Professional Documents
Culture Documents
QUESTIONS their body, and the RF waves that emerge are detected.
From this, a picture of the patient’s insides can be built up
17 Two consecutive peaks in an ultrasound A-scan by computer. As we will see, MRI gives rather different
are separated by a time interval of 0.034 ms. information from that obtained by the other non-invasive
Calculate the distance between the two techniques such as X-rays or ultrasound.
reflecting surfaces. (Assume that the speed of
sound in the tissue between the two surfaces is Principles of nuclear magnetic resonance
1540 m s−1.) The nuclei of certain atoms have a property called spin,
18 Explain why an ultrasound B-scan, rather than and this causes them to behave as tiny magnets in a
X-rays, is used to examine a fetus. magnetic field. In MRI, it is usually the nuclei of hydrogen
atoms that are studied, since hydrogen atoms are present
in all tissues. A hydrogen nucleus is a proton, so we will
consider protons from now on.
Magnetic resonance imaging A proton has positive charge. Because it spins, it behaves
like a tiny magnet with N and S poles. Figure 32.30a shows
Magnetic resonance imaging, or MRI, is a diagnostic
a number of protons aligned randomly.
technique used in medicine. It can provide images
(including moving images) of the insides of a patient. It does
N
not rely on exposing patients to ionising radiation such as
S
N
S
N
N
S
X-rays; rather, it relies on the fact that some atomic nuclei
S
a N
N N
S
S
behave like tiny magnets in an external magnetic field. S
N
S
N
S
S
N
(MRI was originally known as nuclear magnetic N
N
S
S
S
N
S
S
N
resonance imaging, but the word ‘nuclear’ was dropped
N
S
S
because it was associated in patients’ minds with bombs
522 and power stations. To emphasise: MRI does not involve b N
radioactive decay, fission or fusion.) S N S S S
S S S S
N N N N
electromagnet
A proton does not align itself directly along the resonance requires a system with a natural frequency of
external field. In practice, its magnetic axis rotates around vibration; when it is stimulated with energy of the same
the direction of the external field (Figure 32.31), just like frequency, it absorbs energy. In MRI, protons precessing
the axis of a spinning top. This rotation or gyration action about the strong external field are exposed to a burst or
is known as precession. pulse of RF waves whose frequency equals the frequency
of precession. Each proton absorbs a photon of RF energy
path of path of and flips up into the higher energy state; this is nuclear
precession spin precession
magnetic resonance (Figure 32.32).
Now we come to the useful bit. The RF waves are
switched off and the protons gradually relax into their
lower energy state. As they do so, they release their excess
energy in the form of RF waves. These can be detected,
axis of axis of spin spin and the rate of relaxation tells us something about the
spin environment of the protons.
In Figure 32.32, you can see that the relaxation of the
protons follows an exponential decay pattern. Curves like
this are characterised by two relaxation times:
gravitational field magnetic field
■■ T1, the spin–lattice relaxation time, where the energy of the
Figure 32.31 A spinning top (left) rotates about its axis; at the
spinning nuclei is transferred to the surrounding ‘lattice’ of
same time, its axis precesses about the vertical, which is the
nearby atoms
direction of the gravitational field. In a similar way, a proton
(right) spins and its axis of rotation precesses about the ■■ T2, the spin–spin relaxation time, where the energy is
direction of the external magnetic field. transferred to other spinning nuclei.
λB0
f0 =
2π
In an MRI scanner, the external magnetic field is very
strong, of the order of 1.5 T (thousands of times the
strength of the Earth’s field). The precession frequency f 0
is: 0
0 Time
2.68 × 108 × 1.5 Figure 32.32 In nuclear magnetic resonance, a spinning
f0 = = 6.4 × 107 Hz = 64 MHz
2π nucleus is flipped into a higher energy state when it absorbs
This frequency lies in the radio frequency (RF) region a photon of RF energy; then it relaxes back to its lower energy
of the electromagnetic spectrum. You should recall that state.
Cambridge International A Level Physics
This means that different tissues can be distinguished by value for resonance and so the computer can precisely
the different rates at which they release energy after they locate the source of the RF signal within the patient’s body
have been forced to resonate. That is the basis of medical and construct an image.
applications of nuclear magnetic resonance. ■■ A computer that controls the gradient coils and RF pulses,
and which stores and analyses the received data, producing
and displaying images.
QUESTIONS
longitudinal
19 Protons precess at a frequency of 42.6 MHz in an gradient coil RF receiving coil
external field of magnetic flux density 1.0 T. large
a Determine the frequency at which will they
external
magnet
precess in a field of magnetic flux density 2.5 T.
b State the frequency of RF radiation that will
cause the protons to resonate in this stronger
magnetic field.
20 Figure 32.33 shows how the amplitude of RF
waves coming from watery tissue varies after RF transmitting coil
resonance. Copy the graph and add lines and
labels to show the graphs you would expect to computer
see for cancerous and fatty tissues.
Figure 32.34 The main components of an MRI scanner.
Procedure
Amplitude
MRI scanner
Figure 32.34 shows the main components of an MRI
scanner. The main features are:
■■ A large superconducting magnet which produces the
external magnetic field (up to 2.0 T) needed to align the
protons. Superconducting magnets are cooled to 4.2 K
(−269 °C) using liquid helium.
■■ An RF coil that transmits RF pulses into the body.
■■ An RF coil that detects the signal emitted by the relaxing
protons.
■■ A set of gradient coils. (For clarity, only one pair of gradient
coils is shown in Figure 32.34.) These produce an additional
external magnetic field that varies across the patient’s
body. These coils are arranged such that they alter the
magnitude of the magnetic flux density across the length,
depth and width of the patient. This ensures that the
Larmor frequency of the nuclei within the patient will be Figure 32.35 MRI scan through a healthy human head.
slightly different for each part of the body. This means that Different tissues, identified by their different relaxation times,
only a small volume of the body is at exactly the right field are coloured differently.
Chapter 32: Medical imaging