Software enhancements:
• Allows the system to keep the detector on during the course of the night. Keeping the
detector on during the night will allow the detector to be “conditioned while resting.” The
system still monitors the detector temperature and will shut the detector off in the event
that the room temperature exceeds specified limits to protect the life of the detector.
• Automatically restarts the detector prior to imaging patients in the morning in the event the
system turned off the detector during the course of the night due to excessive temperature
in the room. This will now allow the detector to come up to temperature prior to imaging
patients in the morning avoiding the need for user intervention that may delay the imaging
of patients.
• Allows the system to keep the detector on during the weekend so that there is no need to
come early on Monday morning and turn the system on for early warm-up prior to imaging
patients.
• Adds more troubleshooting tools for Hologic Service use.
The contrast of the area around the skin edge has been increased to allow for better visibility of
tissue structures. This change will only affect screening views (i.e. standard CC and MLO views)
acquired after the installation of this software release. Images already stored into PACS cannot
be changed to allow for this enhancement.
All traces of IONA Technologies products have been removed from this release and replaced with
a new product called OmniOrb. This replacement makes absolutely no changes to the Selenia
user interface.