You are on page 1of 98

INSTRUCTION MANUAL

AUTO LENSMETER
TL-7000/TL-6000

Read this manual thoroughly before using the instrument


to ensure proper and safe operation.
Contact Tomey Corporation or your local distributor if
you have any questions or you encounter any problems
during operation.

■ Always follow the operation procedures


described in this manual.
■ Keep this manual in a readily accessible
location while operating the instrument.
■ Contact your local distributor if you lose
this manual.

24LA9090-0A
i. Important safety information

■ Do not install this instrument in a location where explosives or


flammable substances are used or stored. Otherwise, fire or explosion
may occur.
■ Do not remove the cover of the instrument. Otherwise, direct
exposure to high voltage sections may occur.
■ Do not disassemble or modify the instrument. Otherwise, direct
exposure to high voltage sections may occur.
■ Disconnect the power cord from the instrument before servicing the
instrument. Otherwise, electric shock may occur.

■ Do not place water or chemicals on the instrument. Any water or


chemicals entering the instrument may cause electric shock or failure.
■ Use only the specified terminal for connection of the instrument.
Using another type of terminal may result in failure of the instrument.
■ This instrument is a measuring device used in ophthalmology and in
spectacle glasses retail stores. Never use this instrument for any
purpose other than measurement of spectacle glasses or contact lenses.
■ A serrated blade is attached to the top of the printer cover. Touching
this blade may result in injuries, so be sure to pay utmost attention
when handling the printer cover.
■ The external output terminal is not isolated from the internal circuit.
Inappropriate wiring may damage the internal circuit. Contact us or
your local distributor before using the instrument connected to
another device.
■ Light emitted by the instrument built-in lamp includes light in the UV
wavelength range. This lamp may affect eyes and/or body. Observe
the following precautions during use.
- Do not look directly at the measurement light beam while
operating the instrument.
- The measurement light beam is irradiated from the projector
(upper section of the instrument) toward the nose piece. Do not
insert a mirror or other reflective surface between the projector
and nose piece. The mirror may reflect the measurement light
beam and the beam may be irradiated directly into the eye.

i-1 ■
This page is intentionally left blank.

■ i-2
ii. How to read this manual

Symbols used in this manual


Sentences accompanied by the symbols below indicate the following:
■ This is a precaution that, if unheeded, will result in a hazardous
situation where there is imminent danger of serious injury or death.

■ This is a precaution that, if unheeded, could result in a hazardous


situation where there is a possibility of serious injury or death.

■ This is a precaution that, if unheeded, may result in a situation where


there is a possibility of minor or moderate injury or damage to
property.
■ This is an additional instruction which may contain special
precautions on company policies related, either directly or indirectly,
to the safety of personnel or to the protection of property.

ii-1 ■
This page is intentionally left blank.

■ ii-2
iii. Contents
i. Important safety information ......................................................................... i-1

ii. How to read this manual ............................................................................... ii-1


Symbols used in this manual ........................................................................................ii-1

iii. Contents ........................................................................................................ iii-1

1. PRIOR TO USE ............................................................................................... 1-1


1.1 Precautions for operation ....................................................................................... 1-1
1.2 Checking package contents ................................................................................... 1-4
1.3 Glossary ................................................................................................................. 1-5
1.4 Intended use........................................................................................................... 1-7

2. NAMES AND FUNCTIONS ............................................................................. 2-1


2.1 Front of the main unit ............................................................................................. 2-1
2.2 Sides of the main unit............................................................................................. 2-2
2.3 Rear and underside of the instrument ................................................................... 2-3
2.4 Monofocal lens measurement screen .................................................................... 2-4
2.5 Progressive lens measurement screen ................................................................. 2-5
2.6 Primary setting screen ........................................................................................... 2-6
2.7 ID Information screen ............................................................................................. 2-7
2.8 System setup screen.............................................................................................. 2-8
2.9 Symbols used for marking ................................................................................... 2-20

3. OPERATION PROCEDURES.......................................................................... 3-1


3.1 Safety precautions ................................................................................................. 3-1
3.1.1 Precautions for installing the instrument ......................................................... 3-1
3.1.2 Precautions for connecting the power cord..................................................... 3-1
3.2 Preparation before use........................................................................................... 3-2
3.2.1 Connection of external device ......................................................................... 3-2
3.2.2 Turning the power on ....................................................................................... 3-3
3.2.3 Operation of the marking unit .......................................................................... 3-3
3.2.4 Operation of the lens table .............................................................................. 3-3
3.2.5 Replacing the nose piece ................................................................................ 3-4
3.3 Measurement procedures ...................................................................................... 3-5
3.3.1 Measuring a monofocal lens ........................................................................... 3-5
3.3.2 Automatic detection of a progressive lens ...................................................... 3-6
3.3.3 Measuring a progressive lens (PROG mode) ................................................. 3-7

iii-1 ■
3.3.4 Measuring bifocal/trifocal lenses ..................................................................... 3-9
3.3.5 Measuring contact lenses (CL mode; CL holder mode)................................ 3-10
3.3.6 Measuring the pupillary distance (PD) (TL-7000 only) ..................................3-11
3.3.7 Measuring prisms .......................................................................................... 3-12
3.3.8 Simple map preparation ................................................................................ 3-15
3.3.9 Visual recognition support for lens marks ..................................................... 3-16
3.4 Calculating refractive index of lenses .................................................................. 3-17
3.5 Printout ................................................................................................................. 3-19
3.5.1 Printing........................................................................................................... 3-19
3.5.2 Contents to be printed ................................................................................... 3-19
3.6 Communication format ......................................................................................... 3-20
3.6.1 RS-232C communication format ................................................................... 3-20
3.6.2 LAN communication format ........................................................................... 3-22
3.7 Export setting (LAN) ............................................................................................. 3-26
3.7.1 Common detailed setting (Windows7/8/10) .................................................. 3-26
3.7.2 Creating a shared folder (Windows7/8/10) ................................................... 3-28
3.7.3 System setup ................................................................................................. 3-29

4. TECHNICAL INFORMATION .......................................................................... 4-1


4.1 Prism ...................................................................................................................... 4-1
4.1.1 Refraction of prism and lens ........................................................................... 4-1

5. INSPECTION AND MAINTENANCE ............................................................... 5-1


5.1 Warranty ................................................................................................................. 5-1
5.2 Operational life ....................................................................................................... 5-2
5.3 Inspection ............................................................................................................... 5-2
5.4 Routine maintenance ............................................................................................. 5-2
5.4.1 Maintenance of main unit ................................................................................ 5-2
5.4.2 Cleaning nose piece ........................................................................................ 5-2
5.4.3 Cleaning protective glass ................................................................................ 5-3
5.4.4 Cleaning the collimeter lens ............................................................................ 5-3
5.5 Replacing consumables ......................................................................................... 5-4
5.5.1 Printer paper .................................................................................................... 5-4
5.5.2 Ink cartridge ..................................................................................................... 5-4
5.5.3 Nose piece....................................................................................................... 5-5
5.5.4 Fuses ............................................................................................................... 5-5
5.6 Storing .................................................................................................................... 5-6
5.7 Disposal.................................................................................................................. 5-6

■ iii-2
6. TROUBLESHOOTING .................................................................................... 6-1
6.1 General operation .................................................................................................. 6-1
6.2 General operation for measuring lenses................................................................ 6-3
6.3 Measuring progressive power lenses .................................................................... 6-4
6.4 Error messages ...................................................................................................... 6-5
6.4.1 Startup screen ................................................................................................. 6-5
6.4.2 Measurement screen ....................................................................................... 6-5

7. CONSUMABLES AND OPTIONAL EQUIPMENT ........................................... 7-1


7.1 Spare parts ............................................................................................................. 7-1

8. SPECIFICATIONS ........................................................................................... 8-1


8.1 Specifications ......................................................................................................... 8-1
8.2 Data control ............................................................................................................ 8-2
8.3 Dimensions and power source............................................................................... 8-2
8.4 Operating environment........................................................................................... 8-2
8.5 Wireless LAN (wireless LAN installed instrument only) ......................................... 8-3
8.6 Classification .......................................................................................................... 8-3
8.7 Declaration of conformity to EMC .......................................................................... 8-4

9. INDEX ............................................................................................................. 9-1

iii-3 ■
This page is intentionally left blank.

■ iii-4
1. PRIOR TO USE
■ Read this manual thoroughly before using this instrument to ensure
proper and safe operation.
■ Always follow the operation procedures as described in this manual.
■ Check that there are no devices that generate a strong magnetic field
near the instrument.
A strong magnetic field may cause noise and affect measurement.

1.1 Precautions for operation


■ Allow only qualified operators to use the instrument.
■ Precautions when installing the instrument
- Install the instrument in a location free of water or chemicals. Any water
or chemicals entering the instrument may cause electric shock or failure.
- Do not install the instrument in a location where chemicals are stored or
gases may occur. Spilt chemicals or vapor may enter the instrument and
result in fire.
- Check that the frequency, voltage, and allowable current (or power
consumption) of the power source are appropriate.
Otherwise, fire or electric shock may occur.
- Connect the power cord to a grounded 3-pin outlet. Otherwise, a short
circuit due to failure of the instrument may result in electric shock.
- Do not place any heavy object on the power cord or compress the power
cord. Fire or electric shock may occur.
- Fully insert the power cord into the outlet. Faulty contact, allowing any
metal to contact the exposed terminal of the cord, or dust accumulated
on the exposed terminal of the cord may result in fire or electric shock.
- Do not connect a device with data transmission specifications that are
not compatible. Fire or electric shock may occur. Be sure to contact your
local distributor before using the instrument if it is connected to another
incompatible device.
- Conduct grounding work correctly to prevent the possibility of electric
shock.
- Do not hold the display when transporting the instrument. The display
may disconnect and the instrument may drop, resulting in injuries.
- Install the instrument in a location not subject to direct sunlight, high
temperature and humidity, or air containing dust, salt, and/or sulfur.
Otherwise, failure or malfunction may occur.

1-1 ■
- Install the instrument in a level, stable location free of vibration or
mechanical impact. Otherwise, measurement cannot be conducted
correctly. Also, the instrument may topple or fall, resulting in fire or
serious accident.
- Check the power supply (discharge condition, polarity, etc.).
- Check that the date set in the instrument conforms to the actual
operation date and time.

■ Precautions before using the instrument

- Check that the instrument works properly by inspecting switch and


button operations, display and printing.
- Check that all cables are connected correctly.
- Check the power supply (discharge condition, polarity, etc.).
- Check that the instrument is correctly grounded.
- Check that the date set in the instrument conforms to the actual
operation date and time.
- Always keep the nose piece tip clean. If the nose piece tip has dust on it,
the lens may be damaged.

■ Precautions during operation

- Do not lean on the instrument or press on the instrument from the top.
The instrument may topple over, resulting in mechanical failure or
injuries.
- If any problem is found with the instrument, take appropriate action such
as halting operation.
- If problems such as smoke, unusual odor or abnormal sound occur, turn
the instrument off immediately, disconnect the power cord from the
outlet, and contact Tomey Corporation or your local distributor.
- When cutting the printer paper, pull the paper downward to press it
against the cutter blade and then draw the paper sideways. If the paper is
not applied to the blade appropriately, the paper may become jammed or
the printer may be damaged.

■ 1-2
■ Precautions after operation

- Do not use organic solvents such as thinner, benzene, or acetone to


clean the instrument. Fire or electric shock may occur. (These solvents
may also corrode the resin or coating of the instrument.)
- Follow the specified procedures to return the instrument to its original
status and turn it off.
- Hold the plug when disconnecting the power cord from the outlet to
avoid placing excessive force on the cord. Pulling the cord may damage
the inner core wires, resulting in electric shock or fire.
- Refer to "5.6 Storing" for instructions on the storage of the instrument.
- Clean and neatly arrange the accessories and cables.
- Clean the instrument appropriately at the end of operation to be ready for
the next use.
Do not cover the instrument with the dust cover while the instrument is
turned on.

■ If any failure occurs in the instrument, immediately halt operation,


identify the failure in the instrument, and contact your local
distributor for repairs.
- Do not disassemble or modify the instrument. Touching the internal
circuit may result in death or serious injuries.
- Disconnect the power cord from the outlet when replacing fuses.
Otherwise, electric shock may occur, resulting in death or serious
injuries.
- Use the power cord and fuses provided with the instrument or specified
by Tomey to ensure safety. Also, do not use the accessories provided
with the instrument for other equipment.
- When any failure occurs in the instrument, identify the failure in the
instrument and contact your local distributor for inspection and repairs.
Do not attempt to repair the instrument yourself.
- Conduct regular inspections of the instrument and components.
- When the instrument is not in use for 1 month or longer, check that the
instrument is operating correctly and safely before starting operation.
Refer to "5.3 Inspection" in this manual for the checking procedures.

1-3 ■
1.2 Checking package contents
Open the package and check that the required quantity of the following items is included,
and check that they are not damaged.
If any item is missing or damaged, contact your local distributor as soon as possible.

■ Keep the box and packing materials for use when moving or
transporting the instrument.
■ Never lift the instrument by the display when taking the instrument
out of the packing box. The instrument may be damaged.
■ Ink may leak from the ink cartridge of the marking unit due to
vibration during transportation or environmental changes during
storage. Check the marker tip when opening the package. If the tip is
smeared with ink or ink has leaked, wipe off the ink with paper or
similar and mark dummy points two or three times before using the
marker for actual operation.

● Auto lensmeter main unit


● Power cord ······················································· 1
● Nose piece ······················································· 1
(Preinstalled on the main unit.)
● CL nose piece ··················································· 1
● Hard contact lens holder ····································· 1
● Fuse ······························································· 4
(Two of these are preinstalled in the main unit.)
● Printer paper ···················································· 3
(1 roll is set in the main unit.)
● Dust cover ······················································· 1
● INSTRUCTION MANUAL ···································· 1

■ 1-4
1.3 Glossary
Main terms used in this manual are described below.

[ADD] : Addition of multifocal lens (including progressive lens)

[AUTO HOLD] : Function to automatically read the following


measurements
- Optical center of a monofocal lens
- Optical center of a contact lens
- Far point and near point of a progressive lens
- Far point and near point of a bifocal lens
- Far point, middle point, and near point of a trifocal
lens

[AUTO TABLE LOCK] : Function to automatically lock the lens table at the
following position
- Optical center in monofocal lens measurement
mode

[AXIS (A)] : Axis that represents the direction of astigmatism (Unit:


deg.)

[BAS] : Direction (angle) of the base of the prism displayed in


polar coordinates (Unit: deg.)

[BLUE] : Blue light. The light containing the wavelength (380 - 500
nm) in the smart phones and PC monitors.

[CL] : Symbol representing contact lenses

[CYL (C)] : Cylindrical refraction (Unit: D)

[D] : Diopter. Unit for expressing the refractive power of a lens.


This is the inverse of the focal length (in meters)

[DCX] : Distance (decentering) in the horizontal direction (DCX)


from the measurement position to the optical center (Unit:
mm)

[DCY] : Distance (decentering) in the vertical direction (DCY) from


the measurement position to the optical center (Unit: mm)

1-5 ■
[D1] : Anterior refractive power of a lens (Unit: D)

[D2] : Posterior refractive power of a lens (Unit: D)

[L VALUE] : Distance from the lower edge of a spectacle frame to the


optical center of the lens.

[LPD] : Distance from the center of a spectacle frame to the


optical center of the left lens

[n] : Refractive index of a measured lens

[ne] : Refractive index of spherometer


(The default setting of the instrument is 1.523.)

[PD] : Pupillary distance (Unit: mm)

[PSM] : Absolute value of prismatic power displayed in polar


coordinates (Unit: ∆)

[PX, PY] : Prismatic power in horizontal direction (PX) and vertical


direction (PY) displayed in Cartesian coordinates (Unit: ∆)

[RPD] : Distance from the center of a spectacle frame to the


optical center of the right lens

[SPH (S)] : Spherical refractive power (Unit: D)

[UV] : Ultraviolet light

[XFN] : Distance between the far point and near point in the
horizontal direction.

[YFN] : Distance between the far point and near point in the
vertical direction.

[ △] : Prism diopter. Unit for expressing prismatic power

■ 1-6
1.4 Intended use
This instrument is designed to measure the refractive and prismatic power of spectacle
lenses and contact lenses.

1-7 ■
This page is intentionally left blank.

■ 1-8
2. NAMES AND FUNCTIONS
2.1 Front of the main unit

(1)

(6)

(8) (7)

(3)
(2)

(5)
(4)

(Fig. 1)

(1) Display/Touch panel


(2) Nose piece (for measuring spectacle glasses or contact lenses)
(3) Nose block
(4) HOLD button
(5) Lens table
(6) Power indicator
(7) Marking unit
(8) Lens holder

2-1 ■
2.2 Sides of the main unit

(2)
(3)

(1)

(Fig. 1) (Fig. 2)

(1) Printer
(2) Table lock button
(3) Power switch

■ 2-2
2.3 Rear and underside of the instrument

(3)

(4)

(6)

(Fig. 2)
(1)
(Fig. 1) (5)
(2)

(1) Power terminal


(2) Fuse holder
(3) External output port (RS-232C): for phoropter connection
(4) External output port (USB)
(5) External output port (LAN): for PC connection
(6) Service switch
* Used when conducting maintenance.

2-3 ■
2.4 Monofocal lens measurement screen
(4) (2) (13)
(3)
(3)

(1) (15)

(11)

(5)

(10) (9) (12) (8) (7) (6) (14) (Fig. 1)

(1) Measured lens Refractive power Prism PD value


Display of S: Spherical refractive power PXY LPD
measurement values C: Cylindrical refractive power rθ RPD
A: Astigmatic axis DXY
(2) UV/BLUE light Purple bar: UV transmissivity
transmissivity Blue bar: Blue light transmissivity
(3) Measured lens Switches the right and left measured lens.
(4) ID information Jumps to the ID information input screen and displays the
input information.
(5) Clear Clears the measurement information.
(6) Print Prints or exports the measurement information.
Export The output mode and icon alter according to the setting.
(7) Visual recognition Jumps to the lens marks visual recognition support screen.
support of lens marks
(8) Refractive power Jumps to the refractive power calculation screen.
calculation
(9) Progressive lens Jumps to the progressive lens measurement screen.
measurement
(10) Setup Jumps to the primary setting screen.
(11) TOTAL PD/L Total PD and L values of the measured lens are shown.
(TL-7000 only)
(12) Total prism Total prism value of the measured lens is shown.
(13) Table LOCK Displayed while the lens table is locked.
(14) Setting contents The contents of the current settings are shown.
(15) Target area Area for centering the lens to be measured using the target.

■ 2-4
2.5 Progressive lens measurement screen

(1)

(3)

(Fig. 1)
(2) (4)

(1) Measured lens Refractive power Prism PD value (TL-7000 only)


Display of S: Spherical refractive PXY LPD
measurement power rθ RPD
values C: Cylindrical refractive DXY
power Display of far point and
A: Astigmatic axis near point
ADD: Addition (TL-7000 only)
XFN: Offset in X
direction
YFN: Offset in Y
direction
(2) Simple map Jumps to the simple map preparation screen.
preparation
(3) S map Displays spherical refractive power map.
(4) Addition Switches the addition (+/-).

2-5 ■
2.6 Primary setting screen
(11)
(1)

(6)
(2)
(7)

(8)
(3)

(4)
(9)

(5)

(10)

(Fig. 1)

(1) Auto Hold Turns the Auto Hold function ON and OFF.
(2) Auto Prog Turns the function of automatically jumping to the progressive
lens measurement screen ON and OFF.
(3) Auto Table Lock Turns the function of automatically locking the lens table
during measurement ON and OFF.
(4) RL/S Switches between spectacle glasses measurement screen
and single lens measurement.
(5) PD (TL-7000 only) Turns PD measurement ON and OFF.
(6) Step Switches the increments of the display of measurement
values.
(7) Prism Switches the prism display mode.
(8) UV/BLUE Measure Switches the display of UV and blue light ON and OFF.
(9) CYL Switches the CYL display mode.
(10) System Setup Jumps to the system setup screen.
(11) Exit Returns to the measurement screen. Applies the primary
settings.

* The settings made on the primary setting screen reset when power is turned OFF.
However, the settings made on the “System Setup” screen do not reset even when power is
turned OFF.

■ 2-6
2.7 ID Information screen

(1) (2)

(Fig. 1)

(1) ID display Displays the ID here.


field
(2) OK Sets the entered ID.

2-7 ■
2.8 System setup screen
Make settings for this instrument. Contents set here will be retained even when the
instrument is turned off.
System Setup consists of 3 major categories.
● General … Language, version information, etc.
● Measurement … Function and operation
● Connection & Print … Connection and printing

(2) (3)

(1)

(Fig. 1)

(1) Menu tab Displays the corresponding setting screens.


(2) Address bar The setting screen name selected in the menu tab is displayed.
(3) Exit Applies the changes and returns to the previous screen.

■ 2-8
a) General

(1)

(2)

(3)

(4)
(5)
(6)

(7)
(Fig. 1)

(1) System Ver. Displays the version of the system.


(2) Language Select the language to be displayed.
(3) Date & Time Set the display format of the date and time.
The key pad appears when the input field is touched.
(4) Sound Set whether screen operations generate a beep.
(5) Power Saving The monitor will automatically go into sleep mode when no
operation is performed for a specified time. Touch the monitor
screen to return to normal status. Set the time until the auto power
saving function is activated.
(6) Brightness Set the brightness of the monitor screen.
(monitor)
(7) Machine No. Set the number to identify each instrument when multiple
Setting instruments of the same model are installed.

2-9 ■
(1)

(2)

(3)

(4)

(Fig. 1)

(1) ID Necessity Set how the ID must be handled when exporting examination data.
Export [Yes]
Select this when the ID numbers are always required.
[No]
Select this when the ID is usually entered, but data can also be
exported without entering the ID. Data is exported with the ID as
“No ID” when the ID is not entered.
[User Discretion]
Select this when the ID does not need to be entered. Data is
exported with the ID as “No ID” when the ID is not entered.
(2) ID INPUT Set whether to display the patient information input screen before
capturing images when starting to capture images for a new
patient.
When (1) ID Necessity Export setting is “User Discretion”, this
setting is automatically set to “User Discretion” and cannot be
selected.
[Pre-Measurement]
Opens the patient information input screen before capturing
images.
[User Discretion]
Opens the capture screen.
(3) Input Mask Sets which part of the character string should be recognized as
the ID when entering the ID number from a barcode.
If it starts from “1”, the characters from the very beginning are
identified as the ID number.
(4) Confirmation Select whether to open the patient information confirmation screen
when exporting the examination data.

■ 2-10
b) Measurement

(1)
(2)

(3)

(4)

(Fig. 1)

(1) Function Make settings related to measurement values.


(2) Auto Make settings related to auto functions.
(3) Lens Mark Make settings related to recognition support screen of progressive
multifocal lens marks.
(4) Key Make settings related to the setting buttons located on the
Arrangement measurement screen.

b-1) Functions
(1) (1)

(2) (Fig. 2) (2) (Fig. 3)

(1) Functions1&2 Switches measurement value setting screens.


(2) Setting buttons Switches settings.

2-11 ■
b-2) Auto

(1)

(Fig. 1)

(1) Auto setting buttons Switches auto settings.

b-3) Lens Mark

(1)

(Fig. 2)

(1) Pattern & Size Switches the patterns and sizes to be displayed.
selection button

■ 2-12
b-4) Key Arrangement

(1) (2)

(4)
(3)

(5)

(6) (7)
(Fig. 1)

(1) Setting button 1 Make settings for the button (6) on the lower side of the capture
screen.
(2) Setting button 2 Make settings for the button (7) on the lower side of the capture
screen.
(3) Setting selection Select the function to make settings for “Setting buttons 1&2”.
Refer to “2.6 Primary setting screen” for detailed functions of
each button.
(4) Default button Press the “Default” button to return the “Setting buttons 1&2” to
the factory default settings.
(5) Reset button Press the “Reset” button with either setting buttons 1 or 2
selected, then return to the factory default settings.

2-13 ■
c) Connection & Print

(1)

(2)

(3) (5)

(4)

(Fig. 1)

(1) Select Function Select “Print,” "Export" or “Print & Export”.


(2) My IP Make settings for your own IP.
(3) PC Select LAN or RS-232C to be set for PC.
(4) Printout Jumps to the print setting screen.
(5) Setting Jumps to the PC setting screen for LAN or RS-232C.

c-1) My IP

(1)

(2)

(Fig. 2)

(1) My IP Address Select either DHCP (dynamic IP) or Manual (static IP) for the IP
Setting setting type. When “Manual” is selected, set My IP Address, Sub
Net Mask, and Default Gateway.
(2) IP Check Performs a duplicate IP address check.

■ 2-14
c-2-1) PC (LAN)

(1)

(2)
(4)
(3)
(5)
(6)

(7)

(8) (9)

(Fig. 1)

(1) Login Set the log-in user ID and password.


(2) Host IP Address Set the IP address of the PC to be connected.
(3) Domain Enter the domain of the PC to be connected.
(4) Folder Enter a folder name to save data in the connected PC.
(5) Connection Test Performs a duplicate IP address check.
(6) File Name Settings Set the file name to export.
(7) Export File Select the communication format.
(8) Wireless LAN (wireless Select connection via wireless LAN.
LAN installed instrument (Wired LAN communication turns OFF.)
only)
(9) Setting Jumps to the wireless LAN setting screen.

2-15 ■
c-2-2) File Name Settings
Unique file names are assured only on a single PC. If there are multiple
PCs, the same file name may exist on other PCs. Be very careful not to use
another patient’s measurement data by mistake.
The file name is preset as follows.
Default setting: Patient ID_Date_Time_Model Name_Eye(L/R)_Extension Number.File
Extension.
Example: 1111_20170421_154803_Model Name_R_001.csv

(1)

(2)

(4)

(3)
(Fig. 1)

(1) Sample display Display a sample of the file name.


(2) Standard/Optional [Standard Settings]
Settings Year-Month-Date_Hour_Minute_Second_1/1000
seconds.Unit name.File Extension
The time indicated in the file name is the time when
the measurement data was received from the
examination unit.

[Optional Settings]
Select items and punctuation to be included in the file
name set in (4).
(3) Extension Number This number is assigned when selecting “Extension
Number” at optional settings. It will be added
automatically after export (001-999).
Pressing the “Reset” button makes “001”.
(4) Select items and When “Optional Settings” is selected, select items and
punctuation
punctuation to be included in the file name from the
pull-down menu, referring to the items below.

■ 2-16
Selection item Details
Patient ID Patient ID (1 - 64 characters)
Date Examination date (yyyymmdd)
Time Examination time (hhmmss)
Eye(L/R) Examined eye (R/L/B/N) *1
Eye(OS/OD) Examined eye (OD/OS/OU/N) *1
Model Name Examination unit
Extension Number Number assigned automatically (001 - 999)
(*1) Assign “B” or “OU” to indicate the information is for both eyes to a file including data
for both eyes. Assign “N” if the eye that the data is for is not known.

2-17 ■
c-2-3) Wireless LAN (wireless LAN installed instrument only)

(1) (2)

(3)
(5)
(4)

(Fig. 1)

(1) SSID Enter an SSID.


(2) Search Searches for the device to be connected.
(3) Password Enter the password.
(4) Encryption setting Select the encryption settings.
(5) Connection Test Check the connection.

c-3) PC (RS-232C)

(1)

(2)

(Fig. 2)

(1) Output setting Set communication conditions.


(2) Format selection Switches the output format.

■ 2-18
c-4) PC (print)

(1)

(2)

(3)

(4)

(5)

(Fig. 1)

(1) Header Select the description to be printed in the header.


(2) Footer Select the description to be printed in the footer.
(3) Text Enter descriptions to be displayed as content.
(4) Line Spacing Select the line space.
(5) Print Layout Switches the print layout.
* Refer to “3.5.2 Contents to be printed” for details of the contents to be printed.

2-19 ■
2.9 Symbols used for marking

Refer to instruction manual.

Grounding (earth)

Caution

Manufacturer and month of manufacture

■ 2-20
3. OPERATION PROCEDURES
3.1 Safety precautions
3.1.1 Precautions for installing the instrument
- Install the instrument in a place where disconnecting the power cord can
be done smoothly, for situations where the power must be cut off due to
a problem with the instrument.
- Install the instrument so that the user can use the system under optimal
conditions. Carefully connect the devices so that the wiring does not
disconnect unintentionally during operation, and does not hinder
operation of the instrument.

3.1.2 Precautions for connecting the power cord


- Check that the frequency, voltage, and allowable current (or power
consumption) of the power source are appropriate. Otherwise, fire or
electric shock may occur.
- Connect the power cord to a grounded 3-pin outlet. Otherwise, a short
circuit due to failure of the instrument may result in electric shock.
- Do not place any heavy object on the power cord or compress the power
cord. Fire or electric shock may occur.
- Fully insert the power cord into the outlet. Faulty contact, allowing any
metal to contact the exposed terminal of the cord, or dust accumulated
on the exposed terminal of the cord may result in fire or electric shock.

3-1 ■
3.2 Preparation before use
3.2.1 Connection of external device
- The external output terminals are not isolated from the internal circuit.
Inappropriate wiring may damage the internal circuit.

- The connector must be inserted in the correct direction. Check the


direction and firmly insert the connector.
- Refer to “3.6 Communication format”.

a) Connecting the RS-232C terminal


An external device can be connected to manage the measurement data.
Connect the male-male adapter to the RS-232C terminal (1). Also connect the
female-female interlink cable and secure the connector with screws.

- The RS-232C terminal on this instrument is a female terminal. When


connecting to a PC, attach a male-male adapter to the female-female
interlink cable.
- Appropriate settings are required to connect the RS-232C terminal on
this instrument.
- The cable is not included with this instrument. Prepare a cable
separately.

b) Connecting the LAN cable


- Appropriate settings are required to connect the LAN terminal on this
instrument.
- Use CAT5e cable for data communications between this instrument and
PC.
- The cable is not included with this instrument. Prepare a cable
separately.

c) Wireless LAN connection (wireless LAN installed instrument only)


- Wireless LAN connection is available only when the instrument is
installed with wireless LAN.
- Appropriate settings are required for connecting this instrument via
wireless LAN.

■ 3-2
3.2.2 Turning the power on
Connect the power cord to the power terminal and insert the power cord into the outlet.
Turn on the power switch. Confirm the following before turning power on.
・ The nose piece is firmly secured.
・ The protective glass under the nose piece is clean.
・ There is no lens set on the nose piece.
・ There is no dust or dirt on the nose piece tip.

- Place the lens on the nose piece after the measurement screen appears.
“INITIAL ERROR” occurs when power is turned on with a lens on the
nose piece.
- When turning power on while the nose piece for measuring contact
lenses is set, a buzzer goes off after the startup screen appears and an
error message appears.
- Enter contact lens measurement mode after turning power on.
(Do not turn power on while the nose piece for measuring contact lenses
is mounted.)

3.2.3 Operation of the marking unit


The marking unit is a device with three ink markers used to mark the optical center of
the lens and the axis direction (in the case of cylindrical lens). Place the marks while the
lens is secured with the lens holder.

- Slowly lower the lens holder. If the lens holder is lowered rapidly, the lens
may be damaged.

- Use the lens holder only when marking points.


- When measuring a coated lens, do not move the lens while secured by
the lens holder; otherwise the coating may be damaged.
- Do not touch the marker tip. Otherwise the marker tip may be deformed
and the marker may become unusable.
- Replace the ink cartridge when dots become too light.

3.2.4 Operation of the lens table


The lens table is used as a reference for the astigmatic axis of glasses.

- The lens table automatically unlocks after 3 minutes.

3-3 ■
3.2.5 Replacing the nose piece
1) Attach the CL nose piece in contact lens measurement mode, and the spectacle nose
piece for measuring spectacle glasses in spectacle glass measurement mode.
2) The measurement mode automatically switches to “CL” mode when the nose piece is
replaced with the CL nose piece in single lens measurement mode.

■ 3-4
3.3 Measurement procedures
- Lenses may be damaged if forced against the nose piece or moved
rapidly. Check that there is no dust on the nose piece tip and that the
nose piece is not worn.
- Because cylindrical lenses and plano lenses do not have an optical
center, the target may not be aligned with the center.
- Touch the “Clear” button to delete all measurement data saved in the
memory.
- When PD is not being measured, place the nose block on the left end.
- The measured value may not be identical to the value indicated on the
lens.

3.3.1 Measuring a monofocal lens

(Fig. 1) (Fig. 2)

1) Place the lens on the nose piece for measuring so that the front faces up, move the lens
table, and measure the lens while the lower edges of both sides of the spectacle frame
are in contact with the lens table.
2) Slowly move the lens so that the target comes to the center of the target area.
3) When the optical center of the lens is aligned with the center of the target area, the
target mark becomes larger. When the “Auto Hold” function is ON, the measured value
is automatically read. If the Auto Hold function is OFF, press the “Hold” button to
manually read the measured data.
4) The target mark disappears when the lens is removed from the nose piece.
When PD measurement is set to ON, moving the nose block to the left and right
switches left and right eyes.
(PD measurement is available only for TL-7000.)
5) When both lenses have been measured, the measurement result appears.
6) Touching the “Clear” button deletes the measurement data and returns to the
measurement screen for the first lens.

3-5 ■
3.3.2 Automatic detection of a progressive lens
When the “Auto Prog” function is ON and a progressive lens is placed on the nose
piece for spectacle glasses while the monolens measurement screen is open, the
instrument automatically detects the lens as a progressive lens and switches the
measurement mode to “PROG”.

(Fig. 1) (Fig. 2)

- Place the center of the lens on the nose piece for spectacle glasses.
The instrument may not be able to correctly detect a progressive lens
from a peripheral section of the lens even when set.
- When the addition (ADD value) is 1D or lower, automatic detection may
not be possible.
- When the “Auto Prog” function is activated, the measurement mode may
switch to PROG mode if the lens is tilted when placed. If the mode is
switched by mistake, select the monofocal lens measurement screen
again to change the mode. When the Auto Prog function is deactivated,
the instrument does not automatically detect the lens type or change the
mode to PROG.

■ 3-6
3.3.3 Measuring a progressive lens (PROG mode)
a) Measurement procedures
(Back) Lens table
Measurement is conducted according to the steps below.
(3) Near point
(1) Detecting the progressive area
(1) Progressive
(2) Measuring the far point
area
(3) Measuring the near point (2) Far point
(Front)
(Fig. 1)
The screen changes according to the measurement points.

(Fig. 2) (Fig. 3)

- When removing the lens while reading the far point or near point, the
screen automatically returns to the far point measurement screen.

3-7 ■
b) Measuring the far point

(2)
(1)

(Fig. 1)

1) The far point is nearest to you. Move the lens forward to bring the target (1) to the
center (2).
2) Align the target with the center. When the “Auto Hold” function is ON, the value of the
far point is automatically read. When the Auto Hold function is OFF, touch the “HOLD”
button to read the value.
3) When the far point value has been read, the screen changes to measure the near point.

c) Measuring the near point

(1)

(2)

(Fig. 2)
1) The near point is on the lower side of the lens. Move the lens to the front to bring the
target (1) to the center (2).
2) Align the target with the center. When the Auto Hold function is ON, the value of the
near point is automatically read. When the Auto Hold function is OFF, touch the “HOLD”
button to read the value.
3) The measurement results are displayed on the screen.

■ 3-8
3.3.4 Measuring bifocal/trifocal lenses
a) Measurement procedures
Take measurements at the far point and then near point for a bifocal lens (far point,
middle point and then near point for a trifocal lens).

(Back) Lens table (Back) Lens table

(3) Near point


(2) Near point (2) Middle point

(1) Far point (1) Far point


(Front) (Front)
(Fig. 1) (Fig. 2)

Take measurements on the monofocal lens measurement screen. The screen changes
according to the measurement points.

(Fig. 3) (Fig. 4)

1) Place the lens on the nose piece for measuring spectacle glasses so that the front faces
up, move the lens table, and measure the far points while the lower edges of both sides
of the spectacle frame are in contact with the lens table.
2) Move the lens to the front to align the near point with the nose piece and measure the
near point. (In the case of a trifocal lens, align the middle point with the nose piece,
measure the middle point, and then measure the near point.)
3) When both lenses have been measured, the measurement result appears. (The
secondary addition is only displayed in the case of a trifocal lens.)

- When removing the lens while reading the far point or near point, the
screen automatically returns to the far point measurement screen.

3-9 ■
3.3.5 Measuring contact lenses (CL mode; CL holder mode)
- Always measure contact lenses in CL mode or CL holder mode with the
nose piece for measuring contact lenses attached. Errors occur in
measured values when measurement is taken in monofocal
measurement mode.
- The cylindrical refraction (“C” value) may appear for a contact lens used
for a long period because the lens has deformed.

a) Measuring a hard contact lens (using the hard CL holder)

(1)

(Fig. 1)
Back

(Fig. 2)
Front

1) Replace the nose piece with the CL nose piece and change the measurement screen.
2) Select the “CL Holder” (1).
3) Place a hard contact lens on the hard CL holder with the outer surface of the lens facing
up.
4) Place the hard CL holder on the CL nose piece so that the “RL” indication shows on the
back side.
5) Move the holder to center the lens.

- The following types of contact lenses can be measured using the hard CL
holder.
When measuring other lenses, place the lens directly on the CL nose
piece and conduct measurement in CL mode; otherwise large
measurement errors may be introduced.
Power : -15D to +15D
Base curve : 7.35mm to 8.25mm
Diameter : 8.0 mm to 9.6 mm

■ 3-10
b) Measuring a hard contact lens (without using the hard CL holder)
1) Replace the nose piece with the CL nose piece and change the measurement screen.
2) Place the lens on the CL nose piece with the outer surface of the lens facing up.
3) Move the lens to center it.

c) Measuring a soft contact lens


1) Replace the nose piece with the CL nose piece and change the measurement screen.
2) Wipe any moisture from the contact lens.
3) Place a soft contact lens on the CL nose piece with the outer surface of the lens facing
up.
4) Move the lens to center it.

- Because it is difficult to maintain the proper shape of the lens due to the
nature of the lens, precise measurement is difficult.

3.3.6 Measuring the pupillary distance (PD) (TL-7000 only)


1) Activate the “PD” button “ON” and move the lens table to the front.
2) The PD value is displayed on the screen when moving the nose block to the right
slightly. If the RPD or LPD value is not displayed while measuring the PD, move the
nose block to the left end and then again to the right.
3) Press the nose pad of the frame to the PD measurement nose block, holding the top of
the spectacle frame toward you.
4) Place the lens to be measured on the nose piece for measuring spectacle glasses, with
the front facing up. The instrument automatically switches right or left (R/L) according to
the position of the nose block.
5) Unlock the lens table and measure the lens while the bottom edges of both sides of the
spectacle frame are in contact with the lens table.

- The PD cannot be measured on the single lens measurement screen.


- Firmly apply the PD measurement nose block to the pad of the spectacle
frame to ensure correct measurement.
- Always take measurements while the bottom edges of both sides of the
spectacle frame are in contact with the lens table to ensure correct
measurement.

3-11 ■
3.3.7 Measuring prisms
1) A prism can be indicated in the following three ways. Select the prism display mode on
the primary setting screen or the “Function” setting screen. Set “Total Prism” display to
show the total prismatic power of right and left lenses.

● Cartesian coordinates: PX, PY (△)


Prism symbols represent the base direction as follows.
O (base out) : Outward
I (base in) : Inward
U (base up) : Upward
D (base down) : Downward

● Polar coordinates : PSM, BAS (△, deg.)


PSM : Absolute prismatic power (△)
BAS : Base direction (angle in degrees)

● Decentering : DCX, DCY (mm)


The distance from the current measurement position of the lens to the optical center is
indicated in the X (horizontal) and Y (vertical) directions.
DCX : Decentering in the X (horizontal) direction (mm)
DCY : Decentering in the Y (vertical) direction (mm)
When “Total Prism” is set to ON, the total prismatic power of the right and left is
displayed.
P-H : Prismatic power in the horizontal direction
P-V : Prismatic power in the vertical direction
* The prismatic power in the vertical direction indicates the value based on the right eye.

- When there is a prism, the target is not aligned with the center of the
target area because of the prismatic power, and the “Auto Hold” function
is not available.

■ 3-12
a) Measuring the prism of a monofocal lens

(Fig. 1)

1) The examiner and the patient face each other, approximately 40 cm apart.
2) The examiner covers his/her right eye and instructs the patient to look at the examiner's
left eye.
3) Mark a small dot (1-mm diameter or smaller) on the right lens to indicate the center of
the pupil.
4) The examiner covers his/her left eye and instructs the patient to look at the examiner's
right eye.
5) Similar to the right lens, mark a small dot (1-mm diameter or smaller) on the left lens to
indicate the center of the pupil.
6) Set the position of the marked dot and press the HOLD button to measure the prismatic
power. The target appears at a point outside the center of the target area according to
the measured prismatic power.
7) Repeat these steps to measure the left lens.
8) When both lenses have been measured, the prism values of the lenses appear. When
“Total Prism” is ON, the total prismatic power of both lenses is also displayed.

b) Measuring the upper/lower prism of a monofocal lens


1) Measure the optical center (position where both the upper and lower prism is 0.0△) of
the right lens. Measure the lens while the bottom edges of both sides of the spectacle
frame are in contact with the lens table.
2) Press the table lock button to lock the lens table at the position where the right lens was
measured.
3) Measure the left lens. While the bottom edges of both sides of the spectacle frame are
in contact with the lens table, place the left lens on the nose piece and align the target in
the horizontal direction only.
4) If the target appears at a point above or below the center of the target area, the lens will
show prisms. When the target is aligned with the center of the target area, there is no
upper or lower prism.
5) Press the HOLD button to measure the prismatic power of the left lens. When the Total
Prism function is ON, the total upper/lower prismatic power of the measured glasses is
shown.

3-13 ■
c) Measuring the prism of a progressive lens
Lens mark

(Fig. 1)

1) Mark a dot at the position for measuring the far point. Find the two lens marks, and
place dots on these marks with a felt pen and the center between them with the marking
unit.
2) Then place a dot at a position approximately 5 - 10 mm above the marked center dot as
the point for measuring the far point.

Far point measurement position

5 – 10 mm

Lens mark
Lens mark (Fig. 2)
Middle point

3) Set the marked position for measuring the far point on the nose piece and touch the
“HOLD” button to take the measurement. The target appears at a point outside the
center of the area according to the measured prismatic power.
* The prismatic degree and power of the far point vary depending on the position of the
measured point. When measuring the prism in this manner, consider possible errors
beforehand.
4) Measure the addition as usual for the near point.

■ 3-14
3.3.8 Simple map preparation
After measuring progressive lenses, the C-value map of the measured lens can be
prepared.

(Fig. 1) (Fig. 2)

1) After measuring the progressive lens, touch the simple map preparation button to open
the simple map preparation screen.
* Measure the PD and “L” value when measuring the progressive lens before preparing
the map. If the PD and “L” value are not used, the position information of the lens
cannot be correctly obtained and the accuracy of the map lowers significantly.
2) Touch the “HOLD” button to start preparation of the map.
3) Take a measurement at an arbitrary point using the nose block and lens table.
Measured sections are colored on the screen.
4) Touch the “HOLD” button again when points to be included in the map have been
measured.
5) The completed C-value map appears.

(Fig. 3) (Fig. 4)

- This function is designed to visualize the clear vision area and area of
astigmatic aberration. However, measurement using the PD and “L”
value needs to be performed to obtain correct positional information.
With certain spectacle glasses, if measurement is performed without using
the PD or “L” value, the simple map preparation may not be available.
(PD and L value measurement functions are available only for TL-7000)

3-15 ■
3.3.9 Visual recognition support for lens marks
The visual recognition support screen for lens marks of the progressive multifocal lens is
opened.

(1) (Fig. 1)

1) Touch the lens mark visual recognition support button to enter the lens mark visual
recognition support screen.
2) Hold the lens on which lens mark needs to be checked toward the screen, and check
the lens mark.
* The pattern and size can be changed by swiping vertically and horizontally on the
screen.
3) Touch the button (1) to return to the progressive multifocal lens measurement screen.

- There are individual differences in visibility of the lens marks.


Change the pattern and its size to find and use the most visible screen.

■ 3-16
3.4 Calculating refractive index of lenses
Use a spherometer to obtain the refractive power of the lens to be measured.

(3)

(1)
(4)
(2)

(Fig. 1)

(Fig. 2)
Measuring the front dioptry (D1)

- The refractive power of the spherometer here (D1 and D2) is the value
obtained when the refractive index of the lens is 1.523. If the refractive
index of the lens measured by your spherometer is not 1.523, set the
appropriate refractive index for your spherometer.
The refractive index of the spherometer is able to be varied within the
range of 1.400 < ne < 1.900.
The new values are saved in the memory and are retained even after
turning power off.
- When measuring an aspherical lens, errors of 0.03 - 0.05 may occur in the
calculated refractive index.
- When measuring a progressive lens or progressive lens exclusively for
reading, enter the value read by the spherometer at the measured far
point or near point.
However, if a progressive area covers the measured far point or near
point, measurement cannot be completed correctly by the spherometer
and errors may occur in the refractive index obtained.

3-17 ■
1) Set “Auto Hold” “OFF”.
2) Secure the lens with the lens holder when the target is aligned with the center.
3) Touch the “HOLD” button to read the measurement data and then mark points on the
lens.
4) Touch the “CAL” button to change the screen to the refractive index measurement
screen.
5) Touch the refractive index field (1) to activate it when you need to change the refractive
index of a spherometer. Touch the “Default” button (2) to reset the value to the default
(1.523).
6) Remove the lens from the nose piece and measure the diopter of the lens using the
spherometer. Place the lens on the level table, apply the spherometer in the direction of
the marked points, and read the value of the spherometer (D1: front dioptry). Repeat to
measure the diopter of the opposite side (reverse side) (D2: back dioptry).
7) Touch the D1 or D2 entry field (3) to select an item and enter a value using the keypad
to display the refractive index of the measured lens (4). Enter values from +25D to -25D
for D1 and D2. Otherwise an error occurs and "ERROR!" appears in the calculation
result. An error also occurs when the sum of D1 and D2 is 0.

■ 3-18
3.5 Printout
3.5.1 Printing
1) After measuring the lens, touch the “Print” button (1) to print out the data.

(1)

(Fig. 1)

- Check that the paper roll is set in the built-in printer.


- No data is printed if “Print” is set to “OFF” on the setup screen. Set
“Print” “ON” to activate the printout function.
- The printer in this instrument is thermosensitive. This means the
measurements printed on the paper will fade away over time. Make a
copy when storing the measurement data for long periods.

3.5.2 Contents to be printed

(1) Measurement date and


(1)
time
(2) ID/Name
(2) (3) Measurement results
(4) Date and time of
printing/Content

(3)

(4)
(Fig. 2)

3-19 ■
3.6 Communication format
3.6.1 RS-232C communication format
a) FORMAT1

(Fig. 1)

(Fig. 1)

■ 3-20
b) FORMAT2

(Fig. 1)

3-21 ■
3.6.2 LAN communication format
a) FORMAT1

Tag Name Explanation Data (Character type: ASCII)


of the tag Name of The maximum Detail Unit
fields number of the
characters
[POWER_R] S, C, A SPH 6 Signed decimal, (-99.99 to +99.99). When D
nd
values of 2 digit is 0, blank is filled up by changing
Right lens field position.
CYL 6 Signed decimal, (-99.99 to +99.99). When D
nd
2 digit is 0, blank is filled up by changing
field position.
nd
AXIS 3 Unsigned decimal, (0 to 999). When 2 deg.
rd
and 3 digit is 0, blank is filled up by
changing field position.
[ADD_R] ADD power ADD1 4 Unsigned decimal, (0.00 to 9.99) D
of Right ADD2 4 Unsigned decimal, (0.00 to 9.99) D
lens
[PRISM_R] Prism PX 6 Signed decimal, “+” is displayed as “O △
value of (out)”. “-“ is displayed as “I (in)”. (-99.99 to
nd
Right lens 99.99). When 2 digit is 0, blank is filled up
by changing field position.
PY 6 Signed decimal, “+” is displayed as “U △
(up)”. “-“ is displayed as “D (down)”. (-99.99
nd
to 99.99). When 2 digit is 0, blank is filled
up by changing field position.
nd
[PD_R] Right PD RPD 4 Unsigned decimal, (0.0 to 99.9). When 2 mm
digit is 0, blank is filled up by changing field
position.
nd
[OC_L_R] L value of L 2 Unsigned decimal, (0 to 99). When 2 digit mm
Right is 0, blank is filled up by changing field
position.
[POWER_L] S, C, A SPH 6 Signed decimal, (-99.99 to +99.99). When D
nd
values of 2 digit is 0, blank is filled up by changing
Left lens field position.
CYL 6 Signed decimal, (-99.99 to +99.99). When D
nd
2 digit is 0, blank is filled up by changing
field position.
nd
AXIS 3 Unsigned decimal, (0 to 999). When 2 deg.
rd
and 3 digit is 0, blank is filled up by
changing field position.
[ADD_L] ADD power ADD1 4 Unsigned decimal, (0.00 to 9.99). D

of Left lens ADD2 4 Unsigned decimal, (0.00 to 9.99). D

■ 3-22
[PRISM_L] Prism PX 6 Signed decimal, “+” is displayed as “O △
value of (out)”. “-“ is displayed as “I (in)”. (-99.99 to
nd
Left lens 99.99). When 2 digit is 0, blank is filled up
by changing field position.
PY 6 Signed decimal, “+” is displayed as “U △
(up)”. “-“ is displayed as “D (down)”. (-99.99
nd
to 99.99). When 2 digit is 0, blank is filled
up by changing field position.
nd
[PD_L] Left PD RPD 4 Unsigned decimal, (0.0 to 99.9). When 2 mm
digit is 0, blank is filled up by changing field
position.
nd
[OC_L_L] L value of L 2 Unsigned decimal, (0 to 99). When 2 digit mm
Left is 0, blank is filled up by changing field
position.

b) FORMAT2

Tag Name Explanatio Data (Character type: ASCII)


n of the tag Name of The maximum Detail Unit
fields number of the
characters
[POWER_R] S, C, A SPH 6 Signed decimal, (-99.99 to +99.99). When D
nd
values of 2 digit is 0, blank is filled up by changing
Right lens field position.
CYL 6 Signed decimal, (-99.99 to +99.99). When D
nd
2 digit is 0, blank is filled up by changing
field position.
nd
AXIS 3 Unsigned decimal, (0 to 999). When 2 and deg.
rd
3 digit is 0, blank is filled up by changing
field position.
[ADD_R] ADD power ADD1 5 Signed decimal, (-0.00 to +9.99). D
of Right ADD2 5 Signed decimal, (-0.00 to +9.99). D
lens
[PRISM_SE Prism Prism 1 0: No PRISM, 1: PX/PY, 2: PSM/BAS, 3:
L_R] display display DCX/DCY
format of format
Right lens
[PRISM_R] Prism PX 6 Signed decimal, “+” is displayed as “O △
value in (out)”. “-“ is displayed as “I (in)”. (-99.99 to
nd
orthogonal 99.99). When 2 digit is 0, blank is filled up
coordinate by changing field position.
s of Right PY 6 Signed decimal, “+” is displayed as “U (up)”. △
lens “-“ is displayed as “D (down)”. (-99.99 to
nd
99.99). When 2 digit is 0, blank is filled up
by changing field position.

3-23 ■
nd
[PRISM Prism PSM 5 Unsigned decimal, (0 to 99.99). When 2 △
_PC_R] value in digit is 0, blank is filled up by changing field
Polar position.
nd
coordinate BAS 3 Unsigned decimal, (0 to 999). When 2 and deg.
rd
s of Right 3 digit is 0, blank is filled up by changing
lens field position.
nd
[PRISM Eccentricit DCX 5 Signed decimal, (-99.9 to +99.9). When 2 mm
_EC_R] y from the digit is 0, blank is filled up by changing field
measurem position.
nd
ent position DCY 5 Signed decimal, (-99.9 to +99.9). When 2 mm
to the digit is 0, blank is filled up by changing field
optical position.
center of
Right lens
nd
[PD_R] Right PD RPD 4 Unsigned decimal, (0.0 to 99.9). When 2 mm
digit is 0, blank is filled up by changing field
position.
nd
[OC_L_R] L value of L 2 Unsigned decimal, (0 to 99). When 2 digit mm
Right is 0, blank is filled up by changing field
position.
[POWER_L] S, C, A SPH 6 Signed decimal, (-99.99 to +99.99). When D
nd
values of 2 digit is 0, blank is filled up by changing
Left lens field position.
CYL 6 Signed decimal, (-99.99 to +99.99). When D
nd
2 digit is 0, blank is filled up by changing
field position.
nd
AXIS 3 Unsigned decimal, (0 to 999). When 2 and deg.
rd
3 digit is 0, blank is filled up by changing
field position.
[ADD_L] ADD power ADD1 5 Signed decimal, (-0.00 to +9.99). D
of Left lens
ADD2 5 Signed decimal, (-0.00 to +9.99). D
[PRISM_SE Prism Prism 1 0: No PRISM, 1: PX/PY, 2: PSM/BAS, 3:
L_L] display display DCX/DCY
format of format
Left lens
[PRISM_L] Prism PX 6 Signed decimal, “+” is displayed as “O △
value in (out)”. “-“ is displayed as “I (in)”. (-99.99 to
nd
orthogonal 99.99). When 2 digit is 0, blank is filled up
coordinate by changing field position.
s of Left PY 6 Signed decimal, “+” is displayed as “U (up)”. △
lens “-“ is displayed as “D (down)”. (-99.99 to
nd
99.99). When 2 digit is 0, blank is filled up
by changing field position.

■ 3-24
nd
[PRISM_PC Prism PSM 5 Unsigned decimal, (0 to 99.99). When 2 △
_L] value in digit is 0, blank is filled up by changing field
polar position.
nd
coordinate BAS 3 Unsigned decimal, (0 to 999). When 2 and deg.
rd
s of Left 3 digit is 0, blank is filled up by changing
lens field position.
nd
[PRISM_EC Eccentricit DCX 5 Signed decimal, (-99.9 to +99.9). When 2 mm
_L] y from the digit is 0, blank is filled up by changing field
measurem position.
nd
ent position DCY 5 Signed decimal, (-99.9 to +99.9). When 2 mm
to the digit is 0, blank is filled up by changing field
optical position.
center of
Left lens
nd
[PD_L] Left PD RPD 4 Unsigned decimal, (0.0 to 99.9). When 2 mm
digit is 0, blank is filled up by changing field
position.
nd
[OC_L_L] L value of L 2 Unsigned decimal, (0 to 99). When 2 digit mm
Left is 0, blank is filled up by changing field
position.
[P_H] Horizontal P-H 6 Signed decimal, “+” is displayed as “O △
prism value (out)”. “-“ is displayed as “I (in)”. (-99.99 to
nd
+99.99). When 2 digit is 0, blank is filled
up by changing field position.
[P_V] Vertical P-V 6 Signed decimal, “+” is displayed as “U (up)”. △
prism value “-“ is displayed as “D (down)”. (-99.99 to
nd
+99.99). When 2 digit is 0, blank is filled
up by changing field position.
[FILES_N] Attached File 3 Unsigned integer (0 - 999)
file number
number Encryptio - “Blank” : encrypted
n “no encryption”: not encrypted
[FILE] File name File 255 Letter(***.jpg)
and name
function File type 4 Letter "COPY":Screen shot
Classifica 2 Letter Refer to *1) About classification code
tion code

*1) About Classification code

Classification code Category A (RL/S) Category B(Type of Measurement)

R Right 0 Normal Lens

L Left 1 Progressive Lens

D Both eyes 2 Contact Lens

S Single X No indication

X No indication

3-25 ■
3.7 Export setting (LAN)
Create a shared folder on the PC, and export the data to the folder.

3.7.1 Common detailed setting (Windows7/8/10)

(1)

1) Select “Network and Internet” from “Control Panel”.

(2)

2) Select “Network and Sharing Center”.

■ 3-26
(3)

3) Select “Change advanced sharing settings”.

(4)

(5)

4) Select “Turn on network discovery”.


5) Select “Save changes”.

3-27 ■
3.7.2 Creating a shared folder (Windows7/8/10)
Create a folder in an arbitrary location.
Right-click on the folder and select “Properties”.

(1)

(2)

1) Select the “Sharing” tab.


2) Select the “Advanced Sharing” button.

(3)

(4)

3) Check “Share this folder”.


4) Select the “Permissions” button.

■ 3-28
(5)

(6)

5) Check the “Allow” box of “Full Control”.


6) Select the “OK” button.

3.7.3 System setup

(1)

(2)

1) Select “Setting” of My IP.


2) Select the “LAN” button after setting (1), and select the “Setting” button.

3-29 ■
(3)

3) Select whether “DHCP” or “Manual”.


When selecting DHCP, IP address will be allotted from DHCP server automatically.
When selecting Manual, set up the IP address arbitrarily. If default gateway is not required,
input “0.0.0.0”.

(4)

4) Make arbitrary settings.


User ID: User ID on the PC side
Password: Password on the PC side
Host IP Address (when the “DNS” button is not pressed): ID address on the PC side
Host computer name (when the “DNS” button is pressed): PC name on the PC side

■ 3-30
Domain: Enter here when exporting to the PC in the domain of the network.
This field does not need to be filled if the PC is not in the domain.
Folder: Enter only the folder name. Path does not need to be entered even if it is in the
hierarchy.

(5)

5) Make arbitrary settings.


SSID: Access point name
Search: Search access point
Password: Password of access point
Encryption Key: NONE, WPA-AES, WPA-TKIP, WPA2-AES, WPA2-TKIP

3-31 ■
This page is intentionally left blank.

■ 3-32
4. TECHNICAL INFORMATION
4.1 Prism
4.1.1 Refraction of prism and lens

1) Two non-parallel refracting surfaces generate


prismatic power. The incident light is refracted toward
the base.
*1 prism diopter (1 ∆): The unit measuring the
deviation of light passing through a prism equal to a
deviation of 1 centimeter at a distance of 1 meter.

Base
2) A lens can be regarded as a group of prisms. A convex
(Fig. 1)
lens is a group of prisms with the base in the center,
so the light is refracted toward the center and
converged.

Light converges Base


toward the center.

(Fig. 2)

3) The point where light is not refracted by the prism is


called the optical center and the prismatic power at
this point is 0∆. The prismatic power becomes larger
at points farther from the optical center.
4) Because a concave lens is a group of prisms with the
base around the outside, the light is refracted
outwards and diverged.
Base

Light is refracted
outwards.
Base

(Fig. 3)

5) Also for a concave lens, the point where light is not


refracted by the prism is called the optical center and
the prismatic power at this point is 0∆. The prismatic
power becomes larger at points farther from the
optical center.

4-1 ■
This page is intentionally left blank.

■ 4-2
5. INSPECTION AND MAINTENANCE
5.1 Warranty
One-Year Limited Warranty
The seller warrants this product to be free from defects in material and
workmanship under the normal use of this product for one (1) year or other
term complying with local regulations from the date of invoice issued by Seller
to the original purchaser.
Lamps, paper and other consumable items shall not be covered by this
warranty.
This warranty also shall NOT apply if the product has not been installed,
operated or maintained in accordance with the INSTRUCTION MANUAL of
Tomey Corporation (here in after called “Tomey”). Neither seller not Tomey
shall be liable for any damages caused by purchaser's failure to follow
instruction for proper installation, use and maintenance of product.
This warranty is only applicable to the new product and DOES NOT cover any
damage resulting from or caused by accident or negligence, abuse, misuse,
mishandling, improper modification of this product, by persons other than
personnel duty authorized by Tomey, not to a product whose serial number or
batch number is removed, altered or effaced.

THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER


WARRANTIES, EXPRESSED OR IMPLIED (INCLUDING SPECIFICALLY,
WITHOUT LIMITING THE GENERALITY OF THE FOREGOING,
ALL WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE), AND ALL OTHER OBLIGATION AND LIABILITY
ON THE PART OF SELLER AND TOMEY. NEITHER SELLER NOR TOMEY
SHALL BE LIABLE FOR INCIDENTAL,
CONSEQUENTIAL OR SPECIAL DAMAGES UNDER ANY
CIRCUMSTANCES OR FOR MORE THAN REPAIR,
REPLACEMENT OR REFUND OF THE PURCHASE PRICE OF DEFECTIVE
GOODS.

5-1 ■
5.2 Operational life
This instrument is designed to have an operational life of 8 years when operated in the
appropriate environment and adequately inspected and serviced.

5.3 Inspection
1) The instrument cannot be started correctly if the protective glass is damaged or
contaminated. Check that the protective glass is free of damage or contamination.
2) Check that the UV bar graph on the measurement screen indicates 100% without a lens
placed on the nose piece after the instrument is started. Restart the instrument if the
screen does not indicate 100%.

5.4 Routine maintenance


- Hold the plug when disconnecting the power cord from the outlet to
avoid applying excessive force on the cord. Pulling the cord may damage
the inner core wires, resulting in electric shock or fire.
- Turn power off and disconnect the power cord from the outlet before
starting maintenance.
- Do not use organic solvents such as thinner, benzene, or acetone to
clean the instrument. Solvents such as these may corrode the surface of
the instrument.
- Disconnect the power cord from the outlet and place the dust cover over
the main unit when not in use for 1 month or longer.
- Do not touch the optical parts of the measurement window, etc. and keep
these parts free from dust. Fingerprints and dust will affect measurement
accuracy.
- Place the dust cover over the main unit when not in use.

5.4.1 Maintenance of main unit


Wipe the main unit's case, LCD, and operation panel with a dry cloth if dirty. When very dirty,
clean the surface using a cloth dampened with diluted neutral detergent.

5.4.2 Cleaning nose piece


Lightly wipe the tip of the nose piece with cloth, etc. If dust or dirt is stuck to it, it may
damage the lens.

■ 5-2
5.4.3 Cleaning protective glass
Remove the nose piece and blow dust from the protective glass off using a commercial
blower or similar. If dust or dirt is stuck to it, lightly wipe the glass with lens cleaning paper,
etc.

- Do not damage the protective glass.


Otherwise, the reliability of the measurement data lowers significantly.
- Periodically clean the protective glass under the nose piece with a soft
cloth or equivalent.

5.4.4 Cleaning the collimeter lens

(1)

(Fig. 1)

1) Lift the lens holder lever slightly to unlock and lower it.
2) Use a commercial blower to blow on the collimeter lens (1) and remove dust from it. If
dust or dirt is stuck to it, lightly wipe the glass with lens cleaning paper, etc.

5-3 ■
5.5 Replacing consumables
5.5.1 Printer paper
Replace the roll of printer paper when red lines appear on the sides of the printout.
1) Press the button to open the printer cover.
2) Take out the old roll of printer paper.
3) Install a new roll of printer paper. If the roll is installed in the wrong direction, nothing will
be printed.
4) Close the printer cover with the end of the paper slightly protruding from the outlet.
Press the cover firmly until you hear a click.
5) Cut off any unnecessary paper.

- Never touch the cutter in the printer. Otherwise, injury may occur.

- Always use the specified paper for the printer.


Using other types of paper may cause printer failure.
- Do not start printing without paper set in the printer.
The printer head may be damaged.
- Do not pull paper forcibly.
Trying to pull the paper out may cause printer failure.

5.5.2 Ink cartridge


Replace the ink cartridge when dots become faded. Replacement ink cartridges are optional
parts. Cartridges are available from your local distributor.
1) Lift the lens holder.
2) Lower the marking unit without turning the marking lever.
3) Hold the ink cartridge with one hand and remove the securing screw using a Phillips
screwdriver.
4) Remove the spring and ink cartridge together from the marking unit.
5) Attach the spring to a new ink cartridge, install it in the marking unit, and secure it with
the screw.

- Always use genuine Tomey ink cartridges.


- Do not touch the marker tip of the ink cartridge. Otherwise the marker tip
may deform and the cartridge may become unusable.
- Do not tighten the securing screw of the ink cartridge too tightly.
The ink cartridge may break.

■ 5-4
- Be very careful not to drop the spring or screw while replacing the ink
cartridge.
It is recommended to take appropriate precautions so that the spring or
screw can be easily found if dropped.
- Only use the provided securing screw and spring.
If the screw and/or spring is lost while replacing the ink cartridge,
contact Tomey Corporation or your local distributor.

5.5.3 Nose piece


The tip of the nose piece may be damaged due to rubbing by lenses, so it is recommended
to keep a spare nose piece for measuring spectacle glasses.

- Do not use a damaged nose piece.


The lens being measured may be damaged.

5.5.4 Fuses
1) Turn power off and disconnect the power cord from the outlet.
2) Disconnect the power cord from the power terminal.
3) Take out the fuse holder case under the power socket.
4) Remove the blown fuse from the fuse holder case and install a new fuse.
5) Return the holder case to the fuse holder in the reverse order of removal.

- Disconnect the power cord from the outlet when replacing fuses.
Otherwise electric shock may occur, resulting in death or serious
injuries.
- Use fuses specifically designed for this instrument.
- When the instrument does not work correctly after fuses have been
replaced, there may be other causes of the problem. Turn off the
instrument immediately and contact your local distributor.

5-5 ■
5.6 Storing
- Install the instrument in a location free of water or chemicals. Any water
or chemicals entering the instrument may cause electric shock or failure.

- Do not store the instrument in a location where chemicals are stored or


gases may occur.
Spilt chemicals or vapor may enter the instrument and result in fire.
- Disconnect the power cord from the outlet to ensure safety when the
instrument is not operated for 1 month or longer.
- Install the instrument in a location not affected by direct sunlight, high
temperature and humidity, and air containing dust, salt, and/or sulfur.
- Store the instrument in a level, stable place free of vibration or
mechanical impact.
Otherwise, measurement cannot be conducted correctly.
The instrument may also topple or fall, resulting in fire or a fatal accident.
- Place the dust cover over the main unit when not in use.
The measurement accuracy will deteriorate if the optical section in the
instrument becomes dirty.

5.7 Disposal
- Keep the box and packing materials for use when moving or transporting
the instrument.
- Keep the packing materials and box together.
- When disposing of the packing materials, sort them by type and dispose
of them as directed by relevant laws and local rules and regulations.
- A lithium battery is used in the instrument. Handling of lithium batteries
varies depending on governing bodies. Follow relevant laws and local
rules and regulations, or contact your local distributor.

■ 5-6
6. TROUBLESHOOTING
Check the following first in the event of any problems.
If the problem is not solved even after checking the applicable item listed below, contact
your local distributor to request inspection and/or repair.

- Do not remove the cover of the instrument.


Exposure to high voltage sections may occur.

- Do not take any actions other than those specified below.


- If the problem is not solved even after checking the applicable item listed
below, contact your local distributor to request inspection and/or repair.

6.1 General operation


● Nothing appears on the display when the power switch is turned on.
Faulty connection of the power cord
Firmly connect the power cord to the power terminal.
Connect the power cord to the power socket on the wall again.

Faulty fuse
Check that fuses are not blown.
If blown, replace the fuse.
If the new fuse blows again, the instrument may be faulty.
Contact your local distributor to request inspection and/or repair.

● The instrument stops with the initial error message displayed on the initial screen.
A lens or other object is on the nose piece when starting the instrument.
Remove the lens or object from the nose piece and restart the instrument.

The nose piece is tilted when set.


Set the nose piece appropriately and restart the instrument.

The protective glass under the nose piece is dirty.


Remove the nose piece, clean the protective glass, and restart the instrument.
(“5.4.3 Cleaning protective glass”)

6-1 ■
● Values of "S," "C," and "A" are not "0" although there is no lens on the nose piece.
Measurement is conducted but the data is faulty.
The nose piece is tilted when set.
Set the nose piece appropriately and restart the instrument.

The protective glass under the nose piece is dirty.


Remove the nose piece, clean the protective glass, and restart the instrument.
(“5.4.3 Cleaning protective glass”)

● Nothing is printed when the "PRINT" button is pressed.


”PRINT” is set to “OFF.”
Set "PRINT" "ON."

The printer paper is not set in the correct direction (reversed).


Set the printer paper correctly.

● The printed data is too light or dark.


The printer paper is old.
Replace the printer paper with a new roll.

The specified printer paper is not used.


Replace the printer paper with the specified type.

● The PD value is not displayed when the PD measurement nose block is moved.
The PD measurement nose block is not reset.
Move the PD measurement nose block to the left end (stopper position).

The PD measurement function is deactivated.


Activate the PD measurement function.

● The “Auto R/L” function does not work when the next lens is set.
The “Auto R/L” function on the “Measurement” tab is set to "OFF."
Set “Auto R/L” “ON”.

The instrument does not detect that the lens has been replaced.
Remove the lens from the nose piece for measuring spectacle glasses when
measurement is complete and set another lens after the target disappears.

■ 6-2
6.2 General operation for measuring lenses
● The target is not aligned with the center of the target area. / The “Auto Hold” function
does not work.
A monofocal lens with a spherical power ("S" value) of 0.0D and which only
has a cylindrical power ("C" value) was measured.
When a monofocal lens only has a cylindrical power ("C" value) and its
spherical power ("S" value) is 0.0D, the optical center is not a single point but
exists as a line. Therefore, the target cannot be aligned with the center of the
target area and the Auto Hold function does not work. In this case, touch the
“HOLD” button to read the value measured in the physical center of the lens.

A progressive power lens with a spherical power ("S" value) of 0.0D was
measured.
No optical center exists on a progressive power lens with a spherical power
("S" value) of 0.0D. Therefore, the target cannot be aligned with the center of
the target area and the Auto Hold function does not work. In this case, move
the lens slightly back, forth, to the right and left at a point approximately 10 -
15 mm above the center of the spectacle frame, and touch the "HOLD" button
at the position where there is least fluctuation of the SPH value to take the
measurement.

● At the time of measurement, the target was aligned with the center when the lens is
significantly offset to the nasal or temporal side within the spectacle frame.
A prism lens was measured.
Refer to “3.3.7 Measuring prisms”.

6-3 ■
6.3 Measuring progressive power lenses
● The screen does not change to the far point measurement screen.
The addition is small for a progressive lens with an addition (ADD) of 1D or
lower, or for progressive lenses for computers or for intermediate and near
points, so the target may not be aligned with the center of the target area when
detecting the progressive area.
There is a progressive zone near the center of the lens.
Set the lens at the approximate area and touch the "HOLD" button to display
the far point measurement screen.

● Measurement values vary.


The lens is not set correctly.
Push the lens firmly against the nose piece for measuring spectacle glasses.
The lens must not be tilted or separated from the nose piece for measuring
spectacle glasses.
Firmly apply the lens to the nose piece and take measurements again.

■ 6-4
6.4 Error messages
6.4.1 Startup screen
● “Internal error. Reboot. If not successful, contact your local distributor.”
Initial error
Refer to “6.1 General operation”.

6.4.2 Measurement screen


● “LT ERROR!”
Insufficient light
Clean the lens or the protective glass under the nose piece.
Check that no part of the spectacle frame is on the nose piece.
Dark-tinted lenses or lenses with excessive scratches cannot be measured.

● “Printer in trouble. Contact your local distributor.”


”PRINT” is set to “OFF”.
Set "PRINT" to "ON".

● “Communication in trouble. Check communication. Retry communication again. If not


successful, contact your local distributor.”
Communication with an external device is not available. (Communication
cable is not connected.)
Connect the communication cable to the external output terminal.

6-5 ■
This page is intentionally left blank.

■ 6-6
7. CONSUMABLES AND OPTIONAL EQUIPMENT
The following spare parts and accessories are available from your local distributor of this
instrument.
Contact your local distributor to order them.

7.1 Spare parts


● Nose piece / CL nose piece
Specify the instrument name and the part type as "nose piece” or “CL nose piece."
It is recommended to keep a spare nose piece.

● Printer paper
Specify the instrument name and "printer paper."

● Ink cartridge for marking unit


Specify the instrument name and "ink cartridge."

● Fuse
Specify the instrument name and "fuse."

7-1 ■
This page is intentionally left blank.

■ 7-2
8. SPECIFICATIONS
8.1 Specifications
● Ranges
Spherical Power (SPH) ±25D
Cylindrical Power (CYL) ±10D
Axial angle (Axis) 0 to 180゜
Additional Power (ADD) -2 to 10D
Prism Power 0 to 15∆

● Increments
Diopter 0.01/0.06/0.12/0.25D
Prism 0.01/0.06/0.12/0.25∆

● Modes
Cylinder +/±/-
Prism Rectangular Coordinates / Polar Coordinates /
Displacement

● Measuring principle Measurement using multiple points captured by


CMOS image sensor
● Wavelength 535 nm
● Diameter of the lens φ20 to 120mm (more than φ5mm for CL)
● Interpupillary Distance 40 to 86mm (step: 0.5mm)
● Measurement Object Spectacle lens / Contact lens
● Transmittance of UV The peak of the wavelength is 375nm.
● Transmittance of BLUE The peak of the wavelength is 465nm.

8-1 ■
8.2 Data control
● Marking unit Ink cartridge
● Display 7inch wide color TFT-LCD
● Printer Thermal Printer
58mm wide x 25M
● External output port RS-232C: 1port
USB3.0: 1port
10/100BASE-T Ethernet: 1port

8.3 Dimensions and power source


● Dimensions 188(W)×240(D)×430(H) mm (when LCD is tilted)
● Weight Approx. 5.5kg
● Voltage AC100V - 240V
● Frequency 50/60Hz
● Power consumption 40 - 50VA

8.4 Operating environment


Operate the instrument under the following environmental conditions.
● Installation Indoors, not in direct sunlight
● Temperature +10℃ to +40℃
● Power fluctuation Less than ±10% of nominal voltage
● Atmospheric pressure 800 - 1060hPa
● Humidity 30 to 75%

Store and/or transport the instrument in the instrument’s box under the following
environmental conditions.
● Temperature -20℃ to +60℃
● Humidity 10 to 95%

■ 8-2
8.5 Wireless LAN (wireless LAN installed instrument only)
● Frequency 2400MHz - 2483.5MHz
● Maximum output 17dBm

8.6 Classification
● Protection against electrical shock : Class I ME equipment
● IP Code : IP10
● Mode of Operation : Continuous operation

8-3 ■
8.7 Declaration of conformity to EMC
- The TL-7000/TL-6000 belongs to Group 1 and is designated as Class
B equipment in accordance with IEC/EN60601-1-2.
- The TL-7000/TL-6000 is a Group 1, Class B product according to
EN55011(CISPR11). This means that the TL-7000/TL-6000 does not
generate and/or intentionally use ratio-frequency energy, in the form
of electromagnetic radiation, inductive and/or capacitive coupling, for
the treatment of material or inspection/analysis purposes. It is
suitable for use in domestic establishments and in establishments
directly connected to a low voltage power supply network which
supplies buildings used for domestic purposes.
- The TL-7000/TL-6000 requires special precautions regarding EMC
and needs to be installed, put into service and used according to the
following information.
- Do not use any cables other than the cables provided or specified by
Tomey.
- The use of accessories, transducers and cables other than those
specified, with the exception of transducers and cables sold by
Tomey as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of the TL-7000/TL-6000.
- Portable and mobile RF communications equipment can affect the
TL-7000/TL-6000.
- The TL-7000/TL-6000 should not be used adjacent to or stacked on
other equipment and if it needs to be adjacent to or placed on other
equipment, the TL-7000/TL-6000 should be observed to verify normal
operation in the configuration in which it will be used.
- Anyone who configures a medical system by connecting additional
equipment to the signal input/output parts is responsible for ensuring
that the system complies with the requirements of IEC/EN60601-1-2.

 Cable List
No. Cable Minimum length [m] Remarks
1 AC code 2.5 Unshielded
 Accessory List
No. Name Model Manufacturer
1 AC code 36625B KP-300C/VCTF KDK ELECTRIC WIRE

■ 8-4
Caution: Medical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical equipment is installed or put
into service. Follow the information in the accompanying documentation when installing and operating the
TL-7000/TL-6000.
Caution: Portable or mobile RF communication equipment can affect Medical equipment.
Guidance and manufacturer's declaration -
electromagnetic emissions Table 201
The TL-7000/TL-6000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TL-7000/TL-6000 should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions Group 1 The TL-7000/TL-6000 uses RF energy only for its
CISPR 11 internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference to
nearby electronic equipment.
RF emission Class B The TL-7000/TL-6000 is suitable for use in all
CISPR 11 establishments, including domestic establishments
Harmonic emissions Class A and those directly connected to the public low
IEC 61000-3-2 voltage power supply network that supplies
Voltage fluctuation/ Complies buildings used for domestic purposes.
flicker emissions
IEC 61000-3-3

8-5 ■
Guidance and manufacturer's declaration -
electromagnetic immunity Table 202
The TL-7000/TL-6000 is intended for use in the electromagnetic environment specified below. The
customer or user of the TL-7000/TL-6000 should ensure that it is used in such an environment.
Immunity test IEC 60601 test Compliance level Electromagnetic environment -
level guidance
Electrostatic ± 6kV contact ± 6kV contact Floors should be wood, concrete
Discharge(ESD) ± 8kV air ± 8kV air or ceramic tile. If floors are
IEC 61000-4-2 covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast ± 2kV for power ± 2kV for power Mains power quality should be
transient/ Burst supply lines supply lines that of a typical commercial or
IEC 61000-4-4 ± 1kV for ± 1kV for hospital environment.
input/output lines input/output lines
Surge ± 1kV differential ± 1kV differential Mains power quality should be
IEC 61000-4-5 mode mode that of a typical commercial or
± 2kV common ± 2kV common hospital environment.
mode mode
Voltage dips , < 5 % UT < 5 % UT Mains power quality should be
short (> 95 % dip in UT ) (> 95 % dip in UT ) that of a typical commercial
interruptions and for 0.5 cycle for 0.5 cycle or hospital environment. If the
voltage variations 40 % UT 40 % UT user of the TL-7000/TL-6000 requires
in power supply (60 % dip in UT ) (60 % dip in UT ) continued operation during power
input lines for 5 cycles for 5 cycles mains interruptions,
IEC 61000-4-11 70 % UT 70 % UT it is recommended that the
(30 % dip in UT ) (30 % dip in UT ) TL-7000/TL-6000 be powered from
for 25 cycles for 25 cycles an uninterruptible power supply or
< 5 % UT < 5 % UT a battery.
(> 95 % dip in UT ) (> 95 % dip in UT )
for 5 sec for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic
(50/60 Hz) fields should be at levels
magnetic field characteristic of a typical
IEC 61000-4-8 location in a typical commercial
or hospital environment.
NOTE : UT is the a.c. mains voltage prior to application of the test level.

■ 8-6
Guidance and manufacturer's declaration electromagnetic
immunity Table 204
The TL-7000/TL-6000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TL-7000/TL-6000 should ensure that it is used in such an environment.
Immunity test IEC 60601 test Compliance Electromagnetic environment - guidance
level level
Portable and mobile RF communication
equipment should not be used closer to any
part of the TL-7000/TL-6000, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance

Conducted RF 3 V rms 3 V rms d =1.2


IEC 61000-4-6 150kHz to 80MHz d =1.2 80 MHz to 800 MHz

d =2.3 800 MHz to 2.5 GHz


3 V/m Where P is the maximum output power rating
Radiated RF 3 V/m
IEC 61000-4-3 80MHz to 2,5GHz of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than the compliance
level in each frequency range. Interference
may occur in the vicinity of equipment marked
with the following symbol:

Note1 : At 80MHz and 800MHz, the higher frequency range applies.


Note2 : These guidelines may not apply in all situations. Electromagnetic distribution is
affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the TL-7000/TL-6000 is used
exceeds the applicable RF compliance level above, the TL-7000/TL-6000 should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the TL-7000/TL-6000.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

8-7 ■
Recommended separation distances between
portable and mobile RF communications equipment and the
TL-7000/TL-6000 Table 206
The TL-7000/TL-6000 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the TL-7000/TL-6000 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the TL-7000/TL-6000 as recommended below,
according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
power of transmitter
W d =1.2 d =1.2 d =2.3

0.01 0.12 0.12 0.23


0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 : At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 : These guidelines may not apply to all situations. Electromagnetic distribution is affected by
absorption and reflection from structures, objects and people.

■ 8-8
9. INDEX

△ ............................................................ 1-6 Fuses ...................................................... 5-5

A H

ADD ........................................................ 1-5 HOLD button ........................................... 2-1


Auto ............................................. 2-11, 2-12

AUTO HOLD ........................................... 1-5
ID ............................................................ 2-4
Auto Prog................................................ 2-6
AUTO R/L ............................................... 6-2 L
AUTO TABLE LOCK ............................... 1-5
L VALUE .................................................. 1-6
AXIS (A) .................................................. 1-5
LAN ....................................................... 2-15
B Lens holder ............................................. 2-1
lens holder lever...................................... 5-3
BAS ........................................................ 1-5
lens mark ....................................... 2-4, 3-16
bifocal lens ....................................... 1-5, 3-9
lens table .......................................... 1-5, 2-1
BLUE ........................................ 1-5, 2-4, 2-6
LPD .................................................. 1-6, 2-4


CAL ....................................................... 3-18
Marking unit ............................................ 2-1
Cartesian coordinates .......................... 3-12
Middle point ...................................... 1-5, 3-9
CL ........................................................... 1-5
monofocal lens ....................... 1-5, 3-5, 3-13
CL Holder.............................................. 3-10
multi-focal lens ............................... 1-5, 3-16
communication format ................. 2-18, 3-20
contact lens ................................... 1-5, 3-10 N
CYL (C) ................................................... 1-5
n ............................................................ 1-6
D ne ............................................................ 1-6
near point ......................................... 1-5, 3-9
D ............................................................ 1-5
Nose block .............................................. 2-1
D1 ........................................................... 1-6
Nose piece .............................................. 2-1
D2 ........................................................... 1-6
DCX ........................................................ 1-5 P
DCY ........................................................ 1-5
PD .................................................... 1-6, 2-4
Decentering .......................................... 3-12
P-H ........................................................ 3-12
F Polar coordinates .................................. 3-12
Power indicator ....................................... 2-1
Far point .......................................... 1-5, 3-9
Power switch ........................................... 2-2
Function ................................................. 2-11
Power terminal ........................................ 2-3
Fuse holder............................................. 2-3

9-1 ■
PRINT ..................................................... 6-2 V
Printer ..................................................... 2-2
Visual recognition support of lens marks ... 2-4
prism .............................................. 3-12, 4-1

PROG .............................................. 3-6, 3-7
progressive area ................................... 3-17 Wi-Fi ...................................................... 2-18
progressive lens ..................................... 1-5 Wireless LAN ............................... 2-15, 2-18
progressive lens exclusively ................. 3-17

PSM ............................................... 1-6, 3-12
XFN ......................................................... 1-6
P-V ........................................................ 3-12
PX, PY .................................................... 1-6 Y

R YFN ......................................................... 1-6

refractive index ..................................... 3-17


RL ......................................................... 3-10
RPD ................................................. 1-6, 2-4
RS-232C ........................................ 2-3, 2-18

S map ..................................................... 2-5


Simple map.................................... 2-5, 3-15
SPH (S) .................................................. 1-6

Table lock button..................................... 2-2


trifocal lens ...................................... 1-5, 3-9

USB ........................................................ 2-3


UV .................................................... 1-6, 2-6

■ 9-2
Manufacturer
Tomey Corporation
2-11-33 Noritakeshinmachi
Nishi-ku, Nagoya 451-0051 JAPAN
Tel: +81 52-581-5327
Fax: +81 52-561-4735

EC-Representative
Tomey GmbH
Wiesbadener Straße 21
90427 Nürnberg, Germany
Tel: +49 911-9385462-0
Fax: +49 911-9385462-20

AUTHORIZED TOMEY SERVICE CENTERS

Headquarters, Pacific Rim

Tomey Corporation
2-11-33 Noritakeshinmachi
Nishi-ku, Nagoya 451-0051 JAPAN
Tel: +81 52-581-5327
Fax: +81 52-561-4735

Europe

Tomey GmbH
Wiesbadener Straße 21
90427 Nürnberg, Germany
Tel: +49 911-9385462-0
Fax: +49 911-9385462-20

Original instructions
20170607

You might also like