COLENTA Labortechnik – TK 21-01-2012

Einstellwerte für INDX 900E (RDM8-2/SO/S2/CH) mit Prog. MM V 2.8a (NDT)

PROCESSOR: COLENTA INDX 900 E


with PDB Mini_05_2007_P89C668
Displayanzeige bei Initialisierung: P89c668 MM_V2.8a
Type: NDT

Program P1 P2 RANGE

Tank1 time [sec] 100 120 032 - 188


T1 [°C] 26,0 26,0 18,0 - 43,0
R1 [ml/sqm] 0700 0700 0040 - 2000

P3 P4 P5 RANGE

Tank1 time [sec] 150 150 150 032 - 188


T1 [°C] 26,0 26,0 26,0 18,0 - 43,0
R1 [ml/sqm] 0700 0700 0700 0040 - 2000

Setup [Werte für HA - Z12 MINI - NDT] (Setup Bridge neccessary)

Tank1 [pls] 0094 0010 - 1000


Machine [pls] 0560 0030 - 9999
Gear [pls/m] 499 050 - 999
Pump [ml/s] 02,0 00,5 - 99,9
Sensor distance [mm] 060 000 - 300
Replenish after each [sq.m.] 0,125 0,100 - 0,999
Ready [pls.] 0041 0000 - 1000
after film left
Add [pulses] 28 00 - 99
before start
Additional run [cm] 001 000 - 200
Motor error 132 000 - 255
Motor stop 120 000 - 255

Standby

SB replenishment [ml] 100 000 - 999


each [h] 6 0 - 9
SB self-cleaning [cm] 20 05 - 40
each [min] 10 01 - 20
Switch ON Dryer for [min] 05 00 - 20
DR On each [min] 10 00 - 50
for [min.] 01 01 - 10

Setup Bridge: PDB ST11 Pin 5+6


Jumper have to be placed on ST11 PIN 3+4 for NDT processor

G:\Technik\Elektro\DATEN\ART\MINI_MICKEY_MOUSE\Drawings\RDM8\2SO\S2\CH\Testwerte_01_RDM8_2_SO_S2_CH.doc
Page 1/2
COLENTA Labortechnik – TK 21-01-2012

Important:
It’s very important to check all the listed values (Factory settings) after any kind of
software updating and/or loading Medical/NDT applications.

Service
Inputs
Cov= 0 = Cover closed Cover switch
1 = Cover open
Lev= 0 = closed Option level switch
1 = open
SP= 1 = Setup disabled Setup Bridge
0 = Setup enabled
(wire in)
Th= Motor pulse
Id= output for speed regulation
Vd= output for speed regulation

Outputs
H1= 1 = off Heater tanks
0 = on
Cir= 1 = off Circulation pump
0 = on
Rep= 1 = off Replenishment pump
0 = on
Col= 1 = off Cooling magnetic valve
0 = on
Wsh= 1 = off Wash magnetic valve
0 = on
Hd= 1 = off Dryer heater and fan
0 = on

Rep
Rep1 0ml OFF Manual Rep.
LED
Testing LED.Look Display led check
at little disp.

Mem
Load default val Exit Factory settings

Motor
Speed: 0 +- for Motor test
Range 0 - 60
Ph_Del: xx for motor service
T= xxx measured Motor speed

SB
Sensor: 0 +- for sensorbar test
S-bar: ______ 012345

G:\Technik\Elektro\DATEN\ART\MINI_MICKEY_MOUSE\Drawings\RDM8\2SO\S2\CH\Testwerte_01_RDM8_2_SO_S2_CH.doc