pfm Rotary 3005 E
Please read this operating manual
carefully before using the instrument
for the first time and make certain
that you are familiar with the method
of operation as well as the instrument’s
operating elements to ensure that it
is used in the correct way.
www.pfmmedical.com
Note
Note
Dear customer,
Please read this instruction manual carefully before using the instrument for the first time and
make certain that you are familiar with the method of operation as well as the instrument‘s
operating elements to ensure that it is used in the correct way.
Serial number:
Above, please enter the serial number of your instrument, which can be found on the nameplate.
Part 2 – Introduction 7
2.1 Views of the pfm Rotary 3005 E 7
2.2 Details of the pfm Rotary 3005 E 8
2.3 Technical data for the pfm Rotary 3005 E 8
2.4 Device type, nameplate 9
2.5 Scope of delivery 9
Rev. 2016-12 3
Content
Part 7 – Accessories 24
1 Finger guard
Handwheel lock
When object and knife are being
mounted the handwheel lock should
be activated to prevent injury during
the mounting process. The M-STOP
field lights up when the handwheel is
1 locked.
M -ST O P
1 Handwheel lock
Rev. 2016-12 5
Safety instructions
Part 2 – Introduction
9 Pressure plate
10 Knife holder
11 Section waste tray
12 Handwheel
13 Adjustment screws for
object orientation
14 Handwheel lock
15 Clamping lever for
12
pressure plate
13 16 Knife angle scale
10 14
15
11 16
Rev. 2016-12 7
Introduction
1 Nameplate
2 Mains connection/
instrument fuse
Including
1 Disposable blade holder complete, low profile (narrow blade)
1 Section waste tray
1 Protective cover
1 Operating manual IFU2003005
Without object clamp
Rev. 2016-12 9
Operating instructions
The operator is responsible for checking that the instrument is working properly and safely. The operator shall assume
responsibility for any use of special equipment and/or materials for operating the instrument.
1 Connection socket
2 Instrument fuse
3 Power switch
8 9 10
Rev. 2016-12 11
Operating instructions
3.7 Function and adjustment of the Trimming and Coarse Feed modes
By pressing the Trimming/Coarse Feed control wheel you WA R NING
can switch between the Trimming and Coarse Feed modes. C aution! The advance via Trimming AUT is interrupted
Press the T r immI ng button to select the required value by only when the function is deactivated again using the
turning the Settings control wheel. Press the T rimmIn g control wheel (turning it forwards).
button again or press the Settings control wheel to save the
set value. Manual trimming (Trimming MAN)
In manual trimming, infeed occurs only when the Trimming/
Trimming function Coarse Feed control wheel is turned forwards during sectioning.
The Trimming function can be used to select either manual or If the control wheel is released the infeed automatically
automatic sectioning of the specimen. The required function is reverts to the preselected section thickness.
activated by pressing the m a n or A U T button. When
the T r immI n g AUT function has been activated the Coarse feed functions
AUT button blinks. The coarse feed function is intended to quickly change the
distance from the sample to the knife or away from the
Adjusting the trimming thickness during the trimming knife. In the Coarse Feed mode the advance feed only works
It is also possible to change the preset trimming thickness when the Trimming/Coarse Feed control wheel is turned
during the trimming by turning the Settings control wheel. forward.
3.12 Memo
By pressing the M e mo button (acoustic signal) the sample’s
present horizontal position is saved. By pressing the Memo
button again the saved sample position is restored.
The remaining feed amount is indicated acoustically and end position “Front” the value 0 is shown. An acoustic
optically. The potential remaining feed amount in μm is signal sounds between the positions 29000 and 30000 and
shown optically on the display. In the reference position between 1000 and 0 to indicate that only 1000 μm (1 mm)
“Rear” the value 30000 is shown on the display. In the of feed remains.
1 WARNNING
Caution! A red marker on the
2 object support also indicates
that the front or rear end
position of the horizontal
stroke has been reached.
Rev. 2016-12 13
Operating instructions
1 Adapter orientable
1 Clamping lever
2 Adjusting screws for object
orientation
1 Adapter hole
2 Nut
3 Threaded brass bolt
Rev. 2016-12 15
Operating instructions
1 Handwheel
2 Handwheel lock
3.26 Disposable blade holder upper jaw, low profile (narrow blade)
In order to fix the microtome blade in via the scale on the knife holder base
the upper jaw of the disposable blade and the index on the disposable blade
holder with the pressure plate, the gap holder upper jaw. The clearance angle
1
between the pressure plate and the adjustment is released or clamped
mounting rail must first be opened. To using the clamping lever on the left
2 3
do this, move the clamping lever for hand side. Under usual circumstances
4 the pressure plate (right hand side) all a clearance angle setting of 10° is
the way to its stop. Insert the blade. recommended for the pfm Rotary 3006
During insertion of the blade, the EM. The upper jaw of the disposable
pressure plate should be pushed down blade holder is available either as a low
1 Pressure plate at the lower end to make best possible profile blade holder for narrow blades
2 Clamping lever for the pressure plate use of the gap. The clamping of the or as a high profile blade holder for
3 Clamping lever for clearance angle blade is also made using this clamping wide blades.
adjustment lever. The clearance angle is adjusted
4 Knife angle scale
1 1
1 1
Disposable blade holder upper jaw Disposable blade holder upper jaw
Low profile (narrow blade) High profile (wide blade)
Rev. 2016-12 17
Operating instructions
Rev. 2016-12 19
Working with the microtome
4.3 Troubleshooting
Incorrect knife holder clearance angle Check, adjust the clearance angle (10°)
Hand wheel does not run
Handwheel lock Check operation/latching and clean
smoothly
Vibration while sectioning Object holder is insufficiently clamped Check the clamping and clean
Knife holder/pressure plate is insufficiently
Check the clamping and clean
clamped
Sections compressed/wrinkled Incorrect knife holder clearance angle Check, adjust the clearance angle (10°)
Crooked sections Lateral edges of the block are not parallel Move them into a parallel position
Blade edge and block are not parallel Align the block
Sections rolls up Incorrect knife holder clearance angle Check, adjust the clearance angle (10°)
Sections too thick Object holder is insufficiently clamped Check the clamping and clean
Knife holder/pressure plate is insufficiently
Check the clamping and clean
clamped
5.1 Maintenance
In order to guarantee the sectioning quality and operation of remains in faultless condition. Details can be obtained from
the rotary microtome, annual routine maintenance should your local sales representative or directly from pfm medical
be carried out by a qualified service technician. pfm medical ag.
ag can service your instrument to ensure that it always
1 2 3
1 Connection socket
2 Instrument fuse
3 Power switch
Rev. 2016-12 21
Maintenance, cleaning and care
5.5 Warranty
We have checked and tested this device thoroughly. To assert a claim under this undertaking, the device invol-
However, in the event of any defect, apart from minor defects, ved in the complaint must be sent to us, at the customer‘s
occurring in a device or spare part distributed by us up to expense and risk, with a detailed description of the defect
one year from delivery, we undertake to remedy the defect complained of, specifying the purchase order number, delivery
to the extent specified below, without prejudice to any note number and invoice number. Our address is as follows:
statutory or other contractual claims, even where the defect pfm medical ag, Wankelstrase 60, 50996 Köln, Germany.
was not present on delivery to the customer, provided that
there has been no unauthorized interference with the Claims under this undertaking expire 6 months from assertion
device by third parties. This does not include any assurance of the claim, but not before the end of the claim period. This
of the equipment being free of wear and tear, or any guarantee undertaking and any claims, rights and obligations arising
of the constitution or durability of the device. therefrom are s ubject solely to the material law of the Fe-
deral Republic of Germany, to the exclusion of standards of
The remedying of defects under this undertaking shall be International Private Law and UN Law on the Sale of Goods.
carried out, at our discretion, by the replacement or repair Our general s upply and payment terms apply as
of defective parts. Spare parts may be supplied for up to 5 supplementary conditions.
years after the delivery of the device. Any warranty claims
on the part of the customer beyond the preceding are hereby
excluded, without prejudice to any statutory or other
contractual claims. This applies in particular to any claims
for compensation for damage, consequential damage or
costs relating to the defect. The transport of the device to
us and return transport during this period shall be at the
customer‘s risk.
The WEEE directive 2002/96/EC (WEEE = Waste Electrical and Electronic Equipment; German: Elektro- und Elektronikge-
räte-Abfall) has been issued in Europe with the aim of reducing the increasing quantity of electronic scrap and electrical
and electronic devices that are no longer required in addition to disposing of them in an environmentally friendly manner.
In Germany, the directive was anchored in the 2005 Electrical and Electronic Equipment Act (ElektroG), that not only for-
med the basis for the implementation of the WEEE directive, but also for the RoHS (Limitation of Certain Dangerous Subs-
tances in Electrical and Electronic Equipment).
We, pfm medical ag, feel responsible for our products, from the initial idea to their disposal and see the situations as being
an opportunity for us to make our contribution to the conserving of resources. As the manufacturer, we assume responsi-
bility for the disposal of the decommissioned equipment on your behalf.
Please take the following action in order to return the electrical device:
2. Prepare the device for transportation (please only use the original packaging) and commission a freight forwarder of
your choice with the return transport of the equipment to the following address:
pfm medical ag
Abteilung Histotechnology
Wankelstr. 60
50996 Köln, Germany
3. We bear the costs of the disposal as the manufacturer. Should nothing to the contrary be agreed to, the owner/user of
the device bears the transport and packaging costs.
Rev. 2016-12 23
Accessories
Part 7 – Accessories
Universal cassette clamp Universal cassette clamp,
REF 010003 orientable
▸ Object clamp to be used with REF 013019
orientable adapter or fixed ▸ Universal cassette clamp,
adapter orientable
▸ Quick-release clamp system- ▸ Quick-release clamp system
for all available cassettes for all available cassettes
▸ Crossways and lengthways- ▸ Crossways and lengthways
clamping possible clamping possible
Rev. 2016-12 25
Notes
Rev. 2016-12 27
Contact
Should you have any questions our Customer
Solutions Team is happy to advise you.
service@pfmmedical.com
+49 (0)2236 9641-220
+49 (0)2236 9641-51
www.pfmmedical.com – Your information source for the products of pfm medical ag IFU2003005/2016-12