Professional Documents
Culture Documents
Intrinsically safe
Externally powered
pulse input
panel mounting
rate totaliser
issue 14
Certification
information
label
Total
display Rate
display
Issue: 14
25th March 2014
2
CONTENTS
1. DESCRIPTION
The BA338C is an intrinsically safe, externally
powered pulse input rate totaliser primarily
intended for use with flowmeters. The instrument
simultaneously displays the rate of flow and the
total flow in engineering units on two separate
displays.
Total Shows the total flow on the The input safety parameters of terminals 1 and 2
display larger eight digit display. Set are:
to zero when local or remote Ui = 28V dc
reset switch is operated. Ii = 100mA dc
Pi = 0.7W
Flow This disc in the lower left
indicator hand corner of the display Any certified Zener barrier or galvanic isolator with
'rotates' output safety parameters within these limits may
for two seconds each time an be used. The two system certificates provide
input pulse is received. guidance to help selection of suitable devices.
Therefore when the pulse
input frequency exceeds The maximum equivalent capacitance and
0.5Hz the disc appears to inductance between terminals 1 and 2 is:
rotate continuously
Ci = 20nF
Hold Activated when input Li = 20µH
Annunciator frequency is below the clip-off
threshold. To determine the maximum permissible cable
parameters the above figures should be subtracted
Reset Activated while instrument is from the maximum permitted cable parameters
Annunciator being reset. specified for the Zener barrier or galvanic isolator
powering the BA338C.
5
The device can withstand a 500V rms In all cases the device and circuit connected to
insulation test to earth. terminals 5 & 6 must be capable of withstanding a
500V rms insulation test to earth for one minute.
The BA338C EC-Type Examination Certificate The maximum input safety parameters for
specifies that the maximum equivalent capacitance terminals 5 & 6 are:
and inductance between the two pulse input Ui = 28V dc
terminals 3 and 4 is: Ii = 100mA dc
Pi = 0.7W
Ci = 20nF
Li = 20H and the internal capacitance and inductance is:
Interface Certificate No
Zener barriers Ex01E2005
operating frequency and the input terminal pulse output. The BA338C has two programme
numbers. For reliable counting the input signal selectable voltage switching thresholds, ‘COIL’ for
must fall below the lower threshold and rise above magnetic pick-offs and ‘UOLtS’ for higher voltage
the upper threshold. pulses.
4.2.4 2-wire proximity detector input Fig 3 Pulse source in safe area
Any certified intrinsically safe 2-wire proximity
detector complying with NAMUR switching
thresholds may be used, providing the input safety
parameters are greater than the output safety 4.2.6 Remote reset
parameters of terminals 5 & 6. The system The BA338C total display is reset to zero when
certificates list some of the acceptable devices. terminals 7 & 8 are connected together for more
than one second. Permanent interconnection
When programmed to operate with a proximity inhibits totalisation. Remote resetting may be
detector, the BA338C maximum input frequency is accomplished by any mechanically operated
5kHz. switch located in the hazardous area providing it
can withstand a 500V rms insulation test to earth.
No Zener barrier is required.
4.2.5 Voltage pulse input
Voltage pulse sources should be connected to The BA338C may be reset to zero from the safe
terminals 3 and 4. These terminals comply with area. Any switch may be used but a Zener
the requirements for simple apparatus and may be barrier is required to transfer the contact closure
directly connected to any certified intrinsically safe into the hazardous area. Any one of the diode
voltage source within the hazardous area which return barriers specified on the system certificate
can withstand a 500V rms insulation test to earth can be used for this application, it may be
for one minute and has output parameters equal to combined with the supply barrier so that only one
or less than: package is required. Fig 2 illustrates how the
BA338C may be reset from both the safe and the
Uo = 28V dc hazardous area.
Io = 100mA dc
Po = 0.7W Note: The BA338C can be programmed to reset
the total display to zero when the Up and Down
push-buttons are operated simultaneously for more
This enables the BA338C to be connected directly than two seconds – see 6.12.
to most flowmeters incorporating a certified
intrinsically safe magnetic pick-off, or a certified
intrinsically safe amplifier producing a high level
9
The following table shows the switching thresholds This enables the BA338C to be connected directly
for the various transducers, plus the maximum to most flowmeters incorporating a certified
operating frequency and the input terminal intrinsically safe magnetic pick-off, or a certified
numbers. For reliable counting the input signal intrinsically safe amplifier producing a high level
must fall below the lower threshold and rise above pulse output. The BA338C has two programme
the upper threshold. selectable voltage switching thresholds, ‘COIL’ for
magnetic pick-offs and ‘UOLtS’ for higher voltage
Switching Freq Input pulses.
thresholds max terminals
Switch 100 1000 100Hz 5&6
Proximity 1.2mA 2.1mA 5 kHz 5&6
Maximum voltage pulse counting frequency is
detector 5kHz.
Open 2k 10k 5kHz 5&6
collector
Magnetic 0mV 20mV 5kHz 3&4 4.3.6 Remote reset
pick-off peak
Voltage 1.0V 3.0V 5kHz 3&4 The BA338C total display is reset to zero when
pulse terminals 7 & 8 are connected together for more
than one second. Permanent interconnection
inhibits totalisation. Remote resetting may be
accomplished by any mechanically operated
4.3.3 Switch contact input switch located in the hazardous area providing it
Any mechanically activated switch contact located can withstand a 500Vrms insulation test to earth
in the hazardous area may be directly connected for one minute. No galvanic isolator is required.
to terminals 5 & 6 providing it can withstand a
500V rms insulation test to earth for one minute. The BA338C may be reset to zero from the safe
This includes most magnetically activated reed area. Any switch may be used but a galvanic
relays used in turbine flowmeters. The BA338C isolator is required to transfer the contact closure
contains filtering to prevent contact bounce being into the hazardous area. Fig 4 illustrates how the
counted which limits the maximum operating BA338C may be reset from both the safe and the
frequency for a switch contact to 100Hz. hazardous area.
Uo = 28V dc
Io = 100mA dc
Po = 0.7W
11
5.3 EMC
The BA338C complies with the requirements of the
European EMC Directive 2004/108/EEC. For
specified immunity all wiring should be in screened
twisted pairs.
All new BA338C rate totalisers are supplied Display Description of function
calibrated as requested at the time of ordering. If
calibration information is not supplied, the 'SCALE-r' Rate Scale Factor
instrument will be conditioned for an open collector Defines the arithmetic relationship
input, with SCALE-r and SCALE-t set to 1 and a between the pulse input frequency
timebase of seconds. and the rate display. May be adjusted
between 0.001 and 99999999.
When used with a flowmeter, SCALE-
6.3 Summary of programmable functions r should be set to the flowmeter K
This section summarises all the programmable factor (pulses per unit of
functions. When read in conjunction with Fig 9 it measurement) if the display is
provides a quick aid for programming the required in the same flow units as the
instrument. If more detail is required, each section K factor.
contains a reference to a full description of the See section 6.8
function.
't-bASE' Timebase
Selectable multiplier to display flow
SUMMARY rate in units per second, per minute or
per hour..
Display Description of function Select:
tb-1 for flow / second
‘UpdAtE’ Display update time tb-60 for flow / minute
Allows the interval between display tb-3600 for flow / hour
updates to be selected. See section 6.9
See section 6.4
6.4 Display update interval: ‘UpdAtE’ 6.6 Position of the decimal points: ‘d.P.’
Six different intervals between display updates The rate display can have up to six digits and the
varying between 0.5 and 5 seconds may be decimal point may be positioned between any of
selected. If either the rate or the total displays are them, or omitted. Similarly, the total display has
likely to change rapidly, a longer interval between eight digits and the decimal point may be
updates may simplify reading the instrument positioned between any of them, or may be
display. The selected update interval does not omitted.
affect the update time of any other instrument
function such as the optional 4/20mA output. To adjust the position of either decimal point select
'd.P.' from the main menu and press P. This will
To define the update interval select ‘UpdAtE’ from activate both displays with one digit plus the
the main menu and press P to reveal the current following decimal point of the total display flashing.
time. Pressing the Up or Down button will scroll If only the least significant digit is flashing, this
through the six times. When the required interval indicates that the decimal point is omitted.
has been selected press E to return to the main
menu. The position of the decimal point can be moved by
pressing the Up or Down push-button. Operating
the P push-button will toggle control between the
6.5 Type of input: ‘InPut’ two displays. When both decimal points have
The BA338C may be programmed to accept pulse been correctly positioned press E to return to the
inputs from a wide variety of sensors. To define main menu.
an input type select ‘InPut’ from the menu and
press P which will reveal the current input type. If the application requires flow rate to be the
Pressing Up or Down will scroll through the five primary display i.e. shown on the large display, the
options: d.P. function enables the rate and total displays to
be interchanged. The rate display will continue to
have six digits but the total display will be reduced
from eight to six digits.
17
The rate and total annunciators will also be To check or change the timebase, select 't-
interchanged. Pressing the Up and Down buttons bASE' from the main menu and press P which will
simultaneously will interchange the displays. reveal the current setting. Pressing the Up or
Down button will index through the three options:
Note: Both decimal points must be repositioned tb-1 for flow / second
after the displays have been interchanged. tb-60 for flow / minute
tb-3600 for flow / hour
6.7 Lower display: ‘dISPLAY.2’ Select the required multiplier and press E to return
This function turns the lower display on or off. to the main menu.
When turned off, the BA338C will only have one
eight digit display which may be programmed to
show total flow or rate of flow. 6.10 Rate filter: ‘FILtEr’
This digital filter has two independent adjustable
To check the status of the lower display, select parameters enabling the rate display frequency
‘dISPLAY.2’ from the menu and press P which will response to be tailored for optimum performance.
reveal if the function is ‘On’ or ‘OFF’. The setting
can be changed by pressing the Up or Down The filter parameters are controlled by a two digit
button followed by the E button to return to the number. The first digit defines the amount of
main menu. filtering applied to the display as shown below.
During commissioning it is recommend that initially When the required total scale factor has been
the second digit is set to 0 (off) and the first digit is entered, press E to store the number and return to
adjusted to provide acceptable rate display the main menu.
stability. The second digit should then be
increased until the selected step size is greater
than the noise on the display signal, at which 6.12 Clip-off: ‘CLIP-OFF’
setting the rate display will again become stable. To prevent totalisation of very low flow rates that
These will be the optimum filter parameters for over long periods may result in significant
acceptable rate display stability and a fast totalisation errors, the BA338C may be
response to a large rate signal change. programmed to stop totalising when the flow rate
falls below an adjustable threshold.
To check or change the filter select 'FILtEr' from
the main menu and press P to reveal the current To check or change the clip-off threshold select
settings. Pressing the Up or Down button will 'CLIP-OFF' from the main menu and press P which
change the flashing digit and P will transfer control will reveal the current setting. The threshold is
to the second digit. While making adjustments the shown in the units already selected for the flow
filtered rate display is shown on the lower display rate display. One digit will be flashing. The value
so that stability can be assessed. When set as of the flashing digit may be changed by pressing
required, press the E button to enter the revised the Up or Down buttons. When this digit is correct
parameters and return to the main menu. pressing P will transfer control to the next digit.
When clip-off is set as required, press the E button
to enter the revised figure and return to the main
6.11 Total scale factor: ‘SCALE-t’ programme menu.
This factor defines the arithmetic relationship
between the number of input pulses and the total If the flow rate falls below the entered threshold,
display. SCALE-t is a dividing factor that may be the rate display will show zero flow, totalisation will
adjusted between 0.001 and 99999999. See Fig stop and the HOLD annunciator will be activated.
8. The flow indicator will rotate for 2 seconds each
time an input pulse is received i.e. at input
When used with a flowmeter SCALE-t should be frequencies above 0.5Hz it will appear to rotate
set to the K-factor of the flowmeter i.e. the number continuously.
of pulses the flowmeter produces for a unit of flow.
If the BA338C total display is required in units Note: When the rate scale factor SCALE-r, the
different from those in which the flowmeter K factor timebase t-bASE, or the position of the rate display
is specified, a conversion factor will be required. decimal point are changed, clip-off will
automatically be reset to zero. If required, a new
e.g. If the flowmeter has a K factor of 45.6 pulses clip-off threshold may then be entered.
per litre and the BA338C total display is required in
units of 1,000 gallons, then SCALE-t should be set
to number of pulses produced by the flowmeter per 6.13 Local reset of total display:
1,000 gallons: ‘LOC. rSEt’
When activated this function enables the operator
45.6 x 4.54609 x 1000 to reset the BA338C total display to zero in the
display mode by operating the Up and Down push-
207,301.7 pulses per 1000 gallons buttons simultaneously for more than two seconds.
To check the status of the local reset function
. select 'LOC.rSEt' from the menu and press P
(there are 4.54609 litres in a UK gallon) which will reveal if the function is 'On' or 'OFF'. If
necessary press the Up or Down button to change
the setting, followed by the E button to return to the
To check or change the total scale factor select main menu.
'SCALE-t' from the main menu and press P to
reveal the existing setting, one digit will be The total display may also be remotely reset by
flashing. The value of the flashing digit can be connecting terminals 7 and 8 together for more
changed by pressing the Up or Down buttons. than one second. See sections 4.2.6 and 4.3.6 of
When this digit is correct pressing P will transfer this manual.
control to the next digit.
WARNING
After resetting the grand total
counter to zero the old grand total
can not be recovered.
Step 1 Enter the programming mode Finally press E to return to the 'd.P’.
Put the BA338C in the programming prompt in the main menu.
mode by simultaneously pressing P
and E. Assuming a security code has
not already been entered the Step 5 Enter the rate scaling factor
instrument will respond by displaying Select ‘SCALE-r’ from the main menu
'UPdAtE' which is the first function in and press P to show the current figure.
the main menu. See Fig 9. The K factor of the flowmeter in this
example is 1050.0 pulses per litre
which should be entered as the rate
Step 2 Select the interval between display scaling factor.
updates
With 'UPdAtE' displayed, press P to Firstly to position the decimal point,
reveal the existing interval between operate the P push-button to move the
display updates. If this is not as flashing digit to second least significant
required, press the Up or Down button position. Pressing the Up and Down
until ‘0.5’ is displayed. (0.5 seconds i.e. buttons simultaneously will then
2 display updates per second). Enter position the decimal point in front of the
the revised time and return to the least significant digit.
'UPdAtE' prompt in the main menu by
pressing E. Using the Up and Down buttons to
adjust each digit in turn and the P
button to transfer control between
Step 3 Select the type of input digits, enter 1050.0 Finally return to
Using the Up or Down button scroll the 'SCALE-r' prompt in the main menu
through the main menu until ‘InPut’ is by pressing E.
displayed, then press P to reveal the
existing setting. Select ‘COIL’, the
input for a magnetic pick-off, using the
Up or Down button and return to the
'InPut' prompt in the main menu by
pressing E.
21
Step 6 Enter the rate timebase Select ‘SCALE-t’ from the main menu
Select ‘t-bASE’ from the main menu and press P which will reveal the
and press P to show the current setting. current setting with one digit flashing.
In this example the rate display is To position the decimal point, operate
required in litres per hour. Using the Up the P push-button to move the flashing
or Down button scroll through the three digit to the least significant position.
options and select ‘tb-3600’ which will Pressing the Up and Down buttons
multiply the rate display by 3600. simultaneously will then remove the
Return to the ‘t-bASE’ prompt in the decimal point.
main menu by pressing E.
Using the Up and Down buttons to
adjust each digit in turn and the P
Step 7 Adjust the rate filter button to transfer control between
The rate display filter parameters digits, enter 1050000 Finally return to
should be adjusted experimentally after the 'SCALE-t' prompt in the main menu
installation to provide a stable rate by pressing E.
display with an acceptable step
response.
Step 9 Define clip-off
During commissioning it is recommend In this example totalisation is to be
that initially the second digit of the rate inhibited at flow rates below 300 litres
parameters is set to 0 (step response per hour.
off) and the first digit (amount of
filtering) is adjusted to provide Select ‘CLIP-OFF’ from the main menu
acceptable rate display stability. The and press P which will reveal the clip-
second digit should them be increased off threshold in litres per hour i.e. the
until acceptable rate display stability is same units already selected for the rate
once again achieved. display. The most significant digit of
the display will be flashing, indicating
To adjust the filter parameters select that this digit may be adjusted.
‘FILtEr’ from the main menu and press Pressing P five times will move the
P to reveal the current setting. The first flashing digit to the third lease
digit will be flashing and may be significant digit.
adjusted using the Up or Down button.
The P button will transfer control to the Using the Up and Down buttons set the
second digit. When both are set as third least significant digit to 3 i.e. a
required, return to the ‘FILtEr’ prompt in clip-off threshold of 300. Finally
the main menu by pressing E. return to the 'CLIP-OFF' prompt in the
main menu by pressing E.
Note: While adjusting the filter, the rate is
shown on the lower display so that .
stability can be assessed. Step 10 Turn local reset on
In this example the operator needs to
reset the total display to zero by
Step 8 Enter the total scaling factor pressing the Up and Down buttons
The K factor of the flowmeter in this simultaneously. To activate the local
example is 1050.0 pulses per litre and reset function select ‘LOC.rSEt’ from
the total display is required in cubic the main menu and press P to display
metres. the current setting. Pressing the Up or
Down button will toggle the display
There are 1000 litres in a cubic metre between ‘On’ and ‘OFF’. Select ‘On’
so the BA338C is required to display 1 and press E to return to the ‘LOCrSEt’
cubic metre for every 1,050 x 1,000 prompt in the main menu.
pulses received. ‘SCALE-t’ should
therefore be set to 1,050,000
22
c. Gently pull and then slide the blank scale Ron = 5 + 0.6V
card towards the display window until the Roff = greater than 180k
card is free. Mark the card with the required
legend and replace in the slot.
Note: Because of the series protection diode
d. Reassemble the instrument. some test meters may not detect a closed
alarm output
Alarm annunciators on the instrument display No system certificate has been issued for the
indicate the status of both alarms. If an alarm delay BA338C alarm outputs, as the system certificate
or silence time has been selected the annunciator for the circuit to which the alarms are connected
will flash during the delay or silence period. remains valid.
25
For simplicity Fig 13 only shows the programmable 'SIL' Alarm silence time
functions on the rate option of alarm AL1. Alarm 2 Defines the time that the alarm output
and the total options are identical except that the remains in the non-alarm condition
total alarms do not have hysteresis. following acceptance of an alarm.
See section 9.3.11
The following table summarises each of the alarm
programme functions and includes a cross 'AcSP' Access setpoint
reference to more detailed information. Again Sub-menu that enables direct access to
only the functions on alarm AL1 are listed. the alarm setpoints from the display
mode and defines a separate security
code.
See section 9.3.12
26
Note: The alarms will be automatically disabled Note: When ‘tYPE’ is changed, the alarm
after the rate or total displays have been parameters are automatically set to the default
recalibrated, or the alarm type has been changed. values and the alarm is disabled. It must therefore
Both alarms must therefore be enabled after be reprogrammed before use.
calibration changes.
27
9.3.6 Setpoint adjustment: ‘SPx1’ and To adjust the hysteresis, select 'HStr' from the
‘SPx2’ alarm menu and press P which will reveal the
The rate alarm setpoints SPr1 and SPr2 may be existing figure. Each digit can be adjusted using
positioned anywhere between 0 and 999999, and the Up and Down push-buttons, and the P button
the total alarm setpoint SPt1 and SPt2 anywhere to move to the next digit. When the required
between 00000000 and 99999999. hysteresis has been entered, press E to return to
the alarm menu.
To adjust the setpoint of alarm 1 select 'SPr1' or
‘SPt1’, depending upon how it has been e.g. A BA338C calibrated to display a flow of 0 to
programmed, from the alarm menu. Press P which 5000, with a high alarm set at 4000 and hysteresis
will reveal the existing value with one digit flashing. of 100 will perform as follows:
Each digit of the setpoint can be adjusted using the
Up and Down push-buttons, and the P button to High alarm will be activated when flow
move to the next digit. The position of the decimal equals or exceeds 4000, but will not reset
point is defined by the d.P. function which should until the flow falls below 3900.
be set first. When the required value has been
entered press E to return to the alarm menu.
9.3.10 Alarm delay: ‘dELA’
Note: ‘SPr1’ is displayed when alarm 1 has been This function enables activation of the alarm output
programmed as a rate alarm and ‘SPt1’ when to be delayed for a fixed time following the alarm
programmed as a total alarm. condition occurring. The delay can be
programmed in 1 second increments up to 3600
seconds. If a delay is not required zero should be
9.3.7 Alarm function: ‘HI.LO’ entered. To adjust the delay select 'dELA' from the
Each alarm must be conditioned as a high or low alarm menu and press P which will reveal the
alarm. To check or change the alarm function existing delay time. Each digit of the delay can be
select 'HI.LO' from the alarm menu and press P to adjusted using the Up and Down push-buttons,
reveal the current setting. The function can be and the P button to move to the next digit. When
changed by pressing the Up or Down buttons the required delay has been entered, press E to
followed by the E button to return to the alarm return to the alarm menu.
menu.
The alarm annunciator will start flashing
immediately an alarm occurs and continue for the
9.3.8 Alarm output status: ‘no.nc’ delay time after which the alarm output will be
The two alarm outputs may be open or closed in activated.
the non-alarm condition. When the BA338C power
supply is turned off or disconnected, both the
alarm outputs will open irrespective of whether 9.3.11 Alarm silence time: ‘SIL’
normally open or normally closed outputs have The alarm silence function is primarily intended for
been selected. Therefore, when designing an use in small installations where the alarm output
alarm system normally closed ‘nc’ should be directly operates an annunciator such as a
selected so that the output opens when an alarm sounder. When the alarm silence time is set to
occurs or if the power supply fails. any figure other than zero, the P push-button
becomes an alarm accept button.
To check the alarm output status select 'no.nc'
from the alarm menu and press P to reveal the After an alarm has occurred, operating the P
current setting. The function can be changed by button will cause the alarm output to revert to the
pressing the Up or Down button followed by the E non-alarm condition for the programmed alarm
button to return to the alarm menu. silence time. When an alarm is silenced the alarm
annunciator will flash until the silence time expires.
9.3.9 Hysteresis: ‘HStr’ To adjust the silence time select 'SIL' from the
Hysteresis is only available on rate alarms. During alarm menu and press P which will reveal the
programming hysteresis is shown in the units of existing time. The time may be adjusted to
rate the BA338C has been calibrated to display. between 0 and 3600 seconds in 1 second
increments. Each digit can be adjusted using the
Up and Down push-buttons, and the P button
transfers control to the next digit. When the
required time has been entered press E to return
to the alarm menu.
28
9.3.12 Access Setpoint: ‘AcSP’ flashing digit and the P push-button to move
This function activates a separate menu that control to the next digit. If the correct code is
provides direct access to the alarm setpoints when entered pressing E will then cause alarm setpoint
the instrument is in the display mode. See section prompt 'SPx1' to be displayed. If an incorrect
9.3.13 for a full description. An operator may security code is entered, or a button is not pressed
therefore adjust the alarm setpoints without having within ten seconds, the instrument will
access to the programme and alarm menus. automatically return to the display mode.
Further protection is provided by a separate
security code. Once within the menu pressing the Up or Down
buttons will toggle the display between the two
This direct access menu may be enabled and a alarm setpoint prompts 'SPx1' and 'SPx2'.
separate security code entered from the 'AcSP'
function in the main programme menu as shown in
Fig 13. To change the menu parameters select
'AcSP' from the programme menu and press P.
This will display the enable prompt 'EnbL'.
Note: Direct access to the alarm setpoints is only The display brilliance depends upon the current
available when the menu is enabled - see section flowing through the backlight. This is determined
9.3.12 by the supply voltage and the end-to-end
resistance of the Zener barrier or output resistance
of the galvanic isolator. Brilliance will not be
9.4 Display backlight significantly reduced until the current falls below
The BA338C can be supplied with LED 20mA.
backlighting to improve display contrast when the
instrument is installed in a poorly illuminated area. Backlight current = Vsupply -18
End-to-end resistance of barrier#
The backlight is segregated from the measuring
circuit and has been certified as a separate
# or output resistance of galvanic isolator
intrinsically safe circuit. System certificates for use
with Zener barriers and galvanic isolators have
therefore been issued. 9.5 Pulse output
The BA338C may be supplied with an opto-
The backlight must be powered from the safe area isolated solid state pulse output. The output is a
via a Zener barrier or a galvanic isolator as shown polarised current sink that closes for a
in Fig 15. Any certified Zener barrier may be programmable duration each time the least
used, providing the output parameters do not significant digit of the BA338C display changes. A
exceed: programmable divider enables the pulse output
frequency to be reduced by 10, 100, 1,000 or
Uo = 28V dc 10,000. See Fig 17.
Io = 159mA
Po = 0.8W Ron = 60 + 3V
Roff = 1M
Ci = 40nF Isink max = 10mA
Li = 30µH
The output pulse is non-synchronous and the
Alternatively a galvanic isolator may be used In maximum pulse frequency is determined by the
place of the barrier. For UK installations only the programmed pulse width.
galvanic isolators specified on the system
certificate should be used. Max frequency = _______1_______ Hz
2 x (pulse duration)
The pulse output will sink up to 10mA. When To determine the maximum permissible cable
closed the voltage between the pulse output parameters, these figures should be subtracted
terminals P1 & P2 will be less than 3V when from the maximum cable capacitance and
switching 5mA. inductance specified by the system certificate of
the circuit connected to the BA338C. Fig 16
shows how the pulse output circuit may be
powered from the BA338C supply to drive a
counter in the safe area via a diode return barrier.
Uo = 28V dc
Io = 100mA
Po = 0.7W
Uo = 28V dc
Io = 100mA
Po = 0.7W
Ci = 20nF
Li = 20µH Fig 18 Application of 4/20mA output
31
The maximum equivalent capacitance and The 4/20mA output may be calibrated to increase
inductance of the optional 4/20mA output is: or decrease as the rate display increases, and
most instruments will produce a linear output
Ci = 2nF between 3.5 and 22mA.
Li = 8µH
10. INDEX
Zone 0, 1 or 2
Class 1 Group IIC
Group IIB
Group IIA
34
35
36
37
38
39
40
41