Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

Field display user guide


for

Roxar Multiphase meter 2600

Page 1 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

TABLE OF CONTENTS

1.  Purpose ........................................................................................................................................... 3 


2.  Abbreviations / Definitions ............................................................................................................ 3 
3.  Revision history .............................................................................................................................. 3 
4.  Documentation ................................................................................................................................ 3 
4.1  Standard documentation and records ....................................................................................... 3 
4.2  Location and Precautions .......................................................................................................... 3 
4.3  Storage time .............................................................................................................................. 3 
5.  Procedure ........................................................................................................................................ 4 
5.1  General description ................................................................................................................... 4 
5.2  Power – up ................................................................................................................................ 4 
5.3  Display & Control unit ................................................................................................................ 5 
5.4  Start – up ................................................................................................................................... 5 
5.5  Real – time flow data ................................................................................................................. 6 
5.6  Well test operations ................................................................................................................... 6 
5.7  Configuration ............................................................................................................................. 7 
5.7.1  Setup by using Service Console Software ......................................................................... 7 
5.7.2  Setup by using Field Display .............................................................................................. 8 
5.8  Starting well test ........................................................................................................................ 8 
5.9  Stop well test ............................................................................................................................. 8 
5.10  Well test flow chart .................................................................................................................... 9 

Page 2 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

1. PURPOSE
The purpose of this document is to inform the user on the functions of the Field Display Unit for the
Roxar Multiphase Meter

2. ABBREVIATIONS / DEFINITIONS
Roxar MPFM 2600 Roxar Multiphase meter 2600

3. REVISION HISTORY
Revision Description
A New customer documentation
B New template

4. DOCUMENTATION

4.1 Standard documentation and records

Document name Type Doc. Reference


[1] Well Test Record Sheet Appendix 1
[2] Operating Instructions Roxar MPFM 2600. TCE 091983/B

4.2 Location and Precautions


Doc. level: 4
Document Classification: Open
All records from this process shall be stored in the document system under relevant folder.

4.3 Storage time


The records shall be stored during the products life time or as specified in the clients’ contract, a
minimum of 20 years is required.

Note: If the client requires notification before records are discarded, this must be stated on the record itself.

Page 3 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

5. PROCEDURE

5.1 General description


In addition to the flow computer, the enclosure contains a display unit and push-buttons for local
operations.

The enclosure also includes a power supply, computer assembly, barriers to the intrinsically safe
units, converter with galvanic separation, communication switch to peripheral units, and HART
modem.

All wiring and connections to the system should be finalized and checked by authorized Roxar
personnel before external power is applied to the EExd enclosure.

Appendix 1: Well Test Record Sheet may be used to record the measured values.

The electronics components in the flow computer are not repairable and may be subject to
replace. The services should be performed by authorized Roxar personnel only!

5.2 Power – up
The system power is switched ON by setting the changeover switch marked MAIN POWER on the
lower left side of the enclosure (not shown on the picture), to the position marked ON. The system
is switched OFF by setting changeover switch back to the position marked OFF.

Page 4 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

5.3 Display & Control unit


The display unit controlled by the push-button panel is installed in the front door of the EExd
enclosure. The display shows flow rates, primary parameters, totalized flow rates and well test
operations. Push-buttons have a navigation/select function used to toggle between the display
modes as well as for well test setup. The following table describes the labelling and functions of
the push-buttons:

Label Function Description


Used to activate and deactivates the Local Display
DISPLAY
Display ON/OFF Unit. The LCD screens will automatically show
POWER
ACTUAL FLOW.
(1) When enabled, ACTUAL FLOW will appear on
Display screen.
MODE
Enable/Disable (2) Toggles between Actual and Standard conditions
in the WELL TEST MENU.
(1) Changes between:
 ACTUAL FLOW
 ACT. TOTALISERS
 STANDARD FLOW
Scroll / Escape /  STD. TOTALISER
PAGE
Select  PRIMARY PARAMETERS
 WELL TEST MENU
(2) Used as Escape /Select duration /Select well
number
in WELL TEST MENU.
Resets totaled Used to RESET TOTALISERS
RESET
values (Qoil, Qwater, Qgas, Water Cut and GVF)
Navigate /
AUX 1 & 2 Used in WELL TEST MENU as instructed on screen.
Select

5.4 Start – up
Press POWER to activate the local display.
Press MODE until ACTUAL FLOW screen appears.

ACTUAL FLOW 2002-12-17 09:05:58

Q Oil 0,0 m3/hr


Q Water 0,0 m3/hr
Q Gas 0,0 m3/hr
Water Cut 0,0 %
GVF 0,0 %
Velocity 0,0 m/s

Made by Roxar Flow Measurement AS


Bergen / Stavanger, Norway 01

Run the production flow through the system for at least 15-20 minutes and make sure that there is
no “Process Alarm” or “Technical Alarm”.

Page 5 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600
5.5 Real – time flow data
Read the real-time measurements available in ACTUAL FLOW. Press PAGE four (4) times to read
the real-time measurements available in PRIMARY PARAMETERS and record the measured
values in Appendix 1: Well Test Record Sheet.

PRIMARY PARAMETERS 2002-12-17 09:05:58

Temperature 0,0 degC


Pressure 0,0 bar(g)
Dp 0,0 mbar
Density 0,0 g/ml
Perm. 0,0
Cond. 0,0 S/m

Made by Roxar Flow Measurement AS


Bergen / Stavanger, Norway 01

5.6 Well test operations


The first well test should be performed under the supervision of an authorized Roxar
representative. Make sure that the Roxar MPFM 2600 Flow Computer power is ON and has the
correct configuration, parameter set-up, and all of the isolating valves are in the right position.

Page 6 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

5.7 Configuration
5.7.1 Setup by using Service Console Software
Well identification and the fluid parameter setup should be completed using the Service Console
Software.

All of the wells should be named with a number between 1 and 50. Alphanumeric characters
cannot be used.

The flow computer can store up to five (5) different fluid parameter sets (PVT data and fluid
properties). The fluid parameter set that is used during a well test must be pre-selected [2].

Page 7 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

5.7.2 Setup by using Field Display


The well name (identification number) and the corresponding fluid parameter set (number) should
be selected as follows:

If the push button is not activated within 15 seconds, the display will exit to the
default menu “ACTUAL FLOW”

Press PAGE five (5) times to display the WELL TEST MENU.
Press MODE to toggle between standard and actual conditions.
Press AUX2 to start new well test. The NEW WELL TEST menu will appear.
Press AUX1 to change setup. The WELL TEST SETUP menu will appear.
Press AUX1 to change well number. The WELL NUMBER SELECTION menu will appear.
Press AUX1 or AUX2 to increase or decrease the well that is to be tested, respectively.
Press PAGE to return to the WELL TEST SETUP menu.
Press AUX2 to change well test duration. The WELL TEST DURATION menu will appear.
Press AUX1 or AUX2 to decrease or increase the selected well test duration, respectively.
When starting a well test from the field display, the well number and the corresponding fluid
parameter set will then automatically be loaded from the pre-selected list into the flow computer.

5.8 Starting well test


Press RESET two (2) times to reset the accumulated values.

ACT. TOTALISERS 2002-12-17 09:05:58


RESET TIME 2002-12-17 09:00:00
If you really want to reset accumulated values,
Press RESET once again within Q Oil 0,0 m3/hr
15 seconds Q Water 0,0 m3/hr
Q Gas 0,0 m3/hr
Watercut 0,0 %
GVF 0,0 %

Made by Roxar Flow Measurement AS Made by Roxar Flow Measurement AS


Bergen / Stavanger, Norway 01 Bergen / Stavanger, Norway 01

Press PAGE five (5) times to display the WELL TEST MENU.
Press AUX2 to start new well test. The NEW WELL TEST menu will appear.
Press AUX2 two (2) times to start well testing. WELL TEST IN PROGRESS will be displayed.

The well identification number, test duration, remaining time, and measured values will be
displayed along with options to STOP and Escape.

5.9 Stop well test


The well test will stop after the requested amount time and WELL TEST COMPLETED will appear.
Test results can be read from the display or remotely.

Page 8 of 9
Roxar Template doc. no. / rev.: 034708/D
PO: 4500195123 UM_MPFM-2193-13 ROX000230195/AA

Field display user guide


Document no. / rev.: 106425 / B
Roxar Multiphase meter 2600

5.10 Well test flow chart

WELLTEST

AUX1 = View previous


AUX2 = Start new
MODE = Change from ACT to STD (or STD to ACT)

PAGE =

COMPLETED Well NEW


Start time -0 -13 09:16 Duration Well number = 1, Test duration = 2
----- AUX1 = Change setup. AUX2 = START
Measured
----- PAGE =
AUX1=previous test. AUX2=next
PAGE =

WELLTEST
If you really want to START a
AUX1 = Change well
Press AUX2
AUX2 Change
PAGE =
PAGE =

WELLTEST IN PROGRESS (WELLTEST COMPLETED)


WELL NUMBER WELLTEST Well 1 Duration 00:53 Remains
Well number = Duration = 2 -------
AUX1 = lower well AUX1 = shorter Measured
AUX2 = higher well AUX2 = longer -------

PAGE = select PAGE = select AUX2 = STOP well testing (PAGE =

8
If you really want to STOP
press AUX2

PAGE =

Page 9 of 9