You are on page 1of 12

Released

Augest. 2010

AXIOM Iconos MD
SSME AX

Proposals for
Organ Program Settings

with
FLUOROSPOT Compact
Version VD 45 and higher
Last change Augest. 2010
Status: released
(C) 2010, s AG

Augest. 2010

Print No.: AXD3-360.820.02.02.02 General notes AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 1 of 12 valid for SW VD onwards
History
Document History

Version Date of Issue Author Change & Reason of Change


Change Request
01 3/29/2007 Guo Qiang Document creation AX_00182304
02 8/02/2010 Zhang Lifeng Document creation AX_00236941

Augest. 2010

Print No.: AXD3-360.820.02.02.02 General notes AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 2 of 12 valid for SW VD onwards
Released
Augest. 2010

General remarks
This document describes proposals for the FLUOROSPOT Compact organ
program
settings based on experiences of the applications specialists. Settings for cassette
technique are described in a separate document
(AXB0-000.820.01.01.02)
Dependent on the options available on the system, the following organ programs
(OGP's) are applicable:
DFR (Digital Fluoro Radiography)
Fluoro - Contin. and SV* (Supervision)

*option

Notes

Not all options are availabe with all systems!

For pediatric work the grid should be removed in general to save dose.
This is required by law in several countries anyway!

For international use the OGP's are labelled with medical names. Tables with
translations are enclosed.

Window values as well as edge enhancement and harmonization* values are


normally necessary to adjust according to the users preferences. Be aware that
window values might need to be adjusted when harmonization values are modified.

All settings are recommendations!

Augest. 2010

Print No.: AXD3-360.820.02.02.02 General notes AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 3 of 12 valid for SW VD onwards
Generator Released
Layout + Translations

Digital Acquisition Programs

Labelling and arrangement at the generator tableau

1 2 3 4 5
Oesophagus Barium Barium Barium Iodine
1 Intestinum
2 /s Gaster Colon Jejunum

Renal
2 Venogram Myelogram Arthrogram
Biliar

Pulmo MCU
3 Trachea HSG 2 /s
spot

4 Abdomen Pelvis Coxa Genu

Paediat Paediat
6
Barium MCU

10

* Remarks:

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DIGITAL AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 4 of 12 valid for SW VD onwards
Generator Released
Layout + Translations

Translations

Organ English German French Italian


1-1 Oesophagus 2/s Esophagus 2 fps Oesophagus 2 B/s Oesophage 2 I/s Esofago 2 I/s
1-2 Barium Gaster Barium Meal MDP Barium Estomac Baryté Stomaco Bario
1-3 Barium Colon Barium Enema Colon KE Barium Colon Baryté Colon Bario
1-4 Barium Jejunum Barium Small Bowel Sellink Grêle Duodeno Bario
1-5 Iodine Intestinum Iodine Intestine Jodkontr. Intestinal Digestif Iodé Iodio Intestinum
2-1 Venogram Venogram Phlebo Phlebo Flebo
2-2 Renal / Biliar Renal / Biliary Nieren / Galle Reins / Vesicule Reni
2-3 Myelogram Myelogram Myelo Myelo Myelo
2-4 Arthrogram Arthrogram Arthro Arthro Artro
3-1 Trachea Trachea Trachea Trachée artère Trachea
3-2 Pulmo spot Lung spotfilm Lunge Zielaufn. Poumon Torace
3-4 HSG HSG HSG HSG HSG
3-5 MCU 2/s MCU 2 fps MCU 2 B/s MCU 2 I/s MCU 2 I/s
4-1 Abdomen Abdomen Abdomen Abdomen Abdomen
4-2 Pelvis Pelvis Becken Bassin Bacino
4-3 Coxa Hip Hüfte Hanche Anca
4-4 Genu Knee Knie Genou Ginocchio
6-1 Paediat. Barium Pediatric Barium Paediatrie Barium Enfant Baryté Bambini Bario
6-2 Paediat. MCU Pediatric MCU Paediatrie MCU Enfant Cysto Bambini Cysto

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DIGITAL AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 5 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 1-1 1-2 1-3 1-4


Program Name Oesophagus Barium Barium Barium
2 f/s Gaster Colon Jejunum
Fluoro Mode Fluoro 1 Fluoro 1 Fluoro 1 Fluoro 1
continous continous continous continous
SDM Dominant m c c c
Dose Level 100 100 100 100
Exposure Data from Fluoro Yes Yes Yes Yes
kV auto auto auto auto
Characteristic Curve (C) C20 C60 C60 C22
kV-Dose Reduction [kV] 125 125 125 125
Focus auto auto auto auto
Regulation stop no no no no
Max. Pulse Width [ms] 100 100 100 100
Blackening Correction 0.0 0.0 0.0 0.0
Grid on on on on
Acquisition filter auto No No No No
Filter type Cu [mm] 0.0 0.0 0.0 0.0

Acquisition Mode Native series Single Single Single


Mark image No Yes Yes Yes
Autoshutter Yes Yes Yes Yes
Bones Black Black Black Black

Scene length [s] 20


Frame Rate 1 [fps] 2
Frame Rate 2 [fps] 1
Frame Rate 3 [fps] 1

Contrast (C) *2 740 760 760 760


Brightness (B) 440 440 440 440
Edge Filter [%] 30 30 30 30
Kernel Size 5 5 5 5
Harmonization [%] *1 30 30 30 30
Harmonization Kernel *1 127 127 127 127

alternative display white white white white


Contrast (C) *2 740 760 760 760
Brightness (B) 316 316 316 316

*1 Only with option Harmonisation

*2 When "auto contrast" is set to "on", use '-60' contrast values

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 6 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 1-5 2-1 2-2 2-3


Iodine Renal
Venogram Myelogram
Program Name Intestinum Biliar
Fluoro Mode Fluoro 1 Fluoro 1 Fluoro 1 Fluoro 1
continous continous continous continous
SDM Dominant lmr m lmr c
Dose Level 100 100 100 100
Exposure Data from Fluoro Yes Yes Yes Yes
kV auto auto auto auto
Characteristic Curve (C) C20 C74 C74 C74
kV-Dose Reduction [kV] 125 96 96 85
Focus auto auto auto auto
Regulation stop no no no no
Max. Pulse Width [ms] 100 100 100 100
Blackening Correction 0.0 0.0 0.0 0.0
Grid on on on on
Acquisition filter auto No No No No
Filter type Cu [mm] 0.0 0.0 0.0 0.0

Acquisition Mode Single Single Single Single


Mark image Yes Yes Yes Yes
Autoshutter Yes Yes Yes Yes
Bones Black Black Black Black

Scene length [s]


Frame Rate 1 [fps]
Frame Rate 2 [fps]
Frame Rate 3 [fps]

Contrast (C) *2 760 740 760 760


Brightness (B) 500 470 450 500
Edge Filter [%] 30 40 30 30
Kernel Size 5 5 5 5
Harmonization [%] *1 30 40 40 30
Harmonization Kernel *1 127 127 127 127

alternative display white white white white


Contrast (C) *2 760 740 760 760
Brightness (B) 256 286 306 256

*1 Only with option Harmonisation


*2 When "auto contrast" is set to "on", use '-60' contrast values
Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 7 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 2-4 2-5 3-1 3-2


Arthrogram Trachea Pulmo
Program Name spot
Fluoro Mode Fluoro 1 Fluoro 1 Fluoro 1
continous continous continous
SDM Dominant c c c
Dose Level 100 100 100
Exposure Data from Fluoro Yes Yes Yes
kV auto auto auto
Characteristic Curve (C) C63 C17 C20
kV-Dose Reduction [kV] 85 125 96
Focus auto auto auto
Regulation stop no no no
Max. Pulse Width [ms] 100 100 100
Blackening Correction 0.0 0.0 0.0
Grid on on on
Acquisition filter auto No No No
Filter type Cu [mm] 0.0 0.0 0.0

Acquisition Mode Single Single Single


Mark image Yes Yes Yes
Autoshutter Yes Yes Yes
Bones Black White White

Scene length [s]


Frame Rate 1 [fps]
Frame Rate 2 [fps]
Frame Rate 3 [fps]

Contrast (C) *2 760 760 760


Brightness (B) 500 316 316
Edge Filter [%] 40 30 30
Kernel Size 5 5 5
Harmonization [%] *1 40 30 40
Harmonization Kernel *1 127 127 127

alternative display white black black


Contrast (C) *2 760 760 760
Brightness (B) 256 440 440

*1 Only with option Harmonisation


*2 When "auto contrast" is set to "on", use '-60' contrast values

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 8 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 3-3 3-4 3-5 4-1


HSG MCU Abdomen
Program Name 2 /s
Fluoro Mode Fluoro 1 Fluoro 1 Fluoro 1
SV SV SV
SDM Dominant lr lr lmr
Dose Level 50 50 100
Exposure Data from Fluoro Yes Yes Yes
kV 81 77 auto
Characteristic Curve (C) C16
kV-Dose Reduction [kV] 125
Focus large large auto
Regulation stop no yes no
Max. Pulse Width [ms] 100 100 100
Blackening Correction 0.0 0.0 0.0
Grid off off on
Acquisition filter auto No No No
Filter type Cu [mm] 0.0 0.0 0.0

Acquisition Mode Single Native series Single


Mark image Yes Yes Yes
Autoshutter Yes Yes Yes
Bones Black Black White

Scene length [s] 10


Frame Rate 1 [fps] 2
Frame Rate 2 [fps] 1
Frame Rate 3 [fps] 1

Contrast (C) *2 760 760 760


Brightness (B) 450 450 306
Edge Filter [%] 30 30 30
Kernel Size 5 5 5
Harmonization [%] *1 30 30 40
Harmonization Kernel *1 127 127 127

alternative display white white black


Contrast (C) *2 760 760 760
Brightness (B) 306 306 450

*1 Only with option Harmonisation


*2 When "auto contrast" is set to "on", use '-60' contrast values

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 9 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 4-2 4-3 4-4 4-5


Pelvis Coxa Genu
Program Name
Fluoro Mode Fluoro 1 Fluoro 1 Fluoro 1
SV SV SV
SDM Dominant lr c c
Dose Level 100 100 100
Exposure Data from Fluoro Yes Yes Yes
kV auto auto auto
Characteristic Curve (C) C16 C15 C12
kV-Dose Reduction [kV] 125 125 125
Focus small small small
Regulation stop no no no
Max. Pulse Width [ms] 100 100 100
Blackening Correction 0.0 0.0 0.0
Grid on on on
Acquisition filter auto No No No
Filter type Cu [mm] 0.0 0.0 0.0

Acquisition Mode Single Single Single


Mark image Yes Yes Yes
Autoshutter Yes Yes Yes
Bones White White White

Scene length [s]


Frame Rate 1 [fps]
Frame Rate 2 [fps]
Frame Rate 3 [fps]

Contrast (C) *2 760 760 760


Brightness (B) 316 316 316
Edge Filter [%] 40 40 40
Kernel Size 5 5 5
Harmonization [%] *1 40 40 40
Harmonization Kernel *1 127 127 127

alternative display black black black


Contrast (C) *2 760 760 760
Brightness (B) 440 440 440

*1 Only with option Harmonisation


*2 When "auto contrast" is set to "on", use '-60' contrast values

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 10 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

DFR 6-1 6-2 6-3 6-4


Paediat. Paediat.
Program Name Barium MCU
Fluoro Mode Fluoro 2 Fluoro 2
SV SV
SDM Dominant c lr
Dose Level 50 50
Exposure Data from Fluoro Yes Yes
kV 81 75
Characteristic Curve (C)
kV-Dose Reduction [kV]
Focus small small
Regulation stop no no
Max. Pulse Width [ms] 100 100
Blackening Correction 0.0 0.0
Grid off off
Acquisition filter auto Yes Yes
Filter type Cu [mm] 0.2 0.2

Acquisition Mode Single Single


Mark image Yes Yes
Autoshutter Yes Yes
Bones Black Black

Scene length [s]


Frame Rate 1 [fps]
Frame Rate 2 [fps]
Frame Rate 3 [fps]

Contrast (C) *2 740 740


Brightness (B) 440 450
Edge Filter [%] 30 30
Kernel Size 5 5
Harmonization [%] *1 30 30
Harmonization Kernel *1 127 127

alternative display white white


Contrast (C) *2 740 740
Brightness (B) 316 306

*1 Only with option Harmonisation


*2 When "auto contrast" is set to "on", use '-60' contrast values

Augest. 2010

Print No.: AXD3-360.820.02.02.02 DFR AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 11 of 12 valid for SW VD onwards
FLUOROSPOT Compact Released
OGP - Proposals

Contin. Fluoro Fluoro 1 Fluoro 2


Program Name Antiisowatt Pead

Durchleuchtungsmodus Continuous Continuous


Mark Image no no
Dose Rate level 31 20

Fluoro Contrast (C) 620 620


Fluoro Brightness (B) 400 400
Edge Filter (F) [%] 6 6
Kernel Size 5 5
Noise Reduction % 30 30
Fluoro Filter Auto No No
Fluoro Filter Type [mm Cu] 0.0 0.0

SUPERVISION Fluoro 1 Fluoro 2


Program Name Antiisowatt SV Paed SV

Fluoro Mode Supervision Supervision


Mark Image no no
Dose Rate level 15 10

Fluoro Contrast (C) 620 620


Fluoro Brightness (B) 400 400
Edge Filter (F) [%] 6 6
Kernel Size 5 5
Noise Reduction % 15 15
Fluoro Filter Auto No No
Fluoro Filter Type [mm Cu] 0.0 0.0

Important Note
Fluoro curves and dose rate values are information only.

Augest. 2010

Print No.: AXD3-360.820.02.02.02 Fluoro Cont + SUPERVISION AXIOM Iconos MD


Replaces: AXD3-360.820.02.01.02 12 of 12 valid for SW VD onwards