LIASYS 330 CLINICAL CHEMISTRY SYSTEM
DESCRIPTION
ASSAY METHODS:
PROCESSING UNIT:
USER INTERFACE:
TEST SELECTION MODES:
‘TEST CYCLE:
‘THROUGHPUT:
OPTICAL SYSTEM:
LIGHT SOURCE
CUVETTE OPTIC LENGTH:
PHOTOMETER SENSITIVITY:
PHOTOMETER RESOLUTION:
SAMPLING SYSTEM
REACTION VOLUMES:
DILUTER SYSTEM:
WATER SUPPLY:
SAMPLE CAPACITY:
SAMPLE LOADING:
QC capaciry:
STAT POSITIONS:
REAGENT CAPACITY:
REACTION PLATE:
INCUBATION SYSTEM:
CUVETTE WASH STATION:
SAMPLING ROBOTIC ARM:
VOLUME LEVEL SENSING:
ISE MODULE!
BAR CODE READER:
OPERATING ENVIRONMENT:
POWER REQUIREMENTS:
DIMENSIONS:
WEIGHT:
DISTRIBUTED IN USA BY:
SYSTEM SPECIFICATIONS
WALK-AWAY BENCH-TOP, DISCRETE, FULLY AUTOMATED RANDOM ACCESS SYSTEM
CHEMISTRY AND IMMUNOTURBIDIMETRY - UP TO R1+R2+R3 REAGENTS
REAL TIME, EXTERNAL PC WITH MONITOR, KEYBOARD AND MOUSE
GRAPHICAL, USER FRIENDLY, WINDOWS XP BASED
DISCRETE, BATCH, PROFILE AND STAT.
FROM 18 TO 746 SECONDS - TEST RESULTS EVERY 18 SECONDS
UP To 330 TESTS PER HOUR = 200 CHEMISTRY TESTS PLUS 130 ISE
DOUBLE BEAM PHOTOMETER WITH 8 INTERFERENTIAL FILTERS FROM 340~620mm,
6V/10W HALOGEN BULB
10MM
RANGE:0.001 TO 2.5000 Abs
0.0001 Abs:
PRECISION: CV < 1% AT 3uL
300uL Min, - 670uL Max.
SYRINGLESS INTEGRATED; UP TO 3 REAGENT AT 1uL INCREMENTS EACH
(ON BOARD DI WATER TANK
64 SAMPLES IN 4 SLIDING RACKS
CONTINUOS, SAMPLE CUPS OR PRIMARY TUBES
1 SLIDING RACK FOR 14 CONTROLS AND CALIBRATION POSITIONS
1 SLIDING RACK FOR 14 STAT POSITIONS.
33 REAGENT BOTTLES IN 2 SLIDING RACKS (20 REFRIGERATED)
60 DURABLE, CHANGEABLE CUVETTES, INCUBATED AT 379C
WATERLESS AT 37°C- REAGENT PREHEATING CAPABILITIES
QC MONITORED 5 CANNULA SYSTEM FOR CUVETTE EMPTYING, WASHING AND DRYING
INTEGRATED ROBOTIC UNIT FOR SAMPLING OPERATION WITH OWN WASHING WELLS.
VIA INTEGRATED ROBOTIC UNIT PROBE
BUILT-IN EXTERNAL MOUNTED FOR RAPID ELECTRODE REPLACEMENT
OPTIONAL - BUILT-IN FOR SAMPLE PRIMARY TUBES - MANUAL FOR REAGENTS
RELATIVE HUMIDITY 20%-85% TEMPERATURE 18°C ~ 32°C
100-230v, 63Hz, 300W
16.5 H x25.5 Wx39.5 L
143 Ibs
CAMS) Diagnostics
AMS Dlagnosties
2410 Settlers Street, Unit A
Charieston, SC 29492
Tel 1 (866) 419-7839 (Tol Free)
Faxt (843) 277-0903
Service 1 (866) 369-2553 (Toll Free)
femal Info@amsclagnostcs.com
www.amsdiagnostics.com