Professional Documents
Culture Documents
0112-1642/88/000 1-0057/$06.00/0
© ADIS Press Limited
All rights reserved.
Summary Maximal oxygen intake d' 0:: /IIaJ may be the most physiologically sign(ficant and.
therefore, the most commonly measured parameter in the physiological assessment a/well
trained runners. To insure \'alid and reliable i' 0:: /IIax ralues. there are a number of prac-
tical concerns that must be addressed. An exercise modality must be selected that activates
the spec({ic muscle mass and duplicates the motion utilised in the runners' training pro-
gramme. Motorised treadmill running has been shown to allow the most efficient. valid
and reproducible (-0:: /II'" I'alues in a controlled testing em·ironment. A testing protocol
must be used that is easily administered. comfortablefor the runner. allows test completion
within IO minutes. uses grade increments at a constant speed, and is reproducible. Speeds
should be selected that approximate the runner's training pace. The criterion for the at-
tainment of i'o: /II"., include the lel'elling 0.0' 0.( C"O: with an increase in work stage. the
onset 0.( extreme exhaustion. the respiratory exchange ratio exceeding 1.1. blood lactate
lel'els approaching or exceeding 10 mmol/L. and ratings 0.( perceired exertion 0.( 19 or
lO. Qualit.\" control in terms 0.( standardisation 0.( procedures and calibration and main-
tenance 0.(equipment is essentialfor the minimisation o.(technological error..Heasurement
eqllipment must be a.O'ordable, dependable. easily calibrated. .\"ield reproducible data. and
hI.' I\'ithin the technical skill lerels 0.( the operating personnel.
v~ mu Testina in Runners S8
The measurement of maximal oxygen intake not undermine the importance of V02 max but must
(V02 max) determines an individual's respiratory, be interpreted in terms of the degree of homo-
cardiovascular, and skeletal muscular capabilities geneity of the sample population and type of ath-
for the uptake, transport, and utilisation of oxygen letes being tested. For example, the correlation be-
(fig. 1). If this integration of physiological mech- tween V0 2 max and marathon race pace (r = 0.78)
anisms is impaired due to pathology or as a result indicates a wide variation and poor predictability
of a sedentary lifestyle, V02 max will be reduced among a group of elite marathon runners (SjOdin
from expected values. & Svedenbag 1985). Even though there is a sub-
Because of its functional sianificance, V02 max stantial spread in V02 max and poor correlations
provides a good estimate of the potential of the among individuals in such a homogeneously trained
cardiovascular system and endurance exercise ca- population, a good relationship is seen between
pacity. In spite of the physiological significance of V02 max and performance when mean data are re-
V02 max, a wide variation in the relationship be- viewed (table I) or a population more heteroge-
tween V0 2 max and running performance (r - 0.08 neous in terms of V02 max is tested. Even though
to 0.91) bas been demonstrated (Astrand & Saltin other variables have been shown to have as high
1961; Conley & Krahenbuhl 1980; Costill et aI. or higher relationship with running performance
1976; Daniels et al. 1978; Davies et JLI. 1984; Froe- (Conley & Krahenbuhl 1981; Costill et aI. 1973,
lieher et al. 1974; Gibson et al. 1979; Hammond 1976; Daniels 1985; Daniels et aI. 1978; Farrell et
& Froelieher 1984; Londeree 1986; Moreira-Da- al. 1979; Kenney & Hodgson 1985; Kumagi et aI.
Costa et aI. 1984; Shephard 1984; Stamford 1976; 1982; Londeree 1986), V0 2 max is felt to give an
Sjodin & Svedenhag 1985). This variability does indication of the runner's overall maximal cardio-
fig. 1. A scheme illustrating the gas transport mechanisms for coupling cellular (Internal) to pulmonary (external) respiration. WIth
pennIsaIon from Wauerman at aI. (1887).
V01 max Testing in Runners 59
Table I. V02 max and personal records (PR) of elite marathon ance. Testing protocols and principles for assessing
runners other performance parameters have been reviewed
V0 2max PR Reference
by Londeree (1986).
(ml/kg/min) (h:min)
1. Exercise Modalities
70.9 2:23 Costill & Fox (1969)
74.1 2:16 Svedenhag & SjOdin (1984)
74.1 2:15 Pollock (1977) The specific exercise task employed for testing
79.0 2:13 Davies & Thompson (1979) has a definite influence on the V0 2 max response.
McConnell et al. (1984) found that the V0 2 max ob-
tained on a motorised treadmill was 10.2, 26.4, and
respiratory capabilities and sets the limits on the 28.6% higher than that obtained during upright leg
amount of oxygen that can be utilised for the dur- ergometry, supine leg ergometry and arm ergo-
ation of an event (Snell & Mitchell 1984). Since metry, respectively (table II). McArdle &. Magel
pacing is a function of both race distance and rel- (1970) found that subjects had a 9.9% lower V0 2 max
ative intensity, knowledge of V0 2 max assists the during leg ergometry than during treadmill testing.
runner and coach in determining the appropriate Of their subjects, the only 2 who had higher V0 2 max
fractional utilisation of aerobic capacity (%V0 2 max) values during leg ergometry were heavy recrea-
required to maintain a desired race pace .(Costill et tional cyclists. Other investigators have shown
al. 1973; Snell & Mitchell 1984). If the V0 2 max of similar differences between treadmill and leg er-
a competitive runner is low relative to the antici- gometry testing ranging from 5 to 11.2% (Astrand
pated need for a given event, a correction in train- 1967; Astrand & Saltin 1961; Diaz et al. 1978; Fair-
ing emphasis must then be addressed that will re- shter et al. 1983; Faulkner et al. 1971; Glassford et
sult in an enhancement of V0 2 max as well as the al. 1965; Hammond & Froelicher 1984; Harrison
ability to maintain exercise at high relative intens- et al. 1980; Hermansen & Saltin 1969; Hermansen
ities (Shephard 1978; Sjodin & Svedenhag 1985; et al. 1970; Kamon & Pandolf 1972; McArdle et
Williams et al. 1967). al. 1973; McKay & Banister 1976; Miles et al. 1980;
Since the measurement of V0 2 max is para- Neiderberger et al. 1974; Shephard et al. 1968;
mount for the assessment of running performance Smod1aka 1982; Stamford 1975, 1976).
capabilities (Daniels et al. 1978), many articles have When testing trained athletes, the cardiorespir-
discussed appropriate testing methodology. Inves-
atory training response is best demonstrated when
tigations involving runners have used testing pro-
the athletes activate the specific muscle mass util-
cedures that have been based on recommendations ised in their training programmes; therefore,
from the literature, have duplicated others or have
devised their own protocols that meet the specific
objectives of their study. Concerns and practical Table II. Mean maximal exercise values
considerations that must be addressed when meas-
Treadmill Leg Supine Arm
uring V0 2 max in runners include: (a) exercise mo- ergometry ergometry ergometry
dality; (b) protocol selection; (c) laboratory en-
vironment; (d) subject preparation and motivation; V0 2max 48.9 43.9 36.0 34.9
(ml/kg/min)
and (e) accuracy of the measure.
HRmax (beats/min) 187.1 181.2 164.6 175.8
Shephard (1984) published a detailed review of SBP max (mm Hg) 174.3 180.0 178.4 148.8
tests of maximal oxygen intake. The intent and RPP max (X 1(2) 328.4 326.8 294.0 261.3
scope of the present review is more limited and
Abbreviations: V02 max = maximal oxygen consumption; HRmax
will discuss specific practical concerns for testing
= maximal heart rate; SBP max = maximal systolic blood pres-
runners. Also. V0 2 max is only one physiological sure; RPP max = maximal rate pressure prOduct.
parameter that is predictive of running perform-
V~mu Testing in Runners 60
V02 max is dependent upon the mode of exercise track running. Davies et al. (1984) tested trained
and the specific trainilll of the athlete (Snell & runners on motorised and non-motorised tread-
Mitchell 1984). The more closely a laboratory can mills and found no differences in the V0 2 max
simulate the specific muscular action involved in values.
training during the test, the more objective and
Running must be employed to obtain valid
valuable the V02 max assessment (Davies et al.
1984). Corry & Powers (1982) demonstrated the
V02 max values when testing trained runners. Even
though a few studies have demonstrated similar
importance of using specifically trained muscle
groups by comparilll trained runners to trained V02 max values when motorised treadmills were
swimmers. They found that the runners had sig- compared to track running or non-motorised
nificantly higher V0 2 max values during treadmill treadmills, the vast majority of the literature sup-
running (70.2 ml/kg/min) when compared with ports the use of motorised treadmill running as the
swimmers (60.0 ml/kg/min). But, when an arm most effective modality for the laboratory analysis
pulling motion similar to that of swimming was of V0 2 max in runners. When using a motorised
used, the swimmers obtained significantly greater treadmill, running on a grade has been shown to
values (47.1 ml/kg/min vs 37.4 ml/kg/min). elicit greater and more reliable V02 mo values when
McArdle et al. (1978) found that following a nlD- compared with that obtained with flat treadmill
ning programme individuals improved their tread- running and speed increments. It is felt that run-
mill-obtained V02 miX by 6.3%, while their swim- ners will eventually be unable to maintain higher
ming V02 mix improved by only 2.6%. Earlier, they speeds without having reached their actual V0 2 mo.
had demonstrated that swim training did not sig- V02 values during submaximal and maximal track
nificantly improve V0 2 max values obtained during running have been shown to be highly correlated
a running test (Magel et al. 1975).
and not significantly different from that obtained
Moreira-Da-Costa et al. (1984) investigated
during same speed motorised treadmill running
V02 max in trained runners, cyclists and untrained
(McMiken & Daniels 1976). But the difficulty of
individuals to compare trained to untrained muscle
measuring V0 2 during track running and other
groups. The non-athletes and runners had II % and
12% greater V02 max values during treadmill ex- confounding variables (i.e. temperature, humidity,
ercise when compared with leg ergometry. On the and wind velocity) makes motorised treadmill run-
other hand, the cyclists obtained 7% greater V02 max ning the more reliable measurement procedure.
values during leg ergometry. They concluded that Leger and Mercier (1984) reinvestigated published
an ergometer which requires approximately the data and derived an average equation for the pre-
same muscular activity as is usually performed by diction of the gross energy cost of treadmill and
the subject should always be employed to evaluate track running:
the quantitative effects of training on cardiovas- Treadmill: V02 (ml/kg/min) = 2.209 + 3.163 speed
cular and respiratory functions. (km/h)
Others have also demonstrated that the differ-
Track: V0 2 (ml/kg/min) = 2.209 + 3.163 speed +
ence in V0 2 max values and the physiological re-
0.000525542 speed3
sponse to submaximal exercise can be due to the
utilisation of specifically trained versus untrained They concluded that there are wide variations
muscle groups (Jacobs & Sjodin 1985; Kamon & in the energy costs of running and specific equa-
Pandolf 1972; Pelchar et al. 1974; StrBDlme et al. tions need to be developed for male and female
1977). and good and bad runners. If the energy cost of
In terms of different running modalities, running is needed for research purposes, it should
McMiken and Daniels (1976) found no differences be measured because of the wide variations dem-
in the V0 2 max measured between treadmill and onstrated between individuals.
V0 2 max Testing in Runners 61
(IBP) recommended a standard protocol using a elevations (greater than 15%) that may result in a
constant speed and 2.S% grade increments every 2 loss of running efficiency or posterior leg or lower
minutes (lshiko 1978; Shephard et a1. 1968). Others back muscle soreness.
have agreed with using gradient increases and con-
stant running speeds throughout the test (Gass et 2.3 Warm-Ups
al. 1981; Harrison et al. 1980; Pollock 1977; Saltin
& Astrand 1967).
Warming up for 15 minutes at the testing speed
When runners were investigated, testing proto-
resulted in higher heart rate, V0 2 and muscle tem-
cols were selected that use the same speed for all
peratures, and lower lactic acid levels when com-
runners or different speeds and stage increments
pared with tests that did not use a warm-up (Mar-
according to the subject's predicted V0 2 max
tin et a1. 1975). Even a brief (2- to 3-minute) warm-
(Astrand & Rodahl 1977; Costill & Fox 1969).
up of a moderate intensity « 70% V0 2 mu) has
Taylor et al. (19SS) and Gibson et a1. (I979) sug-
been shown to minimise the risks of ECG and other
gested constant testing speeds of 4.3 km/h and 6.2
cardiovascular abnormalities (Bernard et al. 1973;
km/h, respectively, because it was the slowest
Foster et al. 1982) and musculoskeletal injuries (de
speeds at which all subjects had to maintain a run-
Vries 1980). Astrand and Rodahl (1977) agreed and
ning stride. Running V0 2 max is also interrelated
suggested a warm-up of 5 to 10 minutes at SO% of
with speed of running and the state of training.
the estimated V0 2 max.
Harrison et al. (1980) agreed that a constant speed
of 10 km/h was an optimal speed for those unac-
customed to running and 12 km/h was more suit- 2.4 Recommendations for Selection
able for experienced runners. Pollock (1977) sug- of Protocol
gested that a speed be selected that results in a
testing time of 7 to 10 minutes and felt that IO:S Valid and reliable V0 2 max values are not af-
to 12 km/h for untrained and 16.1 to 19.4 km/h fected by minor adaptations in protocols. This al-
for trained runners were appropriate. Davies et al. lows the investigator some latitude for selecting a
(1984) stated that it is essential for laboratories to protocol that meets the objectives of the investi-
simulate as closely as is practically possible the gation and ensures adequate data acquisition. The
training and competitive environments of the ath- following recommendations summarise the above
letes during the test. Clark and McConnell (1986) discussion:
attempted to duplicate the training environment I. A speed should be selected that is comfort-
by comparing protocols that allowed the runners able and requires all subjects to run.
to run at their training pace and at manually self- 2. Speed should be held constant with gradual
selected speeds to protocols that used the same increments in elevation.
speed for all runners. Although there was no sig- 3. Total testing time should be from 7 to 10
nificant difference in the V0 2 max values obtained, minutes.
the runners had lower heart ~tes and lower rating 4. Continuous protocols result in equivalent
of perceived exertion values at every level of V0 2 V02 mu values when compared to discontinuous
when their training pace (P3) was used. protocols and result in shorter testing times.
The consensus of the literature recommends a S. Protocols that employ short work stages of I
constant running speed throughout the test with minute, 15 seconds or continuously incremented
increments in elevation. Speeds should be selected intensities have been shown to elicit reliable
according to the runners' level of training with faster V02 max values, are time efficient and may be more
speeds (approximating training pace) being more desirable to the runner.
appropriate for well trained runners. Also, jf the 6. The test should be preceded by a warm-up
speed selected is too slow, the runner may obtain at a relative intensity of at least 50% of VOl max.
V0 2 rna. Testing in Runners 63
1955) with the day-to-day variability ranging from may limit applicability. Users must also ensure the
4 to 6% (Boileau et al. 1977). 10% ofthis variability accuracy and correctness of the computational and
can be attributed to technical difficulties (Shephard correction equations within the program by check-
1984) and at least a 10% day-to-day variation in ing a list of software program steps.
the level of physical conditioning of subjects Under carefully controlled and standardised
(Wright et al. 1978). conditions. the variation in V0 2 max should reflect
Investigators must address the following con- the biological error. Current state of the art rapid
siderations to help minimise the variability of responding gas analysers have capabilities of ±
V0 2 max· 0.0 I % of full scale when properly calibrated. An-
I. Since test termination points are subjectively alyser specifications should be obtained to ensure
determined by the runner, the reasons for test ter- that they meet the needs of the laboratory, espe-
mination must be consistent for all tests. The in- cially if a rapid response time is desired. The in-
vestigator should use consistent motivational tech- vestigator must also ensure the accuracy of their
niques for encouraging the runners to continue as calibration gases by using manual gas analysis or
long as possible. Suggestions include intensive ver- crosscalibration with an already verified gas con-
bal motivation and informing the subject of the centration. Calibration gas concentration should be
amount of time left to complete a stage or allow known to within ± 0.03% (Weber & Janicki 1986).
for data acquisition. The automated metabolic systems also neces-
2. It is advisable to perform tests under con- sitate the integration of a volume flow device that
sistent conditions with regards to time of day. pos- has the capability of continuously reporting accu-
tabsorptive state. ambient conditions. and length rate flow volumes ranging from the resting state to
of time since previous exercise. peak exercise (5 to 150 L/min). The flow-volume
3. Problems with gas analysis. systematic errors meter selected cannot be sensitive to the phasic na-
and technical personnel must also be given con- ture of respiration. turbulent flow at higher vol-
stant attention. Staff must be given frequent in- umes. moisture content, changes in flow rate,
service training to ensure consistent procedures. changes in temperature. and must offer minimal
Calibration gases must be accurate. The gas ana- resistance, linearity. and the desired frequency re-
Iysers must be calibrated frequently particularly sponse (Froelicher 1983; Wasserman et al. 1987;
immediately preceding and following each test. Weber & Janicki 1986).
Respiratory and sampling tubing must be checked Automated systems commonly use dry gas me-
for leaks daily. ters, turbine flow meters, or pneumotachs with a
The inception of automated metabolic units pressure ditTerentiating transducer. The major dif-
have eliminated some of the technical problems ference between systems is whether they directly
such as contaminated samples due to erroneously measure volume or flow. If flow is measured, it
turned valves or bags with undetected leaks. They must be integrated over time. The accuracy of ven-
also allow metabolic responses to be continually tilation is ± 4% by measuring volume over time.
monitored and updated throughout the test. This There is a 2% deviation from linearity when flow
affords the investigator the opportunity to contin- is integrated over time (Harrison et al. 1980). For
uously monitor values and may prevent the need information concerning flow-volume meters and
for retesting because of technical errors that were gas analysers see Weber and Janicki (1986) and
previously unable to be detected until the test was Wasserman et al. (1987). Recommendations for
completed and all collected samples were analysed. selecting metabolic equipment include:
On the other hand, automated systems can be ex- I. Investigators should obtain the specifications
pensi ve. faulty assumptions on the part of the user for all components of a metabolic system from the
may result in inaccuracies and software limitations manufacturer.
in terms of timing sequences and lack of flexibility 2. Peers who are using the system being con-
V02 mu Testina in Runnen 66
up on left ventricular response to sudden strenuous exercise. up on metabolic responses to strenuous exercise. Medicine and
Journal of Applied Physiology 53: 380-383. 1982 Science in Sports 7: 146-149. 1975
Foster VL. Hume JE. Dickinson AL. Chatfield SJ. Byrnes WC McArdle WD. Katch FI. Pechar GS. Comparison of continuous
The reproducibility of VOl max. ventilation. and lactate thresh- and discontinuous treadmill and bicycle tests for max V0 2•
old in elderly women. Medicine and Science in Sports and Medicine and Science in Sports 5: 156-160. 1973
Exercise 18: 425-430. 1986 McArdle WD. Magel JR. Physical work capacity and maximum
Froelicher Jr VF. Exercise testing and training: clinical applica- oxygen uptake in treadmill and bicycle exercise. Medicine and
tions. Journal of the American College of Cardiology I: 114- Science in Sports 2: 118-123. 1970
125. 1983 McArdle WD. Magel JR. Delio DJ. Toner M, Chase JM. Speci-
Froelicher Jr VF. Brammel H. Davis G. Noguera I. Stewart A. ficity of run training on VO e max and heart rate changes during
Lancaster MC A comparison of three maximal treadmill ex- running and swimming. Medicine and Science in Sports 10:
ercise protocols. Journal of Applied Physiology 36: 720-725. 16-20. 1978
1974 McConnell TR. Swett DD, Jeresaty RM. Missri JC AI-Hani AJ.
Gass GC'. Rogers S. Mitchell R. Blood lactate concentration fol- The hemodynamic and physiologic differences between exer-
lowing maximal exercise in trained subjects. British Journal of cise modalities. Journal of Sports Medicine and Physical Fit-
Sports Medicine 15: 172-176. 1981 ness 24: 238-245. 1984
Gibson TM . Harrison MH. Wellicome RM. An evaluation of a McKay GA, Banister EW. A comparison of maximum oxygen
treadmill work test. British Journal of Sports Medicine 13: 6- uptake determination by bicycle ergometry at various pedall-
II. 1979 ing frequencies and by treadmill running at various speeds.
Glassford RG. Baycroft GHY. Sedgwick AW. Comparison of European Journal of Applied Physiology 35: 191-200. 1976
maximal oxygen uptake values determined by predicted and McMiken DF. Daniels JT. Aerobic requirements and maximum
actual methods. Journal of Applied Physiology 20: 509-513. aerobic power in treadmill and track running. Medicine and
1965 Science in Sports 8: 14-17. 1976
Hammond HK. Froelicher VF. Exercise testing for cardiorespir- Miles DS. Critz JB, Knowlton RG. Cardiovascular, metabolic.
atory fitness. Sports Medicine I: 234-239. 1984 and ventilatory responses of women to equivalent cycle er-
Harrison MH. Brown GA. Cochran LA. Maximal oxygen uptake: gometer and treadmill exercise. Medicine and Science in Sports
its measurement. application. and limitations. Aviation and and Exercise 12: 14-19. 1980
Space Environmental Medicine 51 : 1123-1127. 1980 Mitchell JH . Sproule BJ , Chapman C8. The physiological mean-
Hermansen L. Saltin B. Oxygen uptake during maximal treadmill ing of the maximal oxygen intake test. Journal of Clinical In-
and bicycle exercise. Journal of Applied Physiology 26: 31-37. vestigation 37: 538-547. 1958
1969 Moreira-Da-Costa M, Russo AK. Picarro IC Silva AC Leite-De-
Hermansen L. Ekblom B. Saltin B. Cardiac output during sub- Barros-Neto JK. et al. Maximal oxygen uptake during exercise
maximal and maximal treadmill and bicycle exercise. Journal using trained or untrained muscles. Brazilian Journal of Med-
of Applied Physiology 29: 82-86. 1970 icine Biological Research 17: 197-202. 1984
Ishiko T . Merits of various standard test protocols: a comparison Morgan WP. Psychogenic factors and exercise metabolism. Med-
between I.CP.F.R .. W.H.O .. I.B.P. and other groups. In Shep- icine and Science in Sports and Exercise 17: 309-316. 1985
hard & Lavellee (Eds) Physical fitness assessment - principles. Neiderberger M, Bruce RA. Kusumi F. Disparities in ventilatory
practice and application. pp. 7-17. CC Thomas. lIIinois. 1978 and circulatory responses to bicycle and treadmill exercise.
Jacobs I, Sjodin B. Relationships of ergometric specific VOl max British Heart Journal 36: 377-382. 1974
and muscle enzymes to blood lactate during submaximal ex- Pelchar GS. McArdle WD. Katch Fl. Magel JR. DeLuca J. Spec-
ercise. British Journal of Sports Medicine 19: 77-80, 1985 ificity of cardiorespiratory adaptation to bicycle and treadmill
Jones NL. Haddon RWT. Effect of a meal on cardiopulmonary training. Journal of Applied Physiology 36: 753-756. 1974
and metabolic changes during exercise. Canadian Journal of Pollock ML. Submaximal and maximal working capacity of elite
Physiology and Pharmacology 51: 445-450. 1973 distance runners. Part I: Cardiorespiratory aspects. Annals of
Kamon E. Pandolf KB. Maximal aerobic power during laddermill the New York Academy of Sciences 301 : 310-321. 1977
climbing. uphill running. and cycling. Journal of Applied Phys- Reilly T. Brooks GA. Circadian constancy in work stress indices
iology 32: 467-473. 1972 and physiologic capacity. Ergonomics 25: 329-330. 1982a
Katch VL. Sady S. Freedson P. Biological variability in maxi- Reilly T. Brooks GA. Investigation of circadian rhythms in meta-
mum aerobic power. Medicine and Science in Sports and Ex- bolic responses to exercise. Ergonomics 25: 1093-1107. 1982b
ercise 14: 21-25 . 1982 Rowell LB. Taylor HL. Wang Y. Limitations to prediction of
Kenney WL. Hodgson JL. Variables predictive of performance maximal oxygen intake. Journal of Applied Physiology 19: 919-
in elite middle-distance runners. British Journal of Sports 927. 1964
Medicine 19: 207-209. 1985 Saltin B. Astrand P-O. Maximum oxygen uptake in athletes. Jour-
Kumagi S. Tanaba K. Matsuara y, Matsuzaka A. Hirakoba K. nal of Applied Physiology 23: 353-358. 1967
Relationship of the anaerobic threshold with the 5km. 10km. Shephard RJ. The prediction of athletic performance by labora-
and 10 mile races. European Journal of Applied Physiology tory and field tests - an overview. In Shephard & Lavellee
49: 13-23. 1982 (Eds) Physical fitness assessment - principles. practice, and ap-
Leger L. Mercier D. Gross energy cost of horizontal treadmill and plication. pp. 113-141. CC'. Thomas. lIIinois. 1978
track running. Sports Medicine 1:270-277. 1984 Shephard RJ. Tests of maximum oxygen intake - a critical re-
Londeree BR . The use oflaboratory test results with long distance view. Sports Medicine I: 99-124, 1984
runners. Sports Medicine 3: 201-213. 1986 Shephard RJ. Allen C Benade AJS. Davies C'TM. de Prampero
Magel JR. Foglia GF. McArdle WD. Gutin B. Pelchar GS. et al. PE. et al. The maximal oxygen intake - an international ref-
Specificity of swim training on maximum oxygen uptake. Jour- erence standard of cardiorespiratory fitness. Bulletin of the
nal of Applied Physiology 38: lSI-ISS. 1975 World Health Organization 38: 757-764. 1968
Maksud MG. Coutts KD. Comparison of a continuous and dis- Sjodin B. Svedenhag J. Applied physiology of marathon running.
continuous graded treadmill test for maximal oxygen uptake . Sports Medicine 2: 83-99. 1985
Medicine and Science in Sports 3: 63-65. 1971 Smodlaka VN . Treadmills \'S bicycle ergometers. Physician and
Martin 81. Robinson S. Wiegman DL. Aulick LH. Effect of warm- Sportsmedicine \0: 75-79. 1982
V02 mu Testing in Runners 68
Snell PG. Mitchell JH. The role of maximal oxyaen uptake in objective measure of cardiorespiratory performance. Journal
exercise peri"onnanc:e. Clinics in Sports Medicine S: SI~2. 1984 of Applied PhysiolOlY 8: 73-80. 19S5
Stamford BA. Maximal oxyaen uptake duri", tradmill walki", Wasserman K. Hansen JE, Sut DY, Whipp BJ. Principles of ex-
and runnina at various speeds. Journal of Applied PhysiolOlY ercise testina and interpretation, Lea " Febiaer, Philadelphia,
39: 386-389. 1975 1987
Stamford BA. Step increment versus constant load tests for de- Weber KT. Janicki JS. Cardiopulmonary exercise testing: physi-
termination of maximal oxyaen consumption. European Jour- oloaic principles and clinical applications, W.B. Saunders
nal of Applied PhysiolOlY and Occupational PhysiolOl)' 3S: 89- Company, Philadelphia. 1986
93. 1976 Whipp BJ. Davis J. Torres F. Wasserman K. A test to determine
Stamford BAI Rowland R. Moffatt RJ. Effect of severe prior ex- parameters of aerobic function durina ellercise. Journal of Ap-
ercise on assessment of maximal oxyaen uptake. Journal of plied PhysioJo&y SO: 217-221. 1981
Applied PhysioJo&y 44: SS9-S63. 1978 Williams 00. Wyndam CH. Kok R. Rehden MJE. Effect oftrain-
Stremme SB. Inlier F. Meen HD. Assessment of maximal aerobic ina on maximal oxyaen intake and on anaerobic metabolism
power in specifically trained athletes. Journal of Applied Phys- in man. Internationale Zeitschrift fUr Anaewandte Physiologie
iolOlY 42: 833-837. 1977 einschlieshich Arbeitsphysiologie 24: 18-23. 1967
Wojtczak-Jaroszowa J. Banaskiewicz Z. Physical work capacity
Strydom NB. Environmental variables affectina fitness testina. In
durina the day and at niabt. Eraonomics 17: 193-198. 1974
Shephard" Lavellee (Eds) Physical fitness assessment - prin-
Wriaht GR. Sidney KH. Shephard RJ. Variance of direct and
ciples. practice and application. pp. 94-101. C.c. Thomas.
indirect measurements of aerobic power. Journal of Sports
Illinois. 1978
Medicine and Physical Fitness 18: 33-42. 1978
SvedenhqJ. Sjodin B. Maximal and submaximal oxyaen uptakes Yamaji K. Sakamoto H. Nakaauchi M. Kitamura K. Shephard
and blood lactate levels in elite male middle and lona distance RJ. BiolOlical rhythms of PWCI70 and maximal oxyaen in-
runners. International Journal of Sports Medicine S: 2SS-261. take. Journal of Human ErgolOlY 10: 213-219. 1981
1984
Taylor CI. Some properties of maximal and submaximal exercise
with reference to physiolOlY variation and the measurement
of exercise tolerance. American Journal of PhysiolOlY 142: 200- Author's address: Dr Timothy R. McConn~II, Director. Cardiac
212. 1944 Rehabilitation. Department of CardiolOlY, Geisinaer Medical
Taylor HL. Buskirk E, Henschel A. Maximal oxyaen intake as an Center, Danville, PA 17822 (USA).