Professional Documents
Culture Documents
03
50103643 PROline Prowirl 72
Valid as of software version:
V 1.01.01 (amplifier)
Vortex Flow Measuring System
Operating Instructions
Brief operating instructions PROline Prowirl 72
Installation Page 11
Wiring Page 21
You can commission your measuring device quickly and easily using the special
“Quick Setup” menu. It allows you to configure important basic functions via the
local display, for example display language, measured variables, engineering
units, signal type etc.
2 Endress+Hauser
PROline Prowirl 72 QUICK SETUP for quick commissioning
HOME-POSITION Language
Application
Operation
Mode
Automatically
parameterization
of the display
! Note!
The display returns to the QUICK SETUP COMMISSIONING cell if you press the ESC key combination
F06-72xxxxxx-19-xx-xx-en-000
during interrogation.
➀ Only the output (current output or pulse/status output) not yet configured in the current Quick Setup is
offered for selection after the first cycle.
➁ The “YES” option appears as long as a free output is still available. "NO" is the only option displayed
when no further outputs are available.
➂ When "YES" is selected, the flow is assigned to line 1 of the local display and the totalizer to line 2.
Endress+Hauser 3
QUICK SETUP for quick commissioning PROline Prowirl 72
4 Endress+Hauser
PROline Prowirl 72 Contents
Contents
Endress+Hauser 5
Contents PROline Prowirl 72
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
6 Endress+Hauser
PROline Prowirl 72 1 Safety instructions
1 Safety instructions
Resulting from incorrect use or from use other than that designated the operational
safety of the measuring devices can be suspended. The manufacturer accepts no
liability for damages being produced from this.
• The device must be operated by persons authorised and trained by the facility’s
owner-operator. Strict compliance with the instructions in these Operating Instructions
is mandatory.
• In the case of special fluids (incl. fluids for cleaning), Endress+Hauser will be happy
to assist in clarifying the material resistance properties of wetted parts.
However, the user is responsible for the choice of wetted materials as regards their
in-process resistance to corrosion. The manufacturer refuses to accept liability.
• The installer must ensure that the measuring system is correctly wired in accordance
with the wiring diagrams.
• Invariably, local regulations governing the opening and repair of electrical devices
apply.
• The manufacturer reserves the right to modify technical data without prior notice.
Your Endress+Hauser distributor will supply you with current information and
updates to these Operating Instructions.
Endress+Hauser 7
1 Safety instructions PROline Prowirl 72
1.4 Return
The following procedures must be carried out before a flowmeter requiring repair or
calibration, for example, is returned to Endress+Hauser:
• Always enclose a fully completed “Declaration of Contamination” form with the device.
!
Only then can Endress+Hauser transport, examine and repair a returned device.
Note!
A copy of the “Declaration of Contamination” can be found at the end of these Operating
Instructions.
• Enclose special handling instructions if necessary, for example a safety data sheet as
per the European Directive 91/155/EEC.
• Remove all fluid residues. Pay special attention to the grooves for seals and crevices
which could contain fluid residues.
This is particularly important if the fluid is hazardous to health, e.g. flammable, toxic,
#
caustic, carcinogenic, etc.
Warning!
• Do not return a measuring device if you are not absolutely certain that all traces of
hazardous substances have been removed, e.g. substances which have penetrated
crevices or diffused through plastic.
• Costs incurred for waste disposal and injury (caustic burns, etc.) due to inadequate
cleaning will be charged to the owner-operator.
#
these Operating Instructions by the following symbols:
Warning!
“Warning” indicates an action or procedure which, if not performed correctly, can result
in injury or a safety hazard. Comply strictly with the instructions and proceed with care.
" Caution!
“Caution” indicates an action or procedure which, if not performed correctly, can result
!
in incorrect operation or destruction of the device. Comply strictly with the instructions.
Note!
“Note” indicates an action or procedure which, if not performed correctly, can have an
indirect effect on operation or trigger an unexpected response on the part of the device.
8 Endress+Hauser
PROline Prowirl 72 2 Identification
2 Identification
In the compact version, the transmitter and sensor form a mechanical unit; in the remote
version they are mounted separate from one another.
➈ ➉
ENDRESS+HAUSER
PROWIRL 72
Order Code:72WXX-XXXXXXXXXXXX IP67/NEMA/Type 4X
➀ Ser.No.: 12345678901
TAG No.: ABCDEFGHJKLMNPQRST
➁ 12-36VDC 1.2W
➂ Version: 4...20mA, HART
A
i
-40°C<Ta<+70°C
Ta +10°C/ 50°F
-40°F<Ta<+158°F
➀ Ser.No.: 12345678901
Sensor data:
➃ PN40 / p test = 85bar
➄ K-factor: 1.0000 P/L
Materials: CF3M(1.4404) 316L(1.4435)
➅ Gasket: Graphite
B
F06-72xxxxxx-18-06-xx-xx-000
➆ TM: -40°C...+260°C / -40°F...+500°F
➇
Endress+Hauser 9
2 Identification PROline Prowirl 72
➆ ➇
ENDRESS+HAUSER
PROWIRL 72
Order Code:72WXX-XXXXXXXXXXXX IP67/NEMA/Type 4X
➀ Ser.No.: 12345678901
TAG No.: ABCDEFGHJKLMNPQRST
➁ K-factor: 1.0000 P/dm³
➂ Materials: CF3M(1.4404), 316L(1.4435)
➃
F06-72xxxxxx-18-06-xx-xx-001
Gasket: Graphite
➄ TM: -40...+260°C / -40...+500°F
➅
i
-40°C<Ta<+85°C
-40°F<Ta<+185°F
Fig. 2: Nameplate specifications for transmitter remote version “PROline Prowirl 72” (example)
1 Order code / serial number: see the specifications on the order confirmation for the meanings of the
individual letters and digits.
2 Calibration factor
3 Material sensor
4 Material gasket
5 Medium temperature range
6 Reserved for information on special products
7 Permitted ambient temperature range
8 Degree of protection
10 Endress+Hauser
PROline Prowirl 72 3 Installation
3 Installation
3.1.2 Transport
Please note the following when unpacking or transporting to the measuring point:
• The devices must be transported in the container supplied.
• Devices with nominal diameter DN 40...300 may not be lifted at the transmitter housing
or at the connection housing of the remote version when transporting (see Fig. 3). Use
carrier slings when transporting and put the slings around both process connections.
#
Avoid chains as these could damage the housing.
Warning!
Risk of injury if the measuring device slips.
The centre of gravity of the entire measuring device might be higher than the points
around which the slings are slung. Therefore, when transporting, make sure that the
device does not unintentionally turn or slip.
3.1.3 Storage
Note the following points:
• Pack the measuring device in such a way as to protect it reliably against impact for
storage (and transportation). The original packaging provides optimum protection.
• The permissible storage temperature is –40...+80 °C
(ATEX II 1/2 GD version/dust ignition-proof –20...+55°C).
• When in storage, the device should not be exposed to direct sunlight in order to avoid
impermissibly high surface temperatures.
Endress+Hauser 11
3 Installation PROline Prowirl 72
3.2.1 Dimensions
The dimensions and the lengths of the sensor and the transmitter are on Page 64 ff.
F06-7xxxxxxx-04-xx-xx-xx-002
Esc
- + E
3.2.3 Orientation
The device can generally be installed in any position in the piping.
In the case of liquids, there should be upward flow in vertical pipes to avoid partial pipe
filling (see orientation A).
In the case of hot fluids (e.g. steam or fluid temperature ≥ 200 °C), select orientation C
or D so that the permitted ambient temperature of the electronics is not exceeded.
Orientations B and D are recommended for very cold fluids (e.g. liquid nitrogen) (see
Page 13).
Orientations B, C and D are possible with horizontal installation (see Page 13).
12 Endress+Hauser
PROline Prowirl 72 3 Installation
The arrow indicated on the device must always point in the direction of flow in all
orientations.
" Caution!
• If fluid temperature is ≥ 200 °C, orientation B is not permitted for the wafer version
(Prowirl 72 W) with a nominal diameter of DN 100 and DN 150.
• In case of vertical orientation and downward flowing liquid, the piping has always to
be completely filled.
Esc
- + E
A B
Esc
- + E
C D
F06-7xxxxxxx-04-xx-xx-xx-002
Esc
- + E
When insulating, please ensure that a sufficiently large area of the housing support is
exposed. The uncovered part serves as a radiator and protects the electronics from
overheating (or undercooling).
The maximum insulation height permitted is illustrated in the diagrams. These apply
equally to both the compact version and the sensor in the remote version.
1 2 -
Esc
+ E
Esc
- + E F06-7xxxxxxx-16-00-00-xx-001
" Caution!
Danger of electronics overheating!
• Therefore, make sure that the adapter between sensor and transmitter and the
connection housing of the remote version is always exposed.
• Note that a certain orientation might be required, depending on the fluid temperature
→ Page 12.
• Information on permissible temperature ranges → Page 60.
Endress+Hauser 13
3 Installation PROline Prowirl 72
1 15 x DN 5 x DN 2 18 x DN 5 x DN
A B A B
Esc Esc
- + E - + E
4
20 x DN 5 x DN 40 x DN 5 x DN
3
A B A B
Esc Esc
- + E - + E
5 6
25 x DN 5 x DN 50 x DN 5 x DN
A B A B
Esc Esc
- + E - + E
F06-7xxxxxxx-04-xx-xx-xx-000
Fig. 7: Minimum inlet and outlet runs with various flow obstructions
A = Inlet run
B = Outlet run
1 = Reduction
2 = Extension
3 = 90° elbow or T-piece
4 = 2 x 90° elbow, 3-dimensional
5 = 2 x 90° elbow
!
6 = Control valve
Note!
A specially designed perforated plate flow conditioner can be installed if it is not
possible to observe the inlet runs required (see Page 15).
PT TT
F06-7xxxxxxx-04-xx-xx-xx-003
-
Esc
+ E
3...5 x DN
4...8 x DN
Fig. 8: Installation of pressure measuring point (PT) and temperature measuring point (TT)
14 Endress+Hauser
PROline Prowirl 72 3 Installation
F06-7xxxxxxx-04-xx-xx-xx-00
2 x DN 8 x DN 5 x DN
Esc
- + E
3.2.6 Vibrations
The correct operation of the measuring system is not influenced by plant vibrations up
to 1 g, 10...500 Hz. Consequently, the sensors require no special measures for
attachment.
Endress+Hauser 15
3 Installation PROline Prowirl 72
" Caution!
Please note the following prior to mounting:
• Prior to installing the measuring device in the piping, remove all traces of transport
packaging and any protective covers from the sensor.
• Make sure that the internal diameters of seals are the same as, or greater than, those
of the measuring pipe and piping. Seals projecting into the flow current have a
negative effect on the vortex formation after the bluff body and cause inaccurate
measurement. For this reason, the seals supplied by Endress+Hauser have a slightly
larger internal diameter than the measuring pipe.
• Ensure that the arrow on the measuring pipe matches the direction of flow in the
piping.
• Lengths:
– Prowirl W (wafer version): 65 mm
– Prowirl F (flanged version) → Page 66 ff.
Mounting Prowirl W
The centering rings supplied are used to mount and center the wafer-style devices.
A mounting kit consisting of tie rods, seals, nuts and washers can be ordered
separately.
4
5
2
3
xxx-17-00-06-xx-000
16 Endress+Hauser
PROline Prowirl 72 3 Installation
! Note!
There are recesses in the rotating groove at 90° stages (only compact version).
These help you align the transmitter easier.
180°
F06-7xxxxxxx-17-00-xx-xx-001
180°
Endress+Hauser 17
3 Installation PROline Prowirl 72
The transmitter and the sensor must be mounted separate in the following
circumstances:
• Poor accessibility
• Lack of space
• Extreme ambient temperatures
" Caution!
If the device is mounted to warm piping, make certain that the housing temperature
does not exceed the max. permissible value of +80 °C.
Mount the transmitter as illustrated in the diagram.
A
232 (*226)
Esc
- + E
B
227 (*221)
Esc
- + E
F06-10xxxxxx-17-06-xx-xx-001
18 Endress+Hauser
PROline Prowirl 72 3 Installation
3. Turn the display to the desired position (max. 4 x 45° in each direction) and reset it
onto the retaining rails.
4. Screw the cover of the electronics compartment firmly back onto the transmitter
housing.
4 x 45°
F06-72xxxxxx-07-xx-06-xx-000
Es
c
–
+
E
Installation Notes
Does the arrow on the pipe stand resp. on the meter body match the −
direction of flow through the pipe?
Is the orientation chosen for the sensor correct, in other words suitable See Page 12 ff.
for sensor type, fluid properties (outgassing, with entrained solids)
and fluid temperature?
Endress+Hauser 19
3 Installation PROline Prowirl 72
20 Endress+Hauser
PROline Prowirl 72 4 Wiring
4 Wiring
# Warning!
When connecting Ex-certified devices, please refer to the notes and diagrams in the Ex-
specific supplement to these Operating Instructions. Please do not hesitate to contact
your Endress+Hauser representative if you have any questions.
! Note!
• The remote version must be grounded. In doing so, the sensor and transmitter must
be connected to the same potential matching.
• When using the remote version, always make sure that you connect the sensor only to
the transmitter with the same serial number. Compatibility errors (e.g. the incorrect
K-factor will be used) can occur if the devices are not connected in this way.
1. Remove the cover of the connection compartment of the transmitter (a).
2. Remove the cover of the connection compartment of the sensor (b).
3. Feed the connecting cable (c) through the appropriate cable entries.
4. Wire the connecting cable between the sensor and transmitter in accordance with
the electrical connection diagram:
→ Fig. 14
→ Wiring diagram in the screw caps
5. Tighten the glands of the cable entries on the sensor housing and transmitter
housing.
6. Screw the cover of the connection compartment (a/b) back onto the sensor housing
or transmitter housing.
GROUND
TEMP 1
TEMP 2
TEMP 3
DIFF +
+ 5 VA
DIFF –
– 5 VA
Esc
- + E
1 2 3 4 5 6 7 8
GN
WT
BN
GY
BU
RD
PK
YL
a
e
b
c
f
d
GN
WT
BN
GY
BU
RD
PK
YL
1 2 3 4 5 6 7 8
F06-72xxxxxx-04-xx-xx-xx-000
TEMP 1
DIFF +
GROUND
TEMP 3
TEMP 2
+ 5 VA
– 5 VA
DIFF –
Endress+Hauser 21
4 Wiring PROline Prowirl 72
! Note!
• When connecting Ex-certified devices, please refer to the notes and diagrams in the
Ex-specific supplement to these Operating Instructions.
• The remote version must be grounded. In doing so, the sensor and transmitter must
be connected to the same potential matching.
• When connecting the transmitter, use a connecting cable with a continuous service
temperature of at least –40...(permitted max. ambient temperature +10 °C).
Procedure for connecting the transmitter, Non-Ex/ Ex-i version (see → Fig. 15)
1. Unscrew the cover (a) of the electronics compartment from the transmitter housing.
2. Remove the display module (b) from the retaining rails (c) and refit onto right
retaining rail with the left side (this secures the display module).
3. Loosen screw (d) of the cover of the connection compartment and fold down the
cover.
4. Push the cable for the power supply/current output through the cable gland (e).
Optional: push the cable for the pulse output through the cable gland (f).
6. Pull the terminal connector (g) out of the transmitter housing and connect the cable
for the power supply/current output (see → Fig. 17).
Optional: Pull terminal connector (h) out of the transmitter housing and connect the
cable for the pulse output (see → Fig. 17).
! Note!
The terminal connectors (g / h) are pluggable, i.e. they can be plugged out of the
transmitter housing to connect the cables.
22 Endress+Hauser
PROline Prowirl 72 4 Wiring
! Note!
The connectors are coded so you cannot mix them up.
8. Only remote version: Secure the ground cable to the ground terminal
(see Fig. 17, c).
9. Fold up the cover of the connection compartment and tighten the screws (d).
10. Remove the display module (b) and fit on the retaining rails (c).
11. Screw the cover of the electronics compartment (a) onto the transmitter housing.
d e
g h
c f
a
d
F06-73xxxxxx-04-06-00-xx-001
Fig. 15: Procedure for connecting the transmitter Non-Ex / Ex-i version
a Cover of electronics compartment
b Retaining rail for display module
c Display module
d Connection compartment cover threaded connection
e Cable gland for power supply/current output cable
f Cable gland for pulse output cable (optional)
g Terminal connector for power supply/current output
h Terminal connector for pulse output (optional)
Procedure for connecting the transmitter, Ex-d version (see → Fig. 16)
! Note!
When connecting Ex-certified devices, please refer to the notes and diagrams in the
Ex-specific supplement to these Operating Instructions.
1. Open the clamp (a) securing the cover of the connection compartment.
2. Unscrew the cover (b) of the connection compartment from the transmitter housing.
3. Push the cable for the power supply/current output through the cable gland (e).
Optional: push the cable for the pulse output through the cable gland (f).
5. Pull the terminal connector (g) out of the transmitter housing and connect the cable
for the power supply/current output (see → Fig. 17).
Optional: Pull terminal connector (h) out of the transmitter housing and connect the
cable for the pulse output (see → Fig. 17).
! Note!
The terminal connectors (g / h) are pluggable, i.e. they can be plugged out of the
transmitter housing to connect the cables.
Endress+Hauser 23
4 Wiring PROline Prowirl 72
! Note!
The connectors are coded so you cannot mix them up.
7. Only remote version: Secure the ground cable to the ground terminal (Fig. 17, c).
8. Screw the cover (b) of the connection compartment onto the transmitter housing.
9. Engage the clamp (a) to hold the cover of the connection compartment (b) in
position and tighten the threaded fastener of the clamp.
d
c
f
e
a
b
F06-73xxxxxx-04-06-00-xx-001
Wiring diagram
B D
F06-72xxxxxx-11-00-00-xx-000
+ - + - + - + -
1 2 3 4 1 2 3 4
C C
Pulse/status output:
Open collector, passive, galvanically isolated, Umax = 30 V, with 15 mA current limiting, Ri = 500 Ω,
can be configured as pulse or status output
24 Endress+Hauser
PROline Prowirl 72 4 Wiring
! Note!
• The measuring circuit's minimum load must be at least 250 Ω .
• After commissioning, make the following setting:
– Switch HART write protection on or off (see Page 38)
• For connecting, please refer also to the documentation issued by the HART
Communication Foundation, in particular HCF LIT 20: “HART, a technical summary”.
250
b
+ - + -
1 2 3 4
F06-7xxxxxxx-04-xx-xx-xx-005
6
9 Page
Up
Bksp Delete
Page
On
a
PQRS TUV W XY Z
7 8 9
,()‘ _<> +*/
. 0 -
375
FIELD COMMUNICATOR
250
b
+ - + -
1 2 3 4
F06-7xxxxxxx-04-xx-xx-xx-006
c
RS 232
Endress+Hauser 25
4 Wiring PROline Prowirl 72
F-xxxxxxxx06-04-xx-xx-xx-005
Fig. 20: Installation instructions for cable entries
Do the cables used comply with the specifications? See Page 22, 59
Are the cables for power supply/current output, pulse output/status See Page 22
output (optional) and grounding connected correctly?
Are all the cable entries installed, tightened and sealed? See Page 26
Cable run with “water trap”?
26 Endress+Hauser
PROline Prowirl 72 5 Operation
5 Operation
The display consists of two lines; this is where measured values and/or status variables
(e.g. bar graph) are displayed. You can change the assignment of the display lines to
different variables to suit your needs and preferences (→ see USER INTERFACE
function group on Page 84).
1
+48.25 m 3 /h
+3702.6 m 3
Esc
- + E
F06-72xxxxxx-07-xx-xx-xx-000
2 3
Endress+Hauser 27
5 Operation PROline Prowirl 72
! Note!
• Please refer to the general notes on Page 29.
• Function matrix overview → Page 75
• Detailed description of all functions → Page 76 ff.
The function matrix is a two-level construct: the function groups form one level and the
groups’ functions the other. The groups are the highest-level grouping of the control
options for the measuring device. A number of functions is assigned to each group.
You select a group in order to access the individual functions for operating and
configuring the measuring device.
1. HOME position → F
→ enter the function matrix
2. Select a function group (e.g. CURRENT OUTPUT)
3. Select a function (e.g. TIME CONSTANT)
Change parameter / enter numerical values:
OS → select or enter: release code, parameters, numerical values
F → save your entries
Esc
- + E
➃
➀ –
Esc
+ –
Esc
+
>3s
E
➂
E E E E E
Esc
– +
➁
+
F06-x0xxxxxx-19-xx-xx-xx-000
28 Endress+Hauser
PROline Prowirl 72 5 Operation
!
seconds following return to the HOME position.
Note!
• The transmitter continues to measure while data entry is in progress, i.e. the current
measured values are output via the signal outputs in the normal way.
• If the power supply fails, all preset and configured values remain safely stored in the
EEPROM.
" Caution!
All functions are described in detail, as is the function matrix itself on Page 75 ff.
Endress+Hauser 29
5 Operation PROline Prowirl 72
1 P XXXXXXXXXX
F06-x0xxxxxx-07-xx-xx-xx-000’
#000 00:00:05
2 4 5 3
Serious system errors, e.g. electronic module defects, are always categorised and
displayed as “fault messages” by the measuring device.
Fault message ( ) $
$
• Displayed as → lightning flash( ), error designation (S: system error, P: process
error)
• The error in question has a direct effect on the inputs or outputs.
The response of the inputs/outputs (failsafe mode) can be defined by means of
!
functions in the function matrix (see Page 50).
Note!
Error messages can be output via the current output in accordance with NAMUR NE 43.
30 Endress+Hauser
PROline Prowirl 72 5 Operation
The HART protocol allows the transfer of measuring and device data between the HART
master and the field devices for configuration and diagnostics purposes. HART
masters, such as a handheld terminal or PC-based operating programs (such as
FieldTool), require device description (DD) files. They are used to access all the
information in a HART device. Such information is transferred solely via “commands”.
There are three different command classes:
• Universal commands:
All HART devices support and use universal commands. The following functionalities
are linked to them:
– Recognising HART devices
– Reading off digital measured values (flow, totalizer, etc.)
• Common practice commands:
Common practice commands offer functions which are supported and can be
executed by many but not all field devices.
• Device-specific commands:
These commands allow access to device-specific functions which are not HART
standard. Such commands access individual field device information, (among other
!
things), such as low flow cut off settings etc.
Note!
Prowirl 72 has all three command classes. Page 32 ff. provides you with a list of all the
supported “Universal commands” and “Common practice commands”.
Endress+Hauser 31
5 Operation PROline Prowirl 72
Device variables:
The following device variables are available via the HART protocol:
1 Flow
250 Totalizer
Process variables:
At the factory, the process variables are assigned to the following device variables:
• Primary process variable (PV) → flow
• Secondary process variable (SV) → totalizer
• Third process variable (TV) → not assigned
• Fourth process variable (FV) → not assigned
Universal commands
0 Read the unique device None The device identifier provides information on the device
identifier and manufacturer; it cannot be altered.
1 Read the primary process None – Byte 0: HART unit ID of the primary process variable
variable – Byte 1-4: primary process variable
!
Primary process variable = flow
Access type = Read
Note!
Manufacturer-specific units are represented using the
HART unit ID “240”.
2 Read the primary process None – Byte 0-3: current current of the primary process variable
variable as current in mA in mA
and percentage of the set – Byte 4-7: percentage of the set measuring range
measuring range
Primary process variable = flow
Access type = Read
32 Endress+Hauser
PROline Prowirl 72 5 Operation
Factory setting:
• Primary process variable = flow
• Secondary process variable = totalizer
• Third process variable = not assigned
• Fourth process variable = not assigned
! Note!
Manufacturer-specific units are represented using the
HART unit ID “240”.
6 Set HART short-form Byte 0: desired address (0...15) Byte 0: active address
address
Factory setting:
Access type = Write 0
! Note!
With an address > 0 (multidrop mode),
the current output of the primary process
variable is fixed to 4 mA.
11 Read the unique device Byte 0-5: TAG The device identifier provides information on the device
identifier using the TAG and manufacturer; it cannot be altered.
The response consists of a 12-byte device ID if the given
Access type = Read TAG matches the one saved in the device:
Endress+Hauser 33
5 Operation PROline Prowirl 72
14 Read sensor information None – Byte 0-2: serial number of the sensor
on the primary process – Byte 3: HART unit ID of the sensor limits and measuring
variable range of the primary process variable
– Byte 4-7: upper sensor limit
Access type = Read – Byte 8-11: lower sensor limit
– Byte 12-15: minimum span
! Note!
• The data relate to the primary process variable (= flow).
• Manufacturer-specific units are represented using the
HART unit ID “240”.
! Note!
Manufacturer-specific units are represented using the
HART unit ID “240”.
17 Write user message You can save any 32-character long text Displays the current user message in the device:
in the device with this parameter: Byte 0-23:
Access = Write current user message in the device
Byte 0-23:
desired user message
18 Write TAG, TAG You can save an Displays the current information in the device:
description and date 8-character TAG, a 16-character TAG
description and a date with this – Byte 0-5: TAG
Access = Write parameter: – Byte 6-17: TAG description
– Byte 18-20: date
– Byte 0-5:
TAG
– Byte 6-17:
TAG description
– Byte 18-20:
date
34 Endress+Hauser
PROline Prowirl 72 5 Operation
34 Write attenuation constant Byte 0-3: attenuation constant of the Displays the current attenuation constant in the device:
for primary process primary process variable in seconds Byte 0-3: attenuation constant in seconds
variable
Factory setting:
Access = Write Primary process variable = flow
35 Write measuring range of Write the desired measuring range: The measuring range currently set is shown as the
the primary process – Byte 0: HART unit ID for the primary response:
variable process variable
– Byte 1-4: end of measuring range, – Byte 0:
Access = Write value for 20 mA HART unit ID for the set measuring range of the primary
– Byte 5-8: start of measuring range, process variable
value for 4 mA
– Byte 1-4:
Factory setting: end of measuring range, value for 20 mA
Primary process variable = volume flow
! Note!
– Byte 5-8:
start of measuring range, value for 4 mA (is always at “0”)
If the HART unit ID does not suit the
process variable, the device will continue
with the last valid unit.
! Note!
Manufacturer-specific units are represented using the
HART unit ID “240”.
Access = Write
40 Simulate output current of Simulation of the desired output current of The current output current of the primary process variable
the primary process the primary process variable. An entry is displayed as a response:
variable value of 0 exits the simulation mode: Byte 0-3: output current in mA
Factory setting:
Primary process variable = flow
Access = Write
44 Write unit of the primary Specify the unit of the primary process The current unit code of the primary process variable is
process variable variable. Only units which are suitable for displayed as a response:
the process variable are accepted by the Byte 0: HART unit ID
Access = Write device:
! Note!
• If the written HART unit ID does not suit
the process variable, the device will
continue with the last valid unit.
• If you change the unit of the primary
process variable, this has an impact on
the 4...20 mA output.
Endress+Hauser 35
5 Operation PROline Prowirl 72
48 Read extended device None The current device status is displayed in extended form as
status the response:
50 Read assignment of the None Display of the current variable assignment of the process
device variables to the variables:
four process variables
– Byte 0:
Access = Read device variable ID to the primary process variable
– Byte 1:
device variable ID to the secondary process variable
– Byte 2:
device variable ID to the third process variable
– Byte 3:
device variable ID to the fourth process variable
Factory setting:
• Primary process variable: ID 1 for flow
• Secondary process variable: ID 250 for totalizer
• Third process variable: ID 0 for OFF (not assigned)
• Fourth process variable: ID 0 for OFF (not assigned)
53 Write device variable unit This command sets the unit of the given The current unit of the device variables is displayed in the
device variables. Only those units which device as a response:
Access = Write suit the device variable are transferred:
– Byte 0: device variable ID
– Byte 0: device variable ID – Byte 1: HART unit ID
– Byte 1: HART unit ID
! Note!
If the written unit does not suit the device
variable, the device will continue with the
last valid unit.
59 Specify number of This parameter specifies the number of As a response, the current number of the preambles is
preambles in message preambles which are inserted in the displayed in the response message:
responses message responses:
Byte 0:
Access = Write Byte 0: Number of preambles
Number of preambles (2...20)
109 Burst mode control This parameter switches the burst mode The value set in byte 0 is shown as the response.
on and off.
Access = Write
Byte 0:
0 = burst mode off
1 = burst mode on
36 Endress+Hauser
PROline Prowirl 72 5 Operation
!
command “48”. The command delivers bit-encoded information (see table below).
Note!
Detailed information on the device status messages and error messages, and how they
are rectified, can be found on Page 46 ff.!
Endress+Hauser 37
5 Operation PROline Prowirl 72
1. Unscrew the cover of the electronics compartment from the transmitter housing.
2. Remove the display module (a) from the retaining rails (b) and refit onto right
retaining rail with the left side (this secures the display module).
! Note!
The current status of the HART write protection is displayed in the WRITE
PROTECTION function (see Page 98).
Esc
– + E
F06-72xxxxxx-03-00-00-xx-000
B
A
38 Endress+Hauser
PROline Prowirl 72 6 Commissioning
6 Commissioning
6.2 Commissioning
!
values and/or status variables appear on the display (HOME position).
Note!
If start-up fails, an appropriate error message is displayed, depending on the cause.
Endress+Hauser 39
6 Commissioning PROline Prowirl 72
You will find a flowchart of the “Commissioning” Quick Setup menu on Page 41 and the
function description on Page 81.
40 Endress+Hauser
PROline Prowirl 72 6 Commissioning
++ Quick Setup
E Quick Setup E+
-
Esc
+ E
Commission
HOME-POSITION Language
Application
Operation
Mode
Automatically
parameterization
of the display
! Note!
The display returns to the QUICK SETUP COMMISSIONING cell if you press the ESC
key combination X
during interrogation.
F06-72xxxxxx-19-xx-xx-en-000
➀ Only the output (current output or pulse/status output) not yet configured in the
current Quick Setup is offered for selection after the first cycle.
➁ The “YES” option appears as long as a free output is still available. “NO” is the only
option displayed when no further outputs are available.
➂ When “YES” is selected, the flow is assigned to line 1 of the local display and the
totalizer to line 2.
Endress+Hauser 41
7 Maintenance PROline Prowirl 72
7 Maintenance
Exterior cleaning
When cleaning the exterior of measuring devices, always use cleaning agents that do
not attack the surface of the housing and the seals.
!
are incompatible with the seal material.
Note!
• The time span between the individual replacements depends on the fluid properties.
• Replacement seals (accessory) → Page 43.
Only Endress+Hauser sensor seals may be used.
!
The seals must be dried, cleaned or replaced if necessary.
Note!
If the device is used in a dust atmosphere, only the associated Endress+Hauser
housing seals can be used.
42 Endress+Hauser
PROline Prowirl 72 8 Accessories
8 Accessories
Mounting kit for Mounting kit for remote version, suitable for pipe and DK5WM − B
transmitter wall mounting.
HART Field Handheld terminal for remote configuration and for DXR275 − * * * * * *
Communicator DXR 275 obtaining measured values via the current output
HART (4...20 mA).
HART Field Handheld terminal for remote configuration and for DXR375 − * * * * * * *
Communicator DXR 375 obtaining measured values via the current output
HART (4...20 mA) and FOUNDATION Fieldbus (FF).
Endress+Hauser 43
8 Accessories PROline Prowirl 72
44 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
9 Trouble-shooting
No display visible but output 1. Check whether the ribbon-cable connector of the display module is
signals are present correctly plugged into the amplifier board → Page 52
2. Display module defective → order spare part → Page 51
3. Electronics defective → order spare part → Page 51
Display texts are in a foreign Switch off power supply. Press and hold down both the +/– keys and
language. switch on the measuring device again. The display text will appear in
English and is displayed at 50% contrast.
Measured value indicated, Electronics board defective → order spare part → Page 51
but no signal output at the
current or pulse output
" Caution!
• Please refer also to the information on Page 30 ff.!
• The measuring system interprets simulations and positive zero return as system errors, but displays
them as notice messages only.
Endress+Hauser 45
9 Trouble-shooting PROline Prowirl 72
" Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. In such cases, the procedures on Page 8 must be carried out before you
return the measuring device to Endress+Hauser.
Always enclose a fully completed “Declaration of Contamination” form with the device.
A copy of the form can be found at the end of these Operating Instructions.
$
S CRITICAL Serious device error Replace the amplifier board.
FAIL. # 001 Spare parts → Page 51
$
S AMP HW EEPROM Amplifier: Replace the amplifier board.
# 011 Faulty EEPROM Spare parts → Page 51
$
S AMP SW EEPROM Amplifier: Contact your E+H service organisation.
# 012 Error when accessing data of
the EEPROM
$
S COM HW EEPROM COM module: Replace COM module.
# 021 Faulty EEPROM Spare parts → Page 51
$
S COM SW EEPROM COM module: Contact your E+H service organisation.
# 022 Error when accessing data of
the EEPROM
$
S CHECKSUM TOT. Totalizer checksum error Replace the amplifier board.
# 111 Spare parts → Page 51
3. Reduce flow.
46 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
$
S RESONANCE DSC Device being operated in its Reduce the flow.
# 379 resonance frequency.
" Caution!
If the device is operated in its
resonance frequency, this can
result in damage which can
lead to complete device failure.
$
S DSC SENS DEFCT The DSC sensor is defective, Contact your E+H service organisation.
# 394 measurement no longer takes
place.
S DSC SENS LIMIT The DSC sensor is being If this message is permanently displayed,
! # 395 operated near application contact your E+H service organisation.
limits, device failure is
probable soon.
$
S SIGNAL>LOW PASS The device finds the signal • Check whether the device is correctly
# 396 outside the set filter range. mounted in flow direction.
• Check whether the right choice was
Possible causes: made in the APPLICATION function.
• The flow is outside the • Check whether the operating conditions
measuring range. are within the specifications of the
• The signal is caused by a device (e.g. flow is above measuring
strong vibration which is range, i.e. the flow may have to be
intentionally not measured reduced)
and outside the measuring
range. If the checks do not solve the problem,
please contact your E+H service
organisation.
$
S PREAMP. DISCONN. Pre-amplifier disconnected. Check whether the connection between
# 399 the preamplifier and amplifier board is
established and correct and establish
connection if necessary.
S SW.-UPDATE AKT. Loading a new amplifier Wait until the procedure is complete and
! # 501 software version or data into then restart the device.
the device.
No other commands possible
at this point.
S UP./DOWNLOAD Uploading the device data. Wait until the procedure is complete.
! AKT. No other commands possible
# 502 at this point.
S POS. ZERO-RET. Positive zero return active. Switch off positive zero return.
! # 601
" Caution!
This message has the highest
display priority.
S SIM. CURR. OUT Current output simulation Switch off simulation.
! # 611 active
S SIM. FREQ. OUT Pulse output simulation active Switch off simulation.
! # 631
S SIM. STAT. OUT Status output simulation active Switch off simulation.
! # 641
S SIM. FAILSAFE Simulation of failsafe mode Switch off simulation.
! # 691 (outputs) active
S SIM. MEASURAND Simulation of a measured Switch off simulation.
! # 692 variable active (e.g. mass flow)
S CURR. ADJUST Current adjustment is active. Quit current adjustment.
! # 698
S CURR. ADJUST Current adjustment is active. Quit current adjustment.
! # 699
Endress+Hauser 47
9 Trouble-shooting PROline Prowirl 72
Remark:
You may have to change or correct settings in certain functions of the function matrix in order to rectify
faults. The functions outlined below, such as FLOW DAMPING etc. are described in detail in the section
»Description of device functions« on Page 75 ff.
Flow signal even though Check whether the device is exposed to particularly strong vibrations. If
there is no flow so, a flow can be displayed even if the fluid is at a standstill, depending
on the frequency and direction of the vibration.
Faulty or highly-fluctuating • The fluid is not adequately single-phase or homogeneous. The piping
flow signal must always be completely filled and the fluid must be single-phase
and homogeneous for accurate and reliable flow measurement.
• In many instances, the following measures can be taken to improve the
measurement result even under non-ideal conditions:
– For liquids with a low gas content in horizontal pipework, it helps to
install the device with the head pointing downwards or to the side.
This improves the measuring signal since the sensor is not in the
area where gas accumulates when this type of installation is used.
– For liquids with a low solids content, avoid installing the device with
the electronics housing pointing downwards.
– For steam or gases with a low liquid content, avoid installing the
device with the electronics housing pointing downwards.
• The inlet and outlet runs must be present as per the installation
instructions (see Page 14).
• Suitable seals with an internal diameter not smaller than the pipe
internal diameter must be installed and correctly centered.
• The static pressure must be large enough to rule out cavitation in the
area of the sensor.
48 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
Faulty or highly-fluctuating • Check whether the correct fluid was selected in the APPLICATION
flow signal (contd.) function (see Page 99). The setting in this function determines the filter
settings and can thus have an effect on the measuring range.
• Check whether the data for the K-factor on the nameplate match the
data in the K-FACTOR function. (see Page 104).
• Check whether the device is correctly mounted in flow direction.
• Check whether inner diameter of the mating pipe accord with the inner
diameter of the flowmeter (see Page 101).
• The flow must be in the measuring range of the device (see Page 57).
The start of measuring range depends on the density and the viscosity
of the fluid. Density and viscosity depend on temperature. Density also
depends on the process pressure in the case of gases.
• Check whether the operating pressure is affected by pressure
pulsations (e.g. from piston pumps). The pulsations can affect vortex
shedding if they have a frequency similar to the vortex frequency.
• Check whether the correct engineering unit was selected for the flow or
totalizer.
• Check whether the current output or pulse value was correctly set.
The fault cannot be rectified The following options are available for tackling problems of this nature:
or some other fault not
described above has Request the services of an E+H service technician
occurred. If you contact our service organisation to have a service technician sent
In these instances, please out, please be ready with the following information:
contact your E+H service – A brief description of the error with information on the application.
organisation. – Nameplate specifications (Page 9 ff.): order code and serial number
Endress+Hauser 49
9 Trouble-shooting PROline Prowirl 72
! Note!
The failsafe mode of totalizers and current, pulse and frequency outputs can be
configured by means of various functions in the function matrix.
" Caution!
System or process errors defined as “notice messages” have no effect whatsoever on the inputs and
outputs. Please refer also to the information on Page 30 ff.
MAX. CURRENT
22.6 mA
HOLD VALUE
Measured value output is based on the last measured
value saved before the error occurred.
ACTUAL VALUE
Measured value output is based on the current flow
measurement. The fault is ignored.
HOLD VALUE
Measured value output is based on the last valid flow
data before the error occurred.
ACTUAL VALUE
Measured value output is based on the current flow
measurement. The fault is ignored.
Status In the event of a fault or power supply failure: No effect on the status output
output Status output → not conductive
HOLD VALUE
The totalizer continues to count the flow on the basis of
the last valid flow data (before the fault occurred).
ACTUAL VALUE
The totalizer continues to count the flow on the basis of
the current flow data. The fault is ignored.
50 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
!
The illustration below shows the available scope of spare parts.
Note!
You can order spare parts directly from your E+H service organisation by quoting the
serial number printed on the transmitter nameplate (see Page 9).
4
5
2
F06-7xxxxxxx-03-06-06-xx-000
6
Endress+Hauser 51
9 Trouble-shooting PROline Prowirl 72
! Note!
• When connecting Ex-certified devices, please refer to the notes and diagrams in the
Ex-specific supplement to these Operating Instructions.
• Risk of damaging electronic components (ESD protection).
Static electricity can damage electronic components or impair their operability. Use a
workplace with a grounded working surface, purpose-built for electrostatically
sensitive devices!
" Caution!
Use only genuine Endress+Hauser parts.
52 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
d
e
a b
f
g
e h
i
g j
k
q
l n s
o
r
m n
l
F06-72xxxxxx-17-xx-xx-xx-001
Fig. 26: Installing and removing electronics boards Non-Ex, Ex-i version
a Cover of electronics compartment
b Local display module
c Retaining rails of local display module
d Fixing screws for cover of connection compartment
e Connection compartment cover
f Terminal connector
g Plastic cover
h Signal cable connector
i Retainer for signal cable connector
j Display module ribbon-cable connector
k Retainer for ribbon-cable connector
l Board holder threaded connection
m Board holder
n Board holder latches
o Board body
p I/O board (COM module) threaded connection
q I/O board (COM module)
r Amplifier board threaded connection
s Amplifier board
Endress+Hauser 53
9 Trouble-shooting PROline Prowirl 72
! Note!
• When connecting Ex-certified devices, please refer to the notes and diagrams in the
Ex-specific supplement to these Operating Instructions.
• Risk of damaging electronic components (ESD protection).
Static electricity can damage electronic components or impair their operability. Use a
workplace with a grounded working surface, purpose-built for electrostatically
sensitive devices!
" Caution!
Use only genuine Endress+Hauser parts.
54 Endress+Hauser
PROline Prowirl 72 9 Trouble-shooting
c
c f
a b e
d
k
i
h
p q
m r
l t
n
s
q
o
n F06-73xxxxxx-17-xx-xx-xx-000
Endress+Hauser 55
9 Trouble-shooting PROline Prowirl 72
Amplifier
! Note!
Upload/download between different software versions is normally only possible with
special service software.
56 Endress+Hauser
PROline Prowirl 72 10 Technical data
10 Technical data
10.1.1 Application
The measuring system is used to measure the volume flow of saturated steam,
overheated steam, gases and liquids. If the process pressure and process temperature
are constant, the measuring device can also output the flow as the calculated mass flow
and corrected volume flow.
Measuring principle Vortex flow measurement on the principle of the Karman vortex street.
• Remote version:
Sensor is mounted separate from the transmitter.
10.1.3 Input
Measured variable Volumetric flow (volume flow) is proportional to the frequency of vortex shedding after
the bluff body.
The output variables are volume flow or, if the process conditions are non-varying,
calculated mass flow or corrected volume flow.
Measuring range The measuring range depends on the fluid and the pipe diameter.
6 7
DN 15...25
v min. = [m/s] DN 40...300
v min. = [m/s]
[kg/m³] [kg/m³]
F06-72xxxxxx-19-xx-06-xx-002
Endress+Hauser 57
10 Technical data PROline Prowirl 72
! Note!
By using the selection and planning program “Applicator”, you can determine the exact
values for the fluid you use. You can obtain the Applicator from your Endress+Hauser
sales centre or on the Internet under www.endress.com.
K-factor range
The table is a guide. The range in which the K-factor can be is indicated for individual
nominal diameters and designs.
Nominal diameter K-factor range [pul./dm³]
DIN ANSI 72 F 72 W
DN 15 ½" 390...450 245...280
DN 25 1" 70...85 48...55
DN 40 1½" 18...22 14...17
DN 50 2" 8...11 6...8
DN 80 3" 2.5...3.2 1.9...2.4
DN 100 4" 1.1...1.4 0.9...1.1
DN 150 6" 0.3...0.4 0.27...0.32
DN 200 8" 0.1266...0.1400 –
DN 250 10" 0.0677...0.0748 –
DN 300 12" 0.0364...0.0402 –
10.1.4 Output
Pulse/status output:
Open collector, passive, galvanically isolated
• Non-Ex, EEx d - version:
Umax = 36 V, with 15 mA current limiting, Ri = 500 Ω
• EEx i - version:
Umax = 30 V, with 15 mA current limiting, Ri = 500 Ω
• Status output: Can be configured for error messages or flow limit values
Signal on alarm • Current output: Failsafe mode can be selected (e.g. in accordance with NAMUR
Recommendation NE 43)
• Pulse output: Failsafe mode can be selected
• Status output: “Not conductive“ in event of fault
58 Endress+Hauser
PROline Prowirl 72 10 Technical data
Load
R B [] R B [] R B []
1100 1100 1100
1000 1000 1000
900 900 900
800 800 800
700
Non-Ex 700
Ex d Ex i
700
600 600 600
500 500 500
400 400 400
300 300 300
200 200 200
100 100 100
0 0 0 U S [V]
10 15 20 25 30 36 10 15 20 25 30 36 10 15 20 25 30
18 21 18
F06-73xxxxxx-05-xx-xx-en-000
The area marked in grey indicates the permissible load (with HART: min. 250 Ω)
The load is calculated as follows:
Low flow cut off Switch points for low flow cut off can be selected as required
Galvanic isolation The electrical connections are galvanically isolated from one another.
Power supply failure • Totalizer stops at the last value determined (can be configured).
• All settings are kept in the EEPROM.
• Error messages (incl. value of operated hours counter) are stored.
Endress+Hauser 59
10 Technical data PROline Prowirl 72
• Gas/steam:
< 1% o.r. for Re > 20,000
< 1% o.f.s for Re between 4,000...20,000
o.r. = Of reading
o.f.s = Of full scale
Re = Reynolds number
Installation
Environment
• Remote version:
Sensor –40...+85 °C
(ATEX II 1/2 GD version/dust ignition-proof: –20...+55°C)
Transmitter –40...+80 °C
(EEx-d version: –40...+60°C; ATEX II 1/2 GD version/dust ignition-proof: –20...+55°C)
Display can be read between –20 °C...+70 °C
When mounting outside, we recommend you protect from direct sunlight with a
protective cover (order number 543199), especially in warmer climates with high
ambient temperatures.
60 Endress+Hauser
PROline Prowirl 72 10 Technical data
Process
• Seals:
Grafoil (graphite) –200...+400 °C
Viton –15...+175 °C
Kalrez –20...+275 °C
Gylon (PTFE) –200...+260 °C
bar
PN 10...40 bar
PN 64...160
45 180
40 PN 40 160 PN 160
35 140
30 120
25 PN 25 100 PN 100
20 80
15 PN 16 60 PN 64
10 PN 10 40
5 20
0 °C 0 °C
-200 -100 0 100 200 300 400 -200 -100 0 100 200 300 400
F06-7xxxxxxx-05-xx-xx-xx-000
• JIS B2238:
10...20K → Prowirl 72 W and 72 F
40K → Prowirl 72 F
bar
Class 150...600 bar
10...40 K
120 45
40 40 K
100 Cl. 600
35
80 30
25
60
Cl. 300 20 20 K
40 15
10 10 K
20 Cl. 150
5
0 °C 0 °C
-200 -100 0 100 200 300 400 -200 -100 0 100 200 300 400
F06-7xxxxxxx-05-xx-xx-xx-001
Endress+Hauser 61
10 Technical data PROline Prowirl 72
Pressure loss The pressure loss can be determined with the aid of the Applicator. The Applicator is
software for selecting and planning flowmeters. The software is available both via the
Internet (www.applicator.com) and on a CD-ROM for local PC installation.
• Sensor:
– Flanged version:
Stainless steel, A351-CF3M (1.4404), in conformity with NACE MR 0175;
Alloy C-22 version → Alloy C-22 2.4602 (A 494-CX2MW/N 26022)
– Wafer version:
Stainless steel, A351-CF3M (1.4404), in conformity with NACE MR 0175
• Flanges:
– EN (DIN) → Stainless steel, A351-CF3M (1.4404), in conformity with NACE MR 0175
(DN 15...150 with pressure rating up to PN 40: as of 2004 changeover from fully cast
construction to construction with weld-on flanges in 1.4404)
– ANSI und JIS → Stainless steel, A351-CF3M, in conformity with NACE MR 0175
( ½"...6" with pressure rating up to Cl 300 and DN 15...150 with pressure rating up
to 20 K: as of 2004 changeover from fully cast construction to construction with
weld-on flanges in 316/316L, in conformity with NACE MR 0175)
– Alloy C-22 version (EN/DIN/ANSI/JIS) → Alloy C-22 2.4602 (A 494-CX2MW/
N 26022)
62 Endress+Hauser
PROline Prowirl 72 10 Technical data
• Seal:
– Graphite (Grafoil)
– Viton
– Kalrez 6375
– Gylon (PTFE) 3504
Display elements • Liquid crystal display, two-line, plain text display, 16 characters per line
• Display can be configured individually, e.g. for measured variables and status
variables, totalizers
Ex approval More information on the Ex approvals can be found in the separate Ex documentation.
Pressure measuring Devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3
device approval (3) of the EC Directive 97/23/EC (Pressure Equipment Directive). For larger nominal
diameters, certified flowmeters to Category III are optionally also available if necessary
(depends on fluid and operating pressure). All devices are applicable for all fluids and
instable gases on principle and have been designed and manufactured in accordance
to sound engineering practice.
Endress+Hauser 63
10 Technical data PROline Prowirl 72
Ordering information Your Endress+Hauser service organisation can provide detailed ordering information
and information on the order codes on request.
10.1.10 Accessories
10.1.11 Documentation
❑ Related Ex documentation
❑ Related documentation for Pressure Equipment Directive
❑ System Information PROline Prowirl 72
❑ System Information PROline Prowirl 72/73
232 (226)*
150 (163)*
Esc
- + E
345 (368)*
Ø 8.6 (M8)
F06-72xxxxxx-06-03-00-xx-000
195
100
133
100
25
123
64 Endress+Hauser
PROline Prowirl 72 10 Technical data
161...181**
149 (142)* 141...151** 105
A 121 B
95
Esc
- + E
H – 42.5
H
65
d 65
D
SPANNUNG
TER
UN VR IR SOUS ÖF FN
F06-72xxxxxx-06-00-00-xx-000
HT OU TEN
IC PAS SI E
WARNUNG
N
N
ON
NE
AVERTISSEMENT
WARNING
KEE
IT
IN P TI ALIVE CU
EXP GHT WHE N CIR E RE
LOSIV H
E ATM OSP
Esc
- + E
H+7
!
** The dimension depends on the cable gland used.
Note!
• In the following tables, dimension H increases by 29 mm in the version with extended
temperature range (high/low temperature version) and in the version with a DSC
sensor made of Alloy C-22.
• The weight data refer to the compact version. The weight increases by 0.5 kg for the
version with extended temperature range.
DN d D H Weight
DIN/JIS ANSI [mm] [mm] [mm] [kg]
Endress+Hauser 65
10 Technical data PROline Prowirl 72
161...181**
149 (142)* 141...151** 105
A 121 B
95
Esc
- + E
H – 42
H
d
X
D L
F06-72xxxxxx-06-00-00-xx-001
SPANNUNG
TER
UN VR IR SOUS ÖF FN
HT OU TEN
IC PAS SI E
WARNUNG
N
N
ON
NE
AVERTISSEMENT
WARNING
KEE
IT
IN P TI ALIVE CU
EXP GHT WHE N CIR E RE
LOSIV H
E ATM OSP
Esc
- + E
H+7
!
** The dimension depends on the cable gland used.
Note!
• In the following tables, dimension H increases by 29 mm in the version with extended
temperature range (high/low temperature version) and in the version with a DSC
sensor made of Alloy C-22.
• The weight data refer to the compact version. The weight increases by 0.5 kg for the
version with extended temperature range.
66 Endress+Hauser
PROline Prowirl 72 10 Technical data
d D H L x Weight
DN Pressure rating
[mm] [mm] [mm] [mm] [mm] [kg]
PN 40 17.3 95.0 248 200 16 5
15
PN 160 17.3 105.0 288 200 23 7
PN 40 28.5 115.0 255 200 18 7
25 PN 100 28.5
140.0 295 200 27 11
PN 160 27.9
PN 40 43.1 150.0 263 200 21 10
40 PN 100 42.5
170.0 303 200 31 15
PN 160 41.1
PN 40 54.5 165.0 270 200 23 12
PN 64 54.5 180.0 17
50
PN 100 53.9 310 200 33
195.0 19
PN 160 52.3
PN 40 82.5 200.0 283 200 29 20
PN 64 81.7 215.0 24
80
PN 100 80.9 323 200 39 27
230.0
PN 160 76.3
PN 16 107.1 220.0
295 250 32 27
PN 40 107.1 235.0
100 PN 64 106.3 250.0 39
PN 100 104.3 265.0 335 250 49
42
PN 160 98.3
PN 16 159.3 285.0
319 300 37 51
PN 40 159.3 300.0
150 PN 64 157.1 345.0 86
PN 100 154.1 359 300 64 88
355.0
PN 160 146.3
PN 10 207.3 340.0 63
PN 16 207.3 340.0 62
200 348 300 42
PN 25 206.5 360.0 68
PN 40 206.5 375.0 72
PN 10 260.4 395.0 88
PN 16 260.4 405.0 92
250 375 380 48
PN 25 258.8 425.0 100
PN 40 258.8 450.0 111
PN 10 309.7 445.0 121
PN 16 309.7 460.0 129
300 398 450 51
PN 25 307.9 485.0 140
PN 40 307.9 515.0 158
Endress+Hauser 67
10 Technical data PROline Prowirl 72
68 Endress+Hauser
PROline Prowirl 72 10 Technical data
Endress+Hauser 69
10 Technical data PROline Prowirl 72
161...181**
149 (142)* 141...151** 105
A 121 B
95
Esc
- + E
H – 42.5
H
2 x H – 85
2xH
D d
Esc
- + E
L
C 151 (144)* 161
SPANNUNG
TER
UN VR IR SOUS ÖF FN
HT OU TEN
IC PAS SI E
F06-72xxxxxx-06-00-00-xx-002
WARNUNG
N
N
ON
NE
AVERTISSEMENT
WARNING
KEE
IT
IN P TI ALIVE CU
EXP GHT WHE N CIR E RE
LOSIV H
E ATM OSP
Esc
- + E
H+7
70 Endress+Hauser
PROline Prowirl 72 10 Technical data
! Note!
The weight data refer to the compact version. The weight increases by 0.5 kg for the
version with extended temperature range.
Endress+Hauser 71
10 Technical data PROline Prowirl 72
Weight
DN Pressure rating d [mm] D [mm] H [mm] L [mm] x [mm]
[kg]
Cl. 150 102.4 228.6
Schedule 40
Cl. 300 102.4 254.0
4" Cl. 150 97.0 228.6 335 250 49 42
Schedule 80 Cl. 300 97.0 254.0
Cl. 600 97.0 273.1
Cl. 150 154.2 279.4
Schedule 40
Cl. 300 154.2 317.5
6" Cl. 150 146.3 279.4 359 300 64 80
Schedule 80 Cl. 300 146.3 317.5
Cl. 600 146.3 355.6
72 Endress+Hauser
PROline Prowirl 72 10 Technical data
D
2
F06-7xxxxxxx-06-00-06-xx-002
D1
Fig. 32: Dimensions of flow conditioner according to EN (DIN)/ANSI, material 1.4435 (316L)
D1 : The flow conditioner is fitted at the outer diameter between the bolts.
D2 : The flow conditioner is fitted at the indentations between the bolts.
Endress+Hauser 73
10 Technical data PROline Prowirl 72
74 Endress+Hauser
Endress+Hauser
PROline Prowirl 72
▼
(P. 76) (P. 76) (P. 76)
11
11.1
SYSTEM UNITS MEAS. UNIT TYPE UNIT FLOW UNIT DENSITY UNIT TEMPERATURE UNIT LENGTH TEXT VOL. UNIT FACTOR VOL. UNIT
▼
(P. 77) (P. 77) (P. 78) (P. 79) (P. 79) (P. 80) (P. 80) (P. 80)
▼
(P. 81) (P. 81)
OPERATION LANGUAGE ACCESS CODE DEFINE PRIVATE ACCESS CODE C. ACCESS CODE C.
▼
(P. 82) (P. 82) (P. 82) CODE (P. 82) (P. 83) (P. 83)
USER INTERFACE ASSIGN LINE 1 ASSIGN LINE 2 100% VALUE FORMAT DISPLAY DAMPING CONTRAST LCD TEST DISPLAY
▼
(P. 84) (P. 84) (P. 84) (P. 84) (P. 85) (P. 85) (P. 85) (P. 85)
▼
(P. 86) (P. 86) (P. 86) (P. 86) (P. 87) (P. 87)
CURRENT OUTPUT CURRENT RANGE VALUE 20 mA TIME CONSTANT FAILSAFE MODE ACTUAL CURRENT SIMULATION VALUE SIMUL.
▼
(P. 88) (P. 88) (P. 88) (P. 88) (P. 88) (P. 89) CURRENT (P. 89) CURRENT (P. 89)
PUL./STATUS OUTP. OPERATING MODE PULSE VALUE PULSE WIDTH OUTPUT SIGNAL FAILSAFE MODE ACTUAL PULSE SIMULATION PULSE VALUE SIMULAT. PUL.
▼
(P. 90) (P. 90)
→ PULSE option (P. 90) (P. 90) (P. 91) (P. 92) (P. 92) (P. 92) (P. 93)
ASSIGN STATUS SWITCH-ON POINT SWITCH-OFF POINT TIME CONSTANT ACTUAL STATUS SIMUL. SWITCH PT. VALUE SIM. SWITCH
→ STATUS option (P. 94) (P. 94) (P. 95) (P. 95) (P. 95) (P. 96) PT. (P. 96)
COMMUNICATION TAG NAME TAG DESCRIPT. BUS ADDRESS WRITE PROTECTION BURST MODE MANUFACTURER ID DEVICE ID
▼
(P. 98) (P. 98) (P. 98) (P. 98) (P. 98) (P. 98) (P. 98) (P. 98)
PROCESS APPLICATION OPERATING DENSITY REFERENCE DENSITY OPERAT. TEMP. MATING PIPE D. ON-VALUE LF CUT OFF OFF-VALUE LF CUT
▼
PARAMETER (P. 99) (P. 99) (P. 99) (P. 100) (P. 100) (P. 101) (P. 102) OFF (P. 102)
▼
(P. 103) RETURN (P. 103) (P. 103)
SENSOR DATA K-FACTOR K-FACTOR COMPENS. NOMINAL DIAMETER METER BODY TYPE TEMP. COEFF. AMPLIFICATION
▼
(P. 104) (P. 104) (P. 104) (P. 104) MB SENSOR (P. 105)
Illustration of the function matrix
SUPERVISION ACT. SYST. CONDIT. PREV. SYST. COND. ASSIGN SYS. ERR. ERROR CATEGORY ALARM DELAY SYSTEM RESET OPERATION HOURS
▼
(P. 106) (P. 106) (P. 106) (P. 106) (P. 106) (P. 106) (P. 107) (P. 107)
Description of device functions
▼
(P. 108) (P. 108) MEASURAND (P. 108)
▼
(P. 109) (P. 109) (P. 109) (P. 109)
▼
(P. 109) (P. 109) (P. 109) (P. 109)
75
11 Description of device functions
11 Description of device functions PROline Prowirl 72
Display:
5-digit floating-point number, including unit
!
(e.g. 5.545 dm³/min; 1.4359 kg/h; 731.63 gal/d, etc.)
Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78).
VORTEX FREQUENCY The vortex frequency currently measured appears on the display.
Display:
5-digit floating-point number, including unit Hz
!
(e.g. 120.23 Hz)
Note!
This function is only used for a plausibility check.
76 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
!
(see Page 99) and in the REFERENCE DENSITY function (see Page 100).
Note!
The “calculated mass flow” and “calculated corrected volume flow” unit types
are calculated with fixed values (specified OPERATING DENSITY
and REFERENCE DENSITY). For this reason only select these types of units if
the process conditions are known and do not change.
If the process conditions are not known or if the process conditions are subject
to change, we recommend you use a flow computer (e.g. Compart DXF 351 or
RMC 621). Even when the process conditions change, these flow computers
can calculate the flow correctly by means of pressure and temperature
compensation.
Options:
VOLUME FLOW
CALCULATED MASS FLOW
CORRECTED VOLUME FLOW (calculated)
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
! Note!
If the type of unit is changed, you are asked whether the totalizer should be
reset to 0. The new type of unit is only accepted if you confirm this message
with yes. If the message is not confirmed, the measuring system continues
working with the type of unit previously active.
Endress+Hauser 77
11 Description of device functions PROline Prowirl 72
! Note!
The following units of time can be selected:
s = second, m = minute, h = hour, d = day
Options:
(with MEASURING UNIT TYPE function = VOLUME FLOW)
Metric:
Cubic centimetre → cm³/time unit
Cubic decimetre → dm³/time unit
Cubic metre → m³/time unit
Millilitre → ml/time unit
Litre → l/time unit
Hectolitre → hl/time unit
Megalitre → Ml/time unit MEGA
US:
Cubic centimeter → cc/time unit
Acre foot → af/time unit
Cubic foot → ft³/time unit
Fluid ounce → ozf/time unit
Gallon → US gal/time unit
Million gallon → US Mgal/time unit
Barrel (normal fluids: 31.5 gal/bbl) → US bbl/time unit NORM.
Barrel (beer: 31.0 gal/bbl) → US bbl/time unit BEER
Barrel (petrochemicals: 42.0 gal/bbl) → US bbl/time unit PETR.
Barrel (filling tanks: 55.0 gal/bbl) → US bbl/time unit TANK
Imperial:
Gallon → imp. gal/time unit
Mega gallon → imp. Mgal/time unit
Barrel (beer: 36.0 gal/bbl) → imp. bbl/time unit BEER
Barrel (petrochemicals: 34.97 gal/bbl) → imp. bbl/time unit PETR.
Factory setting
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
Options:
(with MEASURING UNIT TYPE function = CALCULATED MASS FLOW)
Metric:
Gram → g/time unit
Kilogram → kg/time unit
Metric ton → t/time unit
US:
Ounce → oz/time unit
Pound → lb/time unit
Ton → ton/time unit
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
78 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
US:
Standard cubic meter → Sm³/time unit
Standard cubic feet → Scf/time unit
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
!
these Operating Instructions)
Note!
The unit for the totalizer is independent of the option selected here; it is
selected in the UNIT TOTALIZER function (see Page 86).
UNIT DENSITY
! Note!
This function is not available unless the CALCULATED MASS FLOW or
CORRECTED VOLUME FLOW (calculated) option was selected in the
MEASURING UNIT TYPE function (see Page 77).
Use this function to select the unit displayed for the density to be entered in the
OPERATING DENSITY function (see Page 99) and REFERENCE DENSITY
function (see Page 100).
Options:
Metric:
g/cm³; g/cc; kg/dm³; kg/l; kg/m³; SD 4 °C, SD 15 °C, SD 20 °C; SG 4 °C,
SG 15 °C, SG 20 °C
US:
lb/ft³; lb/US gal; lb/US bbl NORM (normal fluids); lb/US bbl BEER (beer); lb/US
bbl PETR. (petrochemicals); lb/US bbl TANKS (filling tanks)
Imperial:
lb/imp. gal; lb/imp. bbl BEER (beer); lb/imp. bbl PETR. (petrochemicals)
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
UNIT TEMPERATURE Use this function to select the unit displayed for the temperature to be entered
in the OPERATING TEMPERATURE function (see Page 100) .
Options:
°C (CELSIUS)
K (KELVIN)
°F (FAHRENHEIT)
R (RANKINE)
Factory setting:
Depends on country, see P. 110 (Metric system units) or Page 112 (US units)
Endress+Hauser 79
11 Description of device functions PROline Prowirl 72
Options:
MILLIMETER
INCH
Factory setting:
Depends on country, see Page 110 (Metric system units) or Page 112 (US
units)
TEXT ARBITRARY
VOLUME UNIT ! Note!
This function is not available unless the VOLUME FLOW option was selected in
the MEASURING UNIT TYPE function (see Page 77).
Use this function to enter a text for a selectable volume flow unit. You define
only the text, the associated unit of time is selected in the UNIT FLOW function
(see Page 78).
User input:
xxxx (max. 4 characters)
Valid characters are A-Z, 0-9, +, -, decimal point, white space or underscore
Factory setting:
“– – – –” (no text)
!
Example: see FACTOR ARBITRARY VOLUME UNIT function.
Note!
The volume unit defined in this function is offered as a possible option (arbitrary
volume unit) in the UNIT FLOW function (see Page 78).
FACTOR ARBITRARY
VOLUME UNIT ! Note!
This function is not available unless a text was entered in the TEXT ARBITRARY
VOLUME UNIT function.
Use this function to define a quantity factor (without time) for the selectable
volume flow unit. The volume unit on which this factor is based is one liter.
User input:
5-digit floating-point number
Factory setting:
1
Unit:
Text arbitrary volume unit / litre
Example:
You want to measure saturated steam at 180°C constant and display the heat
flow. Take the following values from a table document (e.g. IAPWS-IF97):
a. density: 5.158 kg/m³
b. enthalpy: 2777.22 kJ/kg
In the TEXT ARBITRARY VOLUME UNIT function, you could enter “JOUL”, for
example, as the name for the volume unit (appears as an option in the UNIT
FLOW function). The value 538.9 would have to be specified in the FACTOR
ARBITRARY VOLUME UNIT function.
80 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Factory setting:
NO
! Note!
Please refer to Page 40 for a detailed description of the “Commissioning” Quick
Setup menu.
Endress+Hauser 81
11 Description of device functions PROline Prowirl 72
Options:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
NEDERLANDS
NORSK
SVENSKA
SUOMI
POLSKI (in preparation)
CESKI (in preparation)
Factory setting:
!
Depends on country, see P. 110 (Metric system units) or Page 112 (US units)
Note!
If you press the P keys simultaneously at startup, the language defaults to
“ENGLISH”.
ACCESS CODE All data of the measuring system are protected against inadvertent change.
Programming is disabled and the settings cannot be changed until a code is
P
entered in this function. If you press the keys in any function, the measuring
system automatically goes to this function and the prompt to enter the code
appears on the display (when programming is disabled). You can enable
programming by entering the private code (factory setting = 72, see DEFINE
PRIVATE CODE function).
User input:
!
Max. 4-digit number: 0...9999
Note!
• The programming levels are disabled if you do not press a key within
60 seconds following a return to the HOME position.
• You can also disable programming in this function by entering any number
(other than the private code).
• The Endress+Hauser service organisation can be of assistance if you mislay
your private code.
DEFINE PRIVATE CODE Use this function to specify the private code for enabling programming.
User input:
Max. 4-digit number: 0...9999
Factory setting:
!
72
Note!
• Programming is always enabled if the code defined = 0.
• Programming has to be enabled before this code can be changed. When
programming is disabled this function cannot be edited, thus preventing
others from accessing your personal code.
82 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Display:
ACCESS CUSTOMER (parameters can be modified)
LOCKED (parameters cannot be modified)
ACCESS CODE The number of times the private and service code was entered to access the
COUNTER device appears on the display.
(in preparation)
Display:
Integer (delivery status: 0)
Endress+Hauser 83
11 Description of device functions PROline Prowirl 72
Options:
OFF
FLOW
FLOW IN %
Factory setting:
FLOW
ASSIGN LINE 2 Use this function to define which display value is assigned to the additional line
(bottom line of the local display) for display during normal measuring operation.
Options:
OFF
FLOW
FLOW IN %
TOTALIZER
TAG NAME
OPERATING/SYSTEM CONDITIONS
FLOW BARGRAPH IN %
Factory setting:
TOTALIZER
100% VALUE
! Note!
This function is not available unless
• the FLOW IN % option was selected in the ASSIGN LINE 1 function
or
• the FLOW IN % or FLOW BARGRAPH IN % option was selected in the
ASSIGN LINE 2 function.
Use this function to define the flow value to be shown on the display as the
100% value.
User input:
5-digit floating-point number
Factory setting:
Depends on the nominal diameter, application and country,
see Page 110 (Metric system units) or Page 112 (US units)
! Note!
If a value was specified for the VALUE 20 mA function when ordering, this value
is also used here as the factory setting.
84 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Options:
XXXXX. - XXXX.X - XXX.XX - XX.XXX -X.XXXX
Factory setting:
!
XX.XXX
Note!
• Note that this setting only affects the reading as it appears on the display, it
has no influence on the accuracy of the system's calculations.
• The places after the decimal point as computed by the measuring device
cannot always be displayed, depending on this setting and the engineering
unit. In these instances an arrow appears on the display between the
measured value and the engineering unit (e.g. 1.2 → kg/h), indicating that
the measuring system is computing with more decimal places than can be
shown on the display.
DISPLAY DAMPING Use this function to enter a time constant defining how the display reacts to
severely fluctuating flow variables, either very quickly (enter a low time
constant) or with damping (enter a high time constant).
User input:
0...100 s
Factory setting:
5s
! Note!
• The setting 0 seconds switches off damping.
• The reaction time of the function depends on the time specified in the FLOW
DAMPING function (see Page 103).
CONTRAST LCD Use this function to optimise the display contrast to suit local operating
conditions.
User input:
10...100%
Factory setting:
50%
! Note!
P
If you press the keys simultaneously at startup, the language defaults to
“ENGLISH” and the contrast is reset to the factory setting.
TEST DISPLAY Use this function to test the operability of the local display and its pixels.
Options:
OFF
ON
Factory setting:
OFF
Test sequence:
1. Start the test by selecting ON.
2. All pixels of the main line and additional line are darkened for minimum 0.75
seconds.
3. The main line and additional line show an “8” in each field for minimum 0.75
seconds.
4. The main line and additional line show a “0” in each field for minimum 0.75
seconds.
5. The main line and additional line show nothing (blank display) for minimum
0.75 seconds.
6. When the test is completed, the local display returns to its initial state and
the displays the option OFF.
Endress+Hauser 85
11 Description of device functions PROline Prowirl 72
Display:
!
Max. 7-digit floating-point number, including unit (e.g.15467.4 m³)
Note!
The totalizer's response to errors is defined in the “FAILSAFE MODE” function
(see Page 87).
OVERFLOW The total for the totalizer's overflow aggregated since measuring commenced
appears on the display.
Example:
Reading after 2 overflows: 2 E7 kg (= 20,000,000 kg)
The value displayed in the SUM function = 196,845.7 kg
Effective total quantity = 20,196,845.7 kg
Display:
Integer with exponent, including sign and unit, e.g. 2 E7 kg
UNIT Use this function to define the unit for the totalizer.
TOTALIZER Depending on what is selected in the MEASURING UNIT TYPE function (see
Page 77), only the associated units (volume or mass or corrected volume) are
displayed here for selection.
Options:
(with MEASURING UNIT TYPE function = VOLUME FLOW)
Metric:
Cubic centimetre → cm³
Cubic decimetre → dm³
Cubic metre → m³
Millilitre → ml
Litre → l
Hectolitre → hl
Megalitre → Ml
US:
Cubic centimeter → cc
Acre foot → af
Cubic foot → ft³
Fluid ounce → ozf
Gallon → gal
Million gallon → Mgal
Barrel → bbl (normal fluids)
Barrel → bbl (beer)
Barrel → bbl (petrochemicals)
Barrel → bbl (filling tanks)
Imperial:
Gallon → imp. gal/...
Mega gallon → imp. Mgal/...
Barrel (beer: 36.0 gal/bbl) → imp. bbl/... BEER
Barrel (petrochemicals: 34.97 gal/bbl) → imp. bbl/... PETR.
86 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Factory setting:
Depends on country, see Page 110 (Metric system units) or Page 112 (US
units)
Options:
(with MEASURING UNIT TYPE function = CALCULATED MASS FLOW)
Metric:
Gram → g
Kilogram → kg
Metric ton → t
US:
Ounce → oz
Pound → lb
Ton → ton
Factory setting:
Depends on country, see Page 110 (Metric system units) or Page 112 (US
units)
Options:
(with MEASURING UNIT TYPE function = CORRECTED VOLUME FLOW)
Metric:
Standard litre → Nl
Standard cubic meter → Nm³
US:
Standard cubic meter → Sm³
Standard cubic feet → Scf
Factory setting:
Depends on country, see P. 110 (Metric system units) or Page 112 (US units)
RESET TOTALIZER Use this function to reset the sum and the overflow of the totalizer to
0 (= RESET).
Options:
NO
YES
Factory setting:
NO
FAILSAFE MODE Use this function to define the response of the totalizer to an alarm condition.
Options:
STOP
The totalizer does not continue to count the flow if a fault is present. The
totalizer stops at the last value before the alarm condition occurred.
ACTUAL VALUE
The totalizer continues to count the flow on the basis of the current flow data.
The fault is ignored.
HOLD VALUE
The totalizer continues to count the flow on the basis of the last valid flow data
(before the fault occurred).
Factory setting:
STOP
Endress+Hauser 87
11 Description of device functions PROline Prowirl 72
Options:
4-20 mA HART NAMUR
4-20 mA HART US
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
User input:
5-digit floating-point number
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
!
these Operating Instructions)
Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78).
TIME CONSTANT Use this function to select a time constant defining how the current output
signal reacts to severely fluctuating measured variables, either very quickly
(low time constant) or with damping (high time constant).
User input:
Fixed-point number: 0...100 s
Factory setting:
!
5s
Note!
The reaction time of the function also depends on the time specified in the
FLOW DAMPING function (see Page 103).
FAILSAFE MODE The dictates of safety render it advisable to ensure that the current output
assumes a predefined state in the event of a fault. Use this function to define
the response of the current output to fault. The setting you select here affects
only the current output. It has no effect on other outputs or the display (e.g.
totalizers).
Options:
MIN. CURRENT
Depends on the option selected in the CURRENT RANGE function (see
Page 88). If the current range is:
4-20 mA HART NAMUR → output current = 3.6 mA
4-20 mA HART US → output current = 3.75 mA
MAX. CURRENT
22.6 mA
HOLD VALUE
Measured value output is based on the last measured value saved before the
error occurred.
ACTUAL VALUE
Measured value output is based on the current flow measurement.
The fault is ignored.
Factory setting:
MAX. CURRENT
88 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Display:
3.60...22.60 mA
SIMULATION CURRENT Use this function to activate simulation of the current output.
Options:
OFF
ON
Factory setting:
!
OFF
Note!
• The notice message #611 “SIMULATION CURRENT OUTPUT” indicates that
simulation is active.
• The value which should be output at the current output is defined in the
VALUE SIMULATION CURRENT function.
• The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the other outputs
and the display.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
CURRENT ! Note!
This function is not available unless the ON option was selected in the
SIMULATION CURRENT function.
Use this function to define a selectable value (e.g. 12 mA) to be output at the
current output. This value is used to test downstream devices and the
measuring device itself.
User input:
Floating-point number: 3.60...22.60 mA
Factory setting:
3.60 mA
" Caution!
!
The setting is not saved if the power supply fails.
Note!
Simulation is started by confirming the simulation value with the key. If the
F F
Endress+Hauser 89
11 Description of device functions PROline Prowirl 72
Options:
PULSE
STATUS
VORTEX FREQUENCY (unscaled pulses)
PFM
Factory setting:
!
PULSE
Note!
• After PFM is selected, the group current output is no longer available
(see Page 88 ff.) If PFM is selected, current simulation with a simulation value
of 4 mA is automatically activated. If the transmitter was wired for pulse-
frequency modulation (see Page 24), the HART protocol is not available.
• If VORTEX FREQUENCY and PFM are selected, the vortex pulses are
passed on directly. The low flow cut off is taken into account.
PULSE VALUE
! Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
Use this function to define the flow at which a pulse should be output. These
pulses can be totalled by an external totalizer and in this way the total flow
!
since measuring commenced can be registered.
Note!
Select the pulse value in such a way that the pulse frequency does not exceed
a value of 100 Hz with maximum flow.
User input:
5-digit floating-point number
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
! Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78).
PULSE WIDTH
! Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
Use this function to enter the maximum pulse width of the output pulses.
User input:
5...2000 ms
Factory setting:
20 ms
Pulse output is always with the pulse width (B) entered in this function.
The intervals (P) between the individual pulses are automatically adjusted.
However, they must at least correspond to the pulse width (B = P).
90 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
P P
F06-xxxxxxxx-05-xx-xx-xx-012
B = Pulse width entered (the illustration applies to positive pulses)
P = Intervals between the individual pulses
! Note!
When entering the pulse width, select a value that can still be processed by a
connected totalizer (e.g. mechanical totalizer, PLC, etc.).
" Caution!
If the pulse number or frequency resulting from the pulse value entered (see
PULSE VALUE function on Page 90) and from the current flow is too large to
maintain the pulse width selected (interval P is smaller than the pulse width B
entered), a system error message (#359, PULSE RANGE) is generated after
buffering/balancing has occurred.
OUTPUT SIGNAL
! Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
Use this function to configure the pulse output in such a way that it suits an
external totalizer, for example. Depending on the application, you can select
the direction of the pulses here.
Options:
PASSIVE - POSITIVE
PASSIVE - NEGATIVE
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
these Operating Instructions)
PASSIVE:
Umax = 30 V DC
Open
External power supply
Collector
1 5 7 8 3
+
=
-
!
F-xxxxxxx-04-xx-xx-en-000
Note!
For continuous currents up to 15 mA
PASSIVE-NEGATIVE
Pulse (B = Pulse width)
transistor
B
conducting
non conducting
t
F06-7xxxxxxx-05-xx-xx-en-002
PASSIVE-POSITIVE
Pulse (B = Pulse width)
transistor
B
conducting
non conducting
t
F06-7xxxxxxx-05-xx-xx-en-003
Endress+Hauser 91
11 Description of device functions PROline Prowirl 72
The dictates of safety render it advisable to ensure that the pulse output
assumes a predefined state in the event of a fault. Use this function to define
this state. The setting you select here affects only the pulse output. It has no
effect on other outputs or the display (e.g. totalizers).
Options:
FALLBACK VALUE
Output is 0 pulse.
HOLD VALUE
Measured value output is based on the last measured value saved before the
error occurred.
ACTUAL VALUE
Measured value output is based on the current flow measurement. The fault is
ignored.
Factory setting:
FALLBACK VALUE
ACTUAL PULSE
! Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
The current computed actual value of the output frequency appears on the
display.
Display:
0...100 pulse/second
SIMULATION PULSE
! Note!
This function is not available unless the PULSE option was selected in the
OPERATING MODE function.
Options:
OFF
COUNTDOWN
The pulses specified in the VALUE SIMULATION PULSE function are output.
CONTINUOUSLY
Pulses are continuously output with the pulse width specified in the PULSE
WIDTH function. Simulation is started once the CONTINUOUSLY option is
F
confirmed with the key.
! Note!
Simulation is started by confirming the CONTINUOUSLY option with the key. F
If the key is pressed again afterwards, the prompt “End simulation” (NO/YES)
F
appears.
If you choose “NO”, simulation remains active and the group selection is called
up. The simulation can be switched off again via the SIMULATION PULSE
function.
If you choose “YES”, you end the simulation and the group selection is called
up.
92 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
!
(contd.) OFF
Note!
• The notice message #631 “SIM. PULSE” indicates that simulation is active.
• The on/off ratio is 1:1 for both types of simulation.
• The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
PULSE ! Note!
This function is not available unless the COUNTDOWN option was selected in
the SIMULATION PULSE function.
Use this function to specify the number of pulses (e.g. 50) which are output
during the simulation. This value is used to test downstream devices and the
measuring device itself. The pulses are output with the pulse width specified in
the PULSE WIDTH function. The on/off ratio is 1:1.
Simulation is started once the specified value is confirmed with the F key. The
display remains at 0 if the specified pulses have been output.
User input:
0...10,000
Factory setting:
0
! Note!
Simulation is started by confirming the simulation value with the key. If the
F F
" Caution!
The setting is not saved if the power supply fails.
Endress+Hauser 93
11 Description of device functions PROline Prowirl 72
Options:
OFF
ON (operation)
FAULT MESSAGE
NOTICE MESSAGE
FAULT MESSAGE or NOTICE MESSAGE
LIMIT FLOW
LIMIT TOTALIZER
Factory setting:
!
FAULT MESSAGE
Note!
• The status output displays quiescent current behaviour, in other words the
output is closed (transistor conductive) when normal, error-free operation is
in progress.
• Please pay particular attention to the illustrations and detailed information on
the switching behaviour of the status output (see Page 97).
• If you select OFF, the only function shown in this function group is this
function (ASSIGN STATUS).
SWITCH-ON POINT
! Note!
This function is not available unless the LIMIT FLOW or LIMIT TOTALIZER
option was selected in the ASSIGN STATUS function.
Use this function to assign a value to the switch-on point (status output pulls
up). The value can be greater or less than the switch-off point. Only positive
values are permitted.
User input:
5-digit floating-point number
Factory setting:
• If LIMIT FLOW was selected in the ASSIGN STATUS function:
Depends on nominal diameter, application and country, see Page 111
(Metric system units) or Page 112 (US units)
• If LIMIT TOTALIZER was selected in the ASSIGN STATUS function:
!
0
Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78) or
UNIT TOTALIZER function (see Page 86).
94 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Use this function to assign a value to the switch-off point (status output drops
out). The value can be greater or less than the switch-on point. Only positive
values are permitted.
User input:
5-digit floating-point number
Factory setting:
Depends on nominal diameter, application and country, see Page 111
!
(Metric system units) or Page 112 (US units)
Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78).
TIME CONSTANT
! Note!
This function is not available unless the LIMIT FLOW option was selected in the
ASSIGN STATUS function.
Use this function to select a time constant defining how the measuring signal
reacts to severely fluctuating measured variables, either very quickly (low time
constant) or with damping (high time constant).
The purpose of damping, therefore, is to prevent the status output changing
state continuously in response to fluctuations in flow.
User input:
0...100 s
Factory setting:
0s
! Note!
The reaction time of the function depends on the time specified in the FLOW
DAMPING function (see Page 103).
ACTUAL STATUS
OUTPUT ! Note!
This function is not available unless the STATUS option was selected in the
OPERATING MODE function.
Display:
NOT CONDUCTIVE
CONDUCTIVE
Endress+Hauser 95
11 Description of device functions PROline Prowirl 72
Options:
OFF
ON
Factory setting:
OFF
! Note!
• The notice message #641 “SIMULATION STATUS OUTPUT” indicates that
simulation is active.
• The measuring device continues to measure while simulation is in progress,
i.e. the current measured values are output correctly via the other outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT ! Note!
This function is not available unless the ON option was selected in the
SIMULATION SWITCH POINT function.
Use this function to define the switching behaviour of the status output during
the simulation. This value is used to test downstream devices and the
measuring device itself.
User input:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
!
NOT CONDUCTIVE
Note!
You can change the switching behaviour of the status output during the
simulation. The prompt “CONDUCTIVE” or “NOT CONDUCTIVE” appears if the
O S or key is pressed. Select the desired switching behaviour and start the
F
simulation with the key.
If the key is pressed again afterwards, the prompt “End simulation” (NO/YES)
F
appears. If you choose “NO”, simulation remains active and the group selection
is called up. The simulation can be switched off again via the SIMULATION
SWITCH POINT function.
If you choose “YES”, you end the simulation and the group selection is called
up.
" Caution!
The setting is not saved if the power supply fails.
96 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Measured variable
➀
➁
F-xxxxxxx-05-xx-xx-xx-005
➂ ➂
t
➀ = ON ≤ SWITCH-OFF POINT (maximum safety)
➁ = ON > SWITCH-OFF POINT (minimum safety)
➂ = Status output switched off (not conductive)
Endress+Hauser 97
11 Description of device functions PROline Prowirl 72
User input:
Max. 8-character text, permitted characters are: A-Z, 0-9, +,–, punctuation
marks
Factory setting:
“– – – – – – – –” (no text)
TAG DESCRIPTION Use this function to enter a tag description for the measuring device. You can
edit and read this tag name via the local display or the HART protocol
User input:
Max. 16-character text, permitted characters are: A-Z, 0-9, +,–, punctuation
marks
Factory setting:
“– – – – – – – – – – – – – – – –” (no text)
BUS ADDRESS Use this function to define the address for the exchange of data with the HART
protocol.
User input:
0...15
Factory setting:
0
! Note!
A constant 4 mA current is applied with addresses 1...15.
WRITE PROTECTION Use this function to check whether the measuring device can be write-
accessed.
Display:
OFF (execution status) = Data exchange possible
ON = Data exchange disabled
! Note!
Write protection is activated and deactivated by means of a DIP switch on the
amplifier board (see Page 38).
BURST MODE Use this function to activate cyclic data exchange of the process variables flow
and sum to achieve faster communication.
Options:
OFF
ON
Factory setting:
OFF
MANUFACTURER ID The manufacturer number in decimal numerical format appears on the display.
Display:
17 = (11 hex) for Endress+Hauser
Display:
56 = (86 dec) for Prowirl 72
98 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Options:
GAS/STEAM
LIQUID
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
!
these Operating Instructions)
Note!
• If the selection in this function is changed, the values in the following
functions have to be adapt:
– VALUE 20 mA, see Page 88
– PULSE WIDTH, see Page 90
– 100% VALUE (line 1), see Page 84
– 100% VALUE (line 2), see Page 84
• If the selection is changed in this function, you are asked whether the
totalizer should be reset to 0. We recommend you confirm this message and
reset the totalizer.
OPERATING DENSITY
! Note!
This function is not available unless the CALCULATED MASS FLOW or
CORRECTED VOLUME FLOW option was selected in the MEASURING UNIT
TYPE function.
Use this function to enter a fixed value for the density at process conditions.
This value is used to calculate the calculated mass flow and corrected volume
flow (see MEASURING UNIT TYPE function on Page 77).
User input:
5-digit floating-point number
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
!
these Operating Instructions)
Note!
• The appropriate unit is taken from the UNIT DENSITY function (see Page 78).
• If the selection is changed in this function, you are asked whether the
totalizer should be reset to 0. We recommend you confirm this message and
reset the totalizer.
Endress+Hauser 99
11 Description of device functions PROline Prowirl 72
Use this function to enter a fixed value for the density at reference conditions.
This value is used to calculate the corrected volume flow (see MEASURING
UNIT TYPE function on Page 77).
User input:
5-digit floating-point number
Factory setting:
See parameter printout supplied (the parameter printout is an integral part of
!
these Operating Instructions)
Note!
• The appropriate unit is taken from the UNIT DENSITY function (see Page 78).
• If the selection is changed in this function, you are asked whether the
totalizer should be reset to 0. We recommend you confirm this message and
reset the totalizer.
OPERATING Use this function to specify a fixed value for the process temperature.
TEMPERATURE
! Note!
• The sensor (measuring pipe and bluff body) expands differently depending
on the existing process temperature. This has a proportional effect on the
accuracy of the measuring system since the device was calibrated at a fixed
calibration temperature of 20 °C (293 K). However, this effect on the current
measured value and the internal totalizer can be compensated by entering
an average process temperature in this function.
User input:
5-digit floating-point number
Factory setting:
20°C / 293.16 K / 68 °F / 527.67 R
! Note!
The appropriate unit is taken from the UNIT TEMPERATURE function (see
Page 79).
" Caution!
This setting does not change the permitted temperature range of the
measuring system. Please pay particular attention to the temperature
application limits specified in the product specification (see Page 61).
100 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
If the inner diameter of the mating pipe (d1) and the inner diameter of the flow-
meter (d2) are different, this results in a distorted flow profile.
A mismatch of the inner pipe diameter can be present, if the mating pipework is
different from the flowmeter in:
• pressure rate.
• Schedule (ANSI pipes, e.g. Sched. 80 instead of 40).
• material (DIN pipes).
In order to correct the resulting shift of the instrument‘s calibration factor, enter
the actual inner diameter of the mating pipework (d1) in this function.
F06-72xxxxxx-16-00-00-xx-000
Esc
- + E
d1 d2
d1 > d2
d1 = diameter mating pipework
d2 = diameter flowmeter
User input:
5-digit floating-point number
Factory setting:
0
! Note!
• The inlet correction is switched off if 0 is entered in the function.
• The appropriate unit is taken from the UNIT LENGTH function (see Page 80).
• Mismatches between inner diameter of piping and inner diameter of the
flowmeter can only be corrected within the same diameter class
(e.g. DN 50, ½“).
• If the standard inner diameter of the process connection ordered for the
device and the inner diameter of the pipe line are different, then an additional
uncertainty of typically 0.1% (of measurement) per 1 mm difference of
diameter has to be allowed for.
Endress+Hauser 101
11 Description of device functions PROline Prowirl 72
User input:
5-digit floating-point number
Factory setting:
Below the standard measuring range
! Note!
• The appropriate unit is taken from the UNIT FLOW function (see Page 78).
• The on-value can be set at a value corresponding to a Reynolds number of
Re = 20,000. This has the effect that measurements in the non-linear range
are not evaluated. The Reynolds number and the flow (at Reynolds number =
20,000) can be determined using the Endress+Hauser “Applicator” software.
The Applicator is software for selecting and planning flowmeters. The values
required can be determined without having to connect the transmitter
beforehand. The “Applicator” is available both via the Internet
(www.applicator.com) and on a CD-ROM for local PC installation.
OFF-VALUE LOW FLOW Use this function to enter the off-value for low flow cut off.
CUT OFF Enter the off-value as a positive hysteresis from the on-value.
User input:
Integer 0...100%
Factory setting:
50%
Example:
Q
2 2
b
F06-80xxxxxx-05-xx-xx-xx-007
H
a
1 1
t
c c
Q = Flow [volume/time]
t = Time
a = ON-VALUE LOW FLOW CUT OFF = 20 m³/h
b = OFF-VALUE LOW FLOW CUT OFF = 10%
c = Low flow cut off active
1 = Low flow cut off is switched on at 20 m³/h
2 = Low flow cut off is switched off at 22 m³/h
H = Hysteresis
102 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Options:
OFF
ON (signal output is set to the value for zero flow).
Factory setting:
OFF
FLOW DAMPING Use this function to set the filter depth of the digital filter. This reduces the
sensitivity of the measuring signal to interference peaks (e.g. in the event of
high solids content, gas bubbles in the fluid, etc.). The measuring system
reaction time increases with the filter setting.
User input:
0...100 s
Factory setting:
!
1s
Note!
The flow damping acts on the following functions and outputs of the measuring
device:
Function Display
DISPLAY DAMPING
F06-72xxxxxx-19-xx-xx-en-001
Endress+Hauser 103
11 Description of device functions PROline Prowirl 72
" Caution!
Under normal circumstances these settings may not be changed because changes affect numerous
functions of the entire measuring system, and the accuracy of the measuring system in particular.
Please contact your E+H service organisation if you have any questions on these functions.
K-FACTOR The current calibration factor of the sensor appears on the display.
Display:
e.g. 100 P/l (pulse per litre)
! Note!
The K-factor is also given on the nameplate, the sensor and the calibration
report under “K-fct.”.
K-FACTOR The current compensated calibration factor of the sensor appears on the
COMPENSATED display.
Display:
e.g. 102 P/l (pulse per litre)
NOMINAL DIAMETER The nominal diameter of the sensor appears on the display.
Display:
e.g. DN 25
METER BODY TYPE MB The type of meter body (MB) of the sensor appears on the display.
Display:
!
e.g. 71
Note!
• Use this function to specify the nominal diameter and the sensor type.
• The Meter Body Type MB is also given on the parameter printout.
TEMPERATURE The temperature effect on the calibration factor appears on the display. Due to
COEFFICIENT changes in temperature, the meter body expands differently, depending on the
material. The expansion has an effect on the K-factor
Display:
4.8800*10-5 / K (stainless steel)
2.6000*10-5 / K (Alloy C-22)
104 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
" Caution!
Incorrectly configured amplification can have the following effects:
• The measuring range is limited in such a way that small flows cannot be
recorded or displayed. In this instance, the value for the amplification must
be increased.
• Undesired interference signals are registered by the device which means
that a flow is recorded and displayed even if the fluid is at a standstill. In this
instance, the value for the amplification must be reduced.
Options:
1...5 (1 = smallest amplification, 5= largest amplification)
Factory setting:
3
Endress+Hauser 105
11 Description of device functions PROline Prowirl 72
PREVIOUS SYSTEM The last 16 fault and notice messages appear on the display.
CONDITIONS
ASSIGN SYSTEM $
All system errors and the associated error categories (fault message or notice
ERROR message !) appear on the display. If you select a single system error you can
change its error category.
Display:
List of system errors
! Note!
O
• Each individual message can be selected using the and key. S
F
• If the key is pressed twice, the ERROR CATEGORY function is called up.
• Use the key combination or select “CANCEL” (in the system error list) to
P
ERROR CATEGORY Use this function to define whether a system error triggers a notice message or
a fault message. If you select “FAULT MESSAGES”, all outputs respond to an
error in accordance with their defined failsafe mode.
Options:
NOTICE MESSAGES (display only)
!
FAULT MESSAGES (outputs and display)
Note!
F
• If the key is pressed twice, the ASSIGN SYSTEM ERROR function is called
up.
• Use the key combination to exit the function.
P
ALARM DELAY Use this function to define a time span for which the criteria for an error have to
be satisfied without interruption before a fault or notice message is generated.
Depending on the setting and the type of error, this suppression acts on the
display, the current output and the pulse/status output.
User input:
0...100 s (in steps of one second)
Factory setting:
0s
" Caution!
If this function is used, fault and notice messages are delayed by the time
corresponding to the setting before being forwarded to the higher-level
controller (PCS, etc.). It is therefore imperative to check in advance whether a
delay of this nature could affect the safety requirements of the process. If fault
and notice messages may not be suppressed, a value of 0 seconds must be
entered here.
106 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Options:
NO
Factory setting:
NO
OPERATION HOURS The hours of operation of the device appear on the display.
(in preparation)
Display:
Depends on the number of hours of operation elapsed:
Hours of operation < 10 hours → display format = 0:00:00 (hr:min:sec)
Hours of operation 10...10,000 hours → display format = 0000:00 (hr:min)
Hours of operation < 10,000 hours → display format = 000000 (hr)
Endress+Hauser 107
11 Description of device functions PROline Prowirl 72
Options:
OFF
ON
Factory setting:
OFF
SIMULATION Use this function to set all inputs, outputs and the totalizer to their defined flow-
MEASURAND response modes, in order to check whether they respond correctly. During this
time, the message “#692 SIMULATION MEASURAND” appears on the display.
Options:
OFF
FLOW
Factory setting:
OFF
" Caution!
• The measuring device can only be used for measuring to a certain extent
while the simulation is in progress.
• The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND ! Note!
This function is not available unless the SIMULATION MEASURAND function is
active.
Use this function to specify a selectable value (e.g. 12 dm³/s). This value is
used to test downstream devices and the measuring device itself.
User input:
5-digit floating-point number
Factory setting:
0
! Note!
The appropriate unit is taken from the UNIT FLOW function (see Page 78).
" Caution!
The setting is not saved if the power supply fails.
108 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
SENSOR TYPE The sensor type (e.g. Prowirl F) appears on the display.
SERIAL NUMBER The serial number of the DSC sensor appears on the display.
DSC SENSOR
SOFTWARE REVISION The software revision number of the amplifier appears on the display.
NUMBER AMPLIFIER
HARDWARE REVISION The hardware revision number of the I/O module appears on the display.
NUMBER
I/O MODULE
Endress+Hauser 109
11 Description of device functions PROline Prowirl 72
110 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Endress+Hauser 111
11 Description of device functions PROline Prowirl 72
112 Endress+Hauser
PROline Prowirl 72 11 Description of device functions
Endress+Hauser 113
11 Description of device functions PROline Prowirl 72
114 Endress+Hauser
PROline Prowirl 72 Index
Index
A D
Access Damping
Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Declaration of conformity (CE mark) . . . . . . . . . . . . . 10
Actual Define private code . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Degree of protection
Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Mounting information . . . . . . . . . . . . . . . . . . . . . . 26
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Technical information . . . . . . . . . . . . . . . . . . . . . . 60
System condition . . . . . . . . . . . . . . . . . . . . . . . 106 Density
Alarm delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . 60 Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Amplification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Application Device
Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Functions (matrix) . . . . . . . . . . . . . . . . . . . . . . . . 75
Applicator (selection software) . . . . . . . . . . . . . . . . 43 ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Arbitrary volume unit Diameter
Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Jump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Mating pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Assign Dimensions
Display lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Flow conditioner . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Prowirl 72 F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
System error . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Prowirl 72 F, Dualsens . . . . . . . . . . . . . . . . . . . . . 70
Prowirl 72 W . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
B Remote version . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Burst mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Display
Bus address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Display and operating elements . . . . . . . . . . . . . 27
C Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Cable entry Rotating the local display . . . . . . . . . . . . . . . . . . 19
Degree of protection . . . . . . . . . . . . . . . . . . . . . . 26 Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Cable specifications (remote version) . . . . . . . . . . . 22
CE mark (declaration of conformity) . . . . . . . . . . . . 10 E
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Electrical connection
Commissioning Cable specifications (remote version) . . . . . . . . . 22
Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Commubox FXA 191 . . . . . . . . . . . . . . . . . . . . . . 25
Quick Setup Flowchart . . . . . . . . . . . . . . . . . . . . 41 Degree of protection . . . . . . . . . . . . . . . . . . . . . . 26
Switching on the measuring device . . . . . . . . . . 39 HART handheld terminal . . . . . . . . . . . . . . . . . . . 25
Commubox FXA 191 (electrical connection) . . . . . . 25 Post-connection check (checklist) . . . . . . . . . . . . 26
Communication (HART) . . . . . . . . . . . . . . . . . . . . . . 31 Remote version . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Contrast LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 Terminal assignment . . . . . . . . . . . . . . . . . . . . . . 24
Current output Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Actual value . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Electronics boards, installation/removal . . . . . . . . . . 52
Current range . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Electrical connection . . . . . . . . . . . . . . . . . . . . . 24 Error category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Error messages
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 System error (device error) . . . . . . . . . . . . . . . . . 46
Value simulation . . . . . . . . . . . . . . . . . . . . . . . . . 89 Types of error message . . . . . . . . . . . . . . . . . . . . 30
Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Types of error (system and process errors) . . . . 30
Current range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 European Pressure Equipment Directive . . . . . . . . . 63
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Endress+Hauser 115
Index PROline Prowirl 72
F I
Factor arbitrary volume unit . . . . . . . . . . . . . . . . . . . 80 Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . 11
Factory settings Inlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Metric system units . . . . . . . . . . . . . . . . . . . . . . 110 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
US units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 Installation
Failsafe mode Mounting
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Installation conditions
Inputs/outputs, general . . . . . . . . . . . . . . . . . . . . 50 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Pulse output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Inlet and outlet run . . . . . . . . . . . . . . . . . . . . . . . 14
Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 Installation location . . . . . . . . . . . . . . . . . . . . . . . 12
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Orientation (vertical, horizontal) . . . . . . . . . . . . . 12
FieldCheck (tester and simulator) . . . . . . . . . . . . . . 43 Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Flow
Conditioner . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 K
Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 K-factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 K-factor compensated . . . . . . . . . . . . . . . . . . . . . 104
Function
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 L
matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Function and system design . . . . . . . . . . . . . . . . . . 57 LCD contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Lengths
G See Dimensions
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Limiting flow
Group See Measuring range
Amplifier version . . . . . . . . . . . . . . . . . . . . . . . . 109 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Local display
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 See Display
Measured values . . . . . . . . . . . . . . . . . . . . . . . . . 76 Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Off-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Process parameter . . . . . . . . . . . . . . . . . . . . . . . 99 On-value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Pulse/status output . . . . . . . . . . . . . . . . . . . . . . . 90
Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 M
Sensor data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Sensor version . . . . . . . . . . . . . . . . . . . . . . . . . 109 Manufacturer ID . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Simulation system . . . . . . . . . . . . . . . . . . . . . . . 108 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Supervision . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Mating pipe diameter . . . . . . . . . . . . . . . . . . . . . . 101
System parameter . . . . . . . . . . . . . . . . . . . . . . . 103 Maximum measured error . . . . . . . . . . . . . . . . . . . . 60
System units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Measurand simulation . . . . . . . . . . . . . . . . . . . . . . 108
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . 57
User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Measuring
Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
H Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Hardware revision number System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Unit type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
I/O module . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Medium
HART Pressure range . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Command classes . . . . . . . . . . . . . . . . . . . . . . . 31 Temperature ranges . . . . . . . . . . . . . . . . . . . . . . 61
Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Meter body (MB) . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Communicator DXR 275, DXR 375 . . . . . . . . . . . 31 Mounting
Device status, error messages . . . . . . . . . . . . . . 37 Sensor (compact version) . . . . . . . . . . . . . . . . . 16
Device variables . . . . . . . . . . . . . . . . . . . . . . . . . 32 Sensor (remote version) . . . . . . . . . . . . . . . . . . . 18
Electrical connection . . . . . . . . . . . . . . . . . . . . . . 25
Operating options . . . . . . . . . . . . . . . . . . . . . . . . 31 N
Process variables . . . . . . . . . . . . . . . . . . . . . . . . 32 Nameplate
Hazardous substances . . . . . . . . . . . . . . . . . . . . . . . 8 Sensor remote version . . . . . . . . . . . . . . . . . . . . 10
Heat insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
HOME position (operating mode display) . . . . . . . . 27 Nominal diameter . . . . . . . . . . . . . . . . . . . . . . . . . 104
Nominal pressure . . . . . . . . . . . . . . . . . . . . . . . . . . 61
116 Endress+Hauser
PROline Prowirl 72 Index
O Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Off-value Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . 102 Replacing
On-value Electronics boards (installation/removal) . . . . . . . 52
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . 102 Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operating Reset
Density (function) . . . . . . . . . . . . . . . . . . . . . . . . 99 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Temperature (function) . . . . . . . . . . . . . . . . . . . 100 Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Operating mode, selection . . . . . . . . . . . . . . . . . . . 90 Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operation
Display and operating elements . . . . . . . . . . . . . 27 S
General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Safety
Operation hours . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operational Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Seals
Order code Replacing, replacement seals . . . . . . . . . . . . . . . 42
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Seals (replacing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Sensor type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Serial number
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . 63 DSC sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Output Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Signal pulse, configuring . . . . . . . . . . . . . . . . . . 91 Simulation
Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Technical information . . . . . . . . . . . . . . . . . . . . . 58 Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
P Pulse output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Perforated plate flow conditioner . . . . . . . . . . . . . . . 15 Switch point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Performance characteristics . . . . . . . . . . . . . . . . . . 60 Software
Positive zero return . . . . . . . . . . . . . . . . . . . . . . . . 103 Amplifier display . . . . . . . . . . . . . . . . . . . . . . . . . 39
Post-installation check (checklist) . . . . . . . . . . . . . . 19 Revision number amplifier . . . . . . . . . . . . . . . . . 109
Power supply (supply voltage) . . . . . . . . . . . . . . . . 59 Versions (history) . . . . . . . . . . . . . . . . . . . . . . . . . 56
Pressure Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Standards, guidelines . . . . . . . . . . . . . . . . . . . . . . . . 63
Measuring device approval (PED) . . . . . . . . . . . 63 Status access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Previous system conditions . . . . . . . . . . . . . . . . . . 106 Status output
Process errors without display messages . . . . . . . . 48 Actual status . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Programming mode (enable/disable) . . . . . . . . . . . 29 Assign . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Pulse General information . . . . . . . . . . . . . . . . . . . . . . . 97
Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Limit value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Simulation switch point . . . . . . . . . . . . . . . . . . . . 96
Pulse output Switching behaviour . . . . . . . . . . . . . . . . . . . . . . 97
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Switch-off point . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 Switch-on point . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Pulse actual value . . . . . . . . . . . . . . . . . . . . . . . . 92 Time constant . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Pulse value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Value simulation switch point . . . . . . . . . . . . . . . . 96
Pulse width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Storage
Simulation pulse . . . . . . . . . . . . . . . . . . . . . . . . . 92 Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Value simulation . . . . . . . . . . . . . . . . . . . . . . . . . 93 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Pulse/status output Supplementary Ex documentation . . . . . . . . . . . . . . . 7
Operating mode . . . . . . . . . . . . . . . . . . . . . . . . . 90 Supply
Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
R Voltage (power supply) . . . . . . . . . . . . . . . . . . . . 59
Reference Switch point
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Operating conditions . . . . . . . . . . . . . . . . . . . . . 60 On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Registered trademarks . . . . . . . . . . . . . . . . . . . . . . 10 System
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Endress+Hauser 117
Index PROline Prowirl 72
System condition W
Actual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Weight
Previous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 Flow conditioner . . . . . . . . . . . . . . . . . . . . . . . . . 73
System error (assign) . . . . . . . . . . . . . . . . . . . . . . . 106 Prowirl 72 F . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Prowirl 72 F, Dualsens . . . . . . . . . . . . . . . . . . . . 70
T Prowirl 72 W . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Tag Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Write protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Technical data at a glance . . . . . . . . . . . . . . . . . . . . 57 Values
Temperature coefficient . . . . . . . . . . . . . . . . . . . . . 104 100% value (line 1 and 2) . . . . . . . . . . . . . . . . . . . . 84
Temperature ranges 20 mA value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Ambient temperature range . . . . . . . . . . . . . . . . 60
Medium temperature . . . . . . . . . . . . . . . . . . . . . . 61
Storage temperature . . . . . . . . . . . . . . . . . . . . . . 60
Test display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Time constant
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Status output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Tof Tool-FieldTool Package . . . . . . . . . . . . . . . . . . . 31
Totalizer
Failsafe mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Overflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Sum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Transmitter
Electrical connection . . . . . . . . . . . . . . . . . . . . . . 22
Rotating housing . . . . . . . . . . . . . . . . . . . . . . . . . 17
Transport, sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
U
Unit
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Text arbitrary volume unit . . . . . . . . . . . . . . . . . . 80
Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
V
Value simulation
Current output . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Measurand . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Pulse output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Status output switch point . . . . . . . . . . . . . . . . . . 96
Value 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Version
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . 60
Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Vortex frequency
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
118 Endress+Hauser
'HFODUDWLRQRIFRQWDPLQDWLRQ
Dear customer,
Because of legal determinations and for the safety of our employees and operating equipment we need this
“Declaration of contamination” with your signature before your order can be handled. Please put the completely
filled in declaration to the instrument and to the shipping documents in any case. Add also safety sheets and/or
specific handling instructions if necessary.
:DUQLQJKLQWVIRUPHGLXPXVHG
SAFE
o o o o o o o o
radioactive explosive caustic poisonous harmful to biologically inflammable safe
health hazardous
Please mark the appropriate warning hints.
5HDVRQIRUUHWXUQ
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB
&RPSDQ\GDWD
I hereby certify that the returned equipment has been cleaned and decontaminated acc. to good industrial prac-
tices and is in compliance with all regulations. This equipment poses no health or safety risks due to contamination.
_______________________________ ___________________________________
(Date) (company stamp and legally binding signature)
http://www.endress.com
BA084D/06/en/12.03
50103643
FM+SGML 6.0