You are on page 1of 24

Home Search Collections Journals About Contact us My IOPscience

The reproducibility of some thermometric fixed points and the accuracy of temperature

measurements using platinum resistance thermometers

This content has been downloaded from IOPscience. Please scroll down to see the full text.

1999 Metrologia 36 117

(http://iopscience.iop.org/0026-1394/36/2/7)

View the table of contents for this issue, or go to the journal homepage for more

Download details:

IP Address: 200.23.51.1
This content was downloaded on 17/07/2014 at 21:32

Please note that terms and conditions apply.


metrologia

The reproducibility of some thermometric fixed points


and the accuracy of temperature measurements
using platinum resistance thermometers

J. Ancsin and E. Mendez-Lango

Abstract. The reproducibility of some thermometric fixed points and the accuracy of four platinum resistance
thermometers (PRTs) were studied. It was found that the fixed points of aluminium (Al), zinc (Zn), tin (Sn),
indium (In) and gallium (Ga) were realized reproducibly within ± 0.17 mK; ± 0.11 mK; ± 0.10 mK; ± 0.13 mK
and ± 0.12 mK, respectively. Because the actual impurities and their concentration in our samples (of 99.9999 %
or 99.999 99 % purity) are unknown, the systematic uncertainty due to impurities cannot be estimated. However,
any of the samples of Ga, In, Sn, Zn and Al is consistent with the rest within ± 0.2 mK, using a cubic or quadratic
deviation function, in the temperature range 0 C to 660 C. This indicates that the effect of impurities is negligible.
Four PRTs were selected at random. They were calibrated repeatedly, first up to the Zn point and then up to
the Al point. The resistance of each PRT drifted. From time to time, for each PRT, a seemingly well-established
resistance drift suddenly and unpredictably changed to a different rate of drift. Occasionally, the resistance of the
PRTs shifted. Such unpredictable changes obviously limit the accuracy of temperature measurements using PRTs
no matter what the accuracy of their calibrations. In the case of our four PRTs, the uncertainty of temperature
measurements near 660 C ranged from about ± 1 mK to about ± 2.5 mK even though they were all calibrated at
all fixed points well within ± 0.25 mK uncertainty. Possible explanations are offered for the apparently permanent
drifts and the erratic shifts in the resistance of the PRTs. Some comments are made concerning the ambiguity of
“immersion tests” in general. The furnaces of the National Research Council of Canada used in this work are
high-temperature adiabatic calorimeters.

1. Introduction standards of 99.9999 % purity. It was found that the


above-mentioned fixed points of the NRC and the
In 1996 the fixed points of the Centro Nacional de CENAM are equivalent to within 0.5 mK.
Metrologı́a (CENAM) used for calibrating platinum In a recent exercise, several PRTs were used
resistance thermometers for domestic clients were with the aim of studying the reproducibility of these
compared with similar working standards used at the working fixed points at both the NRC and the CENAM.
National Research Council of Canada (NRC). The fixed To combine these reproducibility studies, the water
points involved were those of Hg, Ga, In, Sn and Zn. triple-point cells and the standard resistors used at
A portable calibration laboratory, developed at the the CENAM and the NRC were also compared using
NRC a few years ago, consisting of two sets of small Model F-18 bridges (Automatic Systems Laboratories).
cells (2.5 cm diameter and 22.5 cm high) and a small The fixed points involved were those of Ar, Hg,
furnace (20 cm 30 cm 35 cm high, accommodating H2 O, Ga, In, Sn, Zn and Al. However, as the work
all cells except Hg), was used to compare the fixed progressed, emphasis gradually shifted from study of
points [1]. The cells of one set are metal-clad graphite- the reproducibility of the fixed points to study of the
encased (for industrial use) and those of the other reproducibility of the PRTs, because it became apparent
set are sealed in quartz (for laboratory use). As all that it is the capability of our PRTs, rather than the
cells contain 99.9999 % pure samples they all have fixed points, that limits the accuracy of the temperature
similar melting ranges to those of the working primary measurements.

2. General behaviour of PRTs


J. Ancsin: Institute for National Measurement Standards, National
Research Council of Canada (NRC), Ottawa, K1A 0R6 Let us assume that a PRT is repeatedly calibrated and
Ontario, Canada.
E. Mendez-Lango: Centro Nacional de Metrologı́a (CENAM),
that all fixed points are realized reproducibly within the
Division Termometrı́a, km 4.5 de la Carretera a los Cués, measuring resolution of an F-18 bridge. If the PRT is
Municipio de El Marqués Querétaro, CP 76900 QRO, Mexico. always placed at the same location within the cell, and

Metrologia, 1999, 36, 117-139 117


J. Ancsin and E. Mendez-Lango

always reaches thermal equilibrium with it, its electrical The temperature is, however, high enough to cause
resistance should be reproducible within the measuring considerable quenching: if a PRT is removed from the
resolution unless its physical or chemical state has been Al point and cooled naturally at ambient temperatures,
altered by then its will be higher than if it were cooled very
slowly [2]. The difference between the “quenched” and
(a) contamination
“annealed” values of is fairly reproducible for a
(b) rearrangement of impurities within the platinum given cooling rate during annealing and is equivalent to
lattice of the sensor about 1.5 mK for most PRTs. The temperature is also
high enough for the accumulated platinum oxide to
(c) quenching or annealing
decompose rapidly. Figure 1 shows an example of the
(d) oxidation, or rate of deoxidation (reduction) of a PRT at the Al point.
This PRT was used extensively in our oxidation studies
(e) (unpredictable) changes in dimensions of the sensor (not yet published). By soaking at various temperatures
wire. (between 200 C and 520 C) for several months, it has
We now briefly review the behaviour of a PRT during gradually built up sufficient oxide to change its from
calibration. 25.494 to 25.514 , a change equivalent to 0.2 C.
Yet at the Al point it lost most of its oxygen in less
2.1 Calibrating up to the Zn point than 1 hour. Thus, repeated calibrations at the Al point
will keep PRTs deoxidized.
Below the Zn point, temperatures are sufficiently low
that neither sizeable diffusion of impurities within the
sensor wire nor quenching takes place. However, once
oxidation begins the measured resistance at the fixed
points will gradually increase as the sensor wire of
the PRT is oxidized further. (In practice, oxidation
is considered to begin when it can be detected by
resistance measurements.) This increased oxidation
state will be accompanied by a slight decrease in
the fixed-point resistance ratios, . Here
and are the electrical resistance of the PRT Figure 1. Deoxidation of a PRT at the Al point.
at a given fixed point and at the triple point of water,
respectively. If, on the other hand, the PRT remains in 3. Experimental procedure and apparatus
a given state of quench and oxidation, then at the Zn
point its oxidation will proceed, leading to an increase in 3.1 At the NRC
resistance, while soaking at the Zn point will gradually
anneal the PRT, leading to a decrease in resistance. During these experiments our water triple-point cell
Data from repeated calibrations suggest that some was frequently compared with another reference cell
reversible and some irreversible changes take place and with the cell used at the CENAM. The working cell
during sizeable temperature changes between the sensor was housed in a separate ice-box. We noted that when
wire of a PRT and its supporting structure. Such a our water triple-point cell was used many times during
change is clearly noticeable at the Zn point and it is one a day, its temperature sometimes increased gradually
of the most serious causes of inaccuracy of our PRTs by up to 0.33 mK by the end of the day. This drift
because its magnitude and direction are unpredictable. was detected by a PRT used exclusively to compare
Due to the difference in thermal expansion between the working water triple-point cell with the reference
platinum and quartz, Pyrex or mica, the expansion or cell at the beginning and at the end of the day. Fixed-
contraction of the (quite soft) platinum wire, especially point cells of Al, Zn, Sn (both triple and freezing
at elevated temperatures, may be hindered by the former point), and In were located in their respective furnaces,
or by the capillaries; thus the Pt may be prevented from and were melted overnight. All cells (except the water
achieving its true equilibrium position relative to these triple-point and Ga cells) and furnaces at the NRC
components. In such cases the behaviour of the PRT are home-made. All samples are of 99.9999 % purity
can be erratic (irreproducible). except the Hg, which is of 99.999 99 % purity, and all
are commercially available.
2.2 Calibrating up to the Al point Traditionally, furnaces are used either to melt or
to freeze the ingots of fixed points during melting-
At the Al point (660 C) the temperature is still or freezing-point realizations. The disadvantage of this
low enough that contamination and rearrangement of technique is that samples are forced into transient states.
impurities by diffusion within the platinum sensor Thus, during a fixed-point realization the sample is not
remain negligible during several hours of soaking. in a state of static thermal equilibrium. It seems more

118 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

appropriate to use the furnace only as a heat shield surface is then achieved by inserting a cold metal rod
to provide the sample with an isothermal environment. 15 cm long and 6 mm diameter for 1 min into each of
A separate sample heater can then be used to supply the two thermometer wells of the ingot. (This ingot has
the sample with a known amount of heat, as required, two thermometer wells directly in the Zn, i.e. without
either to measure its heat capacity, to determine its a graphite sheath, in order to allow either calibration of
latent heats, or simply to drive it to any given state two PRTs simultaneously or observation of the ingot’s
of solid/liquid composition and then allow it to reach behaviour during PRT insertion, using a monitor PRT
thermal equilibrium. in the second well. Also, one of the thermometer wells
can be used to house an internal sample heater. Thus
3.1.1 Aluminium the sample can be heated with negligible heat loss to the
surroundings, allowing precise measurements of heats
The Al furnace has a Na heat-pipe liner 10 cm i.d. of fusion and heat capacity. Of course, for accurate
and 46 cm high, used as an adiabatic heat shield. The heat-capacity measurements a pressure-controlled heat
space between the heat pipe and the 5 cm diameter pipe is more suitable because its temperature can easily
ingot is insulated using loose Fibrefrax insulation. The be made to follow instantaneous temperature variations
only role of the heat pipe is to provide the sample with of the sample.) The nucleating metal rods are short and
isothermal surroundings. The sample is heated using hang on metal wires because, during inside nucleation,
a separate sample heater. Complete melting curves we want to extract heat only from the ingot without
can thus be obtained adiabatically. After supplying a cooling the furnace throat. After realization of the Zn
given fraction of the total heat of fusion by the sample point a cold PRT is inserted and is ready for reading in
heater, the sample can approach a state of thermal about half an hour. If the PRT is left in the Zn ingot,
equilibrium, unhindered by the heat pipe. Such heating, then its resistance will gradually rise (by up to about
followed by temperature observations during the time 0.5 mK) until the sample melts completely in about
it takes for the sample to achieve thermal equilibrium, 10 hours. For adiabatic measurements the heat-pipe
can provide an entire equilibrium melting curve [3], heater voltage is, of course, adjusted to keep the heat
i.e. versus amount of sample melted. The shield (i.e. the heat pipe) at the sample temperature.
sample can easily be cooled: by inserting cold metal
rods into the thermometer well, known amounts of heat 3.1.3 Tin
can be removed from the sample (after determinining
the enthalpy of the rod, which is a relatively simple The Sn furnace is also an adiabatic furnace. Its shield
procedure). Realization of the Al point for calibrations is, however, only an Inconel-encased copper cylinder,
is performed conventionally: the liquid ingot (at about 10 cm i.d. and 30 cm high. (If the copper were not
3 C above its melting point) is lifted out of the furnace protected by its Inconel casing, then it would gradually
for 20 s, re-inserted, and the furnace temperature disintegrate due to oxidation, especially when the
lowered to that of the melting point of Al by turning the furnace is used at higher temperatures.) To realize the
heat-pipe heater off for 1.5 min. After inserting the cold Sn point, the liquid sample, kept overnight in another
PRT, Al point readings may be taken after about 20 min. furnace 3 C to 4 C above the melting point, is removed
At this setting of the heat pipe (+ 3 C relative to the from the furnace for 4 min (recalescence following a
melting point), if the PRT is left within the thermometer reproducible supercool of 14 C ± 1 C normally begins
well of the Al ingot then its resistance will gradually in about 3 min) and then inserted into the adiabatic
rise by the equivalent of about 0.5 mK until the sample furnace kept permanently at the melting point of Sn.
melts off in about 5 hours. For an adiabatic melting A cold metal rod (6 mm diameter and 15 cm long) is
curve determination the heat-pipe voltage is of course then inserted into the thermometer well for 3 min. After
adjusted to, and kept at, the melting point. inserting a cold PRT, its reading can be taken in 20 to
30 min. The melting range of the Sn sample is less
3.1.2 Zinc than 0.5 mK. As a matter of curiosity, this tin sample
was acquired about thirty years ago, under the name of
The Zn furnace is, similarly, an adiabatic heat-pipe “spectro tin”, from the Vulcan Materials Company. We
furnace using a 10 cm i.d. Cs heat pipe from which have recently purchased some 99.9999 % pure Sn but
the sample is insulated. Again, starting with the molten its melting range is a little broader (about 1 mK).
sample (prepared overnight by maintaining the heat pipe
about 3 C above the melting point), nucleation of the 3.1.4 Indium
external surface is induced by lifting the ingot out of the
furnace for 1 min and lowering the furnace temperature The realization of the In point consists of an outside
to near the melting point by switching the heat-pipe nucleation of the molten In ingot for 3 min followed
heater voltage off for 3 min (the temperature of the heat by inside nucleation using a 6 mm diameter metal
pipe is monitored by Pt-Au thermocouples spot-welded rod for 2 min. Readings can be taken 10 min after
to it near the top and middle). Nucleation of the inside insertion of the cold PRT. This fixed point is housed in

Metrologia, 1999, 36, 117-139 119


J. Ancsin and E. Mendez-Lango

a conventional furnace. A block heater is wound around


a sizeable Al cylinder whose bore is slightly larger than
the diameter of the In ingot.

3.1.5 Gallium

The Ga calorimeter and the melting behaviour of the Ga


sample are described in [4]. During the course of this
work, the Ga point was normally realized on Monday
mornings and maintained at its melting point all week.
PRTs were calibrated in the Ga cell whenever required.
The regulator, controlling the temperature of the bath
containing the Ga calorimeter, was turned off on Friday
afternoons. The Ga point is realized as follows: the
temperature controller of the water in a metal Dewar
that contains the home-made calorimeter housing the
commercial Ga cell, is turned on. In about 1 hour
the bath temperature reaches near 30 C. The sample
heater is then turned on for 20 min, supplying 18 W
of power. This energy of 21.6 kJ is sufficient to raise
the temperature of the Ga cell to the melting point and
additionally to supply about one-third of its total heat
of fusion (about 35 kJ).

3.1.6 Mercury

Our quartz Hg cell (see Figure 2) is home-made. It


contains 340 g of 99.999 99 % pure Hg. The cell was
specifically designed for simple, rapid, and accurate Figure 2. Mercury cell.
calibration of PRTs and not for the study of the
melting properties of Hg, which involves lengthy
experimentation. To realize the Hg point the cell is small enough to be carried as personal luggage.
with the PRT within its thermometer well (containing The only auxiliary items required for a realization of
alcohol) is inserted into a small Dewar of alcohol the Ar point are a rotary pump, liquid nitrogen, and
cooled to about – 60 C using dry ice. (It is important a Dewar (minimum 8 cm i.d. and 45 cm deep) to
that the portion of the cell that contains the Hg be accommodate the vacuum chamber. No temperature
entirely submerged in the cold alcohol bath, otherwise control of any kind is needed. The sample chamber
the sample will be only partially frozen, leaving large (1.75 cm diameter and 18 cm high, concentrically
vertical temperature gradients along the sample and surrounding the lower part of the 7.9 mm diameter
yielding only a poor-quality partial melting curve.) The and 40 cm long stainless-steel thermometer well) is
Hg freezes in 3 to 5 min, as indicated by the PRT. (Also, insulated from its surroundings only by a modest
the frozen top of the Hg is clearly visible.) The cell is vacuum and the many layers of thin Mylar wrapped
then removed from the bath, wiped clean of the dripping around the sample chamber. The Ar point is realized
alcohol, kept at ambient temperature for a minute or as follows: with the PRT within the thermometer well
so, and inserted into a hole drilled into a rectangular and the vacuum chamber open to ambient pressure, the
Styrofoam block, 10 cm 10 cm 30 cm high. The apparatus is lifted and the vacuum chamber is inserted
cell reaches the melting point in 5 to 10 min and melts into a Dewar containing liquid nitrogen. Then, using a
for 30 to 40 min. During melting, the temperature rises small rubber tube, the cooling coil wound around the
gradually by about 0.2 mK then the Hg suddenly melts sample chamber is connected to a can of liquid nitrogen.
off. Because the melting range of high-purity Hg is so The flow of liquid nitrogen through the coil rapidly
narrow, there is no need to realize the entire melting cools the sample chamber, as indicated by the pressure
curve during each calibration. Thus the entire process of gauge of the 5.5 litre reservoir (the absolute pressure
realization and calibration can be completed accurately at room temperature is 3 atmospheres, which is similar
in about 20 min. to that in a car tyre). The Ar condenses and freezes in
less than 5 min starting from room temperature. After
3.1.7 Argon cooling the sample to near the temperature of liquid
nitrogen, the flow of liquid nitrogen through the cooling
A small Ar triple-point realization cryostat was coil is stopped, the vacuum chamber is evacuated and
developed recently. The apparatus is self-contained and the sample heater, wound around the sample chamber,

120 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

is turned on (supplying about 0.5 W of power). The 3.2.4 Gallium


sample is heated until the PRT indicates a constant
temperature, and heating continues until slow run-off The Ga point is realized by heating the solid sample
begins, in 5 to 10 min. The heater is then switched in a furnace at 35 C for 1 hour and then reducing
off. After a few minutes the sample, overheated by the furnace temperature to 30 C. A steel bar heated
2 mK to 3 mK, cools back to the melting curve, the to 90 C is then inserted into the thermometer well
temperature stabilizes, and the resistance of the PRT can for 1 min. Following insertion of the PRT at room
be determined. The PRT can then be removed, replaced temperature, resistance readings may be taken after
by another, the sample re-frozen using the cooling coil, about 20 min.
and the procedure repeated for this new PRT. With a
little practice, a complete realization and calibration can 3.2.5 Mercury
be performed in about half an hour. The accuracy of
the calibration depends on the purity of the sample and, The Hg cell is made of quartz. It is similar to that
of course, on the reproducibility of the PRT. shown in Figure 2 except that it has no spacers
and contains more Hg. The realization of the triple
point of Hg is achieved as follows: the Hg is frozen
3.2 At the CENAM
within a liquid-nitrogen-cooled double-walled copper
tube located within a Styrofoam block. The sample
All furnaces were purchased. Three-zone furnaces are freezes in about 20 min. Then the flow of liquid nitrogen
used for In, Sn, Zn. A potassium heat-pipe furnace through the double-walled copper tube is stopped and
houses the Al point. Cells of Al, Sn, In, and Ga were heating begins, using the heater wound around the
purchased; those of H2 O and Hg are home-made. Water copper tube. When the sample approaches the triple
triple-point cells are housed in a commercial liquid bath. point within 1 C the heater is turned off. Melting
A cryostat used for realizing the Ar point was developed begins in a few minutes and lasts for about 40 min.
at the NRC. It can house four PRTs. The fixed-point The melting range is about 0.2 mK.
realization procedures are as follows:
4. The combined reproducibility
3.2.1 Aluminium of fixed points and PRTs

The uncertainty of the realization of a fixed point cannot


The liquid sample, 3 C above the melting point of Al, be determined directly. It is usually described in terms
is removed from the furnace for 1 min, then reinserted. of the resistance ratio of the PRT, .
The furnace temperature is set at 0.1 C above the The relative uncertainty of this quantity is
melting point. The PRT is inserted cold into the cell
and resistance readings may be taken in 30 min. If a (1)
completely frozen sample is melted continuously with
the furnace setting 0.1 C above the melting point, the One should bear in mind, however, that ,
sample has a total melting time of about 3.5 hours and i.e. the uncertainty in the measured resistance at a given
a melting range of 2.5 mK. fixed point, arises from both the irreproducibility of the
resistance of the PRT and the irreproducibility of the
realized fixed point at the site of the sensor of the PRT;
3.2.2 Zinc i.e. . Similarly, at
the water triple point, .
Outside nucleation of the liquid sample is performed Thus
for 1.5 min. Inside nucleation is performed by inserting
a 6 mm steel rod for 1.5 min. Resistance readings may
be taken 50 min after inserting the cold PRT. This ingot
has a melting range of about 0.5 mK.
(2)
3.2.3 Tin, indium
The uncertainty of the realization of the fixed point
Both the Sn and the In points were realized by itself is the second term of (2). There is a disadvantage
nucleating freezing of the external surface of the molten in using the above expression: one cannot study the
samples for 4 min and nucleating the inside surface, reproducibility of a fixed point without including the
using a 6 mm steel bar, for 2 min. Following insertion uncertainty associated with the determination. It
of the cold PRT, resistance readings can be taken after is possible, however, to take a different approach. If
40 min in the case of Sn and after 30 min in the case the triple point of water is treated as only one (very
of In. stable) member of a set of fixed points, then, as shown

Metrologia, 1999, 36, 117-139 121


J. Ancsin and E. Mendez-Lango

below, the last two terms can be eliminated from 0.5 mK (see Figure 13, from 7 to 17 July). Thus
the above expression. Yet, because one can observe it is extremely unlikely, using consistent realization
only the combined reproducibility of the PRT and procedures, that the realized sample temperatures at the
the fixed points, if the fixed-point cell has only one usual site of location of the PRT within the thermometer
thermometer well (represented by the first two terms on well will lie outside the melting range. It is of course
the right-hand side of (2)) it is not possible to separate, possible to induce sizeable temperature gradients along
with certainty, the reproducibility of the PRT from the thermometer well (for example, by inadvertently
those of the fixed points. However, as a set of fixed subjecting the sample to excessive vertical heat losses),
points is realized repeatedly, a certain pattern in the and hence to entirely upset the desired equilibrium
measured data becomes apparent, allowing some major temperatures. An “immersion test” would reveal such
uncertainties arising from the PRT to be separated from problems.
those arising from the fixed points. Because no melting In our case all our adiabatic furnaces and sample
point can lie outside the melting range, data points containers were specifically designed to minimize
obtained during repeated calibrations that lie outside vertical heat losses, either by keeping part of the
a consistently observed melting range are probably insulation above the sample (within the sample tube) at
caused by irreproducibility of the PRT and not the the sample temperature using a heater and thermocouple
sample. (If the fixed-point cell has two thermometer arrangement, or by leaving a few centimetres of the
wells then, of course, one can observe the first term space above the sample uninsulated (and by removing
of (2) exclusively.) We now consider the experimental the insulation between the heat pipe and the sample
data for the following PRTs. tube in this region), thus exposing it directly to the
temperature of the adiabatic shield. The space above
4.1 Experimental data the sample is thus kept near the sample temperature
and at the same time this section of the PRT stem is
4.1.1 YSI 1172 preheated.
We now consider the Al data. We have already
Figure 3 shows the overall resistance change during the noted that the resistance of this PRT gradually increases
indicated period at the indicated fixed points. It may be on thermal cycling. This is shown clearly in Figure 5.
seen that during the course of these measurements the We also noted that discrete changes in resistance take
resistance of the PRT gradually increased at all fixed place unpredictably: for example, the Al points of
points by the equivalent of several millikelvins. It is 30 April to 5 May and of 5 to 11 May appear to
improbable that all of our fixed points (including the lie along different parabolas.
water triple point) drifted several millikelvins during One might imagine such behaviour if the sensor
this period. Therefore it must be the PRT that gradually wire were prevented by its supporting structure from
drifted away from its initial state as a result of returning to its equilibrium position after thermal cycles.
repeated thermal cycling. On careful perusal of the As the thermal cycles are repeated, the position of the
data accumulated during the period when a set of fixed sensor might, accumulatively, move further and further
points was realized daily (or almost daily), see Figure 4, from its equilibrium location until the accumulated
one can see that these changes took place in discrete stresses become larger than the restraining influence
steps. Consider for example the measured values of of the supporting structure and then, like a compressed
, , , and on 29 and 30 April spring, the sensor wire would gradually begin to drift
and 1 May. These points all lie along straight lines (with back towards its equilibrium position, approaching
very little scatter). The trend of the values measured on closer and closer with each thermal cycle (like the points
4, 5, 6, 7 May, on the other hand, is entirely different. of 29 April to 1 May and 4 to 7 May of the upper two
It is more parabolic than linear. It is unlikely that all of graphs of Figure 5). After each drift period the sensor
these fixed points would have changed simultaneously might retain a small, permanent, residual deformation,
in such a strikingly similar way. It is more likely that perhaps caused by creep, accumulating it indefinitely.
the PRT drifted in such a manner. If one were to This may be what is seen in Figure 7 (the Al graph in
carry out a least-squares fit of the data to a linear particular). If this is more or less the case, then it is
equation during the period 29 April to 1 May, and to a difficult to see how the reproducibility characteristics of
quadratic equation during the period 4 to 7 May, then PRTs can be much improved upon in their present form.
the residuals would represent the maximum combined Some PRTs may well be remarkably reproducible
irreproducibility of both the fixed points and the PRT on thermal cycling, yet others, apparently made the
(first two terms in (2)) during its well-behaved stages. same way (and sold for the same price), would not
Such an analysis is carried out below. be. This would make the manufacture of PRTs an art,
Before discussing the results for Al it should be not a science. The difference between reproducible
emphasized that the melting range of our Al sample and irreproducible PRTs might be caused by such
from the realized state of about 30 % solid fraction small details as the quality of flame-polishing of the
to melt-off was observed to be consistently about edges of some quartz capillaries, the sensor-supporting

122 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 3. Drift of YSI 1172 during repeated calibrations.

arrangement, the detailed shape of the sensor wire, etc. (see Figure 5). All points between 30 May and 16 June
Scientific (reproducible) improvements could perhaps lie within the 0.75 mK scatter band, although the PRT
be made either by drastically reducing the differential still drifts slowly. The drift pattern of both and
thermal expansion between the sensor and the rest from 6 to 9 June and from 13 to 16 June is similar
of the PRT structure, or by arranging the supporting (see Figure 5). As a result the values in this
structure in such a way that it does not interfere period have a similar scatter band, and the scatter of
with the dimensional changes of the sensor caused by points has a similar pattern. Figure 6 shows a summary
temperature changes. of all values obtained during these reproducibility
In time, the drift of this PRT gradually slowed. studies.
By June, its resistance at the Zn point had more or To briefly summarize the work with YSI 1172, all
less stabilized (see Figure 3) and it became relatively fixed points were realized to within about ± 0.2 mK or
reproducible apart from the slight periodic drift pattern less, as shown below, yet the values of ,

Metrologia, 1999, 36, 117-139 123


J. Ancsin and E. Mendez-Lango

Figure 4. Similarity of drift pattern of YSI 1172 at several fixed points.

and spread across ranges of about 5 mK, 2 mK as seen in Figure 7. From 12 April the calibrations
and 1 mK, respectively. This clearly shows that the also included the Al point. The first Al point soak for
accuracy of temperature measurements beyond the In 30 min reduced the of the PRT from 25.4730 to
point using this particular PRT is limited by the PRT 25.4724 , a drop of 6 mK. An additional 20 hours of
and not by the fixed points. soaking at the Al point reduced its by a further 1 mK.
Clearly, the gradual resistance increase had been caused
4.1.2 YJ 152 (a Chinese PRT) mainly by considerable oxidation of the sensor prior to
its insertion into the Al point. Afterwards its resistance
This PRT was brought from the CENAM. It was remained more or less constant, within the equivalent
repeatedly calibrated at the NRC up to the Zn point of a few millikelvins, apart from the periodic drift
until 12 April. During this time its resistance gradually pattern visible at all fixed points. As seen from these
increased at all fixed points, in an erratic manner, data, this PRT is not capable of reliable temperature

124 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 5. Drift of YSI 1172 at the Al, Zn and H2 O points.

measurement with sub-millikelvin accuracy if it is Its resistance also increased gradually, at all fixed
subjected to sizeable temperature changes following points, after a few days of relative stability. Figure 10
its calibration. None the less, the measurements carried shows its behaviour at the indicated fixed points during
out at the CENAM between 18 and 30 May blend into this time interval. However, unlike PRT YJ 152, the
the overall data of the Al, Zn, Sn, In, Ga, H2 O, and increase in its resistance was more or less unaffected
Hg points within about 1 mK (Figure 8, see also by the inclusion of the Al point in the calibrations.
graphs, Figure 9). This shows that most of the resistance increase was not
caused by oxidation.
4.1.3 YJ 243 (a Chinese PRT) Many users do not bother to anneal their PRT
following its use around 650 C. In order to simulate
This PRT was also brought from the CENAM. It was this situation, from 11 to 19 April the PRT was
repeatedly calibrated up to the Zn point until 11 April. calibrated in a quenched state. Figure 10 includes

Metrologia, 1999, 36, 117-139 125


J. Ancsin and E. Mendez-Lango

Figure 6. Fixed-point resistance ratios for YSI 1172. (N.B. The freezing points of all these substances were realized
within ± 0.2 mK or better.)

the results of those measurements. As shown in Perhaps the sensor, at one point, became restrained
the figure, the quenched is, as expected, higher by the quartz components while subjected to repeated
than the annealed one following determination. thermal cycles, resulting in such an unpredictable shift
Paradoxically, however, the quenched values of , of its electrical resistance. Similarly, the initial set of Sn
, and are lower than the annealed and readings at the CENAM (from 18 to 22 May, see
ones. The initial drop of values on 11 April could be Figure 11), before the resistance of the PRT at any other
attributed to slight deoxidation of the PRT during the Al fixed points was determined (an attempt to use it as an
point calibration. However, the sudden increase of the Sn monitor), indicates another shift relative to the value
quenched values on 19 April, clearly seen in the of 15 May last determined at the NRC. This, however,
and graphs of Figure 10, is probably caused by may be attributed to a change of dimensions of the
a sudden change in dimensions of the PRT sensor. sensor during its transport to Mexico. It is worth noting

126 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 7. Drift of YJ 152 at some of the fixed points. Until 12 April the PRT was calibrated up to the Zn point only.

that during this temporarily stable state of the PRT (18 determined at the NRC from 13 to 15 May and 5 to
to 22 May) it gave a highly reproducible value for the 11 June in spite of the visible drift at all fixed points.
Sn point at the CENAM. The values obtained in the (The agreement of the Zn, Sn, In, Ga and Hg points
period 18 to 22 May were 48.617 93 ; 48.161 794 ; between the CENAM and the NRC has already been
48.161 787 ; 48.161 784 and 48.161 800 , respec- shown [1] to be within 0.5 mK.) It should be recalled
tively (a total scatter band of 0.16 mK) with the corre- that all resistance values in all figures are relative to
sponding values of 25.445 005 ; 25.445 017 ; the same standard resistor and all values are relative
25.445 017 ; 25.445 018 and 25.445 026 (a to the same water triple-point cell. A summary of the
scatter band of 0.21 mK). On the other hand, the values values is given in Figure 12. Again, the overall
of and following Al point determinations scatter indicates the capability of the PRT rather than
(26, 27 May) blend well into the corresponding values the reproducibility of the fixed-point temperatures.

Metrologia, 1999, 36, 117-139 127


J. Ancsin and E. Mendez-Lango

Figure 8. Drift of YJ 152 during repeated calibrations up to the Al point. The Zn, Sn, In and 0 graphs are
those of Figure 7 plotted with increased sensitivity.

4.1.4 T-5 The current and potential leads are of the same gauge
as those of the sensor. The PRT is 60 cm long and it
This PRT was made at the NRC about one year ago. has only six baffles altogether: two within the sensor
Its sensor consists of eight axially parallel Pt wires and four along the leads. The third baffle is about 1 cm
0.38 mm diameter and about 2.5 cm long. The wires are above the sensor, the fourth is 1 cm higher, the fifth
threaded through eight holes drilled near the perimeter is about half-way up the stem and the last one is just
of two quartz baffles, 5 mm in diameter, about 0.5 mm below the head. The PRT is open to ambient air.
thick, and about 2.3 cm apart. Every second lead of It may be of interest to describe our attempt to
the sensor is located within a ceramic tube in order to stabilize this PRT for high-temperature use. After its
prevent the leads from short-circuiting with one another. assembly, it was soaked overnight at 960 C and in the

128 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 9. Fixed-point resistance ratios for YJ 152. The large scatter of values is caused by the erratic behaviour of the PRT.

morning its quenched value of was determined in units of 10–6 ) 22; 15; 7; 6; 0; – 1. The PRT was
to be 3.375 703. It was then soaked again at 960 C for then soaked from the Al to the Ag point in 50 C
460 hours (from time to time its was determined steps, spending 70 to 120 hours at each temperature.
to check for changes that may have taken place) and The value was determined at each temperature
then it was calibrated in a quenched state. The following and was found to decrease during the soaking from
values were found for , , , 3.375 166 to 3.375 099 (the equivalent of about 20 mK).
, and , respectively: 3.375 144; Unfortunately, after an additional 230 hours of soaking
2.568 360; 1.892 487; 1.609 590; 1.118 100; 0.844 193. at 960 C, the sheath developed unusual, foamy sections
The PRT was then left at the Al point for 4 days near the top of the sensor and around the first baffle
and recalibrated again. This soaking caused the above above the sensor. Eventually, after further soaking at
values to increase by (from Al to Hg respectively, 960 C, the sheath cracked. It may be that the clearance

Metrologia, 1999, 36, 117-139 129


J. Ancsin and E. Mendez-Lango

Figure 10. Drift of YJ 243 during repeated calibrations. Until 10 April the PRT was calibrated up to the Zn point only.

between the baffle and the sheath was too small. Or, may be that the contamination was caused by our
during one of the thermal cycles, the baffle may have rather casual cleaning of our newly acquired re-bored
become lodged at an angle, rupturing the sheath during sheath and by the handling of the PRT components
cooling. We have since obtained some 7 mm quartz during re-assembly. Certainly, the experience of making
tubing of slightly oversized bore (over 5.2 mm) in and using PRTs provides an excellent opportunity to
order to avoid similar ruptures in the future. The broken observe at first hand the relation between cleaning
sheath was changed to that of the oversized bore and procedures and the resulting drift of values, the
the PRT was put back to use on 22 May, as shown in dependence of the immersion characteristics on the
Figure 13. The change in sheath resulted in inadvertent (strategic) location of baffles, and the effect of the
contamination of the sensor (note the decrease in gap between the sensor and the sheath on self heating,
by the equivalent of 0.4 C at the Al point). It etc.

130 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 11. Drift of YJ 243. Some of the data are those of Figure 10 plotted with increased sensitivity.
Time interval of measurements at the CENAM is indicated.

4.2 Reproducibility of fixed points As shown in the top and bottom graphs of Figure 13,
the PRT remained quite stable. The elimination of large
As shown in Figure 13, the resistance of PRT T-5 temperature changes apparently removed much of the
also changed with use. Periods of relative stability are irreproducibility. This would of course be expected if
followed by periods of increase in resistance. In order most of the irreproducibility of the resistance of our
to find out if the resistance changes are caused by the PRTs is caused by the difference in thermal expansion
relatively high-temperature soaking or by thermal cy- between the sensor and the rest of the PRT. The rel-
cling, the PRT was left permanently in the Al point cell atively stable temperatures (compared with the change
during the period 7 to 17 July (240 hours). The Al point of 660 C when is determined) left stable.
was realized daily, in our customary way, except that The data plotted in Figure 13 (7 to 17 July) were
the PRT was not removed from the thermometer well. least-squares fitted to a linear equation. The residuals,

Metrologia, 1999, 36, 117-139 131


J. Ancsin and E. Mendez-Lango

Figure 12. Fixed-point resistance ratios for YJ 243. Time interval of measurements at the CENAM is indicated.

i.e. the difference between the measured and the In order to test the reproducibility of the Zn point,
calculated values using the deduced equation, are we inserted the PRT into the Zn cell on 21 July in an
plotted in Figure 16 for both the realized and the melt- “annealed” state (after the Al point determination the
off (liquidus point) values, obtained during the above temperature of the PRT was reduced from 660 C to
period. The slope of the linear equation indicates a PRT 450 C in 1 hour in an annealing furnace). The PRT was
drift of 0.02 mK/day at the Al point. The residuals show left in the Zn cell permanently until 27 July, except to
that the combined standard Type A uncertainty of our determine its each time the Zn point was realized. As
Al point, as we customarily realize it for calibrating shown in Figure 14, there is very little scatter relative to
PRTs, i.e. the first two terms of (2), is within a scatter the smooth lines drawn through these points of
band of ± 0.14 mK and that the melt-off temperature is and . The points (from 21 to 27 July) show a
reproducible within ± 0.12 mK. parabolic drift pattern and were therefore least-squares

132 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 13. Drift of T-5 during Al point soak and repeated calibrations (see text for details).

fitted to a quadratic equation. The residuals of cycling and of oxidation is apparent (see Figures 14
for these six successive realizations of the Zn point lie and 15). The residuals of quadratic fits of the data
within a band of uncertainty of ± 0.07 mK (see Figure of 28 July to 7 August lie within scatter bands of
16). This quantity represents the combined standard ± 0.11 mK, ± 0.1 mK and ± 0.1 mK, respectively, for
Type A uncertainty (the first two terms of (2)) of our the Zn, Sn and In point realizations (see Figure 16),
routine Zn point realization. indicating these values for their combined standard
Type A uncertainties.
4.3 Repeated calibrations up to the Zn point
4.4 Repeated calibrations up to the Al point
From 28 July to 7 August, the values of ,
and were also repeatedly determined, leaving Between 10 and 15 August, the PRT was calibrated
the PRT soaking in the Zn cell overnight. Again, the daily up to the Al point. The initial drop in resistance
drift of the fixed-point resistances as a result of thermal at all fixed points relative to the values of 7 August

Metrologia, 1999, 36, 117-139 133


J. Ancsin and E. Mendez-Lango

Figure 14. Drift of T-5 during Zn point soak and repeated calibrations (see text for interpretation of data).

(see Figures 14 and 15) was due to deoxidation of the likely that the PRT shifted. A perusal of the values of
sensor wire during the first Al point calibration. With 13 August in Figures 13 and 14 indicate the direction
repeated calibrations, the PRT resistance at all fixed of the deviation of the values from the established drift
points began to drift (linearly) due to thermal cycling pattern.
as seen in Figures 13, 14 and 15. The residuals of a If, for example, the sensor wire rapidly expanded
linear fit of the data from 10 to 15 August are shown in when inserted (on 13 August) into the Al cell up to
Figure 16. It is interesting to note that values obtained some temperature between 420 C and 660 C and
on 13 August, at all fixed points, show a remarkably then, perhaps as a result of friction between the
large deviation from the rest (see Figure 16). It is sensor wire and the ceramic spaghetti insulation, was
unlikely that all fixed points (including the water triple unable to reach its unrestricted equilibrium position,
point) changed simultaneously on that day. It is more being longitudinally compressed at the Al temperature,

134 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 15. Figure 14, continued.

it would then have a slightly larger diameter and sizes between thermal cycles. Because the anomalous
thus a lower resistance than its equilibrium value values of 13 August in Figure 16 are clearly due to
(see Figure 13). At lower temperatures, again such the irreproducibility of the PRT (first term of (2)) they
friction might prevent the PRT sensor from reaching its should not be included when estimating the realization
equilibrium size. Being stretched, the wire would have uncertainties of the fixed points (second term of (2)),
a smaller diameter and thus a higher resistance than whatever the reason for the irreproducibility of the PRT.
its equilibrium value at all fixed points below Al. This
type of behaviour seems to be indicated in Figures 13,
14 and 16. The next heating cycle to the Al point (on 5. “Immersion test”
14 August) probably dislodged the sensor wire from its
restrained position, slightly changing its location within In order to investigate the temperature distribution along
the spaghetti and allowing it to resume its equilibrium the ingots, the resistance of the PRT was measured

Metrologia, 1999, 36, 117-139 135


J. Ancsin and E. Mendez-Lango

Figure 16. Least-squares fitting the data of Figures 13, 14 and 15 (see text for details).

along the thermometer well. Starting from the bottom Although near the bottom of the wells the trend of
of the well, the PRT was raised in 1.2 cm steps, and at the temperature variations along the samples is similar
each step a reading was taken after its value stabilized to that expected due to the hydrostatic pressure effect,
about 5 min later. For Hg the immersion test was elsewhere the values seem to indicate more closely
performed in reverse order: the Hg point was realized either vertical heat exchange between the samples and
with the PRT held 10 cm above the bottom of the well, their surroundings or heat leaks along the leads of the
which was entirely filled with alcohol. As the PRT PRT to or from the sensor, or possibly a combination
was lowered in 2.5 cm steps the excess alcohol was of both effects. The larger the temperature difference
expelled, leaving the well filled. Figure 17 shows the between the sample and room temperature, the larger
measured values. the observed temperature variation along the sample, as

136 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

Figure 17. “Immersion test”: observed temperature variations along the ingots using PRT T-5.

seen by the PRT. This apparent temperature variation As may be seen from the values in Table 1,
could be caused by heat leaks, mostly along the PRT, the vertical heat leaks are orders of magnitude larger
and by the hydrostatic pressure effect within the sample. than the hydrostatic pressure effects near the top of
If the temperature difference is very large between the the samples. By thermally anchoring the leads a few
sample and room temperature, then the effect of heat centimetres above the sensor, one hopes to dramatically
leaks will be very much larger than the hydrostatic reduce these vertical temperature gradients and thus the
pressure effect. If the sample temperature is relatively heat flow along the leads to or from the sensor. Still,
close to ambient temperatures as for Ga and H2 O it is a matter of opinion if the observed temperature
variation along the sample is considered to be real or
(compared with Zn or Al), then the hydrostatic pressure
only apparent and whether it is attributed mostly to
effect should make a relatively larger contribution to
heat leaks (primarily along the PRT) or to hydrostatic
the detected temperature variation along the sample. pressure effects within the sample.
Table 1 indicates some relative magnitudes. The vertical Our attempt to thermally anchor the PRT leads
temperature gradients are the values of [ (fixed point) is described above. During the “immersion test”, the
– (room temperature)]/distance from the top of the further the PRT is from the bottom of the well the
samples to the top of the furnace. The dependencies shorter the distance from the top of the sensor to the
of the hydrostatic pressures are those of Table 2 in the tempering region. Because the tempering region is near
text of the ITS-90 [5]. to, but not at, the sample temperature, the magnitude
Metrologia, 1999, 36, 117-139 137
J. Ancsin and E. Mendez-Lango

Table 1. Contrasting the magnitude of vertical heat leaks “pure” substances, one expects samples of 99.9999 %
with hydrostatic pressure effects. purity or better, with narrow melting ranges, to have
liquidus points within a temperature range of about one
Substance Vertical temperature Temperature variation along millikelvin. This is only a crude estimate, however,
gradients/(K/cm) the sample arising from
hydrostatic pressure/
which may turn out to be wrong in specific instances.
(mK/cm) Because the actual impurities and their concentrations
Al 21 0.016 in our samples used as thermometric fixed points are
Zn 13 0.027 unknown, we cannot directly estimate the component
Sn 7 0.022 of the uncertainty of a realized fixed point due to these
In 4 0.033
Ga 0.1 –0.012
impurities. What we can do, however, is to estimate
H2 O –1.3 –0.0073 the uncertainty within which a given fixed point is
Hg –10 0.071 consistent with the others, spanning a given temperature
range.
of the heat flow along the leads to or from the sensor Table 2 contains a summary of values obtained
becomes larger and larger as the distance increases from between 10 and 15 August using PRT T-5. The first
the bottom of the well to the PRT. Also, there is no two columns are self-explanatory. The third column is
way of knowing the efficiency of the thermal contact the average of six realizations in the above-mentioned
between the leads and the tempering region. For sample time interval. (There is a discrepancy between column 3
tubes with no tempering region above the sample, of Table 2 and the values in Figures 13, 14 and 15. The
i.e. where the sample tube is packed with insulating data in column 3 are the averages of values in the figures
material above the top of the graphite crucible, the corrected for hydrostatic pressures.) The fourth column
heat flow along the leads of the PRT should have a is the deviation of the resistance ratios in column three
more pronounced effect on the detected “immersion” from the reference function values, and the last two
values. Even if the numerical values coincide with columns are the residuals of least-squares fits of the
the expectation of either effect, it still remains to values of column four versus column three to quadratic
be proven by unambiguous measurements that the and cubic equations, respectively. The sensitivity of
observed numbers are due to one effect and not to values of varies from about 4 10–6 /mK at
the other. For capsule-type PRTs the leads are always the Ga point to about 3.2 10–6 /mK at the Al point.
very carefully thermally anchored to some components A set of careful realizations of a fixed point using a
of the cryostat that are maintained as close as possible slightly impure sample would yield a of small
to the sample temperature. For long-stem PRTs no such realization uncertainty, i.e. a narrow scatter band in
lead anchoring is possible. However, one has to attempt
Figure 16 indicates only that a well-behaved PRT was
to anchor the leads to near sample temperatures by some
used and that a given fixed point was carefully realized
means, even if the method turns out to be inferior to
the cryogenic routine. For example, a baffle cylinder using consistent procedures. However, the of
of substantial surface area, located a few centimetres an impure sample would deviate conspicuously from
above the sensor (within the PRT sheath), could be used a low-degree polynomial drawn through a set of
to enhance the heat contact between the leads and their versus points obtained for pure samples. Thus,
immediate surroundings. Also, such a baffle cylinder if any of the samples of Table 2 contained a significant
would probably reduce the minimum “immersion” amount of impurities, then such a graph would identify
necessary for accurate temperature measurements. the sample at a glance. These general ideas and some
of their implications are discussed in [6].
6. Consistency of fixed points The last two columns of Table 2 indicate that any of
the fixed points in the table is consistent with the rest,
High-purity samples obtained from different manufac- within ± 0.2 mK, using either a cubic or a quadratic
turers will probably have different types of impurities deviation function in the temperature range 0 C to
present in different trace amounts. Based on experience 660 C. In other words, the uncertainties arising from
gained from studying the melting/freezing behaviour of impurities are insignificant for all practical purposes.

Table 2. Fixed-point consistency.

Fixed point 1 (ref) 2 (av.) 1 2 Residuals Residuals


of quadratic fit of cubic fit
Al 3.376 0086 3.373 7399 0.002 2687 0.000 0000 0.000 0000
Zn 2.568 9173 2.567 4398 0.001 4775 –0.000 0002 0.000 0002
Sn 1.892 7977 1.891 9647 0.000 8330 0.000 0008 0.000 0008
In 1.609 8018 1.609 2358 0.000 5660 0.000 0007 –0.000 0007
Ga 1.118 1389 1.118 0340 0.000 1049 0.000 0001 –0.000 0001

138 Metrologia, 1999, 36, 117-139


The reproducibility of some thermometric fixed points and the accuracy of temperature measurements using PRTs

References 4. Ancsin J., Metrologia, 1993, 30, 105-106.


5. Preston-Thomas H., Metrologia, 1990, 27, 3-10, 107.
1. Ancsin J., Mendez-Lango E., Metrologia, 1996, 33, 6. Ancsin J., Metrologia, 1995/96, 32, 295-300.
415-420.
2. Ancsin J., Zhang Y., Metrologia, 1994, 31, 85-91.
3. Ancsin J., In Temperature: Its Measurement and Control
in Science and Industry, Vol. 6 (Edited by J. F. Schooley), Received on 23 November 1998 and in revised form
New York, American Institute of Physics, 1992, 343-348. on 18 February 1999.

Metrologia, 1999, 36, 117-139 139

You might also like