Professional Documents
Culture Documents
List of tasks:
1. Measure the resistance of the probe and probe cable
2. Acquire data, create a probe curve with a matched transducer system, and find the
transducer system average scale factor
4. Acquire data, create a probe curve with an unmatched transducer system, and observe
the results
2. Measure the resistance of the extension cable and probe between the center conductor
and the connector casing as shown below.
Probe Length Resistance from the Center Conductor to the Outer Conductor (RPROBE) (ohms)
0.5 7.45 ± 0.50
1.0 7.59 ± 0.50
1.5 7.73 ± 0.50
2.0 7.88 ± 0.50
5.0 8.73 ± 0.70
9.0 9.87 ± 0.90
A. You will be using the spindle micrometer to check the voltage versus distance
characteristics of the probe and Proximitors.
1. Secure the probe in position using the correct probe sleeve. With the spindle
micrometer set to zero, position the probe into the spindle micrometer so that the probe
tip lightly contacts the observed sample.
Adjust the spindle micrometer so that the target moves away from the probe tip. Then
return it to zero to ensure the probe is positioned properly.
2. Connect the probe to the Proximitor using the appropriate extension cable.
3. Connect the DMM between the Proximitor OUTPUT and COM connectors. For this
workshop you are using the monitor as a load. If a monitor were not attached, you would
attach a 10K Ohm resistor across the DMM input terminal to simulate monitor input
impedance.
4. Starting at zero, adjust the spindle micrometer out in 5 mil increments. Log the
voltage values at each increment in the data sheet. Be sure to also document transducer
part and serial numbers. Using a transducer system that generates 200mV/mil will give us
approximately one-volt change for each five mils. Calculate the average scale factor
(ASF).
Notice that you can either log the data in the table on page 5 of this workshop or you can
log the data into the spreadsheet called “proximitor curve”.
5. Using the data collected and logged, go to the page following the data log sheet, and
plot the voltage vs gap graph until the response starts to become non-linear. Notice that
if you use the spreadsheet, the data will be plotted for you.
NOTE
The first and last 10 mils of the curve cannot be used due to non-linearity. A sharp
change in the slope of the curve will occur at a probe to target gap of about 10 to 20 mils.
This change will depend upon the type of probe and Proximitor used and the type of
material being observed. The response curve should be linear starting at approximately
10 mils and remain so through 90 mils or more.
2. Enter the values from the 10 mil level to the 90 mil level in your data log to the
appropriate locations in your spreadsheet. Confirm this matches your hand calculated
value.
2. Use the data log sheet on the next page to collect your measured values.
3. Use the graph page following the data log sheet to plot the response curve of your
transducer system.
Data Table
160
90 100 110 120 130 140 150
80
70
60
50
40
30
20
10
24
22
20
18
16
14
12
10
Student:
Date: