DATE 2018. 2. 1
TEAM TS Team
AUTHOR Park kitae
Software upgrade procedure
양식-GE02-27 (Rev.00)
Preparation SD card
양식-GE02-27 (Rev.00)
System configuration – SW update(1)
양식-GE02-27 (Rev.00)
System configuration – SW update(1)
양식-GE02-27 (Rev.00)
System configuration – SW update(2)
양식-GE02-27 (Rev.00)
System configuration – SW update(2)
양식-GE02-27 (Rev.00)
System configuration – SW update(2)
양식-GE02-27 (Rev.00)
Service engineer mode
양식-GE02-27 (Rev.00)
1. Enter hidden mode
Spot Calibration #*spotcal# Can adjust the result value for each item and lots.
Set up Axis
11
2. Supervisor mode
12
3. Spot calibration
Item will be show after Adjust use the “+,-” button. Spot calibration works
testing The range from 50% to for each Lot No.
150%
13
4. Engineering mode Case: When replacing the main board of the device.
scroll
14
4. Engineering mode Case: When replacing the main board of the device.
Or A6_xxxxxxxx_SI.txt
16
4. Engineering mode Case: When replacing the main board of the device.
scroll
17
4. Engineering mode Case: When replacing the main board of the device.
All finished processing, Turn off the device and turn on.
18
Backup Raw data
양식-GE02-27 (Rev.00)
4. Engineering mode
4.3 Backup raw data(1)
Backup menu Function
Raw data Tested data log
Common log Error message log
Databases info History, Cal/QC log
CalF info Calibrated log
Crash log System error log
Take out the SD Card from AFIAS-1 or AFIAS-6 Make a Zip file and
and insert it to a PC send to by e-mail
20
4. Engineering mode
4.3 Backup raw data(2)
21
4. Engineering mode
4.3 Backup raw data(2)
Take out the SD Card from AFIAS-1
and 6 and insert it to a PC
22
Data transfer or restore history
양식-GE02-27 (Rev.00)
4. Engineering mode
4.4 Data transfer or Restore history
Go to “boditech folder”
24
4. Engineering mode
4.4 Data transfer or Restore history
Confirm the number of files from Copy and paste “update.dat“ file to
“D:\boditech\database_log” folder. boditech folder
25
4. Engineering mode
4.4 Data transfer or Restore history
26
Creating ID chip
양식-GE02-27 (Rev.00)
1. Connection
Connection PC to AFIAS
양식-GE02-27 (Rev.00)
2. Procedure
4. Choose the ID Chip text file which users wish to create and click “Open”
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.1 Preparation for disassemble the AFIAS-6
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.2 Right and left side case (M3 x 8mm)
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.3 Top case (M3 x 6mm)
LED
LCD
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.4 Top printer case (M3 x 6mm)
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.5 Tip adaptor and tubing
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.5 Tip adaptor and tubing
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.6 Tip sensor
Loosen the screws. Disconnecting cable connector, and loosen the ground screw.
양식-GE02-27 (Rev.00)
1. Removal and Replacement procedures
1.8 LCD parts
양식-GE02-27 (Rev.00)
Axis set up method
양식-GE02-27 (Rev.00)
4. Engineering mode
4.4 Set up Axis
0. Description
The information contained herein is subject to change without prior notice. The only
warranties for the AFIAS-6 and services are set forth in the express warranty
statement accompanying such products and services. Nothing herein should be
construed as constituting an additional warranty. Boditech Med Inc. shall not be
liable for technical or editorial errors or omissions contained herein. Additionally, any
malfunction in elements including the optical sensor assembly, the system board and
the drive train should be consulted with and serviced by Boditech Med Inc. Any
unauthorized operations on elements not suggested in this document will void the
existing warranty.
양식-GE02-27 (Rev.00)
1. Turn on the device
Connect power cable Turn on the main power switch Press the power button
Optics-F Optics-T
cartridge cartridge
Optics-F
Optics-T
cartridge
cartridge
양식-GE02-27 (Rev.00)
3. FACTORY_SETTING_MODE
양식-GE02-27 (Rev.00)
4. Pump X/Y
1
2
1 2
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X1
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X1
1 4 6 5
2 3
2
3 1
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X2
1 4
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X3
1 4
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X2_Check
1 3
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X3_Check
1
If not correct position,
manually adjust
positions each moving
parts same as page 8.
2 4
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X4
4
1
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X4
1 4 6 5
2 2 3
3 1
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X5
1 4
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X6
1 4
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X5_Check
양식-GE02-27 (Rev.00)
4. Pump X/Y Pump X6_Check
양식-GE02-27 (Rev.00)
4. Pump X/Y Tip remove
1
2
1 4 7 8
2 5 6
Move the tip adaptor
into the center of tip
remove hole.
양식-GE02-27 (Rev.00)
4. Pump X/Y Bay A Tip insert position
2
1
양식-GE02-27 (Rev.00)
4. Pump X/Y Bay B Tip insert position
2
3
1
양식-GE02-27 (Rev.00)
5. Optics-F/X
Insert a cartridge end of the holder.
2
3
4
5
양식-GE02-27 (Rev.00)
6. Optics-F/Y
Insert a cartridge end of the holder.
1 2
Optics-F
3 4 5 cartridge
6 7
Optics-F
cartridge
8 9 10
Optics-F
cartridge
1
2
11
1 2
Optics-T
3 4 5 cartridge
6 7
Optics-T
cartridge
8 9 10
Optics-T
cartridge
1
2
11
양식-GE02-27 (Rev.00)
Q&A
양식-GE02-27 (Rev.00)
TS@boditech.co.kr
양식-GE02-27 (Rev.00)