Applicable for use outside the US

PRIOR TO CHANGE
Technical Specifications
Konelab 20
Konelab 20 is a random access clinical
chemistry system for routine chemistries,
electrolytes and specific proteins.
The broad range of CE marked applications
can be complemented with user-definable
tests.

Throughput Reagents
Workload dependent, being in typi- Continuous access to reagents
cal routine use up to 200 tests/hour. without interrupting test processing.
Time to first result is typically 3 to 12 External barcode reader for identifica-
minutes. tion. Real time reagent status clearly
displayed. Up to four reagent addi-
Samples tions / test possible.
Continuous access to samples with- ON-BOARD STORAGE:
out interrupting test processing. 35 positions in the cooled reagent
MAX ON-BOARD CAPACITY: disk.
84 routine samples and 5 STAT sam- REAGENT VIALS: 10, 20 and 60 ml
ples - 6 segments with 14 positions REAGENT VOLUMES: 2-250 µl; typi-
per segment. Integrated barcode cally 120 - 200 µl.
reader and cup/ tube recognition. Cuvettes
SAMPLE CUPS AND TUBES: Discrete disposable 12-cell cuvette.
0.5 ml and 2.0 ml cups, 5 ml and Continuous access to cuvettes with-
7 ml tubes, 10 ml tubes with tailored out interrupting test processing.
segments. ON-BOARD CAPACITY:
SAMPLE TYPES: Serum, plasma, 600 measurement cells - 50 multicell
urine, CSF. cuvettes with 12 reaction cells, typi-
SAMPLE VOLUMES: Possible range cally up to up to 3 hours walk-away
1-120 µl ; typically 2-15 µl and 50 µl time.
for Na+, K+ and Cl- tests.
Technical Specifications
Konelab 20
Applicable for use outside the US

Calibration Photometric Measurement Data Management


Linear, non-linear or bias calibra- Single channel interference filter Windows® XP workstation (includ-
tion. Method-dependent use of photometer with beam splitting ed) with graphical user interface.
individual calibrators or automati- reference- 11 fixed filters, others Data input online, by mouse, key-
cally diluted series from a stock available upon request. board or barcode reader connected
SPECTRAL RANGE: 340-800 nm. alongside with the keyboard.
calibrator. Automatic repeat for bias
MEASUREMENT TEMPERATURE: Different language versions avail-
correction available 37 oC able.
Quality Control MEASUREMENT PRINCIPLES: SAMPLE BARCODES IN USE:
Real time QC with multiple, Colorimetric, turbidimetric. Code 128, Code 39, USS Codabar,
user-definable Westgard rules. MEASUREMENT MODES: Interleaved 2 of 5.
Programmable control interval. Kinetic, end-point. LIS INTERFACE: ASTM 1394-91 or
QC chart printouts, daily and cumu- KINETIC MEASUREMENT: KONE Online.
lative reports. 30 sec to 60 min, max 12 points HARDWARE INTERFACE: RS-232
LIGHT SOURCE: Halogen lamp. RESULT REPORTS: Collated by
Sample and Reagent with linear absorbance range of patients, manual entry of off-line
Dispensing 0-2.5 A, resolution of 0.001 A and results allowing for fully collated
Externally and internally rinsed reproducibility of SD≤0.005 A result reports, results calculated
single probe dispensers equipped at 2 A. from both measured and off-line
with level sensing.
results. Automatic or ‘on request’
Dispensing with precision syringes ISE Measurement in the printouts, automatic STAT report-
driven by stepping motors. model Konelab 20i
REACTION END VOLUME: Direct potentiometry. Electrodes ing. Abnormal values and repeats
100-250 µl. for Na+, K+ and Cl-; upon request flagged automatically.
SAMPLE CARRY OVER: <1%. Li+. DATA STORAGE: Long term stor-
REPRODUCIBILITY: SAMPLES: Serum, plasma and age of patients with demographics
CV less than 2% for sample urine. including test and QC results and
volumes ≥2 µl and for reagent USABLE RANGE: (mmol/l) calibrations.
volumes >5 µl serum, plasma: Dimensions and Weight
K+ 2.0 - 10 Width 80 cm, depth 79 cm, height
Dilutions Na+ 100 - 200
Automatic sample pre-dilution. 115 cm, weight 130 kg. Separate
Cl- 55 - 150 workstation table.
Automatic post-dilution with both serum:
high and low secondary dilution Li+ 0.2 - 4.0
ratios. Ability to add the value of urine: Power Requirements
manual pre-dilution for the result K+ 20 - 200 100-240 V ±10%, 50-60 Hz ±5%,
calculation. Na+ 20 - 200 300 W Konelab 20, 350 W Konelab
DILUTION RATIOS: up to 1+120
20i.
Environmetal Conditions
Operating temperature range of Distilled Water Consumption
NOTE: The information and techni- 15-32 oC; humidity 40-85% (non Typically <0.5 l/h
cal specifications are subject to condensing). On-board storage with no external
change without notice. connections required.

Regulatory Ordering Codes


Code 895311_N CONFORMITY WITH: 981800 Konelab 20i
04 / 2007 98/79/EC IVDMD Directive. 981801 Konelab 20

Clinical Diagnostics Finland Tel.: +358 9 329 100


Ratastie 2, P.O. Box 100 Fax: +358 9 3291 0300
FI-01621 Vantaa www.thermo.com/konelab