Service Note for E-CUBE 15

1.Release Information
System E-CUBE 15 S/W Version V 3.0.9.378
Release Date 2015-03-05 OS Version V1.1.8.25
Version-Up
Release Type
Patch
New Feature Software Hardware Image
Change Category
New Transducer General Notice
1. Bug fix

2. OS Change (1.1.8.25)
Summary • Applying the new touch controller due to discontinuation of the component.
(This change does not affect the product’s capacity.)
• When booting up the system, sound noise issue is fixed.

Contents

1. Bug fix: Refer to the following 2. Fixed Field Issues.

2. OS Change (1.1.8.25)
• Applying the touch controller.
- OS is changed to apply the new touch controller due to changing supplier.
(The resolution of touch screen is same as previous one. This change does not affect the product’s
capacity.)
- eGalax Controller (Old version) -> DMC Controller (New version)
• When booting up the system, sound noise issue is fixed.

2.Fixed Field Issues


No Fixed Issue

When sending the data to Dicom in No patient, the issue that study date and time information
1
were missing is fixed.

2 The image freezing issue during the scanning is fixed.

1
3. Known issue & Workaround

Category Notes/Workaround

Installation X: You should perform Factory Default.


-Backup/Restore O: Compatible
- SW upgrade Caution :

- The presets marked ‘X’ mean that you cannot use the previous user preset.
Version V2.0.0 -> V2.0.8 V2.0.8 -> V2.5.0 V2.5.0-> V3.0.0 V3.0.0-> V3.0.9
E-CUBE 15

Patient DB
-Image Data O x x O
-Measurement Data

Image Preset O x x O

Service O x x O

Measurement O x x O

Annotation O x x O

Connectivity O x x O

Others
-System
-Monitor Setting
O x x O
-User Setting
-Administration
-Advanced Cardiac

Warning: When upgrading the software V2.5 to V3.0, you must back up the following
data due to the compatibility problem: Patient Data, Image Preset, System Preset

Transducer Transducer Compatibility


Compatibility - Convex : SC1-4H, SC1-4HS, SC1-6H

- Microconvex : C5-8

- Linear: L3-8, L3-12H, L3-12X, L8-17X

- Volume Convex : SVC1-6H

- Endocavity : E3-10H, EC3-10H, EV3-10H


- Phased Array : SP3-8, SP1-5X, MP1-5X

- TEE : TEE3-7

- CW : CW2.0, CW5.0
- IO : IO3-12

2
CP800 Printer When you set the wrong printer paper size and the printing does not proceed,
1) Check the printer paper size. (There is no alert message for the wrong paper size

setting.)

2) Press the System Preset key and go to System > Peripheral, and then click the setting

icon.

3) Select CP800 Printer.

4) Select a waiting print task and select Cancel.

Caution: You must delete the waiting task in the list after you set the paper size.

MITSUBISHI How to set MITSUBISHI printer,


Printer 1) Connect a printer to the system. Copy(1) driver is added.

2) Press the System Preset key and go to System > Peripheral and click Printer and Faxes
list icon.

3) Select MITSUBISHI driver and Copy(1), and then remove the both.

(You must remove the both drivers. The printer list is not updated and the deleted drivers
remain on the list.)
4) Reboot your system.

USB HDD When the backup medium is USB HDD, do not connect it to the USB port on the front
side of the system.
1) Problem: The system makes a noise when the system recognizes the USB HDD and the

data loss for the USB HDD may occur.


2) Solution: You must connect a USB HDD to the USB port on the rear side of the system.

Image Store Save images after registering a patient ID.


1) Problem: When you register a patient ID and select End Exam after saving images with

no patient ID, the saving time takes longer.

3
ECG When ECG Display is set to ON, reboot the system after connecting or disconnecting the
ECG cable.
1) Problem: The ECG waveform is not shown or a crash occurs. (This will be fixed in the

next version)

2) Solution: Reboot the system after connecting or disconnecting the ECG cable.