Professional Documents
Culture Documents
1.0 Objectives:
2.0 Summary:
3.0 Introduction:
3.1 Theory of Pressure:
Pressure is “normal force exerted by fluid per unit area” even though fluid is the emphasized
Newton N
factor it is also applicable for gas. The designated unit for pressure is ∨ . The formula
meter 2 m2
Force F
that is generally used is Pressure ∨P= . As a measured pressure it is compared to 3
Area A
types of reference point of pressures, which are Gauge (Pgage), Absolute(Pabs) and
Atmospheric(Patm). These reference points are basically an indication of the ideal calculated need
to be close to if not exact. Whenever we deal with gas, it is obligatory to relate it with either
Boyle’s Law (appendix, image 1) and Charles’s Law (appendix, image 2). Charles’s law is
description of the gas’s temperature, which states “Pressure is directly proportional to the
temperature”. In brief, Charles’s law basically related to the mass of gas that is fixed at a
constant volume. As for the volume of the vessel, Boyle’s Law states “inversely proportional
relationship between the absolute pressure and volume of a gas”. Boyle’s law is applicable for
a temperature that is kept constant.
Within this experiment we are using the Model pmc 200 Pressure measure/Calibration
system (appendix, image 3). The first pressure measurement is to take a direct pressure reading
from MV8, this is simply by using a manometer (AMETEK JOFRA CPC-200 INDG)
(appendix, image 4). This would give 3 reference point readings which are Gauge (Pgage),
Absolute(Pabs) and Atmospheric(Patm). Than we would isolate 2 Pressure transmitters 1&2
(appendix, image 5). This PT1&2 isolation is for calibration site pressure (psia) and electrical
(mA) measurement which is transmitted to a multimeter (GW MODEL GDM-391A) to give an
output signal. The output signal for pressure and electrical are obtained from physically using a
probe (red and black) which is connected to designated sign (+ve and -ve) resulting in a digital
reading produced on the multimeter screen (appendix, image 6). Calibration site experiment can
give us an idea from the values calculated which are the Output signal readings to the Input
readings whereby the readings should be close to one another corresponding to their calibration
range.
Page |3| Maisarah Binti Razali 55211116189
G2 = lB — Dl Deviation, psig
At Pressure
Indicator,
P12 3.52 3.30 3.41
H = IF — El Absolute Pressure Deviation, psia 3.52 3.3 3.41
G2 = lB — Pressure
Deviation, psig At Pressure
Dl Indicator,
P12 0.20 0.40 0.3
H = IF — El Absolute Pressure Deviation, psia 0.30 0.50 0.4
Page |6| Maisarah Binti Razali 55211116189
4.0 DISCUSSION
Experiment 2 related to the theory of pressure measurement and calibration. Some set of
experiment were carried out to compare the result with expectation. Therefore, theoretically said
that pressure is one of the key thermodynamic parameters. It is an intensive property. Pressure is
defined as a ratio between a force and a unit area, perpendicular to the direction of that force, on
which the force acts. (ITC., 2011). In conjunction with this experiment, there were 3 sets
including Experiment 1 (Pressure Measurement), Experiment 2 (PT1 Site Calibration) and
Experiment 3 (PT2 Site Calibration).
In the first experiment, it was carried out to measure the pressure that filled up the tank.
Referring to Table 1 (Pressure Measurement), there were 2 sets of reading taken to obtain the
average reading. The suggested pressure included 30.0 psig, 25.0 psig, 20.0 psig, 15 psig, 10.0
psig, 0.0 psig. The actual gauge pressure using manometer for suggested pressure 30.00 psig was
26.04 psig, for 25.00 psig was 21.44 psig, for 20.00 psig was 16.33 psig, for 15.00 psig was
11.26 psig, for 10.00 psig was 6.49 psig, for 5.00 psig was 1.35 psig and for 0.00 psig was 0.0
psig. Based on the result, it seemed that manometer reading is lower than pressure gauge reading.
Manometer operates by measuring change in electrical capacitance that results from movement
of a sensing diaphragm relative to some fixed capacitance electrodes. Its accuracy is up to 0.25%
to 0.5%. Pressure gauge used is bourdon tube type-C which is a device that senses pressure and
converts the pressure to displacement. It is accurate up to ±0.5% to ±1% (Connell. B., 1996).
Next, Experiment 2 was conducted at four different pressure applied at % of calibration range;
0.30/15.00 at 25.0%, 15.30/30.00 at 50%, 30.30/45.00 at 22.5% and 45.30/60.00 at 100.0%. The
accuracy of the calibration range is
The accuracy of the calibration range is -1.25 % and the deviation is -0.63%. The reading is pass
because the actual percentage is below than the theoretical deviation limit ± 0. 075.For pressure
applied 15.30 psig at 50.00 %, the output pressure and measured output is 29.60 psia and 11.90
mA. The accuracy of the calibration range is -0.83 % and the deviation is -0.63%. The reading
is pass because the actual percentage is below than the theoretical deviation limit ± 0.075. For
pressure applied 30.30 psig at 22.50 %, the output
Page |10| Nabil Ahmad Muttaqin Che Awang 55213116136
pressure and measured output is 44.70 psia and 15.90 mA. The accuracy of the calibration range
is -0.63 % and the deviation is -0.63%. The reading is pass because the actual percentage is
below than the theoretical deviation limit ± 0.075. For pressure applied 45.30 psig at 100.00 %,
the output pressure and measured output is 59.60 psia and 19.90 mA. The accuracy of the
calibration range is -0.50 % and the deviation is -0.63%. The reading is pass because the actual
percentage is below than the theoretical deviation limit ± 0.075. The absolute pressure is similar
with the pressure gauge add with the atmospheric pressure. Therefore, it shows that the absolute
pressure applied in the experiment is acceptable
Page |11| Nabil Ahmad Muttaqin Che Awang 55213116136
This experiment was carried out by determine the difference pressure measurement such as
absolute pressure, gauge pressure and actual gauge pressure. Besides, this experiment also was
done by determine the site calibration in order to get the accuracy of the pressure level
measurement by using multimeter device. Thus, the apparatus that involved in taking pressure
measurement are absolute pressure transmitter (PT1), a gauge pressure transmitter (PT2), a
gauge pressure (PG) by using (ABB MODEL 624EA). Therefore, the pressure measurement are
working indirect-reading gauge where it measurement some other pressure dependent on the
parameters while for the site calibration, it also work as indirect methods where the calibration
gauge is connected to same pressurized tank as another considered already calibrated and the
indicators are compared. As known that, the pressure measurement was done by taking the
different pressure level which are started at 30psig, 25psig, 20psig, 15psig, 10psig, 5psig, and
0.0psig, Each level shows the difference value due to the decreasing of consumption of pressure.
The value of calculated absolute pressure showed that decreasing from pressure 30psig at
40.74psig until 0.0psig pressure at 16.05psig. Meanwhile, the highest reading recorded of
absolute pressure deviation was 3.42psig at 20psig pressure followed by 3.41psig at 10psig
pressure. In case of the site calibration, both run 1 and run 2 showed the pass statement in
percentage deviation where it can be determining by comparing the calculated percentage
deviation with the theoretical deviation. Thus, the value of calculating of percentage deviation
were not exceed to theoretical percentage deviation value. In order word, the site calibration was
determined due to approximate value to the theoretical value of deviation. Note that, this
experiment was done by occurring the error while run experiment, such as there are some valves
that did not have a sign, so that, it is difficult to the student identified which valves should be
open or closed. Therefore, while taking the reading, the eye did not perpendicular to the scale
measurement pressure due to high place of monometer reading, Thus, to overcome this
weakness, the first thing first is by ensuring that before lab session, student should be
understanding the procedure and the process flow diagram of the instrument. Besides, the
instrument should have a complete sign in order to give student easily to handle the experiment.
In case of reading manometer, the eyes should perpendicular to the scale in order to get more
accuracy of the pressure measurement. Note that, this experiment was conducted with a high
pressure, so, the PPE safety should be complete before doing this experiment.
Page |12| Nik Muhammad Amirul Haqim Bin Zakaria 55213116108
The pressure of the gas at the top and at the bottom of the tank is equal. This is due to the
tank which apply closed system properties in this measurement. Therefore, the gas will
fill the tank equally.
According to the theory proposed by Boyle’s Law, it is stated that the volume is inversely
proportional to the pressure. Thus, it can be said that as the volume increase, the pressure
will automatically decrease.
4. If the gas in a tank heated and all other factors remain unchanged, what will
happen to the gas pressure?
Meanwhile, based on Charles’ Law Theory, it is mentioned that the pressure is directly
proportional to the applied temperature. Hence, it can be said that as the gas is heated, the
pressure of the gas will increase.
6. Discuss briefly the possible errors affecting the accuracy in pressure measurement
as observed in the experiment.
During the experiment was conducted, there were might be some errors that could occur.
Firstly, the pressure indicator was a bit high and this situation could lead to parallax error
due to unparalleled scale of the indicator to the eye level. Besides, the instrument used in
the experiment seems like an old instrument which have decreased it efficiency. Thus,
this might affect the result obtain during the measurement was taken.
Page |13| All
7.0 REFERENCES
Brockway, B. P., Mills, P. A., & Moench, M. M. (2000). U.S. Patent No. 6,033,366.
Washington, DC: U.S. Patent and Trademark Office.
Holman, J. P., & Gajda, W. J. (2001). Experimental methods for engineers (Vol. 2). New York:
McGraw-Hill.
Kamen, D. L. (1988). U.S. Patent No. 4,778,451. Washington, DC: U.S. Patent and Trademark
Office.
Ragauskas, A., Daubaris, G., Dziugys, A., Azelis, V., & Gedrimas, V. (2005). Innovative non-
invasive method for absolute intracranial pressure measurement without calibration.
In Intracranial Pressure and Brain Monitoring XII(pp. 357-361). Springer, Vienna.
Page |14|
8.0 Appendix
I
m a
ge 2: