Customer Support Bulletin

Ph# 608/828-2663 Fax# 608/826-7107 e-mail: support@lunarcorp.com

Volume 24

DPX Series January 17, 2000


The purpose of these bulletins is to keep all DPX Series trained personnel
informed of enhancements and known issues within the DPX Series of products
that affect installation, training and service.

Please note that DPX-NT is covered in the DPX-NT / PRODIGY CSB and that
the items in this bulletin do not apply to DPX-NT.

In this bulletin:
Subject Bulletin #
Software Version 4.7c Release D00049

Software Version 4.7c Release D00049


Please note this is a “bug fix” release only – no new features were added.

Bugs Addressed in DPX Version 4.7c Software:


• The Fracture Risk color bars are missing from the Reference Graph in
version 4.7b, even though the labels “Normal”, “Osteopenia”, and
“Osteoporosis” still appeared.
• The color bars are properly displayed in 4.7c.

• Fracture Risk Graphs were opened to all Females for the following Scan
Types and Regions:
• AP Spine (all sub-regions, but not Body ROIs)
• Femur Neck
• Femur Total
• Forearm Radius 33%
• Fracture Risks Graphs were available for all Females in versions
4.7 and 4.7a.
• In version 4.7b the Fracture Risk Graphs were limited to Caucasian
Females only.

Volume 24 1 January 17, 2000


This document contains confidential or proprietary information of Lunar Corp. neither the document nor the information is to be reproduced, distributed, used or
disclosed, either in whole or in part, except as specifically authorized by Lunar Corp.
Customer Support Bulletin
Ph# 608/828-2663 Fax# 608/826-7107 e-mail: support@lunarcorp.com

• The DPX.BAT files did not always update correctly when a software update
was performed.
• The DPX.BAT file was copied directly from the source directory and often
contains the statement “CD\ (Source directory)”. The new DPX Icon on
the desktop would use the information in the DPX.BAT file and would then
run the older version of software even though the Icon would appear to be
pointing at the new software.
• The work around was to edit the DPX.BAT file with Notepad to point
the DPX.BAT file at the new software.
• 4.7c addresses this bug and will create an Icon that points at the 4.7c
software when the update software is loaded.

Volume 24 2 January 17, 2000