Field Flash

DAN-FF-0038
Date: 12/22/2005 Distribution: Public
Author: Tony Rossetti
System: Selenia Subsystem: AWS and Array
Subject: AWS v3.1.0.X / Array Vx Main compatibility

Problem:
Replacement arrays may be incompatible with system that is operating with
v3.1.0.X software.

Reason:
Incompatible Vx main /Array Software and/or sequence files

Solution:
Systems running 3.1.0.x

For the array to be compatible with systems currently running 3.1.0.X, the array will need
to be equipped with the following:

1. VxMain / Array Software Version 1.2.13.3

2. Array Readout Sequence versions EH5. The version can be identified by the prefix of
the sequence name, there will be an EH5 prefix on the name. There are various flavors
of EH5 and technical support should be consulted to obtain the correct sequence.

CSD-0044-F02, Rev. 001 Page 1 of 1


Hologic Inc. 35 Crosby Drive, Bedford, MA 01730-1401
781-999-7704