You are on page 1of 1

period Task activity Materials needed Personnel

Daly  Clean external panel. User


 Remove waste matter from internal
chamber

weekly Heating element User


 Check internal heating element BME/T
connections are tight
 check the indicating lamp are
working
 check the ratio of
temperature/pressure during
sterilization
 Replace heating element if covered
with lime scale
 If plug, cable or socket are damaged,
replace

Semi  Vacuum test Grease. BME/T


Annually  Test temperature controller Thermometer.
 Lubricate handles, door lock with
grease

Annually  Check bacteria filter and replace Bacteria filter. BME/T


 Check electrical safety Electrical safety
 Replace door gasket sealing analyzer.
 Check, clean and lubricate or replace Gasket.
if necessary
 electromagnetic valve
 safety valve
 drainage valve
PPM for Autoclaves/sterilizers

You might also like