0002 AC POWER FAILURE

Detection Always
Control section & Monitoring section
Detectin Power failure interruption has occurred.
g
condition
Operation Lamp Sound BP stop HP stop Clamp close Bypass UF stop Heater OFF Hyd. line stop
Red blinks Beep ☆ ― ☆ ○ ○ ○ ○
Reset Automatically resets and restarts operation after power failure signal turns OFF.
condition
Note Waits recovery from failure under power failure condition.
☆-mark shows that enable signal is output even when other alarm is being output
Refer to Information 2000.

Possible cause Analyzing/Measures


Power failure has occurred. Measures: Check that power supply source is stable.

Power circuit breaker is turned off. Measures: Check that power supply source is stable.

Plug is disconnected from receptacle. Measures: Reinsert plug into receptacle.

Wire breakage of power harness/Malfunction of Measures: Replace power unit.


power unit.