You are on page 1of 10

Ventricular Cross Stimulation in a Dual

Chamber Pacing System: Phenomenon Analysis


ANDREA PUGLISI. RENATO RICCI, PAOLO AZZOLINI,
CARLO PERALDO NEJA, MASSIMO FIORANELLI, GIULIO SPECIALE,
and GERARDO ANGRISANI
From the Department of Cardiology, Fatebenefratelli Hospital, Isola Tiberina, Rome, Italy

PUGLISI A., ET AL.: Ventricular Cross Stimulation in a Dual Chamber Pacing System: Phenomenon
Analysis. Seven cases o/venlricular cross sfimulation from a group of 23 patients implanted with DDD
devices are presented. In two patients the phenomenon was observed at the moment of DDD programming
at nominal values, and in jive other patients it ivas reproduced by increasing (he atrial output voltage up to
ten volts. In all 23 patients cross stimulation disappeared permanently within 24 hours after implantation.
From the onset of cross stimulation to its end. atrial and ventricular threshold voltages were unchanged.
while the atrial and ventricular impedances significantly decreased. These results suggest that an impor-
tant role in the phenomenon occurs by impedance variation al the interface between the pulse generator
and body tissue. (PACE. Vol. 13, August 1990}

cross stimulation, dual chamber pacing .system, impedance

Introduction
Table I.
In a dual chamber pacing system, stimulation
Threshold (volts/msec)
of a cardiac chamber different from the one to
which the stimulus was directed, is defined as Implant Cross End
"cross" stimulation.
The occurrence of venfricular stimulation by pt1
the atrial lead as a consequence of dislocation into A 1/0.5 1/0.5 1/0.5
the right ventricle, near the tricuspid annulus or V 0.5/0.5 0.75/0.4 0.75/0.4
in the coronary sinus has been reported pre- pt2
viously.^^'' Rarely is fhe phenomenon related fo A 1/0.5 0.9/0.5 0.9/0.5
the reversal of leads due fo incorrect connec- V 0.4/0.5 0.75/0.4 0.75/0.4
tions.^ In this study, seven cases of femporary pt3
ventricular cross sfimulafion are reported. In A 0.9/0.5 1/0.5 1/0.5
V 0.4/0.5 0.75/0.3 0.75/0.3
order to analyze this phenomenon, variations of
pt4
bioelectrical parameters were studied. A 1.1/0.5 1.3/0.5 1.3/0.5
V 0.4/0.5 0.75/0.3 0.75/0.4
Materials and Methods pt5
From January 1988 fo lune 1989, 23 Siemens A 1.3/0.5 1.5/0.5 1.5/0.5
(Siemens-Elema. Sfockholm, Sweden) 544 dual V 0.7/0.5 0.75/0.5 0.75/0.5
chamber pacemakers were implanted. Pacesefter pt6
A 1/0.5 1/0.5 1/0.5
V 0.6/0.5 0.75/0.4 0.75/0.4
Address for reprint.s: Andrea Puglisi. Department of Cardiol- pt7
ogy. Fatehenefratelli Hospita], Isola Tiberina. 39, 00186 Rome, A 1.2/0.5 1.3/0.5 1.5/0.5
Ilaly. V 0.4/0.5 0.75/0.4 0.75/0.4
Received October 30, 1989; revision March 6. 1990: accepted
March 7. 1990. pt = patient; A = atrial; V = ventricular.

PACE, Vol. 13 August 1990 993


PUGLISI, ET AL

1015M ventricular and Pacesetter 1013M leads fime of lead implantation, was performed with a
(Pacesefter Systems, Sylmar, CA, USA) and Med- Medtronic 5311 pacing system analyzer.
tronic 6957-58 (Medfronic, Inc., Minneapolis, MN, In order to exclude extracardiac stimulation
USA) atrial leads were used. Atrial leads had been of surrounding sfructures, fhe afrinm was paced
positioned routinely in the right atrial appendage from the oufpuf threshold value, up fo 10 volts,
and, in some cases, in the high anterior wall of the both in fhe AAI and DDD modes. For the lead, all
righf atrium. measurements were made 30 minutes after tbe
Analysis of stimulation parameters, af fhe atrial fixation into afrial myocardium.

Figure 1. Patient 6. DDD mode, rate = 90 ppm, AA (W} = 667 msec, ventricular refractory period
= 325 msec, AV delay = 140 msec, VA ^ 527 msec, blanking period = 38 msec, atrial output = 8.5
volts, magnet off. Paper speed = 50 mm/sec. Full circle = cross stimulation.

994 August 1990 PACE. Vol. 13


CROSS STIMULATION

,1:' liii
- • 1 •

-- •I - ;;;

k
ilL .:i

- 1 •

:t 1 •• -

T

— ^

T
% ' ""Tr "TTT — —
--I i
1: T , •;
! •. ;
if..
-"1
: J •; 4
, 1,. . . . -• I
-H -1— -~r
• ; i ! : ] ••• •

i 1,.

Ak ; : • '
• T y
i •
y -": A
' :' r
•' ' £ \
'. i- ^ • -

i ^^

y
' • • t ; : • :

A V:
il- 1
%

r7\
^ 1 •
—I
-— ^- -f-
1V
— : • -

'- / • .

i Mlil ii 1!•• •
Xn i I
/^ K
•, : :

JitiW' i; {:: . • ;

/ ^
;
• ;

:. l':
: :
\

•; , :
hJ Y \

-^ |f - • • •
Vd
. •
F hi:
IT! '' , ;

; ; ; •

iiJiii :;;;

. • ' :

V J...

: ....
Figure 2A. Patient 3. (A} DDD mode, rate = 70 ppm, AA (Wj = 857 msec, ventricular refraclory
period = 250 msec. AV delay =165 msec, VA = 692 msec, blanking period = 38 msec, atrial output
= 5 volts, magnet off.

In two patients, cross stimulation of the ven- Results


tricle from fhe atrial lead was observed af the
moment of fhe DDD programming, immediately There were seven patients, aged 68 fo 79
after implant af nominal values. In five patients, years (four males and three females]. Three had
fhe phenomenon was reproduced by increasing sick sinus syndrome and atriovenfricular block;
fhe atrial oufpuf voltage in tho DDD mode. two trifascicular block; one sick sinus syndrome;
From tbe onsef of cross stimulation fo its end and one afrioventricular block. Three had idio-
fhe following procedures were performed, using a pathic heart disease; two hypertensive heart dis-
Paceseffer System Inc. Model 380 programmer ease: and fwo chronic ischemic hearf disease. In
wifh function pack 385B: deferminafion of atrial two patients, the venfricular cross stimulation
and venfricular stimulation threshold voltage; was observed at the moment of DDD program-
atrial sfimulafion af increasing voltage, from ming at nominal value, The minimum afrial out-
threshold value up fo 10 volfs, in the DDD mode; put volfage needed to induce the phenomenon
measurement of afrial and venfricular pacing im- was 3 and 4 volts, respectively.
pedance and current at an oufpuf of 5 volfs and af In fhe remaining five pafienfs. cross sfimula-
a pulse duration of 0.5 msec. tion was induced by increasing fhe outpuf voltage

PACE. Vol. 13 August 1990 995


PUGLISI, ET AL

Figure 2B. Afrial output - 6 volts. Paper speed = 50 mm/sec.

to a mean value of 7.4 ± 1.1 volts (range 6 to 8.5 rate increased: as the pacing cycle length was
volts). In all patients, during cross stimulation, the equal to the programmed VA interval plus the
QRS complex was paced from the atrial stimulus, blanking period (Figs. 3A, 3B, 4A). By increasing
and after the programmed AV delay tbe ventricu- the blanking period from 38 to 50 msec, and leav-
lar stimulus appeared (Figs. 1 and 2). The ventric- ing the atrial output unchanged, only cross stimu-
ular stimulus was ineffective as it fell into the lation was present [Figs. 3C, 4B).
refractory period ofthe ventricular myocardium. None of the seven patients showed cross
Tbe QRS morphology caused by the atrial stimu- stimulation 24 hours after implant or during later
lus and ventricular stimulation were identical. follow-up.
During cross stimulation, the ventricular From the onset of cross stimulation to its end,
stimulus was absent in four patients because of a tbe following variations of bioelectrical parame-
cross-talk phenomenon; consequently, tbe pacing ters were observed: atrial and ventricular thresb-

996 August 1990 PACE, Vol. 13


CROSS STIMULATION

Figure 3A. Patient 4. (A) DDD mode, rale = 90 ppm, AA (W) = 667 msec, venlricuJar refractory
period = 250 msec, AV delay = 165 msec, VA = 502 msec, blanking period = 38 msec, alrial oufpul
= 7 volts, magnet off.

old voltage were unchanged; mean atrial imped- the paced QRS was clearly different from that in-
ance decreased from 606 ± 83 to 529 ± 89 ohms (P duced by the ventricular stimulus. In such cases,
< 0.01): mean ventricular impedance decreased chest X ray shows the malposition.
from 705 ± 144 to 527 ± 85 ohms (P < 0.01); mean In our patients, diagnosis of dislocation could
atrial current increased from 7.4 ± 1.4 to 8.4 ± 1.2 be excluded for the following reasons; the QRS
mA (P < 0.01); mean ventricular current in- morphology during cross phenomenon was iden-
creased from 6,4 ± 1 to 8.5 ± .6 mA {P < 0.01). tical to that recorded in the VVI mode; by pro-
Table I lists atrial and ventricular threshold gramming the unit to the AAI mode at nominal
values. Variation of atrial and ventricular imped- values, normal atrial pacing occurred; chest X
ance and current are shown in Tables II and III. rays taken after pacemaker implantation showed
that both atrial and ventricular leads were in good
position; in all cases the phenomenon disap-
Discussion peared spontaneously within 24 hours, and no
cross stimulation was observed during foIIow-up.^
In the majority of cases reported, ventricular Others have pointed out the possibility of
stimulation by an atrial lead was attributed to cross stimulation when the lead positioned in the
lead dislocation.^'^'^ When this occurred, the phe- right atrial appendage is very close to the right
nomenon was permanent and the morphology of ventricular outflow tract,^ in which case suffi-

PACE, Vol. 13 August 1990 997


PUGLISI. ET AL

Figure 3B. Atriai output = 8.5 vo/Is. Pacing cycle length = 550 msec. (VA + blanking -I- 10 ms.

ciently high pacing energy can be transmitted to ulation is very low and, if present, should either
the ventricle, causing depolarization. This could regress according to stimulation threshold varia-
also be ruled out in our patients because of the tions, or persist. In all patients cross stimulation
paced QRS morphology, and at implantation, disappeared permanently within 24 hours after
using a PSA. temporary stimulation was routinely implantation, and no atrial and ventricular
carried out in AAI and DDD modes with output threshold differences were noticed from the im-
up to 10 volts, without observing the cross stimu- plantation to the end of the phenomenon (Table i).
lation phenomenon. Therefore a circuit defect was very unlikely. Si-
Cross stimulation in these instances de- multaneously with the end of cross stimulation,
pended on a pacemaker circuit defect or on modi- atrial and ventricular impedances decreased sig-
fications of hioelectrical parameters between the nificantly (Tables II and III).
pulse generator case and body surface. Normal As stimulation thresholds were unchanged,
intracircuit resistances are very high. The possi- impedance reduction is related to a better pace-
bility of transferring sufficient energy from the maker/subcutaneous tissue contact that there-
atrial to the ventricular circuit to produce a stim- fore played a role in the cause of the cross stimu-

998 August 1990 PACE. Vol. 13


CROSS STIMULATION

Figure 3C. Blanking period = 50 msec. Pacing cycle Jenglh = 667 msec. Paper speed = 50
mm/sec.

lation.° Levine et aP described cross stimulation (confirmed hy the manufacturer) are in labeling
in DDD pacing system when the pacer was out of on the can and the coating applied to the can. The
the pocket with both atrial and ventricular leads window of the can is considerably smaller on the
connected. In this case because of the high imped- 544 as compared to the 285; postoperatively this
ance, a small amount of current flowed between may not allow a good contact with the body tissue
the implanted atrial and ventricular electrode and produce high impedance.
when atrial or ventricular outputs are delivered, In this case, atrial stimulation will produce a
resulting in atrial and/or ventricular anodal stim- "small" current flow from the atrial to the ven-
ulation. tricular electrode becoming clinically significant
In the clinical setting, a similar condition in the presence of very low ventricular thresholds
could be observed if the impedance at case-body and/or high atrial output voltage, resulting in
tissue level significantly increases. The Siemens cross stimulation.
544 pacemaker is derived from the model 285 The phenomenon is temporary probably be-
Genesis electronically and mechanically identical cause of the following mechanisms: impedance
to Pacesetter model 285. The only differences decrease at the generator-subcutaneous tissue in-

PACE. Vol. 13 August 1990 999


PUGLISI, ET AL

Figure 4. Patient 2. {A} DDD mode, rate = 65 ppm, AA fW)


= 923 msec, ventricular refractory period - 250 msec. AV
delay = 140 msec, VA = 783 msec. Wanking period = 38 msec,
atrial output - 5 volts, magnet off. Pacing cycle length = 830
msec. (VA + blanking + 9 msec.) fB) blanking period = 50
msec. Pacing cycle length = 923 msec. Paper speed; 25 mm/

1000 August 1990 PACE, Vol. 13


CROSS STIMULATION

Table II. Table III.

Impedance (ohms) Current (mA)

Implant Cross End Implant Cross End

pt1 pt1
A 667 617 579 A 1.5 7.7 8.4
V 1000 739 577 V 0.6 6.6 8.4
pt2 pt2
A 440 473 352 A 2.3 4.6 6.1
V 444 437 372 V 0.9 4.9 5.6
pt3 pt3
A 483 707 617 A 3.1 7.5 8.4
V 500 813 551 V 0.7 6.5 9.5
pt4 pt4
A 526 609 537 A 2.1 7.4 8.4
V 500 705 522 V 0.7 6.4 8.6
pt5 pt5
A 376 532 478 A 3.1 9.4 10.4
V 650 674 462 V 0.9 7.7 10.8
pt6 pt6
A 474 608 569 A 1.9 8.1 8.6
V 625 668 620 V 0.8 7.4 7.9
pt7 pt7
A 538 697 571 A 2.6 7.1 8.6
V 800 897 587 V 0.5 5.5 8.4

A = atrial; V = ventricular. A = atrial; V = ventricular.

terface; increase of ventricular stimulation Acknowledgment: We gratefully acknowledge Mr. Rivetti


threshold; impedance increase at lead-tissue in- (Siemens-Italy) and Mr. I.. Sfilznick (Pacesetter Systems, CA,
terface. USA) for technical assistance.

References
1. Nathan AW, Camm AI. Douhle ventricular pacing Bognolo DA. Clinical application of A-V sequential
in a patient with a DDD universal pacemaker. (DVI) pacing. In DA Bognolo (ed.) Practical ap-
PACE 1984; 7:432-435. proach to physiologic cardiac pacing. Tarpon
2. Stark S, Fashidi A, Hager WD, et al. Unusual pre- Springs, Florida, Tampa Tracings, 1983, p. 88.
sentation of DDD pacemaker system malfunction. Barber K, Amikam S. Furman S. Atrial lead malpo-
PACE 1985; 8:255-259. sition in a dual chamber (DDD.M) pacemaker.
3. Levine PA, Rihanek BD, Sanders R, et al. CTORS Chest 1983; 84:766-767.
stimulation: The unexpected stimulation of the 7. Doi Y, Takada K, Nakagaki O, et al. A case of cross
unpaced chamber. PACE 1985; 8:600-606. stimulation. PACE 1989; 12:569-573.
4. Barold SS, Falkoff MD, Ong LS, fit al. Function and Stokes K. Bornzin G. The electrode-biointerface:
electrocardiography of DVI pacing system. In SS stimulation. In SS Barold (ed.): Modern Cardiac
Barold (ed.): Modern Cardiac Pacing. Mount Kisco, Pacing. Mount Kisco, New York, Futura Publishing
New York, Futura Publishing Co., Inc., 1985, pp. Co., Inc., 1985, pp. 33-77.
625-644.

PACE. Vol. 13 August 1990 1001

You might also like