Period Task Activity Materials need personnel

Daily Damp close User


 Remove any dust
 Fill humidifier bottle up to marker
with clean water
 Check all screws, connectors, tubes
and parts tightly fitted
 Check oxygen flow before clinically
use
weekly User
 Wash the filter in warm water and
dry if damaged replace
 Clean humidifier bottle
 Replace humidifier bottle if covered
with lime scale
 If mains plug, cable or socket are
damaged, replace
 Run machine for two minutes and
check no alarms occur
 Check that flow rate varies with flow
control
Semi  Check oxygen concentration Oxygen analyzer BME/T
annually
Annuall  Clean/ replace cabinet filters. HEPA filter BME/T
y  Check outlet HEPA filter. Check Cabinet filter
compressor inlet Filter. Oxygen analyzer
 Check O2% and power loss alarm. Compressor inlet
filter
PPM for oxygen concentrator