Diapact®

CRRT System
Technical Features and Specifications An integrated system for high flow continuous or
intermittent renal replacement and plasma therapies
• Easy access to internal components
• PC-based diagnostics system assists
technician in isolating and correcting
machine problems as well as conducting
calibration and preventative maintenance
procedures

Technical Specifications
Dimensions and weight Blood side
Dimensions (W x H x D).................. 19 in. x 49.5 in. x 19.5 in. Blood flow rate...................................................... 10 − 500 mL/min.
Weight.............................................................................100 pounds Blood volume range.........................................................Up to 999 L
Arterial pressure range................................... -400 — +100 mmHg
Electrical data
Venous pressure range.............................................. 0 − 400 mmHg
Rated voltage.............................................110/240 V - 50/60 Hz
Filter inlet pressure range........................................ 0 − 500 mmHg
Current........................................................................................3.5 A
Safety air detector............................. Ultrasound type connection
Battery power supply (option)............................................... 12 V
External Anticoagulant Pump........................”Smart“ connection
Electromagnetic compatibility........................... EN 60601-1-2
Fluid side
Operating conditions Intermittent Dialysate flow rate.........5 − 400 mL/min. (24L/hr.)
Relative humidity.......................................................30% to 90% Continuous Dialysate flow rate...........5 − 200 mL/min. (12L/hr.)
Temperature....................................................................15 − 40° C Substitution flow rate
Atmospheric pressure......................................700 − 1060 mBar        system control............................. 5 − 100 mL/min. (6L/hr.)
Staff call........................................................................... Connector Ultrafiltration flow rate.........................0 − 300 mL/min. (18L/hr.)
Run Time Clock..................................................................Standard Intermittent Weight loss goal range.....................0 − 2000 mL/h
Continuous Weight loss goal range.......................0 − 2000 mL/h Flexible design allows multiple treatments:
SCUF, CVVH, CVVHD, CVVHFD, PEX, PAP, HD/FD
Fluid balance goal range.................................................... 2 − 25 Kg
Other CRRT Products from B. Braun include Fluid temperature range.................................................. 20 − 40° C
hemofilters, standard dialysate, substitution fluids Inlet/outlet filter pressure range..................-400 − +400 mmHg
and customized fluids from CAPS® Blood leak detector....................................................... Red sensitive
(Central Admixture Pharmacy Services). TMP Monitoring Range......................................... 100—600 mmHg

Schedule a demo today.


Call your B. Braun Representative
at 1-800-848-2066 or email rtd@bbraun.com

www.bbraunusa.com
Rx Only. ©2010 B. Braun Medical Inc., Bethlehem, PA. All Rights Reserved. RTD.001 Rev. 4.10 JH
Versatility and Multifunctionality Innovative Technical Features
The Diapact CRRT System can be used to perform the following therapies:

Continuous Renal Therapies Continuous Renal Therapies Intermittent Dialysis Therapies


SCUF HF
Intermittent Dialysis Therapies BLOOD BLOOD
10-500
BLOOD BLOOD
10-500
mL/min. mL/min.
HEMOFILTER HEMOFILTER

50-2000 0-2000
mL/hr. mL/hr.

Plasma Therapies* ULTRAFILTRATE SUBSTITUTION ULTRAFILTRATE

Slow Continuous Ultrafiltration Hemofiltration

CVVH HD
BLOOD BLOOD BLOOD BLOOD Integrated Warmer Single Weighing System Battery Backup (available option)
10-500
mL/min.
10-500
mL/min. Warming of infusion and dialysate In the Diapact® CRRT System, the solution Diapact® CRRT System can be
HEMOFILTER DIALYZER fluids is sometimes an important and ultrafiltrate bags are connected to equipped with an optional battery
300-6000 0-1000 5-400 requirement for ICU patients which can a single load cell. The simplicity of this pack which allows the system to
mL/hr. mL/min.
mL/hr.
DIALYSATE DIALYSATE
be satisfied by means of the integrated weighing system reduces the possibility continue treatment, even when the
SUBSTITUTION ULTRAFILTRATE
(Fresh) (Spent) plate warmer. of error and increases accuracy. patient is moved or during a power
outage.
Continuous
Veno-Venous Hemofiltration Hemodialysis
Filtration/ Backfiltration System CVVHFD — Filtration-Backfiltration System
CVVHD/HFD HFD Using a high flux filter at low UF rates To patient
Filtration Backfiltration
BLOOD BLOOD BLOOD BLOOD
produces backfiltration at the dialysate
10-500 10-500
mL/min. mL/min. inlet of the filter. This backfiltrated fluid From patient
DIALYZER DIALYZER Blood Pump
serves as a substitution fluid. The patient‘s Weighing system
High-Flux Filter
5-200 5-400
mL/min. mL/min.
fluid balance is maintained by the Diapact® Dialysate
Out
Dialysate
In
DIALYSATE DIALYSATE DIALYSATE DIALYSATE
(Fresh) (Spent) (Fresh) (Spent)
CRRT System’s single weigh scale.
The result of this therapy (CVVHFD —
Continuous Veno-Venous Dialysate
Hemodialysis High Flux Dialysis Continuous Veno-Venous High Flux Dialysis) bags
UF=O
High Flux Dialysis is a combination of convective and diffu- Air sensor
(SAD)
sive clearances combined with the middle Warmer
Clamp
molecule clearance capability of the high
Plasma Therapies
flux filter.
PEX PAP
BLOOD
BLOOD BLOOD BLOOD
10-300 10-300
mL/min. mL/min. Reference
PLASMA PLASMA
FILTER FILTER Fig. CVVHFD – Filtration-Backfiltration System Lonnemann G., Bechstein M., Linnenweber S., Burg M., Koch KM.:
5-50
Clark William R., Ronco C.: Renal replacement therapy in Tumor necrosis factor-alpha during continuous high-flux hemodialysis
0 -60 -2000—3000 0-6000
mL/min. mL mL/min. mL acute renal failure: Solute removal mechanisms and dose in sepsis with acute renal failure. Kidney Int Suppl. Nov; 72:S84-7, 1999
SUBSTITUTION
FLUID
FILTRATE COL-
LECTION
ADSORPTION/
PERFUSION
quantification. Kidney Int 53 (Suppl 66):S133-S137, 1998
Ronco C., Brendoan A., Crespaldi C., Feriani M., LA Greca G.: Continuous
Clark William R., Soltys, Paul J.: Hemodialysis Systems for High Flux Dialysis (CHFD): A New Form of Continuous Renal Replace-
Therapeutic Plasma Exchange Plasma Adsorption/Perfusion Intermittent, Semi-Continuous, and Continuous Therapies ment Therapy in Intensive Care Patients. Proc. 3rd Int. Satel. Symposium
In Acute Renal Failure. Home Hemodialysis Int Vol-2, 30-33, 1998 on ARF, Halkidiki 1993

* Plasma Therapy is optional.