Professional Documents
Culture Documents
Fetal Monitoring
1 Table of Contents
1 General Information 5
2 Site Preparation 13
Introduction 13
Roles and Responsibilities 13
Site Requirements 15
Site Planning 17
3 Installation 25
4 Theory of Operation 43
5 Disassembly/Reassembly 47
Introduction 47
Tools Required 48
Disassembly Top Cover Avalon CL Base Station 48
Disassembly Short Range Radio Boards 52
Disassembly CL Pod Device Charger 54
Disassembly Charging Contacts 57
Disassembly Standby Button/Switchboard 62
Disassembly Bottom Cover Avalon CL Base Station 65
Disassembly CL Transducer Battery 67
3
6 Spare Parts 77
8 Troubleshooting 95
Overview 95
System is Completely Inoperative 95
System Reset 96
Checking the Fetal Monitor 96
Troubleshooting the Top Cover 96
Technical INOPs Avalon CL 98
Common Problems 102
RF Problems 105
9 Specifications 111
10 Upgrades 133
Index 135
4
1
1 General Information
This guide describes:
• how to service and repair the Avalon CL base station (866074) and the transducers CL Toco
(866075), CL US (866076), and CL ECG/IUP (866077) hardware and software
• how to diagnose operating and servicing problems
• how to test the system
The Avalon CL Fetal Transducer System Service Guide supplements the maintenance and
troubleshooting procedures, carried out by the operator that are described in the Instructions for Use,
and the Service Guides of the fetal monitors FM20/30 and FM40/50. Refer to the Instructions for
Use for maintenance and troubleshooting procedures that may be performed during normal operation.
For detailed information on the IntelliVue CL Pods that can be used with the Avalon CL base station,
please refer to their Instruction of Use and Service Guide.
Only authorized and qualified service personnel should attempt to install the system, disassemble the
base station, remove or replace any internal assemblies, or replace the transducer batteries or belt
buttons.
What to Do Next
Familiarize yourself with the contents of this guide and the Instructions for Use before attempting to
service or repair the system.
5
1 General Information
Repair Strategy
The IntelliVue Support Tool software helps you to determine whether a fault is a hardware, software,
or an RF related problem. Any maintenance and repair procedures beyond the level covered in the
Instructions for Use are limited to:
Unit exchange for:
• base station
• transducers
• external power supply
Replacement of:
• top and bottom housing
• transducer battery
Repair or replacement of individual components on the boards is not supported, and should never be
attempted.
WARNING
Do not maintain or repair the device in patient vicinity.
6
1 General Information
Electric Hazards
WARNING
• Do not use additional AC mains extension cords or multiple portable socket-outlets. If a multiple
portable socket-outlet is used, the resulting system must be compliant with IEC 60601-
1:2005+A1:2012 / EN 60601-1:2006+AC:2010.
• Do not connect any devices that are not supported as part of a system.
• Do not use a device in the patient vicinity if it does not comply with IEC 60601-1:2005+A1:2012
/ EN 60601-1:2006+AC:2010. The whole installation, including devices outside of the patient
vicinity, must comply with IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010. Any non-
medical device, including a PC running an OB TraceVue/IntelliSpace Perinatal system, placed and
operated in the patient's vicinity, must be powered via a separating transformer (compliant with
IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010) that ensures mechanical fixing of the
power cords and covering of any unused power outlets.
• Only the power cables provided with the system may be used. For reasons of safety, power (mains)
extension cables or adapters shall not be used.
• Observe ESD (electrostatic discharge) precautions when working within the unit.
• The use of accessories, transducers, and cables other than those specified may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the device.
• Energized circuits are accessible with the covers open. Do not work on the base station with the
covers open while the base station is connected to a monitor. Only authorized and qualified
service personnel should open or disassemble the base station.
• Before attempting to open or disassemble the base station, disconnect it from the fetal monitor.
7
1
• For paced patients: The radiated SRR power of the CL SpO2 and CL NBP Pods, and other
sources of radio-frequency energy, when used in very close proximity of a pacemaker, might be
sufficient to interfere with pacemaker performance. Due to shielding effects of the body, internal
pacemakers are somewhat less vulnerable than external pacemakers. However, caution should be
exercised when monitoring paced patients.
• In order to minimize the possibility of interference, avoid positioning and wearing the Cableless
Measurement Devices in very close proximity to a pacemaker. Consult the pacemaker
manufacturer for information on the RF susceptibility of their products.
Use Environment
WARNING
Liquid Ingress:
• Before operation, ensure that the base station is free from condensation. This can form when
equipment is moved from one building to another, and is exposed to moisture and differences in
temperature.
Never immerse the base station in liquid. You must protect it against water sprays or splashes.
Place the base station where there is no chance of contact with, or falling into water, or other
liquids. The transducers are watertight to a depth of 1 meter/1.64 feet (rated IP 68). The base
station is protected against vertically falling water drops only (rated IP 31 according to
IEC 60529).
Explosion Hazard/Heat Exposure:
• Use the system in an environment which is reasonably free from vibration, dust, corrosive or
explosive gases, extremes of temperature, humidity, and so forth. It operates within specifications
at ambient temperatures between 0°C and +45°C/32°F and +113°F. Ambient temperatures that
exceed these limits can affect the accuracy of the system, the transmitter radio frequency
transmission, and can damage the components and circuits. The system can be stored at ambient
temperatures between -20°C and +60°C/-4°F and +140°F.
Installation
WARNING
• No modification of the fetal monitors, transducers, and the Avalon CL base station is allowed.
• Perform initial inspection of delivery, unpack, and check the shipment.
• The correct and accurate functioning of the equipment is ensured by the successful completion of
the safety tests, performance test, and the system test.
8
1 General Information
• It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling,
wall, or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for
structural integrity and compliance with all local, state, and any other required codes by a
registered, professional, structural, and/or mechanical engineer.
• Although considerable effort has been made to ensure the safety of the ceiling mount installation,
and the mounting guidelines, it is to be understood that the installation itself is beyond the control
of Philips Medical Systems. Accordingly, Philips Medical Systems will not be responsible for the
failure of any such installation.
• Check the mounting of the Avalon CL base station for integrity as part of your safety precautions.
Ensure that the base station is not able to slip, fall, or be pushed over by accident.
• The device should not be used adjacent to, or stacked with, other equipment unless otherwise
specified.
• NEVER run power cables through the same conduit or trunking used for system cables.
9
1 General Information
• If a battery has been dropped or banged against a hard surface, whether damage is visible
externally or not:
– discontinue use
– dispose of the battery in accordance with the disposal instructions above.
• Keep batteries out of the reach of children.
• Do not disassemble, heat above 100°C (212°F), or incinerate the batteries, to avoid the risk of fire
and burns. Keep batteries in their original package until you are ready to use them.
• Do not install or use pre-damaged batteries.
Care and Disinfection:
• To avoid contaminating or infecting personnel, the environment or other equipment, make sure
you disinfect and decontaminate the monitor appropriately before repairing or disposing of it in
accordance with your country's laws for equipment containing electrical and electronic parts.
• For disposal of parts and accessories such as thermometers, where not otherwise specified, follow
local regulations regarding disposal of hospital waste.
Walkthrough
The walkthrough provides an overview and a guideline for the service tasks from the site planing
through the troubleshooting.
10
1 General Information
11
1 General Information
12
2
2 Site Preparation
Introduction
This section describes the procedures you should follow to plan and prepare a site for an Avalon CL
Transducer System installation.
• Roles and responsibilities for local and Philips personnel
• Site requirements
• Site planning
13
2 Site Preparation
Preparing Mounts
Where ceiling, wall, or shelf mounts are required for mounting the equipment, the customer is
responsible for the following:
– Providing and installing all hardware which is required to install the mounting hardware
supplied by Philips as detailed in the Installation Notes.
– Making sure that all ceilings, walls, and mounting rails that supports mounting hardware are
suitable for their proposed load.
WARNING
It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling, wall,
or mounting rail, and the construction of the ceiling, wall, or mounting rail evaluated for structural
integrity and compliance with all local, state, and any other required codes by a registered, professional,
structural, and/or mechanical engineer.
Although considerable effort has been made to ensure the safety of the ceiling mount installation and
or mounting guidelines, it is to be understood that the installation itself is beyond the control of Philips
Medical Systems. Accordingly, Philips Medical Systems will not be responsible for the failure of any
such installation.
• Providing Conduit
– Providing conduit and/or trunking of a sufficient cross-sectional area for the planned cables
and possible future expansion (for additional components or systems).
– Providing and/or installing suitable wall boxes to accommodate the face plates.
• Pulling Cables
WARNING
NEVER run power cables through the same conduit or trunking used for system cables.
14
2 Site Preparation
Site Requirements
The site requirements are listed in this section.
Space Requirements
The situating of the Avalon CL base station should be planned such that the nursing staff are able to
monitor the patient with relative ease, with all patient connectors and controls readily available and the
displays clearly visible. The location should also allow access to service personnel without excessive
disruption and should have sufficient clearance all round to allow air circulation.
Dimensions and weight:
CL Base Station:
Size (W x H x D): 349 x 74 x 183 mm (13.7 x 2.9 x 7.2 in)
Weight; 0.985 kg (2.2 lb)
CL Transducer:
Size (diameter/height): 75.5 mm/ 36.6 mm (3.27 in)
Weight: 123 g (4.3 oz)
Environmental Requirements
The environment where the Avalon CL base station will be used should be reasonably free from
vibration, dust, and corrosive or explosive gases. The ambient operating and storage conditions for the
Avalon CL base station must be observed. If these conditions are not met, the accuracy of the system
will be affected and damage can occur.
15
2 Site Preparation
Safety Requirements
The base station is an electrical Class II device in connection with M2702A/M2703A (FM20/FM30)
and M2704A/M2705A (FM40/FM50) in which the protection against electric shock does not rely on
basic insulation and a protective earth conductor, but on double and/or reinforced insulation.
Electrical Requirements
Line Voltage Connection
The Avalon CL base station is powered by the connected fetal monitor.
Essential Performance
Refer to the fetal monitor Service Guide for details on the essential performance for Avalon CL base
station in combination with the Avalon fetal monitors. The essential performance is maintained under
the electromagnetic conditions specified in chapter “Electromagnetic Compatibility (EMC)” on
page 121. The essential performance is specified separately for non-transient and for transient
electromagnetic interference.
Non-Transient Electromagnetic Phenomena:
• Radiated electromagnetic fields
• Conducted disturbances induced by RF fields
• Conducted disturbances induced by magnetic fields
• Voltage dips/voltage variations
Transient Electromagnetic Phenomena:
• Electrostatic Discharge (ESD)
• Electrical Fast Transients/Bursts
• Surges
• Voltage interruptions
• Electrosurgery (ESU)
Refer to the Service Guides of the FM20/30 and FM40/50 for further details.
16
2 Site Preparation
Site Planning
The careful planning of the site for the Avalon CL Transducer System is essential for its safe and
efficient operation. A consulting schedule should be established between the customer and Philips
sales and support representatives, to ensure that all preparations are completed when the system is
delivered.
The site planning phases prior to equipment installation are:
Location: Planning the location of the system components.
Environment: Confirming and correcting, as necessary, the environment of the proposed installation
site(s).
Radio Frequency and area of reach plan: Consult the local radio frequency plan if available or
create one.
System Capabilities: Explaining the possibilities for system expansion.
Mounting: Referencing the mounting hardware information website for the listing of suitable
mounting hardware recommended for use with the various system components, and all details on the
available mounts and accessories.
Installation Scope
Depending on how many Avalon CL Transducer Systems were purchased, the site planning has to be
differentiated in relation to selecting in which room the systems are placed, and which fixed
frequencies are assigned to them. If the systems are installed in adjoining exam rooms the fixed
frequency channels have to be alternated, or an assigned CL transducer will report an INOP e.g. cl US
Disconnect.
In the frequency plan also include fetal monitors to which the Avalon CL base station may only be
temporarily connected (e.g. moving the base station with the patient to a different room). The
configuration of each fetal monitor determines the frequency settings of the connected Avalon CL
Transducer System. If the OBR settings have not been configured during the installation of a fetal
monitor, an Avalon CL Transducer System cannot be operated with it.
The physical structure of a building has an influence on the available range of the OBR radio. The
following structures can reduce the signal strength:
• steel enforced concrete walls
• metal doors
• elevators
17
2 Site Preparation
Mounting Options
The Avalon CL Transducer System can be mounted:
• on top of a cart or a desk
• in a shelf or on top of an Avalon FM40/FM50 fetal monitor
• a wall channel
• the post of a rail stand
• a mounting plate
....
CAUTION
Check the mounting of the Avalon CL transducer System for integrity as part of your safety
precautions. Ensure that the base station is not able to slip, fall, or be pushed over by accident.
NOTE
Do not mount the Avalon CL Transducer System in an enclosed metal rack, cabinet, or carts drawer.
This will cause a marked decrease in the operating range.
18
2 Site Preparation
Should there not be a frequency plan available for the site, use the Excel sheet Channel_Crosscheck
located on the Fetal Documentation DVD in the same folder as this Service Guide. The following
tutorial will use this Excel sheet to configure the OBR channels of various OBR devices.
19
2 Site Preparation
Example 1
Situation
An OB department has three Avalon CTS and two M1310A. They bought three additional Avalon CL
systems. The Avalon CL systems are marked blue, the Avalon CTS systems green, and the M1310A
systems red.
Frequency Planning
In this example it is assumed that the fixed frequencies of the two M1310A are 610.15 MHz and
608.5 MHz.
1 Open the Excel sheet and write down the frequencies below the green frequency block.
2 Search for 610.15 MHz in the frequency block. You will find 610.1375 MHz and 610.1625 MHz
but not exactly 610.15 MHz. The fixed frequencies of the M1310A are not configurable. Since
their range in this example lies between two listed frequencies, it eliminates both the higher, and
the lower frequency from a possible selection. In this case, both frequencies must not be used by
any other equipment. Mark them with a red font color.
3 Now repeat the procedure with the second frequency 608.5 MHz. This leads to the exclusion of
608.4875 MHz and 608.5125 MHz.
4 Now configure the fixed frequencies for the Avalon CTS. Please refer to Avalon CTS Service
Guide on the Fetal Documentation DVD. One possible approach is, to place the Avalon CTS in
the higher part of the frequency band. Configure the three Avalon CTS as follows:
– 1st: 612.8125 MHz
– 2nd: 612.8875 MHz
– 3rd: 612.9625 MHz
20
2 Site Preparation
5 Mark them also in a red font. Keep in mind, that each base station of the Avalon CTS uses three
frequencies of this block. Mark the consecutive used frequencies also in red font.
6 In the last step, you can choose the channels for the Avalon CL. Just check the appropriate box in
the channel block and verify that the channels are not already occupied. If the channels are already
occupied (you have marked them by setting the font color to red), you will see that those
frequencies are blocked out with a red bar.
7 In this example, channels 4, 6, and 8 were chosen. You can see the resulting spreadsheet below.
The frequencies now reserved for the Avalon CL systems are highlighted in red.
21
2 Site Preparation
Example 2
Situation
An OB department has 12 rooms. They do not have any cableless systems so far, and want to install 12
new Avalon CL Transducer Systems. Part of the frequency band is occupied by the neighboring
department. The floor plan could look like this:
Frequency Planning
The frequencies in use are already determined and marked red in the spreadsheet.
The remaining channels are assigned to meet the two criteria as good as possible:
• Use at least alternate channels (3 and 5) for two adjoining rooms.
• Keep away from other RF equipment as far as possible.
Of course several combinations are possible. The resulting spreadsheet could look like this:
22
2 Site Preparation
23
2 Site Preparation
24
3
3 Installation
Installation should be carried out by authorized and qualified service personnel, by Philips directly, or a
Philips certified service partner/dealer.
As the first step in preparing the Avalon CL Transducer System for use, follow the installation
instructions given in this chapter.
The Avalon CL base station is a system that can only be operated fully functional with the fetal
monitors FM20/30 and FM40/50.
The configuration of the fixed frequency channels should be carried out by authorized and qualified
service personnel, either from the hospital’s biomedical department, or from Philips service personnel.
25
3 Installation
Unpacking/Checking Shipment
The cableless transducer system and any supporting options ordered are delivered packed in protective
shipping cartons.
Initial Inspection
1 Before unpacking, visually check the packaging and ensure that there are no signs of mishandling
or damage.
2 Open the package carefully and remove the instrument and accessories.
3 Check that the contents are complete, and that the correct options and accessories have been
delivered.
System Components, Accessories and Supplies Quantity Comments
Avalon CL Base Station 1 -
CL Ultrasound Transducer 1/2/3 -
CL Toco+ MP Transducer 1 -
CL ECG/IUP Transducer 1/2/3 -
Connection with red connector (connected to the fetal 1 -
sockets at the fetal monitors)
Connection cable with black connector (connected to the 1 Only for use with
telemetry socket at the rear of the FM20/50) FM40/50
Application Note 1 -
Repacking
Retain the original packing carton and material, in case you need to return equipment to Philips for
service. If you no longer have the original packing materials, Philips can advise you on alternatives.
26
3 Installation
Connection Options
The fetal monitors FM20/FM30 and FM40/FM50 are compatible with the Avalon CL and
Avalon CTS Transducer Systems. Regard the following points for cableless monitoring:
• You can connect one Avalon CL base station with a red connector to an FM20/FM30 or an
FM40/FM50 (fetal socket), or one Avalon CL base station with a black connector to an FM40/
FM50 (telemetry socket).
• You can connect two Avalon CL base stations with black connectors to an FM40/FM50
(telemetry sockets)
27
3 Installation
• You cannot connect two Avalon CL base stations to an FM40/FM50, if one Avalon CL base
station has a red connector, and the other Avalon CL base station has a black connector.
• You can connect one Avalon CTS system to an FM20/FM30 or FM40/FM50 at a time (either
fetal or telemetry socket).
28
3 Installation
• You cannot connect an Avalon CTS and an Avalon CL at the same time to the same fetal monitor.
• Monitoring a multiple pregnancy using cableless transducers is supported by the Avalon CL system
only.
• Using a mixture of wired and cableless fetal transducers is not supported. You can use either wired
or cableless fetal transducers.
• If you cannot get sufficient signal quality using Avalon CTS transducers, switch to wired
transducers.
29
3 Installation
The following rules apply with the connection of two base stations to the monitor:
1 The Avalon CL base station connected to the left telemetry socket takes over the radio
communication with the fetal monitor, if both base stations are connected when the FM40/FM50
is switched on.
2 If another base station is added while the first one is in use, the base station in use keeps the radio
communication with the fetal monitor (no matter which telemetry socket it is connected to), and a
message is issued Too many cableless transducer systems connected. When the FM40/50 is
switched off and on again, the base station connected to the left telemetry socket will resume the
radio communication with the fetal monitor again.
3 Both base stations can charge and assign CL transducers equally. You can switch cableless
monitoring to On or Standby toggle button of either base station.
4 When connecting two base stations to the FM40/50, mount the base station connected to the left
telemetry socket optimized for good radio transmission. The second base station can be placed,
for example, in a drawer since it does not have radio communication.
5 When updating the firmware of two connected base stations, update one base station after the
other, and always connect the one to be updated to the left telemetry socket. Also after both
updates have been done, and both base stations are again connected as desired, you have to cycle
the power of the fetal monitor to initialize the setup again.
Telemetry Interfaces
The Avalon CL base station is delivered with:
• OB Radio (OBR)
The base station has a built-in radio with an integrated antenna to communicate with CL transducers.
Host systems (Avalon fetal/maternal monitors) can control the transducers via the base station and
OBR.
30
3 Installation
The base station is shipped with three corresponding frequency bands depending on the region it is
delivered to:
Frequency Bands:
• USA: WMTS 608 .. 614 MHz
• EU: ISM 433.05 .. 434.79 MHz
• Japan: T108 920.5 .. 923.5 MHz
The base station has a built-in radio interface with an integrated antenna to communicate with SRR
devices (IntelliVue CL NBP and CL SpO2 Pods). Host systems (Avalon fetal/maternal monitors) can
control the CL Pods via the base station and SRR.
The OBR has an operating range up to min. 100 m/300 ft (in line of sight). The SRR operating range
is limited to the parameter of the room (ca. 5 m) the base station is placed in. The actual effective area
of reach will vary according to the geographical and physical characteristics of the building where the
system is installed, and is also influenced by the presence of other radio frequency (RF) devices or
interference. We recommend that you define effective operating range prior to putting the system into
operation.
If the area of reach is adequate for the intended monitoring area, and the system’s operation is not
influenced by other RF sources, then the normal delivery configuration is sufficient.
31
3 Installation
OB Radio
You should consult the local radio frequency plan before configuring the OBR channels, or see “Radio
Frequency Plan” in this guide.
Setting Up OB-Radio
OB-Radio is set up at the fetal monitor. The base station is delivered with a fixed frequency band for
your region. Consult the local radio frequency plan for available frequency channels, or scan for an
available frequency channel and see also “OB-Radio Scan” on page 34.
1 Switch to the Configuration or Service Mode of the fetal monitor. Service Mode functions can be
used to identify channel assignment conflicts in the hospital environment.
2 Select Main Setup, Hardware, OBR. The revision shows which OBR setting revision is installed.
3 Select Frequency Band and select one of the options WMTS, ISM, or T108 depending for which
region your base station was configured. Check the base station label for it's type of frequency
band.
32
3 Installation
4 Before you select a channel consult the hospital's frequency plan, or use the OB-Radio scan to
determine the availability of the channels.
OBR provides a total of 11 channels on the ISM band. The channels are labeled 1 to 11. The ISM
band is not exclusively reserved for OBR. It is also used by, for example, SRR, Wireless LAN
(WLAN), and the OB TraceVue/IntelliSpace Perinatal Telemetry network. For this reason, depending
on the hospital’s existing wireless infrastructure, a number of OBR channels might already be occupied
by other wireless applications.
33
3 Installation
OBR provides a total of 14 channels on the T108 band. The channels are labeled 24 to 37. Wherefore
a selection of e.g. channel 24 uses both the channels 24 and 25, and channel 26 also uses 26 and 27,
and so on. The used channels are not available for other wireless devices.
OB-Radio Scan
With an OB-Radio scan you can scan the radio activity from other radio devices on available channels
in the direct vicinity of the fetal monitor for a set duration.
1 Switch to the Service Mode.
2 Select Main Setup, Hardware, OBR.
3 For the Scan Duration select either Single Scan, or a duration from 15 min—24 h.
4 Select Start Scan and the OB-Radio scan is started.
The OB-Radio scan runs until the set duration is reached, and then the recorder prints out an OB-
Radio scan result similar to the example report below.
34
3 Installation
5 You can stop the scan anytime manually by selecting Stop Scan.
The values show how long1 a channel was occupied through the length of the duration, and how
high the traffic of radio devices was in that duration. The best choice for OB-Radio are channels
with low mean and max. values.
To calculate the ideal channel values, use the following formula:
35
3 Installation
Avalon CL
Channel 7: (46/104/167)
Low min., high max., and mean Shows that an Avalon CL system was sending for about 50% of
values, typical for a radio the time within the scan duration.
signature like Avalon CL: not
recommended for channel
selection
Avalon CTS
Channel 30: (117/138/154)
High min., max., and mean Shows an Avalon CTS with a high duty cycle sending
values, typical for a radio continuously.
signature like Avalon CTS:
not recommended for channel
selection
Any RF device
any Channel: (40/50/120)
Low min., high max., and low Shows a short interference of a device using this OBR band
mean value, typical for a short infrequently. Making this channel available for channel selection.
time interferer:
recommended for channel
selection
Indication of a free channel (20/
30/40).
Low min, max., and mean value, Use the best available channel for the OB-Radio channel setting.
typical for a channel which is
not in use at all: highly
recommended for channel
selection
1 The closer the mean value is to the max. value, the longer the channel is occupied.
36
3 Installation
RSSI Trace
With the RSSI Trace you can test the signal strength and range of the CL transducers.
1 Switch to Service Mode
2 Select Main Setup, Hardware, OBR, and Start RSSI Trace.
3 Undock a CL transducer and walk away from the monitor.
The RSSI trace will scan the signal from the CL transducer as you are walking and will print it out
at the recorder. Depending on where the signal trace is printed in the FHR section of the trace, you
can discern the strength and quality of the OB-Radio signal. The RSSI value corresponds to the
FHR scale.
4 You can stop the scan anytime by selecting Stop RSSI Trace
The following RSSI example traces were run on the channels 5 and 6 from the previous OBR scan
example.
Channel 6 has a mean RSSI of 82 and a Channel 5 has a mean RSSI of 72 and a max
max. RSSI of 123. Using the OBR formula, RSSI of 106. Using the OBR formula, the
the result is an RSSI of 102.5 which is >100 result is an RSSI of 89 which is <100 and
and therefore should not be used. therefore very good. The appropriate RSSI
The appropriate RSSI trace also shows a trace shows a stable RF link with single
stable RF link, with a few package losses. package losses (approaching out-of-range).
37
3 Installation
NOTE
Package losses: An Avalon CL base station queries the CL transducers in regular intervals for data. A
package loss occurs when the CL transducer does not answer the query from the Avalon CL base
station. A single loss of data is considered a package loss. It is indicated on the trace as a black line in
the FMP section of the trace. The more package losses occur, the less data reaches the base station and
the monitor. Most package losses are inconsequential, unless the connection is lost completely (out-of
range, or malfunction).
38
3 Installation
For a successful SRR deployment, the SRR channels must be located in the RF spectra where they are
least likely to be interfered with. Choosing appropriate channels after reviewing the Spectrum Analyzer
data is critical. In hospitals, 802.11 systems are most the likely source of interference with SRR
channels. The figures above show the relationship between 802.11, IIT, and DECT devices. For
example, if the site uses European 802.11 channels 1, 7, 13 for WLAN and has no IIT or DECT
devices in the SRR channels 15, 16, 21 or 22, these channels can be used for SRR.
39
3 Installation
NOTE
Short range radio signals are low power signals and therefore have a relatively short range. You can use
this fact if the number of unused channels is low, and you run out of channels. Provided the distance
between two SRR groups is large enough, i.e. none of the short range radio signals transmitted by the
one group can interfere with signals of the other group, you may attempt to assign the same SRR
channel to both groups. Take into consideration that portable components belonging to one group
may be temporarily used within the range of another group.
The range of SRR signals cannot be clearly defined as it depends on external factors such as the
components and structure of walls, ceilings, etc., but it is unobstructed ca.5 m.
Setting Expectations
No matter how good a telemetry system design is, it will always experience occasional loss of radio
communications, resulting in CL US/Toco/ECG drop-outs. A telemetry system will never be as
reliable as a hard-wired monitor that transmits its signal through a wire. If occasional loss of CL US/
Toco/ECG monitoring is not acceptable for certain patients, they should be connected to a hard-
wired bedside monitor.
Following are guidelines to set proper expectations of hospital staff and to improve system
performance:
• Clinicians will tend to see more motion related artifact on the US/Toco/ECG measurements of
ambulatory patients than on patients that are restricted to a bed.
• Patients should be restricted to the designated coverage area. Monitoring performance will
degrade if patients go outside the radius of coverage of the receiving antenna.
• Telemetry system performance will degrade as the system size increases. The larger a system is, the
greater the potential for receiving interfering signals.
• A patient location protocol is critical to a telemetry system. If a life-threatening event occurs, the
clinician must be able to locate the patient quickly.
• Philips Medical Systems has no control over the RF environment in the hospital. If interference
exists at the operating frequencies, telemetry system performance will be affected. Careful
selection of frequencies for all wireless devices used within a facility (telemetry transmitters,
walkie-talkies, ambulance radios, other wireless medical devices, etc.) is important to prevent
interference between them. Frequency management is the responsibility of the hospital.
40
3 Installation
Safety Tests
CAUTION
The correct and accurate functioning of the equipment is ensured by the successful completion of the
safety tests, performance test, and the system test.
41
3 Installation
42
4
4 Theory of Operation
This chapter describes the functional operation of the system, including the CL base station and CL
transducers. It incorporates features of the mechanical design, indicating the physical relationship of
the assemblies and components.
43
4 Theory of Operation
Antenna PCA
The base station has a built in antenna board for ISM, WMTS, and T108 band.
44
4 Theory of Operation
SRR PCA
The Short-Range-Radio (SRR) interface provides wireless connectivity with an IEEE 802.15.4 radio to
SRR enabled host devices. It allows wireless communication of the base station to the monitor with an
IntelliVue Instrument Telemetry Transceiver (e.g. cableless device, CL SpO2, CL NBP).
Switchboard PCA
The switchboard PCA contains the on/stand-by switch with which the operating modes stand-by and
on of the base station are controlled.
45
4 Theory of Operation
46
5
5 Disassembly/Reassembly
WARNING
• Performance verification: do not place the system into operation after repair or maintenance has
been performed, until all performance tests and safety tests listed in “Testing and Maintenance”
on page 79 of this Service Guide have been performed. Failure to perform all tests could result in
erroneous parameter readings, or patient/operator injury.
• Energized circuits are accessible with the covers open. Do not work on the base station with the
covers open while the base station is connected to a monitor. Only authorized and qualified
service personnel should open or disassemble the base station.
• Before attempting to open or disassemble the base station, disconnect it from the fetal monitor.
CAUTION
Observe ESD (electrostatic discharge) precautions when working within the unit.
WARNING
To avoid contaminating or infecting personnel, the environment or other equipment, make sure you
disinfect and decontaminate the monitor appropriately before disposing of it in accordance with your
country's laws for equipment containing electrical and electronic parts. For disposal of parts and
accessories such as thermometers, where not otherwise specified, follow local regulations regarding
disposal of hospital waste.
Introduction
• Remember to store all screws and parts in a safe place for later refitting.
• The disassembly sections detail the step-by-step procedures you use to access replaceable parts of
the base station.
• All the CL transducers are sealed at the factory and are not repairable. Replacement CL
transducers are supplied complete with both battery and belt button fitted.
– Exception: The battery of the CL transducers can be replaced.
47
5 Disassembly/Reassembly
Tools Required
CAUTION
When replacing the front cover, do not over-torque the screws. Excessive torque may damage the
plastic screw mountings.
Long-nosed pliers
48
5 Disassembly/Reassembly
49
5 Disassembly/Reassembly
3 Gently remove the cable by pulling the adapter from its interlock (1), pulling too hard will damage
the connector. Disconnect the cable from the connector (2) located on the main PCA. The
procedure is same for all connector cables (red connector, black connector, and external power
supply).
5 After removing the screws, turn the base station back around, and lift the top housing from the
bottom housing by gently prying them apart.
50
5 Disassembly/Reassembly
1 CL Pod connector
2 Holding frame
3 Short Range Radio
4 Switchboard/Standby
button
5 Switch cable
6 PCA board
7 Charging contacts
8 Antenna holder
9 Antenna board
51
5 Disassembly/Reassembly
52
5 Disassembly/Reassembly
2 Lift up the top holding frame (1), and pick up the SRR board (2) releasing it from its interlock.
53
5 Disassembly/Reassembly
54
5 Disassembly/Reassembly
55
5 Disassembly/Reassembly
56
5 Disassembly/Reassembly
3 With the tool unplug the antenna cable from the antenna board.
4 Place the antenna removal tool into the designated interlock in the bottom housing with the fork
part facing up and bracing against the side wall from the bottom housing.
57
5 Disassembly/Reassembly
5 Remove the antenna board from the antenna holder by extracting it from the bottom interlock
first.
58
5 Disassembly/Reassembly
8 Unwind the cable of the connector from its mounting at the antenna holder.
9 Remove the charging contact from its interlock by pressing the springs on each side of it. Assure
that the silicon sealing stays attached to the contact.
59
5 Disassembly/Reassembly
60
5 Disassembly/Reassembly
5 Assure that all charging connector cables are lined up in the cable guides of the antenna holder.
61
5 Disassembly/Reassembly
7 Replace the antenna board back into its interlock at the antenna holder.
62
5 Disassembly/Reassembly
2 Press with the screw driver on the release catch and gently slide the switchboard from its
mounting.
3 Remove the right SRR board as described in the topic “Removing the Short Range Radio Boards”
on page 52.
4 Remove the three screws (1) and slide the holding frame from its position out the interlock and lift
it up (2).
63
5 Disassembly/Reassembly
NOTE
The Standby button and the light pipe are part of the support kit and are part of one silicon item.
Separate the parts using scissors, and extract the part you need.
64
5 Disassembly/Reassembly
1 Slide the holding frame back in its interlock and fasten the three screws.
2 Reinsert the SRR board as described in “Replacing the Short Radio Boards” on page 53.
3 Slide/replace the switchboard in the interlock in the holding frame and reconnect the switchboard
cable.
65
5 Disassembly/Reassembly
2 To remove the name plate gently press the arrestors of the name plate with a small flat nose pliers
together as you pull it out its interlock.
66
5 Disassembly/Reassembly
3 Insert the name plate into the designated interlock of the new bottom housing.
4 Take the new bottom housing and reassemble the base station as described in “Replacing Top
Cover Assembly CL Base Station” on page 51.
67
5 Disassembly/Reassembly
2 Place the top cover from the tool on the bottom part, and fit the interlock on the sides together.
68
5 Disassembly/Reassembly
3 Turn the top of the tool counter-clock-wise and back to the start position to open the CL
transducer housing within.
4 Take the opened CL transducer out of the tool and gently lift the top housing of the CL transducer
up.
69
5 Disassembly/Reassembly
5 Unplug the charger cable and lift the battery and attached CPU board from the CL transducer
housing.
6 Unplug the battery connector (1), and pry the CPU board gently from the holding frame of the
battery (2). The CPU board and the battery are separated (3).
70
5 Disassembly/Reassembly
3 Place the CPU board on the holding frame (1) so that the clamps hold the board in place. Turn the
holding frame over and plug-in the battery connector (2) at the CPU board.
4 Place the battery with attached CPU board back into the CL transducer housing. Make sure you
place the nose of the CPU board exactly on the shelf, and not against the shelf within the CL
transducer head. This ensures that the connectors are lined up correctly (1).
71
5 Disassembly/Reassembly
5 Push the CPU board lightly down on the left side and center it by wiggling it back and forth (2).
6 Push the CPU carefully down until you hear the connectors in the transducer shell arrest with a
clicking sound.
7 Wind the charger cable back on its mounting as shown in the graphic below:
– lay it over the first nose (1)
– under the second nose to the back (2)
– under the hook (3)
– and plug the connector into the CPU board (4)
72
5 Disassembly/Reassembly
9 Replace it with one of the two sealing rings shipped together with the battery kit.
73
5 Disassembly/Reassembly
10 Take the flat screw driver and move it one time under seal around the whole CL transducer cover
to ensure that the seal lays correctly in its groove, and is not twisted. This ensures the
watertightness of the CL transducer.
CAUTION
Follow these instructions closely, or a part of the CPU board can be
damaged in the process.
74
5 Disassembly/Reassembly
14 Set the still open CL transducer into the bottom of the transducer tool.
15 Place the top cover of the transducer tool with the symbol depicting a closed lock facing upward
onto the bottom of the tool.
75
5 Disassembly/Reassembly
16 Turn the top of the tool clock-wise and back to the start position to close the CL transducer
housing within.
17 The CL transducer is closed, and you can take it out of the tool, and dock it back at the base
station.
76
6
6 Spare Parts
Here all spare parts are listed with part numbers. All transducers and accessories are latex-free, unless
indicated otherwise.
77
6 Spare Parts
78
7
CAUTION
Failure on the part of the responsible individual hospital or institution using this equipment to
implement a satisfactory maintenance schedule may cause undue equipment failure and possible health
hazards.
79
7 Testing and Maintenance
Recommended Frequency
Perform the procedures as indicated in the suggested testing timetable. These timetable
recommendations do not supersede local requirements.
Suggested Testing Time Table:
Test Frequency
Preventive maintenance Battery check Once every two years, or more
often if this is specified by local
laws.
Other regular test Visual inspection Before each use
Power on test
Performance assurance test Transducer functional test • Switch board/ silicon
Battery check button was replaced
• Transducer charging
contacts replaced
• Cableless connection board
replaced
• SRR board replaced
Safety test Visual inspection After each service event
Electrical (base station only) Electrical safety testing is not
required for the Avalon CL
base station. The base station
has the same status as a wired
transducer. The base station is
either connected to the red fetal
sensor socket (FM20/30/40/
50), or the black rear telemetry
socket (FM40/50 only). When
used as a charger station only,
the device is not intended to be
a medical device.
80
7 Testing and Maintenance
81
7 Testing and Maintenance
Testing Sequence
Summary of the recommended sequence of testing:
NOTE
If any single test fails, testing must be discontinued immediately, and the device under test must be
repaired or labeled as defective.
Visual Inspection
Before Each Use
Check all exterior housings for cracks and damage. Check the condition of the interface cable,
especially for splits or cracks, and signs of twisting. If serious damage is evident, the cable should be
replaced immediately. On the CL Toco+ MP and CL ECG/IUP transducer ensure that the adapter
cable socket is not damaged. Check that all mountings are correctly installed and secure. Refer to the
instructions that accompany the relevant mounting solution.
82
7 Testing and Maintenance
Power On Test
1 Connect the base station to a fetal monitor and switch on the base station.
2 Press the On button of the base station and check if the LED lights up.
3 Open at the fetal monitor screen the Tele Info window and check if the monitor registers the
connected base station with the docked transducers.
83
7 Testing and Maintenance
Safety Tests
Electrical safety testing is not required for the Avalon CL base station. The base station has the same
status as a connected wired transducer. The base station is either connected to the red fetal sensor
socket (FM20/FM30, FM40/FM50), or the black rear telemetry socket (FM40/FM50 only).
CAUTION
The correct and accurate functioning of the equipment is ensured by the successful completion of the
safety tests, performance test, and the system test.
Please refer to the Service Guides of the FM20/FM30 and FM40/FM50 for the detailed description of
required safety tests for the fetal monitors.
System Test
The Avalon CL base station does not function independently, and can only be operated if it is
connected to a fetal monitor. Therefore all system tests are done with the connected fetal monitor.
Please refer to the Service Guides of the FM20/30 and the FM40/50 for a detailed description of the
required system tests.
Battery Check
The CL transducers use a Lithium Ion battery (part no. 453564107871).
84
7 Testing and Maintenance
The cableless transducers have a multi-color LED that indicates the status of the transducer with
specific colors. This LED remains visible when the transducer is correctly attached to the transducer
belt (Philips standard belt).
1 LED
85
7 Testing and Maintenance
Active cableless equipment information is displayed at the begin of the list inside the window.
Displayed information per active device is:
– Device symbol
– Assigned parameter label(s) (OBR equipment) or Equipment ID (SRR equipment)
– Radio link quality indicator
– Battery status symbol
– Remaining operating time (HH:MM)
Battery Cycle
Battery replacement is recommended after 500 charge/discharge cycles *2), or if the battery is older
than 4 years whatever is reached first. If the battery of a CL transducer has aged and an exchange of
the battery is highly recommended, a prompt message is displayed at the fetal monitor for ca.
60 seconds. The prompt is repeated whenever the Tele Info window is opened until the battery is
replaced.
Depending on the transducer type the following messages are displayed:
cl US battery has aged. Replacement strongly recommended
cl Toco battery has aged. Replacement strongly recommended
cl ECG/IUP batt has aged. Replacement strongly recommended
The date of manufacture and number of charge/discharge cycles can be inquired on the connected
fetal monitor.
For the battery cycle and battery life specification of the CL Pods (NBP and SpO2) please refer to the
Service Guide of the Cableless Measurements.
Charge Cycle
The CL transducer batteries are designed for frequent recharging. A complete charging cycle is only
reached and counted, when all recharging periods equal a 100% charge (900 mAh equal 8 hours
continued operation).
86
7 Testing and Maintenance
Battery Report
You can generate a battery report for all docked CL devices at the Avalon CL base station. The battery
report lists:
For the fetal monitor:
• Product number
• Serial number
• Software revision
• OBR band and current channel configured
For the Avalon CL base station
• Firmware version
• Serial number
• OBR band
For the Avalon CL transducers:
• Serial number
• Firmware version
• Manufacture date
• Capacity (mAh)
• Charge cycles
1 To generate a battery report switch to the Service Mode.
2 Open the Tele Info window either by selecting Main Setup, Tele Info, or the Tele Info SmartKey
(configurable), or selecting the
87
7 Testing and Maintenance
Battery Exchange
If the CL transducer battery requires replacement, use the transducer battery kit. This contains:
• a replacement battery
• a special tool for removing the battery drawer
• an instruction sheet
For details on how to replace the battery, see the Instruction Sheet “Removing and Replacing the
Transducer Battery” that accompanies the battery replacement kit, or refer to “Disassembly CL
Transducer Battery” on page 67 in this guide.
For details how to exchange the battery of a CL Pod (NBP and SpO2) refer to the Service Guide of the
Cableless Measurements.
Battery Storage
Battery recharging period within Philips’ factories and warehouses
The CL transducer and battery replacement kits require recharge within:
• 12 months for the transducer (with battery integrated), starting with the date of manufacture of
the battery, as shown in the battery report on the fetal monitor or on the battery label (yww).
• 18 months for a replacement battery, starting with the date of manufacture shown after the LOT
number on the battery replacement kit label (yyyy-ww) or on the replacement battery label (yww).
Battery recharging requirements during storage outside of Philips
Stored batteries and stored CL transducers should be partially charged to 40%—50% of their capacity
every 6 months.
Battery Disposal
Batteries should be disposed of in an environmentally-responsible manner. Consult the hospital
administrator or your local Philips representative for local arrangements.
Discharge the batteries and insulate the terminals with tape before disposal. Dispose of used batteries
promptly and in accordance with local recycling regulations.
Testing Alarms
Details of how to test alarms is given in the Instructions for Use.
88
7 Testing and Maintenance
CAUTION
Use of ultrasound gel that is not approved by Philips may reduce signal quality and may damage the CL
transducer. This type of damage is not covered by warranty.
89
7 Testing and Maintenance
7 The CL ultrasound transducer contains seven piezoelectric crystals. Basic functioning of each can
be verified by holding a flat bottomed pen or similar above each crystal and moving it up and
down as shown. A sound should be heard for each crystal tested. The pen should be held two to
three centimeters from the CL transducer surface when the test is carried out.
8 A sound should also be heard when the CL transducer is moved back and forth over a solid
surface, or the hand as shown.
90
7 Testing and Maintenance
CAUTION
The tip of the spiral electrode is sharp. Take care not to injure your fingers.
91
7 Testing and Maintenance
Or cut off the plastic tip of the fetal scalp electrode (containing the spiral and reference electrodes)
from the end of the wires. Strip the insulation from the end of the wires, and connect them to a
patient simulator.
NOTE
We do not recommend the use of a specific patient simulator. The use of a patient simulator does
not allow checking the specification of the ECG-functionality; it allows only a check of the general
function.
1 Result: the DFHR x Leads Off INOP should disappear.
2 Viewing the ECG wave: when configured, you can view the DECG wave on the screen, and any
noise will be visible as additional verification of the effectiveness of the test.
If the test results are not as outlined above, repeat the test with another ECG transducer. If this
does not solve the problem, try the following:
3 Check all connections.
4 If the DFHR x Leads Off INOP is still displayed, the DECG adapter cable may be defective. Replace
the adapter cable.
5 If the problem persists, replace the CL transducer.
NOTE
We do not recommend the use of a specific patient simulator. The use of a patient simulator does
not allow checking the specification of the ECG-functionality; it allows only a check of the general
function.
92
7 Testing and Maintenance
6 Result: You should see MECG values displayed on the maternal display, or annotated on the
recorder trace.
If the test results are not as outlined above, repeat the test with another ECG/IUP transducer. If
this does not solve the problem:
7 The MECG adapter cable may be defective. Replace the adapter cable, and repeat the test.
8 Check all connections.
93
7 Testing and Maintenance
94
8
8 Troubleshooting
A guide for recognizing system error messages and for troubleshooting problems you may encounter
while using the system is given in the Instructions for Use. This chapter provides a guide for qualified
service personnel for troubleshooting problems that cannot be resolved by the user.
CAUTION
If the troubleshooting procedure requires you to disassemble the base station, be certain to follow the
disassembly and reassembly procedures given in “Disassembly/Reassembly” on page 47. Whenever
parts are replaced in the system, be certain to verify the hardware and software compatibility of the
repaired system.
Overview
Troubleshooting a cableless system presents some unique challenges. These arise from several
assemblies acquiring signals, processing signals, and transmitting/receiving signals. Problems can result
from hardware and communications failures, or can be RF-related. In the case of a suspected
malfunction it is important to first verify that the malfunction is a genuine failure. This can be checked
by:
• Resetting the system
• Checking/cleaning the charge/communication contacts
• Checking the connected fetal monitor
95
8 Troubleshooting
System Reset
Always reset the system as the first troubleshooting step. With the transducers docked in the
appropriate slots, disconnect the interface cable from the fetal monitor. Wait until the LED is of the
base station is off, then reconnect the system to the fetal monitor. A persistent problem indicates a
genuine malfunction.
If you think that the display information is suspicious, initiate the power-on self test by switching the
base station to standby, and then switching it on again. All display elements of the CL transducer and
base station should be displayed briefly before the normal registration process starts.
No Tele symbol The base station is either not properly connected to the fetal monitor, or the
displayed fetal monitor cannot recognize it.
Check the connections, and ensure the fetal monitor has the required software
version for the base station installed.
The fetal monitor recognizes the connected base station, but the base station is
still in Stand-by mode.
Switch the base station on.
The base station and the fetal monitor are ready for monitoring.
A base station is connected to the monitor but no CL transducers and no CL
Pods are currently active. All are still docked at the base station, or none are
assigned.
A base station is connected to the monitor, it is on, and minimum one CL
transducer or CL Pod are active and assigned.
The CL transducer is moved away from the base station approaching out-of-
range. The US sound is replaced by an artificial QRS sound (like DECG). The
DECG and MECG waves are no longer displayed.
96
8 Troubleshooting
Checking Contacts
If the CL transducer is not assigned to the base station, or you suspect that the CL transducer is not
behaving normally while it is docked, check that the spring-loaded charge/communication contacts are
clean and free of dirt and ultrasound gel.
Check that the springs function correctly by applying pressure with a finger on the contacts. The
contacts should offer some resistance, deflect and return to their normal position; if they do not, then
they have to be replaced, “Replacing Charging Contacts” on page 59.
97
8 Troubleshooting
cl ECG/IUP BattLow
cl US Batt Empty Battery Transducer The remaining monitoring time with this CL
!!cl US Batt Empty
transducer is below 15 minutes.
!!!cl US BattEmpty
Charge the battery.
!!clToco BattEmpty
!!!clTocoBattEmpty
cl ECG/IUP Empty
!!!clECG/IUP Empty
INOP tone
Check OBR Config Monitor OBR (OB Radio) channel configuration is
INOP Tone invalid or settings are not present.
Check the channel configuration and hardware
option.
Run an OBR scan with the fetal monitor.
cl US Check Temp Monitor/Transducer/Battery The temperature of the CL transducer or its
INOP tone battery is critically high. If the CL transducer is
docked on the base station, charging is
cl Toco Chk Temp interrupted.
INOP tone Check that device is not covered and do not
clECG/IUP Chk Temp expose device to heat sources.
INOP tone If INOP persists, remove device from patient.
98
8 Troubleshooting
cl Toco Malf
Check if the CL transducer is properly assigned
at the base station currently in use.
INOP tone
Check if the CL transducer is out-of-range, or if
cl ECG/IUP Malf
the signal is blocked by a structure in the
INOP tone building.
Check if the CL transducer has interference
from another base station or CL transducer in
direct proximity.
cl US Disconnect Monitor The CL transducer has lost the connection to
INOP tone the base station.
cl Toco Disconnect
INOP tone
cl ECG/IUP Disconn
INOP tone
cl NBP Batt Empty Battery CL NBP Pod The remaining battery time of the
!!cl NBP BattEmpty
CL NBP Pod is below 30 minutes.
INOP tone
cl NBP Batt Incomp Battery CL NBP Pod The battery in use with the CL NBP Pod is
INOP tone incompatible.
Replace it with one approved for use with the
CL NBP Pod.
cl NBP Batt Low Battery CL NBP Pod The remaining battery time of the
INOP tone CL NBP Pod is below 2 hours.
Charge the CL NBP Pod on the base station.
cl NBP Batt Malf Battery CL NBP Pod There is a malfunction in the CL NBP Pod's
INOP tone battery system.
Check if the CL NBP Pod is properly assigned
at the base station currently in use.
Check if the CL NBP Pod is out-of-range, or if
the signal is blocked by a structure in the room.
Check if the CL NBP Pod has interference
from another wireless device.
cl NBP Batt Temp Battery CL NBP Pod The temperature of the battery in the CL NBP
INOP tone Pod is critically high.
Check that the CL NBP Pod is not covered or
exposed to a heat source.
99
8 Troubleshooting
INOP tone
cl SpO₂ Batt Incmp Battery CL SpO2 Pod The battery in use with the CL SpO2 Pod is
INOP tone incompatible.
Replace it with one approved for use with the
CL SpO2 Pod.
cl SpO₂ Batt Low Battery CL SpO2 Pod The remaining battery time of the CL SpO2
INOP tone Pod is below 2 hours.
cl SpO₂ Batt Malf Battery CL SpO2 Pod There is a malfunction in the CL SpO2 Pod's
INOP tone battery system.
Check if the CL SpO2 Pod is properly assigned
at the base station currently in use.
Check if the CL SpO2 Pod is out-of-range, or if
the signal is blocked by a structure in the room.
Check if the CL SpO2 Pod has interference
from another wireless device.
100
8 Troubleshooting
101
8 Troubleshooting
Common Problems
Symptoms Possible Reasons Action
A CL transducer is removed from the The base station cannot establish 1 Check if there are already the
base station, and the base station a connection to the CL maximum of 4 CL transducers active
issues a descending tone sequence. transducer. and removed from the CL base
The parameters of this CL transducer station (check the Tele Info window at
are not activated. the monitor).
2 An ISM band CL transducer is being
used with a WMTS band CL base
station.
3 Open the top cover and check if the
antenna is connected correctly to the
PCA board.
4 Scan for RF interference by making
an OBR scan.
5 Check the hardware type of the CL
transducer and the CL base station
(WMTS, ISM, and T108). If they are
not the same, they cannot
communicate.
The INOP Check OBR Config is General incompatibility: For 1 Check the OBR configuration setting
issued at the monitor. example, you have a CL base at the fetal monitor.
station that is delivered for the use
of the T108 band and the fetal
monitor is configured for the ISM
band.
The OBR configuration has to be
completed, or the OBR
configuration does not match
with the CL base station type
(WMTS, ISM, or T108).
102
8 Troubleshooting
103
8 Troubleshooting
104
8 Troubleshooting
RF Problems
An RF problem can be summarized as a loss of monitoring capability due to signal corruption in the
RF link (transmitter to receiver). A lack of parameter measurements, or the inability of the transducer
to reach the registered, ready state may indicate an RF-related problem. A problem is likely to be an RF
problem if it causes any of the following conditions:
• Weak or intermittent signals, or signal loss (see “Area of Reach” on page 107).
• Interference (see also “RF Interference” on page 108).
Follow the RF troubleshooting operations in this chapter, starting at “Gathering Data” on page 106.
105
8 Troubleshooting
Gathering Data
Gathering the information to troubleshoot a potential RF problem will probably account for around
fifty percent of your job in determining if in fact you do have an RF problem. To gather data, you
should:
• Observe system performance
• Question the user
106
8 Troubleshooting
Area of Reach
The effective operating range, or the area of reach, of the system will vary depending on the
geographical and constructional characteristics of the building where the system is installed. We
recommend that you define the effective operating range prior to putting the system into operation.
Range Definition
The average range of the Avalon CL transducers is min. 100 m/300 ft (in line of sight), the CL Pods
only have a range of ca. 5 m within the parameters of a room. You can define the area of reach by
performing a walk-test wearing the CL transducers on bare skin.
The area of reach can be determined by observing how far from the base station the transducers will
still work normally allowing continuous monitoring.
There are two ways to define the area of reach:
• Watching the LED of the transducers: The transducer LED is off in normal monitoring mode. If
the transducer is out of range an INOP is triggered and displayed at the monitor. The numerics of
the transducer at the monitor are then replaced by ? (a question mark). The LED of the transducer
lights up cyan and a beep tone is emitted to notify the walking patient that she is out of range.
• Running an RSSI (received Signal Strength Indicator) trace at the monitor. The RSSI trace records
the RSSI values of all OBR devices (base station and all active CL transducers as a trace on the
recorder.
1 If you want to test the range and signal quality of the base station and its assigned CL transducers
make an RSSI trace.
2 Click on Start RSSI Trace to start recording the RSSI values of all OBR devices (base station and all
active CL transducers as trace on the recorder).
107
8 Troubleshooting
3 Click on Stop RSSI Trace to stop the recording of the RSSI values.
RF Interference
You may encounter RF-related problems other than those associated with the area of reach. Problems
concerning RF link stability are due to RF interference. There are two categories of interference:
• On-channel or narrow-band interference
• Broadband interference
Broadband Interference
A broadband interference source occupies the bandwidth of many channels causing degradation of the
carrier-to-noise ratio (CNR) on all of the telemetry channels for as long as it is present. This results in
an increase in dropouts on all of the channels. The system has quiet and noisy periods. When the noise
source is on, the system is noisy. When the noise source is off, the system performs well. To try to
eliminate the problem:
• Troubleshoot the problem as it is occurring. Use the RSSI scan to scan the RF environment for
interference). Ask users to determine if there is a pattern, when the episodes occur, and how long
they last. If there is a pattern (for example, every morning between 6 and 8 AM), try to predict
when it will happen again, and prepare to troubleshoot.
108
8 Troubleshooting
• Locate where the interference signal is entering the system. If an antenna system is being used,
remove branches from the antenna system to see when the interference quiets down. Remember
that you will not be able to monitor patients that are being received by antennas that have been
removed.
• Remember that interference sources can move, making troubleshooting a little more difficult.
Work quickly!
• Solve the problem by trying to eliminate or repair the source. In the case of an antenna system,
relocate or permanently remove the antenna where the signal is entering. Things to check for that
cause broadband RF interference are other pieces of electrical equipment, especially RF devices
using the same frequency range.
109
8 Troubleshooting
110
9
9 Specifications
Specifications for the accuracy of heart rate measurements are given in the Instructions for Use of the
fetal monitor.
Environmental Specifications
Avalon CL Base Station 866074
Temperature Range Operating 0°C—45°C (32°F—113°F)
Storage/Transportation -20°C—60°C (-4°F—140°F)
Humidity Range Operating <95% relative humidity @ 45°C (113°F)
Storage/Transportation <90% relative humidity @ 60°C (140°F)
Altitude Range Operating -500—3000 m (-1640—9840 ft.)
Storage/Transportation -500—13100 m (-1640—43000 ft.)
111
9 Specifications
Physical Specifications
Avalon CL Base Station
Dimension and Weight Width min. 349 mm -5%, — max. width 350 mm
Size mm/(in) (min. 13.7 in -5%, — max. width 13.8 in)
Depth 183 mm ± 5% (7.2 in ± 5%)
Height 74 mm ± 5% (2.9 in ± 5%)
Weight 0.985 kg (2.2 lbs)
Electrical Class When the CL base station is connected to the fetal monitors M2702A/M2703A/
M2704A/M2705A it is a Class II equipment.
Electrical Power Source External (powered by fetal monitor)
Mode of Operation Continuous operation
Interface cable connector type Connector color Red for FM20/30 left side or FM40/50 front fetal
connector socket
Black for FM40/50 rear telemetry connector
socket
Interface cable length 1.5 m (4.11 ft.)
Ingress Protection IP 31
112
9 Specifications
Interface Specification
Avalon CL Radio
Avalon CL Radio Interface Specifications
Electronic Article Surveillance (EAS) EAS tag inside the housing (58 kHz)
Short Range Radio Interface* Type Internal SRR interface
Technology IEEE 802.15.4
Frequency Band 2.4 GHz ISM (2.400—2.483 GHz)
Modulation Technique DSSS (O—QPSK)
Effective radiated power max. 0 dBm (1 mW)
Range ca. 5 m without any physical obstructions as walls
and doors
OBR (WMTS)* Frequency Band 608—614 MHz
Effective radiated power <20 mW (base station)
<1 mW (CL transducers)
Range min. 100 m/300 ft (in line of sight)
OBR (ISM)* Frequency Band 433.05—434.79 MHz
Effective radiated power <10 mW (base station)
<1 mW (CL transducers)
Range min. 100 m/300 ft (in line of sight)
OBR (T108)* Frequency Band 920.6—923.4 MHz
Effective radiated power <20 mW (base station)
<10 mW (CL transducers)
Range min. 100 m/300 ft (in line of sight)
*Wireless transmissions are encrypted for security.
Performance Specifications
Battery Specifications
Avalon CL Transducer Battery Performance
Operating time With a new and fully charged battery 8 hours
Charge time From a "low battery" indication to a <3 hours
"fully charged" indication.
Charge time with Avalon From a "low battery" indication to a >6 hours
FM20/30 Battery Option "fully charged" indication.
#E25
113
9 Specifications
114
9 Specifications
Fetal/Maternal Specifications
Ultrasound
Complies with IEC 60601-2-37:2007 / EN 60601-2-37:2008
Performance Specifications
Ultrasound
Measurement Method Ultrasound Pulse Doppler
Measurement Range US 50—240 bpm
Resolution Display 1 bpm
Printer 1/4 bpm
Jitter @ 200 bpm ≤3 bpm
Display Update Rate 1 per second
US Intensity (M2736A/AA) Average output power P = (7.4 ± 0.4) mW
Peak-negative acoustic pressure p_ = (40.4 ± 4.3) kPa
Output beam intensity (Iob) Isata = (2.38 ± 0.59) mW/cm2
(= spatial average - temporal average
intensity)
Spatial-peak temporal average intensity Ispta = (15.0 ± 3.2) mW/cm2
Effective radiating area @ -12 dB (3.11 ± 0.74) cm2
Thermal index (TI) and mechanical index (MI) are always below 1.0.
US Intensity CL (866076) Average output power P = (12.4 ± 0.4) mW
Peak-negative acoustic pressure p_ = (49.1 ± 5.2) kPa
Output beam intensity (Iob) Isata = (2.77 ± 0.56) mW/cm2
(= spatial average - temporal average
intensity)
Spatial-peak temporal average intensity Ispta = (21.1 ± 5.1) mW/cm2
Effective radiating area @ -12 dB A-12dB = (4.47 ± 0.89) cm2
Thermal index (TI) and mechanical index (MI) are always below 1.0.
Signal Quality Indication Poor Quality empty
Acceptable Quality half-full
Good Quality full
Beat-to-Beat change (max.) for Ultrasound 28 bpm
US Frequency 1 MHz ± 100 Hz
US Signal range 3.5 μVpp—350 μVpp @ 200 Hz
US Burst Repetition Rate 3.0 kHz
Duration ≤100 μs
FMP Signal Range @ 33 Hz 200 μVpp—40 mVpp
115
9 Specifications
Toco
Performance Specifications
Toco
Measurement Method Strain Gauge Sensor Element
Sensitivity 1 unit = 2.5 g
Resolution Display 1 unit
Printer 1/4 unit
Measurement Range 400 units
Signal Range 0—127 units
Maximum Offset Range -300 units
Baseline Setting 20 units
Update Rate Display 1 per second
Printer ~4 per second
Auto Offset Correction 3 seconds after connecting the transducer, the Toco value is set
to 20 units
Auto Zero Adjust Toco value is set to zero following a negative measurement
value for 5 seconds
Performance Specifications
Maternal Pulse from Toco
Emitted Light Energy ≤15 mW
Wavelength Range 780—1100 nm
Range 40—240 bpm
Resolution 1 bpm
Display Update Rate 1 per second
Accuracy ± 2% or 1 bpm, whichever is greater
Update Rate every 4 seconds
IUP
Performance Specifications
IUP
Measurement Method Passive Resistive Strain Gauge Elements
Measurement Range -100—+300 mmHg
Signal Range -99—127 mmHg or (-13.2—16.9 kPa)
Resolution Display 1 mmHg
Printer 1/4 mmHg
Sensitivity 5 μV/V/mmHg
Offset Compensation +100—-200 mmHg
Accuracy (not including sensor accuracy) ±0.5% per 100 mmHg
116
9 Specifications
Performance Specifications
Update Rate Display 1 per second
Printer ~4 per second
Auto Offset Correction 3 seconds after connecting the transducer, the IUP value is
set to 0 mmHg
ECG
Complies with IEC 60601-2-27:2011+AC:2012/EN 60601-2-27:2006+AC:2006 except clauses listed
below:
• 201.6.2, 201.8.5.5
• 201.12.1.101
• 202.6.2.101
Sub clause 201.12.1.101.15 on heart rate range and accuracy is passed.
Performance Specifications
ECG
Performance Specifications Type DECG Single Lead ECG (derived from
Fetal Scalp Electrode)
MECG Single Lead ECG (derived from RA
and LA electrodes)
Measurement Range 30—240 bpm
Resolution Display 1 bpm (display update rate 1/s)
Recorder 1/4 bpm
Wave Speed (Global Speed) 6.25 mm/sec, 12.5 mm/sec,
25 mm/sec, 50 mm/sec
Accuracy ±1 bpm or 1%, whichever is greater
(non-averaging)
Beat-to -Beat (max.) MECG: 28 bpm
DECG: 28 bpm (with Artifact
Suppression On)
Differential Input Impedance >15MΩ
Electrode Offset Potential Tolerance ±400 mV
INOP Auxiliary Current (Leads Off Detection) <100 µA
Input Signal Range DECG 20 µVpp—6 mVpp
MECG 150 µVpp—6 mVpp
Dielectric Strength 1500 Vrms
Defibrillator Protection None
ESU Protection None
Paced pulse detection None
117
9 Specifications
WARNING
The fetal/maternal monitor is not a diagnostic ECG device. In particular, the display of fetal/maternal
ECG is intended only for evaluating signal quality for fetal/maternal heart rate as derived from the
ECG waveform.
When in doubt, it can be used to identify sources of compromised signal quality, such as noise or
muscle artifacts. It can subsequently be used to verify the result of measures taken to resolve them
(e.g., checking ECG cable connections or adapting the fetal ArtifactSuppress configuration).
The safety and effectiveness of the displayed fetal/maternal ECG waveform (i.e., P, QRS and T
segments) for evaluation of fetal/maternal cardiac status during labor have not been evaluated.
118
9 Specifications
119
9 Specifications
WARNING
No modification of the Avalon CL base station and the cableless transducers is allowed.
The base station is an electrical Class II device in connection with M2702A/M2703A (FM20/FM30)
and M2704A/M2705A (FM40/FM50). The device complies with the following major international
safety and performance standards:
• IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010
• IEC 60601-1-6:2010 / EN 60601-1-6:2010
• IEC 60601-1-8:2006 / EN 60601-1-8:2007+AC:2010
• IEC 60601-2-49:2011 / EN 60601-2-49:2001
• ANSI/AAMI ES60601-1+C1:2009+A2:2010
• CAN/CSA C22.2#60601-1-08+TC2:2011
• JIS T 1303 2005
• AS/NZS 3200.1.0-1998
The possibility of hazards arising from hardware and software errors was minimized in compliance
with ISO 14971:2012, IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010.
Alarm sounds are compliant with Standard IEC 60601-1-8:2006 / EN 60601-1-8:2007+AC:2010.
120
9 Specifications
Radio
The Avalon CL Transducer System complies with the following major international radio standards:
• ETSI EN 300 220-1:2012
• ETSI EN 300 220-2:2012
• ETSI EN 301 489-1:2011
• ETSI EN 301 489-3:2003
• FCC 47 CFR Part 95
• IC RSS-210 Issue 8
• ARIB STD-T108
• ETSI EN 300 328:2006
• ETSI EN 301 489-17:2009
• FCC 47 CFR Part 2 & 15
• AS/NSZ 4268
• AS/NSZ 4771:2000+A1:2003
• ARIB STD-T66
CAUTION
The use of accessories, transducers and cables other than those specified may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the device.
WARNING
Do not use cordless/mobile phones or any other portable RF communication system within the
patient vicinity, or within a 1.0 m radius of any part of the fetal monitoring system.
121
9 Specifications
WARNING
For paced patients: The radiated SRR power of the CL SpO2 and NBP Maternal Cableless
Measurement Devices, the CL Transmitter, and other sources of radio-frequency energy, when used in
very close proximity of a pacemaker, might be sufficient to interfere with pacemaker performance.
Due to shielding effects of the body, internal pacemakers are somewhat less vulnerable than external
pacemakers. However, caution should be exercised when monitoring paced patients.
In order to minimize the possibility of interference, avoid positioning and wearing the Cableless
Measurement Devices in very close proximity to a pacemaker. Consult the pacemaker manufacturer
for information on the RF susceptibility of their products
EMC Testing
CAUTION
Fetal parameters, especially ultrasound and ECG, are sensitive measurements involving small signals,
and the monitoring equipment contains very sensitive high gain front-end amplifiers. Immunity levels
for radiated RF electromagnetic fields and conducted disturbances induced by RF fields are subject to
technological limitations. To ensure that external electromagnetic fields do not cause erroneous
measurements, it is recommended to avoid the use of electrically radiating equipment in close
proximity to these measurements.
The product and associated accessories can be susceptible to interference from continuous, repetitive,
power line bursts, and other RF energy sources, even if the other equipment is compliant with
EN 60601-1-2 emission requirements. Examples of other sources of RF interference are other medical
electrical devices, cellular products, information technology equipment, and radio/television
transmissions.
When electromagnetic interference (EMI) is encountered, for example, if you can hear spurious noises
on the fetal monitor's loudspeaker, attempt to locate the source. Assess the following:
• Is the interference due to misplaced or poorly applied transducers? If so, re-apply transducers
correctly according to directions in this book, or in the Instructions for Use accompanying the
accessory.
• Is the interference intermittent or constant?
• Does the interference occur only in certain locations?
• Does the interference occur only when in close proximity to certain medical electrical equipment?
122
9 Specifications
Once the source is located, there are a number of things that can be done to mitigate the problem:
1 Eliminating the source. Turn off or move possible sources of EMI to reduce their strength.
2 Attenuating the coupling. If the coupling path is through the patient leads, the interference may be
reduced by moving and/or rearranging the leads. If the coupling is through the power cord,
connecting the system to a different circuit may help.
3 Adding external attenuators. If EMI becomes an unusually difficult problem, external devices such
as an isolation transformer or a transient suppressor may be of help. Your service provider can be
of help in determining the need for external devices.
Where it has been established that electromagnetic interference is affecting physiological parameter
measurement values, a physician, or a suitably qualified person authorized by a physician, should
determine if it will negatively impact patient diagnosis or treatment.
System Characteristics
The phenomena discussed above are not unique to this system, but are characteristic of fetal patient
monitoring equipment in use today. This performance is due to very sensitive high gain front end
amplifiers required to process the small physiological signals from the patient. Among the various
monitoring systems already in clinical use, interference from electromagnetic sources is rarely a
problem.
123
9 Specifications
Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user must ensure
that it is used in the appropriate environment as described below.
124
9 Specifications
125
9 Specifications
de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si
le brouillage est susceptible d'en compromettre le fonctionnement.
L'utilisation de cet appareil de télémesure est permise seulement dans les hôpitaux et établissements de
soins de santé. Cet appareil ne doit pas être mis en marche dans des véhicules (y compris les
ambulances et autres véhicules associés aux établissements de santé). La personne qui installe/utilise
cet appareil doit s’assurer qu’il se trouve à au moins 80 km de l’Observatoire fédéral de
radioastrophysique (OFR) de Penticton en Colombie-Britannique. Les coordonnées de l’OFR sont:
latitude N 49° 19’ 15», longitude O 119° 37 12 ′′. La personne qui installe/utilise un système de
télémesure médicale ne pouvant respecter cette distance de 80 km (p. ex. dans la vallée de l’Okanagan
(Colombie-Britannique), doit se concerter avec le directeur de l’OFR et obtenir de sa part une
autorisation écrite avant que l’équipement ne puisse être installé ou mis en marche. Le directeur de l’
OFR peut être contacté au 250-497-2300 (tél.) ou au 250-497-2355 (télécopieur). (Le Directeur des
Norm es réglementaires d’Industrie Canada peut également être contacté).
This device should not be modified (otherwise the granted designation number will become invalid).
126
9 Specifications
Over the frequency range 800 MHz—2.5 GHz, the recommended separation distance in meters (d) is found by the
following equation:
For a compliance level of 3.0 VRMS: d = 2, 3√P
Field strengths from fixed transmitters, such as base stations, or radio, (cellular, cordless) telephones,
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
127
9 Specifications
location in which the device is used exceeds the applicable RF compliance level above, it should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the device.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
If you require further information or assistance, please contact Philips Support.
Table 5 - Separation Distance (d) in Meters According to Frequency of Transmitter at IEC 60601-1-2 Test
Compliance Level
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
Rated Maximum Output
Power (P) of Transmitter
(in Watts)
128
9 Specifications
129
9 Specifications
130
9 Specifications
131
9 Specifications
132
10
10 Upgrades
The bus master and CL transducer firmware of the base station and the CL transducers can be
upgraded by the M3086A IntelliVue Support tool from a PC (component upgrade).
The Avalon CL base station needs to be connected through the common connection cable to the fetal
monitor.
See the M3086A IntelliVue Support tool Instruction for Use for details on how to perform a
component upgrade.
NOTE
This software option is not available for a FM20/30 #E25. The IntelliVue CL Pods are not supported
for the battery option of the FM20/30.
133
10 Upgrades
134
1 Index
A Configuring SRR Channels 38 I
Connecting Two Avalon CL Base Stations
After Each Service, Maintenance, or Repair at the FM40/50 29 Initial Inspection 26
Event 83 Installation 8, 25
Connection Options 27
Antenna PCA 44 Installation Check List 25
Area of Reach 107 D Installation Scope 17
Avalon CL Radio 113 Interface Specification 113
Disassembly Bottom Cover Avalon CL
Avalon CL with T108 126 Base Station 65 Introduction 13, 47
Avalon CL with WMTS 125 Disassembly Charging Contacts 57 IUP 116
B Disassembly CL Pod Device Charger 54
M
Disassembly CL Transducer Battery 67
Base Station Hardware Overview 43 Disassembly Short Range Radio Boards 52 Maintenance, Repair, and Care 9
Base Station PCA 43 Disassembly Standby Button/ Mounting Options 18
Battery Check 84 Switchboard 62
Battery Cycle 86 Disassembly Top Cover Avalon CL Base O
Battery Disposal 88 Station 48
OB Radio 32
Battery Exchange 88 Disassembly/Reassembly 47
OBR Channel Settings 33
Battery Report 87 E OB-Radio Scan 34
Battery Specifications 113 Observe System Performance 106
Battery Storage 88 ECG 117
One-Channel or Narrow-band
Before Each Use 82 Electric Hazards 7 Interference 108
Broadband Interference 108 Electrical Requirements 16 Other Regular Tests 94
Electromagnetic Compatibility (EMC) 121 Overview 95
C Electromagnetic Emissions and
Immunity 123 P
Cableless Charging Interface 45
Electromagnetic Immunity 124
Care and Cleaning 79 Performance Assurance Test 88
EMC Testing 122
Carrier to Noise Ratio 106 Performance Specifications 113
Environmental Requirements 15
Checking Contacts 97 Physical Specifications 112
Environmental Specifications 111
Checking the Fetal Monitor 96 Power On Test 83
Essential Performance 16
CL ECG/IUP and CL Toco+ MP with Preventive Maintenance Procedures 84
DECG Check 91 Example 1 20
Procedure for Local Staff 13
CL ECG/IUP and CL Toco+ MP with Example 2 22
Procedure for Philips Personnel 14
MECG Check 92
F Q
CL ECG/IUP and CL Toco+MP with IUP
Check 93 Fetal/Maternal Specifications 115
Question the User 107
CL ECG/IUP Front-end Hardware 46 Finding Recommended Separation
CL Toco+ MP Transducer Functional Distances 126 R
Check 91 Firmware Upgrades and Software
CL Toco+MP Front-end Hardware 46 Installation 30 Radio 121
CL Transducer Hardware Overview 45 Frequently Asked Questions 41 Radio Compliance Notice 125
CL Transducers Functional Tests 89 Functional Description of the RF/CPU Radio Frequency Interference 7
CL Ultrasound Front-end Hardware 46 Hardware 45 Radio Frequency Plan 18
CL Ultrasound Transducer Functional Radio Frequency Radiation Exposure
G Information 128
Check 89
Claims for Damage 26 Gathering Data 106 Radio Information Canada 125
Common Problems 102 General Information 5 Range Definition 107
Recommendations for Cableless
135
Monitoring 40 SRR Channel Settings Configuration 38
Recommended Frequency 80 SRR PCA 45
Recommended Separation Distances from Switchboard PCA 45
Other RF Equipment 128 Symbols on the System 128
Reducing Electromagnetic System Characteristics 123
Interference 122
System is Completely Inoperative 95
Regulatory and Standard Compliance 120
System Reset 96
Removing Bottom Cover Assembly CL
Base Station 65 System Test 84
Removing Charging Contacts 57 T
Removing CL Pod Charger 54
Technical INOPs Avalon CL 98
Removing CL Transducer Battery 67
Telemetry Interfaces 30
Removing Standby Button/
Switchboard 62 Terminology and Definition 79
Removing the Short Range Radio Testing Alarms 88
Boards 52 Testing and Maintenance 79
Removing Top Cover Assembly CL Base Testing Sequence 82
Station 48 Theory of Operation 43
Repacking 26 Toco 116
Repair Strategy 6 Tools Required 48
Replacing Bottom Cover Assembly CL Top Cover Assembly Overview 51
Base Station 66
Transducer Charging Contacts 45
Replacing Charging Contacts 59
Troubleshooting 95
Replacing CL Pod Charger 55
Troubleshooting the Top Cover 96
Replacing CL Transducer Battery 71
Tutorial Frequency Planning 19
Replacing Standby Button/Switchboard 64
Replacing the Short Radio Boards 53 U
Replacing Top Cover Assembly CL Base
Station 51 Ultrasound 115
Reporting of Test Results 94 Unpacking/Checking Shipment 26
RF Interference 108 Upgrades 133
RF Problems 105 Use Environment 8
Roles and Responsibilities 13 V
RSSI Trace 37
Visual Inspection 82
S
W
Safety and Performance 120
Walkthrough 10
Safety Requirements 16
Warnings and Cautions 6
Safety Tests 41, 84
Warnings, Cautions, and Safety
Scanning the Available Frequency
Precautions 83
Range 109
What to Do Next 5
Setting Expectations 40
When are Special Configurations
Setting Up OB-Radio 32
Required? 6
Short Range Radio 38
When to Perform Safety Tests 81
Site Planning 17
Who Should Read This Guide 5
Site Preparation 13
WMTS Frequency Registration WMT 31
Site Preparation Responsibilities 13
Site Requirements 15
Software Option for CL Pods 133
Space Requirements 15
Spare Parts 77
Specifications 111
136
Part Number 453564526801
Published in Germany 11/2014
*453564526801*