Program P1 P2 RANGE
P3 P4 P5 RANGE
Standby
G:\Technik\Elektro\DATEN\ART\MINI_MICKEY_MOUSE\Drawings\RDM8\2SO\S2\CH\Testwerte_01_RDM8_2_SO_S2_CH.doc
Page 1/2
COLENTA Labortechnik – TK 21-01-2012
Important:
It’s very important to check all the listed values (Factory settings) after any kind of
software updating and/or loading Medical/NDT applications.
Service
Inputs
Cov= 0 = Cover closed Cover switch
1 = Cover open
Lev= 0 = closed Option level switch
1 = open
SP= 1 = Setup disabled Setup Bridge
0 = Setup enabled
(wire in)
Th= Motor pulse
Id= output for speed regulation
Vd= output for speed regulation
Outputs
H1= 1 = off Heater tanks
0 = on
Cir= 1 = off Circulation pump
0 = on
Rep= 1 = off Replenishment pump
0 = on
Col= 1 = off Cooling magnetic valve
0 = on
Wsh= 1 = off Wash magnetic valve
0 = on
Hd= 1 = off Dryer heater and fan
0 = on
Rep
Rep1 0ml OFF Manual Rep.
LED
Testing LED.Look Display led check
at little disp.
Mem
Load default val Exit Factory settings
Motor
Speed: 0 +- for Motor test
Range 0 - 60
Ph_Del: xx for motor service
T= xxx measured Motor speed
SB
Sensor: 0 +- for sensorbar test
S-bar: ______ 012345