You are on page 1of 10

Adult and paediatric ventilators

Quick Setup Guide


English

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 2 30.04.2015 14:02:42
Setup
Noninvasive use

AC Power cord Mask


Stellar
2 Air tubing

3 4
5
1

H4i heated humidifier


AC Power cord
AC locking clip

Invasive use
Antibacterial filter Catheter mount**
ResMed Leak Valve
HMEF*
2 2

5
1 5
1 3

3 7
6
6
4
4

AC Power cord External


humidifier* Air tubing
AC locking clip

*If not using an external humidifier, connect the HMEF to the ResMed Leak Valve.
** The ResMed Leak Valve or the HMEF can be connected to standardised tracheostomy interfaces
including connector pieces like catheter mounts.
The catheter mount and the external humidifier are not part of the ResMed component.

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 3 30.04.2015 14:02:50
About the control panel
External power supply LED Mains power LED LCD screen
On when external battery is On when operating on mains power.
connected. Menu button
Navigate through the menus
Internal battery LED Monitoring
On when internal battery is in use. Setup
Flashes when device is turned off Info
and battery is charging.
Push dial (rotate/click)
•• Navigate through settings.
Alarm LED •• Change settings’ values.
Red or yellow during an alarm •• Enter and confirm your
condition or during alarm testing. selection.

Therapy LED
Alarm mute button LED
• On during treatment
On when is pressed.
• F lashes during mask fitting
operation.
Alarm mute button
Start/Stop button
•• Silence the alarm
• Clear the alarm message (low •• Start/Stop ventilation
priority alarms). • Start mask-fit (hold down for at
least three seconds).

Rear panel
Oxygen inlet

Air filter cover

XPOD pulse oximeter connection DC power socket


AC power socket
FiO2 sensor connection
Data port for USB stick Power on/stand-by switch

Remote alarm connection

Data port for direct PC connection


3

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 4 30.04.2015 14:02:55
Turning the ventilator on/off
 o turn the ventilator on, press the power switch at the rear of the device. Wait until the Treatment screen
T
is displayed.
 o turn the ventilator off, press the power switch at the rear of the device. Follow the onscreen
T
instructions.

Accessing Clinical mode


Press and simultaneously for three seconds
until the duration options appear. Follow the onscreen
instructions.

To return to the Patient menu, press and


simultaneously for at least three seconds.

Treatment settings
Program
Ramp time

Warm-up humidifier

Yellow arrows display at the top or bottom of the pressure


bar graph will show if a machine trigger or cycle occured. If
the inspiratory phase extends beyond the spontaneous cycle
by Ti Min, a red arrow is displayed.
Selecting the mask type and performing a Learn circuit enables to calculate the impedance and leak for
controlled therapy pressure.

Learn Circuit configuration


Do not connect patient interfaces prior to performing the Learn Circuit. Patient interfaces include any
components placed after the ResMed Leak Valve (eg, HMEF, catheter mount, tracheostomy tube).

With an external humidifier Without an external humidifier


Antibacterial filter ResMed Leak Antibacterial filter ResMed Leak
Valve Valve

Air Air tubing


tubing

External Humidifier
4

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 5 30.04.2015 14:02:58
Using the menus
Menu overview

MONITORING SETUP INFO

Treatment Clinical Settings Event Summary

Monitoring Leak

Pressure / Flow Alarm Settings Minute Ventilation

MV/Resp Rate
or MV/Va (iVAPS Tidal Volume
Mode)

Options Respiratory Rate


Leak

I:E Ratio
Tidal Volume

Pressure Support
Synchronisation

AHI
Oximetry

SpO2

Used Hours

Device Information

Reminder

Note: iVAPS mode is only available in Stellar 150.

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 6 30.04.2015 14:02:58
Overview Clinical Settings

SETUP

Clinical Settings Advanced Settings

CPAP Mode

S Mode

ST Mode

T Mode

iVAPS Mode

PAC Mode

Note: iVAPS mode is only available in Stellar 150.


6

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 7 30.04.2015 14:02:58
Overview Alarm Settings and Options

SETUP

Alarm Settings

Options Configuration Menu

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 1 30.04.2015 14:02:58
Pathology defaults

Setting Mode Obstructive Restrictive Obesity Normal


iVAPS lung lung Hypoventilation lung
disease disease Syndrome mechanics
PAC
ST
S

IPAP [cm H 2 O] 13 11 16 11

EPAP or PEEP [cm H 2 O] 5 5 8 5

PS [cm H 2 O] 8 6 8 6

Rise time [ms]* 150 300 300 200

Fall time [ms]* 200 200 200 200

Trigger sensitivity Medium Medium Medium Medium

Cycle sensitivity High Low Medium Medium

Min PS [cm H 2 O] 4 4 4 2

Max PS [cm H 2 O] 20 20 18 20

Min EPAP [cm H 2 O] 5 5 8 5

Max EPAP [cm H 2 O] 15 15 15 15

Ti Min [sec] 0.3 0.5 0.5 0.5

Ti Max [sec] 1.0 1.5 1.5 1.5


*The Rise/Fall Time milliseconds scale is an approximate only.
Note: iVAPS mode is only available in Stellar 150.

See the Stellar Clinical Guide for full details.


Not all devices are available in all regions.

MANUFACTURER DISTRIBUTED BY
ResMed Germany Inc. ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
Fraunhoferstr. 16 ResMed Corp 9001 Spectrum Center Blvd San Diego CA 92123 USA
82152 Martinsried Germany ResMed (UK) Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK
See www.resmed.com for other ResMed locations worldwide. For patent information, see www.resmed.com/ip. ResMed and Stellar are
trademarks of ResMed Ltd. ResMed and Stellar are registered in U.S. Patent and Trademark Office.
© 2015 ResMed Ltd. 248804/1  2015-05

ResMed.com

248804r1 Stellar 100 & 150 Quick Setup Guide ROW ENG.indd 1 30.04.2015 14:02:41
Designed for versatility
Stellar series
Non-invasive / invasive ventilators
Stellar specifications Accessories
Modes iVAPS (intelligent Volume-Assured Pressure Support) with AutoEPAP*
H4i Heated Humidifier
ST (Spontaneous–Timed) 24930
PAC (Pressure Assisted Control) H4i water chamber
CPAP (Continuous Positive Airway Pressure) (disposable tub)
26952
S (Spontaneous)
Hospital Home T (Timed)
Pressure range IPAP: 2–40 cm H 2 O  tellar DC–DC converter
S
EPAP: 2–25 cm H 2 O 24922

TiControl™ Ti Max 0.3–4 sec


Ti Min 0.1–Ti Max
Respiratory Rate 5–60 bpm Mobility bag
24977
Rise Time Min, 150–900 msec (approx.)
Trigger and Cycle Five sensitivity settings
Adjustable alarms High Leak ResMed Xpod LP
Mobility Non-invasive oximeter 22374
Low Minute Ventilation
Oximetry reusable
Low and High Pressure sensor 22370
Low and High Respiratory Rate Oximetry disposable
sensor 22380
Apnoea
Low and High FiO 2
Low SpO 2 FiO 2 monitoring kit 24955
FiO 2 sensor CAP008617
Non-vented mask
Standard fixed alarms Circuit disconnected
Over pressure Remote Alarm II 27902
2 m cable 27904
Blocked tube 5 m cable 27905
Paediatric Invasive
Internal battery empty 10 m cable 27906
20 m cable 27907
Dimensions (LxWxH) 230 x 170 x 120 mm
Weight 2.1 kg  esMed Power Station II
R
battery and bag 24921
Air filter Electrostatic fibre mesh
DC cable 24959
Adult Air outlet 22 mm taper, compatible with ISO5356–1:2004
Power supply 90W
Internal battery Lithium-ion, 2 hours at EPAP 5, IPAP 15, 20 bpm 36824
Power supply AC 100–240V 50–60Hz,
2.2A AC 110V 400Hz, ResMed leak valve
2.2A DC 24V/3A 24988
IEC 60601–1 Classification Class II (double insulation) Type BF Continuous operation
* iVAPS with AutoEPAP is only available on Stellar 150
Stellar is ResMed’s premium non-invasive ventilator
series (with invasive capabilities) for patients with
changing ventilatory needs or progressive respiratory
Product codes ResMed standard trolley
24905
Stellar series:
the versatile choice
Stellar 100 Stellar 150
conditions, including chronic obstructive pulmonary Asia Pac 24164 24162
disease (COPD), neuromuscular disease (NMD) and Asia Pac Group 2 24170 24169
See the ventilation accessories catalogue for the full suite
obesity hypoventilation syndrome (OHS). Asia Pac Group 3 24168 24167 of ResMed accessories.

Easy to set up and use, Stellar carefully strikes a balance between ResMed Germany Inc. Fraunhoferstr. 16 82152 Martinsried Germany Stellar Distributed by: ResMed (UK) Ltd, 96 Jubilee Ave, Milton Park, Abingdon, Oxfordshire OX14 4RW. See ResMed.
providing you with pre-programmed settings that make your job easier, ResMed Humidification Technologies GmbH Karneolstr. 4 HumiCare D900 com for ResMed locations worldwide. CounterStream, H4i, HumiCare, ResScan, Stellar and TiControl are trademarks and/
or registered trademarks of the ResMed family of companies. For patent information, see ResMed.com/ip. Specifications
72250 Freudenstadt Germany
and the opportunity to fine-tune settings when needed to suit each ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia All others may change without notice. Product availability may vary across regions. © 2015 ResMed Ltd. 1018255/2 2015-07

individual patient – a versatile system with its end-users in mind.


ResMed.com/Stellar ResMed.com/Stellar
Meeting the demands of fast-paced hospitals Delivering optimal therapy to many types of patients Transitioning home and staying mobile
Stellar 100 and Stellar 150 are packed Access critical setup options Providing the modes you’d expect to see for Flexible therapy options Stellar has been thoughtfully engineered
with technologies to deliver a versatile quickly and easily excellent positive pressure ventilation, Stellar to not only streamline patient setup, but
Integrated oxygen: up to 30 L/min
ventilation solution, but not at the Critical setup options like Learn
also includes the auto-adjusting, volume- also streamline a patient’s transition to
High pressure delivery: up to 40 cm H2O at the mask
expense of easy setup and use. Circuit and Mask Type are now assured mode, iVAPS, with an intelligent NIV therapy at home.
also displayed on the first clinical Extensive alarms: including FiO2 and SpO2 to provide
Taking into consideration the busy menu screen you see, so you Backup Rate, and optional AutoEPAP feature added assurance for ‘at risk’ patients
can access important settings Small, lightweight and portable, Stellar is a suitable
hospital environment and typical immediately. – delivering what we call IntelligentAir.* Paediatric use: for children weighing down to as little ventilator for patients to return home with, fitting in
workflows, Stellar’s intuitive menus as 13 kg seamlessly with their lifestyle as they move about
IntelligentAir continuously monitors and automatically and carry on with their daily activities.
simplify the setup process, while adjusts to deliver personalised therapy to patients with Pass-over or CounterStream™ humidification: heated
its real-time data display makes changing ventilatory needs. humidification with the H4i™ or HumiCare™ D900
monitoring quick and always available. Dual programs: set and save up to two programs per
intelligent Volume-Assured Pressure patient for day/night ventilation or rest/active periods
Support (iVAPS): Targets each patient’s
Continuous monitoring at-a-glance Choose optimised settings using alveolar ventilation.
Personalised efficacy and comfort: using the following Powered to move
features, you can easily fine-tune settings on a patient-to-
Stellar’s large colour LCD screen displays essential Pathology Defaults patient basis Stellar gives patients who need extended ventilation the
data (such as pressure and flow curves, leak, tidal freedom to move – whether in hospital, at home or beyond.
A number of values are pre-set
volume, SpO2 and FiO2) in real-time. And for even and organised by disease intelligent Backup Rate (iBR): Provides backup Two hours capacity with an internal battery
greater insight, high-resolution (25Hz) flow and category to help facilitate breaths only when needed to give patients
therapy initiation. By optimising a chance to spontaneously trigger the ventilator. The power of one – or two – RPS II external batteries,
pressure data is also available via the ResScan™ settings per condition, staff are
patient management software. Vsync TiControl 5 Trigger Rise and
each providing an extra eight hours’ capacity
given an accurate starting point,
from which they can quickly fall times
Take it anywhere with a DC outlet with the new
fine-tune (if necessary) to
suit the individual patient’s AutoEPAP: Auto-adjusts the expiratory DC–DC converter
specific needs. pressure in response to partial and full
upper airway obstructions. Custom-built mobility bag protects the ventilator
when out and about
Ramp Ramp Down ResMed Power Station II Stellar mobility bag Stellar DC/DC converter

* iVAPS with AutoEPAP is only available on Stellar 150

You might also like