You are on page 1of 4

ORIGINAL ARTICLE

Biomechanical Assessment of Locked Plating for the


Fixation of Patella Fractures
Simone Wurm, MD,* Peter Augat, PhD,†‡ and Volker Bühren, MD*

the transformation of tension into compressive forces; however,


Objective: To analyze the mechanical stability of locked plating in under cyclic loading particularly in knee extension, fragmentary
comparison with tension-band wiring for the fixation of fractures of compression is neutralized and the fracture might gap.6 In addi-
the patella. tion, tension-band wiring often leads to loosening of the wires,
dislocation of the fracture, and failure of the osteosynthesis re-
Methods: Biomechanical tests were performed on artificial foam
sulting in knee stiffness and posttraumatic osteoarthritis.4,7,8
patella specimens comparing an angular stable plate and mono-
Transverse patella fractures have recently been treated with in-
cortical screws with tension-band wiring. Tests were performed
terfragmentary screw fixation alone or in combination with a wire
under combined tension and bending until failure simulating
looped over the screws. Although these techniques create a stable
physiological loading of the tibia during walking.
osteosynthesis, the anatomic reduction of the fracture is often
Results: Tension-band wiring failed at 66% of the failure load of problematic, especially with comminuted fractures. Therefore,
plating (1052 N, P = 0.002) and had 5 times larger fracture gap complications like knee stiffness and posttraumatic osteoarthritis
displacements (P = 0.002). remain still a problem after patella fractures.3,7
An alternative approach is the application of bone
Conclusions: Based on the biomechanical advantages, locked plating for patella fractures. Plating of the patella became
plating of the patella may constitute a reasonable alternative in the feasible with the development of the locked plating technique.
treatment of patella fractures. In 2006, Matejcic et al9 evaluated a basket plate in the treat-
Key Words: patella plate, locking, patella fracture, osteosynthesis, ment of comminuted fractures of the distal pole of the patella
comminuted fracture and showed good results. Recently, biomechanical studies
have demonstrated the feasibility of using locked plating for
(J Orthop Trauma 2015;29:e305–e308) the fixation of certain types of patella fractures with or with-
out the additional use of cable wires.10,11 We hypothesized
that the biomechanical stability of patella fracture fixation will
INTRODUCTION be improved by using an anatomic locking plate as compared
The patella is the largest sesamoid bone in the human with tension-band wiring. We specifically addressed the load
body. It increases the effectiveness of the quadriceps muscle to failure and the dislocation after loading.
and enables flexion of the knee, which requires substantial
forces in load-bearing situations such as squatting or
kneeling.1–3 Patella fractures are not unusual, accounting for MATERIALS AND METHODS
0.5%–1.5% of all skeletal injuries.4 Because of the functional The study was conducted with the arrow plate design of
importance of the patella, a patella fracture requires an accu- the patella plating system by Arthrex (Arthrex, Karlsfeld,
rate reduction and rigid fixation. Only nondisplaced fractures Germany, Fig. 1). This device has recently been approved for
without retinacular disruption should be treated nonopera- the treatment of patella fracture in Europe and is under con-
tively. Fracture gaps of more than 2 mm and/or comminuted sideration by the FDA. The plate is anatomically shaped and
fractures are considered indications for operative treatment.3,5 provides unidirectional angular stability and suture holes for
Tension-band wiring has long been the standard treatment the fixation of soft tissue to the plate. Twelve dome-shaped
for displaced patella fractures. This technique includes the use of test specimens were custom-made from Sawbone foam blocks
2 parallel Kirschner wires over the anterior aspect of the patella with a density of 30 pound per cubic foot (pcf) (Sawbones,
acting as a tension band. The principle of tension-band wiring is Malmoe, Sweden). The test specimens were cylindrically
shaped with a diameter of 76 mm and a height of 38 mm.
Accepted for publication February 2, 2015. The top of the cylinders had a curvature according to that of
From the *Department of Traumatology, BG Trauma Center Murnau, Murnau, a human patella. To fix the simulated patella, the cylinders
Germany; †Institute of Biomechanics, BG Trauma Center Murnau, Murnau,
Germany; and ‡Institute of Biomechanics, Paracelsus Medical University were trimmed off with a turning lathe. The upper area imi-
Salzburg, Austria. tated the human patella. With a diameter of 50 mm, it had the
Presented in part at the ESTES Annual Meeting, April 29, 2011, Milan, Italy. dimensions of an average human patella.12 The lower area
The patella plates used in the biomechanical test setup were provided by served as fastening for mechanical fixture of the testing setup.
Arthrex. The authors report no conflict of interest.
Reprints: Simone Wurm, MD, BG-Unfallklinik Murnau, Prof. Küntscher-Str
After preparing the Sawbone dome, a transverse
8, 82418 Murnau, Germany (e-mail: Simone.Wurm@bgu-murnau.de). osteotomy was performed in the middle of the cylinders using
Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. a bandsaw with a width of 1 mm. Then, the 12 cylinders were

J Orthop Trauma  Volume 29, Number 9, September 2015 www.jorthotrauma.com | e305

Copyright © 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Wurm et al J Orthop Trauma  Volume 29, Number 9, September 2015

maximum fracture separation after patella osteosynthesis


was observed from 60 degrees to 30 degrees of flexion.
The lower area of the foam block was inserted in the
fixture with steel pins so that the upper area representing the
patella did not have to be clamped. The patella fixture was
adjusted to the load cell and the movable adapter to the base
plate. The adapters and the patella attachment were connected
with flat rods. A rounded plastic pulley was used to imitate the
femoral condyle pushing on the back of the patella at the level
of the osteotomy. This allowed the fracture gap to open without
interference from the fixture. The tests were performed on an
electromechanical testing machine (Z010; Zwick, Einsingen,
Germany). Movements between the fracture fragments were
recorded with an optoelectronic movement tracking system
(Pontos, Braunschweig, Germany), providing an accuracy of
0.01 mm. Opening of the fracture gap was measured at the
center of the patella at the anterior aspect.
FIGURE 1. Anatomically shaped locking plate with mono- Tension was applied until failure of the osteosynthesis.
cortical screws for the fixation of patella fractures. Load was applied in displacement-control at a velocity of 1 mm/
min. Failure of the osteosynthesis was defined as a sudden
decrease in load. The biomechanical performance was assessed
divided into 2 groups. One group was equipped with the patella by the relative fracture gap movement and the force at failure.
plate. The plate was fixed with 3 monocortical locking screws
Statistical comparisons between both techniques were performed
(3.5 mm diameter, 16 mm length) per bone fragment. The other
with SPSS Statistics 18.0 using a Wilcoxon signed-rank test.
group was equipped with a tension-band wiring using K-wires of
2 mm diameter (Aesculap, Germany). Two Kirschner wires were
inserted just under the cupola of the upper area of the Sawbone
dome with a space of 2.8 mm. A steel cable with a diameter of 2 RESULTS
mm was tightened around the K-wires with 222 N (50 lbs). The failure mechanism in the biomechanical test was
The biomechanical test setup combined tension and always failure of the bone substitute, which occurred on
bending to simulate the physiological load of the patella average at 627 N (SD: 231 N) using the tension-band wiring
(Fig. 2). Stretching the leg, the patella is exposed to tension and at 1052 N (SD: 32 N) using the patella plate (P = 0.002)
caused by the quadriceps and patellar tendon. During flexion, (Figs. 3, 4).
an additional bending moment is generated by the fulcrum Because the patella Sawbones with tension-band wiring
caused by the femoral condyles. This bending moment was started to show failure at 370 N, we compared fracture gap
simulated by a plastic pulley moving on the foam cylinder displacements between the 2 groups at a load of 350 N at
with mounting. Tension was simulated by the use of 2 rods which all specimens were still intact. With tension-band
adjusted to an angle of 45 degrees. The angle of 45 degrees wiring, the fracture gap at 350 N load level was more than
was chosen because Burvant et al13 demonstrated the 5 times larger than with plate fixation (Table 1, P = 0.002).

FIGURE 2. Biomechanical testing


device: a view from anterior with
patella plate (A) and from lateral (B).

e306 | www.jorthotrauma.com Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved.

Copyright © 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
J Orthop Trauma  Volume 29, Number 9, September 2015 Plate Fixation of Patella Fractures

FIGURE 3. Loads to failure (median,


interquartile range, minimum, and
maximum) of patella plating com-
pared with the tension-band wiring.
The tension-band wiring failed at
significantly lower loads (P = 0.001).

The measured displacement at the load level before failure biomechanical assessment of locked plates under challenging
of the fracture fixation construct was almost 3 times larger conditions of knee flexion demonstrated less interfragmentary
for tension-band wiring compared with plate fixation gap movement and increased failure strength compared with
(Table 1, P = 0.002). tension-band wiring, which is the current gold standard for
the fixation of patella fractures. Compared with tension-band
wiring, locked plating resulted in increased fixation strength
DISCUSSION and reduced distraction of the fracture gap. The tension-band
The findings of our study suggest that osteosynthesis of wiring failed at significantly lower loads than the patella plate
patella fractures with this particular locked plate design with some failures occurring at load levels below 400 N. These
provides sufficient mechanical stability to prevent fracture values are in accordance with the results by Carpenter et al,7
separation with postoperative functional loading. The who biomechanically tested tension-band wiring with screw

FIGURE 4. Displacement of the fracture


gap at the maximal load after plate osteo-
synthesis (A) and tension-band wiring (B).

Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved. www.jorthotrauma.com | e307

Copyright © 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Wurm et al J Orthop Trauma  Volume 29, Number 9, September 2015

this was the first study that assessed the biomechanical prop-
TABLE 1. Displacement of Fracture Gap at 350 N Load Level
erties of a locking patella plate in comparison with tension-
and Immediately Before Failure
band wiring for the fixation of patella fractures.
Displacement, Arrow Plate Tension-Band Wiring One of our limitations was the use of Sawbones, which
mm (Mean 6 SD) (Mean 6 SD) P
were too brittle. If we had used less brittle bone substitutes, we
Displacement at 0.5 6 0.1 2.4 6 1.1 0.01 could probably have observed the cutting through of the screws
350 N
or steel cables. That would have been a more physiologically
Displacement 1.8 6 0.3 5.1 6 1.8 0.01
before failure failure mechanism. In addition, using a bone substitute with
a dense cortical layer might have been improved the results of
tension-band wiring. However, the observed failure forces in our
fixation in human patellae and found failure at a load level of foam model were in agreement with previous biomechanical
395 N, at a knee flexion of 45 degrees. Although screw fixation studies in human cadaveric patellae.7 Another limitation was
alone failed at 554 N, screw fixation in combination with a wire that our tests were performed at 1 fixed angle of knee flexion.
placed through the cannulated screws failed at significantly We chose the flexion angle at which the maximum fracture
higher loads of 732 N. Together with our findings, this would separation was observed in previous studies.13 Furthermore,
suggest that failure loads with plate fixation are nearly twice we inserted the K-wires just under the cupola of the upper area
those observed with tension-band wiring. of the Sawbone dome. We chose this position because the inser-
In our study, the failure mechanism in both groups was tion depth corresponded best to an in vivo situation. Because of
failure of the bone substitute. The patella Sawbones fractured at the fact that we tested only 1-layered bone substitutes without
the screw–sawbone interface, or at the K-wire–sawbone inter- a cortical layer, the K-wires presumably broke out earlier.
face. That failure mechanism could be explained by the mate- In conclusion, locked plating of patella fractures with an
rial properties of the bone substitute. We used a Sawbones anatomically contoured plate provides a biomechanical stable
model with a density of 30 pcf because this most closely fixation. Compared with tension-band wiring, locked plating
represents the characteristics of the cancellous nature of the resulted in higher mechanical strength and reduced fragment
human patella. Cancellous bone has a compressive strength dislocation under loading. Based on our biomechanical findings,
between 3 and 30 MPa and a tensile strength of 2–20 MPa. locked plating may constitute a reasonable alternative to tension-
The compressive strength of a 30 pcf Sawbone is 18 MPa, and band wiring in the treatment of patella fractures in the future.
its tensile strength is 12 MPa as defined by the manufacturer.
Contrary to our expectations, the foam was more brittle than
the human bone, which led to specimen fracture instead of REFERENCES
cutting through the specimen with cables or screws. 1. Benjamin J, Bried J, Dohm M, et al. Biomechanical evaluation of various
Probably, more important than the increase in fixation forms of fixation of transverse patellar fractures. J Orthop Trauma. 1987;
strength was the reduction of fracture gapping by the locked 1:219–222.
plate. Retention of fracture reduction under load is an important 2. Koval KJ, Kim YH. Patella fractures. Evaluation and treatment. Am J
Knee Surg. 1997;10:101–108.
requirement for successful fracture repair. With the use of 3. Wild M, Windolf J, Flohé S. Patellafrakturen. Unfallchirurg. 2010;113:
patella plating, fracture gap widening under loading was always 401–412.
,2 mm, even at maximum load levels before construct failure. 4. Galla M, Lobenhoffer P. Frakturen der Patella [in German]. Chirurg.
In contrast, using tension-band wiring, fracture gapping ex- 2005;76:987–999.
5. Carpenter JE, Kasman RA, Patel N, et al. Biomechanical evaluation of current
ceeded 2 mm at load levels of 350 N and was almost 5 times patella fracture fixation techniques. J Orthop Trauma. 1997;11:351–356.
larger compared with the locked plate construct. Load levels of 6. Labitzke R. Zuggurtungen—Richtiges und Falsches am Beispiel der Pa-
350 N are likely to occur with partial weight-bearing, whereas tellafraktur [in German]. Chirurg. 1997;68:638–642.
full load-bearing during level walking generates loads in excess 7. Carpenter JE, Kasman RA, Matthews LS. Fractures of the patella. J Bone
of 850 N.14 Fracture gapping .2 mm during loading is likely to Joint Surg Am. 1993;75:1550–1561.
8. Wild M, Khayal T, Miersch D, et al. Die Zuggurtungsosteosynthese nach
negatively affect fracture healing and may to a certain extent Patellafraktur [in German]. Unfallchirurg. 2008;111:892–897.
explain the clinical observations of wire loosening, fracture dis- 9. Matejcic A, Smiljanic B, Bekavac-Beslin M, et al. The basket plate in the
location, and healing delays after tension-band wiring of patella osteosynthesis of comminuted fractures of distal pole of the patella. Injury.
fractures.4,7,8 The reduced amount of deformation under loading 2006;37:525–530.
10. Banks KE, Ambrose CG, Wheeless JS, et al. An alternative patellar frac-
may result in faster consolidation of the fracture and earlier ture fixation: a biomechanical study. J Orthop Trauma. 2013;27:345–351.
return to function.15 Our findings therefore suggest that gapping 11. Thelen S, Schneppendahl J, Jopen E, et al. Biomechanical cadaver testing
of the fracture under load-bearing is more likely to occur with of a fixed-angle plate in comparison to tension wiring and screw fixation
tension-band wiring as compared with locked plating. in transverse patella fractures. Injury. 2012;43:1290–1295.
This study not only has several strengths, but also 12. Yoo JH, Yi SR, Kim JH. The geometry of patella and patellar tendon
measured on knee MRI. Surg Radiol Anat. 2007;29:623–628.
limitations. One important strength of our study was the 13. Burvant JG, Thomas KA, Alexander R, et al. Evaluation of methods of
custom-designed testing device, which combined tension and internal fixation of transverse patella fractures: a biomechanical study.
bending in a physiological way. Furthermore, this testing J Orthop Trauma. 1994;8:147–153.
device and the use of Sawbones instead of human cadavers led 14. Curtis MJ. Internal fixation for fractures of the patella. J Bone Joint Surg
Br. 1990;72:280–282.
to a high reproducibility. The fracture gap could be measured 15. Augat P, Simon U, Liedert A, et al. Mechanics and mechano-biology of
with a high degree of accuracy by using a 3-dimensional fracture healing in normal and osteoporotic bone. Osteoporos Int. 2005;
image–based movement analysis. Finally, to our knowledge, 16:36–43.

e308 | www.jorthotrauma.com Copyright © 2015 Wolters Kluwer Health, Inc. All rights reserved.

Copyright © 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.

You might also like