Maintenance of AFIAS-6

DATE 2018. 2. 1
TEAM TS Team
AUTHOR Park kitae
Software upgrade procedure

양식-GE02-27 (Rev.00)
Preparation SD card

Copy the “boditech” folder under root directory in SD card

양식-GE02-27 (Rev.00)
System configuration – SW update(1)

양식-GE02-27 (Rev.00)
System configuration – SW update(1)

양식-GE02-27 (Rev.00)
System configuration – SW update(2)

양식-GE02-27 (Rev.00)
System configuration – SW update(2)

양식-GE02-27 (Rev.00)
System configuration – SW update(2)

양식-GE02-27 (Rev.00)
Service engineer mode

양식-GE02-27 (Rev.00)
1. Enter hidden mode

Mode Engineering Supervisor Spot calibration


Password #*boditech# #*supervisor# #*spotcal#
10
1. Enter hidden mode

Mode Password Function

Supervisor #*supervisor# History delete function

Spot Calibration #*spotcal# Can adjust the result value for each item and lots.

Backup/Restore setting information.

Engineering #*boditech# Raw data backup

Set up Axis

11
2. Supervisor mode

Shown delete function


button after apply in history

12
3. Spot calibration

Item will be show after Adjust use the “+,-” button. Spot calibration works
testing The range from 50% to for each Lot No.
150%
13
4. Engineering mode Case: When replacing the main board of the device.

4.1 Backup setting information

scroll

14
4. Engineering mode Case: When replacing the main board of the device.

4.1 Backup setting information

1. find “axis_config.txt” or “A6_xxxxxxxx_SI.txt” file in a SD card


that will be inside “boditech” folder. If not inserted SD card, will
2. Send to us “axis_config.txt” or “A6_xxxxxxx_SI.txt” be show message above.
15
4. Engineering mode Case: When replacing the main board of the device.

4.2 Restore setting information

Or A6_xxxxxxxx_SI.txt

Store in the SD card for the restore setting value


(in the Boditech folder)

16
4. Engineering mode Case: When replacing the main board of the device.

4.2 Restore setting information

scroll

17
4. Engineering mode Case: When replacing the main board of the device.

4.2 Restore setting information

All finished processing, Turn off the device and turn on.
18
Backup Raw data

양식-GE02-27 (Rev.00)
4. Engineering mode
4.3 Backup raw data(1)
Backup menu Function
Raw data Tested data log
Common log Error message log
Databases info History, Cal/QC log
CalF info Calibrated log
Crash log System error log

Take out the SD Card from AFIAS-1 or AFIAS-6 Make a Zip file and
and insert it to a PC send to by e-mail

20
4. Engineering mode
4.3 Backup raw data(2)

21
4. Engineering mode
4.3 Backup raw data(2)
Take out the SD Card from AFIAS-1
and 6 and insert it to a PC

Make a Zip file and


send to by e-mail

22
Data transfer or restore history

양식-GE02-27 (Rev.00)
4. Engineering mode
4.4 Data transfer or Restore history

Take out the SD Card from AFIAS-1/6 and insert it to a


PC

Go to “boditech folder”

24
4. Engineering mode
4.4 Data transfer or Restore history

Confirm the number of files from Copy and paste “update.dat“ file to
“D:\boditech\database_log” folder. boditech folder

25
4. Engineering mode
4.4 Data transfer or Restore history

Insert the SD Card to AFIAS-1/6

**The App Version of the AFIAS-6 which the results need


to be restored to should be above AA6.WW.02.31.01. Reboot AFIAS-1/6
**The SW Ver. does not matter in AFIAS-1.
**Do not remove the SD Card until the Log in
pop-up appears

26
Creating ID chip

양식-GE02-27 (Rev.00)
1. Connection

1. AFIAS ID Chip Creator Program


2. Rs232 serial cable

Connection PC to AFIAS

양식-GE02-27 (Rev.00)
2. Procedure

1. Software open 2. Match the com port number


If connected with AFIAS, will be show “Create” button.

3. Insert ID chip in the No.1 ID chip slot.

4. Choose the ID Chip text file which users wish to create and click “Open”

5. Click “Create” button

6. Click “Check” button

7. “Success” will be displayed and ID Chip is ready for use.


양식-GE02-27 (Rev.00)
Removal and Replacement procedures

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.1 Preparation for disassemble the AFIAS-6

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.2 Right and left side case (M3 x 8mm)

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.3 Top case (M3 x 6mm)

LED

LCD

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.4 Top printer case (M3 x 6mm)

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.5 Tip adaptor and tubing

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.5 Tip adaptor and tubing

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.6 Tip sensor

Loosen 2 screws left/right side and inner 4 screws

Disconnect the ID Chip PCB cable from the Main PCB

양식-GE02-27 (Rev.00) Disassembling the case


1. Removal and Replacement procedures
1.6 Tip sensor

Disconnecting cable connector Loosen the screws


양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.7 Back case

Loosen the screws. Disconnecting cable connector, and loosen the ground screw.

Remove the back case

양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.8 LCD parts

양식-GE02-27 (Rev.00)
Axis set up method

양식-GE02-27 (Rev.00)
4. Engineering mode
4.4 Set up Axis
0. Description
The information contained herein is subject to change without prior notice. The only
warranties for the AFIAS-6 and services are set forth in the express warranty
statement accompanying such products and services. Nothing herein should be
construed as constituting an additional warranty. Boditech Med Inc. shall not be
liable for technical or editorial errors or omissions contained herein. Additionally, any
malfunction in elements including the optical sensor assembly, the system board and
the drive train should be consulted with and serviced by Boditech Med Inc. Any
unauthorized operations on elements not suggested in this document will void the
existing warranty.

This setting is working in higher than software version AA6.WW.02.31.00

양식-GE02-27 (Rev.00)
1. Turn on the device

Connect power cable Turn on the main power switch Press the power button

2. Login and Entering engineering mode


2.1 Accessary
Optics-F Optics-T
cartridge cartridge

Optics-F Optics-T
cartridge cartridge

Optics-F
Optics-T
cartridge
cartridge

양식-GE02-27 (Rev.00)
3. FACTORY_SETTING_MODE

Entered Pump X/Y for the setting tip adaptor position

양식-GE02-27 (Rev.00)
4. Pump X/Y

1
2
1 2

Insert setting cartridge in the BAY A,B

Press the “move” button


양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X1
1

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X1

Move the tip adaptor in the


correct position using by
hand.
3 If not correct position,
manually adjust positions
each moving parts.

Check the “unlock”

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X1

1 4 6 5

2 3
2

3 1

Check the “lock”

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X2

1 4

Tap Pump_X1 value + 200


Ex) 74+200 = 274

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X3

1 4

Tap Pump_X1 value + 400


Ex) 74+400 = 474

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X2_Check

1 3

If not correct position,


manually adjust
positions each moving
parts same as page 8.
2 4

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X3_Check

1
If not correct position,
manually adjust
positions each moving
parts same as page 8.
2 4

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X4

4
1

Move the tip adaptor in the


correct position using by
3 hand.
If not correct position,
manually adjust positions
Check the “unlock” each moving parts.

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X4

1 4 6 5

2 2 3

3 1

Check the “lock”

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X5

1 4

Tap Pump_X4 value + 200


Ex) 1175+200 = 1375

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X6

1 4

Tap Pump_X4 value + 400


Eg) 1175+400 = 1575

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X5_Check

If not correct position,


1 manually adjust
positions each moving
parts same as page 8.
2 4

양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X6_Check

If not correct position,


manually adjust
1 positions each moving
parts same as page 8.
2 4

양식-GE02-27 (Rev.00)
4. Pump X/Y Tip remove

1
2

1 4 7 8

2 5 6
Move the tip adaptor
into the center of tip
remove hole.

양식-GE02-27 (Rev.00)
4. Pump X/Y Bay A Tip insert position

2
1

2 Move the tip adaptor 4 5 6


on the top of the
cartridge.

양식-GE02-27 (Rev.00)
4. Pump X/Y Bay B Tip insert position

2
3
1

Move the tip adaptor


2 on the top of the 4 5 6
cartridge. 7

양식-GE02-27 (Rev.00)
5. Optics-F/X
Insert a cartridge end of the holder.

Change this value to 50.


After the setting is completed, change back
1 to the existing setting value.

2
3

4
5

양식-GE02-27 (Rev.00)
6. Optics-F/Y
Insert a cartridge end of the holder.

1 2

Optics-F
3 4 5 cartridge
6 7
Optics-F
cartridge

8 9 10
Optics-F
cartridge

1
2
11

Change this value to 50.


After the setting is completed, change back
양식-GE02-27 (Rev.00)
to the existing setting value.
7. Optics-T/Y
Change the setting cartridge and insert a cartridge end of the holder.

1 2

Optics-T
3 4 5 cartridge
6 7
Optics-T
cartridge
8 9 10
Optics-T
cartridge

1
2
11

양식-GE02-27 (Rev.00)
Q&A

양식-GE02-27 (Rev.00)
TS@boditech.co.kr
양식-GE02-27 (Rev.00)