This document provides artwork and signature files for release notes describing version 3.1.3.0 of the Selenia medical imaging system. The single page of artwork contains release notes describing fixes for issues that could cause images to appear out of order or skewed. The release also fixes issues with exposure index calculation, USB drive access, backup functionality, and communication errors. References to IONA Technologies products were replaced with OmniOrb without changing the user interface.
This document provides artwork and signature files for release notes describing version 3.1.3.0 of the Selenia medical imaging system. The single page of artwork contains release notes describing fixes for issues that could cause images to appear out of order or skewed. The release also fixes issues with exposure index calculation, USB drive access, backup functionality, and communication errors. References to IONA Technologies products were replaced with OmniOrb without changing the user interface.
This document provides artwork and signature files for release notes describing version 3.1.3.0 of the Selenia medical imaging system. The single page of artwork contains release notes describing fixes for issues that could cause images to appear out of order or skewed. The release also fixes issues with exposure index calculation, USB drive access, backup functionality, and communication errors. References to IONA Technologies products were replaced with OmniOrb without changing the user interface.
A. FAND 12/16/05 REV DRAFTED BY DATE DANBURY, CT 06810 A. FAND 12/16/05 PROPRIETARY INFORMATION: TITLE DOCUMENT NUMBER REV The content of this document is the exclusive property of Lorad and may AW, REL NOTES, SELENIA 3.1.3.0 AW-01203 001 not, without prior written permission of Lorad, be reproduced, copied or used for any purpose whatsoever. ARTWORK SIZE A SHEET 1 OF 1 ENG-0034-T33, Rev. 001 Selenia 3.1.3 Release Notes 1.0 Introduction This document is provided to give you a quick overview of the new features of the Selenia Version 3.1.3. It is not meant to replace the User’s Manual.
2.0 Issues Fixed in Selenia 3.1.3
This release fixes the circumstance that can lead to images that have portions that are out of order or skewed in appearance.
3.0 Issues Fixed in Selenia 3.1.2
This release fixes the following circumstances: • The exposure index had been incorrectly calculated if a manual AEC sensor position is selected for an exposure and if the chest wall is on the right side. • Some circumstances when the USB CD-RW drive could not be accessed have been resolved. • The configuration backup functionality has been restored. • Some circumstances of intermittent Error 51 (error communicating with the HTC grid) have been resolved.
All traces of IONA Technologies products have been removed from this release and replaced with a new product called OmniOrb. This replacement makes absolutely no changes to the Selenia user interface.
MAN-00449 Revision 001 Manufactured by Hologic, Inc Page 1 of 1
36 Apple Ridge Road, Danbury, CT 06810 (203) 207-4500