Reprint

LITHOSTAR Modularis –
A Proven System with New, Innovative Features

B. Granz, M. Lanski, R. Nanke, M. Mahler, A. Rohwedder, C. Chaussy, S. Thüroff


Reprint from electromedica 1, 2003, page 53-57
Innovative Methods – Advanced Modalities

LITHOSTAR Modularis –
A Proven System with New, Innovative Features
B. Granz 1, M. Lanski 1, R. Nanke 1, M. Mahler 1, A. Rohwedder 1, C. Chaussy 2, S. Thüroff 2

1
Siemens AG, Medical Solutions, Erlangen, Germany
2
Munich-Harlaching Teaching Hospital, Ludwig Maximilian University, Munich, Germany

Keywords
80
• lithotripsy module • shock wave parameter
• shock wave head • C-arm coupling 60

Pressure [MPa]
40

20

Abstract -20 Steep edge caused


by nonlinear effects
Two innovations by the Siemens Medical Solutions -40
division Med SP expand and optimize the potential 0 1 2 3 4 5 6
applications for the clinical workflow of the known Time [µs]
urological platform MODULARIS Uro Plus.
First, another lithotripsy module, the LITHOSTAR™ Fig. 1
Typical time-dependent behaviour
Modularis Vario (Siemens AG, Medical Solutions), of pressure in shock wave focus.
was developed in addition basing on the conventional
LITHOSTAR Modularis, which features a special
mechanical coupling system that now also allows the
adaptation of the SIREMOBIL Compact L, along with Introduction
the use of numerous X-ray C-arms from other manu- Extracorporeal shock wave lithotripsy (ESWL), the
facturers. Second, a new shock wave therapy head was disintegration of urinary stones using focused ultrasound
developed for both lithotripsy modules, the LITHOSTAR shock waves, has been the method of choice in uro-
Modularis and LITHOSTAR Modularis Vario, with the lithiasis since its introduction over 20 years ago. The
objective of increasing efficiency relative to safety of electromagnetic shock wave head, first introduced on
treatment, ease of use and patient comfort. the market with the LITHOSTAR from Siemens, has
Mathematical modeling has made it possible to gained universal acceptance among the three physical
further improve the shock wave parameters. Increasing principles for shock wave generation. Even many users
the penetration depth to 14 cm makes it easier to adjust of the “first-generation” electrohydraulic shock wave
the shock wave focus on the concretion to be treated, generation principle or the piezoelectric principle have
especially in adipose patients, thereby resulting in a switched to the electromagnetic shock wave. After three
distinctly elevated rate of therapeutic effect, and hence successive lithotripter systems generations that were
efficacy. introduced in 1980, 1987 and 1998, respectively, there
appears to be little leeway for new development as loca-
In addition, the conventional spark gap in the electri- lization, operation and patient comfort became increas-
cal circuit was replaced by the new semiconductor switch ingly perfected. While there might be no revolutions,
for triggering the shock wave, reducing the maintenance
consistent progress requires that we take numerous small
considerably.
steps in the right direction. To this end, we always draw
These innovations and new products allow for better on verified research results, analytically classifying and
customer benefits of the modular urological platform utilizing other worldwide results of our customers in
Siemens MODULARIS. addition to our own studies.
53 electromedica 71 (2003) no. 1
Lithotripters generate pressure pulses that are Materials and Methods
steepened up by nonlinear propagation to shock waves Modeling Methods for Optimizing Shock Wave
in the focus, which is directed at the urinary stones, and Generation
appear impressive in their steepness and maximal peak
pressure values (Fig. 1). Pressure peaks normally range Mathematically modeling the formation and propa-
from 50 to 150 MPa (500 to 1500 bar). And yet, this gation of the shock wave from the electromagnetic source
often propagated and advertised peak pressure value [2] up to the focus is indispensable for fine tuning and
alone is irrelevant for stone disintegration. It has been utilizing optimal components. How an electromagnetic
scientifically proven that the energy content of an shock wave source works is described in detail in [6]
expanded focus with moderate peak sound pressure along with the physical correlations. A charged high-
results in extensive stone disintegration, and not the voltage capacitor is instantaneously discharged via a flat
coil, in front of which a metal membrane is situated as a
narrow, high-pressure peak [1]. We have always used
secondary coil. The mutual magnetic repulsion between
this principle in our LITHOSTAR products [2, 3], and
the two coils (Lenz’ rule) pushes the membrane in
this with great success [4]. Current studies that describe
an adjacent water volume, thereby generating a plane
the process of stone disintegration in lithotripsy [5]
pressure wave, which is then concentrated in the focus
confirm this approach. We are currently offering the
by an acoustic lens. This plane wave directly after the
LITHOSTAR Multiline and LITHOSTAR Modularis
membrane already embodies all characteristics essential
(Fig. 2), two lithotripters that permit all urological stone for stone disintegration, e. g., energy.
treatments.
In principle focusing merely concentrates the initial
We are also continuing to consistently pursue this sonic energy on a small area, thus increasing the field
successful course with our latest shock wave therapy parameters pressure and intensity by the focal gain.
head, which is being implemented in the LITHOSTAR Absorption and reflection are neglected in this context.
Modularis and LITHOSTAR Modularis Vario lithotripsy The local effect of the shock wave in the focus solely ori-
modules. Described below are the novel methods and ginates from the raising of the pressure and the intensity
components that have been incorporated in the new by the focal gain above the stone disintegration threshold
shock wave module: Modeling methods for optimizing [7], a value characteristic for kidney stones.
shock wave generation, increase in the penetration depth
to 14 cm, use of a semiconductor switch and the mecha- It is sufficient to consider the initial plane pressure
nical coupling system for various X-ray C-arms. wave at the membrane to optimize shock wave energy.

Fig. 2
Presentation of
MODULARIS Litho with
MODULARIS Uro
endourology table,
SIREMOBIL Iso-C X-ray
C-arm and LITHOSTAR
Modularis lithotripsy module.

electromedica 71 (2003) no. 1 54


Innovative Methods – Advanced Modalities
To this end, a finite element program is used to model
the flat coil with its individual windings and the mem-
brane. Since we assume an axially symmetrical system, a
Membrane two-dimensional calculation is sufficient. The tightly
packed magnetic field between the coil and membrane
Axis of Symmetry

along with the low ambient stray field emerge as the


initial result (Fig. 3).
Additional results of this calculation include macros-
Flat coil
Magnetic
copic values like stray inductance and electrical resis-
field lines tance of the flat coil. The time-dependent currents and
magnetic forces between the flat coil and membrane
can then be calculated for the oscillator comprised of
the high-voltage capacitor, flat coil and additional
Fig. 3 cable inductances (Fig. 4). Solving Newton’s differential
Magnetic field distribution
between flat coil and membrane equation of motion, which includes the membrane mass
as well as in environment. to be accelerated and the water compression adjacent
to the membrane as additional components, results in the
time dependent displacement z(t) of the membrane. All
other quantities, like membrane pressure or energy, can
Switch
be derived from this displacement z(t) and the known
surface area. As a result, the ultrasound characteristics of
the system are set.
High voltage

Coil

Capacitor
Fig. 5 shows the time-dependent displacement z(t)
Resistor of the membrane for three different coil configurations.
As evident, there is rapid movement within the first 5 µs,
until a plateau with a displacement of about 10-30 µm is
reached after about 15 µs when the damped oscillation of
the oscillator has decayed. A high displacement means
Fig. 4
Schematic diagram of an electro-
both high pressure at this location and high mechanical
magnetic shock wave source. tensile stress on the membrane. As an alternative, the
Closing the switch initiates the desired high energy can be achieved by increasing the
shock wave generation. winding surface area at a smaller membrane displace-
ment. Varying the sole input variable “flat coil winding”
hence makes it possible to find the optimum pressure
and energy values for the target values high source
30
service life and stone disintegration.
Configuration A
Membrane extension [µm]

25

20 Increase of Penetration Depth to 14 cm


15
Configuration B The above statements would suggest that it makes
no difference if the stone to be disintegrated lies in the
10
Configuration C axial focus or not, since the shock wave energy is identi-
5 cal in areas in front of and behind it (disregarding tissue
0
absorption). By contrast, it is true that disintegration is
0 5 10 15 20 25 30 most effective in the focus, since the energy here is
Time [µs] concentrated on an area tailored to the stone size, and
a high percentage of the pressure is above the disinte-
Fig. 5 gration threshold. As shown by the axial distribution of
Time-dependent behaviour of positive pressure peak p + in Fig. 6, the source of the two
membrane extension for various
coil configurations. MODULARIS lithotripsy modules now has a focus
distance of 14 cm instead of the previous 12 cm. With the
matched penetration depth of up to 14 cm, the new shock
wave source conforms to the anatomical realities and
expanded indications of ESWL, and can simplify treat-
ment courses in obese patients.
electromedica 71 (2003) no. 1 55
In addition to this application-related advantage,
closer examination of the basic physical relations
reveals another positive effect of increasing the axial 100

Normalized pressure [%]


New
focus position. The following applies by approximation Previous Modularis
for the full width half maximum (fwhm) ∆x of the lateral 80 shock wave shock wave
source source
focal pressure distribution of a monofrequent acoustic
60
wave:
∆x = 1.4 •  • F / A (1) 40

where  is the wavelength in the propagation medium, 20


180
100 120 140 160
F is the focus distance and A is the effective aperture of
the focusing lens. The axial fwhm ∆z is about 10 times Axial distance [mm]

greater than ∆x. Even though an ultrasound shock wave


pulse in the focus is far from monofrequent (Fig. 1), this Fig. 6
approximation is helpful. As evident from formula (1), Axial pressure distribution and
location of focus.
shifting the focus position from 12 to 14 cm increases
the fwhm by the same ratio, i. e., a greater active focus
is obtained for stone disintegration, when the energy
of the shock wave is increased concurrently. In addition
to simplifying adjustment, this also increases the shock Mechanical Coupling System for
wave treatment efficiency, and a lower retreatment rate Various X-Ray C-Arms
is prognosed [8]. All effects achieved by increasing The modular technology in the LITHOSTAR
the focus position are therefore of the same positive Modularis for using other high-quality equipment
nature. components (MODULARIS Uro endourology table,
SIREMOBIL Iso-C X-ray C-arm, SONOLINE Adara
Implementation of a Semiconductor Switch ultrasound device) represents an attractive enhancement
for use even outside ESWL applications. To increase
Fig. 4 shows the schematic diagram of an electro- flexibility and variability even further for the customer,
magnetic shock wave source. When the switch is closed, the LITHOSTAR Modularis Vario lithotripsy module
the charged capacitor is discharged via the flat coil, can also be combined with the SIREMOBIL Compact L
during which the entire capacitor current flows through and numerous X-ray C-arms of other manufacturers, as
the switch. Capacitor voltages of 20 kV and peak currents opposed to the LITHOSTAR Modularis.
of 10 kA place high demands on the switch. Spark gaps
triggered by an additional high-voltage pulse are com- A mechanical coupling system adjustable for many
monly used today for these switches. The disadvantage X-ray C-arms is clamped onto the extension of the chassis
lies in the limited service life, which means the spark of the X-ray C-arm. The lithotripsy module is docked and
gaps need frequent replacement, and in the added elec- latched to this mechanical coupling system of the X-ray
tronics complexity involved in ignition with another C-arm by means of a precision yoke. The adjustment
high-voltage pulse. phantom connected to the shock wave head and a cross-
hair on the image intensifier are used to manually align
In using a semiconductor switch, we opted for a novel the central beam of the C-arm to the isocenter via X-ray
solution. Its advantages are smaller package size, longer imaging. The base of the MODULARIS Uro table is pro-
service life and simpler triggering by a 5 V signal. Of vided with another positioning yoke that enables the
course, this switch must fulfill very special conditions, alignment of the endourology table relative to the X-ray
e. g., a low on-resistance or a fast switching phase in the C-arm. Therapy can then take place in the usual manner.
ms range. Semiconductor switches demonstrated their Due to this innovative mechanical coupling system, the
superior properties in service life tests, so that this part lithotripsy module makes it possible to treat stones to-
must no longer be replaced. Because these switches are gether with numerous different X-ray C-arms, including
electrically indistinguishable from the ignited spark those made by various manufacturers.
gaps in the ON state, the oscillator according to Fig. 4
exhibits the same time and amplitude behavior, and Consequently, we are also giving even those customers
hence an identical magnetic force field between the flat who already own a X-ray C-arm the opportunity to treat
coil and membrane, resulting in an identical shock wave stones, which is particularly interesting for customers
generation. As a consequence, this part can be imple- on a limited budget. Therefore, we are providing price-
mented instead of the spark gap resulting in the same sensitive markets as India, Brazil or Asia with access to
therapeutic efficacy. These new semiconductor switches modern medical technology, and are entering into com-
are available to all customers worldwide. petition with the low-end manufacturers in these regions.
56 electromedica 71 (2003) no. 1
Innovative Methods – Advanced Modalities
The conventional LITHOSTAR Modularis litho- Literature
tripsy module continues to offer a convenient, simple [1] Granz B, Köhler G. What makes a Shock Wave Efficient in
and reliable coupling to the SIREMOBIL Iso-C and Lithotripsy. J. Stone Disease 1992; 4:123-128.
SIREMOBIL Iso-C3D X-ray C-arms. Other advantages [2] Reichenberger H, Naser G. Electromagnetic Acoustic Source
of this established interface include central operation for the Extracorporal Generation of Shock Waves in Lithotripsy.
using only a single hand-held remote control and moto- Siemens Forsch.- u. Entwickl.-Ber. 1986; 15(4): 187-194.
rized X-ray C-arm angulation for localizing stones. This [3] Folberth W et al. Pressure Distribution and Energy Flow in the
permits easy operation of the complete system even Focal Region of two different Electromagnetic Shock Wave Sources.
from a greater distance, e. g., behind a lead glass wall. J. Stone Disease 1992; 4(1): 1-7.
[4] Teichmann JMH et al. In Vitro Comparison of Shock Wave
Result Lithotripsy Machines. J. Urol. 2000; 164: 1259 -1264.
The LITHOSTAR Modularis modular platform has [5] Xi X, Zhong P. Dynamic Photoelastic Study of the Transient Stress
Field in Solids during Shock Wave Lithotripsy. J. Acoust. Soc. Am. 2001;
for years proven itself highly effective in clinical practice 109(3): 1226-1239.
as a diagnostic and therapeutic urological device. The
[6] Eisenmenger W. Elektromagnetische Erzeugung von ebenen
new LITHOSTAR Modularis Vario lithotripsy module Druckstößen in Flüssigkeiten. Akustische Beihefte, ACUSTICA, 1962;
combines the proven attributes of the previous module 12: 185-202.
with the following advantages: The effective ultrasound [7] Granz B, Köhler G. Quantitative Assignment of Shock Wave Para-
shock wave with the increased penetration depth ensures meters to Stone Disintegration in Lithotripsy. Conference Proceedings
a high level of patient comfort and an improved ease of Ultrasonics International, Butterworth Heinemann, 1991: 61- 64.
operation. The increased service life and optional inter- [8] Lingeman JE. Extracorporeal Shock Wave Lithotripsy – what
face for various X-ray C-arms also represents an impor- happened?. J. Urol. 2003; 169: 63.
tant step in the right direction with regard to the constant
discussion of costs in health care.
The LITHOSTAR Modularis Vario lithotripsy module
presents an advanced solution that continues on the path
of success achieved by the LITHOSTAR product family.
Author’s address
Dr. Bernd Granz
Calculations relating to the magnetic field and pressure pulse generation Siemens AG, Medical Solutions
were performed by Mr. Röckelein at Siemens Corporate Technology, Special Systems
Erlangen. Hartmannstr. 48
D-91052 Erlangen
Germany
The information about both the LITHOSTAR Modularis Vario and Tel.: +49 (0) 9131 84-4987
new therapy head is being provided for planning purposes. The product Fax: +49 (0) 9131 84-6387
requires 510(k) review and is not commercially available in the U.S. e-mail: bernd.granz@siemens.com

The views expressed by the author do not necessarily


coincide with the opinion of the Siemens company.

Original images always lose a certain amount


of detail when reproduced.

Siemens AG, Medical Solutions


Henkestr. 127, D-91052 Erlangen
Germany
Order No. A91100-M1500-C505-1-7600
Telephone: ++49 9131 84-0 Printed in Germany
www.SiemensMedical.com CC 65505 SD 07032.