil-free Moisture and oil-free Moisture and oil-free Moisture and oil-free
requirements ISO8573-1 ISO8573-1 ISO8573-1 ISO8573-1
dependent on Preferably Class IV Preferably Class IV Preferably Class IV Preferably Class IV
system configuration
Space Requirements: *SorterA Configuration 2659 mm / 105 in ErgonomicA 1000 mm / 40 in Dependent on Stand-Alone 750 mm /
Width Range: 18–3 to 24–6 165: 1650 mm / 65 in customizable track ErgonomicA 29.5 in
*Size dependent 2281 mm–2720 mm 200: 2000 mm / 79 in configuration 165: 1650 mm / 65 in
on configuration 90 in–107 in 220: 2200 mm / 87 in 200: 2000 mm / 79 in
220: 2200 mm / 87 in
Space Requirements: *SorterA Configuration 933 mm / 37 in 850 mm / 33.5 in 1594 mm / 63 in Dependent on 850 mm / 33.5 in 1130 mm /
Depth Range: 18–3 to 24–6 customizable track 44.4 in
*Size dependent 830 mm–1380 mm configuration
on configuration 33 in–54 in
Space Requirements: 1950 mm / 76.77 in 1950 mm / 76.77 in 1950 mm / 76.77 in 2300 mm / 76.55 in 1950 mm / 76.77 in 1950 mm / 76.77 in 1015 mm /
Height (with signaling lamp) (with signaling lamp) (with signaling lamp) (with signaling lamp) (with signaling lamp) 39.9 in
2196.66 mm / 86.48 in
with biosafety hood
Weight *300–500 kg / *700 kg / 1543 lbs *130–160 kg / *500kg / 1100 lbs *500–700 kg / *150 kg / 330 lbs *262 kg / 577 lbs
*Approximately 660–1100 lbs 285–350 lbs 1100–1540 lbs
**All Configurations come with a System Control Unit (SCU).

Media Plate Specifications IT Requirements Environmental Conditions


Diameter (bottom) 85–91 mm 2x 1GBps CAT 6 - preferably load balanced. Installation Category II and
Diameter (lid) 89–93 mm Remote access through BD Assurity Link™. Pollution Degree 2 as per IEC 664
Height (including lid) 13–16.2 mm Access to ports 443; 8531; 10050; 10051.
Noise Level
*All Configurations come with a System Control
Sample Container Specifications* Unit (SCU) <67 dBA
Container Diameter 10–50 mm
LIS Interface Environmental Requirements
Container Height 60–125 mm
Bidirectional interface based on ASTM Clearance for Operator Service:
Cap/Lid Diameter 10–54 mm
protocol Floor: 800 mm / 31.5 in arround all sides
Cap/Lid Height 11–30 mm
Height: minimal floor height
*Some exceptions apply. Please check with your BD ReadA Compact Incubation
representative for additional container compatibility 2780 mm / 109.5 in
information. Specifications Temperature Requirements:
Incubation Atmosphere: 02 or CO2
Biosafety Cabinet 18°C–27°C / 64.4°F–80.6°F
Temperature Range: 30°C–40°C ± 1°C CO2
Requires additional electric power Relative Humidity:
Consumption: 50 L/h per ReadA Compact
208–230V, +/ – 10% 20%–80% non-condensing
Note: Air compressors are not included in equipment
13A / 50–60Hz, Max 16A configurations and need to be ordered separately. Operating Elevation: 0–2000 m

BD, Sparks, MD 21152-0999 USA


Tel: 800.638.8663

bd.com
© 2016 BD. BD, the BD Logo and all other trademarks are property of Becton, Dickinson and Company.
3-223576 May 2017 Printed in the U.S.