Application 241-00

Agilent Natural Gas/Sulfur Analyzer


Technical Overview

Application Highlights
• A Flame Photometric Detector (FPD) to determine
hydrogen sulfide, carbonyl sulfide, methyl mercaptan,
and ethyl mercaptan to a lower detection limit of
0.5 ppm.
• A Thermal Conductivity Detector (TCD) to identify
isobutane, n-butane, isopentane, n-pentane, propane,
ethane, hydrogen sulfide, oxygen/argon composite,
nitrogen, methane, and carbon monoxide with an initial
composite C6+ backflush to detector. The lower detec-
tion limit on the TCD is 200 ppm.
• Approximate analysis time on the FPD is 15 minutes,
and on the TCD is 25 minutes.

Optional Configurations 3 4 5 6

• Detailed hydrocarbon analysis of extended natural gas


• TCD/FID/FPD or TCD/FID/SCD for extended natural FPD output
1 Methane

gas with trace sulfur analysis 2


3
Ethane
H2S 2 3
5
1
2
C6+
Nitrogen
4 COS 3 Methane
4 Carbon dioxide
• TCD/FCD/FID for extended natural gas with helium or 5
6
Methyl mercaptan
Ethyl mercaptan 6 5
6
Ethane
Propane
hydrogen 7
8
Isobutane
N-butane
9 Isopentane
1 10 N-pentane
1
For More Information 2 7
8

9 10
4

For more information on our products and services, visit our


Web site at www.agilent.com/chem.
www.agilent.com/chem

3 4 5 6

FPD output
1 Methane
2 Ethane 1 C6+
2 3 2 Nitrogen
3 H2S 5
4 COS 3 Methane
5 Methyl mercaptan 4 Carbon dioxide
6 Ethyl mercaptan 6 5 Ethane
6 Propane
7 Isobutane
8 N-butane
9 Isopentane
1 10 N-pentane
1
8

2 7 9 10
4

FPD output from the Agilent natural gas/sulfur analyzer.

Agilent shall not be liable for errors contained herein or for incidental or consequen-
tial damages in connection with the furnishing, performance, or use of this
material.

Information, descriptions, and specifications in this publication are subject to change


without notice.

© Agilent Technologies, Inc. 2002

Printed in the USA


December 3, 2002
5988-6730EN