UV Persulfate Combustion

Limit of detection: 0.2ppb C Limit of detection 50ppb C


Max concentration: 4,000ppm C Max concentration: 30,000ppm C
Sample size:2-10mL Sample size: 100uL-2.0mL
Measures: TC, IC, TOC, TC-IC Measures: TC, IC, TOC, TC-IC and optional TN

Applications requiring high sensitivity Good for difficult to oxidize compounds (carbonaceous,
Particulate samples ( less than 0.5mm in diam.) (Particulates are filtered nitrogenous)
prior to analysis) Particulates (0.5-0.8mm diam) (where TOC in particulate must be quantitated)
Drinking water (if method calls for UV Persulfate oxidation (ex. 5310C)) Ex. cellulose
Ground Surface water (SW846-9060A)
Environmental Industry
Pharmaceutical Industry Waste water (suspended solids)
Pharmaceutical grade waters Industrial waste effluents (non-clean matrices)(ex. Paper mill effluent)
Cleaning validations (validation for equipment cleaning procedures) Drinking water (if method calls for combustion (ex.5310B))
(evaluation of rinse waters used for cleaning procedures) Ground Surface water (SW846-9060A) (Low TOC with high TIC content)
Water for Injection ( Method 643, EP 2.2.44) (Evaluates quality of water Brines (high salt content or intrusion ground water samples) (coastal areas)
used for pharmaceutical injections)
Ultra pure water (trace level TOC concentrations, 500ppb or lower) TNb and TOC/TNb Analysis
replacement for Kjeldahl nitrogen wastewater/(used in industrial monitoring
Target Market: Pharmaceutical facilities)
Official methods: EPA 415.1-415.3, 9060A, STD method Target Market: Environmental
5310C,ASTM D4779 and D4839,Cleaning validation/ USP Official methods: EPA 415.5, 415.3 and 9060A, STD Method
TOC Method <643> 5310B, ISO 8245,USP 643 (chapter 24), ASTM D2579, AOAC
Carrier Gas: CO2 free air 973.47,
Nitrogen Carrier Gas: CO2 free air
Oxygen