HK-51 Sodium Benchtop Huakeyi
HK-51 Sodium Benchtop Huakeyi
Instruction Manual
Version: 3.9
PREFACE
HK-51 Benchtop Sodium Meter is the latest laboratory instrument developed by Beijing Huakeyi
Power Plant Instrument Research Institute. With the powerful technical superiority, Huakeyi
developed HK-51 Benchtop Sodium Meter. It incorporates practice experiences of about 10 years
and advanced technologies of the whole world, can measure sodium of aqueous solutions for
applications in institutes and universities, as well as electric power, chemical/petrochemical,
paper/paper pulp, pharmaceutical, food and environmental protection and other related fields. The
meter has such characteristics as fast response, accurate analysis and easy operation.
The indicator is 192×64 lattice back-lighted LCD displays in English or Chinese. There are 8
buttons on the instrument keypad; they can realize all operation and configuration. HK-51 has
four-digit security code for the meter programming in order to prevent accidental or unwanted
changes to program settings, display and calibration.
The Instruction Manual provides information about HK-51 Benchtop Sodium Meter. It is compiled
for users to familiarize the structure, principle of operation and operation procedures of HK-51
Benchtop Sodium Meter. The potentially fallible and easy-to-ignore problems during operation are
detailed for preventing unnecessary troubles and failures for long-term safe and reliable operation.
WARNING
Be sure to read this Instruction Manual carefully before proceeding with operation and
maintenance of this instrument.
CONTENT
1. INTRODUCTION ................................................................................................................................. 1
2. INSTALLATION .................................................................................................................................. 3
3. SPECIFICATIONS .............................................................................................................................. 5
4. PROGRAMMING ................................................................................................................................ 6
1
HK-51 Benchtop Sodium Meter CONTENT
2
HK-51 Benchtop Sodium Meter INTRODUCTION
1. INTRODUCTION
HK-51 Benchtop Sodium Meter is consisted of the meter, sodium electrode, reference electrode, ATC
temperature electrode and the electrode holder. It adopts the high sensible electrode system to make
sure the stability and accuracy of sodium measurement.
The meter has one temperature input channel. When automatic temperature compensation is needed,
the sample temperature can be measured by a Pt1000 thermometer, otherwise manual temperature
compensation will replace the automatic temperature compensation
The effect of the sodium concentration variation to the electric potential of the sodium electrode can be
depicted by the Nernst equation:
RT
E = E 0 + 2.3 lg a Na +
nF
where:
E — The electric potential generated by the sodium electrode, mV;
a Na +
E0 —The electric protential of the electrode when is 1mol/L, mV;:
-1 -1
R — Ideal gas constant; 8.314JK mol ;
T — Absolute temperature of the sample water, K;
F — Faraday constant, 9.649×104C·mol-1
n — The electron number participating the reaction, for sodium ion it is +1;
a Na +
— The activity of the sodium ion in the solution, mol/L.
1
HK-51 Benchtop Sodium Meter INTRODUCTION
The above equation had indicated that, the measured electric potential will variate with the variation of
the temperature and the related ion concentration. In order to eliminate the errors caused by the water
temperature variation, HK-51 sodium meter will amend the temperature compensation value timely
according to the temperature measured by the ATC temperature electrode.
From the Nernst equation we can see that, at the temperature of 25℃, the theoretical response value
of the selected sodium electrode to the 10 times of ion concentration is 59.16mV, which is called
electrode slope (S). However, most of electrodes will not manifest as theoretical slope. So it is
necessary to calibrate the instrument to determine the actual slope of the electrode. In the actual
application, we calculate out the real slope value and the E0 of the electrode by using two kinds of
standard solution.
When the concentration of sodium ion is in a low level, the existence of the hydrogen ion will cause
magnificent interference. In order to eliminate this interference, basifier should be added in the sample
water and making the pH value be increased up to over 10.5.
2
HK-51 Benchtop Sodium Meter INSTALLATION
2. INSTALLATION
2.2 Mounting
Please read the instruction manual carefully before installing and pay attention to the treatment method
of the electrode.
It is necessary to mount the new sodium electrode provided by Huakeyi. And the particular sodium
electrode is matched with the particular reference electrode.
1. Put the electrode on the bracket.
2. Insert BNC cable connector of sodium electrode into corresponding hole on the meter back panel,
then counterclockwise rotate and lock it. Refer to the figure 2.1 below.
Every reference electrode is provided together with a filling liquid bottle of 60ml. The reference
electrode of the sodium meter is specially made and can not be alternated with other reference
electrodes.
Note: the sodium electrode and the reference electrode shall be used as a fitted couple.
Insert the temperature electrode into the ATC input hole on the back panel of the meter. Refer to the
figure 2.1 below.
3
HK-51 Benchtop Sodium Meter INSTALLATION
Hole for mounting Sodium Electrode Hole for mounting Temperature Electrode
Power Switch
Hole for mounting Reference Electrode Power Supply
4
HK-51 Benchtop Sodium Meter SPECIFICATIONS
3. SPECIFICATIONS
5
HK-51 Benchtop Sodium Meter PROGRAMMING
4. PROGRAMMING
Display
DISPLAY: The indicator is 192×64 lattice back-lighted LCD display in English or Chinese.
KEYPAD: There are 4 function keys (SAVE, LOG, ENTER and ESC) and 4 direction keys (UP, DOWN,
RIGHT and LEFT) on the instrument keypad.
Function Keys:
SAVE : Press the key under measurement screen, “Now Saving, Waiting...” will display. The
meter saves the current data.
LOG : Press the key under the measurement screen, inquiry history data directly.
ENTER : a. During programming, press the key after select some menu or function, the meter will
enter the corresponding one.
b. After update some parameters, press the key to store numbers and settings, and return
to the previous operation screen.
ESC : a. Press the key under measurement screen, Security Code screen will appear. After input
the correct access code, press ENTER key to enter main menu.
b. During programming, press the key to return to the previous menu or screen without
storing changes
6
HK-51 Benchtop Sodium Meter PROGRAMMING
During programming, after select some parameter (display in reverse color), press UP and DOWN key
to update its value, and then press ENTER key to save the revision and return to the previous screen,
or press ESC key to return to the previous screen without saving the revision.
During programming, after select some digit of numerical value (display in reverse color) press UP and
DOWN key to increase or decrease the value of the digit, press LEFT and RIGHT key to move the
cursor from digit to digit. Then press ENTER key to save the revision and return to the previous screen,
or press ESC key to return to the previous screen without saving the revision.
4.2 Power-on
WARNING
· DO NOT apply power to the meter until all electrical connections are verified and secure.
· If the meter is used for the first time, its internal parameters should be set by programming
and the instrument calibration is necessary.
Apply power to the meter and turn on the power switch, the following welcome interface will appear
first:
This interface will last for 5s, then shift to the measurement interface automatically as the below screen:
7
HK-51 Benchtop Sodium Meter PROGRAMMING
Besides the measurement screen, HK-51 sodium meter also has a key unit-Program Unit. Almost all
the parameter setting, calibration, maintenance, information are realized by the program unit. It is
protected by the security code. Use the security code to prevent accidental or unwanted changes for
the program unit.
Access Method:
Under measurement screen, press ESC key to enter Security Code screen:
Operation:
After enter one code with the range from 0000 to 9999, press ENTER key. If the entry is correct, the
meter will move to the main menu screen, otherwise it will return to the measurement screen. Press
ESC key to return to the measurement screen.
8
HK-51 Benchtop Sodium Meter PROGRAMMING
Access Method:
Under measurement screen, press ESC key to enter Security Code screen. Enter correct code, press
ENTER key to move to the below main menu screen.
1. Setup
2. Calibration
3. Log
Choice:
The Main Menu includes 3 sub-menus; they are Setup, Calibration and Log.
Operation:
The meter only has four lines indication. Press UP and DOWN keys to show more choices. After select
some sub-menu, press ENTER key to enter corresponding one.
Access Method:
Choose Setup under the main menu screen. Press ENTER key to enter the below screen:
1. Language
2. Password
3. Adjust Contrast
4. System Time ︾
Choice:
The Sub-menu includes 7 choices; they are Language, Password, Adjust Contrast, System Time,
Temp Comp Type, Setup Comp Temp and Reset.
The details as follows:
¾ Language: Two options, English and Chinese.
¾ Password: The security code prevents accidental or unwanted changes for the meter. It can be
changed.
9
HK-51 Benchtop Sodium Meter PROGRAMMING
Operation:
Press UP, DOWN, LEFT and RIGHT key to move the cursor to the requested place. Press ENTER key
to access the corresponding choice. Update parameters or values as mentioned method in 4.1.1 and
4.1.2. After complete changes, press ENTER key to store the revision and return to the previous
operation level.
Press ESC key to cancel the revision and return to the previous operation screen.
Access Method:
Choose Calibration under the main menu screen. Press ENTER key to enter the below screen:
Choice:
The Sub-menu includes 4 choices; they are Na+ standard solution calibration, Sample calibration,
Temperature calibration and Electric calibration.
The details as follows:
¾ Na+ standard solution calibration:Use the Na+ standard solution to calibrate the meter;
¾ Sample calibration:Use the measuring solution as the standard solution to calibrate the
instrument;
¾ Temperature calibration:Calibrate the temperature measurement system of the meter.
¾ Electric calibration:Calibrate the electric elements of the meter.
Operation:
Press UP, DOWN, LEFT and RIGHT key to move the cursor to the requested place. Press ENTER key
to access the corresponding sub-menu. According to the menu instruction, perform the corresponding
10
HK-51 Benchtop Sodium Meter PROGRAMMING
calibration.
Press ESC key to return to the main menu screen.
REMARK
The detail operating steps of several calibration or trim refer to chapter 5 OPERATION.
Access Method:
Choose Data Log under the main menu screen. Press ENTER key to enter the below screen:
Choice:
The Sub-menu includes 3 choices; they are Data Log, Event Log and Calibration Log. The details as
follows:
¾ Data Log: Display the storage of measured data. It can record 256 items. Under the measurement
screen, press LOG key can also enter this interface.
¾ Event Log: Display the events of operation. These events include power-on, power-off, update
parameter and so on. It can record 256 items
¾ Calibration Log: Display the recorded calibration data. Include standard solution and temperature
calibration. It can record 64 items.
11
HK-51 Benchtop Sodium Meter PROGRAMMING
Operation:
Press UP, DOWN, LEFT and RIGHT key to move the cursor to the requested place. Press ENTER key
to access the corresponding sub-menu and data.
Press ESC key to return to the main menu screen.
12
HK-51 Benchtop Sodium Meter OPERATION
5. OPERATION
5.1 Power-on
WARNING
1. Ensure to carefully read the chapter 4 PROGRAMMING of the instruction manual
before applying power to the meter.
2. Ensure all connections are correct and reliable.
3. The diisopropylamine is of hazardous goods, please wear protection glasses and
gloves while operating. Refer to the appendix for the detailed protection standard
After the meter is normal operation, it is necessary to perform the calibrations, including:
• Na+ standard solution calibration
• Sample calibration;
• Temperature calibration;
• Electric calibration
Among which, the Temp Cal and Electric Cal are factory calibrated, there is no need to recalibrate. In
order to ensure the accurate and reliable measurement, Na+ standard solution calibration (2-points Cal)
is necessary prior to the initial operation.
Na+ standard solution calibration is based on the usage of a known pNa value standard solution. 1-point
or 2-points calibration is available, among which, 1-point calibration is to correct the curves stored in
the memory, while 2-points calibration is to create a curve for mV voltage values to pNa values.
During normal measurements, the meter results the measured mV voltage values in accordance with
the curves stored in the memory. And then it is necessary to perform temperature compensation, and
display the pNa value at 25℃.
CAUTION
2-points calibration is required for the initial operation, electrodes replacement and
long-term out of operation.
13
HK-51 Benchtop Sodium Meter OPERATION
Preparation:
¾ According to the application to prepare the standard solutions, each solution is at least 500mL.
¾ If take two point calibration, the difference between the concentration of the two solution should be
at least 10 times, for example: 230μg/L and 2300μg/L;
¾ The beaker should be rinsed by the standard solution at least once;
¾ Basifier solution:diisopropylethylamine solution(the concentration not lower than 98%),every
30ml sample use one drop of the diisopropylethylamine solution, both the normal measurement
and calibration should take this proportion.
Note:DO NOT use Ammonia as the basifier solution, so as to avoid the interference brought by
NH4+.
Measurement interface
Input Password
Security code prevents accidental or unwanted
changes for the meter. Enter correct code via UP,
DOWN, RIGHT and LEFT keys.
Press ENTER key to enter main menu.
Calibration Sub-menu
Move the cursor to Na+ std solution calibration
via DOWN and UP keys.
14
HK-51 Benchtop Sodium Meter OPERATION
Standard solution 1
Use the basification standard solution 1 to rinse
the electrodes at least two times, then immerse
the electrode to the beaker that filled with the
basification standard solution 1
Press ENTER key to continue
Calibrating …
When to end calibration is decided by settings.
Na+ voltage and the temperature voltage are new
measured ones
Wait ……
Standard solution 2
Use the basification standard solution 2 to rinse
the electrodes at least two times, then immerse
the electrode to the beaker that filled with the
basification standard solution 2.
Press ENTER key to continue
15
HK-51 Benchtop Sodium Meter OPERATION
Calibrating …
When to end calibration is decided by settings.
Na+ voltage and the temperature voltage are
new measured ones
Wait ……
Cal Fail
Calibration success
If success, the meter will automatically return to
the calibration place.
Na+ standard solution calibration completed.
Calibration Fail
If this screen appears, the calibration is invalid,
without storing the calibration data.
Sample calibration is using the known concentration sample solution to do the 1-point calibration,
and has the feature of flexible and quicker.
Measurement interface
16
HK-51 Benchtop Sodium Meter OPERATION
Input Password
Security code prevents accidental or unwanted
changes for the meter. Enter correct code via UP,
DOWN, RIGHT and LEFT keys.
Main Menu
Move the cursor to Calibration via DOWN and UP
keys.
Calibration Sub-menu
Move the cursor to Na+ std solution calibration
via DOWN and UP keys.
Calibrating …
When to end calibration is decided by settings.
Na+ voltage and the temperature voltage are new
measured ones
Wait ……
Cal Fail
Calibration success
If success, the meter will automatically return to
Sample Calibration place.
17
HK-51 Benchtop Sodium Meter OPERATION
Calibration Fail
If this screen appears, the calibration is invalid,
without storing the calibration data.
Temperature calibration is factory calibrated, there is no need to recalibrate. If any large temperature
deviations are found during operations, it may use the temperature calibration to overcome the
resulting errors.
Measurement interface
Input Password
Security code prevents accidental or unwanted
changes for the meter. Enter correct code via
UP, DOWN, RIGHT and LEFT keys.
Main Menu
Move the cursor to Calibration via DOWN and UP
keys.
Calibration Sub-menu
Move the cursor to Temperature Calibration via
DOWN and UP keys.
18
HK-51 Benchtop Sodium Meter OPERATION
Calibrating ……
When to end calibration is decided by settings.
The values of temperature & voltage are new
measured ones.
Waiting ……
Calibrating ……
When to end calibration is decided by settings. The
values of temperature & voltage are new measured
ones.
Waiting ……
Cal Fail
Calibration Fail
If this screen appears, the calibration is invalid,
without storing the calibration data.
WARNING
The electrical calibration is completed prior to shipment, and it is no need to repeat this
operation. If necessary, the electrical calibration should be performed by qualified technicians
with proper equipment.
Preparations:
¾ Potentiometer (0 – 1.2000V, increment 0.0001V).
¾ Turn off power supply, and remove electrode from the meter.
¾ Connect the BNC connector with the potentiometer output. Wire the internal lead to positive (+),
the external lead to negative (-).The potentiometer will be used for simulating electrode input
signal.
¾ Turn on power switch.
Measurement interface
Input Password
Security code prevents accidental or unwanted
changes for the meter. Enter correct code via
UP, DOWN, RIGHT and LEFT keys.
20
HK-51 Benchtop Sodium Meter OPERATION
Main Menu
Move the cursor to Calibration via DOWN and UP
keys.
Calibration Sub-menu
Move the cursor to Electric Cal via DOWN and UP
keys.
Volt1 Calibration
Enter Volt1 value with the potentiometer
(between ±190 mV), and then use direction keys
to make the screen reading match for that of the
potentiometer.
Calibrating…
When to end calibration is decided by settings.
The value of electric voltage is new measured one.
Please wait.
21
HK-51 Benchtop Sodium Meter OPERATION
Volt2 Calibration
Enter Volt2 value with the potentiometer
(between ±190 mV), the difference of Volt1 &
Volt2 is more than 100mV. Then use direction
keys to make the screen reading match for that of
the potentiometer.
Press ENTER key to continue.
Calibrating ……
When to end calibration is decided by settings.
The value of electric voltage is new measured
one.
Waiting …………….
Cal Fail
Calibration Fail
If this screen appears, the calibration is invalid,
without storing the calibration data.
After calibrations completed, the meter can carry out normal measurements.
1. Dribble 3-4 drop of the basification solution to the plastic beaker, add about 100ml of the sample to
the beaker and then shake the beaker gently for several seconds;
2. Use the above sample to wash the electrode.
3. Repeat the procedure 1 and 2 to rinse the electrode for at least two times.
4. Prepare the sample solution got from procedure 1 and then immerse the electrode into the beaker
and wait for the measuring value become stable. (about 2 minutes).
5. When the reading is stable, the value can be used.
22
HK-51 Benchtop Sodium Meter OPERATION
REMARK
+
When the concentration of Na is in a high level, rinse the electrode for two times; and if the
concentration is in a low level (less than pNa6), it is recommended to rinse the electrode for
more times so as to reduce the interference and get a more accurate result.
The newer the electrode is, the faster its stabilization. When the electrode has been used
for some time, its stabilizing time will correspondingly be increased.
23
HK-51 Benchtop Sodium Meter MAINTENANCE
6. MAINTENANCE
When the meter is temporary out of use, the electrode should be immersed in the basified pNa4
solution. Place the meter and beakers in an area where is dustproof for future use.
When the meter is long time out of use, the following operation should be taken:
Power-off meter; remove the electrode, put it into the protection bottle which is filled with
immersion solution; Insert the short circuit cap into the hole for electrode connector.
Weekly
Monthly
Do 2-points calibration once a month.
Usage and maintenance of the electrode refer to Electrode Instruction Manual for details.
24
HK-51 Benchtop Sodium Meter APPENDIX
Appendix I
1. Accurately weigh 1.1690g (NaCl) primary reagent (or high-grade pure) which has been baked for
1~2 hours under 250~350℃, dissolve it in 1st grade reagent water, and then shift into a 2 liter
volumetric flask and dilute to the scale, shake up.
Absorb pNa4 standard solution, by using the 1st grade reagent water and dilute to 10 times volume and
then get pNa5 standard solution.
Absorb pNa5 standard solution, by using the 1st grade reagent water and dilute to 10 times volume
and then get pNa6 standard solution.
With the same method, pNa7 (2.3μg/L) standard solution can be prepared according to actual use.
NOTE: Actually the 1st grade reagent water for preparing the standard solution is not really
contain no Na+, therefore, when the concentration is lower than 2.3μg/L(pNa7), the Na+
contained in the water should be deducted. When prepare for 2.3μg/L (pNa6) standard solution
or concentration below it, we should use plastic or quartz flask, all the standard solutions
should be stored in polythene plastic container to prevent any pollution.
25
HK-51 Benchtop Sodium Meter APPENDIX
Appendix Ⅱ
26
HK-51 Benchtop Sodium Meter APPENDIX
27
HK-51 Benchtop Sodium Meter APPENDIX
NOTE: The description above are our understanding of the product, only refer to the
characteristic of safety precautions, but not guarantee the behalf of product feature. Huakeyi
assumes no responsibility for any hazard caused by the product usage.
28
HK-51 Benchtop Sodium Meter APPENDIX
Appendix Ⅲ
29
HK-51 Benchtop Sodium Meter APPENDIX
A3-8 Stabilization
Incompatibility: acid
Don’t contain in the glassware, avoid any contact with the metal. And it is sensitive to light, heat and
alkali metal.
Burn and decomposed product. Hydrofluoric gas
NOTE: The description above are our understanding of the product, only refer to the
characteristic of safety precautions, but not guarantee the behalf of product feature. Huakeyi
assumes no responsibility for any hazard caused by the product usage.
30
HK-51 Benchtop Sodium Meter APPENDIX
Appendix Ⅴ
Meter accessory
31
HK-51 Benchtop Sodium Meter PRODUCTS & TECHNICAL SUPPORT
Laboratory Instruments
Technical Support
Essential Instruction!
On-line Instruments
32