Professional Documents
Culture Documents
INTRODUCTION
1.1 Introduction
The Adsorbent, which is the solid that is highly porous in nature is key in any adsorption
process. It is introduced to selectively carryout the distribution and separation of molecules in
a bulk fluid mixture. Adsorbents are substances with an amorphous or microcrystalline
structure that might be natural or artificial. In terms of sales volume, the most often utilized
ones are activated carbon, molecular sieves, silica gel, and active alumina (Keller et al.,
1987).
Adsorption takes place when forces coming from a neighbouring surface hold molecules that
are diffusing in the fluid phase for a while. Atoms at the surface of a solid contain residual
molecular forces that are not met by surrounding atoms, such as those in the structure's body.
This surface reflects a significant discontinuity in the solid's structure (Perry and Green,
1999).
The only reason some materials are referred to as "adsorbents" is that they can be made in a
very porous form, giving rise to a large internal surface, because these residuals or van der
Waals forces are universal to all surfaces and responsible for withholding molecules (Coulson
and Richardson, 2002).
1
1.2 Aims and Objectives
The study of the principle of UV-spectroscopy is significant in so many ways, some of which
include:
CHAPTER TWO
2
LITERATURE REVIEW
2.1 Uv-spectrophometer
Due to its accessibility, ease of use, versatility, and wide range of applications in a number of
fields, including analytical chemistry and biochemistry, the UV-Vis analytical method has
grown significantly in importance and acceptance in several scientific fields throughout the
world (Porcu, 2018).
The amount of discrete wavelengths of UV or visible light that are absorbed by or transmitted
through a sample in contrast to a reference or blank sample is measured by the analytical
technique known as UV-spectroscopy. The sample composition has an impact on this
attribute, which could potentially reveal what is in the sample and at what concentration
(Tom, 2021)
3
When light passes through a sample solution in a cuvette, a UV spectrophotometer detects the
intensity of the light and compares it to the intensity of the light before the sample as shown
below
A stable source that can emit light over a variety of wavelengths is necessary for this light-
based method. It is typical and most common to employ a single xenon lamp as a high
intensity light source for both UV and visible wavelengths. However, compared to tungsten
and halogen lamps, xenon lamps are more expensive and less stable. A tungsten or halogen
lamp is frequently used for visible light in devices with two lamps, while a deuterium lamp is
typically the source of UV light. The light source in the device must swap during the
measurement since it takes two different light sources to scan both the UV and visible
wavelengths. (Tom, 2021).
The next step is to choose from among the variety of light wavelengths that the light source
emits those that are suitable for the sample type and analyte for detection for sample
inspection. There are several options for doing this which includes: monochromators,
absorption filters, interference filters, cut off filters and band pass filters. (Tom, 2021) of all
these monochromators are mostly used because it separates light into a narrow band of
wavelength and can be used with a filter for the purpose of being more precise.
4
2.2.3 Sample Analysis
A blank sample, a solvent, such as water or alcohol, is poured into a cuvette, a suitable,
transparent, and impermeable container. The light source's light beam travels through the
solvent-filled cuvette and is used as a reference for the sample whose absorbance and
transmittance are to be detected. A detector placed behind the cuvette containing the solvent
then measures and records the intensity of the transmitted light at a certain wavelength
(Cosimo and Haller, 2015)
A light beam generated by the light source passes through the cuvette with the sample after
the sample has been dissolved in the solvent and added. The sample molecules in the solution
partially absorb the light as it travels through the cuvette. The detector then measures the
transmitted light and the transmitted intensity of the sample solution is divided by the
equivalent values of the blank to determine the light intensity at a given wavelength. Finally,
a recorder stores this ratio. The detector is based on semiconductors and photoelectric coating
and then recorded by a computerized system (Tom, 2021).
A UV-spectrophotometer's detector monitors the light's intensity after it has passed through
the sample solution. The transmitted intensity, I, represents the percentage of light that the
detector has detected. For example, light at particular wavelength being absorbed, the sample
solution reduces the intensity of the transmitted light. Its value is consequently less than the
initial intensity I0 at the light source (Cosimo and Haller, 2015)
Transmittance T, which has the unit %, is the ratio of the two intensities I to I0
5
The primary, but not the only value that UV spectroscopy can determine is the transmittance.
Also, a second value can be determined while recording UV spectra, the absorbance A.
Absorbance represents an additional outcome that is frequently used. It is known as the
transmittance's negative logarithm and has a significant advantage. Absorbance is a
dimensionless value, that is a unitless value (Cosimo and Haller, 2015).
This relationship is called the Lambert-Beer law where the sample concentration c is
expressed in mol/L, the path length d of the cuvette is expressed in cm and the extinction
coefficient e (epsilon) is a sample specific constant describing how much the sample is
absorbing at a given wavelength in L/(cm*mol) or mL/(cm*g).
The sample concentration can be calculated using the obtained absorbance value and the
Lambert-Beer law. Concentration c can be determined from absorbance A as shown below if
the extinction coefficient and the route length d are known (Cosimo and Haller, 2015).
6
CHAPTER THREE
METHODOLOGY
3.1 Materials
1. Distilled water
2. Methylene blue
3.2 Apparatus
7
Table 3.1: Apparatus Used during the Experimentation
3.3 Diagrams
8
Figure 3.2: Beaker Figure 3.3: Spatula
9
3.4 Procedures
1. A clean and dry beaker was placed on a weigh balance and 0.1 gram of methylene
blue was weighed.
2. The 0.1 gram of methylene blue was mixed with distilled water in a beaker and
poured into a volumetric flask which was then topped with water up to the 1000ml
mark which would serve as the stock solution.
3. Then 10ppm of a solution was prepared in a 100ml cylinder by mixing 1.0ml of the
stock solution with 99ml of distilled water.
4. The uv-spectrophotometer was turned on and allowed to normalize for about
5minutes.
5. Then the sample chamber of the uv-spectrophotometer was opened and the rack was
pulled out and the cuvette was filled with a reference sample (water).
6. Then the sample chamber was closed and the reference sample was aligned to the rays
of UV light.
7. The transmittance button was pressed and the reading from the indicator was recoded
as 100% and using the mode button to navigate to the absorbance, the reading of 0%
was recorded.
8. Then the 10ppm solution sample prepared was placed in the sample chamber and
aligned with the position of the rays of light and its transmittance and absorbance
value were indicated and recorded.
9. The above step was repeated for the prepared 15ppm, 20ppm, 25ppm and 30ppm
solutions and their corresponding transmittance and absorbance readings indicated
and recorded.
10
CHAPTER FOUR
4.1 Results
Aborbance vs concentrati on
0.4
0.35
0.3 f(x) = 0.00644 x + 0.126
0.25
Absorbance
Absorbance
0.2
Linear (Absorbance)
0.15
0.1
0.05
0
5 10 15 20 25 30 35
Concentration
11
Figure 4.1: A graph of Absorbance Vs Concentration
After successfully carrying out the experiment by accurately following the methodology, the
results above were obtained. From the data above, it is observed that there exists a directly
proportional relationship between the concentration of a solution (sample) and absorbance.
that is as the concentration increased the absorbance increases. This relationship is supported
by the Absorbance Vs Concentration graph. Also, there exist an inversely proportional
relationship between the concentration of a solution (sample) and transmittance. That is as the
concentration increased the transmittance decreases.
12
CHAPTER FIVE
5.1 Conclusion
Following the successful completion of this study and careful examination of the results
obtained it is concluded that the concentration, transmittance and absorbance of a solution
were determined and there exists a relationship between these values. The relationship
between concentration and transmittance and concentration and absorbance is an inversely
proportional relationship and a directly proportional relationship. Hence, the aims of the
experiment were achieved.
5.2 Recommendation
In view of the success of this experiment, a couple of recommendations are made to further
enhance a more accurate result. Firstly, the liquid solution can be treated to remove impurities
so as to prevent the scattering of light caused by the presence of impurities in the absorbing
liquid. Secondly, the sample chamber should be clean and dry and the cuvette free from
finger prints. Thirdly, samples should be prepared according to specifications by avoid
random and parallax errors during measurements and the sample should be homogenous.
Fourthly, the procedure can be repeated and an average of the values used. Lastly, other
samples such as chromium salt can yield better and more accurate results hence
recommended for use.
13
REFERENCES
14
APPENDIX
x
x 1000 ppm=10 ppm
100 ml
x = 1.0ml
x
x 1000=1 5 ppm
100 ml
x = 1.5ml
x
x 1000 ppm=20 ppm
100 ml
x = 2.0ml
15
x
x 1000 ppm=2 5 ppm
100 ml
x = 2.5ml
x
x 1000 ppm=30 ppm
100 ml
x = 3.0ml
A= e.c.d
c = A/e.d
where c = concentration
A = absorbance
d = l = path length
The product e.L = slope of the calibration curve of the graph of Absorbance Vs Concentration
= 0.0064.
16
c = 0.354/0.0064 = 55.3125ppm
c = 0.760/0.0064 = 118.75ppm
c = 0.910/0.0064 = 142.1875ppm
17