You are on page 1of 5

Registration Number ECN-000969

Revision Number 1st

Subject FDR nano/FDR AQRO (DR-XD 1000) :Releasing Service Manual (5th edition)

Date 2018/11/27

Department Medical Systems Business Division

Originator kayo okuda

Edited by TAKAHIRO IWAMOTO


NAOKO SHIMOKAWA
Approved by MASAHIRO MATSUSHITA

Affected Products
DR-XD1000

Scope of Application

Content

Applicable To (Serial Nos. Etc.)

Design Change Reason


Serviceability
Content

Action
Not Required(infomation only)
Content

Time Frame
Next Maintenance,Upon Failure,Other
Content
At installation

Completion Deadline
N/A

Notes

Estimated Work Time

ECN-000969 1/ 2
Implementation H
Completion H

Charge
Not Applicable

Affected Documents
Yes
Content
DR-XD 1000 Service Manual 018-241-05E

Description Of Change

Content
We are releasing the 5th edition of the Service Manual for FDR nano/FDR AQRO (DR-XD 1000).

Supply Of The Parts


Yes
When
With this ECN

Name/Code
DR-XD 1000 Service Manual / 018-241-05E

Quantity
Download from SIAS

Return Of The Replaced Parts To FUJI


No

Special Tools / Instruments


No
Content

Details
*Refer to the attachment for details.

(For US Only)C&R Assessment

SIAS Published
WW
Content

ECN-000969 2/ 2
393C193133E74174A925F6749354121A.xls

【Revision list】
Section Chapter Revision
Added descriptions for X-ray tube device label and
2. Labels
Safety Precaution rating/standard/identification label for the other overseas switching
2.1 Rating Indication Labels and Caution
(Safe) machine
Labels
Added label of In-vehicle kit (option)
Added the description for the Wet tissue holder
Product Specifications 1. Commodity Composition and
Added NOTE for the Area dose meter fixation kit
(Spec) Optional Items
Added the In-vehicle kit
Added NOTE for the "Collimator installation reference surface"
2. Weight and Dimensions
Added the dimensions of carry cart
Added the description for FUJIFILM-made AP
8. Other Specifications Added the Bluetooth BCR and Bluetooth USB Adapter Selection
Standards
9. Supported Disinfectants Added REFERENCE for the DR-ID 12xxSE.
Machine Description
1.2 System Configuration Examples Added NOTE for DR-ID 1200MC
(MD)
2. Block Diagram Updated to block diagram and signal diagram of new configured machine
5. Changes in PC-GNC Connection
Revised NOTE for new configured machines
Specifications
7. Handling the SE Battery Newly added
8. Setting the mAs Value Newly added
9. Software Overview Newly added the description for GNC Software and MC Software
Appendix 1. Block Diagram (For Older
Moved block diagram of older configuration machine from MD main part.
Configured Machines)
Revised the description for "Battery log: BT"
Troubleshooting (MT) 1.4 How to Use the Cart Part Log
Added "1.4.1 Example of Log Analysis"
Revised Checks and Workarounds on aging for the cart unit error
2. Error Code Table Add the description of all the error codes for the MC error
Added MC Manager error code
For each phenomenon occurring in the DR-XD 1000, sorted causes and
3. Troubleshooting
remedies in the table
Added "4.7 If the SE Cannot Be Charged", "4.8 If Irradiating Cannot Be
4. Parse Tree
Done", "4.9 When the Caster Lock Pedal Does Not Rise All the Way Up"
5.1 GNC board IP Address Resetting
Added note when using RU PC-TOOL
Procedures
6. Troubleshooting from FUJIFILM-made AP Added NOTE on password input at login
Log Added the description for checking DFS operation
Appendix 1. Aging Added "Warm-up aging log example".
Added the isolation flow
Appendix 2. 10753 Error Countermeasures Revised phenomenon and countermeasure
Added "3. Confirming Analysis via the Log"
Added the description for the structure of log file acquired by DR
Appendix 3. Procedures for Acquiring the Logs
maintenance tool
Appendix 4. How to Obtain the Image Files
Newly added
When There are A Abnormal Images
Appendix 5. Analysis Procedures When 11240
Line Disconnecting Errors Occur in Wireless Newly added
Communications
Appendix 6. Windows OS Recovery Procedures
Newly added
Using a System Repair Disc
Appendix 7. How to Handle a 1-LAN Port
Newly added
Which Was Set for a 2-LAN Port Configuration
Checks, Replacement
and Adjustment of ALL Revised screws for all items.
Parts (MC)
2.7 Left-side Cover (INV Board Side) Added NOTE for installing sequence for the screws
2.23 Caster Added the replace procedure for the new caster
2.24 Panel Lock Assembly Added NOTE before removal and for removal units.
2.30 AC Inlet Cover Assembly Newly added
2.31 DR Slot Base Cover Newly added
Revised the procedure.
3.1 Battery Assembly
Revised the check item after installation.
3.5 FET Board Revised INSTRUCTION for installation torque
3.6 INV Board Revised INSTRUCTION for installation torque
Added the description for the new type construction parts of the guide
4.1 Guide Rails rail.
Added NOTE for installation.
4.10 Arm Lock Release Handle Assembly Newly added
5.7 Operation Panel Relay LAN Connector Kit Newly added
6.8 LED Board Assembly Revised the description on increase and decrease the size of the light
Added NOTE for applying the screw lock during installation
6.10 Collimator Assembly
Revised NOTE for applying the grease during installation

1 / 3 ページ
393C193133E74174A925F6749354121A.xls

Added descriptions for "The label is required when purchasing the mono-
7.1 Top Cover of Mono-tank
tank top cover"
Added NOTE for applying the screw lock during installation
Revised INSTRUCTION for installation torque
7.7 Mono-tank Assembly
Added NOTE that the number of clamps is different depending upon the
machine.
Revised INSTRUCTION for installation torque
7.10 Torque Hinge (Large) Added NOTE when loosening the screws for attaching the torque hinge
(large) and the bracket
8.2 Panel Approaching Unit Added NOTE for installation.
8.3 Roller Assembly Added NOTE for installation.
8.6 SE Charging Terminal Assembly Newly added
8.7 DSC65A Board Assembly Newly added
8.8 DR Slot Housing Newly added
8.9 DR Slot Securing Member Newly added
8.10 DR Slot Damper Newly added
8.11 ESD67A Board Newly added
Appendix 1. Setting Country Code for the
Added requirements for setting the FUJIFILM-made AP country code
FUJIFILM-made AP
Added NOTE for the country code for the MC V16.5 or later along with
Appendix 2. Changing the SE Country Code
change in communication method between MC and SE
Appendix 5. Backup/Restoration for Windows 7 Added NOTE when "System restore disc" is available
Newly added the following replacement procedure
- Brake Wire
Appendix 10. Carry Cart - Brake Lever
- Disc Brake Caliper
- Wheel Assembly
Newly added
Appendix 11. Updating MC Software Version
(Including VA expanded security compliant)
Maintenance Utility
3.10 Adjusting the Filament Preheating Voltage Added NOTE when the exposure switch is released part way through
(MU)
5.9.10 Install SE CPU (0x0220) Newly added
5.9.11 Replace MP Revised the procedure
5.9.13 Battery Threshold Setting Newly added
5.10.6 LOG Backup Added the description for the log
6. Smart Log Acquisition Tool (SLAT) Newly added
Revised the parts and circuit diagrams to the latest information
Added the references of replacement procedure for each part
Service Parts List (SP)
Revised "List of Service Parts for Securing and Wiring" and "Table of
Screws / Washers Indication Symbols"
18A
Newly added
Parts that Need Labels (For USA)
18B
Parts that Need Labels (For Overseas Other Newly added
than USA)
Preventive
3. Removing the Cover Revised the procedures for the second and fourth years
Maintenance(PM)
4.1 Checking the Cart Handle Added the check item
4.2 Checking the Casters Added descriptions for non-slip sheet
4.4 Replacing the Monitor Arm Added the check item
4.8 Checking the Arm's Horizontal Rotation Added the check item
4.9 Checking the Arm's Center Pin Receiver Added the check item
4.13 Checking the Swiveling Mechanism Newly added
4.14 Checking the Battery’s Tightness Newly added
4.15 Checking the Round Terminal of the
Newly added
INV/FET Board
7.2 Checking the Rotation Added the check item
8. Installing the Cover Revised the procedures for the second and fourth years
10.3 Adjusting the Filament Preheating Voltage Added NOTE for the exposure switch
11. Visual Inspection and Cleaning Added cleaning place

2 / 3 ページ
393C193133E74174A925F6749354121A.xls

Revised the items to the latest information


Installation (IN) 5.1 Items Supplied with DR-XD 1000PX
Added the references for each item
Changed the chapter configuration
6. Removing the Cover
Revised the procedure
Changed the chapter configuration
7. Installing and Confirming for the Battery
Revised the procedure
Added recommended DAP information
8. Installing the DAP Kit (Optional) Revised "Configuration of the DAP Kit"
Revised the reinstallation procedure of DAP
9. Attaching the Optional Items 1 Newly added
Changed the chapter configuration
10. Attaching the Cover
Revised the procedure
14.1 Installing the Operation Panel Assembly Revised the procedure in the case of new configured machine
16.8 Checking the Light Irradiation Field Revised the procedure
17. Difference between Default Setup of
Added "Software applications that is installed in PC for DR-XD 1000
DR-ID 1200 and DR-XD 1000 system
system".
(Reference)
18.2.1 Network Settings Added NOTE for the network adapter is connected to the GNC board.
18.2.3 Changing the Wireless Settings of
Added NOTE and REFERENCE on FUJIFILM-made AP setting items
FUJIFILM-made AP
18.2.4 Confirming the MAC Address of
Added the procedure
FUJIFILM-made AP
19. Upgrading MC Application Revised the procedure
Added NOTE that Internet Explorer 11 is needed in order to use the DR
20.1 Installing the DR Maintenance Software Maintenance Software.
Added NOTE for the VA expanded security mode
20.5 Updating MP Application Software Newly added
Added the description that the settings become effective after restarting
21. Configuration Setup the MC Manager if the settings changed
Added the setting item list
22.1 Registering SE on DR-ID 300CL Revised the procedure
Revised the procedure
22.3 Registering SE Serial ID on DR
Added NOTE that only the SE to be registered is set in the DR slot, and
Maintenance Software
do not power on the other SEs
Added NOTE that the communications method between the MC and the
SE is changed for the MC V16.5 or later, and the country code
22.4 Changing the SE Country Code corresponding to the new communications method is needed for setting
the SE country code
Revised the procedure
22.8 Installing the Machine-Specific Data Revised the procedure
Added NOTE to implement only when the network address of the system
23. Changing the IP Address of the System
is to be changed from the default value.
(SE/MP/MC)
Added procedures for changing the MP and MC IP address
Added INSTRUCTION when the defect corrections are implemented
27.4 Defect Calibration via the gain calibration conditions, the number of defects and the defect
sizes may not be correctly detected, and calibration errors may occur
28. Attaching the Optional Items 2 Newly added
Appendix Installation
Appendix 2. Recommended Wifi Adapter Updated available Wifi adapter
(AppxIN)
Appendix 4. Installing and Setting the Magnetic Added NOTE when using both the NFC card reader and the magnetic card
Card Reader reader
Appendix 11. Installing the NFC Card Reader
Newly added
for Contactless IC Cards
Appendix 12. Wireless Channel Used for the
Newly added
Access Point
Appendix 13. Installing the Operation Panel
Moved from IN main part
Assembly (For Older Configured Machines)
Appendix 14. Initializing the
Newly added
FUJIFILM-made AP
Appendix 15. Precautions When Attaching the
Newly added
In-Vehicle Kit
Appendix 16. Measuring Method of mAs Value Newly added
Appendix 17. Installing the Internet Explorer 11 Newly added

3 / 3 ページ

You might also like