Professional Documents
Culture Documents
User's Manual
FREJA 300/ FREJA Win
GE Energy
User’s Manual
for
Programma Electric AB
Eldarvägen 4
SE-187 75 TÄBY
Sweden
Tel +46 8 510 195 00
Fax +46 8 510 195 95
E-mail programma@ge.com
Internet www.geenergy.com
Programma Electric AB
ZP-CF02E R101 Doc.No: CF0280LE 2005 3
FREJA 300 / FREJA Win
Contents
1 Safety precautions .................................... 8 Installation of Freja Win software
version 5.2 .................................................................. 28
2 Instrument panels ................................... 10 Hard disk space needed for installation of
Front panel ........................................................................ 10 Freja Win 5.2 .............................................................. 28
Analog input LOW ................................................. 10 Installation procedure .......................................... 28
Multiconnector ......................................................... 10 Local SW ...................................................................... 29
Rear panel ......................................................................... 11 Software keys ............................................................ 29
Recheck key ................................................................ 29
Extension connector ............................................. 11
Getting ready for use ................................................ 30
CA3 Connector ......................................................... 11
Remote contact ........................................................ 12 Starting Freja Win .................................................. 30
PC contact & Relay contact ............................ 12 User's registration ................................................. 30
Remote control (option) .......................................... 12 Optimizing the system per-formance
running FREJA Win ............................................... 30
Local mode menu ........................................................ 13
FREJA Win Control Center - menu .................. 31
3 How to use Freja 300 – Stand-alone . 14 Configuration ............................................................ 31
V i e w ................................................................................. 32
Local Mode (version R04R and later) ............ 14
Register ......................................................................... 32
General .......................................................................... 14
Import/Export ............................................................. 32
RX (I) ................................................................................ 14
Open ................................................................................ 32
RX (%Z).......................................................................... 14
Start test and save ...................................................... 32
How to navigate in Local mode ......................... 14
New test ........................................................................ 33
Make the settings ................................................... 14
Report ............................................................................. 33
Save the settings .................................................... 15
Second test ................................................................. 33
Change language .................................................... 15
Exit .................................................................................... 34
Change phase name ............................................. 15
Running Freja DOS ..................................................... 34
Main menu .................................................................. 15
How to select the Local version ................... 34
General instrument – Local ................................... 16
Read the measured values .............................. 17 5 General instrument ................................ 36
Set Binary Out ........................................................... 18 Basic ..................................................................................... 36
Special Binary Input - Binary Output fea-
General – Toolbar ........................................................ 37
ture .................................................................................... 19
Testing transducers ............................................... 19 General – Menu............................................................. 39
Set Uref ......................................................................... 20 Config – menu................................................................ 39
RX(I) Instrument............................................................ 21 Test Conditions ......................................................... 39
RX(%Z) Instrument ..................................................... 22 Connect – menu ............................................................ 41
Phase to earth ........................................................... 22 Sequence – menu ........................................................ 42
Application example ................................................. 23 State Settings ............................................................ 42
Testing distance and impedance protection Ramp – menu ................................................................. 45
relays in Local mode ........................... 23 Ramp type ................................................................... 45
Freja 300 together with a current amplifier 24 Start .................................................................................. 46
External Amplifier setup .................................... 24 S t o p .................................................................................. 46
Setting up the ROGOW amplifier Result .............................................................................. 46
(Rogowski) .................................................................. 25 Run Ramp .................................................................... 46
Check list ........................................................................... 47
4 How to use Freja 300 – PC-controlled... Notes ............................................................................... 47
.......................................................................... 28
Programma Electric AB
4 ZP-CF02E R101
FREJA 300 / FREJA Win
Programma Electric AB
ZP-CF02E R101 5
FREJA 300 / FREJA Win
Programma Electric AB
6 ZP-CF02E R101
FREJA 300 / FREJA Win
Programma Electric AB
ZP-CF02E R101 7
1 Safety precautions
1 Safety precautions
When connecting, please note:
Always connect protective earth.
This is normally done via the
cable for the mains supply but can
also be done via the ground terminal on
the rear panel of Freja 300.
Warning:
Do not apply voltage to the outputs.
Important:
Refer all servicing to qualified
service personnel.
Programma Electric AB
8 ZP-CF02E R101
1 Safety precautions
Programma Electric AB
ZP-CF02E R101 9
2 Instrument panels
2 Instrument panels
Front panel
1. Binary inputs 8. DC-supply, connect to (11) to read the
2. Binary outputs (normally-closed and nor- values (in General mode page 5/6 on the
mally-open) display)
3. Display and buttons used in the Local Mode 9. Analog input LOW for measurement trans-
ducers
4. Dial, press for Enter
10.Fuse (50 mA) at the bottom, for Analog
5. Multiconnector for voltage and current
input LOW (9)
6. Current and voltage outputs
11.Analog inputs HIGH for volt- and ammeter
7. Switch, PC to Freja 300 or relay
1 2 3 4
11 10 9 8 7 6 5
1 I in 1- NU
2 I0 2+ NI
3 NC 2- L3U
4 U0 3+ L3I
5 U in 3- L2U
4+ L2I
4- L1U
Programma Electric AB
10 ZP-CF02E R101
2 Instrument panels
Rear panel
1. Mains supply ON/OFF 5. Remote control
2. Connector for mains supply 6. Relay connection
3. Extension connector (used for the calibration) 7. Serial computer interface, to the serial port
4. Connector for external current amplifier (PC)
CA3, CA1 or CA1H 8. Ground (earth) terminal
1 2 3 4 5
85-264V AC
47-63Hz EXT. CA3
600VA
REMOTE
CONTROL
PC RELAY
Programma
8 7 6
Programma Electric AB
ZP-CF02E R101 11
2 Instrument panels
2
PC contact & Relay
contact
3
PC Relay
Freja 300 Signal
Contact pin Contact pin
1 GND 1 1
2 RX 2 2
3 TX 3 3
5 0V 5 5
Programma Electric AB
12 ZP-CF02E R101
2 Instrument panels
Programma Electric AB
ZP-CF02E R101 13
3 How to use Freja 300 – Stand-alone
You can run three instruments in Local mode If you press you go to a higher level
in the menu tree.
General
Above button you can read 1/6, this
The General instrument is a ”toolbox” where you
means that you are on page 1 of 6 in
can change the amplitude and phase angles for each
this part of the menu tree. If you press
generator, you can also change the frequency. It is
you will jump to next page, 2/6.
possible to generate 2 different states, 1ST
(Prefault) and 2ND (Fault). You can also ramp
between the states. Above button you can read Start. If
you press it you will start to generate
the values you have set on this page.
RX (I)
The RX (I) instrument is designed for testing The button is used to toggle be-
distance relays in the RX plane. It uses a fixed
current and the built-in computer calculates the tween SET (Set values) and MEAS
voltage and phase angles for each generator. (Show measuring values). More about
this button in section ”Read the meas-
ured values”.
RX (%Z)
There is no any text above button ,
The RX (%Z) instrument is also used for testing
distance relays, but here you define the line with the but by pressing this you always go to
values for the line and source impedance. Then you previous page. In this example you
vary the percentage of Z from 0-100%. 100% means would come to page 6/6. (You go back
the end of the line and 0% the beginning. In this from first page and come to last page) .
way it is easy to make a quick check of the relay by
just testing in a few points along the line. This test
is good for routine testing.
Make the settings
1. Turn the Dial to move the marker < > to
the field that you want to change.
Programma Electric AB
14 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
2. Press the Dial. more digit after the decimal (0.00 - 89.00, 90.0 -
150.0 V, 0.000 - 8.999, 9.00 - 15.0 A).
3. Turn the Dial again, whereupon the values
shown on the display will change. The Dial Default is set to "No" because "Yes" also means
senses the speed at which it is turned. that you need to turn the dial ten times more to
When you turn the Dial slowly, the values make the settings.
are changed in small increments. As you
turn the Dial faster, the increments become
Memory
larger.
Freja 300 automatically saves the last settings in the
4. Press the Dial to enter the desired value.
memory. You can also save settings in one of the 15
5. Turn the Dial to the next field, and repeat memory places whenever you wish.
the procedure. Then turn to next field, etc.
Save If you select a place in the memory
that already has a content it will be
Save the settings replaced with the new one.
Main menu
System setup
Programma Electric AB
ZP-CF02E R101 15
3 How to use Freja 300 – Stand-alone
Programma Electric AB
16 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
* =
0 =
– = or
Programma Electric AB
ZP-CF02E R101 17
3 How to use Freja 300 – Stand-alone
Page 5/7
Set Binary Out
The time for the Binary outputs are defined here,
you can set up to 4 events, one event on each
horizontal line. A ”*” (star) means that you activate
an output. On the panel you can select if you want
to use the normally open or the normally closed
contact. The Binary outputs will start to work when On the Mode page you can see the measured values
you press “ Start” in the “Sequence” menu, page from the analoge inputs, high or low. This is useful
3/6. if you want to measure the voltage of the
DC supply. You can also select AC, DC or OFF
output for the generators.
To measure the voltage of the DC supply
1. Connect the cables externally.
2. Change the measurement from AC to DC.
Page 4/6
Programma Electric AB
18 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
Programma Electric AB
ZP-CF02E R101 19
3 How to use Freja 300 – Stand-alone
U-ref t
L1U
10°
L2U
250°
t
U-ref L3U
130°
L1U
360° (0)
Note! In Local Mode the normal settings is 0°,
240°, 120°. In Freja Win you can select
L2U 0°, 240°, 120° or 0°, 120°, 240° as
240°
phase rotation.
L3U
120°
Programma Electric AB
20 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
Page 2/3
2N D 5 0. 00Hz I: <1.000> U: 45 .0 V
* - - V OL T R: 45.000 Z: 45. 00 0
o o o -- -- --ms X: 0.000 Z ϕ : 0.0
2/3 S tart RST Run: S eq
Programma Electric AB
ZP-CF02E R101 21
3 How to use Freja 300 – Stand-alone
Phase to earth
Phase to phase
Programma Electric AB
22 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
2N D 5 0. 00Hz I: <1.000> U: 45 .0 V
* - - V OL T R: 45.000 Z: 45. 00 0
o o o -- -- --ms X: 0.000 Z ϕ : 0.0
2/3 S tart RST Run: S eq
Programma Electric AB
ZP-CF02E R101 23
3 How to use Freja 300 – Stand-alone
As L3U below
L2I:
L3I:
As Φ2 below
As Φ3 below
Programma Electric AB
24 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
Programma Electric AB
ZP-CF02E R101 25
3 How to use Freja 300 – Stand-alone
Programma Electric AB
26 ZP-CF02E R101
3 How to use Freja 300 – Stand-alone
Programma Electric AB
ZP-CF02E R101 27
4 How to use Freja 300 – PC-controlled
Installation procedure
1. Insert the FREJA Win CD. It will auto start (or
run frejinst.exe).
2. Select the language to use during installa-
tion.
3. Next you can choose to either update the
FREJA Local (internal software) or to install IMPORTANT!
the FREJA Win (or cancel the installation). You must install “MDAC” once on every
computer.
Programma Electric AB
28 ZP-CF02E R101
4 How to use Freja 300 – PC-controlled
Programma Electric AB
ZP-CF02E R101 29
4 How to use Freja 300 – PC-controlled
This icon will appear if: By doing this we can give you a better support and
keep you informed on new software releases.
• No Freja is connected
• If you try to start Freja Win before Freja start- Optimizing the system per-
up is ready ( you should hear a beep - beep
when Freja is ready) formance running FREJA Win
• To be able to communiucate again, disable The performance of the computer depends not only
this “off-line-mode”. Click on the icon or on the CPU clock speed, it is also dependent on the
unmark the “off-line” in ”Configurations”. CPU type, hardware configuration and other soft-
ware components (power saving, disk caching etc.)
in the system.
Lower CPU performance needs more support from
This icon shows that FREJA 300 is in online mode and is
other components to get better system performance!
ready to communicate. You can switch to offline mode by Here are some adjustments you can do to get better
clicking on it.
system performance running FREJA/Win:
Note! If the communication between FREJA and • Don't use disk compression.
the PC is not working correct, please se • Don't use power saving.
the following fault list.
• Don't use network connection.
• Don't use screen saver.
Fault list
• Don’t use IR-communication.
Fault Freja Win is set to ”Off-line”.
• Minimise graphic acceleration.
What to do Click on the icon in the • Minimise disk cache.
FREJA Win Control Center. • Adjust COM port FIFO buffer to low.
Fault No FREJA connected. • Antivirus SW can decrease system performance.
What to do Connect FREJA to PC, or keep on if you The “System Monitor”, click Start/Program/Acces-
just want to work with the FREJA Win sories/System Tools/System Monitor, can help you
without the FREJA.
visualise CPU workload. If it is 100% most of the
Fault FREJA was not ready with the start-up time the communication might hang up. Try to close
procedure. other programs or use a faster PC.
Programma Electric AB
30 ZP-CF02E R101
4 How to use Freja 300 – PC-controlled
Options - menu
Configuration
Diagnose
The Diagnose will start Doctor FREJA. This will
result in a file “Diagnose.txt” saved under
C:\PETOOLS\FREJAWIN. In this file you can get
information about the system and about the SW-
keys loaded into the FREJA. You can e.g. see which
instruments are NOT possible to run.
Key load
Use this to download new SW-keys. If you have
bought a new FREJA from Programma the
Phase names
download is already done. This function is only
Here you can change the default phase
names. used if you want to add some SW-key (e.g. for
Transient)
Programma Electric AB
ZP-CF02E R101 31
4 How to use Freja 300 – PC-controlled
Recheck key
In some special cases it can be necessary to use the Start test and save
function ”Recheck key”.
1. Start the instrument you want to use. For
Example: more information see the respective instru-
You have SW-keys only for e.g. General and ment. Fill in the report header information
Distance in a certain FREJA (No.1) and you use this for the test object.
FREJA with your PC. If you connect your PC to
another FREJA (No.2), which has the SW-keys for Note! If one or more instruments (e.g. Trans-
e.g. General, Distance, Sync, Current, and Trans- ducer) is not possible to select (the text
ducer you will only be able to use General and is gray) you have not the software key
Distance, because the PC “remember” that last time for that instrument. See section “Soft-
it was only these instruments available on FREJA ware keys” in this manual. If you do not
(No.1). To solve this problem you use the function want an instrument to be displayed you
”Recheck key”. FREJA (No.2) will then ”tell” the can in make this selection in the “View”
PC that there are more instruments available. menu.
Import special keys 2. Make the test in e.g. Distance, when you
Use this to import special keys, e.g. Trial keys. are ready you can make a printout direct
from Distance or do it later on.
3. Close Distance.
View 4. Go back to the Control Center.
Here you select which instrument to be shown.
5. Click ”Save test”, check the information for
the report header.
Register
Here you can fill in the User´s registration. Note! It is important that you set the correct
information here, all your reports will
have this information in the header. This
Import/Export also includes the date and time, press
This menu option will only be visible if you have “Get system time” to update the Date
the software “WinZip” with “Command Line Add- and Time fields.
On” installed on your computer.
This feature makes it easy to export/import
complete tests.
Open
Under the “Open” menu is "FREJA local
maintenance" which has a number of entries, e.g.
“Download software”, “Upload FREJA test
settings memory”, “Set FREJA Date & Time” and
make a “FREJA Calibration Report”.
Programma Electric AB
32 ZP-CF02E R101
4 How to use Freja 300 – PC-controlled
Report
1. To make a report on an old test, click “Open
Test”.
2. Select the directory where you have saved
the test.
3. Click the printer button when the instrument
has opened.
7. Now you will see the directory structure of
your PC. You can save the test wherever Note! In ”Configuration” you can select print
you want, you decide the file structure out in colour or black & white.
yourself. Create the structure with the
Windows Explorer or use the “Create Direc-
tory” button. Second test
8. Select the directory where to save, click
If you want to test the same object again e.g. one
“Proceed”. On the screen you will see the
year later.
content of the selected directory, to help
you avoid overwriting an old file. 1. Click “Open test” and select the test direc-
tory.
2. The instrument e.g. General will open with
Note! A Freja icon will be shown on each the settings and results from last test.
directory containing a Freja test.
3. Click the button “New test”. This will delete
the old test results, but the settings will
remain.
New test
1. To make a new test in General, click the
General button and wait until General have
started. 4. Make the test.
5. Close the instrument.
6. Go back to the Control Center and click
“Save”.
2. Start a new test by clicking the button “New
test”. This will delete the old test results, 7. Click “Get system time” to update date and
but the settings will remain. time.
Programma Electric AB
ZP-CF02E R101 33
4 How to use Freja 300 – PC-controlled
Programma Electric AB
34 ZP-CF02E R101
4 How to use Freja 300 – PC-controlled
Programma Electric AB
ZP-CF02E R101 35
5 General instrument
5 General instrument
Basic
The General Instrument is like a big ”toolbox” were
you can control the generators as you like in ampli-
tude, phase angle and frequency. The General
instrument is useful for many different types of
relays. Under the first tab you find the Config
submenu.
Programma Electric AB
36 ZP-CF02E R101
5 General instrument
General – Toolbar
Relations Set the relations in amplitude and
phase angle.
By clicking the button "Standard" the
following will be set as default values.
Amplitude: Equal voltage and current
New test Start a new test by clicking the button magnitudes.
“New test”. This will delete the old test Phase: Phase displacement is set as a
results, but the settings will remain. balanced 3-phase system.
By clicking the button "None" you can
change only the values required for the
specific element under test.
Report Print -out of a report, see section
Report
Programma Electric AB
ZP-CF02E R101 37
5 General instrument
Analogue inputs
Shows the measurement of the ana-
logue inputs. (Not available in Ramp.)
Harmonic curve
Here you can create a harmonic curve.
Up to the 25th harmonic and up to 5 Trash can Using the trash can icon you can delete
components. You can save up to 6 or undo delete of any trip results.
different curves in one test. Here you
Programma Electric AB
38 ZP-CF02E R101
5 General instrument
Programma Electric AB
ZP-CF02E R101 39
5 General instrument
Programma Electric AB
40 ZP-CF02E R101
5 General instrument
Connect – menu
In the “Connect” submenu you can write informa-
tion on how to connect the relay. This information
will be saved together with the object. This means
that the connection information is ready next time
you shall test this object.
You can copy information, with text and pictures,
from another program e.g. from Word and then
paste it here with the command <Ctrl> + <V>.
Programma Electric AB
ZP-CF02E R101 41
5 General instrument
Sequence – menu
State Settings
Programma Electric AB
42 ZP-CF02E R101
5 General instrument
If the “Stop if Goal” is not marked, the program relay. Grey means open contact or no voltage from
will jump to next state, if more than 2 states are the relay.
selected.
The top right lamp shows the Binary Goal.
In this example “Stop if Goal” is grey because it is
When Binary Goal = status, the goal is reached and
in the last state 2 of 2 and then it will stop at goal.
the timer stops (or jumps to next state, see above).
If you want it to continue generate after goal in this
case you can add one more state with the same
Binary Goal
settings.
Grey = Open
You can make the settings in five ways:
Red = Closed
A. Click in a field, type in the new value, press
<Enter> Red/grey = Change, from open to close or from
close to open (stop on any event).
B. Click in a field, type in the new value, press
<Tab> Black = Not used, this means this binary input is not
C. Click in a field, turn the dial on Freja 300. used for the moment and is not used to check if
Binary goal = status.
D.Using the mouse, drag the vector of any
generator. The most common BI Goal is “Close”.
E. Click in the field, change the values using the
arrow keys, up /down. Note! The Binary Goal is an AND function, this
This can also be done while generating (if manual means all signals on the Binary Inputs
mode is selected). must come before the goal has reached.
But you can use Black = Not used, this
When the setting of state 1 is ready, change to state signal is not used to check if
2 and make the setting for it. Binary Goal = status.
Programma Electric AB
ZP-CF02E R101 43
5 General instrument
Vector diagram
The vector diagram shows the amplitude and phase
angle for each generator. Red for phase 1, yellow for 2. Back in the "Sequence" menu, click the
phase 2, blue for phase 3. The thicker vectors are for "Frequency" button and set the frequency
the currents. If you right-click the mouse you can for each generator.
change the colours of the vectors and background. You
can drag the vectors using the left mouse button and
then change the amplitude and phase angle. The
number in the upper left corner shows the actual state.
Programma Electric AB
44 ZP-CF02E R101
5 General instrument
Ramp – menu
Ramp type
Programma Electric AB
ZP-CF02E R101 45
5 General instrument
Start
Set the start value of the ramp.
Stop
Set the stop value of the ramp.
Run Ramp
1. When all settings are done, click button
"START".
Programma Electric AB
46 ZP-CF02E R101
5 General instrument
Check list
Notes
In the ”Notes” menu you can make notes (but only
as text, no pictures) about the specific test e.g. if
some part of the relay is broken and need repairs.
This information will then be printed in the report.
Programma Electric AB
ZP-CF02E R101 47
6 Distance instrument
6 Distance instrument
Basic
The Distance instrument is designed for testing
distance relays. Distance contains Zt for time test
and RX-Ramp and RX-point for testing the reach,
whirlwind and RX for manual/semiautomatic test in
the impedance plane.
Programma Electric AB
48 ZP-CF02E R101
6 Distance instrument
Distance – Toolbar
BO Set the Binary outputs manually.
Programma Electric AB
ZP-CF02E R101 49
6 Distance instrument
RIO Converter
The RIO converter converts RIO files to FREJA
reference graphs. RIO files contains information
about the settings of the relay. You can get the RIO
files out of e.g. SIEMENS ”Digsi”, and some other
relays (mostly in Germany).
1. Click this button to import the RIO-file into
FREJA.
Programma Electric AB
50 ZP-CF02E R101
6 Distance instrument
Programma Electric AB
ZP-CF02E R101 51
6 Distance instrument
Note! If you want to make a new test or need If the trip signal is a wet contact you
to change the settings in the “Config”
select voltage sense with this button.
menu after you have started a test you
must first click the button “Start a new BO Here you select which binary outputs to
test” to clear the data in log list. use. You can also type in a name for
each Binary output.
Current Direction
Change the direction with 180°, forward
or backward direction.
Programma Electric AB
52 ZP-CF02E R101
6 Distance instrument
Connect – menu
Check this box and set the value if your
relay setting has the definition of Z
related to the nominal current.
Ground factor
The ground factor can be defined in two
ways. One way is to define Ke and the
other way is to unmark “Use standard”
and define Re/RI and Xe/XI.
lKel Ground fault factor, used in the network
models. This factor is used for some
manufactures distance relays to get the
1-phase fault values into ohm/phase. If
Ke = 0 the 1 phase fault is presented
as ohm/loop.
Kphi The ground factor angle.
In the “Connect” submenu you can write information on
Re/RI Xe/Xl how to connect the relay. This information will be saved
This is an alternative setting possibility together with the object. This means that the connection
to lKel and Kphi that is used by some information is ready next time you shall test this object.
protection manufacturers. Hereby the You can copy information, with text and pictures, from an
protection's ground impedance determi- other program e.g. from Word and then paste it here with
nation is done via resistance and the command <Ctrl> + <V>.
reactance ratio. In that case the protec-
tion settings must be entered here.
This setting possibility should only be
used when the protection to be tested
is using this definition. In all other
cases lKel and Kphi should be used.
Programma Electric AB
ZP-CF02E R101 53
6 Distance instrument
RX-Ramp – menu
Programma Electric AB
54 ZP-CF02E R101
6 Distance instrument
Tolerance lines
Programma Electric AB
ZP-CF02E R101 55
6 Distance instrument
Results
The result for each fault type can be seen in the 3. Click on the Ramp Start button
“BI” list by clicking on the BI list ikon. You will
only see one fault type in each list. 4. Use the mouse to get the position (in the
oscilloscope) for the start of the ramp, then
Report click the left mouse button.
Report, see section Report. 5. If you want the ramp to stop some where
else than the origin.
Paus
You can make a paus in the testing by clicking the Click on the Ramp Stop button
Programma Electric AB
56 ZP-CF02E R101
6 Distance instrument
Run ramps
A. You can select one ramp to run by clicking
on the ramp, it will then be red, or by using
the arrows at the bottom of the “Ramp
Points” window.
B. To run all the ramps. Unmark all marked
ramps by clicking somewhere in oscillo-
scope without marking any ramp.
Origin offset
To move the origin outside of the oscilloscope type
in the desired values.
To move the origin inside the oscilloscope use the
“Adjust axis” in the toolbar.
Programma Electric AB
ZP-CF02E R101 57
6 Distance instrument
RX Point – menu
Programma Electric AB
58 ZP-CF02E R101
6 Distance instrument
RX – menu
Programma Electric AB
ZP-CF02E R101 59
6 Distance instrument
Programma Electric AB
60 ZP-CF02E R101
6 Distance instrument
Fault
If you want to use the dial and change the value
while generating you can not use the sequence
mode, you must select the Fault mode. In some cases you might want to save the position
of the marker even if the relay do not send a trip
Import ramp results signal. This will be shown as a blue marker both in
To import Ramp Results, click this button: the oscilloscope and in the binary list.
Log list
With the log list you can jump between the test
points to view the results or to delete this test points
from the results that will be shown in the BI list and
the report. For report, see section Report.
Programma Electric AB
ZP-CF02E R101 61
6 Distance instrument
Whirlwind – menu
Programma Electric AB
62 ZP-CF02E R101
6 Distance instrument
Zt – menu
Programma Electric AB
ZP-CF02E R101 63
6 Distance instrument
Programma Electric AB
64 ZP-CF02E R101
6 Distance instrument
Zoom
If you want to zoom some specific details:
1. Decrease the scale.
2. Change the origin offset.
3. If necessary also increase the number of
test points/division.
4. Click “Start”.
Results
To view the results for the fault type shown in the
oscilloscope, click “BI list” on the toolbar.
Report
See section Report.
Programma Electric AB
ZP-CF02E R101 65
6 Distance instrument
Power Swing
Blocking (PSB) –
menu
The PSB page is designed for testing the Power 4. Click the ”Start” button.
Swing Blocking function. The fault type is only 3-
phase (ABC). Note! If any ramp is marked (red) when you
1. Import a reference graph (if you want). click ”Start”, this is the only ramp that
will run. If you want to run all ramps you
must first unmark the ramp by clicking
2. Add a new ramp, using the icon on the somewhere else in the black oscillo-
toolbar. scope.
3. Set the speed of the ramp ”dZ/dt”, or the
”Delta time”. Note! The PSB function is not included in the
If you set the ”PSB out” and the ”PSB in” you can report.
see how long time ”PSB delta” will take to pass
these borders with the speed ”dZ/dt”. Normally you
set ”PSB out” and ”PSB in” to the borders of the
relays setting for PSB.
Programma Electric AB
66 ZP-CF02E R101
6 Distance instrument
Check list
Notes – menu
In the ”Notes” menu you can make notes about the
specific test e.g. if some part of the relay is broken
and need repairs. This information will then be
printed in the report.
Programma Electric AB
ZP-CF02E R101 67
7 Sync instrument
7 Sync instrument
Basic
Programma Electric AB
68 ZP-CF02E R101
7 Sync instrument
Programma Electric AB
ZP-CF02E R101 69
7 Sync instrument
Pulses
Programma Electric AB
70 ZP-CF02E R101
7 Sync instrument
Synchro check –
menu
Programma Electric AB
ZP-CF02E R101 71
FREJA 300 / FREJA Win
8 Current instrument
Basic
The current instrument is designed for testing over
current relays. Here in the config menu you set up
the conditions for the test.
Programma Electric AB
72 ZP-CF02E R101
8 Current instrument
Current – Toolbar
Clear the results
This will clear the results from the
selected fault type. If you want to clear
the result from all fault types, use the
”Clear all results” button.
Binary out Set the Binary out list. Edit/create a reference graph
Edit/create a reference graph
Analog inputs
Show the measured values from the
analog inputs.
Programma Electric AB
ZP-CF02E R101 73
FREJA 300 / FREJA Win
Pass/Fail settings
Set up the tolerance, in percent, for the time test and
as an absolute minimum value in milliseconds. The
time is compared with the reference graph. Set up
the I>, the tolerance for the pick-up and the hyster-
esis.
Test of electromechanically
relays
In the "Config" page you set the following param-
eter values.
Principle of operation
Pickup test will start with an initial energizing
phase. The Tap setting current multiplied with the
Ramp multiplier value will be the initial current
during the specified time.
Then the current level will be 1.5 times the Tap
setting stepping down towards 0.5 times the Tap
setting (in 25 steps). The duration of each step is set
in the pickup page, "Wait time (ms)".
Programma Electric AB
74 ZP-CF02E R101
8 Current instrument
In this menu you test the pick-up and dropout In the time test you will test the trip time of the
values, then the program calculate the hysteresis. overcurrent relay and display the result with current
The test is done automatically. and time on the axis.
Programma Electric AB
ZP-CF02E R101 75
FREJA 300 / FREJA Win
Programma Electric AB
76 ZP-CF02E R101
8 Current instrument
Programma Electric AB
ZP-CF02E R101 77
9 Voltage instrument
9 Voltage instrument
Basic
The voltage instrument is designed for testing over/
under voltage relays. Here in the config menu you
set up the conditions for the test.
Programma Electric AB
78 ZP-CF02E R101
9 Voltage instrument
Voltage – Toolbar
Clear the results
This will clear the results from the
selected fault type. If you want to clear
the result from all fault types, use the
”Clear all results” button.
Binary out Set the Binary out list. Edit/create a reference graph
Edit/create a reference graph
Analog inputs
Show the measured values from the
analog inputs.
Programma Electric AB
ZP-CF02E R101 79
9 Voltage instrument
Pass/Fail settings
Set up the tolerance, in percent, for the time test and
as an absolute minimum value in milliseconds. The
time is compared with the reference graph. Set up
the U>, the tolerance for the pick-up and the hyster-
esis.
Test of electromechanically
relays
In the "Config" page you set the following param-
eter values.
Principle of operation
Pickup test will start with an initial energizing
phase. The Tap setting voltage multiplied with the
Ramp multiplier value will be the initial voltage
during the specified time.
Then the voltage level will be 1.5 times the Tap
setting stepping down towards 0.5 times the Tap
setting (in 25 steps). The duration of each step is set
in the pickup page, "Wait time (ms)".
Programma Electric AB
80 ZP-CF02E R101
9 Voltage instrument
Programma Electric AB
ZP-CF02E R101 81
10 Frequency instrument
10 Frequency instrument
Basic
The frequency instrument is designed for testing
frequency relays. To use the frequency instrument,
you set a fixed voltage and vary the frequency,
either slowly to check the frequency limit or by
switching rapidly from a prefault frequency to a
faulty frequency to measure tripping time.
Here in the Config menu you set up the conditions
for the test.
Programma Electric AB
82 ZP-CF02E R101
10 Frequency instrument
Frequency – toolbar
Clear the results
This will clear the results from the
selected fault type. If you want to clear
the result from all fault types, use the
”Clear all results” button.
Binary out Set the Binary out list. Edit/create a reference graph
Edit/create a reference graph
Analog inputs
Show the measured values from the
analog inputs.
Programma Electric AB
ZP-CF02E R101 83
10 Frequency instrument
Sequence
1. Select “Sequence”.
2. Select state 1 and set the Prefault values.
Select state 2 and set the Fault values. Do
not forget to set BI-goal for each state.
3. Connect the relay between phase-neutral (L-
N) or phase-phase (L-L) depending on the
voltage level.
4. Click “Start”.
Manual
1. Select “Manual”
2. Select the settings from state 1 or state 2
3. Connect the relay between phase-neutral (L-
N) or phase-phase (L-L) depending on the
voltage level.
4. Click “Start”
Programma Electric AB
84 ZP-CF02E R101
10 Frequency instrument
In the report (click the Printer icon) you will see all
the results.
Programma Electric AB
ZP-CF02E R101 85
10 Frequency instrument
Ramp – menu
The “Ramp” menu is used to search for a frequency
border.
1. Set the voltage to generate.
2. Connect the relay (Line-Line or Line-Neutral,
depending on the required voltage). Con-
nect the BI signal to a contact without any
delay.
3. Set Ramp settings.
4. Adjust the scale.
5. Set BI-goal.
6. Connect the relay between phase-neutral (L-
N) or phase-phase (L-L) depending on the
voltage level.
7. Click “Start”.
Result
If you run more than one test you can jump between
the results using the arrows in the log.
In the report (click the Printer icon) you will see all
the results.
Programma Electric AB
86 ZP-CF02E R101
10 Frequency instrument
Programma Electric AB
ZP-CF02E R101 87
11 Transducer instrument
11 Transducer instrument
Basic
Transducers are used to measure e.g. current,
voltage, power, phase angle or frequency. The
output from the transducer is either a DC voltage or
a DC current. Normal ranges are 0-10 V or 4-20
mA, and in some cases also 0-1 mA.
Programma Electric AB
88 ZP-CF02E R101
11 Transducer instrument
Programma Electric AB
ZP-CF02E R101 89
11 Transducer instrument
Results
When the test is ready you can display the result as:
• “Full Scale error” i.e. the fault in % compared
to the Full scale value.
• ”Relative error”, i.e. the fault in % relative to
each test point.
• ”Absolute error”, i.e. the fault in V or mA
relative to each test point.
Report
You can see the result of the test in table by clicking
on the binary list icon.
Programma Electric AB
90 ZP-CF02E R101
11 Transducer instrument
Programma Electric AB
ZP-CF02E R101 91
12 Autoreclose instrument
12 Autoreclose
instrument
Basic
The autoreclose instrument is designed for testing of
relays with built in autoreclose and synchrocheck
function. Here in the configuration menu you set up
some of the start conditions for the test.
Programma Electric AB
92 ZP-CF02E R101
12 Autoreclose instrument
Autoreclose –
Graphic Graphic display of binary events and
Toolbar wave forms.
Analogue inputs Log List of all trips. If you use the arrows
Shows the measurement of the ana- and change to another number, the
logue inputs. setting values on the screen will also
change to the value you had at that trip.
Programma Electric AB
ZP-CF02E R101 93
12 Autoreclose instrument
Config – menu
Network Model
You should start by selecting a network
model since the protective relay system
will normally be set for ohms per Ground Factor
phase, but Freja must generate ohms The ground factor can be defined in two
per loop in order to supply the system ways. One way is to define Ke and the
correctly. This means that a factor of 2 other way is to unmark “Use standard”
must be included (for 2-phase faults). and define Re/Rl and Xe/Xl. For more
Either by generating half the current or information see ground factor under
double the voltage. For more informa- Distance instrument config menu.
tion see network model under Distance
instrument config menu.
Correction factor
Check the box if your relay settings
have the definition of Z related to the
nominal current.
Time You can select ms or ccs (cycles, not
implemented yet). BI (1-5), BI (6-10)
Here you select which binary inputs to
use. You can also type in a name for
each Binary input.
Max settings
These are the maximum values that will
be generated.
Programma Electric AB
94 ZP-CF02E R101
12 Autoreclose instrument
Connect - menu
Programma Electric AB
ZP-CF02E R101 95
12 Autoreclose instrument
Reclose – menu
Number of states
Set the number of states you will
generate
Programma Electric AB
96 ZP-CF02E R101
12 Autoreclose instrument
Application example –
Distance relay with
synchrocheck function
By using Impedance you can set the type of fault 1 Vector diagram, used 10000
prefault values
and your impedance values. You can either type the
such as normal
values or use the mouse and click in the Impedance load conditions
plane for your test point. You can also choose what
2 Impedance used 2000 close
state that should be your reference time (the clock diagram, fault
will start counting from 0 in this state). If you are values in zone 1
testing relays with synchrocheck function you can 3 Circuit breaker used 20-40
use Uref. For Uref it’s possible to change amplitude, opening time.
phase angle and frequency. Impedance
diagram, fault
values in zone 1
4 Vector diagram, used 200-500
zero state (line
not energized)
5 Vector diagram, used 2000 close
only voltage
applied and with
the same
amplitude, phase
angle and
frequency as U
ref
6 Circuit breaker used 60-150
State settings/Vector closing time.
Vector diagram,
By using Vector you can set the type of fault the only voltage
same way as in the General instrument. Type in applied and with
your values, it’s easier if you use the relation the same
amplitude, phase
function. You can also choose what state that should angle and
be your reference time (the clock will start counting frequency as U
from 0 in this state). If you are testing relays with ref
synchrocheck function you can use Uref. For Uref 7 Vector diagram,
it’s possible to change amplitude, phase angle and postfault
frequency. conditions such
as normal load.
Programma Electric AB
ZP-CF02E R101 97
13 Auto 300 instrument
Programma Electric AB
98 ZP-CF02E R101
13 Auto 300 instrument
Copy
You can use “Copy” instead of “Add Test” if you
want to use the same test directory, by use an other
tab in this instrument.
1. Select the “Tab” to copy
2. Click “Copy”
3. The new “Tab” will now be placed in the
end of the list.
Programma Electric AB
ZP-CF02E R101 99
13 Auto 300 instrument
Fail
Pass
Not tested
Print
This will open the Print dialogue, where you can
select:
• Design – design the first page in the
AUTO300 report. For more info how to do,
see section “Quick guide for report
generator”.
Programma Electric AB
100 ZP-CF02E R101
13 Auto 300 instrument
Report
configuration
Programma Electric AB
ZP-CF02E R101 101
13 Auto 300 instrument
Advanced settings
Extra parameter
Here you can send an extra parameter. This can be
useful if your own program needs special command
lines (switches). Example “/read only”
Programma Electric AB
102 ZP-CF02E R101
13 Auto 300 instrument
Programma Electric AB
ZP-CF02E R101 103
13 Auto 300 instrument
Programma Electric AB
104 ZP-CF02E R101
13 Auto 300 instrument
Programma Electric AB
ZP-CF02E R101 105
14 Transient instrument
14 Transient instrument
Basic
Programma Electric AB
106 ZP-CF02E R101
14 Transient instrument
Generators
In the drop down menu you can select “AC”, “DC”
or “OFF”. This is only used for the prefault or post
fault states when you select “Sine”. It is possible to
set a maximum generated value in the right fields.
BI (1 – 5) & BI (6 – 10)
Here you select which binary inputs to use in the 1-5
group and the 6-10 group. You can also type in a
name for each Binary input e.g. Trip. You select
contact sense or voltage sense with the button.
BO
Select which binary outputs to use. You can also
type in a name for each Binary output.
Programma Electric AB
ZP-CF02E R101 107
14 Transient instrument
Converter – menu
Converter
Before you can generate the transient file with
FREJA you must convert it into a format that
FREJA can understand. It will be converted to SDF
files (Sample Data Files).
Select the converter you want to use.
Name Works with files
Comtrade .DAT and .CFG
Ascii .DAT
Emtp .PL4
Programma Electric AB
108 ZP-CF02E R101
14 Transient instrument
Ascii
Ascii format is normally a text file. The columns
represent the different generators and the lines
represent each sample (up to 2000). The text file
must only contain numbers with data. If the text file
contains additional text information as in the picture
below (the text marked), you can use a normal text
editor to delete it. Then save it with extension DAT.
Programma Electric AB
ZP-CF02E R101 109
14 Transient instrument
Programma Electric AB
110 ZP-CF02E R101
14 Transient instrument
Sequence – menu
Result
Graphic
1. Click the icon “a graphic presenta-
tion…” in the toolbar to show the graphical
result.
Programma Electric AB
ZP-CF02E R101 111
15 AUTO21 instrument
15 AUTO21 instrument
Basic Config – menu
Test conditions
Edit testplans
“Make a PAUSE between the tests”
You must edit the testplans in FREJA RTS11, 21,
When this is marked the program will stop and
21D or FREJA 300 DOS. It is not possible to edit
make a pause after each test is ready. This can be
the testplans in AUTO21.
good if you want to look at the graphical picture
before you continue with next test.
“Do not show the INFO part between tests”
If this is marked the program will not show the info
form (Type I) between the tests.
“Stop if Goal”
For T-POINT test. If this is marked the program will
stop generate at BI-GOAL, if it is NOT marked it
will continue to generate until Tmax (from the tgr-
file). By default it is marked otherwise it will heat
up thermal relay too much.
“Stop after first FAIL”
If this is marked the program will stop after the first
test is failed.
“Restart the selected test if FAIL”
If this is marked the program will rerun a test that
fail until the test is OK. The report will contain only
the last test.
Programma Electric AB
112 ZP-CF02E R101
15 AUTO21 instrument
Connect – menu
This page is used to describe how to connect etc.
Normally this is already described in the DOS
testplans.
Programma Electric AB
ZP-CF02E R101 113
15 AUTO21 instrument
TestPlan – menu
Test conditions
• Start from the beginning
• Start from selected position
• Selected only
Programma Electric AB
114 ZP-CF02E R101
15 AUTO21 instrument
Start
1. Select “Test conditions”
2. Click “START”. The test will run the same
way as with FREJA DOS. When a form
opens up read it and if nessesary write
comments.
3. Close it with by clicking the window “X”.
View results
When the test is ready you can view the results.
1. Click on a specific test group. The graphical
picture will then be shown for this group. If
any testpoint fails it will be marked with a
red square otherwise a green circle.
2. To look at the result in a tabular form, click
on the BI-list icon in the toolbar.
Programma Electric AB
ZP-CF02E R101 115
15 AUTO21 instrument
Complete report
1. Select “Test conditions”, e.g. “Selected
only” will print only the selected test.
2. Click the printer icon to get a complete
report of all test groups. If a testpoint has
failed it will be marked with a “F”.
Programma Electric AB
116 ZP-CF02E R101
15 AUTO21 instrument
Programma Electric AB
ZP-CF02E R101 117
16 Reference graph
16 Reference graph
Reference graph –
Toolbar Clear tolerances
Clear tolerance curves
Grid on/off
About Shows information on the reference
graph
Create zone
Show tolerances
Hide or show tolerances
Programma Electric AB
118 ZP-CF02E R101
16 Reference graph
Manual method
1. Click on the icon “Edit create reference
Recommended sequence for entering elements
graph” on the toolbar.
This will open the “Reference graph editor”
Programma Electric AB
ZP-CF02E R101 119
16 Reference graph
You can select different types of elements (LINE, CIRCLE VIA 0: This is used if you want a circle,
CIRCLE, LENS etc.) using the icons to the left or in which goes through the origin. Define it with
the “Element” box. ”VER” and “HOR”.
LINE: Defined as an angle and a point on horizon-
tal or vertical axis.
Programma Electric AB
120 ZP-CF02E R101
16 Reference graph
LENS: Define this with Alpha, “a”, “b” and “d”. Example 1
If we want to create a reference graph like this:
Second element 2:
4. Set “Alpha” to 75, press <Enter>
5. Set “Hor” to 4, press <Enter
6. Click “Save element”
Third element 3:
7. Set “Alpha” to 180, press <Enter>
8. Set “Ver” to 3, press <Enter>
9. Click “Save element”
Fourth element 4:
10.Set “Alpha” to 255 or 75 press <Enter>
Programma Electric AB
ZP-CF02E R101 121
16 Reference graph
Programma Electric AB
122 ZP-CF02E R101
16 Reference graph
Programma Electric AB
ZP-CF02E R101 123
16 Reference graph
Programma Electric AB
124 ZP-CF02E R101
16 Reference graph
Editing/adding tables
Use the buttons "Paste", "Copy" and "Clear" to edit
tables. This enables to copy and paste data from/to
other tables e.g. Microsoft Excel. You can also enter
new values directly into the table.
Other buttons
"Remove" Delete selected entry in the list.
"Update" Writes back the table data to the
selected entry.
"Add new" Makes a new entry.
"Import time test"
Imports the latest opened time test.
Programma Electric AB
ZP-CF02E R101 125
17 Report
17 Report
To make reports you use different templates de-
pending on from which instrument you make the
report. Reports can contain a summary page, tables
and graphical pictures of the result. There are also
templates with only summary or only summary and
tables. The templates are different types of list files
(LST-files). You can edit the templates your self,
some parts is quite easy to edit some parts is not that
easy.
Change logotype
Exchange the file logo.bmp in
C:\PETOOLS\REPORT
This is normally the only change of the templates
you need to do. To do some more changes yourself
is possible but it will take more time. Read the
separate manual from List & Label for more infor- 3. Click on “Info” to edit the report information
mation. or to change to monochrome printout.
4. Select a printer (if no printer is connected
How to make a printout. select a printer that have LPT1 as the port,
even if no printer is physically connected to
Click the “Report” icon on the toolbar. LPT1), and click ”Start”.
Print on paper
1. Click the “Report” button.
Preview on screen 2. Click on “Info” to edit the report information
or to change to monochrome printout.
1. Click the Report button and you will see this
picture: 3. Click “Print”.
4. Select printer and click “Start”.
Programma Electric AB
126 ZP-CF02E R101
17 Report
Programma Electric AB
ZP-CF02E R101 127
17 Report
General
GENERAL.LST
This template contains tables and summary, no
graphical pictures.
GENERALL.LST
This template contains tables and summary, no
graphical pictures. Paper size landscape.
Distance
DISTANCE.LST
This template contains the full report with summary
page, tables and graphical pictures. From RX there
is however no graphical picture.
TABLES03.LST
This template contains the full report with summary
page and tables.
Sync
ASYNC.LST
Full report template with summary page, tables and
graphics.
TABLES01.LST
Tables and summary template.
Current
CURRENT.LST
Full report template with summary page, tables and
graphics.
TABLES02.LST
Tables and summary template.
Programma Electric AB
128 ZP-CF02E R101
17 Report
Evaluation of reports
Programma Electric AB
ZP-CF02E R101 129
17 Report
Batch report
To make an automatic print out of several tests.
1. Click the button “Batch report”.
2. You will now have a screen like this (see
below).
Programma Electric AB
130 ZP-CF02E R101
17 Report
Search for all tests – from 2. The reports will be printed on the printer
selected as the default printer in Windows.
one part of the directory If the printer is a colour printer make sure
you have not marked “Monochrome printer”
tree, and down in the “Control center”, “Configuration”.
1. Make sure that the Freja Win is in “Off-line” 3. Click on the directory you want to start the
mode. search from.
Each report will start up the used instru- You can make the window a bit wider and
ment and this will be faster if Freja Win is in easier to see the tree structure by dragging
“Off-line” mode. the right edge with the mouse.
2. The reports will be printed on the printer 4. Mark “Enable date selection”.
selected as the default printer in Windows. 5. Define Start and stop date.
If the printer is a colour printer make sure 6. Click “Scan for date”. Each FREJA test, which
you have not marked “Monochrome printer” will be reported, will now be marked “✔”.
in the “Control center”, “Configuration”. You can see how many tests found up in the
3. Click on the directory you want to start the left corner. If you click on this “bar” and then
search from. press an” arrow key” (left or right) on the
You can make the window a bit wider and keyboard, you can toggle between the tests.
easier to see the tree structure by dragging The selected test is indicated with “an open
the right edge with the mouse. FREJA icon”. The description of each selected
4. Click “Check all”. Each FREJA test, which test is shown in the “Test info view” (if you
will be reported, will now be marked “✔”. have filled them in).
You can see how many tests found, up in 7. You can unmark a test or more, if you do
the left corner. If you click on this “bar” and not want to report it/them. Click on these
then press an“ arrow key” (left or right) on tests with the mouse, or use the “Uncheck
the keyboard, you can toggle between the all” button if you have selected the wrong
tests. The selected test is indicated with part of the directory.
“an open FREJA icon”. The description of Each instrument will use the shown report
each selected test is shown in the “Test templates.
info view” (if you have filled them in). 8. Fill up the printer with paper.
5. You can unmark a test or more, if you do 7. Click “Start print” to print all.
not want to report it/them. Click on these You can see how many reports that have
tests with the mouse, or use the “Uncheck been printed on the “bar” up in the left
all” button if you have selected the wrong corner.
part of the directory.
Each instrument will use the shown report
templates. Select and mark the test-
6. Fill up the printer with paper. files using the mouse
7. Click “Start print” to print all. 1. Click on the tests you want to report.
You can see how many reports that have Each instrument will use the shown report
been printed on the “bar” up in the left template
corner.
3. Fill up the printer with paper
4. Click “Start print” to print all.
Search for all tests within a You can see how many reports that have
specific date range been printed on the “bar” up in the left
corner.
1. Make sure that the Freja Win is in “Off-line”
mode.
Each report will start up the used instru-
ment and this will be faster if Freja Win is in
“Off-line” mode.
Programma Electric AB
ZP-CF02E R101 131
17 Report
Save a report as a
text file
When you want to have the report as a text-file.
1. Run the test as normal and close the
instrument
2. Save the test as normal as a directory.
A “Result.txt” is now created in that direc-
tory.
Below is an example of a report text file.
Programma Electric AB
132 ZP-CF02E R101
17 Report
Programma Electric AB
ZP-CF02E R101 133
18 Application examples
5. Select “Current Direction” L1 > NI or NI >
Testing distance L1, press spacebar to change.
6. Set “No. of Zones” to be tested (1-6).
and impedance 7. Set time margins and tolerance for each
protection relays zone according to the relay settings.
8. Set “Prefault Settings”. For instance normal
load impedance at the tested object.
9. Set “Max. Fault Settings” of the generators
depending of your relay settings.
10. Select “BI 1” (binary input) and name it,
e.g. “Trip”, select contact sense or
voltage .
Programma Electric AB
136 ZP-CF02E R101
18 Application examples
or
1. Select number of states (1-28).
7. Select binary output if you need any
changes in breaker position during the
sequence.
2. Select state 2.
Connect menu
3. Set duration.
Programma Electric AB
ZP-CF02E R101 137
18 Application examples
5. Set what kind of “goal” you want. 8. Set if necessary in witch state you need
changes in circuit breaker position. In this
example we had order a change after 50 ms
in state 3.
10
Programma Electric AB
138 ZP-CF02E R101
18 Application examples
Programma Electric AB
ZP-CF02E R101 139
19 Calibration
19 Calibration
Basic Calibration using the
Calibration of FREJA 300 is recommended once a
year.
Calibration Box 2
To perform a calibration you use a calibration box.
There are two models of the calibration box, the Important!
new called ”Calibration Box 2” and the old Disconnect amplifiers during calibration.
”Calibration Box”.
Channels
Ana Low
1. Connect the Calibration box as shown in
the figure below.
Programma Electric AB
140 ZP-CF02E R101
19 Calibration
LU/LI
1. Select LU/LI and follow the instructions on
the display.
2. When the calibration is finished there will be
an ”*”over LU/LI on the display.
* *
Ana.Low LU/LI Ana.High
Ana.High
1. Disconnect the Calibration box.
2. Select “Ana.High” (Analog high).
Programma Electric AB
ZP-CF02E R101 141
19 Calibration
4. Press button .
5. Press button .
6. Press the Start button on the Calibration box.
7. When the test is finished there will be three
”*” on the display.
* * *
Programma Electric AB
142 ZP-CF02E R101
19 Calibration
5. Press button .
6. Wait until ”BO3 Calibration” is displayed.
7. Connect as shown below.
8. Press button .
9. Wait until ”BO4 Calibration” is displayed.
10.Connect as shown below.
11.Press button .
12.If any of the ”BI/BO” has failed (due to
bounces), press the button and run the
test again.
13.The calibration is finished, press to
return to the Main menu.
Programma Electric AB
ZP-CF02E R101 143
19 Calibration
Channels
Note! If you get an Error message ”Unsuccess-
Ana Low ful calibration” on the display, run the
calibration again and be sure to set the
1. Connect the Calibration box as shown in the right code on the calibration box.
figure below.
2. Select “Calib” from the display main menu. LU/LI
3. Then select “Channels”.
1. Select LU/LI and follow the instructions on
the display.
Channels BI/BO Recall 2. When the calibration is finished there will be
an ”*”over LU/LI on the display.
* *
4. Select ”Ana.Low” (Analog low). Ana.Low LU/LI Ana.High
Programma Electric AB
144 ZP-CF02E R101
19 Calibration
4. Press button .
5. When ”Analog I High” is displayed , change
the connections as described on the dis-
play:
L1I – ANALOG INPUTS HIGH I
NI – ANALOG INPUTS COM
6. Press button .
1b.An alternative is to connect to the ”Ext”
connector on the rear panel. See figure 1b
below.
8. Press button .
Programma Electric AB
ZP-CF02E R101 145
19 Calibration
2. Select “BI/BO”.
5. Press button
6. Wait until ”BO3 Calibration” is displayed.
7. Connect as shown below.
Programma Electric AB
146 ZP-CF02E R101
19 Calibration
Programma Electric AB
ZP-CF02E R101 147
19 Calibration
Programma Electric AB
148 ZP-CF02E R101
19 Calibration
Calibration
A. Voltage precision
1. Turn on the Calibration box and wait 5
minutes.
2. Set the Code to 2167.
3. Measure the voltage at U ref, with 21 kΩ
load.
4. Note the results of step 3.
B. Current generation
The calibration box is to be tested in five steps at 1. Set the Code to 0943.
23(±5)oC. 2. Measure the current at I ref, with 50 Ω
A.Voltage precision load.
3. Note the results of step 2.
B. Current generation
C. Time-pulse generation C. Time-pulse generation
1. Connect a cable between Uref red and Out
D.Voltage divider resistance black
E. Shunt resistance 2. Connect a timer between Uref black and
Thumbwheel switch
Out red.
The following codes are used for the thumbwheel: 3. Set code 2167.
2167 +10 V
4. Press Start.
6567 (150 V) measurement 7.5 V (approx. 1/20) 5. Note the result with one decimal. (The value
8767 (10 A) measurement 0.5 V (approx. 0.05 Ω)
shall be between 245 and 255 ms).
0943 +20 mA D. Check the resistance of the voltage
Nominal values for Calibration box
divider
1. Set the Code to 6567.
Tolerance limit
2. Measure the resistance between U out red
Uref 10 V 0.01% (1 mV)
and U ref red.
Iref 20 mA 0.01% (2 µA)
3. The value should be 500 kΩ ±500 Ω
Time 250 ms ±5 ms 0.1 ms
R Voltage divider 500 kΩ 0.1%
E. Check the shunt resistance
R Shunt 50 mΩ 0.1%
1. Set the Code to 8767.
Measuring instruments for 2. Measure the resistance between I out black
and U ref red.
calibration
3. The value should be 250 mΩ ±50 mΩ
Volt ammeter
The stability of the volt ammeter must be better than
what is set forth above and it must operate within
closer tolerance limits.
Programma Electric AB
ZP-CF02E R101 149
19 Calibration
Label
Programma Electric AB
150 ZP-CF02E R101
19 Calibration
Programma Electric AB
ZP-CF02E R101 151
20 Troubleshooting
20 Troubleshooting
Training
Problem Remedy
For information about training courses contact your
You only get half the Select network model local distributor or the Programma head office. You
current for a 2- phase fault 3PZSTD2A which provides can also get some hints studying the Freja demo CD.
1xI and 2xU
Communication problems
PC – FREJA
Fault No FREJA connected
What to do Connect FREJA to PC, or keep on if you
just want to work with the FREJA Win
without the FREJA.
Programma Electric AB
152 ZP-CF02E R101
20 Troubleshooting
Programma Electric AB
ZP-CF02E R101 153
21 Specifications
21 Specifications
Specifications are valid for resistive load, nominal DC current measuring input, LOW
voltage supply and ambient temperature +25°C ±3°C, Measuring range ±20 mA
(+77°F ±5.4°F), after 30 minutes warm up time. All
hardware data are for full scale values. Specifications Resolution SW 0.1 µA
are subject to change without notice. HW 0.6 µA
Inaccuracy 0.01% typical,
Environmental conditions 0.03% guaranteed (= 6 µA)
Application field For use in high-voltage DC voltage measuring input, LOW
substations and industrial Measuring range ±10 V
environments.
Resolution SW 0.1 mV
Temperature HW 0.3 mV
Operating 0 to +50°C (+32 to +122°F) Inaccuracy 0.01% typical,
Storage/transport -40 to +70°C (-40 to +158°F) 0.03% guaranteed (= 3 mV)
Humidity 5% – 95% RH, non-condensing 1)
AC/DC current measuring input, HIGH
Measuring range ±14 A DC, 10 A ACRMS
CE-marking
Inaccuracy DC <0.1%, AC <0.3%
LVD Low Voltage Directive 73/23/ 1)
EEC am. by 93/68/EEC AC/DC voltage measuring input, HIGH
EMC EMC Directive 89/336/EEC Measuring range ±220 V DC, 150 V ACRMS
am. by 91/263/EEC, 92/31/ Inaccuracy DC <0.05%, AC <0.2%
EEC and 93/68/EEC
Measurement, internal generated values
General Accuracy
Voltage AC/DC <1% ±1 digit
Mains input (nominal) 90 – 264 V AC, 47 – 63 Hz
Current AC/DC <2% ±2 digits
Power consumption 600 VA typical (1200 VA max)
Binary outputs
Dimensions
Number 2 x 4 (NO & NC)
Instrument 450 x 160 x 410 mm
Type zero-potential contacts, con-
(17.7 x 6.3 x 16.1")
trolled via software
Transport case 560 x 240 x 575 mm
Break capacity AC 240 V AC, max 8 A,
(22” x 9.5” x 22.6”)
max load 2000 VA
Weight
Break capacity DC 275 V DC, max 8 A,
Instrument 15 kg (33.1 lbs) max load 240 W
Transport case 7.5 kg (16.5 lbs)
Low Level outputs (ROGOW)
Display LCD
Setting range
Available languages English, French, German,
LLU 3 X 0...2 Vrms
Spanish, Swedish
LLI 3 X 0...3 Vrms
Measurement section Max. output current 5 mA
Inaccuracy <0.1% typ. (<0.2% guaranteed)
Binary inputs Resolution 250 µV
Number 10 Inputs (2 groups of 5 inde- Distortion (THD+N)2) <0.05% typ. (<0.1% guaranteed)
pendent)
Max generating time 5 minutes
Type Dry or wet contacts,
275 V DC, 240 V AC Generator Section
Internal resolution time 50 µs
Voltage outputs
Galvanic isolation Galvanically separated from Range
the amplifier section. Two 4-phase AC 4 x 150 V
galvanically separated groups: 1-phase AC (L-L) 2 x 300 V
1 to 5 and 6 to 10 DC (L-N) 180 V
Max measuring time 15264 h (636 days) Power
Range Resolution 3-phase AC 3 x 82 VA at 150 V
0 – 9.9 ms 0.1 ms 1-phase AC (L-L) 1 x 140 VA at 300 V
10 ms – 60 minutes 1 ms DC (L-N) 87 W
1 h – 15264 h 1s Resolution SW 10 mV
HW 6.5 mV
Inccuracy3) (guaranteed) ±0.01% of range + 0.05% of
reading
Programma Electric AB
154 ZP-CF02E R101
21 Specifications
Generators, general
Frequency range
Continuous signals DC – 2000 Hz
Transient signals DC – 3.5 kHz
Frequency resolution 1 mHz
Frequency inaccuracy 0.01%
Phase angle range 0 – 360°
Phase resolution 0.1°
Phase inaccuracy4) ±0.1°
Connection (Amplifier outputs)
4 mm stackable safety plugs
or 8-pin amplifier multiconnector
All seven generators are continuously and independently
adjustable in amplitude and phase. No switching of range is
necessary. All current and voltage outputs are fully overload-
and short-circuit-proof and protected against external high
voltage transient signals and overtemperature.
Note! To allow continuous generation of high DC current
(12 - 15 A), a minimum load impedance of 0.2 Ohm is
required. For lower load impedances, e.g. short-circuit,
the time is limited to 1 minute.
Programma Electric AB
ZP-CF02E R101 155
Others
Abbreviation list
”.” Indication that test was passed
”1st” First state, normally Prefault
”2nd” Second state, normally Fault
BI Binary Input
Binary Goal Goal for Binary Input. Timer stops
when BI = Binary Goal
BO Binary Output
DUT Device Under Test = Relay
Prefault Generated values before the fault
Ramp There are two different types of
Ramp. In General the Ramp extends
continuously from 1st to 2nd. But in
Distance RX-Ramp the rampsare not
continuous. These ramps jump back
to Prefault. Otherwise it would be
impossible to test distance relays
without turning of the time zones.
REFGRAPH Library (directory) for reference
graphs.
Programma Electric AB
156 ZP-CF02E R101
Others
Programma Electric AB
ZP-CF02E R101 157
Others
Index
Symbols Comtrade ........................................................................ 109
Config .................................................................................. 69
- - .......................................................................................... 129
Connect ............................................................................... 41
|Z| tolerance ..................................................................... 52
Connect menu ................................................................ 53
1ST state ............................................................................ 16
Connector .......................................................................... 11
2 N D ........................................................................................ 17
Contact ................................................................................ 12
A Control Center ................................................................ 31
Convert ............................................................................. 108
Add manual ramps ..................................................... 56
Converter ........................................................................ 108
Adjust axis ........................................................................ 49
Copy zone ...................................................................... 122
Application example ................................................. 23
Create zone ................................................................... 122
Application examples ............................................ 134
Current ................................................................................. 72
ARC VIA 0 ....................................................................... 120
Current - Toolbar ........................................................... 73
Ascii .................................................................................... 109
Current Direction .......................................................... 52
Auto 300 ............................................................................. 98
Current instrument ...................................................... 72
A U T O 2 1 ........................................................ 78, 82, 112
Config - menu 74
AUTO21 instrument ............................. 78, 82, 112
Pickup - menu 75
Automatic mode ........................................................... 60
Time test - menu 75
Autoreclose ...................................................................... 92
Autoreclose – Toolbar ............................................... 93 D
Autoreclose instrument ........................................... 92
Delete all log data ....................................................... 40
B Delete ramp ..................................................................... 56
Distance instrument ................................................... 48
Batch report .................................................................. 130
Config menu 51
Binary Goal ...................................................................... 43
Connect menu 53
Binary goal ....................................................................... 17
Network model 3PZ3I1B 52
Binary Input ..................................................................... 43
Network model 3PZSTD1A 51
Binary list ........................................................................... 38
Network model 3PZSTD2A 50
Binary outputs ................................................................ 18
D I S TA N C E . L S T ............................................................ 128
B O ........................................................................................... 37
Dry contact ....................................................................... 52
C
E
C A 1 ........................................................................................ 24
E-mail ................................................................................ 127
C A 1 H ..................................................................................... 24
Elements ......................................................................... 120
C A 3 ............................................................................ 24, 140
Evaluation of reports .............................................. 129
Calibration ..................................................................... 140
External Amplifier ....................................................... 24
Ana.High 141, 145
Binary inputs & outputs 142, 145 F
LU/LI 141, 144
F A I L .................................................................................... 129
Recall 143, 146
FREJA Win Control Center .................................... 31
Calibration Box ........................................................... 144
Freja Win software ...................................................... 28
Calibration Box 2 ...................................................... 140
Installation 28
Calibration using the Calibration Box ........ 144
System requirements 28
Calibration using the Calibration Box 2 .... 140
Frequency instrument ............................................... 82
Change language ......................................................... 15
Front panel ........................................................................ 10
Change logotype ....................................................... 126
Fuse ....................................................................................... 10
Change phase name .................................................. 15
Check list ........................................................................... 67 G
C I R C L E .............................................................................. 120
General instrument ..................................................... 36
CIRCLE VIA 0 ................................................................ 120
Binary Goal 43
Communication problems .................................. 152
Connect 41
Programma Electric AB
158 ZP-CF02E R101
Others
Programma Electric AB
ZP-CF02E R101 159
Others
Programma Electric AB
160 ZP-CF02E R101
Others
Programma Electric AB
ZP-CF02E R101 161
Others
Appendices
1. List & Label Quick guide
3. Freja reports
Example 1
Example 2
4.
5. CD ROM
FREJA 300 Demonstration
FREJA Win
Programma Electric AB
162 ZP-CF02E R101
Others
Programma Electric AB
ZP-CF02E R101 163